welldana Check Kit Moicro Salt Instruction Manual

July 12, 2024
welldana

welldana Check Kit Moicro Salt

Product Usage Instructions

  • Measurement of the Parameters:
    • Carefully pull off the protective cap and rinse the electrode.
    • Select conductivity mode (COND).
    • Immerse the electrode in the recommended calibration solution.
    • Press and hold for 3 seconds until the screen turns green (Calibration mode).
    • Stir the tester briefly and wait for the calibration value to stabilize.
    • Press and hold for 1 second to start calibration.
  • Changing Between Measurement Modes:
    • To change between conductivity (COND), TDS, and salinity (SAL) modes for new measurements, follow these steps:
    • Press the designated key to switch between modes.
    • Ensure the measuring range and resolution settings are appropriate for the desired mode.
  • Replacing Batteries:
    • To replace the batteries, follow these steps:
    • Open the battery cover and remove the old batteries.
    • Insert four AAA batteries, ensuring the positive terminals face up.
    • Close the battery cover securely to maintain water tightness.

FAQs

  • Error Messages and Resolutions:
    • Err 1: Reason: Old calibration. Procedure: Use new calibration solutions.
    • Err 2: Reason: Key pressed too early. Procedure: Wait for the appropriate time before pressing the key.
    • Contamination/Error Fluctuations: Procedure: Clean the electrode with detergent, and rinse with deionized water (avoid rubbing black pins).

Display

Keyboard function

  • Short keystroke (1 sec.)
  • Long keystroke (3 sec.)

welldana-Check-Kit-Moicro-Salt-FIG-1 \(2\)

Settings

  • In off mode, press for 3 seconds: Opening Settings.
  • Navigate
  • Confirm

Blue marked = factory setting

  Parameter Selec tion Meaning
P1 Auto-Hold a) On

b)  Off

| when: Measured value min. 10 s stable Auto-Hold switched off
P2| Backlight| a) On

b)  1 min

c) Off

| Switched on continuously Switched on for 1 minute Continuously switched off
P3| Temperature compensation coefficient| 0 – 4%

2%

| Setting the linear temperature compensation
P4| TDS conversion factor| 0.4 – 1

0.71

| Setting of the TDS conversion factor
P5| Salinity Unit| a) ppt

b)  g/L

| Display in ppt Display in g/L
P6| Temperature unit| a) °C

b)  °F

| Display in °C Display in °F
P7| Factory calibration| a) Yes

b)  No

| Calibration is reset

Measurement of the parameters

welldana-Check-Kit-Moicro-Salt-FIG-1 \(6\)

Note:

  • Change between conductivity (COND), TDS, and salinity (SAL)
  • Press “Hold“ for new measurement recording.

Calibration conductivity

Note:

  • Recommendation: 1-point calibration depending on the measuring range.
Measuring range Calibration solution
0 -100 µS/cm 84 µS/cm
100 – 2000 µS/cm 1413 µS/cm
2 – 20 mS/cm 12,88 mS/cm

After successful 1, 2 or 3-point calibration, is shown on the display:

L 84 µS/cm
M 1413 µS/cm
H 12,88mS/cm

Note: 1000 μS/cm = 1 mS/cm

Specifications

SD 30 Con – TDS – Salt/T Specifications

Conductivity (mS/ cm)| Measuring range: 0.01-20; resolution: 0.01;

accuracy: ±1% F.S.

---|---
TDS (ppt)| Measuring range: 0.01-20; resolution: 0.01
Salinity (ppt)| Measuring range: 0.01-10; resolution: 0.01
Temperature (°C)| Measuring range: 0-50; resolution: 0.1; Accuracy: ±0.5
Background light| Blue : Measurement; Green : Calibration
HOLD| The measured value is frozen
| Flashing, and battery change is required
Auto. switch off| after 8 minutes without operation
Housing protection type| IP67
Batteries| 4 x AAA Batteries

Exchange Batteries

  • Caution: Insert all batteries with the positive terminal facing up. Incorrect insertion of the batteries could result in damage to the meter.
  • Note: Make sure that the battery cover is completely closed to ensure water tightness.

Error messages and procedure

  Reason Procedure
Err 1 old calibration Use of new calibration solutions
Err 2 -Key pressed too early Wait for then press key.
Measured value fluctuates Contamination Cleaning the electrode with

detergent Rinse the electrode with deionized Water
Important: Do not rub the black pins as the platinum coating may be damaged.

Accessories

  • Art.-No. Description
  • 726654 Conductivity solution 1413 μS/cm (NIST), 90 ml
  • 726684 Conductivity solution 12.88 mS/cm (NIST), 90 ml
  • 720035 SD30 Con/TDS/Salt/T replacement electrode
  • 1950026 AAA batteries, set of 4
  • 384801 Measuring beaker 100 ml
  • 457022 Deionised water for rinsing

Warranty

  • The warranty period for the SD30 is 2 years from the date of purchase and 6 months for the electrode. Excluded from the warranty are damages caused by improper handling or mechanical damage.
  • Complete instructions with an accessories list can be found at www.lovibond.com (Download Centre).

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals