InBody BPBIO210 High Blood Pressure Monitor User Manual

July 26, 2024
InBody

BPBIO210/220
User’s Manual

BPBIO210 High Blood Pressure Monitor

BPBIO210/220 User’s Manual for Measurement Guide and Setup
Thank you for purchasing the BPBIO210/220. This user’s manual describes all the features of the BPBIO210/220.
Please read before use and keep it in a safe place. By following the manual instructions, you will be able to use the
BPBIO210/220 more safely and effectively.

Please note the important information below before reading this manual.

Warning
Failure to comply with safety warnings and regulations can cause serious injury or death.
Caution
Failure to comply with safety cautions and regulations can cause injury or property damage.
Note
Referring to notes can help improve equipment use.
Not MR safe
Warnings and cautions described in the user’s manual should be observed all times.

InBody Co., Ltd. [HEAD OFFICE]
For product support or clinical questions, please contact us at the following:
625, InBody Bldg., Eonju-ro, Gangnam-gu, Seoul 06106 KOREA
TEL: +82-2-501-3939
FAX: +82-2-578-5669
Website: inbody.com
E-mail: info@inbody.com

InBody Co., Ltd.
15, Heugam-gil, Ipjang-myeon, Seobuk-gu,
Cheonan-si, Chungcheongnam-do 31025 KOREA
TEL: +82-41-581-3003 FAX: +82-41-581-3103
Website: inbody.com E-mail: info@inbody.com
©2024 InBody Co., Ltd. All rights reserved.

Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd. under the copyright laws. This manual may be subject to change without notice. InBody Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred by not complying with the content of the User᾽s Manual.
Visit our website (inbody.com) to view and download additional information about the BPBIO220/210. InBody
Co., Ltd. reserves the right to modify the appearance, specifications, etc. of this product to improve its quality, without prior notice.

Representative & Sponsor Information
InBody Europe[EUROPE]
InBody Europe B.V.
Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands
TEL: +31-20-238-6080 FAX: +31-6-5734-1858 Website: nl.inbody.com E-mail: info.eu@inbody.com

Australian Sponsor. [AUSTRALIA]
Emergo AUSTRALIA.
Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA

USA CONTACT
BIOSPACE INC. dba INBODY 13850 CERRITOS CORPRT DR STE C, Cerritos, California, 90703
TEL: 323-932-6503

Customer Service Information
InBody USA [USA] 13850 Cerritos Corporate Dr. Unit C Cerritos, CA 90703 USA
TEL: +1-323-932-6503 FAX: +1-323-952-5009 Website: inbody.com E-mail: info.us@inbody.com

InBody Japan [JAPAN]
InBody Japan Inc.
Tani Bldg., 1-28-6, Kameido, Koto-ku, Tokyo 136-0071 Japan
TEL: +81-3-5875-5780 FAX: +81-3-5875-5781 Website: www.inbody.co.jp E-mail: inbody@inbody.co.jp

InBody China [CHINA]
903/904, XingDiPlaza, No.1698 YiShanRoad, Shanghai 201103 China
TEL: +86-21-64439705 FAX: +86-21-64439706 Website: inbodychina.com E-mail: info@inbodychina.com

InBody India [INDIA]
57/57/ A, 1st Floor, Raj Industrial Complex, Military Road, Marol, Andheri (East). Mumbai-400059,
Maharashtra, India
TEL: +91 22-6223-1911 Website: inbody.in E-mail: india@inbody.com

InBody Asia [ASIA]
InBody Asia Sdn. Bhd.
Unit 3A-11, Oval Damansara, 685 Jalan Damansara Kuala Lumpur, WP KL 60000 Malaysia
TEL: +60-3-7732-0790 FAX: +60-3-7733-0790 Website: inbodyasia.com E-mail: info@inbodyasia.com

Intended Use (Indications for Use):
The BPBIO210/220 is a digital device intended for use in measuring blood pressure and pulse rate in user population with upper arm circumference ranging from 17cm to 42cm (7-inch to 17-inch).
This device should be used by medical professionals. The systolic blood pressure and diastolic blood pressure shall be measured by the auscultatory method. Pulse rate will be shown on the LCD screen after the examination is finished. This device may provide useful clinical information about the current health status of not only the users who are diagnosed with hypertension but also those who are not diagnosed with hypertension. Warnings and cautions described in the user’s manual should be observed at all times.

  • This device is designed for use in medical facilities only, such as hospitals, clinics, and physicians’ offices. This device is not intended for use at home by non-medical professionals.

BPBIO210/220
User’s Manual Contents

Introduction for the BPBIO210

This blood pressure monitor can be used by a medical professional to check blood pressure with auscultatory method by using the pressure indicator bar on the LCD screen. The back-light function allows blood pressure monitoring even under dark conditions.

A. Standard Components
❶ Main unit (1 EA)
❷ M-size cuff (1 EA) – for the use of 22 to 32 cm (9 to 13 in) in arm circumference
❸ Air bulb (1 EA)
❹ User’s manual (1 EA)
❺ “AA” alkaline batteries 1.5V (4 EA)

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
1

* Optional Components (Sold Separately)
❻ Cart
❼ AC Adapter
– Input Power: AC 100-240V, 50/60Hz, 0.32-0.19A MAX
– Output Power: DC 6.2V, 2.0A
❽ S, L size cuffs (S-size: Suitable for arm circumference between 17 to 22cm (7 to 9 in),
L-size: for the use of 32 to 42cm (13 to 17 in) in arm circumference.

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
2

* Main Unit Parts

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
3

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
4

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
5

* Cuff Parts

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO210
6

* Air Bulb Parts

* Display Screen Parts

B. Installation Environment
Please make sure the environment is appropriate for the BPBIO210 installation.
This equipment is designed for indoor use. If installing the equipment outdoors, the following requirements must be met.

Temperature range 10 to 40ºC (50 to 104ºF)
Relative humidity 30 to 85% RH
Atmospheric pressure range 70 to 106kPa

C. Installation Instructions

  1. Unboxing the BPBIO210
    After opening the BPBIO210 packaging box, take out the BPBIO210 from the buffer and the protective vinyl package. Open the BPBIO210 and take out the cuff and the air bulb from the protective vinyl package.
    InBody BPBIO210 High Blood Pressure Monitor - Installation Instructions
1 Caution
    • Do not inflict excessive force when opening the BPBIO210.
    • The main unit cannot be bent over a certain angle. Opening the BPBIO210 with excessive force may damage the device.

  2. Connecting cuff and air bulb
    1. Plug the cuff tube into the cuff jack of the main unit. ( ).
    2. Plug the air bulb tube into the air bulb jack of the main unit. ( ).
    3. Push the tubes all the way in to the end.
    InBody BPBIO210 High Blood Pressure Monitor - Installation Instructions
2

  3. Inserting / replacing batteries
    1. Remove the battery cover.
    2. Insert four (4) “AA” batteries in the right direction as indicated below.
    3. Close the battery cover.
    InBody BPBIO210 High Blood Pressure Monitor - Installation Instructions
3 Warning
    • Use the “AA” size alkaline or manganese batteries.
    • Do not mix new and used batteries. Replace four batteries with all of the same brand and the same model new batteries.
    • Do not insert batteries with wet hands.
    • Make sure that the polarities (+) and (-) are correct.
    • Remove the batteries from the battery compartment when it will not be in use for long period.
    • Dispose used batteries only in designated locations.
    • Improper use of batteries may cause degradation of the performance of BPBIO210 and may cause fire.

  4. Use of the AC adapter (Optional)

  • Input Power: AC 100–240V, 50/60Hz, 0.32-0.19A
    Output Power: DC 6.2V 2.0A
    ❶ Insert the AC adapter plug into the AC adapter jack on the side of the main unit as shown in the image below.
    ❷ Plug in the AC adapter to an electrical outlet.

< Plug Type F>

InBody BPBIO210 High Blood Pressure Monitor - Installation Instructions
4

< Plug Type B>

InBody BPBIO210 High Blood Pressure Monitor - Installation Instructions
5

Warning

  • Do not connect or disconnect the AC adapter with wet hands.
  • Do not put heavy objects on the AC adapter cable.
  • Do not use the AC adapter if it is damaged.

Caution

  • Plug the AC adapter into the electrical outlet securely.
  • When unplugging the AC adapter, grasp the body of the adapter and pull out.
  • To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter from the main unit.
  • Do not apply excessive force to the AC adapter body and the plug.
  • Keep the adapter away from foreign objects.
  • Use only the AC adapter provided by InBody.
  • Do not use the AC adapter for any other purposes other than for the operation of BPBIO210.

Note

  • The AC adapter is sold separately.

D. Features and Functions

  1. Automatic power off
    • The BPBIO210 will automatically shut down after 5 minutes has elapsed without pressing any buttons.
    • The BPBIO210 will not shut down if it is connected to the AC adapter.
    Main unit will switch off automatically if idle for more than 5 minutes to prevent battery drain. Main unit will not switch off automatically if connected to the AC adapter.
    The automatic power off function can be turned off by following the settings below.
    ❶ Press the button to turn on the power.
    ❷ When the image below is displayed on the LCD, the power is on.
    ❸ Press and hold the button for about 5 seconds. If ‘OFF’ is displayed on the screen as shown below, the automatic power off function is disabled.
    ❹ You can turn the automatic power off function back on by repeating the steps above.

  2. Protection from excessive pressurization
    • If the pressure applied to the cuff exceeds 320mmHg, the BPBIO210 will automatically exhaust the pressure from the cuff to protect the patient.

  3. Residual air detection
    • If there is any remaining air left in the cuff, the error code ‘E1’ will appear on the screen when the product power is switched on.
    • For accurate measurement result, exhaust the air in the cuff entirely before switching the power on.

  4. Pulse measurement
    • Pulse is detected when the pressure applied to the cuff is greater than 50mmHg. The pulse rate is measured only when the pulse is detected at least 5 times.
    • If deflation is performed too quickly, pulse may not be detected. Deflate at the rate of 2 to 4mmHg/sec. to detect the pulse.
    • The measured heart rate will be displayed when the pressure is less than 20mmHg.
    • If pulse is not detected or is not constant, “- – -” will be displayed.

  5. Back-light
    • The back-light is automatically turned on or off depending on the brightness of the surrounding environment.
    • The back-light function can also be manually operated by pressing the back- light button.
    • If the back-light is operated manually with the back-light button, the back- light will not turn on or off automatically. When the product is turned off and on, the back-light automatically turns on or off again depending on the brightness of the surrounding environment.

  6. Mark
    • Mark function allows to display the measured value on the screen momentarily once the button is pressed.
    • Up to five marks can be displayed.

Blood Pressure Measurement

A. Precautions before Measurement
Caution

  • The BPBIO210 must be operated by those who are well trained of the auscultatory method of measuring the blood pressure.
  • Do not operate the BPBIO210 immediately if it has been left for a long period of time in an environment where the temperature is lower than 10ºC (50ºF) or higher than 40ºC (104ºF).
  • Do not use the BPBIO210 in a highly humid environment.
  • Do not use the BPBIO210 near any equipment which generates strong magnetic fields, electrostatic, or high frequency.
  • Patients with artificial heart or lungs may obtain inaccurate results.

Warning

  • Do not wrap the cuff on an injured arm. Injury may worsen or can be infected.
  • Do not wrap the cuff in an arm with the intravenous injection or IV plugged in.
  • Please consult a doctor before using the blood pressure monitor if you have any cardiovascular diseases.
  • Do not wrap the cuff around your neck.
  • Do not wrap the cuff tube around your neck.
  • Do not wrap the air bulb tube around your neck.

B. Measurement Instructions

  1. How to wear the cuff
    ❶ Wrap the cuff around the arm, about 2-3 cm (1 in) above the elbow.
    ❷ Make sure that the index (▲) on the cuff is located within the RANGE bar.
    Replace the cuff with the right size if the index deviates from the RANGE.
    ❸ Place the cuff so that the artery mark (▼) is located on the artery.
    ❹ Do not wrap the cuff too tightly. Leave a margin of a finger to get the best result.
    InBody BPBIO210 High Blood Pressure Monitor - Blood Pressure Measurement
1 Caution
    • For correct measurement, please wear the cuff properly.
    • If the cuff is too loosely wrapped, the measured blood pressure values may not be correct.
    Note
    • The cuff provided will be M-sized cuff which is suitable for individuals with an arm circumference of 22 to 32cm (9 to 13 in). If the arm circumference exceeds 32cm (13 in), use the L-size cuff.
    If the arm circumference is smaller than 22cm (9 in), use the S-size cuff.
    • L-size or S-size cuff is sold separately.

  2. Proper measurement posture
    ❶ Place the arm so that the middle of the cuff is at the patient’s heart position.
    ❷ Be relaxed during the measurement.
    Caution
    • The measurement posture may affect the blood pressure values.
    • Please keep the proper measurement posture for accurate measurement.

  3. Turning the power on
    ❶ Press the button to turn on the power.
    ❷ The symbols on the screen and the back-light will be turned on and off with a beeping sound.
    ❸ When there are no changes detected in the pressure within the cuff, number “0” will be displayed on the screen, which means it’s ready for measurement. If the pressure of the cuff is not stabilized within 20 seconds, the error code ‘E1’ will be displayed in the pressure/pulse display column on the screen.

  4. Battery level indicator
    ❶ The remaining battery level can be checked in the bottom right corner of the screen while using the blood pressure monitor.
    ❷ When using the AC adapter, the battery level is not displayed, instead the AC adapter connection symbol ( ) appears.
    | : The capacity of the batteries is sufficient.
    ---|---
    | : Battery level is low.
    | : Battery level is low.

Please replace with new batteries.

  • Refer to “Inserting / Replacing Batteries” on page 12.

    1. Adjusting the back-light
      ❶ The back-light is automatically turned on or off depending on the brightness of the surrounding environment.
      ❷ The back-light function can also be manually operated by pressing the back- light button.

    2. Pressurization
      ❶ Rotate the exhaust valve clockwise to close the valve completely.
      ❷ Squeeze the air bulb repeatedly to pressurize.
      ❸ The current pressure value is immediately displayed in the pressure bar on the screen in units of 1 step / 2 mmHg, and is displayed as a number in the Pressure/pulse display column on the screen.
      ❹ Stop pressurization when it reaches the targeted inflation pressure.

  • The maximum pressurization can be up to 320mmHg, and if pressurized at a higher rate, it will depressurize automatically and ‘—’ will be displayed in the Pressure / Pulse Rate Display on the screen. 7. Depressurization
    ❶ Rotate the exhaust valve counterclockwise to open the valve. Continue the blood pressure measurement at a proper exhaust rate.
    ❷ If the button on the main unit is pressed, a bar symbol will be displayed on the pressure display bar.

  • Marking is displayed up to 5 times cumulatively, and if you mark more than 5 times, it will be erased sequentially from the first mark in the past.
    InBody BPBIO210 High Blood Pressure Monitor - Blood Pressure Measurement
6

    1. Completing the measurement
      ❶ When the blood pressure measurement is completed, open the exhaust valve and completely exhaust the air in the cuff.
      ❷ When the pressure drops below 20mmHg, the pulse rate will be displayed in the pressure/pulse display on the screen.
      The pulse rate is measured when the pressure is reduced at a rate of 2 to 4 mmHg/sec.
      If the pulse is not detected or unstable, ‘- – -’ is displayed on the screen.
  • Please review ‘B. Regarding the equipment – Pulse is not being measured’ on page 39. 9. Turning the power off
    ❶ Press the button to switch off the power.

C. Storage

  1. Wipe any stains on the main unit with a smooth cloth.
  2. Place the cuff and air bulb as shown below.
    The BPBIO210 should be transported or stored under the following conditions.Temperature range| -20 to 60ºC (-4 to 140ºF)
    ---|---
    Relative humidity| 10 to 85% RH (No Condensation)
    Atmospheric pressure range| 50 to 106 kPa

Caution
• Use UV sterilizer to disinfect the cuff.
Note
• If the cuff is contaminated and can no longer be used, please replace with a new cuff.
• Contact your local InBody representative to purchase the cuff.

BPBIO220

Introduction for the BPBIO220

This blood pressure monitor can be used by a medical professional to check blood pressure with auscultatory method by using the pressure indicator bar on the LCD screen. The back light function allows blood pressure monitoring even under dark conditions.
The user may control the speed of the exhaust by using the controller and mark the systolic and the diastolic blood pressure values on the pressure-bar display.

A. Standard Components
❶ Main unit (1 EA)
❷ M-size cuff (1 EA) – for the use of 22 to 32 cm (9 to 13 in) in arm circumference
❸ Controller (1 EA)
❹ User’s manual (1 EA)
❺ “AA” alkaline batteries 1.5V (4 EA)

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
1

* Optional Components (Sold Separately)
❻ Cart
❼ AC Adapter
– Input Power: AC 100-240V, 50/60Hz, 0.32-0.19A MAX
– Output Power: DC 6.2V, 2.0A
❽ S, L size cuffs (S-size: Suitable for arm circumference between 17 to 22 cm (7 to 9 in),
L-size: for the use of 32 to 42 cm (13 to 17 in) in arm circumference.

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
2

* Main Unit Parts

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
3

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
4

* Cuff Parts

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
5

* Controller Parts

* Display Screen Parts

B. Installation Environment
Please make sure the environment is appropriate for the BPBIO220 installation.
This equipment is designed for indoor use. If installing the equipment outdoors, the following requirements must be met.

Temperature range 10 to 40ºC (50 to 104ºF)
Relative humidity 30 to 85% RH
Atmospheric pressure range 70 to 106kPa

C. Installation Instructions

  1. Unboxing the BPBIO220
    After opening the BPBIO220 packaging box, take out the main unit from the buffer and the protective vinyl package. Open the BPBIO220 and take out the cuff and the air bulb from the protective vinyl package.
    InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
8 Caution
    • Do not inflict excessive force when opening the BPBIO220.
    • The main unit cannot be bent over a certain angle. Opening the BPBIO220 with excessive force may damage the device.

  2. Connecting cuff and air bulb
    1. Plug the cuff tube into the cuff tube jack of the main unit ( ).
    2. Plug the controller 4-pole plug and the controller tube into the corresponding jack ( ) of the main unit.
    3. Push the controller 4-pole plug and the controller tube all the way in to the end.
    InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
9

  3. Inserting / replacing batteries
    1. Remove the battery cover.
    2. Insert four (4) “AA” batteries in the right direction as indicated below.
    3. Close the battery cover.
    InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
10 Warning
    • Use the “AA” size alkaline or manganese batteries.
    • Do not mix new and used batteries. Replace four batteries with all of the same brand and the same model new batteries.
    • Do not insert batteries with wet hands.
    • Make sure that the polarities (+) and (-) are correct.
    • Remove the batteries from the battery compartment when it will not be in use for long period.
    • Dispose used batteries only in designated locations.
    • Improper use of batteries may cause degradation of the performance of BPBIO220 and may cause fire.

  4. Use of the AC adapter (Optional)

  • Input Power: AC 100-240V, 50/60Hz, 0.32-0.19A
    Output Power: DC 6.2V, 2.0A
    ❶ Insert the AC adapter plug into the AC adapter jack on the side of the main unit as shown in the image below.
    ❷ Plug in the AC adapter to an electrical outlet.

< Plug Type F>

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
11

< Plug Type B>

InBody BPBIO210 High Blood Pressure Monitor - Introduction for the BPBIO220
12

Warning

  • Do not connect or disconnect the AC adapter with wet hands.
  • Do not put heavy objects on the AC adapter cable.
  • Do not use the AC adapter if it is damaged.

Caution

  • Plug the AC adapter into the electrical outlet securely.
  • When unplugging the AC adapter, grasp the body of the adapter and pull out.
  • To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter from the main unit.
  • Do not apply excessive force to the AC adapter and the plug.
  • Keep the AC adapter from foreign objects.
  • Use only the AC adapter provided by InBody.
  • Do not use the AC adapter for any other purposes other than for the operation of BPBIO220.

Note

  • The AC adapter is sold separately.

D. Features and Functions

  1. Automatic power off
    • The BPBIO220 will automatically shut down after 5 minutes has elapsed without pressing any buttons.
    • The BPBIO220 will not shut down if it is connected to the AC adapter.
    Main unit will switch off automatically if idle for more than 5 minutes to prevent battery drain. Main unit will not switch off automatically if connected to the AC adapter.
    The automatic power off function can be turned off by following the settings below.
    ❶ Press the button to turn on the power.
    ❷ When the image below is displayed on the LCD, the power is on.
    ❸ Press and hold the button for about 5 seconds. If ‘OFF’ is displayed on the screen as shown below, the automatic power off function is disabled.
    ❹ You can turn the automatic power off function back on by repeating the steps above.

  2. Protection from excessive pressurization
    • If the pressure applied to the cuff exceeds 320mmHg, the BPBIO220 will automatically exhaust the pressure from the cuff to protect the patient.

  3. Residual air detection
    • If there is any remaining air left in the cuff, the error code ‘E1’ will appear on the screen when the product power is switched on.
    • For accurate measurement result, exhaust the air in the cuff entirely before switching the power on.

  4. Pulse measurement
    • Pulse is detected when the pressure applied to the cuff is greater than 50mmHg.
    The pulse rate is measured only when the pulse is detected at least 5 times.
    • If deflation is performed too quickly, heart rate may not be checked.
    Deflate at the rate of 2 to 4mmHg/sec. to detect the pulse.
    • The measured heart rate will be displayed when the pressure is less than 20mmHg.
    • If pulse is not detected or is not constant, “- – -” will be displayed.

  5. Back-light
    • The back-light is automatically turned on or off depending on the brightness of the surrounding environment.
    • The back-light function can also be manually operated by pressing the back- light button.
    • If the back-light is operated manually with the back-light button, the back- light will not turn on or off automatically. When the product is turned off and on, the back-light automatically turns on or off again depending on the brightness of the surrounding environment.

  6. One-hand mark function
    • One-hand mark function allows display the measured value of the blood pressure on the screen momentarily when the jog switch is pressed during the measurement.
    • One-hand mark function allows display the measured value of the blood pressure on the screen momentarily once the button is pressed.
    • Up to five marks can be displayed.

  7. Exhaust speed control function
    • If the jog switch is rotated clockwise while holding the air bulb, the exhaust speed will be accelerated. A downwards arrow symbol (▼) will appear each time the jog switch is rotated clockwise. If the jog switch is rotated counterclockwise, the exhaust speed will be decelerated, and the arrow symbol (▼) will disappear every time the jog switch is rotated counterclockwise.
    • To perform “rapid exhaust,” hold the jog switch clockwise while three arrow symbols (▼) are shown on the screen. To stop the “rapid exhaust,” rotate the jog switch counterclockwise.
    • Each arrow symbol indicates the speed of exhaust at 2mmHg per second. Accordingly, 2 arrow symbols represent 4mmHg/sec, 3 arrow symbols represent 8mmHg/sec.

Blood Pressure Measurement

A. Precautions before Measurement
Caution

  • The BPBIO220 must be operated by those who are well trained of the auscultatory method of measuring the blood pressure.
  • Do not operate the BPBIO220 immediately if it has been left for a long period of time in an environment where the temperature is lower than 10ºC (50ºF) or higher than 40ºC (104ºF).
  • Do not use the BPBIO220 in a highly humid environment.
  • Do not use the BPBIO220 near any equipment which generates strong magnetic fields, electrostatic, or high frequency.
  • Patients with artificial heart or lungs may obtain inaccurate results.

Warning

  • Do not wrap the cuff on an injured arm. Injury may worsen or can be infected.
  • Do not wrap the cuff in an arm with the intravenous injection or IV plugged in.
  • Please consult a doctor before using the blood pressure monitor if you have any cardiovascular diseases.
  • Do not wrap the cuff around your neck.
  • Do not wrap the cuff tube around your neck.
  • Do not wrap the air bulb tube around your neck.

B. Measurement Instructions

  1. How to wear the cuff
    ❶ Wrap the cuff around the arm, about 2-3cm (1 in) above the elbow
    ❷ Make sure that the index (▲) on the cuff is located within the bar. Replace the cuff with the right size if the index deviates from the RANGE.
    ❸ Place the cuff so that the artery mark (▼) is located on the artery.
    ❹ Do not wrap the cuff too tightly. Leave a margin of a finger to get the best result.
    Caution
    • For correct measurement, please wear the cuff properly.
    • If the cuff is too loosely wrapped, the measured blood pressure values may not be correct.
    Note
    • The cuff provided will be M-sized cuff which is suitable for individuals with arm circumference of 22 to 32cm (9 to 13 in). If the arm circumference exceeds 32cm (13 in), use the L-size cuff.
    If the arm circumference is smaller than 22cm (9 in), use the S-size cuff.
    • L-size or S-size cuff is sold separately.

  2. Proper measurement posture
    ❶ Place the arm so that the center of the cuff is level with the patient’s heart position.
    ❷ Be relaxed during the measurement.
    Caution
    • The measurement posture may affect the blood pressure values.
    • Please keep the proper measurement posture for accurate measurement.

  3. Turning the power on
    ❶ Press the button to turn on the power.
    ❷ The symbols on the screen and the back-light will be turned on and off with a beeping sound.
    ❸ When there are no changes detected in the pressure within the cuff, number “0” will be displayed on the screen, which means it’s ready for measurement. If the pressure of the cuff is not stabilized within 20 seconds, the error code ‘E1’ will be displayed in the pressure/pulse display column on the screen.

  4. Battery level indicator
    ❶ The remaining battery level can be checked in the bottom right corner of the screen while using the blood pressure monitor.
    ❷ When using the AC adapter, the battery level is not displayed, and instead the AC adapter connection symbol ( ) appears.
    | : The capacity of the batteries is sufficient.
    ---|---
    | : Battery level is low.
    | : Battery is exhausted.

Please replace with new batteries.

  • Refer to “Inserting / Replacing Batteries” on page 27.

    1. Adjusting the back-light
      ❶ The back-light is automatically turned on or off depending on the brightness of the surrounding environment.
      ❷ The back-light function can also be manually operated by pressing the back- light button.

    2. Pressurization
      ❶ Rotate the exhaust valve clockwise to close the valve completely.
      ❷ Press the air bulb repeatedly to pressurize.
      ❸ The current pressure value is immediately displayed in the pressure bar on the screen in units of 1 step / 2 mmHg, and is displayed as a number in the pressure/pulse display column on the screen.
      ❹ Stop the pressurization when it reaches the targeted inflation pressure.

  • The maximum pressurization can be up to 320mmHg, and if pressurized at a higher rate, it will depressurize automatically and ‘—’ will be displayed in the Pressure / Pulse Rate Display on the screen. 7. Depressurization
    ❶ Rotate the exhaust valve counterclockwise to open the valve. Or you may adjust the exhaust speed by operating the jog switch. Maintain an adequate exhaust rate when
    conducting a blood pressure test.
    ❷ When the systolic blood pressure and the diastolic blood pressure are examined by auscultation, and the jog switch or the on the main body is pressed, marking will be displayed on the display bar corresponding to the instantaneous pressure value. The displayed bar does not disappear while the pressure drops.

  • Marking is displayed up to 5 times cumulatively, and if you mark more than 5 times, it will be erased sequentially from the first mark in the past.
    InBody BPBIO210 High Blood Pressure Monitor - Blood Pressure Measurement
21| Clockwise rotation: increase exhaust speed
    ---|---
    | Counter-clockwise rotation: decrease exhaust speed
    | Pressing: same as pressing the mark button
    8. Completing the measurement
    ❶ When the blood pressure measurement is completed, open the exhaust valve and exhaust the air compressed in the cuff completely.
    ❷ When the pressure drops below 20mmHg, the pulse rate will be displayed in the Pressure / Pulse Rate Display on the screen.
    The pulse rate is measured when the pressure is reduced at a rate of 2 to 4 mmHg/sec.
    If the pulse is not detected or unstable, ‘- – -’ is displayed on the screen.

  • Please review ‘B. Regarding the equipment – Pulse is not being measured’ on page 39. 9. Turning the power off
    ❶ Press the button to switch off the power.

C. Storage

  1. Wipe any stains on the main unit with a smooth cloth.
  2. Place the cuff and controller as shown below.
    The BPBIO220 should be transported or stored under the following conditions.Temperature range| -20 to 60°C (-4 to 140°F)
    ---|---
    Relative humidity| 10 to 85% RH (No Condensation)
    Atmospheric pressure range| 50 to 106 kPa

Caution
• Use UV sterilizer to disinfect the cuff.
Note
• If the cuff is contaminated and can no longer be used, please replace with a new cuff.
• Contact your local InBody representative to purchase the cuff.

BPBIO210/220 Common

FAQ

A. Error code

| Cause| Solution
---|---|---
| Excess air detected in the cuff| Exhaust all the air from the cuff completely, and restart.
Pressure sensor malfunction| If the same error message occurs continuously after exhausting the air from the cuff, the pressure sensor may be defective. Please contact Product Support.
| Detection of adapter abnormality| Use the AC adapter provided with the product after disconnecting the current adapter.
| AC adapter malfunction| Please contact Product Support.
| Low battery level| Replace the batteries.

B. Regarding the Equipment

Status Cause Solution
Power does not turn on. Check the inserted batteries.
Polarities may not be matching. Check the battery polarities and reinstall

the batteries to match with ‘+’ and ‘-’ marks.
Refer to the User’s Manual (page 10, page 26), (3) Inserting/replacing batteries.
Batteries are exhausted.| Replace all batteries with new AA batteries.
The controller does not work, or it works  abnormally.| The controller 4-pole plug is not connected properly.| Push the controller 4-pole plug all the way in tightly.
The controller is damaged.| Please contact Product Support.
Pulse is not being measured.| Blood pressure is measured too rapidly.| Keep the exhaust speed at 2 to 4 mmHg/sec.
Patient may be moving during the blood pressure measurement.| Make sure that the patient is relaxed and stable. Patient should refrain from talking during the measurement.
Pulse may be irregular or very weak.| Impossible to measure the pulse. Please use another device to measure the pulse in these cases.

Others

A. Optional Equipment

Product Name Explanation
AC adapter Dedicated adapter
L-size cuff For arm circumference of 32 to 42 cm (13 to 17 in)
M-size cuff For arm circumference of 22 to 32 cm (9 to 13 in)
S-size cuff For arm circumference of 17 to 22 cm (7 to 9 in)
Air bulb Available for BPBIO210

Does NOT contain the jog switch
Controller| Available for BPBIO220
Contains the jog switch
Cart| Dedicated mobile cart

B. Safety Information

Safety Symbols

| Warning/Caution
---|---
| Note
| Power adapter connection terminal
| BF-type equipment
| The User’s manual must be read
| CLASS II equipment (AC adapter)

Other Symbols

| Disposal guidelines
---|---
| Manufacturer
| Serial number
| Authorized representative in the EUROPEAN COMMUNITY
| European conformity
| Unique Device Identification

  • MET symbols apply to products sold in the United States and Canada.

Caution

Storage
Do not store the blood pressure monitor in the following places:

  • Where water or other liquids can reach
  • Exposure to direct sunlight
  • Strong vibration or shock
  • Dust, salt, and sulfur content
  • Hot and humid environment
  • Any equipment which generates strong magnetic field, static, or high frequency
  • Flammable objects or environment

Maintenance

  1. Clean the device with a dry cloth and do not use organic solvents.
  2. The cuff cannot be washed. Be careful not to expose the cuff to any liquid.
  3. Turn off the device completely by taking the batteries out or disconnecting the AC adapter if the device malfunctions.
  4. For inspection / repairs, please contact the local InBody representative.
  5. Do not disassemble or modify the equipment arbitrarily. Any disassembly or modification will void the warranty and may cause a fire.

Disposal

When disposing any components of the device, follow the local regulations.

C. Classification

  • Type of Protection Against Electric Shock: Class-2 equipment or internal power source-typed equipment
  • Level of Protection Against Electric Shock: BF-type
  • Classification for Protection Against Electromagnetic Wave: Category 1 / Class B
  • Level of Protection Against Flooding: General equipment (No special protection against external water infiltration)

D. Specifications

Product name
(Model name)| BPBIO210/220
---|---
Test
Instructions| Auscultatory method
Measuring Part| Left or Right Upper Arm
Test Range| Pressure (Value): 0 to 320mmHg
Pressure (Bar):  20 to 300mmHg
Pulse:  30 to 240 bpm
Degree of
Precision| Pressure (Value):  ±3mmHg
Pressure (Bar):  ±4mmHg
Pulse: Within ±5%
Minimum Scale
Unit| Pressure (Value): 1mmHg
Pressure (Bar): 2mmHg
Cuff| Cuff for the Use of 22 to 32cm (9 to 13 in) in Arm Circumference
(M size cuff)
Pressurization| Manual Pressurization Method by Air Bulb
De-pressurization| BPBIO210: Variable Exhaust Valve Method
BPBIO220: Variable Exhaust Valve Method/Variable Jog Switch Method
Rated Power| DC 6.2V, 2A (four 1.5V AA-size batteries or AC adapter)
Rated Input
(AC Adapter)| AC 100-240V , 50/60Hz, 0.32-0.19A
Operating
Environment| Temperature 10 to 40ºC (50 to 104ºF), Humidity 30 to 85% RH,
Pressure 70 to 106 kPa
Storage
Environment| Temperature -20 to 60°C (-4 to 140°F), Humidity 10 to 85% RH,
Pressure 50 to 106 kPa (No Condensation)
Electricity
Protection| Internal Power Source-Typed Equipment BF Mounting
(Mounting Unit: Cuff)
Weight| Approx 923g (Batteries excluded)
Dimension| 320(L) × 98(W) ×  73(H) mm
12.6(L) × 3.9(W) × 2.9(H) inch
Manufacturing
Country| Republic of  Korea
Manufacturing
Company| InBody Co., Ltd.

  • Specifications are subject to change without notice.
  • This product is a medical device. Please read the guide and precautions before using this blood pressure monitor.

E. EMC Declaration

Phenomenon| Basic EMC standard or test method| Operating mode| Port tested| Test Voltage| Test level/requirement
---|---|---|---|---|---
Mains terminal disturbance whop| CISPR 11:2015 +A1:2016+A2:20I9
EN 55011:2016 +A1:2017| Continuous operation mods| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| Group 1. CLus B
Radiated distudiance| CISPR 112015 +A1:20164A2:2019
EN 55011:2016 +A1:2017| Continuous operation mode| Enclosure| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| Group I.Class B
Harmonic Current Emission| IEC 61000-3-2:2018 +A1:2020
EN 61000-3-22014| Continuous operation mode| AC Mains| 230 V. 50 Hz| Class A
%Wage change. Voltage
fluctuations and Flicker Emission| IEC 61000-3-3:2013 +A1:2017
EN 61000-3-32013 +A1:2019| Continuous operation mode| AC Mains| 230 V.50 Hz| Pst: 1
Plt: 0.65
dmax: 4%
dc: 3.3%
Electrostatic Discharge Inummity| IEC 61000-4-2:2008 EN 61000-4-2.1009| Continuous operation mode| Enclosure| 100 V. 50 Hz
100 V.60 Hz
120 V. 6011z
220 V. 6011
230 V.50 Hz| ± 8 kV/Contact
± 2, ± 4, ± 8,
± 15 kV/Air
Radiated RF Electromagnetic Field Immunity| IEC 61000-1-3:2020 EN 610001-32006 +A2:2010| Continuous operation mode| Enclosure| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| 3 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Immunity to Proximity Fields from RF wireless Communications Equipment| IEC 61000-4-3:2020 EN 61000-#32006 +A2:2010| Continuous operation mode| Enclosure| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| Table 9 in
IEC 60601-1-2:
2014
Electrical Fast Transienalurst Immuni ty| IEC 61000-4-4:2012
EN 61000-4-4:2012| Continuous
operation
mode| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| ± 2 kV, 100 kHz
repetition frequency
SIP/SOP| ± 1 kV, 100 kHz
repetition frequency
Surge Immunity| IEC 61000-4-5:2014 EN 610004-52014 +A 1:2017| Continuous operation mode| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| Line to Line
± 0.5 kV, ± 1 kV
Line to Ground
± 0.5 kV, ± 1 kV, ± 2 kV
Immunity to
Conducted
Disturbances
Induced by
RF fields| IEC 61000-4-6:2013
EN 61000-4-6:2014| Continuous
operation
mode| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| 3 V
0.15-80 MHz
6 V in ISM bands
Between 0.15 MHz and
80 MHz
80% AM at 1 kHz
SIP.SOP
Power Frequency
Magnetic Field
Immunity| IEC 61000-4-8:2009
EN 61000-4-8:2010| Continuous
operation
mode| Enclosure| 100 V, 50 Hz
100 V, 60 Hz
120 V, 60 Hz
220 V, 60 Hz
230 V, 50 Hz| 30 A/m
50 Hz & 60 Hz
Voltage dips| IEC 61000-4-11: 2020
EN 61000-4-11:2004
+A1:2017| Continuous
operation
mode| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
240 V, 50 Hz
240 V, 60 Hz
220 V, 60 Hz| 0 % UT: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0º
Voltage
interruptions| IEC 61000-4-11: 2020
EN 61000-4-11:2004
+A1:2017| Continuous
operation
mode| AC Mains| 100 V, 50 Hz
100 V, 60 Hz
240 V, 50 Hz
240 V, 60 Hz
220 V, 60 Hz| 0 % UT; 250/300 cycle

Electromagnetic immunity
The BPBIO210/220 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the BPBIO210/220. Otherwise, degradation of the performance of this equipment could result.

Immunity test| Band a)| Service a)| Modulation b)| IEC60601 test level| Compliance level
---|---|---|---|---|---
Proximity fields from
RF wireless Communications IEC61000-4-3| 380 – 390 MHz| TETRA 400| Pulse modulation b)18Hz| 27 V/m| 27 V/m
430 -470 MHz| GMRS 460
FRS 460| ±-5 kHz
deviation
1 kHz sine| 28 V/m| 28V/m
704 – 787 MHz| LTE Band 13, 17
‘| Pulse modulation b) 217 Hz| 9 V/m| 9 V/m
800 -960 MHz| GSM800:900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5| Pulse modulation b) 18 Hz| 28 V/m| 28 V/m
1700 -1990 MHz| GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band
1,2,4,25
UMTS| Pulse modulation b) 217 Hz| 28 V/m| 28 Vhn
2400 – 2570 MHz| Bluetooth
WLAN
802.11b/g/n
RFID 2450
LTE Band 7| Pulse modulation b)
217 Hz| 28V/m| 28 V/m
5100 – 5800 MHz| WLAN 802.11a/n| Pulse modulation b)
217 Hz| 9 V/m| 9 V/m

NOTE : If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The 1m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.

inbody.com
©2024 InBody Co., Ltd. All rights reserved.