ARROW DLX-35541-VPSC Pressure Injectable One Lumen PICC pre Loaded Instructions

July 1, 2024
Arrow

ARROW DLX-35541-VPSC Pressure Injectable One Lumen PICC pre-loaded

Product Specifications

  • Model: PB-10000-850
  • Size: 8.5 X 10.5
  • DLX Model: DLX-35541-VPSC (IPNXXXXXX)
  • Catheter Length: 55 cm
  • Guide Diameter: .018 inch
  • Trademark: GlideThruTM Peel-Away Sheath, Nitinol Spring-Wire Guide

Product Usage Instructions

Pressure Injectable One-Lumen PICC pre-loaded with Arrow® VPS Precision™ Stylet

Contents:

  • 1: One Lumen TaperFree® Catheter‡‡‡: 4 Fr. (1.45 mm OD) x 55 cm, Pressure Injectable, Contamination Guard, Arrow® VPS Precision™ Stylet and Tuohy- Borst Adapter
  • 1: GlideThru™ Peel-Away Sheath: 4 Fr. x 2-3/4″ (7 cm) Radiopaque over 4 Fr. Dilator
  • 1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow Advancer
  • 1: Stiffening Stylet (0.36mm dia. x 75cm with straight distal tip)
  • 1: Injection Needle: SafetyGlide™1 25 Ga. x 5/8″ (1.60 cm)
  • 1: Safety Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
  • 1: Blunt Fill Needle: 18 Ga. x 1-1/2″ (3.81 cm)
  • 1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
  • 1: Syringe: 5 mL Luer-Lock
  • 4: 10 mL Luer-Lock Pre-Filled Saline Syringe
  • 1: 5 mL 1% Lidocaine HCl Solution and Alcohol Prep
  • 1: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution
  • 2: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution with Hi-Lite Orange™ Tint
  • 1: Dust Cap: Non-Vented
  • 1: SharpsAway® II Locking Disposal Cup
  • 1: SharpsAway® Disposal Cup
  • 1: Catheter Trimmer
  • 1: Towel: 24″ x 36″ (61 cm x 91 cm)
  • 1: Maximal Barrier Drape™ with two 13 cm (5″) x 15 cm (6″) fenestrations
  • 1: Safety Scalpel: #11
  • 1: Sterile Procedure Sign
  • 1: Patient ID Card
  • 1: Patient Information Booklet
  • 1: Medication Label:
  • 1% Lidocaine
  • 2: Paper Tape Measure
  • 1: Tourniquet
  • 1: Foam Electrodes (3 per pack)
  • 10: Gauze Pad: 2″ x 2″ (5 cm x 5 cm)
  • 10: Gauze Pad: 4″ x 4″ (10 cm x 10 cm)
  • 1: Surgical Apparel: Impervious Gown
  • 1: Dressing: Tegaderm®3 CHG 3-1/2″ x 4-1/2″ (8.5 cm x 11.5 cm)
  • 1: Transducer Cover with Gel and Rubber Bands
  • 1: Surgical Apparel: Mask with Eye Shield
  • 1: Dressing: STATLOCK®4 PICC Plus Catheter Stabilization Device
  • 1: Surgical Apparel: Bouffant Cap
  • 1: Surgical Apparel: Mask
  • 1: Tape: Steri-Strip®3
    • 1A trademark of Becton, Dickinson and Company.
    • 2A registered trademark of CareFusion or one of its subsidiaries.
    • 3A registered trademark of 3M Company.
    • 4A registered trademark of C. R. Bard, Inc

Preparation:

  1. Ensure the product is sterile and undamaged.
  2. Verify that the patient has no contraindications as mentioned in the manual.

Insertion Process:

  1. Insert the GlideThruTM Peel-Away Sheath followed by the Spring-Wire Guide into the intended vessel.
  2. Use the guide with the Arrow Advancer for accurate placement.

Priming and Flow Rates:

  1. Priming Volume: 0.63 mL
  2. Gravity Flow Rate: 970 mL/hr
  3. Pump Flow Rate: 5300 mL/hr
  4. Maximum Pressure Injection Flow Rate: 5 mL/sec

Cleaning and Storage:

  1. After use, clean the product as per recommended guidelines.
  2. Store in a clean and dry environment away from direct sunlight.

ǂǂǂMR Safe
Not made with natural rubber latex.

Store between 20 – 25°C (68 – 77°F).
Arrow International provides the enclosed medication label(s) for your convenience. Please ensure that these labels are applied to the correct syringe and corresponding medication.

Contraindications: The Pressure Injectable PICC is contraindicated wherever there is the presence of device-related infections or the presence of thrombosis in the intended insertion vessel or catheter pathway. Clinical assessment of the patient must be completed to ensure no contraindications exist.

ARROW-DLX-35541-VPSC-Pressure-Injectable-One-Lumen-PICC-pre-Loaded-fig-
\(2\)

  • Priming volumes are approximate and are done without accessories.
  • Flow rate values are approximate and are determined using deionized water at 100 cm head height.
  • Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities.
  • Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.

Frequently Asked Questions (FAQ):

Q: What are the contraindications for using the product?
A: The product is contraindicated in cases of device-related infections or thrombosis in the insertion vessel or catheter pathway.

Q: What are the flow rate values based on?
A: Flow rate values are approximate and determined using deionized water at specified conditions as mentioned in the manual.