FUTEK IAC200 1-4 Inch Channel Summing Junction Box User Guide

July 3, 2024
FUTEK

FUTEK IAC200 1-4 Inch Channel Summing Junction Box

Specifications

  • Model: IAC200
  • Manufacturer: FUTEK
  • Output: mV/V
  • Capacity: 1779.2 N
  • Website:www.futek.com

Product Usage Instructions

Setup Guide

  • The setup steps involve placing a known load on each sensor and adjusting the output so each sensor has the same output under the same load.
  • The final summed output will be a combined signal of all connected sensors with the electrical output reaching its full capacity when all sensors are fully loaded.

Configuration

  1. Remove the bottom cover of the IAC200 to access the board.
  2. If the amplified version, disconnect the ribbon cable from the center amplifier to the IAC200 board on connection J6.
  3. Enable one sensor channel at a time by setting both DIP switches to ON for the channel being adjusted.
  4. Set the output of the IAC200 to S (mV/V) for direct sensor passthrough.

Wiring

  • Connections to the IAC200 board from sensors and output are routed through black pass-through connections. Ensure proper connections are made as per the setup guide.

Measurement

  • Utilize multiple high-capacity sensors in conjunction with powerful instrumentation for accurate measurement. Ensure each sensor is adjusted as per the setup guide for consistent outputs.

Adjustment

  • Adjust each sensor to have the same output when a known load is applied. Follow the setup guide for precise adjustments.

Setup Finalization

  • After adjusting all sensors, ensure that the final summed output matches the expected values as per the setup guide. Reconfigure the signal select if using an amplified version.

FAQ

Q: Where can I find technical support for the IAC200?

A: For technical support and more information, please visit here.

Q: How can I connect my IAC200 to a computer or instrument for readout?

A: Solutions are available at FUTEK to connect your IAC200 to a computer or instrument. View FUTEK instruments here.

Getting Help TECHNICAL SUPPORT

For more IAC200 support, please visit: https://www.futek.com/summing- junction-box

Setup Guide

  • The setup steps on the following pages involve placing a known load on each sensor and adjusting the output down so each sensor has the same output under the same load.
  • The result of the setup will be a summed signal of all connected sensors with the electrical output at the lowest sensor adjusted down to and will reach its full electrical output change when all sensors are fully loaded.
  • For example, four 444.8 N sensors are adjusted to have the same 1.98 mV/V when 444.8 N is applied. The final summed output will be 1.98 mV/V at 1779.2 N (444.8 N on each sensor).
  • For amplified versions, voltage, or current, an extra step will be needed to set up the amplifier using the final summed signal that will go into the amplifier.
  • In the above example, it would mean setting up an amplifier to a sensor with 1.98 mV/V output at its fully loaded 1779.2 N capacity.
  • Measuring the contents of any industrial tank, silo, or hopper requires a robust and precise system. Utilizing multiple high-capacity sensors in conjunction with powerful instrumentation can make for an effective platform.

SETUP TIPS

  • A system calibration at FUTEK will help save time, confusion, and potential damage by providing a plug-and-play experience with certification.
  • Standard NIST and A2LA calibration requests are started online at: https://www.futek.com/recalibrationterms
  • Solutions are available at FUTEK to connect your IAC200 to the computer or an instrument with a readout. View FUTEK instruments at: https://www.futek.com/store/instruments
  • The spec sheet for the IAC200 contains the wiring, DIP switch locations, and power level information.
  • Measuring the contents of any industrial tank, silo, or hopper requires a robust and precise system.
  • Utilizing multiple high-capacity sensors in conjunction with powerful instrumentation can make for an effective platform.

Configuration

  • Remove the bottom cover of the IAC200 to gain access to the IAC200 board. Connections to the IAC200 board from sensors and IAC200 output are routed to the board through black passthrough connections.
  • A) If present, for amplified IAC200, disconnect the ribbon cable from the center amplifier to the IAC200 board on connection J6. (Applies to FSH04736 Voltage output and FSH04737 Current output versions.)
  • B) Enable one sensor channel at a time by placing both DIP switches in the ON position for the channel and the rest in the OFF position. As each sensor is adjusted its channel will need to be turned on and the prior channel turned off.
  • C) Set the output of the IAC200 for S (mV/V) for direct sensor passthrough. It will later be necessary to reconfigure the signal select to V (VDC) for the FSH04736 amplified voltage version and I (mA) for the FSH04737 amplified current version.

Wiring

  • A) Connect each sensor to the IAC200. +/- Excitation will be power to the sensor and +/- Signal will be the output from each sensor to the IAC200. Refer to the sensor spec sheet online for the sensor’s wiring.
  • B) Connect the power supply to IAC200 for sensor excitation. Refer to the sensor spec sheet online for maximum excitation.
  • C) Connect the Voltage meter to the IAC200 output. Direct sensor output will be in mV.

Measurement

  • A) Place a known load on the first sensor. The load does not have to be the full capacity of the sensor.
  • B) Record mV output on Voltmeter. Full capacity mV output level is a product of the sensor mV/V rating and excitation.
  • C) Adjust Channel Select for next channel ON and switch OFF prior channel.
  • D) Continue to process until all sensor individual outputs are known under the same load.

Adjustment

  • A) Observe the lowest recorded output from connected sensors.
  • B) Set Channel Select DIP switches to the first Channel ON and all other Channels OFF.
  • C) Apply known load to the first sensor.
  • D) Adjust the SPAN potentiometer to adjust sensor mV output down, under load, while observing the sensor output until output is equal to the lowest recorded output. Span potentiometer can only adjust outputs down.
  • E) Continue the process of Channel Select DIP Switches, applied known load, and adjustment of mV output down on each sensor until each sensor has the same mV output under the sample applied load. To ensure best performance re-check the output of each sensor while under load and re-adjust if necessary.

Setup Finalization

  • A) Set Channel Select DIP switches ON for all Channels with a connected sensor. The output of IAC200 will now be one signal of all connected sensors into one output with the same mV output level that each sensor has been adjusted down to. The combined output will now reach the set mV output when all sensors are fully loaded.
  • B) If present, re-connect the ribbon cable from the center amplifier to the IAC200 board on connection J6.
  • C) Set the output of the IAC200 to S (mV/V) for non-amplified IAC200, V (VDC) for voltage amplified IAC200 and I (mA) for current amplified IAC200.
  • FSH04735 is non-amplified S (mV/V), FSH04736 is V (VDC), and FSH04737 is the I (mA) amplified current version.

D) Additional amplified setup:

  • The combined one mV output from all sensors will be the input to the amplifier.
  • To assist with amplifier setup instructions, convert the adjusted one mV output of all connected sensors into mV/V by dividing the mV output by the supply voltage used during the setting of the one mV output.
  • This will be the one mV/V combined signal going into the amplifier.
  • Refer to the FUTEK online support page of the IAA amplifier for further instructions on how to set up the attached amplifier.
  • A Video of the IAA amplifier series is available on the FUTEK website.
  • IAA100 for Amplified IAC200 Voltage V(VDC) FSH04736
  • IAA200 for Amplified IAC200 Current I(mA) FSH04737
  • For legacy IAC200 see the FUTEK online CSG110 amplifier webpage for setup support.
  • Sensor Solution Source
  • Load  Torque Pressure Multi-Axis  Calibration Instruments  Softwarewww.futek.com
  • Drawing Number: EM1067-B
  • FUTEK reserves the right to modify its design and specifications without notice. Please visithttps://www.futek.com/TermsAndConditions for complete terms and conditions.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals