InBody 580 Body Composition Analyzer User Manual
- June 1, 2024
- InBody
Table of Contents
- InBody 580 Body Composition Analyzer
- Product Information
- Product Usage Instructions
- Intended Use
- Medical Indication
- Intended user profile
- Safety
- Product Overview
- Installation
- Setting
- Connecting Compatible Device
- InBody Test
- Maintenance and Storage
- Frequently Asked Questions
- Classifications and Specifications
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
InBody 580 Body Composition Analyzer
Product Information
Specifications
- Age Range: 3+ years
- Weight Range: 2~300kg
- Height Range: 95 ~ 220cm
- Language Support: Basic English and other languages as specified
Product Usage Instructions
Intended Use
The InBody is designed for body composition analysis. It provides detailed information about body fat, muscle mass, and water content.
Medical Indication
The device should not be used by individuals with medical implant devices such as pacemakers or essential support devices. It may cause malfunctioning of the devices or pose a risk to the user’s life.
Contraindication
Individuals with known metal allergies to stainless steel materials should avoid using this equipment.
Intended User Profile
The user should have basic education to understand on-screen instructions, knowledge of the language displayed (English and other supported languages), and no specific experience level is required.
Intended Patient Population and User Profile
The intended patient population includes individuals aged 3 years and above, weighing between 2 to 300kg, and with a height ranging from 95 to 220cm. Patients must be able to stand for 1-2 minutes, be in a woken-up and mentally healthy state, and not have medical implant devices or essential support systems.
FAQs
- Q: Can pregnant women use the InBody?
- A: It is not recommended for pregnant women to use the InBody due to the electrical currents involved in the analysis.
- Q: Can individuals with metal implants use the InBody?
- A: Individuals with metal implants such as pacemakers should avoid using the InBody to prevent any risks associated with electrical currents.
- Q: How often should I use the InBody for accurate results?
- A: For best results, it is recommended to use the InBody consistently at the same time of day, under similar conditions, such as hydration levels and food intake.
“`
User’s Manual
Intended Use
The InBody is mainly used for healthy and acute or chronically ill populations in hospitals, medical practices and inpatient care facilities in accordance with national regulations. It can be used to assist in the assessment of nutritional status, obesity and muscle balance. Body composition analy is important in preventive medicine since it provides the basis of appropriate physical activity and dietary habits for improving personal daily routine. It can be also usefully applied to follow-up studies of patients treated for various diseases.
Medical Indication
· Medical check-up: Four body composition analysis can be identified for the
risk of developing diseases that are highly related to body composition
imbalance like obesity, malnutrition, fluid imbalance and osteoporosis for
medical check-up.
· Obesity: Percent body fat has been recommended rather than BMI to ensure
proper weight loss and improvements in long-term health, tracking changes for
adjusting/developing customized treatments.
· Pediatric obesity: Body composition measurement is an essential part of
health assessments for children and adolescents. Percent Body fat is better
than the indicators of weight status to identify children and adolescents with
unfavorable lipid profile.
· Sarcopenia: InBody provides a quick, easy to perform test that provides a
calculation for skeletal muscle index(SMI), the sum of the lean mass in the
arms and legs, normalized for height. This marker is useful in identifying low
muscle in the appendages, which increases frailty risk.
· Diabetes & endocrinology: Diabetes is often associated with excess fat,
however having insufficient muscle mass is just as detrimental and increases
diabetes risk. And visceral fat plays a key role in the development of
metabolic and cardiovascular disease.
· Edema: Over-hydration as assessed by ECW ratio(ECW/TBW) is prevalent in
dialysis patients, and is associated with loss of residual renal function,
inflammation, malnutrition and hypertension. Monitoring the ECW ratio
(ECW/TBW) provides an assessment of fluid accumulation in the extracellular
space resulting from compromised cardiovascular function. The patients who did
not have ascites originally but have higher ECW/TBW had a higher incidence of
ascites in liver cirrhosis.
· Segmental fluid retention: InBody objectively measures each region of the
body separately and provides segmental ECW ratio measures for each of the
arms, legs and the trunk, and these measures can be used to detect fluid
imbalances resulting from the development or progression of lymphedema.
· Nutrition: The four primary components of the nutritional assessment are
summarized by the mnemonic ABCD, with A standing for anthropometric
measurements including stature, body weight, BMI and body composition. Body
composition analysis can reveal changes in body composition (body water,
protein, minerals and body fat) that cannot be known by changes in body
weight.
· Fitness: Strength training greatly stimulates muscle growth, exercise burn
the calories strengthens cardiorespiratory capacity, which reduce the risk of
diabetes, heart disease, and other health concerns and result in the various
changes in body composition. Body composition analysis shows skeletal muscle
mass and lean in each segment of body, it helps focusing on building more
muscle or correct imbalance.
· The InBody device is not a diagnostic device. To make an accurate diagnosis,
the physician needs to commission thorough examinations and take their results
into account in addition to the results of the InBody.
· The InBody device is not used in home healthcare environment.
3. Contraindication
Individuals with medical implant devices such as pacemakers, or essential
support devices such as patient monitoring systems, must not use this
equipment. Safe, low-level currents will flow through the body during the
test, which may cause malfunctioning of the device or endanger lives.
Individuals with known metal allergies against stainless steel materials shall
not use the equipment.
Intended user profile
1. Education: – At least, the user needs to be able to understand explanation
of words on screen.
2. Knowledge: – At least, the user needs to be able to understand explanation
of words on screen. – No maximum.
3. Language understanding: – Basic language: English – Languages are
supported as specified in the marketing need.
4. Experience: – No minimum and maximum.
5. Intended patient population and user profile
1. Age: 3+ years 2. Weight: 2~300kg 3. Health: Examinee need to be able to
stand for 1~2minutes. 4. Condition: Individuals with medical implant devices
such as
pacemakers, or essential support devices such as patient monitoring systems,
must not use this equipment. The currents will flow through the body during
the test, which may cause malfunctioning of the device or endanger lives. 5.
Nationality: Multiple 6. Patient state: Woken up, mentally healthy 7. Height:
95 ~ 220cm
InBody User’s Manual for Measurement Guide and Setup
Thank you for purchasing the InBody. This user’s manual describes all the
features of the InBody. Please read before use and keep it in a safe place. By
following the manual instructions, you will be able to use the InBody more
safely and effectively.
Headquarters Information
[]
06106 625
TEL: 02-501-3939FAX: 02-6919-2417: 1899-5841
Website: inbody.com
E-mail: info@inbody.com
InBody Co., Ltd. [HQ]
625, InBody Bldg., Eonju-ro, Gangnam-gu, Seoul 06106
Republic of Korea
TEL: +82-2-501-3939
FAX: +82-2-6919-2417
Website: inbody.com
E-mail: info@inbody.com
31025 15
TEL: 041-581-3003
FAX: 041-581-3103
Website: inbody.com
E-mail: info@inbody.com
InBody Co., Ltd. [MANUFACTURER]
15, Heugam-gil, Ipjang-myeon, Seobuk-gu, Cheonan-si,
Chungcheongnam-do 31025 KOREA
TEL: +82-41-581-3003
FAX: +82-41-581-3103
Website: inbody.com
E-mail: info@inbody.com
Representative & Sponsor Information
InBody Europe B.V. [NETHERLANDS]
Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands
TEL: +31-20-238-6080
FAX: +31-6-5734-1858
Website: nl.inbody.com
E-mail: info.eu@inbody.com
InBody Germany [GERMANY]
InBody Europe B.V. Niederlassung Deutschland,
Mergenthalerallee 15-21, 65760 Eschborn, GERMANY
TEL: +49-6196-769-1662 FAX: +49-6196-76916-11
Website: de.inbody.com
E-mail: erfolg@inbody.com
InBody UK [UNITED KINGDOM]
11 Phoenix Park, Telford Way, Stephenson Industrial Estate,
Coalville LE67 3HB, United Kingdom
TEL: +44-1530-569620
Website: uk.inbody.com
E-mail: uk@inbody.com
InBody Oceania [AUSTRALIA]
Main office: Level 8, 1 York Street, SYDNEY, NSW 2000, Australia
Showroom: U2/82-86 Minnie Street, Southport, Queensland
TEL: +61-7-5681-1900
Website: au.inbody.com
Email: oceania@inbody.com
Customer Service Information
InBody USA [USA]
13850 Cerritos Corporate Dr. Unit C Cerritos, CA 90703 USA
TEL: +1-323-932-6503
FAX: +1-323-952-5009
Website: inbodyusa.com E-mail: info.us@inbody.com
InBody BWA Inc. [USA] 2550 Eisenhower Avenue, Suite C 209, Audubon, PA 19403 TEL: +1-610-348-7745 Website: inbodybwa.com E-mail: bwainquiries@inbody.com
[JAPAN]
137-0071 1-28-6
TEL: +81-3-5875-5780 FAX: +81-3-5875-5781
Website: inbody.co.jp
E-mail: inbody@inbody.co.jp
[] [CHINA] [] 1698903904 +86-21-6443-9705 +86-21-6443-9706 inbodychina.com info@inbodychina.com
InBody Asia [MALAYSIA & SINGAPORE]
Unit 3A-11, Oval Damansara, 685 Jalan Damansara Kuala Lumpur,
WP KL 60000 Malaysia
TEL: +60-3-7732-0790
FAX: +60-3-7733-0790
Website: inbodyasia.com E-mail: info@inbodyasia.com
InBody MEXICO [MEXICO] Av. Eugenia 197 Piso 1 Ofic 1-B, Col. Narvarte, Benito Juarez, C.P. 03020, Ciudad de Mexico, Mexico TEL: +52-55-5025-0147 Website: inbodymexico.com E-mail: info.mx@inbody.com
InBody India [INDIA]
57/57 A,1st Floor, Raj Industrial Complex, Military Road, Marol,
Andheri (East). Mumbai- 400059, Maharashtra, India
TEL: +91-22-6223-1911
Website: inbody.in
E-mail: india@inbody.com
Copyright
Reproduction, adaptation, or translation of this manual is prohibited without
prior written consent from InBody Co., Ltd. under the copyright laws. This
manual may be printed incorrectly and subject to change without notice. InBody
Co., Ltd. shall not be liable for any errors, incidental, or consequential
damages that occurred by not complying with the content of the User’s Manual.
Visit our website (inbody.com) to view and download additional information
about the InBody. InBody Co., Ltd. reserves the right to modify the
appearance, specifications, etc. of this product to improve its quality,
without prior notice.
©2024 InBody Co., Ltd. All rights reserved.
Safety
1.1 Safety Symbols used in the User’s Manual
Warning
Failure to comply with safety warnings and regulations can cause serious
injury or death.
Caution
Failure to comply with safety cautions and regulations can cause injury or
property damage.
1.2 Precautions for Use
Warning
· Individuals with medical implant devices such as pacemakers, or essential
support devices such as patient monitoring systems, must not use this device.
Safe, micro alternative currents will flow through the body during the test,
but this may cause malfunctioning of the device or endanger lives. InBody Co.,
Ltd. shall not be liable for any damages to an individual or an equipment that
occurred by not complying with the content above.
· The Bioelectrical Impedance Analysis (BIA) method does not harm the human
body because it uses micro alternative currents. However, if you are pregnant,
please consult your doctor or specialist.
· It is not recommended to use the device with an individual who has a
contagious disease or an infectious disease. If an individual with any kind of
contagious disease or infection tests on the InBody, use an alcohol-based
disinfectant (e.g. 70% ethanol) to clean the device.
· Do not pour the liquid cleaner when cleaning the device. If liquid cleaner
flows into the device, it may cause an equipment failure or an electric shock
due to a short circuit.
· Do not use this device for any purpose other than body composition analysis
or weight measurement.
· This product is not a diagnostic device. To make an accurate diagnosis,
consult your doctor.
· Failure to follow these instructions may result in the user suffering
serious injuries.
· Failure to comply with safety warnings and regulations may result in the
user’s death or serious injury.
· When not in use for an extended period, please disconnect the power plug
from the outlet.
· Please do not apply excessive force when removing the power plug.
Caution
· This is a sensitive device which precisely measures the body composition. If
you test near electronic products such as refrigerators, TVs or right under
fluorescent lights, the test results may be inaccurate. Please use the device
away from electronic devices.
· Do not use the device in a humid space such as a bathroom, as excessively
high or low temperature, humidity, and pressure may affect the operation of
the device. Use in the installation environment specified in the product
specifications.
Caution
· Do not allow any liquid substances to contact the device directly. Keep food
and drinks away from the device. Substances getting inside the device can
cause critical damage to the electronic components.
· Do not disassemble or modify the device including internal parts without
written consent from the manufacturer. This may cause electric shock or
injury, device malfunction, inaccurate test results, and will void the
manufacturer’s warranty.
· Children or people with restricted mobility should be tested with the help
of a instructor or assistant.
· When storing the device for a long period of time, store it on a flat
surface after turning off the device, unplugging the adapter, and packing the
device.
· Dispose of the device and its batteries in accordance with the relevant
local laws and regulations.
· Repairs and inspections can only be performed by InBody’s technician. For
repairs and inspections, contact the customer service.
· Do not support yourself on the equipment when stepping up or down from the
footplate.
· Please be careful not to trip or get your foot caught in the footplate.
· Failure to follow these instructions may result in product damage or
inaccurate test results.
· Failure to comply with safety precautions and regulations may result in the
user suffering injuries or incurring property damage.
6
Product Overview
2.1 Unpacking
Open the packing box and remove the packaging pad. Take the device out of the
packing box.
Packaging Pad
Note · For repackaging the device at a later time, the supplied
packing materials must be kept. Other wastes should be disposed according to
relevant local laws and regulations. · To repack the device, please refer to
“7.4 Repacking and Transportation”.
2.2 Product Components
Please check if the device has any damage prior to installation.
Power Adaptor
Power Cable
UseUr’sseMr’asnMuaanl ua5l 80
InBody580
User’s Manual
Note · The specifications of power cables can vary by country.
2.3 Optional Product
2.4 Name of Each Part
Operation Panel
1
2
7
3
4
5
8
9
6
10
11
12
11
1 Display: Displays each step of test, guide, and test result.
2 Number Button: Enter numeric data such as age and height.
3 Function: Enter the “FAQ” screen directly from the test standby screen.
4 Setup: Enter the “Setup” from the test standby screen.
5 Print: Reprint test results.
6 Decimal Point: Enter the decimal point in ID, height, age and weight.
7 Power Button: Turn the device on and off.
8 Gender: Select the gender. (M: Male, F: Female)
9 Enter: Used when input is completed or changes are saved in the Setup.
10 Delete: Used to delete the saved data.
11 Hand Electrode: Sends a weak current to the upper body and measures the
voltage for the InBody test.
12 Smart Reader: InBody App user can transfer the personal information needed
for the test with QR code.
Serial Distributor (SD400)
7
Product Overview
Footplate
2
2
3
4 1
5
1 Footplate: Connected to loadcells that measures user’s weight.
2 Foot Electrode: Sends a weak current to the lower body and measures the
voltage for the InBody test.
3 Level Indicator: Shows the level status of the footplate.
4 Hinge Cover: Used to cover the connecting part of the upper section and
lower section.
5 Level Supporter: Can fit the level by adjusting height.
Rear View
see what you’re made of
12
3 4 5
1 9-pin Serial Port (Female, RS-232C): Connect to a stadiometer, a blood
pressure monitor and SD400.
2 9-pin Serial Port (Female, RS-232C): Connect to LookinBody120 installed on
the PC.
3 USB HOST Port: Connect printer, USB flash drive, and barcode reader.
4 LAN Port (10/100T-Base): Connect to the Internet or LookinBody120 installed
on the PC.
5 Power Inlet: Connect a power adapter.
Note
· Make sure to connect only InBody stadiometer, Blood Pressure Monitor and
SD400.
· When connecting to a PC where the data management program LookinBody120 is
installed, you can connect with one of the serial port, Bluetooth, LAN, or Wi-
Fi port.
· Serial, USB, and LAN cables are not included in the product.
8
Installation
3.1 Installation Environment
Caution
· If you are operating the device in a place where the altitude is 2,000m or
higher, the weight measurement may be affected.
· Use the device in a location where it is not exposed to direct sunlight. It
may cause discoloration or damage of device.
Check the environment before installing the device.
· Using the device in a dry environment or on a carpet may result in static
electricity and damage. Use an antistatic mat if you need to install in an
environment with static device.
· Install the device on the floor that is flat and vibrationfree. If the
device is installed where the floor is not flat, it may topple during a test
or the test results may be inaccurate.
This device is suitable for indoor use. If installing this device outdoors,
the following requirements must be fulfilled.
Temperature Relative humidity Atmospheric pressure
10 – 40 (50 – 104 °F) 30 – 75 % RH (No Condensation) 70 – 106 kPa
3.2 Installation
Warning
· Always use an outlet connected to the rated power (AC 100 – 240 V). Using
other power rated outlets may result in fire or malfunction.
· When using a power surge protector, make sure that the outlet or the
extension cable has adequate power capacity.
Caution
Hold the footplate when moving the device. When transporting the device, do
not grab upper section (LCD) or hand electrode connection area. Or, the device
may be damaged.
1 Raise the upper part of the device and remove the protective packaging
vinyl.
Protective Packaging Vinyl
2 Push the hinge cover horizontally so that the hinge cover is joined to the end of the stand. Hinge Cover
3 Level the device by rotating the level supporter under the footplate to the
left and right.
· You can check whether the device is level or not checking the air bubble in
the level indicator.
· For accurate weight measurement, the device should be leveled.
Level Indicator
Leveled Not Leveled
Level Supporter
Caution Be careful not to get your hands hurt when handling the Level
Supporter.
O
X
X
9
see what you’re made of see what you’re made of
Installation
4 Connect the power cable to the device. a Connect the power cable to the
power adapter. b Connect the power adapter to the power inlet on
the rear of the device.
c Connect the power cord to a 3-terminal outlet with a ground terminal.
Power Inlet Power Adapter Power Cable
Power Cord
Caution · Use the power adapter provided by InBody. The device
may malfunction when using other adapters. · The device may get damaged or
malfunctioned due to
the electric if plugging into an ungrounded outlet. Or the test results may be
inaccurate.
5 Press the power button to turn it on.
Caution If you are connecting the InBody device to compatible device then turn
on the compatible device first. On the contrary, turn off the power of InBody
device first, and then turn off the power of the compatible device when
turning off the power. This can minimize the electric shock to the InBody
device.
The InBody device can be connected to optional products and PC program such as
stadiometer, blood pressure monitor, printer, smart reader, serial
distributor, and LookinBody120. For more details, refer to “5 Connecting
Compatible Device”.
10
Setting
4.1 Initial Setup
The InBody device automatically starts booting when it is turned on. While
booting, it performs a self weight calibration.
Caution While booting, please do not stand on the footplate or place objects
on the footplate.
When InBody device is booted for the first time, the initial setting screen
appears. Complete the initial setup selecting the appropriate item.
1 Select whether to use the Self Mode.
· Self Mode is suitable for an environment without an administrator, as the
screen is configured for testing alone.
3 Select Weight Measurement Mode.
· Select whether to proceed with the InBody test completing the weight
measurement or to automatically measure the weight during the InBody test.
Note You can change the initial settings at any time in Setup.
4.2 Setup
1 Press Start InBody after completing initial setting.
· You can choose whether to use the membership number or mobile number for the
InBody User ID.
2 Select how to manage the User ID.
2 Press Setup on the test standby screen or Setup button on the keypad of the operation panel before stepping on the footplate.
· You can choose whether to use the membership number or mobile number for the
InBody User ID.
11
Setting
3 Enter the administrator password and press Enter.
· This screen will appear only once for initial password setup.
4 Press the menu in the Setup and change the setting.
Note Be careful not to forget the password you set. If you have forgotten your
password, please contact InBody Customer Service.
4.3 Overview of the Setup Menu
Set your InBody device fits to your facility and manage the data.
General
· Date and Time: Set the date and time displayed at the test standby screen
and on the result sheet.
· Units: Set the unit displayed on the InBody device. · Country: Select the
country where InBody device is
installed. · Language: Select the language to be applied to InBody
device. · Password Lock Setting: You can lock the device to
control the accessibility. · Volume: Adjust the volume of InBody device. ·
Touchscreen Alignment: Calibrate the accuracy of the
touch screen. · Additional Function Setup: Set functions other than
the basic functions.
Cloud Service
Cloud Service is a sevice that transmits InBody results to the member’s mobile
phone so that the members can check the result at any time with the InBody
App.
You can send the InBody results by entering your mobile number during the
test.
Note
The result can be sent only when the Internet connection is set in the Setup >
3.Connect > Internet, and 2.Cloud Service is ‘On’. When the result is not
sent, refer to “8.1 Regarding the Device”.
12
Setting
Connect
· Internet: This option allows connecting the InBody device to the Internet.
Once the device is connected to the Internet, the test results can be shared
to InBody App or the LookinBody120, the personal membership management PC
application without distance limitation. To connect to the Internet, refer to
“5.8 Connecting Internet”.
· Bluetooth: This option allows setting up the Bluetooth so that the InBody
can connect to LookinBody120, or to other compatible devices such as BSM
Stadiometer series, BPBIO Blood Pressure Monitor series. For details to
connect Bluetooth, please refer to “5.7 Connecting Bluetooth”.
· Serial: This option allows connect a PC with member management program,
LookinBody120 or other compatible devices. For details on how to connect
compatible devices, refer to “5 Connecting Compatible Device”.
Data Management
· View/Print/Delete Data: This option allows checking, printing or deleting
the test result stored in the InBody device as a membership number or mobile
number.
· Export Data as a MS Excel File: This option allows copying the test result
saved in the InBody device to USB flash drive as an Excel file format. You can
check the test results opening the Excel file on your computer. A file that
calls the test results saved in InBody device into the member management
program LookinBody120 is created together in USB flash drive.
· Data Backup/Restoration/Combine: This option allows saving the test results
to USB flash drive for backup or restores the backed up test results to the
InBody device. If you are using multiple devices, the data can be combined.
Printer and Print Settings
· Printer Setup: This option allows setting the printer to connect to the
InBody. Connect the printer to the InBody device to print your Result Sheet.
· Automatic Printing Options: This option allows setting the Result Sheet to
be printed automatically after completing the test. You can print up to two
Result Sheets per each Result Sheet type at one time.
· Print Results on A4 paper: You can choose to print out the Result Sheet on
the plain A4 paper after test. However, you can only print in black and white
if you print the result on the plain A4 paper.
· Printing Alignment: This option allows adjusting the position of the result
to be printed on the Result Sheet.
Detailed Result Sheet Settings
· InBody Result Sheet: You can set whether to use the InBody Result Sheet and
the Result Sheet Items (Body Composition History, Right Side Outputs,
Segmental Lean Analysis).
· InBody Result Sheet for Children: You can set whether to use InBody Result
Sheet for Children and the Result Sheet Items (Body Composition History, Right
Side Outputs, Standard Child Growth Curve). You may also set the Result Sheet
Setting (Change the Age Range of Child).
· Custom Logo: You can preview the logo printed on the upper right of the
Result Sheet.
Note
Please contact Customer Service for help with uploading or modifying a logo.
Test Mode & Normal Range
· Self Mode: Select whether to use Self Mode. Self Mode is suitable for an
environment without an administrator, as the screen is configured for testing
alone. When using Self Mode, InBody test can be started without entering ID.
· Member Management Method: You can choose whether to use the ID or mobile
number for the member ID.
· Weight Input Method: You can select whether to automatically measure the
weight as soon as the subject stands on the footplate, to measure the weight
before the test, or to enter the weight.
· Adjust Weight: Adjust measured weight by a fixed value on the InBody device.
(Example: Workout clothes at the gym are approximately 0.2kg; most people are
assumed to be wearing workout clothes, so the instructor may adjust the set
value to -0.2kg.)
· Bypass Age/Gender: The subject can bypass inputting their age or gender if
the test environment is designed for testing a specific age group or gender.
· Normal Range: This option allows setting the normal range of BMI, Percent
Body Fat, and Waist-Hip Ratio. The ideal value of BMI may also be set.
13
Connecting Compatible Device
To connect a compatible device to InBody device, check the communication
method of the compatible device.
There are two ways of communication; wired connection such as USB or RS232C
(9-pin serial port, Female), and wireless connection (Bluetooth).
5.1 Printer
In order to print InBody Result Sheet, an InBody compatible printer is
required.
1 Turn off the InBody device.
· When InBody device is already turned on, the printer might not properly
connect.
2 Connect the USB cable supplied with the printer to the USB HOST port on the
rear of the InBody device. Connect the other end of the USB cable to the USB
port of the printer.
3 Turn on the printer. 4 Turn on the InBody device. 5 Connect the printer
according to the instructions on
Setup > 5.Printer and Print Settings > Printer Setup.
Note You can set the printing options in Setup > 5.Printer and Print Settings
and 6.Detailed Result Sheet Settings.
5.2 Stadiometer
The height values measured by the stadiometer are sent to InBody when it is
connected to InBody device.
1 Turn off the InBody device.
· When InBody device is already turned on, the stadiometer might not properly
connect.
2 Connect the serial cable supplied with the stadiometer to the right serial
port on the rear of the InBody device. Connect the other end of the cable to
the serial port of the stadiometer.
3 Turn on the stadiometer. 4 Turn on the InBody device. 5 Select Stadiometer
in Setup > 3.Connect > Serial.
When the stadiometer is connected, a message will be displayed on the test
standby screen.
14
Note · Please make sure to connect only BSM Stadiometer
series provided by InBody. · You can also connect the stadiometer via
Bluetooth.
For details on device connection, refer to “5.7 Connecting Bluetooth”.
5.3 Blood Pressure Monitor
The blood pressure values measured by the blood pressure monitor are sent to
InBody device when it is connected to InBody device.
1 Turn off the InBody device.
· When InBody device is already turned on, the blood pressure monitor might
not properly connect.
2 Connect the serial cable supplied with the blood pressure monitor to the
right serial port on the rear of the InBody device. Connect the other end of
the cable to the serial port of the blood pressure monitor.
3 Turn on the blood pressure monitor.
4 Turn on the InBody device.
5 Select Blood pressure monitor in Setup > 3.Connect > Serial.
When the blood pressure monitor is connected, a message will be displayed on
the test standby screen.
Note · Please make sure to connect only the BPBIO Blood
Pressure Monitor series provided by InBody. · If you select blood pressure in
the sub options in
Setup > 6.Detailed Result Sheet Settings > InBody Result Sheet you can print
the blood pressure value on the InBody Result Sheet. · You can also connect
the blood pressure monitor via Bluetooth. For details on device connection,
refer to “5.7 Connecting Bluetooth”.
Connecting Compatible Device
5.4 Barcode Reader
The ID will be inputted automatically if a barcode reader is connected to the
InBody device.
1 Turn off the InBody device.
· When InBody device is already turned on, the barcode reader might not
properly connect.
2 Connect the USB cable of the barcode reader to the USB HOST port on the rear
of the InBody.
3 Turn on the InBody device.
When the barcode reader is connected, a message will be displayed on the test
standby screen.
Note If barcode reader is not recognized, please contact InBody Customer
Service.
5.5 Data Management Software (LookinBody120)
By connecting LookinBody to InBody device, you can manage your InBody data.
1 Turn off the InBody device.
· When InBody device is already turned on, LookinBody120 might not properly
connect.
2 Connect the serial cable provided with LookinBody120 to the right serial
port on the rear of the InBody. Connect the other end of the cable to the
serial port of the PC.
3 Turn on the InBody device.
4 Launch LookinBody120 installed on your PC and follow its instructions to
connect to InBody device.
Note · If LookinBody120 is not recognized, please contact
InBody Customer Service. · You can connect the LookinBody120 via Bluetooth
by connecting the Bluetooth dongle included in LookinBody120 to your PC. For
details on device connection, refer to “5.7 Connecting Bluetooth”. · You can
also connect the LookinBody120 via Internet. For details on device connection,
refer to “5.8 Connecting Internet”.
5.6 Serial Distributor (SD400)
With SD400, you can connect both the stadiometer and blood pressure monitors
at the same time.
1 Turn off the InBody device.
· When InBody device is already turned on, the SD400 might not properly
connect.
2 Connect the serial cable supplied with the SD400 to the left serial port on
the rear of the InBody device. Connect the other end of the cable to the
serial port of the SD400.
3 After connecting the stadiometer or blood pressure monitor to each serial
port of the SD400, turn on the power of each device.
4 Turn on the InBody device. 5 Select SD400 in Setup > 3.Connect > Serial and
press Next.
6 Press Edit to configure the device(s) connected to each port of the SD400.
When each device is connected, the standby screen will display a message that
the product is connected.
5.7 Connecting Bluetooth
Requirements · The compatible device to be connected must
support Bluetooth. · Bluetooth may not operate normally if the
compatible device is more than 10m away from InBody device. · There should be
no obstacles such as walls between the InBody device and the compatible
device.
Connecting Stadiometer/Blood Pressure Monitor
1 Press Setup on the test standby screen or Setup button on the keypad of the
operation panel before stepping on the footplate.
2 Enter the password.
15
Connecting Compatible Device
3 Select 3.Connect > Bluetooth and turn on the Bluetooth by pressing ON/OFF
button.
If the compatible device is paired properly, the Bluetooth ID of the device
connected to the InBody device will appear.
Press the Exit on the upper-left to return to the test standby screen. The
Bluetooth icon will be displayed on the upper-left and a message will let you
know the device is connected.
4 Press ON/OFF button of the compatible device that you want to connect to.
5 Turn the device (stadiometer or blood pressure
monitor) that you want to connect with InBody device.
6 The compatible device’s Bluetooth ID consists of “Product Name-Serial
Number”. Check the product name and serial number on the name plate attached
to the compatible device.
7 Select the Bluetooth ID of the device to be paired and press Connect.
Connecting LookinBody120
1 Press Setup on the test standby screen or Setup button on the keypad of the
operation panel before stepping on the footplate.
2 Enter the password. 3 Select 3.Connect > Bluetooth and turn on the
Bluetooth by pressing ON/OFF button.
4 Launch LookinBody120 installed on your PC and follow its instructions to
connect to InBody device.
Note If the Bluetooth ID does not appear or the connection is not good, please
contact InBody Customer Service.
16
Note
· Refer to the User’s manual of LookinBody120 for more details on how to setup
Bluetooth on LookinBody120.
· You can check the Bluetooth ID of InBody in Setup > 3.Connect > Bluetooth.
Connecting Compatible Device
5.8 Connecting Internet
Once InBody device is connected to the Internet, you can use it to connect the
Cloud Services or LookinBody120.
Note If the Cloud Service does not work or if LookinBody Web is not
recognized, please contact InBody Customer Service.
Connecting LAN
1 Press Setup on the test standby screen or Setup button on the keypad of the
operation panel before stepping on the footplate.
2 Enter the password. 3 Select 3.Connect > Internet and turn on the Internet
by pressing ON/OFF button.
4 Connect the LAN(RJ45) cable to the LAN port of the InBody device. · A LAN
cable must be connected to the terminal(router) registered with Internet
service or to the LAN(RJ45) port on the wall of the building.
5 Press LAN (Wired) and press Next.
Connecting Wi-Fi
1 Press Setup on the test standby screen or Setup button on the keypad of the
operation panel before stepping on the footplate.
2 Enter the password. 3 Select 3.Connect > Internet and turn on the Internet
by pressing ON/OFF button.
4 Press Wi-Fi and press Next.
5 Select the Wi-Fi to be connected and press Select.
· If you need to manually set up the Wi-Fi network, press Input manually.
· Wi-Fi IDs and passwords can be recognized only when they consist of
alphabets, numbers, or symbols.
6 Select Automatic or Input IP Manually.
· Automatic: IP address will be automatically set. · Input IP Manually: You
can manually enter the IP
address.
Once the device is connected to the internet, the IP address will appear.
6 Enter the Wi-Fi password if needed and press Enter.
· Once the device is connected to the Wi-Fi, the network name (SSID) and IP
address will appear.
17
Connecting Compatible Device
5.9 Optional device’s Specification
1 Stadiometer
· DC 6V-1.5V AA battery 4EA or · Input: 100-240V, 50/60Hz, 1.2A · Output: 12V,
3.33A
2 Blood Pressure Monitor
· AC 100-240V, 50/60Hz · AC 100-120/200-240V, 50/60Hz
3 Printer
· AC 220-240V~, 50/60Hz, 2.5A
4 LookinBody120 and PC
· Operating system: Microsoft Windows XP SP3/ Windows Vista SP1/ Windows
7(32bit/64bit) compatible/ Windows 8(32bit/64bit) compatible
· CPU: 1.7GHz or higher processors · Hard Disk: More than 1GB of free space ·
Memory (RAM): 1GB or higher recommended · Graphics cards and monitors:
Resolution 1024×768
5 Interfaces and network protocol
Interfaces
Protocol
Wi-Fi
IEEE 802.11 b/g/n 2.4GHz IEEE 802.11 a/n/ac 5GHz(under CH. 100)
LAN
IEEE 802.3u
USB
USB 2.0
Bluetooth
BLE 5.0, 2.4GHz ~ 2.4835GHz
6 Recommended WiFi settings
Settings
Recommended settings
Consequences in the event of
different settings
Authentication / encryption
WPAPSK/WPA2PSK + TKIP WPA2PSK + TKIP WPAPSK + TKIP WPAPSK + TKIP/AES WPAPSK/WPA2PSK + TKIP/AES WPA2PSK + TKIP/AES WPAPSK + AES WPAPSK/WPA2PSK + AES WPA2PSK + AES(recommend)
No network connection, other encryption methods not supported
Frequency Band
2.4GHz: 2412~2484 MHz
5GHz: 5180~5320 MHz(under CH. 100)
No network connection, above frequency band not supported
5.10 IT security measures
InBody would like to clarify that the user access to the InBody580 is only
granted for the authorized users, who have appropriately registered the
passcode in the system setting menu of the InBody580. The actual steps for
registering the access passcodes which is implemented to grant the access to
only the authorized users are illustrated with the images of the passcode set
ups as shown below:
18
InBody Test
6.1 Precautions for Test
Warning
· Individuals with medical implant devices such as pacemakers, or essential
support devices such as patient monitoring systems, must not use this device.
Safe, micro alternative currents will flow through the body during the test,
but this may cause malfunctioning of the device or endanger lives. InBody Co.,
Ltd. shall not be liable for any damages to an individual or a device that
occurred by not complying with the content above.
· The Bioelectrical Impedance Analysis (BIA) method does not harm the human
body because it uses micro alternative currents. However, if you are pregnant,
please consult your doctor or specialist.
For an accurate measurement, factors affecting the measurement must be
controlled. Please follow the precautions below before the test.
· Test with empty stomach. Food in your stomach affects your weight and is
considered part of the body composition, which may affect your test results.
· Test in the morning if possible. In the afternoon, body water tends to be
driven to your lower body, which can affect your test results.
· Test after using the bathroom. The volume of urine and excrement is included
in the weight measurement affecting accuracy of the test results.
· Test before exercising. Even light exercise can temporarily change body
composition.
· Avoid using the sauna or bath before measuring. The body water is
temporarily unstable as blood flow rate increases and as you sweat.
· Measure at room temperature (20 -25). The human body remains stable at room
temperature, but the body composition may change temporarily in cold or hot
conditions.
· Stand upright for about 5 minutes before testing. Taking the test
immediately after lying in bed or sitting for a long period of time might
result in a slight change in the test results as body water tends to move to
the lower body.
· During menstruation, there may be temporary fluctuations in fluid levels
that can affect the results.
6.2 Test Instructions
The screen displayed during the test varies depending on whether you use Self
Mode in Setup > 7. Test Mode & Normal Range or not.
· Professional Mode: The screen is configured for testing with the help of
experts.
· Self Mode: The screen is configured for testing alone.
1 Stand onto the footplate with bare feet on the standby screen.
· InBody starts weight measurement. · Please make sure that the subject stands
alone on the device. It may affect the weight measurement when other people
lean or touch the device.
19
InBody Test
2 Input the InBody User ID or height and press Next.
· Professional Mode: Input ID. · Self Mode: Input height, age, gender.
3 Maintain test posture.
· For proper test posture, refer to “6.3 Test Posture”. · The InBody test
automatically starts when the body
and the electrode are in precise contact.
Professional Mode
Self Mode
Professional Mode
Self Mode
20
4 The result screen is displayed after completing the test.
· The Result Sheet is printed out according to your setting when the printer
is connected.
· Please refer to “4.3 Overview of the Setup Menu” and “5.1 Printer” for
details of printer and Result Sheet setting.
Professional Mode
Self Mode Note Press the X on the upper-right to end the test.
InBody Test
21
InBody Test
6.3 Test Posture
The subject must maintain proper posture to have accurate test results. · The
InBody test automatically starts when the body and the electrode are in
precise contact.
Your arms must not touch the sides of your body.
The thighs should not touch each other.
Keep the arms straight.
Hold the hand electrode so that the 4 fingers wrap the surface of the bottom
hand electrode, and place the thumb on the oval electrode.
Step on the footplate barefoot.
Place the heels on the rear sole electrodes.
22
Maintenance and Storage
7.1 Precautions for Maintenance
Caution
· Do not bend the handle or turn it in the opposite direction.
· Do not place anything on the footplate when the device is not in use.
· Be careful not to get foreign materials on the bottom of the device. It may
cause weight measurement errors.
· Be careful not to get injured by getting your feet caught in the bottom of
the device.
· Turn off the device if you are not using it for a day or longer.
· If you are not using for a long time, unplug the power cord.
· Do not allow any liquid substances to contact the device directly. Keep food
and drinks away from the device. Substances getting inside the device can
cause critical damage to the electronic components.
7.4 Repacking and Transportation
Once the device is installed, avoid transporting it. If long distance
transportation is needed, repack the product before transporting.
Caution
When repacking the device, the protective packaging materials provided by
InBody must be used.
1 Turn off the device. 2 Remove all connected adapters, cables and the hinge
cover. Press both gaps of the hinge cover, and pull it out horizontally. · If
the stand is lowered with hinge cover attached, it
may cause scratches.
7.2 Cleaning
Use the alcohol-based disinfectant (e.g.70% ethanol) for 1 minute to clean the
surfaces of the device.
7.3 Disinfecting
1 Use the alcohol-based disinfectant (e.g.70% ethanol). 2 Follow the
instructions on the disinfectant. 3 Disinfect the device: Comply with the
intervals specified
in the below table.
Interval
Component
Before every measurement
After every measurement
Hand electrodes and Foot electrodes
Hand electrodes and Foot electrodes
3 Wrap the upper section of the device with protective packaging vinyl. Fold
down lower the upper part.
4 Place the device into the box. Put the packaging pad as shown in the figure
and seal it.
Caution
To transport the device safely, two people should keep it in a horizontal
position.
23
Maintenance and Storage
7.5 Storage Environment
The device should be stored under the following conditions.
Temperature range
10 – 70 (14 ~ 158°F)
Relative humidity
10 – 80 % RH (No Condensation)
Atmospheric pressure 50 – 106 kPa
24
Frequently Asked Questions
If a problem arises with the device, you may first attempt to check the Setup
8.FAQ / Customer Service Information > FAQ. If your problem cannot be resolved through the `FAQ’, please refer to the possible solutions below. If the problem is still not resolved, please contact InBody Customer Service.
8.1 Regarding the Device
Question: InBody Result is not sent to the App.
Answer:
· Check whether the Internet connection icon (Wi-Fi or LAN) and the cloud icon
on the upper-left of the standby screen are displayed.
· When the Internet connection icon is not displayed: Connect to the Internet
again in Setup > 3.Connect > Internet. For details on how to connect to the
Internet, refer to “5.8 Connecting Internet”.
· If the cloud icon is marked with an X : Check the settings in Setup >
2.Cloud Service.
Question: Power does not turn on.
Answer:
· Check if the power cord is correctly plugged into the socket. Make sure to
plug the power cord completely into the socket.
· When using a power surge protector, check the power switch on the power
surge protector is turned off.
· Check if the power cable is completely plugged into the adapter. Make sure
to insert the power cable completely into the adapter.
Answer:
· The problem may occur if you are using a power adapter that was not provided by InBody. Always connect a power adapter (DC 12V, 3.4 A/3.34A) provided by InBody.
Power Adapter
Power Cable
Question: The touch screen does not work well.
Answer:
· Calibrate the touchscreen under Setup > 1.General > 07. Touchscreen
Alignment.
· Press firmly to optimize touchscreen response.
8.2 Regarding of Serious Incidents
If you are aware of a serious incident involving your product, or communicate a corrective action to you clients, you must report this as quickly as possible to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.
The set deadlines in accordance with the MDR (EU) 2017/745 are:
Question: When an accident occurs
Answer:
· No later than within 15 calendar days after you have been informed of a
serious incident.
· No later than within 2 calendar days after you have been informed of a
serious incident which implies a serious threat to public health.
· No later than within 10 calendar days after you have been informed of a
serious incident which has led to a death, or a serious deterioration in
someone’s state of health.
You must report a serious incident before the corrective action to eliminate
the risk is taken, except in an emergency, in which case you must immediately
carry out a field safety corrective action.
8.3 Regarding the InBody Test
Question: Do I have to remove my socks or stockings?
Answer:
If the test is carried out while wearing socks or stockings, the current will not flow smoothly, and the test may not be performed correctly. The skin must be in direct contact with the electrode for testing.
Question: Answer:
Is it OK to test, wearing accessories or metallic materials?
If accessories or metallic objects do not touch the electrodes, they will not
have a significant effect on the test result. However, it is not recommended
to wear it for accurate test results.
25
Frequently Asked Questions (FAQ)
Question: Answer:
Is there any case where I must not take the InBody Test?
A person who is equipped with a medical device that is essential for life
support, such as pacemakers or patient monitoring devices, must not take the
InBody Test. Electronic medical devices may malfunction due to the current
flowing through the human body during the test.
Question: Answer:
Can a person with metal implants in the body take the InBody Test?
A person who has a metallic material inserted in the body may have different
conductivity that may affect the results of the test.
Question: Answer:
I have limited mobility and cannot maintain proper posture for the InBody
Test. How can I still be tested?
· The test is available, but the test result may be inaccurate due to poor
contacting to the electrode surface.
· InBody has a line of products that conduct body composition analysis in
lying posture that allow the patients to stay in bed. For more information,
please contact InBody.
Question: Answer:
Is the current flowing in the test harmless to the human body?
The InBody does not harm the human body because it uses micro alternative
electric current. The safety of the InBody has been proven and is being used
by many medical institutions because the InBody has already obtained the
national and European medical certificate.
Question: Answer:
How often should I take the InBody Test?
· Testing every other week or once a month can effectively track InBody Test
results for exercise prescriptions, hormonal prescriptions, obesity, and
rehabilitation.
· It’s good to check up often, but it’s also important to keep track of your
body’s changes over time through steady tests.
· Body composition changes can be seen on the InBody result screen and the
InBody Result Sheet.
Question: Answer:
What are the precautionary steps to ensure accuracy of the InBody Test?
Refer to “6.1 Precautions for Test”.
26
8.4 Residual Risks and Undesirable Side Effects
Undesirable side effects have been identified as general allergies that can be
associated with the skin contact of the metal surface during the clinical use
of the InBody580. Upon the comprehensive risk management, the metal patient
contacting material of the stainless steel has been evaluated with ISO-10993
biocompatibility testing, particularly with the skin sensitization testing,
which has resulted in the favorable biocompatibility test results. In
addition, the following contraindication statement has been added to this IFU:
Individuals with known metal allergies against stainless steel materials shall
not use the equipment.
Classifications and Specifications
· This device was produced according to InBody’s quality control procedures.
InBody complies with ISO9001 and ISO13485, international quality management
systems.
· This device satisfies IEC60601-1(EN60601-1), an international safety
standard for electronic medical equipment, and IEC60601-1-2(EN60601-1-2), an
international standard for electromagnetic compatibility.
9.1 Classifications
Classifications
Body Composition Analyzer of Direct Segmental Multi-frequency Bioelectrical Impedance Analysis Method
Type of protection against electric shock
Class I
Type of the applied parts BF Type
EMC Emission
CLASS B
Degree of protection
against water
IPX0
infiltration
Mode of operation
Continuous Operation
This equipment is not suitable for use in the presence of flammable anesthetics or oxygen.
9.2 Specifications
Bioelectrical Impedance Analysis (BIA) Measurement
Item
· 20 Impedance Measurements by Using 4 Different Frequencies (5 kHz, 50 kHz,
250 kHz, 500 kHz)at Each of 5 Segments (Right Arm, Left Arm, Trunk, Right Leg
and Left Leg)
· 5 Phase Angle Measurements by Using 1 Frequency (50kHz) at Each of 5
Segments (Right Arm, Left Arm, Trunk, Right Leg and Left Leg)
Electrode · Tetrapolar 8-Point Tactile Electrodes Method
Measurement Method
· Direct Segmental Multi-Frequency Bioelectrical Impedance Analysis (DSMBIA)
· Simultaneous Multi-Frequency Bioelectrical Impedance Analysis (SMFBIA)
Body composition
calculation Method
No Empirical Estimation
Outputs (InBody Result
Sheet)
· Body Composition Analysis (Total Body Water, Protein, Minerals, Body Fat
Mass, Soft Lean Mass, Fat Free Mass, Weight)
· Muscle·Fat Analysis (Weight, Skeletal Muscle Mass, Body Fat Mass)
· Obesity Analysis (Body Mass Index, Percent Body Fat)
· Segmental Lean Analysis · Segmental ECW Ratio – Phase Angle · ECW Ratio –
Phase Angle · Body Composition History (Weight,
Skeletal Muscle Mass, Soft Lean Mass, Body Fat Mass, Percent Body Fat, BMI,
ECW Ratio, InBody Score, Basal Metabolic Rate, Visceral Fat Level, Waist Hip
Ratio, Fat Free Mass, Waist Circumference, Obesity Degree, FFMI, FMI, SMI,
SMM/WT, Whole Body Phase Angle_50kHz) · InBody Score · Whole Body Phase Angle
(History) · SMI (History) · Visceral Fat Area (Graph) · Body Type (Graph) ·
Weight Control (Target Weight, Weight Control, Fat Control, Muscle Control) ·
Nutrition Evaluation (Protein, Minerals, Fat Mass) · Obesity Evaluation (BMI,
Percent Body Fat) · Body Balance Evaluation (Upper, Lower, Upper-Lower) ·
Segmental Fat Analysis (Right Arm, Left Arm, Trunk, Right Leg, Left Leg) ·
Segmental Fat Analysis (Graph) · Segmental Body Water Analysis · Segmental ICW
Analysis · Segmental ECW Analysis · Segmental Circumference (Neck, Chest,
Abdomen, Hip, Right Arm, Left Arm, Right Thigh, Left Thigh) · Waist-Hip Ratio
(Graph) · Visceral Fat Level (Graph) · InBody Score (Graph) · Research
Parameters (Intracellular Water, Extracellular Water, Skeletal Muscle Mass,
Fat Free Mass, Basal Metabolic Rate, Waist-Hip Ratio, Waist Circumference,
Visceral Fat Level, Visceral Fat Area, Obesity Degree, Bone Mineral Content,
Body Cell Mass, Arm Circumference, Arm Muscle Circumference, TBW/FFM, FFMI,
FMI, SMI, SMM/WT, Recommended Calorie Intake, Recommended Calorie Intakemanual
input) · Calorie Expenditure by Activity · Blood Pressure (Sys, Dia, Pulse) ·
Blood Pressure (MAP, PP, RPP) · QR Code · Results Interpretation QR Code ·
Whole Body Phase Angle (50kHz)
27
Classifiaaiiossaons Spiifiaaiios
Outputs (InBody Result
Sheet for Chidren)
· BIVA (Bioelectrical Impedance Vector Analysis)
· Impedance Graph (Each segment and each frequency)
· Body Composition Analysis (Total Body Water, Protein, Minerals, Body Fat
Mass, Weight)
· Muscle·Fat Analysis (Weight, Skeletal Muscle Mass, Body Fat Mass)
· Obesity Analysis (Body Mass Index, Percent Body Fat)
· Growth Curve Outputs (Height, Weight, BMI)
· Body Composition History (Height, Weight, BMI, Skeletal Muscle Mass, Soft
Lean Mass, Body Fat Mass, Percent Body Fat, Basal Metabolic Rate, Fat Free
Mass, Child Obesity Degree, FFMI, FMI, SMI, SMM/WT, Whole Body Phase
Angle_50kHz)
· Whole Body Phase Angle (History) · SMI (History) · Growth Score · Weight
Control (Target Weight, Weight
Control, Fat Control, Muscle Control) · Nutrition Evaluation (Protein,
Minerals,
Body Fat Mass) · Obesity Evaluation (BMI, Percent Body
Fat) · Body Balance Evaluation (Upper, Lower,
Upper-Lower) · Segmental Lean Analysis (Right Arm,
Left Arm, Trunk, Right Leg, Left Leg) · Segmental Body Water Analysis ·
Research Parameters (Intracellular
Water, Extracellular Water, Skeletal Muscle Mass, Fat Free Mass, Basal
Metabolic Rate, Child Obesity Degree, Bone Mineral Content, Body Cell Mass,
FFMI, FMI, SMI, SMM/WT) · Blood Pressure (Sys, Dia, Pulse) · Blood Pressure
(MAP, PP, RPP) · QR Code · Results Interpretation QR Code · Whole Body Phase
Angle (50kHz) · Segmental Phase Angle (50kHz: Right Arm, Left Arm, Trunk,
Right Leg, Left Leg) · Impedance Graph (Each segment and each frequency)
- Blood pressure information can only be printed when the blood pressure monitor is connected.
- “QR Code” is registered trademark of DENSO WAVE INCORPORATED.
Functional specifications
Compatible Device
BSM Series (BSM170B, BSM370, BSM270B), BPBIO Series (BPBIO320, BPBIO750), InBodyBAND Series and Serial Distributor (SD400)
Logo Display Name, Address and Content Information can be shown on the Results Sheet
Digital Results LCD Screen, LookinBody Web, LookinBody120
Types of InBody Result Sheet, InBody Result Sheet Result Sheets for Children
Voice Guidance
Notification sounds (test in progress, saving settings, personal information, etc.) and voice guidance during the test
Data Storage Saves up to 100,000 measurements (When ID is entered)
Backup Data Backup data saved in InBody380 by using an InBody USB
Test Mode Professional Mode / Self Mode
28
Other specifications
Applied Rating 200 A (±20 A) Current
Adapter (Bridge Power)
Power Input Power Output
AC 100-240 V, 50/60 Hz, 1.2 A
DC 12 V, 3.4 A
Adapter (Meanwell)
Power Input Power Output
AC 100-240 V, 50/60 Hz, 1.0-0.5 A
DC 12 V, 3.34 A
Display Type 1280 × 800 10.1inch Color TFT LCD
Internal Interface Keypad, Touch Screen
External Interface Serial(RS-232C): 2 EA, USB (HOST): 2 EA, LAN (10/100T): 1EA
Printer Laser/Inkjet PCL3 or above SPL
Dimension 501.9 (W) × 892.4 (L) × 1124.4 (H) mm 19.8 (W) × 35.1 (L) × 44.3 (H) in
Equipment 24.4 kg (53.8 lb) Weight
Test Duration About 30 seconds
Operation 10 – 40 (50 – 104 °F), 30 – 75%RH, Environment 70 – 106kPa (No Condensation)
Storage 10 – 70 (14 ~ 158°F), 10 – 80%RH, Environment 50 – 106kPa (No Condensation)
Measuring 2 ~ 300kg (4.4 – 661.4lb) Weight
Age Range 3+ years
Height Range 95 ~ 220cm (3ft 1.4in – 7ft 2.6in)
Classifcations and Specifcations
9.3 Symbols used on the Product
Indicators
9-pin Serial Port (Female, RS-232C)
LAN Port (10/100T Base)
USB HOST Port
Safety Symbols
Dangerous High Voltage
Direct Current
BF type equipment
12V , 3.4A / 3.34A
Power Adapter
Power On
Power Off
Etc. Symbols
Operating Instructions
Warning / Caution Do not disassemble the product arbitrarily. European
Conformity
Manufacturer
Medical Device
Unique Device Identification
Serial Number Authorized representative in the EUROPEAN COMMUNTY
29
Classifiaaiiossaons Spiifiaaiios
Operating instructions Catalogue number Importer Country of manufacture
Disposal of old Electrical & Electronic Equipment (Application in the European
Union and other European countries with separate collection system.) This
symbol indicates that this product shall not be treated as household waste.
Instead, it shall be handed over to the applicable collection point for the
recycling of electrical and electronic equipment. By ensuring this product is
disposed of correctly, you will help prevent potential negative consequences
for the environment and human health, which could otherwise be caused by
inappropriate waste handling of this product. For more detailed information
about recycling this product, please refer to local governing ordinances and
recycling plans.
30
Classifcations and Specifcations
9.4 Guidance and Manufacturer’s Declaration
The InBody device is intended for use in the electromagnetic environment
specified below. The customer or the user of the InBody device should ensure
that it is used in such an environment.
Electromagnetic emissions
Emissions test
Compliance Electromagnetic environment
RF emissions CISPR 11
Group 1
The InBody device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Class A Complies
The InBody device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic discharge IEC 61000-4-2
±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 % is recommended.
Electrical fast transient/ burst IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
0% U (100% dip in U) for 0.5/1 cycles 70% U (30% dip in U) for 25/30 cycles 0% U (100% dip in U) for 250/300 cycles
0% U (100% dip in U) for 0.5/1 cycles 70% U (30% dip in U) for 25/30 cycles 0% U (100% dip in U) for 250/300 cycles
Mains power quality should be that of a typical commercial or hospital environment. If the user of this product requires continued operation during power mains interruptions, it is recommended that this product be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels characteristic of a commercial or hospital environment.
31
Classifiaaiiossaons Spiifiaaiios
Recommended separation distances between portable and mobile communication
equipment and InBody device
The InBody device is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
InBody device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the InBody device as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter [m]
Rated maximum output power of transmitter [W] 0.01 0.1 1 10 100
150 kHz to 80 MHz d = 1.2P 0.12 0.38 1.2 3.8 12
IEC 60601-1-2: 2014 80 MHz to 2.7 GHz d = 2.0P 0.20 0.63 2.0 6.3 20
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
32
Classifcations and Specifcations
Electromagnetic immunity
Immunity test IEC 60601 test level
Compliance level
Conducted RF IEC 61000-4-6
3 V 150 kHz to 80 MHz
3 V
6 Vrms
150 kHz 80 MHz In ISM bands1 amateur radio
6 V
bands Bands2
10 V/m 80 MHz to 2.7 GHz
Radiated RF IEC 61000-4-3
10 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should not be used closer to
any part of the Ultrasound System, including cables, than the recommended
separation distance. This is calculated using the equation applicable to the
frequency of the transmitter.
Recommended separation distance d=1.2P
IEC 60601-1-2:2014 d=2.0 80 MHz to 2.7 GHz Where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m). Field
strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,3 should be less than the compliance level in each frequency range.4
Interference may occur in the vicinity of equipment marked with following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations, Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
- 1 The ISM (Industrial, Scientific and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
- 2 The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz,5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
- 3 Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixedRF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the InBody device is used exceeds the applicable RF compliance level above,the InBody device should be observed to verify normal operation. If abnormal performance is observed,additional measures maybe necessary, such as re-orienting or relocating the InBody device.
- 4 When the frequency range exceeds 150 kHz 80 MHz, the electric field strength should be not higher than 3 V/m.
33
Classifiaaiiossaons Spiifiaaiios
Electromagnetic emissions
The InBody device is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. Portable RF communications
equipment should be used no closer than 30 cm (12 inches) to any part of the
InBody device. Otherwise, the performance of this equipment could be impaired.
Immunity test
Band
Service5
Modulation6
IEC60601 test level
Compliance level
380 – 390 MHz
TETRA 400
Pulse modulation 18Hz
27 V/m
27 V/m
430 – 470 MHz
GMRS 460 FRS 460
FM7
±5 kHz deviation
1 kHz sine
28 V/m
28V/m
704 – 787 MHz
LTE Band13, 17
Pulse modulation 217 Hz
9 V/m
9 V/m
Proximity fields from RF wireless Communications IEC61000-4-3
800 – 960 MHz
1700 – 1990 MHz
GSM800:900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5
GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1,2,4,25 UMTS
Pulse modulation 18 Hz
28 V/m
28V/m
Pulse modulation 217 Hz
28 V/m
28V/m
2400 – 2570 MHz
Bluetooth WLAN 802.11b/g/n RFID 2450 LTE Band
Pulse modulation 217 Hz
28V/m
28V/m
5100 – 5800 MHz
WLAN 802.11a/n
Pulse modulation 217 Hz
9 V/m
9 V/m
NOTE
If it is necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The 1m test distance is permitted by IEC 61000-4-3.
- 5 For some services, only the uplink frequencies are included. * 6 The carrier shall be modulated using a 50% duty cycle square wave signal.
- 7 As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be the worst case.
34
Classifcations and Specifcations
Follow instructions for use
Suivez les instructions d’utilisation
WARNING
· Electric shock hazard do not dismantle. · Dismantling will void the
warranty. · Do not touch signal input, signal output or other
connectors, and the patient simultaneously. · External equipment intended for
connection to signal
input, signal output or other connectors, shall comply with relevant IEC
Standard(e.g., IEC60950 for IT equipment and IEC60601-1 series for medical
electrical equipment). In addition, all such combination-system-shall comply
with the standard IEC60601-1 and/or IEC60601-1-1 harmonized national standard
or the combination. If, in doubt, contact qualified technician or your local
representative. · Do not position so that it is difficult to operate the
disconnection device.
AVERTISSEMENT
· Risque de choc électrique – ne pas démonter. · Le démontage annulera la
garantie. · Ne touchez pas l’entrée de signal, la sortie de signal ou
d’autres connecteurs et le patient simultanément. · L’équipement externe
destiné à être connecté à l’entrée
de signal, à la sortie de signal ou à d’autres connecteurs doit être conforme
à la norme IEC pertinente (par exemple, IEC60950 pour les équipements
informatiques et la série IEC60601-1 pour les équipements électriques
médicaux). De plus, tous ces systèmes combinés doivent être conformes à la
norme nationale harmonisée IEC60601-1 et/ou IEC60601-1-1 ou à la combinaison.
En cas de doute, contactez un technicien qualifié ou votre représentant local.
· Ne pas positionner de telle sorte qu’il soit difficile d’actionner le
dispositif de déconnexion.
DANGER
· Do not use this equipment with electrical medical device such as a
pacemaker.
· Ne pas utiliser cet équipement avec des appareils médicaux électriques comme
un stimulateur cardiaque.
CAUTION
· Do not spray any liquid substance directly onto the device. · Ne pulverisez
aucune substances liquids directement sur
l’appareil.
CAUTION
· No excessive force on shoulder joint. · Ne pas appliquer de force excessive
sur les bars articulés.
9.5 Key Performance Claims of InBody580
The key performance claims of InBody580 has been established as the
correlation coefficient ratio (R) of Fat Free Mass (FFM), which is numerically
defined as the R value shall be 0.80 (80%)
Clinical Benefit
Using the InBody580 with the probability of harm occurring is more beneficial
when compared to the severe harm that might occur from not using the Body
Composition Analyzer of InBody (Models: InBody580). The Body Composition
Analyzer of InBody (Models: InBody580) provides clinical benefits to support
the aforementioned intended use, as the of InBody (Models: InBody580) in
mainly used for healthy and acute or chronically ill populations in hospitals,
medical practices and inpatient care facilities in accordance with national
regulations. It can be used to assist in the assessment of nutritional status,
obesity and muscle balance. Body composition analysis is important in
preventive medicine since it provides the basis of appropriate physical
activity and dietary habits for improving personal daily routine. It can be
also usefully applied to follow-up studies of patients treated for various
diseases.
The key performance claims of InBody580 have been established as the
correlation coefficient ratio (R) of Fat Free Mass (FFM), which is numerically
defined as the R value shall be 0.80 (80%). Inaccurate measurements of the Fat
Free Mass (FFM) could have a negative impact on further use of the body
composition analysis data gathered from the clinical use of InBody580.
35
© 2024 InBody Co., Ltd. All rights reserved. IM-ENG-R6-B-240111
References
- InBody - Oceania | InBody Oceania
- 【公式】体成分分析装置InBody(インボディ)
- InBody
- Body Composition Analyzer Machine for Fitness & Medical Industry
- InBody BWA for Medical Professionals – Monitor body water changes and track clinical outcomes for medical professionals
- InBody – Descubre de qué estás hecho
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>