ZEISS AT.Shooter A1-2000 VISCOJECT-BIO 1.8 Cartridge Set User Manual

June 13, 2024
Zeiss

ZEISS AT.Shooter A1-2000 VISCOJECT-BIO 1.8 Cartridge Set

Product Information

  • Specifications:
    • Product Name: AT.Shooter A1-2000 with VISCOJECT-BIO 1.8 Cartridge Set
    • Manufacturer: Carl Zeiss Meditec
    • Model Number: AT.Shooter A1-2000
    • Cartridge Set: VISCOJECT-BIO 1.8
  • Handling Instructions for Non-Preloaded ZEISS MICS IOLs:
    • Place a small amount of OVD* in the cartridge tip, ensuring a line of OVD in each channel without touching the front tip of the cartridge.
    • Pull the lens gently from its holder using forceps with fine tips to prevent drying up. Avoid using fluid from the bottle for ocular irrigation.
    • Center the lens in the opened cartridge, ensuring to mind the meridian marks for high diopter toric IOLs.
    • Maintain the IOL in place with folding forceps while closing the cartridge, ensuring a click is heard upon closure.
    • Check the position of the lens within the closed cartridge, ensuring haptic movement.
    • Mount the tip holder onto the shooter tip, insert the plunger into the blue silicone tip’s recess and remove the tip from its holder by retracting the plunger.
    • Place the cartridge into the loading dock of the shooter and push it forward.
    • Gently push the plunger with the blue silicone tip straight into the cartridge channel to start pushing the lens forward.
    • Verify correct lens positioning by pushing it forward until visible in the tip of the cartridge for lens implantation.
    • OVD: Ophthalmic viscosurgical Device

FAQs

  • Q: Can I reuse the VISCOJECT-BIO 1.8 Cartridge Set?
    • A: No, it is recommended not to reuse the cartridge set to ensure optimal performance and safety during procedures.
  • Q: Is it necessary to use ZEISS OVDs with this product?
    • A: Yes, Carl Zeiss Meditec recommends the use of ZEISS OVDs for compatibility and best results.

Product Usage Instructions

  1. Place a small amount of OVD* in the cartridge tip. Place a line of OVD in each channel of the cartridge. Do not touch the front tip of the cartridge.
  2. Pull the lens gently from its holder using a forceps with fine tips. Do not let the IOL dry up. Do not use the fluid from the bottle for ocular irrigation.
  3. Place the lens in the centre of the opened cartridge. For high diopter toric IOLs, mind the meridian marks.
  4. Maintain the IOL in place with the folding forceps while closing the cartridge. Close the cartridge until you hear a click.
  5. Control the position of the lens within the channel of the closed cartridge. Haptic must be freely movable.
  6. Mount the tip holder onto the shooter tip. Insert the plunger into the recess of the blue silicone tip and remove the tip from its holder by retracting the plunger.
  7. Place the cartridge into the loading dock of the shooter and push it forward.
  8. Gently push the plunger with the blue silicone tip mounted on it straight into the cartridge channel and start to push the lens forward.
  9. Check the correct positioning of the lens. Gently push forward the lens until it becomes visible in the tip of the cartridge. Start implantation of lens.

Please refer to our website for the most up-to-date references.

Carl Zeiss Meditec recommends the use of ZEISS OVDs. en-OUS_32_022_0170I CZ- VIII/2023 International edition: Only for sale in selected countries. The contents of the brochure may differ from the current status of approval of the product or service offering in your country. Please contact our regional representatives for more information. Subject to changes in design and scope of delivery and due to ongoing technical development. VISCOJECT is a trademark of Medicel. © Carl Zeiss Meditec AG, 2023. All rights reserved.