ARROW DLX-35541-VPSC Pressure Injectable One Lumen PICC pre Loaded Instructions
- July 1, 2024
- Arrow
Table of Contents
ARROW DLX-35541-VPSC Pressure Injectable One Lumen PICC pre-loaded
Product Specifications
- Model: PB-10000-850
- Size: 8.5 X 10.5
- DLX Model: DLX-35541-VPSC (IPNXXXXXX)
- Catheter Length: 55 cm
- Guide Diameter: .018 inch
- Trademark: GlideThruTM Peel-Away Sheath, Nitinol Spring-Wire Guide
Product Usage Instructions
Pressure Injectable One-Lumen PICC pre-loaded with Arrow® VPS Precision™ Stylet
Contents:
- 1: One Lumen TaperFree® Catheter‡‡‡: 4 Fr. (1.45 mm OD) x 55 cm, Pressure Injectable, Contamination Guard, Arrow® VPS Precision™ Stylet and Tuohy- Borst Adapter
- 1: GlideThru™ Peel-Away Sheath: 4 Fr. x 2-3/4″ (7 cm) Radiopaque over 4 Fr. Dilator
- 1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow Advancer
- 1: Stiffening Stylet (0.36mm dia. x 75cm with straight distal tip)
- 1: Injection Needle: SafetyGlide™1 25 Ga. x 5/8″ (1.60 cm)
- 1: Safety Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
- 1: Blunt Fill Needle: 18 Ga. x 1-1/2″ (3.81 cm)
- 1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
- 1: Syringe: 5 mL Luer-Lock
- 4: 10 mL Luer-Lock Pre-Filled Saline Syringe
- 1: 5 mL 1% Lidocaine HCl Solution and Alcohol Prep
- 1: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution
- 2: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution with Hi-Lite Orange™ Tint
- 1: Dust Cap: Non-Vented
- 1: SharpsAway® II Locking Disposal Cup
- 1: SharpsAway® Disposal Cup
- 1: Catheter Trimmer
- 1: Towel: 24″ x 36″ (61 cm x 91 cm)
- 1: Maximal Barrier Drape™ with two 13 cm (5″) x 15 cm (6″) fenestrations
- 1: Safety Scalpel: #11
- 1: Sterile Procedure Sign
- 1: Patient ID Card
- 1: Patient Information Booklet
- 1: Medication Label:
- 1% Lidocaine
- 2: Paper Tape Measure
- 1: Tourniquet
- 1: Foam Electrodes (3 per pack)
- 10: Gauze Pad: 2″ x 2″ (5 cm x 5 cm)
- 10: Gauze Pad: 4″ x 4″ (10 cm x 10 cm)
- 1: Surgical Apparel: Impervious Gown
- 1: Dressing: Tegaderm®3 CHG 3-1/2″ x 4-1/2″ (8.5 cm x 11.5 cm)
- 1: Transducer Cover with Gel and Rubber Bands
- 1: Surgical Apparel: Mask with Eye Shield
- 1: Dressing: STATLOCK®4 PICC Plus Catheter Stabilization Device
- 1: Surgical Apparel: Bouffant Cap
- 1: Surgical Apparel: Mask
- 1: Tape: Steri-Strip®3
- 1A trademark of Becton, Dickinson and Company.
- 2A registered trademark of CareFusion or one of its subsidiaries.
- 3A registered trademark of 3M Company.
- 4A registered trademark of C. R. Bard, Inc
Preparation:
- Ensure the product is sterile and undamaged.
- Verify that the patient has no contraindications as mentioned in the manual.
Insertion Process:
- Insert the GlideThruTM Peel-Away Sheath followed by the Spring-Wire Guide into the intended vessel.
- Use the guide with the Arrow Advancer for accurate placement.
Priming and Flow Rates:
- Priming Volume: 0.63 mL
- Gravity Flow Rate: 970 mL/hr
- Pump Flow Rate: 5300 mL/hr
- Maximum Pressure Injection Flow Rate: 5 mL/sec
Cleaning and Storage:
- After use, clean the product as per recommended guidelines.
- Store in a clean and dry environment away from direct sunlight.
ǂǂǂMR Safe
Not made with natural rubber latex.
Store between 20 – 25°C (68 – 77°F).
Arrow International provides the enclosed medication label(s) for your
convenience. Please ensure that these labels are applied to the correct
syringe and corresponding medication.
Contraindications: The Pressure Injectable PICC is contraindicated wherever there is the presence of device-related infections or the presence of thrombosis in the intended insertion vessel or catheter pathway. Clinical assessment of the patient must be completed to ensure no contraindications exist.
- Priming volumes are approximate and are done without accessories.
- Flow rate values are approximate and are determined using deionized water at 100 cm head height.
- Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities.
- Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
Frequently Asked Questions (FAQ):
Q: What are the contraindications for using the product?
A: The product is contraindicated in cases of device-related infections or
thrombosis in the insertion vessel or catheter pathway.
Q: What are the flow rate values based on?
A: Flow rate values are approximate and determined using deionized water at
specified conditions as mentioned in the manual.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>