NONIN 6000CI Disposable Pulse Oximeter Sensors Instruction Manual
- June 25, 2024
- NONiN
Table of Contents
NONIN-6000CI-Disposable-Pulse-Oximeter-Sensors
Specifications
- Model: 6000CI/7000I
- Type: Infant Disposable Single-Patient Use Pulse Oximeter Sensor
- Indicated Use: Non-invasive spot-checking and/or continuous monitoring of infant patients weighing greater than 4 pounds (2 kilograms)
- Intended Environments: Operating room, surgical recovery, critical care, emergency room, long-term care, home use, mobile environments
- Regulatory Compliance: Federal law (USA) restricts sale to licensed practitioners
Product Usage Instructions:
Cautions
- Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Indications for Use:
- Suitable for non-invasive spot-checking and continuous monitoring of infant patients with good or poor perfusion, weighing over 4 pounds (2 kilograms).
- Intended for use in various environments including operating rooms, critical care units, emergency rooms, and home settings.
Model 6000CI/7000I
Infant Disposable Single-Patient Use Pulse Oximeter Sensor
Cautions:
Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Indications for Use
Nonin’s Models 6000CI and 7000I Single-Patient Use Disposable Pulse Oximeter Sensors are indicated for non-invasive spot-checking and/or continuous monitoring of infant patients who are well or poorly perfused, weighing greater than 4 pounds (2 kilograms). It is intended for use in environments including operating room, surgical recovery, critical care, emergency room, long-term care, home use and mobile environments.
Contraindications
- Do not use the device in an MR environment or in an explosive atmosphere.
- This device is not defibrillation proof per IEC 60601-1:1990 clause 17h.
Warnings :
- The use of sensor and oximeter combinations other than Noninbranded products have not been tested for accuracy as a system and may affect performance of the system. Refer to Nonin pulse oximeter operator’s manuals for a complete listing of Nonin-branded oximeters, sensors, and accessories.
- Inspect the sensor application site at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity to sensors may vary due to medical status or skin condition.
- Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
Cautions:
- Do not use a damaged sensor. If the sensor is damaged, discontinue use immediately.
- Do not sterilize, autoclave or immerse in liquid of any kind.
- Do not use caustic or abrasive cleaning agents on the sensor.
- Follow local governing ordinances and recycling instructions regarding disposal or recycling of the sensor and any components.
- A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
- As with all medical equipment, carefully route patient cables and connections to reduce the possibility of entanglement or strangulation.
- Refer to the pulse oximeter operator’s manual for additional warnings and cautions.
- Factors that may degrade pulse oximeter performance include the following:
- excessive ambient light
- excessive motion
- electrosurgical interference
- moisture in the sensor
- improperly applied sensor
- carboxyhemoglobin
- methemoglobin
- blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.)
- incorrect sensor type
- poor pulse quality
- venous pulsations
- anemia or low hemoglobin concentrations
- cardiovascular dyes
- dysfunctional hemoglobin
- artificial nails
- fingernail polish
Emitter
Detector
Center Line
Symbols
Choosing the Sensor Application Site
The preferred application site for infants is the large toe. Other sites may
not give acceptable results because of inadequate light transmission or
perfusion.
Attaching the Infant Disposable Sensor:
- Carefully peel away and discard the adhesive backing (figure A).
- Place the sensor light detector on the bottom of the large toe as shown in figure B.
- Wrap the tape around the toe (figures C, D, and E). When wrapping the tape, ensure the light detector is centered on the bottom of the toe and the light emitter is on the toe nail, directly opposite the light detector. The center line may be used as an alignment guide and should be on the side of the toe.
- For best results, use medical tape to secure the sensor cable independently from the sensor (figure E). Ensure the tape securing the cable does not restrict blood flow.
Note: Proper sensor placement is critical for good performance. If the sensor is not positioned properly, light may bypass the tissue and result in SpO2 inaccuracies.
Specifications
-
SpO2 Accuracy: 70 % to 100 % ±2 digits (Arms*)1, 2, 3
-
SpO2 Low Perfusion Accuracy: 70 % to 100 % ±2 digits (Arms*)3
-
Pulse Rate Accuracy: 18 to 321 BPM ±2 digits (Arms*)3
-
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±2 digits (Arms*)3
-
±1 Arms encompasses 68 % of the population.
- SpO2 accuracy was conducted during induced hypoxia study on healthy subjects over the range of 70 % to 100 %.
- Accuracy specifications based on testing with Models 2500/2500A and 9600.
- Additional accuracy and performance information can be found in the pulse oximeter operator’s manual.
Measurement Wavelengths and Output Power**
- Red: 660 nanometers @ 3 mW nominal
- Infrared: 910 nanometers @ 3 mW nominal
- **This information is especially useful for clinicians.
Compliance
This product complies with ISO 10993-1.
Not made with natural rubber latex.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc. and the competent authority of the EU Member State in which the user and/or patient is established, if applicable.
- Nonin Medical, Inc. Nonin Medical, Inc.
- 13700 1st Avenue North
- Plymouth, MN 55441, USA
- www.nonin.com
- E-mail: info@nonin.com
- +1 763-553-9968
- 800-356-8874 (US and Canada)
- Nonin Medical B.V.
- Doctor Paul Janssenweg 150
- 5026 RH Tilburg, Netherlands
- Email: infointl@nonin.com (Europe)
- +31 (0)13 – 45 87 130 (Europe)
- MPS, Medical Product Service GmbH
- Bomgasse 20
- D-35619 Braunfels, Gemany
- MedEnvoy Switzerland
- Gotthardstrasse 28, 6302 Zug
- Switzerland
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>