NONIN 7500 Fibre Optic Tabletop Pulse Oximeter User Guide
- June 9, 2024
- NONiN
Table of Contents
Model 7500
Quick Start-up Guide
SET USER-DEFINED DEFAULTS
- Turn unit on – Review factory settings below
- Step through ALL settings using LIMITS button
- Change desired settings using or buttons
- MUST push LIMITS button after each setting to confirm & move on
- Scroll to end of settings to 2nd symbol ( =minutes) and press LIMITS button
- Unit will automatically review new settings
- TO SAVE SETTINGS – press and hold ALARM SILENCE button while briefly pressing LIMITS button – “dEF On” will flash
Return to Factory Defaults
Note: User-Defined Default values are lost when returned to Factory
Defaults.
- Turn unit on – “dEF On” will appear
- Press and hold ALARM SILENCE button and briefly press button – “dEF OFF” will flash
- Press POWER button until unit turns OFF, then turn back ON
Factory Default Settings
The values at right come pre-programmed in the Model 7500. The Patient
Security Mode is OFF, unless turned on.
KEY| Alarm Limit| Factory Setting| Adjustment Options|
Increment
---|---|---|---|---
| Sp02 Low Alarm Limit| 85%| Off, 50-95| 1%
| Pulse Rate High Alarm Limit| 200 BPM| Off, 75-275| 5 BPM
| Pulse Rate Low Alarm Limit| 50 BPM| Off, 30-110| 5 BPM
| Sp02 High Alarm Limit| Off| Off, 80-100| 1%
| Alarm Volume| High| Off, Low, High| N/A
| Oximetry Averaging| No| Yes, No| N/A
SET PATIENT SECURITY MODE ON
- Turn unit on – Review factory settings below
- Step through ALL settings using LIMITS button
- Change desired settings using or buttons
- MUST push LIMITS button after each setting
- Scroll to end of settings to 2nd symbol ( =minutes) and press LIMITS button
- Unit will automatically review new settings
- TO SAVE SETTINGS – press and hold ALARM SILENCE button while briefly pressing LIMITS button – “dEF On” will flash
- Press POWER button until unit turns OFF
- TO SET – Press and hold ALARM SILENCE button while turning the unit back ON
- Screen will flash “SEC On”
Note: To set “Low O2” alarm below 85%, must set in User-Defined Defaults first, then follow steps above.
Turn Off Patient Security Mode
- Press POWER button until unit turns OFF
- Together hold down ALARM SILENCE button, then LIMITS button and then POWER button
- Units reads “SEC OFF”
DATA COLLECTION & DOWNLOAD
Data Collection
- Turn unit on – Recording begins automatically
Note: Data less than 1 minute is not saved.
Downloading Data To PC
Note: Patient Security Mode must be OFF
- Press POWER button until unit turns OFF
- Connect unit to PC using 7500SC download cable
- Boot up nVISION and select FILE>New Data Capture, click 7500 oximeter option
- On 7500 unit press and hold and POWER buttons – unit will display “PLA bAC” during download
- Unit returns to normal operating mode after download
TO CLEAR MEMORY
-
Turn unit on
Note: Patient Security Mode must be OFF -
Press LIMITS button until “CLr no” is displayed
-
Press button until “CLr YES” appears
-
Press LIMITS button to confirm
-
“dEL no” appears, press button so “dEL YES” appears
-
Press LIMITS button to confirm
-
“dnE CLr” confirms memory has been cleared
For complete specifications, please refer to the Operator’s Manual.
https://www.nonin.com/support/7500/
Translations of this quick start guide can be found using this QR code or on the USB for the Model 7500 operator’s manual.
©2021 Nonin Medical, Inc.
All trademarks are the property of Nonin Medical, Inc. unless otherwise noted.
114527-001-01 06/2021