988-00087 Electrochemical Multiplexer Gamry Instruments Instruction Manual

June 13, 2024
GAMRY

IMX8™ Multiplexer
Operator’s Manual

If You Have Problems

Please visit our service and support page at https://www.gamry.com/support-2/. This page contains information on installation, software updates, and training. It also contains links to the latest available documentation. If you are unable to locate the information you need from our website, you can contact us via email using the link provided on our website. Alternatively, you can contact us one of the following ways:

Internet https://www.gamry.com/support-2/
Telephone 215-682-9330  9:00 AM−5:00

PM  US Eastern Standard Time
877-367-4267  Toll-free US & Canada Only

Please have your instrument model and serial numbers available, as well as any applicable software and firmware revisions.
If you have problems in installation or use of a system containing an IMX8 Multiplexer, please call from a phone next to your computer, where you can type and read the screen while talking to us.
We will be happy to provide a reasonable level of free support for registered users of the IMX8 Multiplexer.
Reasonable support includes telephone assistance covering the normal installation, use and simple customization of a computerized system containing an IMX8 Multiplexer connected to a Windows®-compatible computer.
A service contract that extends both the hardware warranty and software update period is available at an additional charge. Software updates do not include software enhancements offered to our customers at additional cost.
Enhancements to the IMX8 Multiplexer and Gamry’s standard applications software that require significant engineering time on our part can be performed on a contract basis. Contact us with your requirements.

Limited Warranty

Gamry Instruments, Inc. warrants to the original user of this product that it shall be free of defects resulting from faulty manufacture of the product or its components for a period of two years from the original shipment date of your purchase.
Gamry Instruments, Inc. makes no warranties regarding either the satisfactory performance of the IMX8 Multiplexer including the software provided with this product or the fitness of the product for any particular purpose. The remedy for breach of this Limited Warranty shall be limited solely to repair or replacement, as determined by Gamry Instruments, Inc., and shall not include other damages.
Gamry Instruments, Inc. reserves the right to make revisions to the system at any time without incurring any obligation to install same on systems previously purchased. All system specifications are subject to change without notice.

There are no warranties which extend beyond the description herein. This warranty is in lieu of, and excludes any and all other warranties or representations, expressed, implied or statutory, including merchantability and fitness, as well as any and all other obligations or liabilities of Gamry Instruments, Inc; including but not limited to, special or consequential damages.
This Limited Warranty gives you specific legal rights and you may have others, which vary from state to state.
Some states do not allow for the exclusion of incidental or consequential damages.
No person, firm or corporation is authorized to assume for Gamry Instruments, Inc., any additional obligation or liability not expressly provided herein except in writing duly executed by an officer of Gamry Instruments, Inc.

Disclaimers

Gamry Instruments, Inc. cannot guarantee that the IMX8 Multiplexer will work with all computer systems, operating systems, or third-party hardware/software.
The information in this manual has been carefully checked and is believed to be accurate as of the time of printing. However, Gamry Instruments, Inc. assumes no responsibility for errors that might appear.

Copyrights

IMX8™ Multiplexer Operator’s Manual copyright © 2021—2024, Gamry Instruments, Inc., all rights reserved.
Gamry Framework copyright © 1989—2024, Gamry Instruments, Inc., all rights reserved.
Interface 1010™, Interface 5000™, Interface Power Hub™, EIS Box™ 5000, Reference 620™, Reference 3000™, Reference 3000AE™, Reference 30K™, LPI1010™, eQCM 15M™, IMX8™, Gamry Framework™, Faraday Shield™, and Gamry™ are trademarks of Gamry Instruments, Inc.
Windows® is a registered trademark of Microsoft Corporation.
No part of this document may be copied or reproduced in any form without the prior written consent of Gamry Instruments, Inc.

Chapter 1: Safety Considerations

Your IMX8 Multiplexer is supplied in a safe condition. This chapter of the IMX8 Operator’s Manual contains some information and warnings that you must follow to insure continued safe operation of the IMX8.

Table 1-1
Safety Symbol Definitions

Symbol Description
General Warning sign;

Read throughout manual for all instances
| Warning symbol;
Risk of electric shock
| Warning symbol;
Observe humidity and water ingress requirements

Inspection

When you receive your IMX8 Multiplexer, inspect it for evidence of shipping damage. If you note any damage, please notify Gamry Instruments Inc. and the shipping carrier immediately. Save the shipping container for possible inspection by the carrier.

Warning: An IMX8 Multiplexer damaged in shipment can be a safety hazard. Do not operate damaged apparatus until a qualified service technician has verified its safety. Tag a damaged IMX8 Multiplexer to indicate that it could be a safety hazard.

Product Safety
The IMX8 Multiplexer has been designed, tested, and certified to meet the requirements of EN 61010, Safety requirements for electrical equipment for measurement, control, and laboratory use. The IMX8 Multiplexer circuitry is at voltages low enough to be considered safe.
The IMX8 Multiplexer ’s circuitry does not contain voltages higher than 48 V DC. As a generalization, input and output voltages in the IMX8 Multiplexer are limited to 24 V. This voltage level is considered safe.
The “AC Adapter” supplied with the IMX8 Multiplexer is certified under EN 60950. The AC Adapter converts the AC mains voltage to 48 V DC, which is used to power the IMX8 Multiplexer.
Always use the AC adapter (power brick) supplied with your IMX8 Multiplexer to power the IMX8. One exception to this rule; power supplied by a Gamry IPH Interface Power Hub is also allowed.

Warning: Use only Gamry-approved DC power sources with your IMX8 Multiplexer. Other power sources may alter the performance and/or safety characteristics of the IMX8 Multiplexer.

AC Mains Connection to the Power Brick
The IMX8 Multiplexer does not connect directly to an AC Mains supply. Instead, the mains are connected to desktop AC adapter (power brick), which outputs 48 V DC, which in turn powers the IMX8 Multiplexer.

The IMX8 Multiplexer’s AC Adapter is rated for operation from 100 to 240 V AC, 47 to 63 Hz. It should therefore be useful throughout the world.

The IMX8 Multiplexer is normally provided with an AC line cord suitable for your location. This AC line cord connects the AC mains to the AC power adapter. If your IMX8 Multiplexer has been provided without an AC line cord, or a cord that is not compatible with your local AC mains socket, obtain a line cord certified for use in your country. Contact your local Gamry Representative or e-mail to techsupport@gamry.com if you are uncertain what AC line cord to use.

Grounding in the IMX8 Multiplexer
Most of the IMX8 Multiplexer’s circuitry and its metal case are connected to earth ground via the AC adapter. The cell connections to the Active cell can be isolated from earth ground if the System Potentiostat is isolated from earth. Inactive cells that are Open (not in the Shorted or Local Potentiostat modes) will also be isolated from earth ground.
Most electrochemical cells are isolated from earth ground, so isolation of the IMX8 Multiplexer and System Potentiostat from earth is not required. In these cases, connection of the System Potentiostat chassis to an earth ground may lower the noise seen in an electrochemical test. A Chassis Ground banana jack on the rear panel of the System Potentiostat easily implements this connection. Simply run a wire from this binding post to a suitable source of earth ground. The black 1.2-meter wire usually provided with Gamry Potentiostats facilitates this connection.

Sources of earth ground include:

  • Most metal water pipes,
  • the chassis of most electronic apparatus (which are generally earth-grounded), and
  • the earth ground terminal of an AC mains power plug.

We recommend that you discuss grounding with an electrical or electronics professional prior to making this earth-ground connection.
This connection of the System potentiostat to an earth ground is not a “Protective Earth Ground” as defined in EN 61010. The IMX8 Multiplexer is safe in the absence of this connection.

This binding post is not intended for any use other than connecting the System Potentiostat to a functional earth ground to improve shielding against noise. Connecting this binding post to a hazardous voltage can create a significant safety hazard.

Warning: Do not earth ground the System Potentiostat if your electrochemical cell has earth ground. This is especially dangerous if the cell can source significant power (e.g., a battery, fuel cell, or capacitor). The cell can discharge through the IMX8 and/or System Potentiostat, possibly causing severe damage to the IMX8 or System Potentiostat.

Functional Grounding and Product Safety
As defined in IEC Publication 348, Safety Requirements for Electronic Measuring Apparatus, the IMX8 is a Class I grounded device.
When you use the IMX8 with an approved line cord, the connection to the functional earth ground is automatically made prior to making any power connections. The USB connection can also sometimes connect the IMX8 to functional ground if the computer is earth grounded.

Warning: Do not negate the protection of the IMX8’s earth ground by any means. Do not use the IMX8 with a two-wire extension cord, with an adapter that does not provide for earth grounding, or with an electrical outlet that is not properly wired with an earth ground. If the earth ground is not properly connected, it creates a safety hazard, which could result in personal injury or death.

The IMX8 is supplied with a line cord suitable for use in the United States. In other countries, you may have to replace the line cord with one suitable for your type of electrical outlet. You must always use a line cord with a CEE 22 Standard V female connector on the instrument end of the cable. This is the same connector used on the US standard line cord supplied with your IMX8.

Warning: If you replace the line cord, you must use a line cord with the same polarity as that supplied with the IMX8. An improper line cord can create a safety hazard, which could result in personal injury or death.

The wiring polarity of a properly wired connector is shown in Table 1-2 for both US line cords and European line cords that follow the “harmonized” wiring convention.

Table 1-2
Line Cord Polarities and Colors

Area Line Neutral Earth Ground
USA Black White Green
Europe Brown Light Blue Green/Yellow

If you have any doubts about the line cord for use with your IMX8, please contact a qualified electrician or instrument service technician for assistance. He or she can perform a simple continuity check that can verify the connection of the IMX8 chassis to earth and thereby check the safety of your IMX8 installation.

Temperature and Ventilation
Your IMX8 Multiplexer was designed to operate at ambient temperatures between 0°C and 40°C.

Caution: Be careful when operating the IMX8 in an enclosed space. The temperature within the enclosure must not exceed 40°C. You may need to provide ventilation holes or even forced air cooling for the enclosure if the temperature rises excessively.

Defects and Abnormal Stresses
Treat your IMX8 Multiplexer as potentially hazardous if any of the following is true of the unit:

  • It shows visible damage,
  • It does not operate properly,
  • It has been stored for a long period of time under unfavorable conditions,
  • It has been dropped or subjected to severe transport stress,
  • It has been subjected to environmental stress (corrosive atmosphere, fire, etc.).

Do not use your IMX8 Multiplexer or any other apparatus if you think it could be hazardous. Have it checked by qualified service personnel.

Environmental Limits
There are environmental-limit conditions on the storage, shipping, and operation of this equipment. The IMX8 Multiplexer is not designed for outdoor use.

Storage| Ambient Temperature Relative Humidity| −20°C to 60°C
Maximum 90% non-condensing
---|---|---
Shipping| Same as storage plus Acceleration| Maximum 30 G
Operation| Ambient Temperature Relative Humidity| 0°C to 40°C
Maximum 90% non-condensing

Warning: The IMX8 Multiplexer is not designed for operation in conditions where liquid water may enter the chassis, or water vapor may condense within the chassis. Operation of a IMX8 Multiplexer that has water within the chassis can create a safety hazard, which could result in personal injury or death.

Cleaning
Disconnect the IMX8 Multiplexer from all power sources prior to cleaning.
Use a cloth lightly dampened with either clean water or water containing a mild detergent to clean the outside of the IMX8 Multiplexer enclosure. Alternatively, you can use isopropyl alcohol. Do not use a wet rag or allow fluid to enter the IMX8 Multiplexer enclosure. Do not immerse the IMX8 Multiplexer in any type of cleaning fluid (including water). Do not use any abrasive cleaners.

Service
Your IMX8 Multiplexer has no user-serviceable parts inside. Refer all service to a qualified service technician.

Warning: Never operate the IMX8 Multiplexer with any cover or panel on the chassis open. Dangerous voltages may be present at several points within the IMX8 Multiplexer chassis, including PC board traces. Always remove the power connection before opening the IMX8 Multiplexer case.

RF Warning
Your IMX8 Multiplexer generates, uses, and can radiate radio-frequency energy. The radiated levels are low enough that the IMX8 Multiplexer should not create an interference problem in most industrial laboratory environments. The IMX8 is likely to cause radio-frequency interference if operated in a residential environment.

Electrical Transient Sensitivity
Your IMX8 Multiplexer was designed to offer reasonable immunity from electrical transients. However, in severe cases, the IMX8 could malfunction or even suffer damage from electrical transients. If you are having problems in this regard, the following steps may help:
If the problem is static electricity (sparks are apparent when you touch the IMX8):

  • Placing your IMX8 on a static-control work surface may help. Static-control work surfaces are now generally available from computer-supply houses and electronics-tool suppliers. An antistatic floor mat may also help, particularly if a carpet is involved in generating the static electricity.
  • Air ionizers or even simple air humidifiers can reduce the voltage available in static discharges.

If the problem is AC power-line transients (often from large electrical motors near the IMX8):

  • Try plugging your IMX8 into a different AC-power branch circuit.
  • Plug your IMX8 into a power-line surge suppressor. Inexpensive surge suppressors are now generally available because of their use with computer equipment.

Contact Gamry Instruments, Inc. if these measures do not solve the problem.

CE Compliance
The European Community has instituted standards limiting radio-frequency interference from electronic devices and mandating several safety requirements. Gamry Instruments, Inc. has modified its instruments, including the IMX8, to comply with these standards.
The relevant CE regulations include EN 61010.

RoHS Compliance
The IMX8 Multiplexer has been built using lead-free components and lead-free solder. It is in compliance with the European RoHS initiative.

Chapter 2: Introduction

About this Manual
This manual covers the installation, safety, and use of the Gamry Instruments IMX8 Multiplexer.
This manual describes use of an IMX8 Multiplexer with Revision 7.9 (and later revisions) of the Gamry Framework software. It is equally useful when setting up a newly purchased instrument or modifying the setup of an older instrument for use with new software.
You will find dry technical material such as specifications and connector pin- out diagrams in the Appendices.
This manual does not discuss software installation or software operation in any detail. Software support for the IMX8 Multiplexer is described in the Gamry Framework’s Help system.
All the Gamry Instruments’ applications running under the Gamry Framework control the IMX8 Multiplexer via an IMX object. See the Framework’s Help system for information concerning IMX objects and their functions.

About the IMX8 Multiplexer
The IMX8 Multiplexer was developed to increase the throughput of your electrochemical testing program. It allows one potentiostat/galvanostat/ZRA to drive up to eight electrochemical cells. Under the control of an external computer, the IMX8 connects one cell at a time to the potentiostat, which makes an electrochemical measurement on that cell.
The IMX8 was specially designed for the needs of electrochemists. Extra features such as sense line and shieldswitching, and control of inactive cells separate the IMX8 – designed specifically for electrochemists – from other general-purpose switching devices.
The IMX8 has been integrated into Gamry Instruments’ electrochemical applications relevant to the IMX8 that run under the Gamry Framework. It allows many of the DC Corrosion measurement applications to increase their throughput by testing multiple cells in a single test run. In this system a sophisticated Microsoft Windows®-program is used to control both the IMX8 and a Gamry Instruments potentiostat.

However, the IMX8 is not restricted to use in corrosion testing or use with Gamry Instruments hardware or software. It was designed to work with other potentiostats , including those sourcing as much as 1 ampere of current. Refer to Gamry Instruments’ Echem Toolkit™ for use with 3rd party potentiostats.

This chapter of the IMX8 Operator’s Manual provides an overview of the operation of the IMX8. It discusses the following topics:

  • Operation overview
  • Switching and inactive cell control

The material in this chapter is only an overview. If you are using your IMX8 Multiplexer with Gamry Instruments software, this overview should give you sufficient understanding to make intelligent use of your IMX8.

Operation Overview
Figure 2-1 shows a “typical” electrochemical test system built around an IMX8 Multiplexer. Up to eight cells can be connected simultaneously to an IMX8. The label on the block marked as a potentiostat is slightly misleading. Most potentiostats can also be used as a galvanostat or as a ZRA (zero- resistance ammeter) in an IMX8 system.

In the description of this figure, keywords are in bold type and are defined in the next section.

Figure 2-1
Typical Electrochemical Test System Using an IMX8 Multiplexer

The system computer is a critical component of the system. It controls the IMX8 by sending it simple commands through a USB connection.
The most important function of these commands is to select which cell is active (connected to the system potentiostat). You can think of an IMX8 as a complex switch. One of the eight cell cables on the IMX8 is switched so that it connects to the potentiostat’s cell cable. All of the wires (including sense leads and shields) in the cell cable are switched. Other commands select the IMX8’s control mode for the cells that are inactive (not presently connected to the potentiostat).
The system computer also controls the system potentiostat. Most modern potentiostats generate digital signals. Primarily, a standard USB connection is used to digitally control these potentiostats. In the case of analog potentiostats, analog signals from a D/A and A/D computer interface can be sent. When a cell is connected to the system potentiostat cell leads, the computer causes a test to be run on that cell. The test can be as simple as a single current or potential measurement or as complex as a cyclic scan or acquisition of an EIS spectrum. The computer software and system potentiostat are responsible for taking the measurement and storing it.
The system potentiostat in Figure 2-1 is shown as a “black box”. The IMX8 is designed to work with virtually any potentiostat. The relays in the IMX8 can carry currents up to 1.5 A, allowing use of the IMX8 with most high- performance potentiostats.
In a typical IMX8 experiment, the system computer cycles through the cells. Each cell is connected in turn to the system potentiostat and used to collect a measurement. Usually, the computer delays at the end of each cycle through the cells so that the measurements on each cell are separated by a fixed time period.

Keyword Definitions
The following key terms are used throughout this manual.

  • Active Cell
    The active cell is the cell (if any) currently connected to the system potentiostat. Any electrochemical measurements are made on the active cell.

  • Cell
    The term cell is used to describe either an electrochemical cell or the switches needed to connect an electrochemical cell to the system potentiostat.

  • Channel
    An IMX8 contains eight channels. A channel includes cell-switching, a local potentiostat and a D/A converter. The term cell is used when only the cell- switching portion of a channel is being discussed.

  • Inactive Cell
    The term inactive cell is any cell not currently connected to the system potentiostat. Inactive is a slightly misleading term since an inactive cell can actually be potentiostated by the IMX8’s local potentiostat on that channel. An inactive cell can be also shorted to mimic an active cell in ZRA mode.

  • Inactive Mode
    Inactive cells can be open, connected to a local potentiostat, or shorted. The term inactive mode describes this choice of control mode for inactive cells.

  • Local Potentiostat
    A simple low-current potentiostat that maintains potential control on inactive cells. Unlike the system potentiostat, the local potentiostat is incapable of making electrochemical measurements.

  • System Computer
    The computer responsible for coordinating operation of an IMX8-based electrochemical test system, not to be confused with the simple micro- controller buried within the IMX8.

  • System Potentiostat
    This is the potentiostat responsible for making electrochemical measurements on the cells used in the experiment. The system potentiostat is typically a full-featured, computer-controlled instrument.

Switching and Inactive Cell Control
In the Operation Overview, we said you can think of the IMX8 as a complex switch. This section of the manual describes the IMX8-switching in some detail. In addition, control modes for inactive cells are discussed.
Each of the eight IMX8 channels is made up of an independent set of switches and control circuits. Figure 2-2 is a simplified schematic diagram for one channel of the IMX8. Switches connected with dotted lines are always switched together and are therefore treated as one multi-pole switch. All the switches are under control of the system computer. Shield relays are not shown.

Figure 2-2
Simplified Schematic of a Single IMX8 Channel

The cell cable from the system potentiostat is connected at the left side of the schematic. The system-potentiostat cell cable can be connected to any of the IMX8 channels. The system computer must ensure that only one cell at a time is connected to the system potentiostat.
Switch S1 controls which cell is active. S3 and S4 control the inactive mode of the cell on this channel.
Looking at the schematic, you can see that switch S1 connects this channel’s cell to the system potentiostat. An active channel always has S3 and S4 open. The relays in S1 are rated to carry 1.5 A and switch 500 mA. You can normally use the IMX8 with potentiostats rated at up to 100 V of output compliance and 1.5 A of output current.
To avoid transients, Gamry’s software always turns off the cell at the system potentiostat prior to switching S1.
The IMX8 switches all its relays nearly simultaneously (even those in different channels), but there is still about 1 millisecond in which the relay pattern in the IMX8 is in neither the initial or final configuration. Switching transients as the IMX8 switches channels with a “live” potentiostat can be particularly severe because the system potentiostat’s feedback goes through one of the S1 relays. To avoid the problem, turn off the cell at the system potentiostat before switching channels.
You have a choice of three different control schemes for inactive cells. These are tabulated in Table 2-1 along with the active mode.

Table 2-1
IMX8 Control Modes

Mode Switch S1 Switch S3 Switch S4
Active Closed Open Open
Open Open Open Open
Local Potentiostat Open Closed Open
Shorted Open Open Closed

The inactive control mode labeled “open” is the simplest. In this mode the cell is disconnected and isolated from both the system potentiostat and the IMX8. No cell current can flow between any electrodes. This is the most common inactive mode in corrosion studies, where the cell behavior at open circuit is of interest.
The inactive control mode labeled “local potentiostat” is used to maintain potential control between readings made with the system potentiostat. If you know something about operational amplifiers and potentiostats, you may recognize that the operational amplifier in Figure 2-2 is wired as a simple potentiostat. With S3 closed, the working electrode is connected to the ground on the IMX8 circuit board and the counter electrode is connected to the output of an operational amplifier. The feedback for this amplifier is taken from the cell’s reference electrode. This mode can be used whenever it is important to maintain potential control between readings. One experimental example is tests on amperometric chemical sensors.
Unlike a typical system potentiostat, the local potentiostat is very limited in capability. The amplifier in this local potentiostat is not a large power amplifier. Its output is typically capable of generating ±30 mA at up to ±10 V.
The local potentiostat has no cell switch other than S3, and it cannot measure current.

Caution: You can damage the local potentiostat’s amplifier if the reference electrode potential exceeds ±24 V versus the IMX8 chassis.

The cell can be potentiostated at a voltage that is set via the D/A converter seen in Figure 2-2. This is a 16-bit converter operating at a resolution of 152.59 µV per bit. The maximum voltage available is ±4.9 V.
The final inactive mode is label “shorted” in Table 2-1. In this mode a relay shorts the working and counter electrodes. This mode was added so that the IMX8 could be used in galvanic corrosion studies. In a galvanic corrosion experiment, the current flow between two electrically connected metals is measured. The device that makes this measurement is called a zero-resistance ammeter (ZRA) .
In the shorted inactive mode the two metals are connected, even when the system ZRA is measuring other channels.
Both the local potentiostat and shorted mode have voltage inaccuracies caused by the cell current. The working electrode is nominally at zero volts versus earth ground. When cell current is present, resistance in relays, wires, and connectors causes the voltage to differ from zero. In the shorted mode the counter lead also has resistance that contributes voltage error. With a 1.5 m Gamry Instruments cell cable, the resistance in each cell lead should be less than 0.2 Ω. At 30 mA of current, the voltage error in the local potentiostat mode can be as large as 6 mV. In the shorted mode and at the same current, the total voltage error can be 12 mV.

Notational Conventions
In order to make this manual more readable we have adopted some notational conventions. These are used throughout this manual and all other Gamry Instruments manuals:

  • Numbered lists. A numbered list is reserved for step-by-step procedures, with the steps always performed sequentially.
  • Bulleted list. The items in a bulleted list, such as this one, are grouped together because they represent similar items. The order of items in the list is not critical.
  • File names and folders. Inside paragraphs, references to computer files and Windows® folders are capitalized and placed within quotes, for example: “C:\MYGAMRYDATA\CV.DTA” and “GAMRY.INI”.

Chapter 3: Installation

This chapter of the IMX8 Multiplexer Operator’s Manual covers normal installation of the IMX8. Both installation as part of a Gamry Framework-based electrochemical-measurement system and installation as a component in a user- configured system are discussed.
These instructions assume use with Gamry’s Framework Software Revision 7.9 or higher.

Figure 3-1
Front View of an IMX8 Multiplexer

Initial Visual Inspection
After you remove your IMX8 Multiplexer from its shipping carton, check it for any signs of shipping damage. If you find any damage, please notify Gamry Instruments, Inc., and the shipping carrier immediately. Save the shipping container for possible inspection by the carrier.

Caution: An IMX8 damaged in shipment can be a safety hazard. Do not operate a damaged apparatus until a qualified service technician has verified its safety. Tag a damaged IMX8 to indicate that it could be a safety hazard.

Physical Location
Place your IMX8 Multiplexer on a flat workbench surface. You need access to the rear of the instrument because some cable connections are made from the rear. The IMX8 is not restricted to operation in a horizontal position. You can operate it on its side, or even upside-down.

Caution: The IMX8 has an internal fan which distributes the heat generated by the IMX8 circuitry to the metal chassis. Blocking heat transfer from the chassis to the environment could cause IMX8 overheating.

If you place your IMX8 Multiplexer within an enclosed space, make sure that the temperature within that space does not exceed the 40ºC ambient temperature limit of the IMX8. Be particularly careful if a computer or other heat- dissipating equipment is mounted in the same enclosure as the IMX8.
The IMX8 Multiplexer has not been designed for outdoor use.

Quick-start Guide for System Installation
Your shipment includes a short document entitled Quick-start Installation Guide–USB Potentiostat. It contains the latest instructions for installing Gamry hardware and software onto a computer system. If this document is missing, the following information should be sufficient for you to install the Gamry Framework Software and Gamry potentiostat onto your computer.

Connection to Gamry Instruments Reference 600+ or 620 and Interface 1000 or 1010 potentiostats
Connection of your IMX8 to a Gamry Instruments Reference 600+ or 620 and Interface 1000 or 1010 potentiostat respectively is quite simple. Gamry Instruments supplies a standard cable (part number 985-00080) that properly connects the cell wires. This 1.5 m cable is equipped with a male 9-pin D-connector on the IMX8 end and a male 25-pin D-connector on the Reference 600+ or 620 and Interface 1000 or 1010 end.

Caution: Never substitute a generic 25-to-9-pin cable designed for computer interconnection for this cable. Gamry Instruments’ cable includes individual shielding for several signals in this cable. Use of a generic cable is likely to create excessive noise and system oscillation.

The Reference 600+ or 620 and Interface 1000 or 1010 can identify the cable plugged into its cell connector.
It detects the presence of an IMX8 Multiplexer in your system. The Reference 600+ or 620 and Interface 1000 or 1010 cannot identify the cables used to connect the eight individual cells to the IMX8.
EIS performance with an IMX8 and a Reference 600+ or 620 and Interface 1000 or 1010 is optimal when all the cell cables are 1.5 m cables. All other cables may create errors in phase and magnitude.

Connection to a Third-party Potentiostat/Galvanostat
Connecting the IMX8 to a third-party potentiostat/galvanostat is more of a problem: You have to build a custom cable. See Appendix C of this manual for the pin-layout of the IMX8 Potentiostat connector and for hints on cable construction. We strongly recommend that you contact us prior to constructing the cable.

Cell Cable Installation
Each cell to be tested in an experimental run requires a cell cable.
When you are using the IMX8 with a Gamry Instruments potentiostat, Cell Cable connections to the cell are made in the manner described in the potentiostat’s Operator’s Manual. Cell connections when the IMX8 is used with a third-party potentiostat may not follow the normal Gamry conventions, so you should contact Gamry Instruments to obtain special instructions if this is your situation.
Your IMX8 Multiplexer comes with eight 1.5 m long cell cables with a male 9-pin D-connector on the IMX8 end. The other end of the cable has banana-plug /alligator-clip termination to connect to a cell. Longer cables as well as specialized cables are also available, including cables designed for connection to standard threeterminal corrosion-monitoring probes. Contact support for more information.

If you need to design a cell cable for your specific needs, refer to the pin- layout of the IMX8 cell cable in Appendix B of this manual.

Computer Requirements
Before you connect an IMX8 Multiplexer to a host computer, you must ensure that your computer meets these simple requirements:

  • A computer based on one of the x86 family of Intel™ microprocessors or a 100%-compatible processor from another vendor.
  • Microsoft Windows® 10 or newer is required with Gamry Framework™ software version 7.9.1 and above. Only the 64-bit version of these operating systems is supported.
  • A USB port that supports Full Speed (12 Mbits/second) or High Speed (480 Mbits/second) USB transfers. It must be compatible with Revision 1.1 or Revision 2.0 of the USB specification.

Gamry’s Windows®-based application software packages may impose additional, more stringent requirements.
See the software documentation or contact a Gamry representative for details.

Software Installation
The IMX8 Multiplexer is compatible with the Windows® Plug & Play configuration system. Like most Plug & Play hardware, it is best if you install the software for the IMX8 Multiplexer before you install the instrument hardware.

The latest Gamry software is not provided on a DVD anymore but is now available for download as .exe or .iso file on Gamry Instruments’ Client Portal after creating an account and registering your instrument:
https://www.gamry.com/client-portal/my-account/

An internet download of Gamry Instruments’ software will download a self- extracting file. Running this file will extract the software and begin the installation.

Reboot your Computer after Software Installation
Reboot your computer after the Gamry Setup program is done. The Setup program normally offers you the opportunity to do so. USB device drivers are usually loaded when Windows boots up. Following Setup, you may not be able to use your IMX8 Multiplexer until the drivers are loaded.

Device-driver installation may not occur until a while after the Windows® Desktop appears.
On a slow computer, or a busy computer with lots of active applications, the delay before driver installation can be a minute or more.

USB Cables
The IMX8 Multiplexer connects to the computer using a standard High Speed USB A/B cable. A suitable cable was shipped with your IMX8 Multiplexer. If this cable is lost, you can get a replacement at almost any computer retailer. The replacement cable should be rated for USB 2.0 High Speed USB operation.
An A/B USB cable has different connectors on each end. The end with a wider, rectangular connector plugs into a USB port on your computer (or a similar port on a USB hub). The end with a nearly-square connector plugs into the USB port on the rear panel of the IMX8 Multiplexer (see Figure 3-2).

The USB connection can be “hot-plugged”. This means that both the computer and the IMX8 Multiplexer can be powered up before the USB cable is plugged in. Unlike many other instrument system connections, you need not power down the system before plugging in the USB.
You may also safely remove the USB cable without powering down the IMX8 Multiplexer and your computer.
Be aware however, that this may have undesirable consequences if the system is currently taking data or performing an electrochemical experiment.

Figure 3-2
Rear Panel of the IMX8 Multiplexer

Front Panel USB LED
The front panel USB LED provides a simple test of two aspects of normal IMX8 Multiplexer USB operation. It has three normal states:

Unlit| The USB cable is disconnected, or the USB connection is disabled by the host computer.
---|---
Continuous green| A valid cable connection has been made and the IMX8 USB processor is receiving power from the USB cable.
Continuous green| Valid USB messages are being transferred between the computer and the IMX8 Multiplexer.
Continuous red| A software download is in progress, or a USB communications error. Do not turn off the IMX8 Multiplexer.

The firmware update process is described later in this chapter.

First-time Device Installation in Windows®

These instructions presume you have already installed Gamry software Revision 7.9 or higher.

Running the Framework
Regardless of your electrochemical application, Gamry recommends running the Gamry Framework after you install new Framework software or add an instrument to your system. The Framework Instrument Manager allows you to:

  • Rename instruments,
  • Calibrate instruments,
  • Manage instrument firmware,
  • Authorize specific applications for use with specific instruments

Run the Gamry Framework by clicking on the icon it installed on the Windows® desktop. You can connect and power any Gamry instrument either before or after you start the Framework.

Framework Device Status Bar
By default, the Gamry Framework shows a Device Status Bar under its main menu (see Figure 3-3). If you don’t see the Device Status Bar when you run the Gamry Framework, it has been disabled in the Framework Options menu.

Potentiostat Devices (instruments) that are connected to the computer appear on this bar. The round indicator associated with each device shows its status:

Green The device is available to run experiments.
Orange The device is currently running an experiment.
White The device is connected to the system but is not usable. This is

generally the result of a mismatch between the Framework software and the device’s firmware. You can use the Gamry Instrument Manager to fix the mismatch.

The screen capture below shows a Framework screen with three USB instruments connected.

Figure 3-3
Framework with Three Potentiostats and One Running Test

The Reference 620 (REF620-46052) in this system is shown with a green indicator because it is installed and ready to run. The Interface 1010 labeled IFC1010-01860 has a yellow indicator because it is recording the EIS spectrum shown on the screen. The Reference 620 labeled REF620-chem lab has a white indicator, showing it is plugged in but cannot be used. This is an indication of obsolete firmware.
Though no IMX8 Multiplexer was used in this example, its status indicator behaves in the same manner described above.

Gamry Instrument Manager
Use Gamry’s Instrument Manager (GIM) application make changes to the configuration of your IMX8 Multiplexer system. Get access to this dialog box via the Options menu in the Gamry Framework.
Use the Instrument Manager to:

  • Rename devices
  • Delete devices that are not currently connected to the computer
  • Select the order in which devices appear in menus
  • Update firmware within devices
  • Pair your Multiplexer with a potentiostat

Figure 3-4 shows an example of the Gamry Instrument Manager window with an IMX8 Multiplexer.

Figure 3-4
Instrument Manager Dialog Box

GIM appears automatically when a new instrument is connected to the computer. You can start GIM anytime by opening Framework and selecting Options > Instrument Manager….
GIM is always in its own window, separate from the Framework window.
Each Gamry device in the system appears in the list on the left. All Gamry Instruments devices that are known to the system are displayed in the Instrument Manger. Select an instrument by clicking on its name. Selecting an instrument that is connected and idle blinks its Power LED. After a moment, your potentiostat should appear next to Devices Present along with a green virtual LED. Repeat for additional devices.

There is only one firmware image that you can update on your IMX8 Multiplexer.
• Instrument Firmware
Handles most of the functions of the IMX8 Multiplexer.

Appropriate update files (firmware images) can be obtained from the Gamry Instruments website at www.gamry.com. Download the file containing the new image and save it onto your computer’s hard disk.
Alternatively, every Gamry Software disk (or Gamry Software Flash Drive) contains a firmware folder that contains firmware files compatible with that DVD’s Framework revision.

GIM automatically determines if the software and firmware are compatible. The status of compatibility is shown in the Device Info area.
If there is a conflict, this area shows the following (here using a Reference 3000 as an example):

If the firmware is compatible with software version, the Device Info area appears as:

Auto Update All
If no updates are required to use the instrument, this tool is grayed out. When your firmware revision is incompatible, the Auto Update All button is active. This tool automatically selects the compatible firmware and installs it. Gamry Instruments recommends that you use this button. To use this tool, take the following steps:

Caution: Do not power off or disconnect the instrument during this process. Doing so can cause irreversible damage to the instrument.

Click the Auto Update All button. A window appears giving information on what version of firmware is currently installed, and what version will be installed. The install process starts automatically. After the relevant firmware is updated, a window appears confirming the process is complete and successful. Click the OK button. The instrument is now ready to be used.

Select And Update
The Select And Update button is available at all times, even if an update is not required. Gamry recommends not altering the firmware if it is not necessary. This tool allows you to change individual firmware.

Please contact Gamry Tech Support if you are unsure if firmware modification is required.

Pairing your IMX8 Multiplexer
Pairing your Multiplexer with one of your potentiostats is simple. Two different methods within the Framework software can be used.

  1. Be sure your host computer is on, the multiplexer is on, and your potentiostat is on.

  2. Run the Gamry Framework software.

  3. In the main menu bar, choose Options > Instrument Manager….
    The Gamry Instrument Manager window opens. a. Click the … button under Pair with… to open the selection window. b. Select the potentiostat you want to pair with your IMX8 Multiplexer and press OK
    The Instrument Manager displays now the potentiostat paired with your IMX8 Multiplexer.

  4. In the main menu bar, choose Experiment > Utilities > Multiplexer Configure and Test.
    The Select Function window appears. a. Select Multiplexer Pair.
    Click the OK button.
    b. The Select Multiplexer to Pair window opens.
    Select a multiplexer. Click the OK button.
    c. The Select Potentiostat to Pair with IMX8 window opens.
    Select a potentiostat. Click the OK button.
    d. A window confirms the successful pairing process. Click the OK button.

Test the Communications and Run Your Experiment

  1. Be sure your host computer is on, the multiplexer is on, and your potentiostat is on.
  2. Run the Gamry Framework software.
  3. In the main menu bar, choose Experiment > Utilities > Multiplexer Configure and Test.

The Select Function window appears.
a. Select Multiplexer Test.

Click the OK button.
A window appears which prompts you to select a Multiplexer.

Click the OK button.
Watch the eight amber LEDs on the front of the multiplexer illuminate in sequence two times.
Listen for the simultaneous clicks of the relays.
b. Select Set Channel.

Click the OK button.
A window appears which prompts you to select a Multiplexer.

Click the OK button.
Activate the checkboxes corresponding to the active channels on the multiplexer.

When a particular channel runs, the Channel LED for that channel illuminates.

Table 3-1
Channel LEDs

Status LED color
Active channel Green
Inactive channel Off
Inactive mode of Local Potentiostat Yellow

Chapter 4: Grounding Recommendations for EIS Measurements

The IMX8 Multiplexer has not been configured for use with “floating” potentiostats. The mismatch in ground potentials can result in excess noise and discontinuities in Electrochemical Impedance Spectroscopy (EIS) testing.

The IMX8 Multiplexer is generally not recommended for high-impedance measurements such as those found in high-quality coating systems.
For any other system, it is recommended to operate your system in an earth- grounded configuration. To reduce noise, test cells should be placed in a Faraday cage for all EIS measurements.

The grounding scheme described below will reduce noise and improve the quality of an EIS measurement when using the IMX8 Multiplexer.

  1. Connect the banana-to-banana cable (985-00083) included with the instrument to the Chassis ground at the rear of the IMX8.

  2. Connect the opposite banana plug to the paired potentiostat’s chassis ground.

  3. Connect both ground leads of the IMX8 cell cable for any channel that is in use:
    a. If a Faraday Cage is used, connect both black ground leads to the Faraday cage.
    b. If no cage is used, short both ground leads together and leave them hanging so that no contact to a conductive surface can be made.

  4. Leave all other lead connections the same.

Appendix A: IMX8 Multiplexer Specifications

All specifications are at an ambient temperature of 25°C, with the IMX8 Multiplexer powered using the external power supply shipped with the unit (116 V AC). All specifications are after software calibration.
All specifications are subject to change without notice.

Channel Characteristics

Modes| | Active
Off
Local
Shorted|
---|---|---|---
Channel switching time| Max| 10| ms
Cell current| Max| 1| A
Input voltage (measured versus IMX8 chassis ground) operating no damage| Typ| ±12
±24| V
Counter Electrode voltage (measured versus IMX8 chassis ground)| Typ| ±30| V

Channel Isolation

Current leakage (to active channel pin from any source) Max 2 nA
Impedance (equiv. impedance to IMX8 chassis ground) Typ >500
<20

pF

Local potentiostat

Compliance Current (500 Ω load) Typ ±20 mA
Compliance Voltage (1 kΩ load) Typ ±9.5 V
Applied E-range (Working versus Reference) Typ ±4.9 V
Resolution Typ 152.59 µV/bit
Zero-offset error1 (0 V output into 10 kΩ load) Typ Max <0.2
10 mV
Gain error1 (10 kΩ load) Typ <0.2 %
Reference input current Typ <5 pA

¹Applied voltage error is sum of zero-offset and gain error.

Power Input

Input voltage ranges (selectable) 100, 120, 220, 240 V rms
Allowed voltage variation (all ranges) Typ ±10 %
Power (worst case) Typ <15 VA

Environmental

Operating Temperature Range 0 to 40 °C
Relative Humidity Max 90 (non-condensing) %
Storage and Shipping Temperature –20 to 60 °C
Maximum Shipping Acceleration 30 G

Appendix B: I/O Connections for the IMX8

The IMX8 Multiplexer has various connectors on its front panel. This section discusses these connectors and their pin assignments in detail.

Figure B-1
Front Panel of the IMX8 Multiplexer

Cell Cable Connector
The eight cell connectors (one for each channel) are located on the IMX8 front panel, see Figure B-1. Each cell connector is a 9-pin female D-shaped connector. The pin-layout of all eight connectors is identical. The signal assignment for each pin is shown in Table B-1.
Most of the pins in the IMX8 cell connector are switched to pins on the potentiostat connector when the cell is active. Only one pin is not switched: the ground pin. It is always connected to the system potentiostat’s (floating) ground.

Table B-1
Cell Cable Connector

Pin Name Use
1 Working Sense (WS) Connected to the working electrode in cases where the

system potentiostat uses a differential electrometer. This pin can be left open when the IMX8 is used with a potentiostat that uses a single-ended electrometer.
2| WS Shield| The shield for the working sense input. Left open at the cell end of the cell cable.
3| Working Electrode (WE)| Connected to the working electrode. The cell current flows through this pin.
4| WE Shield| The shield for the working electrode input. Left open at the cell end of the cell cable.
5| Ground| The system potentiostat’s ground. It can be used to shield the cell if very low currents need to be measured. Also recommended as a shield for the counter electrode cable.
6| Reference Electrode (REF)| Connected to the reference electrode.
7| REF Shield| The shield for the reference electrode cable. Left open on the cell end of the cell cable.
8| Counter Sense (CS)| With a potentiostat, used to sense the potential of the counter electrode. Allows accurate ZRA mode  even with significant cell current through counter lead’s resistance.
9| Counter Electrode (CE)| Connected to the cell’s counter electrode. The cell current flows through this connection.

Normal Cell Cable Connections
Normally you can use one of Gamry Instruments’ standard cell cables to connect each channel’s connector to its cell. Gamry supplies cables that are 1.5 m, 4.5 m, or 9 m in length. Special-purpose cables are also available.
The standard cell cable terminates in banana plugs and one pin jack. Table B-2 identifies each terminal of the cable and its connections to the cell.

Table B-2
Normal Cell Cable Connections

Color Type Name Normal Connection
Blue Banana Plug Working Sense Connect to working electrode
Green Banana Plug Working Electrode Connect to working electrode
White Pin Jack Reference Electrode Connect to reference electrode
Red Banana Plug Counter Electrode Connect to counter electrode
Orange Banana Plug Counter Sense Connect to counter electrode
Long Black Banana Plug Floating Ground Leave open or connect to a Faraday

shield
Short Black| Banana Plug| Chassis (Earth) Ground| Leave open or connect to a Faraday Shield

The working electrode is the specimen being tested. You make two connections to the working electrode. One of the two connections, with the blue banana plug, is used to sense the voltage at the working electrode when you use the IMX8 with a system potentiostat that includes a differential electrometer. If you are using a system potentiostat with a single-ended electrometer, you may leave this wire disconnected. The green workingelectrode connection carries the cell current.
Connect the white pin jack to the cell’s reference electrode, generally an SCE or Ag/AgCl reference electrode. Connect the red (and orange) banana plug to the counter or auxiliary electrode.
The longer black lead is connected to the system potentiostat’s ground. The shorter black lead is connected to earth ground. In most cases, leave these terminals disconnected at the cell end. When you do so, take care that they do not touch any of the other cell connections.
When you are measuring very small currents, you may find that a metal enclosure completely surrounding your cell (a Faraday shield) will significantly lower the current noise caused by AC pickup at the cell. Connect this Faraday shield to both black cell leads.
You can add an alligator clip (provided) to each of the cell connections. The alligator clips simply slide onto the banana plugs. If you need to permanently change the terminations on your cell cable, feel free to remove the existing terminations and replace them with your desired termination.
In a few cases, you may find that you need to modify the cell cable or make a special-purpose cable. By far the easiest changes involve modifying a standard cell cable. If you decide to make your own cables, we recommend that you use coaxial cable for all the cell connections. Connect the shield of the coaxial cable to the pin recommended in Table B-1 on the potentiostat end and leave the shield open on the cell end.
Cell cables longer than 3 meters may result in degraded instrument performance. Increased noise and decreased stability both can occur. However, with most cells, the instrument will work acceptably with an extended cell cable, so our advice is going ahead and try it.

Potentiostat Connector
The IMX8 potentiostat connector is used to connect the IMX8 to the system potentiostat. The pin-layout of this connector is very similar to that of the cell connectors.

Table B-3
Potentiostat Connector

Pin Name Use
1 Working Sense (WS) Only needed if the system potentiostat uses a

differential electrometer. Connect this pin to the negative input of the potentiostat’s differential electrometer. This pin can be left open when the IMX8 is used with a potentiostat that uses a single-ended electrometer.
2| WS Shield| The shield for the working sense input. Connect to the shield of the potentiostat’s working sense lead (if any).
3| Working Electrode (WE)| Connected to the potentiostat’s working electrode input.
4| WE Shield| The shield for the working electrode input. Connect to the shield of the potentiostat’s working sense lead.
5| Ground| The system potentiostat’s (floating) ground.
6| Reference Electrode (REF)| Connect to the reference electrode input on the system potentiostat.
7| REF Shield| The shield for the reference electrode cable. Connect to the shield of the potentiostat’s reference lead
8| Counter Sense (CS)| Connect to the potentiostat’s counter sense input (if it has one).
9| Counter Electrode (CE)| Connected to the potentiostat’s counter electrode lead. Shield this wire with the potentiostat ground.

Connecting the IMX8 to Third-party Potentiostats
The cable required for connection to third-party potentiostats varies with the potentiostat. The following suggestions may help you in constructing the cable.

  • In many cases the easiest way to build the cell cable is to modify a cell cable supplied by the potentiostat’s manufacturer. Spare cables are generally available.
  • Connect each of the electrode connections to its corresponding pin on the IMX8 Potentiostat connector. For example, the potentiostat’s working-electrode lead is connected to the IMX8 connector’s working electrode pin.
  • Most commercial potentiostats use shielded leads for all of the cell connections. If the potentiostat uses shielded wire, use shielded wire for your custom cable. Connect the pins labeled shield (Working Electrode Shield, Reference Shield, etc.) on the IMX8 connector to the corresponding potentiostat cable shield. If there is no shield on the potentiostat’s cell lead, you can leave the corresponding shield pin disconnected.
  • If the potentiostat does not have a working electrode sense lead, you can leave the working sense pin on the IMX8 open.
  • You can ignore the counter sense pin on the IMX8 connector (unless the cable is intended for use with a Gamry Instruments potentiostat). Leave this pin disconnected.

If an Auxiliary A/D connection is required, connect the IMX8’s Potentiostat Aux A/D Input connector to the corresponding connector on the potentiostat.

Appendix C: Certifications

Redefining Electrochemical Measurement

Declaration of Conformity: No. DOC-2022-CE-IMX8
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name and Location:
Gamry Instruments
734 Louis Drive
Warminster, PA 18974
USA

This declaration is for the Gamry Instruments product model: IMX8 Multiplexer
The declaration is based upon compliance with the following directives:
• EMC Directive 2014/30/EU

The declaration is based upon product compliance with the following standards as defined in report number R-3542P-1 from Retlif Testing Laboratories for EMC testing and analysis and report number RET 17514-01 from Washington Laboratories, Ltd. for safety analysis.

EMC Standards Title Class/ Criteria
EN 61326-1:2013; Clause 7.2 Emissions requirements B
EN 61326-1:2013; Table 1 All methods except Radiated Immunity
EN 61326-1:2013; Table 3 Radiated Immunity only
General Safety Standard Title
--- ---

IEC 61010-1:2010
IEC 61010-1:2010/AMD1:2016| Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 1: General requirements

Signature
Dr. Gregory A. Martinchek, PhD
Title: President

July 20, 2022
Date
Formal signed declaration is on file at Gamry, Inc.

Certificate of Conformance

Declaration of Conformity: No. DOC-2022-UKCA-IMX8

According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name and Location:
Gamry Instruments
734 Louis Drive
Warminster, PA 18974
USA

This declaration is for the Gamry Instruments product model: IMX8 Multiplexer
The declaration is based upon compliance with the following directives:
• EMC Directive 2014/30/EU
The declaration is based upon product compliance with the following standards as defined in report number R-3542P-1 from Retlif Testing Laboratories for EMC testing and analysis and report number RET 17514-01 from Washington Laboratories, Ltd. for safety analysis.

EMC Standards Title Class/ Criteria
EN 61326-1:2013; Clause 7.2 Emissions requirements B
EN 61326-1:2013; Table 1 All methods except Radiated Immunity
EN 61326-1:2013; Table 3 Radiated Immunity only
General Safety Standard Title
--- ---

IEC 61010-1:2010
IEC 61010-1:2010/AMD1:2016| Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 1: General requirements

Signature
Dr. Gregory A. Martinchek, PhD
Title: President
July 20, 2022
Date
Formal signed declaration is on file at Gamry, Inc.

Appendix D: Comprehensive Index

AC adapter, 7, 8
AC line cord, 8
AC mains, 8
active cell, 14, 15
Auto Update All, 25
CE Compliance, 12, 21
cell, 15, 16, 17
cell cable, 14, 16, 17, 20, 36, 38
Cell Cable connector, 20, 35
channel, 15, 16, 33, 35, 36
channel schematic, 15
cleaning, 11
Client Portal, 21
Compliance Current, 33
Compliance Voltage, 33
computer, 14, 15, 16, 26, 28
computer requirements, 21
contact, 8
contract engineering, 3
conventions
notational, 17
Counter Electrode, 17, 36, 38
Counter Sense, 36, 38
current carrying, 16
current switching, 16
Declaration of Conformity, 39, 41
earth ground, 8, 17, 31, 36
electrical transients, 11
enclosed space, 10
environmental limits, 10, 34
environmental stress, 10
Faraday Shield, 31, 36, 37
firmware update, 24
floating ground, 31, 35, 36, 38
floating ground binding post, 9
Framework software, 19, 23, 26, 28
Gamry Framework, 13
GIM, 24, 25, 26
ground, 17, 35, 36, 38
grounding, 9
Help system, 13
inactive cell, 14, 15, 16
inactive mode, 15, 16, 17
inspection, 7, 19| Instrument Manager, 24, 26
interface cable, 20
line cord, 8, 9
local potentiostat, 15, 17, 33
noise, 37
open circuit, 17
operation, 10
pairing, 26, 27
pin-layout, 20, 21, 35, 37
potentiostat, 13, 14, 15, 16, 17, 26, 28, 35, 36, 37,
38
Potentiostat connector, 20, 35, 37, 38
Power, 34
power brick, 7, 8
Power LED, 24
power line transient, 11
radio frequency, 11
Reference Electrode, 17, 36, 38
Resolution, 33
RFI, 11
RoHS Compliance, 12
safety, 7
Select and Update, 26
service, 11
service contract, 3
shielding against noise, 9
specifications, 33
static electricity, 11
storage, 10
support, 3, 8, 10, 12, 20, 26
system potentiostat, 14, 15
telephone assistance, 3
temperature, 10
USB, 14
USB cable, 21
USB LED, 22
ventilation, 10, 19
visual inspection, 19
Warranty, 3
Windows, 4
Working Electrode, 17, 36, 38
Working Sense, 36, 38
ZRA, 13, 15, 17, 36
---|---

Copyright © 2024 Gamry Instruments, Inc.