INSTRUMENTS IMX8 Electrochemical Multiplexer
Owner’s Manual
Copyright 0 1991-2022
Gamry Instruments, Inc.
Revision 7 December
30,2020 988.00027
If You Have Problems
Please visit our service and support page at www.gamry.com/service-
support/. This page contains
information on installation, software updates, and training. It also contains
links to the latest available documentation. If you are unable to locate the
information you need from our website, contact us via e-mail using the link
provided on our website. Alternatively, you can contact us one of the
following ways:
Internet www.gamty.com/service-support/
Telephone 215-682-9330 9:00 AM-5:00 PM
US Eastern Standard Time 877-367-4267
Toll-free US & Canada Only
Please have your instrument model and serial numbers available, as well as any
applicable software and firmware revisions.
If you have problems in the installation or use of a system containing an IMXB
Electrochemical Multiplexer, please call from a phone next to your computer,
where you can type and read the monitor while talking to us.
We will be happy to provide a reasonable level of free support for registered
users of the IMX8 Electrochemical Multiplexer. Reasonable support includes
telephone assistance covering the normal installation and use of an IMX8
Electrochemical Multiplexer connected to a Windows-10 or newer computer.
A service contract that extends both the hardware warranty and software update
period is available at an additional charge. Software updates do nor include
software enhancements offered to our customers at additional cost.
Enhancements to the IMX8 Electrochemical Multiplexer and Gamry’s standard
applications software that require significant engineering time on our part
can be performed on a contract basis. Contact us with your requirements.
Limited Warranty
Gamry Instruments, Inc. warrants to the original user of this product that it
shall be free of defects resulting from faulty manufacture of the product or
its components for a period of two years from the original shipment date of
your purchase. Gamry Instruments, Inc. makes no warranties regarding either
the satisfactory performance of the IMX8 Electrochemical Multiplexer including
the software provided with this product, or the fitness of the product for any
particular purpose. The remedy for breach of this Limited Warranty shall be
limited solely to repair or replacement, as determined by Gamry Instruments,
Inc., and shall not include other damages. Gamry Instruments, Inc. reserves
the right to make revisions to the system at any time without incurring any
obligation to install the same on systems previously purchased. All system
specifications are subject to change without notice.
There are no warranties that extend beyond the description herein. This
warranty is in lieu of and excludes any and all other warranties or
representations, expressed, implied or statutory, including merchantability
and fitness, as well as any and all other obligations or liabilities of Gamry
Instruments, Inc; including but not limited to, special or consequential
damages.
This Limited Warranty gives you specific legal rights and you may have others
that vary from state to state. Some states do not allow for the exclusion of
incidental or consequential damages. No person, firm, or corporation is
authorized to assume for Gamry Instruments, Inc., any additional obligation or
liability not expressly provided herein except in writing duly executed by an
officer of Gamry Instruments, Inc.
Disclaimers
Gamry Instruments, Inc. cannot guarantee that the IMXB will work with all
computer systems, operating systems, and third-party potentiostat hardware.
The information in this manual has been carefully checked and is believed to
be accurate as of the time of printing. However, Gamry Instruments, Inc.
assumes no responsibility for errors that might appear.
Copyrights
IMXB Operator’s Manual copyright 0 2021, Gamry Instruments, Inc. Gamry
Framework copyright 0 2021, Gamry Instruments, Inc. IMX8IM, Gamry Framework,
and GamryTm are trademarks of Gamry Instruments, Inc. Microsoft and Windows
have registered trademarks of Microsoft Corporation. No part of this document
may be copied or reproduced in any form without the prior written consent of
Gamry Instruments, Inc.
Chapter 1: Safety Considerations
Your IMX8 Electrochemical Multiplexer is supplied in a safe condition. This
chapter of the IMX8 Operator’s Manual contains some information and warnings
that you must follow to ensure the continued safe operation of the IMX8.
An inspection
When you receive your IMXB Electrochemical Multiplexer, inspect it for
evidence of shipping damage. If you note any damage, please notify Gamry
Instruments Inc. and the shipping carrier immediately. Save the shipping
container for possible inspection by the carrier.
Warning:
An IMX8 damaged in shipment can be a safety hazard. Do not operate a damaged
apparatus until a qualified service technician has verified its safety. Tag a
damaged IMXB to indicate that it could be a safety hazard.
Protective Grounding and Product Safety
As defined in IEC Publication 348, Safety Requirements for Electronic
Measuring Apparatus, the IMX8 is a Class I apparatus. Class I apparatus is
only safe from electrical-shock hazards if the case of the apparatus is
connected to protective earth ground. In the IMXB this protective-ground
connection is made via the ground prong in the AC line cord. When you use the
IMX8 with an approved line cord, the connection to the protective earth ground
is automatically made prior to making any power connections.
Warning:
Do not negate the protection of the IMX8’s earth ground by any means. Do not
use the IMXB with a two-wire extension cord, with an adapter that does not
provide for protective grounding, or with an electrical outlet that is not
properly wired with a protective earth ground. If the protective ground is not
properly connected, it creates a safety hazard, which could result in personal
injury or death.
The IMX8 is supplied with a line cord suitable for use in the United States.
In other countries, you may have to replace the line cord with one suitable
for your type of electrical outlet. You must always use a line cord with a CEE
22 Standard V female connector on the instrument end of the cable. This is the
same connector used on the US standard line cord supplied with your IMX8.
Warning:
If you replace the line cord, you must use a line cord with the same polarity
as that supplied with the IMX8. An improper line cord can create a safety
hazard, which could result in personal injury or death.
The wiring polarity of a properly wired connector is shown in Table 1.1 for
both US line cords and European line cords that follow the “harmonized” wiring
convention.
Table 1-1 Line Cord Polarities and Colors
| Line| Neutral| Earth Ground
---|---|---|---
US| Black| White| Green
European| Brown| Light Blue| Green/Yellow
If you have any doubts about the line cord for use with your IMX8, please
contact a qualified electrician or instrument service technician for
assistance. He or she can perform a simple continuity check that can verify
the connection of the IMX8 chassis to earth and thereby check the safety of
your IMX8 installation.
Ventilation
Your IMX8 Electrochemical Multiplexer was designed to operate at ambient
temperatures between 0°C and 40°C.
Caution:
Be careful when operating the IMX8 in an enclosed space (such as an enclosed
relay rack or NEMA enclosure). The temperature within the enclosure must not
exceed 40°C. You may need to provide ventilation holes or even forced air
cooling for the enclosure if the temperature rises excessively.
Defects and Abnormal Stresses
You should treat your IMX8 Electrochemical Multiplexer as potentially
hazardous if any of the following is true of the unit:
- It shows visible damage.
- It does not operate properly.
- II has been stored for a long period of time under unfavorable conditions.
- It has been dropped or subjected to severe transport stress.
- It has been subjected to environmental stress (corrosive atmosphere, fire, etc.).
Do not use your IMX8 or any other apparatus if you think it could be
hazardous. Have it checked by qualified service personnel.
There are limited conditions on the storage, shipping, and operation of this
equipment. The IMXB is not designed for outdoor use.
Storage
Ambient Temperature……………—20°C to 60°C
Relative Humidity…………………………Maximum 90% non-condensing
Shipping
Same as storage plus
Acceleration…………………………….Maximum 30 G
Operation
Ambient Temperature……………………….. 10°C to 40°C
Relative Humidity……………………………… Maximum 90% non-condensing
Cleaning
Disconnect the IMXB from all power sources prior to cleaning.
Use a cloth lightly dampened with either clean water or water containing a
mild detergent, to clean the outside of the IMX8 enclosure. Alternatively, you
may use isopropyl alcohol. Do not use a wet rag or allow fluid to enter the
IMX8 enclosure. Do not immerse the IMX8 in any type of cleaning fluid
(including water). Do not use any abrasive cleaners.
Service
Your IMX8 Electrochemical Multiplexer has no user-serviceable parts inside.
Refer all services to a qualified service technician.
Warning:
Never operate the IMX8 with any cover or panel on the chassis open. Dangerous
AC line voltages are present at several points within the IMX8 chassis,
including PC board traces. Always remove the power connection before opening
the IMX8 case.
RFI Warning
Your IMX8 Electrochemical Multiplexer generates, uses, and can radiate radio-
frequency energy. The radiated levels are low enough that the IMX8 should
present no interference problem in most industrial laboratory environments.
The IMX8 is likely to cause radio-frequency interference if operated in a
residential environment.
Electrical Transient Sensitivity
Your IMX8 Electrochemical Multiplexer was designed to offer reasonable
immunity from electrical transients. However, in severe cases, the IMX8 could
malfunction or even suffer damage from electrical transients. If you are
having problems in this regard, the following steps may help:
If the problem is static electricity (sparks are apparent when you touch the
IMX8):
-
Placing your IMX8 on a static-control work surface may help. Static-control work surfaces are now generally available from computer-supply houses and electronics-tool suppliers. An antistatic floor mat may also help, particularly if a carpet is involved in generating static electricity.
-
Air ionizers or even simple air humidifiers can reduce the voltage available in static discharges.
If the problem is AC power-line transients (often from large electrical motors near the IMX8): -
Try plugging your IMX8 into a different AC-power branch circuit.
-
Plug your IMX8 into a power-line surge suppressor Inexpensive surge suppressors are now generally available because of their use with computer equipment.
Contact Gamry Instruments, Inc. if these measures do not solve the problem.
CE Compliance
The European Community has instituted standards limiting radio-frequency
interference from electronic devices and mandating several safety
requirements. Gamry Instruments, Inc. has modified its instruments, including
the IMX8, to comply with these standards.
The relevant CE regulations include EN55022 Class B and EN60950.
Chapter 2: Introduction
The IMX8 Electrochemical Multiplexer was developed to increase the throughput
of your electrochemical testing program. It allows one
potentiostat/galvanostat/ZRA to drive up to eight electrochemical cells. Under
the control of an external computer, the IMX8 connects one cell at a time to
the potentiostat, which makes an electrochemical measurement on that cell.
The IMX8 was specially designed for the needs of electrochemists. Extra
features such as sense line and shield-switching, and control of inactive
cells separate the IMX8—designed specifically for electrochemists—from other
general-purpose switching devices.
This chapter of the IMX8 Operator’s Manual provides an overview of the
operation of the IMX8. It discusses the following topics:
- Operation overview
- Switching and inactive cell control
- Communications overview
The material in this chapter is only an overview. If you are using your IMX8 with Gamry Instruments software this overview should give you sufficient understanding to make intelligent use of your IMX8.
Operation Overview
Figure 2-1 shows a “typical” electrochemical test system built around an IMX8
Electrochemical Multiplexer. Only four cells are shown in this figure,
although up to eight cells can be connected simultaneously. The label on the
block marked as a potentiostat is slightly misleading. Most potentiostats can
also be used as a galvanostat or as a ZRA (zero resistance ammeter) in an IMX8
system.
In the description of this figure, keywords are in bold type and are defined
in the next section.
Figure 2-1 A Typical Electrochemical Test System Using the IMX8
System computer
The system computer is a critical component of the system. It controls the
IMX8 by sending it simple commands through a USB connection.
The most important function of these commands is to select which cell is
active (connected to the system potentiostat). You can think of an IMX8 as a
complex switch. One of the eight-cell cables on the IMX8 is switched so that
it connects to the potentiostat’s cell cable. All of the wires (including
sense leads and shields) in the cell cable are switched. Other commands select
the IMXB’s control mode for the cells that are inactive (not presently
connected to the potentiostat).
The system computer also controls the system potentiostat. When a cell is
connected to the system potentiostat cell leads, the computer causes a test to
be run on that cell. The test can be as simple as a single current or
potential measurement or as complex as a cyclic scan. The computer software
and system potentiostat are responsible for taking the measurement and storing
it.
In a typical IMX8 experiment, the system computer cycles through the cells.
Each cell is connected in turn to the system potentiostat and used to collect
a measurement. Usually the computer delays at the end of each cycle through
the cells so that the measurements on each cell are separated by a fixed time
period.
Keyword Definitions
The following key terms are used throughout this manual.
Active Cell
The active cell is the cell (if any) currently connected to the system
potentiostat. Any electrochemical measurements are made on the active cell.
Cell
The term cell is used to describe either an electrochemical cell or the
switches needed to connect an electrochemical cell to the system potentiostat.
Channel
An IMX8 contains eight channels. A channel includes cell-switching, a local
potentiostat, and a D/A converter. The term cell is used when only the cell-
switching portion of a channel is being discussed.
Inactive Cell
The term inactive cell is any cell not currently connected to the system
potentiostat. Inactive is a slightly misleading term since an inactive cell
can actually be potentiostat by the IMX8’s local potentiostat on that channel.
Inactive Mode
Inactive cells can be open, connected to a local potentiostat, or shorted. The
term inactive mode describes this choice of control mode for inactive cells.
Local Potentiostat
A simple low-current potentiostat that maintains potential control on inactive
cells. Unlike the system potentiostat, the local potentiostat is incapable of
making electrochemical measurements.
System Computer
The computer is responsible for coordinating the operation of an IMX8-based
electrochemical test system, not to he confused with the very simple micro-
controller buried within the IMX8.
System Potcntiostat
This is the potentiostat responsible for making electrochemical measurements
on the cells used in the experiment. The system potentiostat is typically a
full-featured, computer-controlled instrument.
Switching and Inactive Cell Control
In the Operation Overview, we said you can think of the IMXB as a complex
switch. This section of the manual describes the IMXB-switching in some
detail. In addition, control modes for inactive cells are discussed.
Each of the eight IMXB channels is made up of an independent set of switches
and control circuits. Figure 2-2 is a simplified schematic diagram for one
channel of the IMX8. Switches connected with dotted lines are always switched
together and are therefore treated as one multi-pole switch. All the switches
are under the control of the system computer. Shield relays are not shown.
The cell cable from the system potentiostat is connected at the left side of
the schematic. The system-potentiostat cell cable is also connected to all of
the other channels. The system computer must ensure that only one cell at a
time is connected to the system potentiostat.
Figure 2-2 Simplified Schematic of One IMX8 Channel Switch 51 controls
which cell is active. 53 and S4 control the inactive mode of the cell on this
channel.
Looking at the schematic, you can see that switch Si connects this channel’s
cell to the system potentiostat. An active channel always has 53 and 54 open.
The relays in S1 are rated to carry 2 A and switch 1 A. You can normally use
the IMX8 with potentiostats rated at up to 100 V of output compliance and 2 A
of output current.
To avoid transients, always turn off the cell at the system potentiostat prior
to switching S1. The IMX8 switches all its relays nearly simultaneously (even
those in different channels), but there is still about 1 millisecond in which
the relay pattern in the IMX8 is in neither the initial or final
configuration. Switching transients as the IMX8 switches channels with a
“live” potentiostat can be particularly severe because the system
potentiostat’s feedback goes through one of the S1 relays. To avoid the
problem, turn off the cell at the system potentiostat before switching
channels.
The switch labeled 52 connects one of eight pairs of uncommitted contacts to a
common input/output connector. The system computer must make sure that S2 is
closed for only one channel at a time. Switch S2 is controlled independently
of S1, although they are often switched together.
You have a choice of three different control schemes for inactive cells. These
are tabulated in Table 2-1 along with the active mode.
Table 2-1 IMXB Control Modes
Mode | S1 | S3 | 54 |
---|
Active
Open
Local potentiostat
Shorted| Closed
Open
Open
Open| Open
Open
Closed
Open| Open
Open
Open
Closed
The inactive control mode labeled “open” is the simplest. In this mode the
cell is disconnected and isolated from both the system potentiostat and the
IMX8. No cell current can flow between any electrodes. This is the most common
inactive mode in corrosion studies, where the cell behavior at open circuit is
of interest.
The inactive control mode labeled “local potentiostat” is used to maintain
potential control between readings made with the system potentiostat. If you
know something about operational amplifiers and potentiostats, you may
recognize that the operational amplifier in Figure 2-2 is wired as a simple
potentiostat. With S3 closed, the working electrode is connected to the ground
on the IMX8 circuit board and the counter electrode is connected to the output
of an operational amplifier. The feedback for this amplifier is taken from the
cell’s reference electrode. This mode can be used whenever it is important to
maintain potential control between readings. One experimental example is tests
on amperometric chemical sensors.
Unlike a typical system potentiostat, the local potentiostat is very limited
in capability. The amplifier in this local potentiostat is not a large power
amplifier. Its output is typically capable of generating ±30 mA at up to ±5 V.
The local potentiostat has no cell switch other than S3, and it cannot measure
current.
Notice that the reference electrode is always connected to the local
potentiostat amplifier. You must make sure that the reference electrode
potential does not go beyond ±5 volts with respect to the IMX8 chassis
voltage. This is true in all operating modes including the active mode.
Caution: You can damage the local potentiostat’s amplifier if the
reference electrode potential exceeds ±24 volts versus the IMX8 chassis.
The final inactive mode is labeled “shorted” in Table 2-1. In this mode a
relay shorts the working and counter electrodes. This mode was added so that
the IMX8 could be used in galvanic corrosion studies. In a galvanic corrosion
experiment, the current flow between two electrically connected metals is
measured. The device that makes this measurement is called a zero-resistance
ammeter (ZRA).
In the shorted inactive mode the two metals are connected, even when the
system ZRA is measuring other channels. The shorted mode is subject to some
errors at high currents because of the resistance in relays, wires, and
connectors. With a 1.5 m Gamry Instruments cell cable, the resistance between
the two electrodes should be less than 0.2 0. A good ZRA is much better in
this regard, with electrode-to-electrode impedances in the low ma.
Chapter 3: Installation
This chapter of the IMX8 Electrochemical Multiplexer Operator’s Manual covers
the normal installation of the IMX8. Both installation as part of a Gamry
Framework-based electrochemical-measurement system and installation as a
component in a user-configured system are discussed.
Initial Visual Inspection
After you remove your IMX8 from its shipping carton, check it for any signs of
shipping damage. If you note any damage, please notify Gamry Instruments, Inc.
and the shipping carrier immediately. Save the shipping container for possible
inspection by the carrier.
Warning: An IMX8 damaged in shipment can be a safety hazard. Do not
operate a damaged apparatus until a qualified service technician has verified
its safety. Tag a damaged IMX8 to indicate that it could be a safety hazard.
Physical Location
Place your IMX8 on a normal workbench surface. The IMX8 is not restricted
to operation in a horizontal position. You can operate it on its side, or even
upside-down. It may even be stacked on top of a Gamry Instruments
potentiostat. To the right is an example of stacking the IMX8 on top of your
potentiostat.
Power cord
Caution:
Never substitute a generic 25-to-9-pin cable designed for computer
interconnection for this cable. The Gamry Instruments’ cable includes
individual shielding for several signals in this cable. The use of a generic
cable is likely to create excessive noise and system oscillation.
The Reference 600, 600+, 620, or Interface 1000 or 1010 can identify the cable
plugged into its cell connector. It detects the presence of an IMX8 in your
system. The Reference 600, 600+, 620, or Interface 1000 or 1010 cannot
identify the cables used to connect the eight individual cells to the IMX8.
EIS performance with an IMX8 and a Reference 600, 600+, 620 or Interface 1000
or 1010 is optimal when all the cell cables are 1.5 m cables. All other cables
may create errors in phase and magnitude.
Connection to Gamry Instruments Reference 3000
Connect the dual ends of the Y-cable (part number 985-00108) to the
appropriate jacks on the front of the Reference 3000. Attach the single end of
the Y-cable to the 985-00080 (PCS-to-IMX8 cable). Attach the other end of the
985-00080 cable to the Pstat connector on the front of the IMX8.
Cell Cables
Each cell to be tested in an experimental run requires a cell cable. Newer
IMX8 multiplexers are shipped with eight-cell cables.
When you are using the IMX8 with a Gamry Instruments potentiostat, cell-cable
connections to the cell are made in the manner described in the potentiostat’s
Operators Manual.
When you are using the IMX8 with a Gamry Instruments potentiostat, cell-cable
connections to the cell are made in the manner described in the potentiostat’s
Operators Manual.
1.5 m, 4 m, and 9 m cables are available with banana-plug/alligator-clip
termination. Specialized cables are also available, including cables designed
for connection to standard three-terminal corrosion-monitoring probes. Contact
us for more information.
If you need to design a cell cable for your specific needs, refer to the pin-
out of the IMXB cell cable in Appendix C of this manual.
Software Installation
The installation process for the Gamry Framework also installs Gamry’s
Framework application packages and installs any multiplexed scripts that are
available for those applications. Test the Communications and Run Your
Experiment
- Be sure your host computer is on, the multiplexer is on, and your potentiostat is on.
- Run the Gamry Framework”‘ software.
If you are running AC experiments (e.g., EIS), you must run AC calibration of
your potentiostat with the multiplexer connected. For all DC experiments
calibration is not needed.
To Change the Multiplexer COM Port
1. Be sure your host computer is on, the multiplexer is on, and your
potentiostat is on. 2. Run the Gamry Framework software.
Appendix A: IMX8 Specifications
Specifications apply at 25°C, 116 V AC power, and operation with a Gamry
Instruments potentiostat unless otherwise noted.
Channel Characteristics
Modes
Channel switching time.
Cell current
Input voltage (operating)
Input voltage (no damage)
Counter electrode voltage| Active, Off, Local, Shorted
<10 ms
2A
1A
±12 V
±24 V
±120 V| Maximum carried
Maximum switched
Measured versus IMX8 chassis ground.
Measured versus IMX8 chassis ground.
Measured versus IMX8 chassis ground.
---|---|---
C hannel Isolation
Current leakage………………. <2 nA maximum
Impedance ………………………..>500 MO……………………. <20 pF
Live channel pin from any source. quiv. impedance to IMX8 chassis ground.
Local Potentiostat
Compliance-current
Compliance-voltage
Applied E Range
Resolution
Zero-offset error
Gain error
Reference input current| ±20 mA
±11 V
±5.12 V
2.5 mV/bit
<4 mV typical
<10 mV max
<0.2% <50 pA| 500 iZ load
1 ko load
Working versus Reference
0 V output into
10 kCI load
10 ko load
---|---|---
Note: Applied voltage error is the sum of zero-off set error and gain error.
Power Input
Input voltage ranges………………..100-240 V RMS………………. Selectable
Allowed voltage variation……………..±10%…………………………All ranges
Power…………..<50 VA……………………………………………………..Worst case r-
Environmental
Operating temperature……………………………. 0 to 40°C
Humidity ……………………………. 0 to 90% noncondensing
Appendix B: Grounding and Floating Operation
Overview As shipped from the factory, the IMX8 has not been configured for use
with “floating” potentiostats. As you can see from the specifications, in
normal operation all the electrodes except the counter electrode are limited
to potentials ±12 V versus the IMX8 chassis ground. The IMX8 chassis is
connected to the protective earth line in the power cord, so the potential is
limited to ±12 V versus earth ground.
This limitation prevents the IMX8 from working with cells where the counter
electrode must be earth-grounded. Some autoclaves use the earth-grounded wall
of the autoclave as the counter electrode. You should be able to use the IMX8
with cells that have an earth-grounded working electrode, as long as the
potentials of the IMXB’s earth ground and the cell’s earth ground are
reasonably close together.
The ground pin on the IMXB’s potentiostat connector is not connected to the
IMX8’s chassis ground. It is quite possible to have a several-volt difference
between these grounds without any adverse effects on measurements made through
the IMX8. Larger ground differences are likely to cause a violation of the ±12
V electrode-to-chassis-ground limitation.
Warning: It is very dangerous to defeat the IMXB’s connection to the
protective earth ground. The earth-ground protection is there for operator
safety in case of a severe power-supply problem with the instrument. Even if
you do break the earth-ground connection in the AC power cord, the USB
connection usually re-establishes an earth ground.
If you really need to multiplex cells with earth-grounded counter electrodes,
see the next section. It tells how you can move jumpers in the IMX8 to remove
the ±12 volt limitation. Unfortunately, this change also disables the local
potentiostats in the IMX8.
Disabling the IMX8 Local Potentiostats [Needs more input from R &D]
Under many circumstances, the IMX8 cannot operate with cells that contain
earth-grounded electrodes. See the Overview above. The reference-electrode
connection in the IMX8’s local potentiostats is the cause of the problem.
Fortunately, you can disable the local potentiostats, thereby eliminating the
problem with floating operation. Disable the local potentiostats by moving
eight jumpers inside the IMXB chassis. Follow these instructions to disable
the IMXB’s local potentiostats:
-
First, remove the IMXB power cord from the IMX8’s rear-panel power-entry module.
Warning: Dangerous voltages are exposed if the IMXB chassis is open with the power cord connected to a source of AC power. To prevent a shock hazard that could result in personal injury or even death, always unplug the IMXB from its AC power source before removing any panel or cover. -
The IMXB top cover is held in place by one screw at the middle rear of the cover. Remove this screw, making sure that you do not lose it.
-
Slide the IMXB top cover backward and off the chassis. You may need to loosen (do not remove!) the set of screws on one side of the IMX8 if the cover does not slide off easily
-
All the IMXB circuitry is contained on one large printed circuit board. All locations and directions are given assuming that you are looking over the IMXB’s front panel towards the IMX8’s rear panel. The local potentiostat jumpers are near the back of the instrument. There are eight jumpers, one for each channel. They are located in a gap within the “field” of relays on the IMXB printed circuit board. There are also two jumpers on the right side of the IMX8’s board.
Caution: These jumpers are near the IMXB’s microprocessor, a large 40-pin integrated circuit. Do not change the position of these jumpers.
Each jumper is located on one side of a set of three pins. If the jumper is on the leftmost pair of pins, the local potentiostat is enabled. If the jumper is on the rightmost pair of pins, the local potentiostat is disabled. -
If you need a local potentiostat operation and don’t need a floating operation, move the jumper on each channel to the left side. If you need a floating operation and don’t need the local potentiostats, move the jumper to the right side.
You can enable the local potentiostat on some channels and disable it on others. -
Replace the IMX8 top cover and its retaining screw.
-
Replace the IMX8 power line cord.
-
If you have disabled some or all of the IMX8’s local potentiostats, we recommend that you label the outside of your IMX8 to show that you have done this.
Comprehensive Index
985-00080, 2, 3
985-00108,3
AC line cord, 1-1
active cell, 2-3, 2-4
autoclaves, 3-1
CE Certificate, 3-1
CE Compliance, 1-4
cell definition, 2-4
cell cable, 3
Change the Multiplexer COM Port, 5
channel definition, 2-4
schematic, 2.5
cleaning, 1-3
contract engineering, i CSA Certificate, 3-1
current carrying, 2-5
switching, 2-5
Disabling Local Potentiostats, 3.1
earth ground, 3-1
electrical transients, 1-3
environmental specifications, 1
environmental stress, 1-2
Experiment, 4, 6
Floating Operation, B-1
floating potentiostats, 3-1
Framework software, 3, 5
ground, B-1
host computer, 3, 5
inactive definition, 2-4
modes, 2-6
inactive cell, 2-3
inactive mode, 2-4
installation, 1
Interface 1000, 2
Interface 1010, 2, 3| local potentiostat, 2-4, 2-6, 1
Multiplexer Auto Detect, 4
Multiplexer Auto
Detect window, 4
Multiplexer WM Port, 6
OK button, 4, 5, 6
open, 2-6
operation, 1-3
PC5 to ECMS cable, 3
potentiostat, iii, 2.1, 2-2, 2-3, 24, 2-5, 2.6, 2, 3, 5, 3-1, 3-2
POTENTIOSTAT
connector, 3
power line transient, 1-4
protective grounding, 1-1
Reference 3000, 3
Reference 600+, 2, 3
reference electrode, 2.6 RFI, 1-3
safety, 1-1
schematic channel, 2-5
service, 1-3
service contract, i shipping, 1-3
shorted inactive mode, 2-6
specifications, 1
static electricity, 1 -4
storage, 1-3
support, i system computer, 2-3
definition, 24
system potentiostat, 2-3, 2-4
telephone assistance, i Utilities, 4, 6
ventilation, 1-2
Warranty, ii Which Multiplexer
Do You Want To Configure?
window, 6
Y-cable’ 3
ZRA,2-2,2-6
---|---