Abbott Architect SARS-CoV-2 lgG User Guide

June 7, 2024
Abbott

Abbott Architect SARS-CoV-2 lgG User Guide

Abbott Architect SARS-CoV-2 lgG User Guide

For use under an Emergency Use Authorization (EUA) Only
Prescription Use only.
For In Vitro Diagnostic Use Only. Instructions must be carefully followed. Reliability of assay results cannot be guaranteed if there are any deviations from these instructions. For laboratory professional use only.

NAME

SARS-CoV-2 IgG (also referred to as CoV-2 IgG)

INTENDED USE

The SARS-CoV-2 IgG assay is a chemiluminescent microparticle immunoassay (CMIA) intended for the qualitative detection of IgG antibodies to SARS-CoV-2 in human serum, serum separator tube and plasma (ACD, CPD, CPDA-1, dipotassium EDTA, dipotassium EDTA, lithium heparin, lithium heparin separator tube, sodium citrate, sodium heparin). The SARS-CoV-2 IgG assay is intended for use as an aid in identifying individuals with an adaptive immune response to SARS- CoV-2, indicating recent or prior infection. At this time, it is unknown for how long antibodies persist following infection and if the presence of antibodies confers protective immunity. The SARSCoV-2 IgG assay should not be used to diagnose acute SARS-CoV-2 infection. Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments (CLIA) of 1988, 42 U.S.C 263a, to perform moderate or high complexity test. Results are for the detection of SARS CoV-2 antibodies. IgG antibodies to SARS-CoV-2 are generally detectable in blood several days after initial infection, although the duration of time antibodies are present post-infection is not well characterized. Individuals may have detectable virus present for several weeks following seroconversion. Laboratories within the United States and its territories are required to report all positive results to the appropriate public health authorities. The sensitivity of SARS-CoV-2 IgG early after infection is unknown. Negative results do not preclude acute SARS-CoV-2 infection. If acute infection is suspected, direct testing for SARS-CoV-2 is necessary. False positive results for SARS-CoV-2 IgG assay may occur due to cross-reactivity from pre-existing antibodies or other possible causes.
The SARS-CoV-2 IgG assay is only for use under the Food and Drug Administration’s Emergency Use Authorization.

SUMMARY AND EXPLANATION OF THE TEST

The SARS-CoV-2 IgG assay is designed to detect immunoglobulin class G (IgG) antibodies to the nucleocapsid protein of SARS-CoV-2 in serum and plasma from individuals who are suspected to have had coronavirus disease (COVID-19) or in serum and plasma of subjects that may have been infected by SARS-CoV-2. COVID-19 is defined as illness caused by a novel coronavirus now called severe acute respiratory syndrome coronavirus 2 (SARSCoV-2, formerly called 2019-nCoV), which was first identified in December 2019 during an outbreak of respiratory illness cases in China.1 On March 11, 2020, the World Health Organization (WHO) declared COVID-19 a global pandemic.2 The incubation period of COVID-19 ranges between 1 and 14 days, with the majority of cases manifesting within 3 to 5 days. The most common symptoms of COVID-19 are fever, tiredness, dry cough, and difficulty breathing. A severe acute respiratory distress syndrome (ARDS) may develop.3 Reported case fatality rates depend on geographic location,4 age, and comorbidities.
The causative agent of COVID-19 is a beta coronavirus and belongs to a family of viruses that may cause respiratory symptoms ranging from common cold to severe pneumonia. These viruses are common in animals worldwide and may eventually transfer to humans, as has likely happened with SARS-CoV-2.1
The host immune system reacts to the infection by SARS-CoV-2 by producing specific antibodies. These antibodies have been reported to appear in serum or plasma of infected individuals after the detection of viral ribonucleic acid (RNA) in swabs5 in as early as a few days to 2 weeks after the onset of symptoms.6 Specific IgG antibodies to SARS-CoV-2 may be detectable in COVID-19 patients during the symptomatic phase of the disease after RNA is no longer detectable.5, 6 The persistence of IgG antibodies allows identification of people who have been infected in the past, and likely have recovered from the illness.7 It is unknown if IgG antibodies to SARS-CoV-2 confer immunity to infection. IgG detection and other serological assays will likely play an important role in research and surveillance.8

BIOLOGICAL PRINCIPLES OF THE PROCEDURE

This assay is an automated, two-step immunoassay for the qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology.
Sample, SARS-CoV-2 antigen coated paramagnetic microparticles, and assay diluent are combined and incubated. The IgG antibodies to SARS-CoV-2 present in the sample bind to the SARS-CoV-2 antigen coated microparticles. The mixture is washed. Anti-human IgG acridinium-labeled conjugate is added to create a reaction mixture and incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are added.
The resulting chemiluminescent reaction is measured as a relative light unit (RLU). There is a direct relationship between the amount of IgG antibodies to SARS-CoV-2 in the sample and the RLU detected by the system optics.
This relationship is reflected in the calculated Index (S/C).
The presence or absence of IgG antibodies to SARS-CoV-2 in the sample is determined by comparing the chemiluminescent RLU in the reaction to the calibrator RLU.
For additional information on system and assay technology, refer to the ARCHITECT System Operations Manual, Section 3.

REAGENTS

Kit Contents
SARS-CoV-2 IgG Reagent Kit 06R86
NOTE: Some kit sizes may not be available for use on all ARCHITECT i Systems. Please contact your local distributor.
Volumes (mL) listed in the following table indicate the volume per bottle.

Tests per kit Number of kits per box Tests per box

Abbott Architect SARS-CoV-2 lgG User Guide - REAGENTS Abbott Architect SARS-CoV-2 lgG User Guide -
REAGENTS

(M I) Purified SARS-CoV-2 recombinant antigen coated microparticles in TRIS buffer with surfactant. Minimum concentration: 0.045% solids. Preservatives: Pro CLIN 950 and sodium azido.

(CONJUGATE) Anti-human IgG (mouse, monoclonal) acridinium labeled conjugate in MES buffer with surfactant and protein (bovine) stabilizer. Minimum concentration: 4 ng/lm Preservatives: Proclaim 300 and antimicrobial agents.

(ASSAY I) TRIS buffer and detergent. Preservatives: Proclaim 950 and sodium aside.

Warnings and Precautions
For Use Under An Emergency Use Authorization Only.
This assay is only for in vitro diagnostic use under the FDA Emergency Use Authorization.

  • IVD
  • For In Vitro Diagnostic Use
  • Rx ONLY

Safety Precautions
CAUTION: This product requires the handling of human specimens. It is recommended that all human-sourced materials and all consumables contaminated with potentially infectious materials be considered potentially infectious and handled in accordance with the OSHA Standard on Bloodborne Pathogens. Biosafety Level 2 or other appropriate regional, national, and institutional biosafety practices should be used for materials that contain, are suspected of containing, or are contaminated with infectious agents.9-12

Abbott Architect SARS-CoV-2 lgG User Guide - Warnings and
PrecautionsAbbott Architect
SARS-CoV-2 lgG User Guide - Warnings and Precautions

Follow local chemical disposal regulations based on your location along with recommendations and content in the Safety Data Sheet to determine the safe disposal of this product. For the most current hazard information, see the product Safety Data Sheet. Safety Data Sheets are available at www.corelaboratory.abbott or contact your local representative. For a detailed discussion of safety precautions during system operation, refer to the ARCHITECT System Operations Manual, Section 8.

Reagent Handling

  • Reagents are shipped on wet ice.

  • Do not pool reagents within a kit or between kits.

  • Before loading the reagent kit on the system for the first time, the microparticle bottle requires mixing to resuspend microparticles that may have settled during shipment. For microparticle mixing instructions, refer to the PROCEDURE, Assay Procedure section of this package insert.

  • Septum’s MUST be used to prevent reagent evaporation and contamination and to ensure reagent integrity. Reliability of assay results cannot be guaranteed if septum’s are not used according to the instructions in this package insert. ­
    — To avoid contamination, wear clean gloves when placing a septum on an uncapped reagent bottle.
    —­ Once a septum has been placed on an open reagent bottle, do not invert the bottle as this will result in reagent leakage and may compromise assay results.
    —­ Over time, residual liquids may dry on the septum surface. These are typically dried salts and have no effect on assay efficacy.

  • When handling conjugate vials, change gloves that have contacted human serum or plasma, since introduction of human IgG will result in a neutralized conjugate. For a detailed discussion of reagent handling precautions during system operation, refer to the ARCHITECT System Operations Manual, Section 7.

Reagent Storage

  • Do not freeze.
  • Protect from light.

Abbott Architect SARS-CoV-2 lgG User Guide - Reagent
Storage

Reagents may be stored on or off the ARCHITECT I System. If reagents are removed from the system, store them at 2 to 8°C (with septum’s and replacement caps) in an upright position. For reagents stored off the system, it is recommended that they be stored in their original trays and boxes to ensure they remain upright.
For information on unloading reagents, refer to the ARCHITECT System Operations Manual, Section 5.

Indications of Reagent Deterioration
Deterioration of the reagents may be indicated when a calibration error occurs or a control value is out of the specified range. Associated test results are invalid, and samples must be retested. Assay recalibration may be necessary. For troubleshooting information, refer to the ARCHITECT System Operations Manual, Section 10.

INSTRUMENTS PROCEDURE

The SARS-CoV-2 IgG assay file must be installed on the ARCHITECT i System prior to performing the assay.
In order to prevent potential interactions, perform maintenance prior to and following the batching of SARS-CoV-2 IgG samples.

  • ARCHITECT i2000SR instrument: Perform Daily Maintenance (Operations Manual, 6041 Daily Maintenance Procedure)
  • ARCHITECT i1000SR instrument (upon availability of assay file): Perform Weekly Maintenance Procedure (Operations Manual, 6445 Pipettor/WZ Probe Cleaning Maintenance Procedure)

For detailed information on assay file installation and viewing and editing assay parameters, refer to the ARCHITECT System Operations Manual, Section 2. For information on printing assay parameters, refer to the ARCHITECT System Operations Manual, Section 5. For a detailed description of system procedures, refer to the ARCHITECT System Operations Manual.

SPECIMEN COLLECTION AND PREPARATION FOR ANALYSIS

Specimen Types
The specimen types listed below may be used with this assay.

Abbott Architect SARS-CoV-2 lgG User Guide - Specimen
Types

  • Each laboratory is responsible for following their own procedures to establish the use of additional tube or collection types.
  • Performance has not been established for the use of cadaveric specimens or the use of bodily fluids other than human serum/ plasma.
  • Liquid anticoagulants may have a dilution effect resulting in lower Index (S/C) values for individual specimens.
    The instrument does not provide the capability to verify specimen types. It is the responsibility of the operator to verify that the correct specimen types are used in the assay.

Specimen Conditions

  • Do not use:
    —­ heat-inactivated specimens
    —­ pooled specimens
    — grossly hemolyzed specimens
    —­ specimens with obvious microbial contamination
    —­ specimens with fungal growth

  • For accurate results, serum and plasma specimens should be free of fibrin, red blood cells, and other particulate matter. Serum specimens from patients receiving anticoagulant or thrombolytic therapy may contain fibrin due to incomplete clot formation.

  • To prevent cross contamination, use of disposable pipettes or pipette tips is recommended.

Preparation for Analysis

  • Follow the tube manufacturer’s processing instructions for collection tubes. Gravity separation is not sufficient for specimen preparation.

  • Specimens should be free of bubbles. Remove bubbles with an applicator stick before analysis. Use a new applicator stick for each specimen to prevent cross contamination. To ensure consistency in results, recentrifuged specimens prior to testing if

  • they contain fibrin, red blood cells, or other particulate matter.
    NOTE: If fibrin, red blood cells, or other particulate matter are observed, mix by low speed vortex or by inverting 10 times prior to centrifugation. Prepare frozen specimens as follows:

  • Frozen specimens must be completely thawed before mixing.

  • Mix thawed specimens thoroughly by low speed vortex or by inverting 10 times.

  • Visually inspect the specimens. If layering or stratification is observed, mix until specimens are visibly homogeneous.

  • If specimens are not mixed thoroughly, inconsistent results may be obtained.

  • Recentrifuged specimens. Centrifugation of Specimens

  • Transfer specimens to a centrifuge tube and centrifuge.

  • Transfer clarified specimen to a sample cup or secondary tube for testing. For centrifuged specimens with a lipid layer, transfer only the clarified specimen and not the lipemic material.

Specimen Storage

Abbott Architect SARS-CoV-2 lgG User Guide - Specimen
Storage

If testing will be delayed more than 7 days, it is recommended to store frozen. It is the responsibility of the individual laboratory to determine specific specimen stability criteria for their laboratory per their laboratory workflow. For additional information on sample handling and processing, refer to CLSI GP44-A4.13 The storage information provided here is based on data maintained by the manufacturer. Frozen specimens subjected to up to 2 freeze/thaw cycles have been evaluated.

Specimen Shipping
Package and label specimens in compliance with applicable state, federal, and international regulations covering the transport of clinical specimens and infectious substances. Do not exceed the storage limitations listed above.

PROCEDURE

Materials Provided
06R86 SARS-CoV-2 IgG Reagent Kit

Materials Required but not Provided

  • SARS-CoV-2 IgG assay file found on www.corelaboratory.abbott
  • 06R8601 SARS-CoV-2 IgG Calibrator Kit
  • 06R8610 SARS-CoV-2 IgG Control Kit or other control material containing IgG antibodies to SARS-CoV-2
  • ARCHITECT Pre-Trigger Solution
  • ARCHITECT Trigger Solution
  • ARCHITECT Wash Buffer
  • ARCHITECT Septum
    For information on materials required for operation of the instrument, refer to the ARCHITECT System Operations Manual, Section 1. For information on materials required for maintenance procedures, refer to the ARCHITECT System Operations Manual, Section 9.

Assay Procedure
For a detailed description of how to run an assay, refer to the ARCHITECT System Operations Manual, Section 5.

  • If using primary or aliquot tubes, refer to the ARCHITECT System
    Operations Manual, Section 5 to ensure sufficient specimen is present.

  • Minimum sample cup volume is calculated by the system and printed on the Order List report. To minimize the effects of evaporation, verify adequate sample cup volume is present prior to running the test.

  • When loading the reagents on the reagent carousel or reagent carrier, the conjugate must be placed in the inner ring (yellow), the microparticles must be placed in the middle ring (pink), and the assay diluent must be placed in the outer ring (green).

  • Before loading the reagent kit on the system for the first time, the microparticle bottle requires mixing to resuspend microparticles that may have settled during shipment. After the first time the microparticles have been loaded, no further mixing is required. ­
    — Invert the microparticle bottle 30 times. ­
    — Visually inspect the bottle to ensure microparticles are resuspended. If microparticles are still adhered to the bottle, continue to invert the bottle until the microparticles have been completely resuspended.
    —­ If the microparticles do not resuspend, DO NOT USE. Contact your local Abbott representative. ­
    — Once the microparticles have been resuspended, place a septum on the bottle. For instructions about placing septum’s on bottles, refer to the Reagent Handling section of this package insert.

  • Maximum number of replicates sampled from the same sample cup: 10 ­
    — Priority:

    • Sample volume for first test: 75 µL

    • Sample volume for each additional test from same sample cup: 25 µL
      —­ Routine:

    • Sample volume for first test: 150 µL

    • Sample volume for each additional test from same sample cup: 25 µL

  • Refer to the SARS-CoV-2 IgG calibrator package insert
    REF 06R8601 and/or SARS-CoV-2 IgG control package insert
    REF 06R8610 for preparation and usage.

  • For general operating procedures, refer to the ARCHITECT System Operations Manual, Section 5.

  • For optimal performance, it is important to perform routine maintenance as described in the ARCHITECT System Operations Manual, Section 9. Perform maintenance more frequently when required by laboratory procedures.

  • In order to prevent potential interactions, perform maintenance prior to and following the batching of SARS-CoV-2 IgG samples.
    —­ ARCHITECT i2000SR instrument: Perform Daily Maintenance (Operations Manual, 6041 Daily Maintenance Procedure)
    — ­ ARCHITECT i1000SR instrument (upon availability of assay file): Perform Weekly Maintenance Procedure (Operations Manual, 6445 Pipettor/WZ Probe Cleaning Maintenance Procedure)

Sample Dilution Procedures
Dilution of samples for the SARS-CoV-2 IgG assay has not been verified.

Calibration
For instructions on performing a calibration, refer to the ARCHITECT System Operations Manual, Section 6. Each assay control must be tested to evaluate the assay calibration. Once a calibration is accepted and stored, it may be used for 7 days. During this time, all subsequent samples may be tested without further calibration unless:

  • A reagent kit with a new lot number is used.
  • Daily quality control results are outside of quality control limits used to monitor and control system performance.

This assay may require recalibration after maintenance to critical parts or subsystems or after service procedures have been performed.

Quality Control Procedures
The recommended control requirement for the SARS-CoV-2 IgG assay is that a single sample of each control level be tested once every 24 hours each day of use. Additional controls may be tested in accordance with local, state, and/or federal regulations or accreditation requirements and your laboratory’s quality control policy. To establish statistically-based control limits, each laboratory should establish its own concentration target and ranges for new control lots at each clinically relevant control level. This can be accomplished by assaying a minimum of 20 replicates over several (3-5) days and using the reported results to establish the expected average (target) and variability about this average (range) for the laboratory. Sources of variation that should be included in this study in order to be representative of future system performance include:

  • Multiple stored calibrations
  • Multiple reagent lots
  • Multiple calibrator lots
  • Multiple processing modules (if applicable)
  • Data points collected at different times of the day

Refer to published guidelines for information or general control recommendation, for example Clinical and Laboratory Standards Institute (CLSI) Guideline C24, 4th ed., or other published guidelines, for general quality control recommendations.14

  • If more frequent control monitoring is required, follow the established quality control procedures for your laboratory.
  • If quality control results do not meet the acceptance criteria defined by your laboratory, sample results may be suspect. Follow the established quality control procedures for your laboratory. Recalibration may be necessary. For troubleshooting information, refer to the ARCHITECT System Operations Manual, Section 10.
  • Review quality control results and acceptance criteria following a change of reagent or calibrator lot.

Controls should be used according to the guidelines and recommendations of the control manufacturer. Concentration ranges provided in the control package insert should be used only for guidance. For any control material in use, the laboratory should ensure that the matrix of the control material is suitable for use in the assay per the assay package insert.

Quality Control Guidance
Refer to “Basic QC Practices” by James O Westgard, Ph.D. for guidance on laboratory quality control practices.15
Verification of Assay Claims
For protocols to verify package insert claims, refer to the ARCHITECT System Operations Manual, Appendix B. For protocols to verify package insert claims, follow CLIA recommendations or internal laboratory procedures.

RESULTS

Calculation
The ARCHITECT i System calculates the calibrator mean chemiluminescent signal from 3 calibrator replicates and stores the result. Results are reported by dividing the sample result by the stored calibrator result. The default result unit for the SARS-CoV-2 IgG assay is Index (S/C).

Interpretation of Results
The cutoff is 1.4 Index (S/C). As with all analyte determinations, the result should be used in conjunction with information available from clinical evaluation and other diagnostic procedures.

Index (S/C)
< 1.4
≥ 1.4

Interpretation
Negative
Positive

Flags
Some results may contain information in the Flags field. For a description of the flags that may appear in this field, refer to the ARCHITECT System Operations Manual, Section 5.

LIMITATIONS OF THE PROCEDURE

  • For use under an Emergency Use Authorization only.
  • This assay is for in vitro diagnostic use under FDA Emergency Use Authorization only.
  • This test is for clinical laboratory use only. It is not for home use.
  • Results should be used in conjunction with other data; e.g., symptoms, results of other tests, and clinical impressions.
  • Immunocompromised patients who have COVID-19 may have a delayed antibody response and produce levels of antibody which may not be detected as positive by the assay.
  • Negative results do not rule out SARS-CoV-2 infection, particularly in those who have been in contact with the virus. Testing with a molecular diagnostic should be considered to evaluate for active infection in these individuals.
  • Results from antibody testing should not be used as the sole basis to diagnose or exclude SARS-CoV-2 infection or to inform infection status.
  • Pedigreed specimens with direct evidence of antibodies to nonSARS-CoV-2 coronavirus (common cold) strains such as HKU1, NL63, OC43, or 229E have not been evaluated with this assay.
  • Not to be used to screen units of blood for SARS-CoV-2 infection.
  • Potentially interfering disease states and other cross reactants have been evaluated and are represented in the SPECIFIC PERFORMANCE CHARACTERISTICS section of this package insert.
  • Specimens from patients who have received preparations of mouse monoclonal antibodies for diagnosis or therapy may contain human anti-mouse antibodies (HAMA). Such specimens may show either falsely elevated or depressed values when tested with assay kits such as SARS-CoV-2 IgG that employ mouse monoclonal antibodies.16, 17
  • Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or to animal serum products can be prone to this interference, and anomalous values may be observed.18
  • Rheumatoid factor (RF) in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays.18

CONDITIONS OF AUTHORIZATIONS FOR THE LABORATORIES

The SARS-CoV-2 IgG Letter of Authorization, along with the authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for Patients, and authorized labeling are available on the FDA website: https://www.fda.gov /medical-devices/emergency-situations-medical-devices/emergency-use- authorizations#covid19ivd or at https://www.corelaboratory.abbott/us/en/offerings/segments/infectious-disease /sars-cov-2.
Authorized laboratories using the SARS-CoV-2 IgG (“your product” in the conditions below), must adhere to the Conditions of Authorization indicated in the Letter of Authorization as listed below:

  • A. Authorized laboratories* using your product will include with result reports of your product, all authorized Fact Sheets. Under exigent circumstances, other appropriate methods for disseminating these Fact Sheets may be used, which may include mass media.

  • B. Authorized laboratories using your product will use your product as outlined in the Instructions for Use. Deviations from the authorized procedures, including the authorized instruments, authorized clinical specimen types, authorized control materials, authorized other ancillary reagents and authorized materials required to use your product are not permitted.

  • C. Authorized laboratories that receive your product will notify the relevant public health authorities of their intent to run your product prior to initiating testing.

  • D. Authorized laboratories using your product will have a process in place for reporting test results to healthcare providers and relevant public health authorities, as appropriate.

  • E. Authorized laboratories will collect information on the performance of your product and report to DMD/OHT7-OIR/OPEQ/CDRH (via email: CDRH-EUA-Reporting@fda.hhs.gov) and Abbott Laboratories at https://www.corelaboratory.abbott/us/en/offerings/segments/infectious-disease/sars-cov-2 any suspected occurrence of false reactive or false non-reactive results and significant deviations from the established performance characteristics of your product of which they become aware.

  • F. All laboratory personnel using your product must be appropriately trained in automated immunoassay techniques and use appropriate laboratory and personal protective equipment when handling this kit, and use your product in accordance with the authorized labeling. All laboratory personnel using the assay must also be trained in and be familiar with the interpretation of results of the product.

  • G. Abbott Laboratories, authorized distributors, and authorized laboratories using your product will ensure that any records associated with this EUA are maintained until otherwise notified by FDA. Such records will be made available to FDA for inspection upon request.

  • The letter of authorization refers to, “Laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, to perform moderate and high complexity tests” as “authorized laboratories.”

SPECIFIC PERFORMANCE CHARACTERISTICS

Representative performance data are provided in this section. Results obtained in individual laboratories may vary.

Precision

Within-Laboratory Precision
A study was performed based on guidance from CLSI EP05-A3.19 Testing was conducted using 1 lot of the SARS-CoV-2 IgG Reagent Kit, 1 lot of the SARS- CoV-2 IgG Calibrator Kit, and 1 lot of the SARS-CoV-2 IgG Control Kit and 1 instrument. Two controls and 1 human serum panel were assayed in replicates of 3 at 2 separate times per day on 5 different days.

Abbott Architect SARS-CoV-2 lgG User Guide - Precision

a Includes repeatability (within-run), between-run, and between-day variability.
b Not applicable

Analytical Specificity

Potentially Cross-Reacting Antibodies

The SARS-CoV-2 IgG assay was evaluated for potentially cross-reaction antibodies. A total of 112 specimens from 23 different categories were tested. One hundred eleven (111) specimens were negative and 1 specimen was positive by the SARS-CoV-2 IgG assay. The data are summarized in the following table.

Abbott Architect SARS-CoV-2 lgG User Guide - Analytical
Specificity

  • Testing was performed on an additional 199 specimens collected prior to September 2019 from subjects who were reactive for total (IgG/IgM) antibodies to CMV. Of those 199 specimens, 1 was positive on the SARS-CoV-2 IgG assay.

Potentially Interfering Medical Conditions and Respiratory Illnesses

The SARS-CoV-2 IgG assay was evaluated for potential cross reactivity from individuals with potentially interfering medical conditions and respiratory illnesses. A total of 65 specimens from 12 different categories were tested. Sixty-five (65) specimens were negative by the SARS-CoV-2 IgG assay. The data are summarized in the following table.

Abbott Architect SARS-CoV-2 lgG User Guide - Potentially Interfering Medical
Conditions and Respiratory Illnesses

Clinical Performance

A study was performed to determine the clinical performance of the SARS-CoV-2 IgG assay.

To estimate the positive percent agreement (PPA), between the SARS-CoV-2 IgG assay and the PCR comparator, 122 serum and plasma specimens were collected at different times from 31 subjects who tested positive for SARS-CoV-2 by a polymerase chain reaction (PCR) method and who also presented with COVID-19 symptoms. Each specimen was tested using the SARS-CoV-2 IgG assay. The PPA and the 95% confidence interval (CI) were calculated.

To estimate the negative percent agreement (NPA), 1070 serum and plasma specimens from subjects assumed to be negative for SARS-CoV-2 were tested. Of the 1070 specimens, 997 specimens were collected prior to September 2019 (pre- COVID-19 outbreak). An additional 73 specimens were collected in 2020 from subjects who were exhibiting signs of respiratory illness but tested negative for SARS-CoV-2 by a PCR method. All 1070 specimens were tested using the SARS- CoV-2 IgG assay. The NPA and the 95% CI were calculated.

The results of both groups are presented in the following 2 tables.
Positive Agreement by Days Post-Symptom Onset

Abbott Architect SARS-CoV-2 lgG User Guide - Positive Agreement by Days
Post-Symptom Onset

a Five specimens from 1 immunocompromised patient were excluded from the study. Refer to the LIMITATIONS OF THE PROCEDURE section of this package insert for further information. When the results from these specimens were included, the PPA at 14 days post-symptom onset was 96.77% (95% CI: 90.86, 99.33).

Negative Agreement by Category

Abbott Architect SARS-CoV-2 lgG User Guide - Negative Agreement by
Category

Class Specificity

The anti-human IgG antibody used in the SARS-CoV-2 IgG assay demonstrates class-specific reactivity only to human IgG isotypes. No binding interactions were observed to human IgM, human IgA, or sheep (ovine) IgG.

Longitudinal Study
From the positive agreement study above, a subset of 13 subjects with 2 or more blood draws post-symptom onset were assessed longitudinally. Out of the 13 subjects, 7 presented positive results and 4 presented negative results in all the bleeds while 2 subjects showed SARS-CoV-2 IgG seroconversion as shown in the SARSCoV-2 IgG results provided below.

Abbott Architect SARS-CoV-2 lgG User Guide - CoV-2 IgG
results Abbott Architect SARS-
CoV-2 lgG User Guide - CoV-2 IgG results

BIBLIOGRAPHY

  1. Zhu N, Zhang D, Wang W, et al. A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med 2020;382:727-733. doi:10.1056/NEJMoa2001017
  2. World Health Organization. WHO Director-General’s opening remarks at the media briefing on COVID-19 – 11 March 2020. Accessed April 14, 2020. https://www.who.int/dg/speeches/detail/who-directorgeneral-s-opening-remarks-at-the-media-briefing-on-covid-19—11march-2020
  3. World Health Organization. Report of the WHO-China Joint Mission on Coronavirus Disease 2019 (COVID-19) 16-24 February 2020. Accessed April 14, 2020. https://www.who.int/docs/default-source/coronaviruses/who-china-joint-mission-on-covid-19-final-report
  4. World Health Organization. Coronavirus disease 2019 (COVID-19) situation report-85, Data as received by WHO from national authorities by 10:00 CET, 14 April 2020. Accessed April 14, 2020. 20200414-sitrep-85-covid-19 [PDF]
  5. Zhao J, Yuan Q, Wang H, et al. Antibody responses to SARS-CoV-2 in patients of novel coronavirus disease 2019. Clin Inf Dis Accepted manuscript. Published online 28 March 2020. doi:10.1093/cid/ ciaa344
  6. Xiao AT, Gao C, Zhang S. Profile of specific antibodies to SARSCoV-2: the first report. J Infect 2020. doi:10.1016/j.jinf.2020.03.012
  7. Shen C, Wang Z, Zhao F, et al. Treatment of 5 critically ill patients with COVID-19 with convalescent plasma. JAMA Published online March 27, 2020. doi:10.1001/jama.2020.4783
  8. World Health Organization. Laboratory testing strategy recommendations for COVID-19: interim guidance, 21 March 2020. World Health Organization; 2020.
  9. US Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910.1030, Bloodborne pathogens.
  10. US Department of Health and Human Services. Biosafety in Microbiological and Biomedical Laboratories. 5th ed. Washington, DC: US Government Printing Office; December 2009.
  11. World Health Organization. Laboratory Biosafety Manual. 3rd ed. Geneva: World Health Organization; 2004.
  12. Clinical and Laboratory Standards Institute (CLSI). Protection of Laboratory Workers From Occupationally Acquired Infections; Approved Guideline–Fourth Edition. CLSI Document M29-A4. Wayne, PA: CLSI; 2014.
  13. Clinical and Laboratory Standards Institute (CLSI). Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline–Fourth Edition. CLSI Document GP44-A4. Wayne, PA: CLSI; 2010.
  14. Clinical and Laboratory Standards Institute (CLSI). Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI Guideline C24. Wayne, PA: CLSI; 2016.
  15. Westgard JO. Basic QC Practices. 3rd ed. Madison, WI: Westgard Quality Corporation; 2010.
  16. Primus FJ, Kelley EA, Hansen HJ, et al. “Sandwich”-type immunoassay of carcinoembryonic antigen in patients receiving murine monoclonal antibodies for diagnosis and therapy. Clin Chem 1988;34(2):261-264.
  17. Schroff RW, Foon KA, Beatty SM, et al. Human anti-murine immunoglobulin responses in patients receiving monoclonal antibody therapy. Cancer Res 1985;45(2):879-885.
  18. Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all immunoassays. Clin Chem 1988;34(1):27-33.
  19. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline–Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.

Key to Symbols

Abbott Architect SARS-CoV-2 lgG User Guide - Key to
Symbols
Note for number formatting:

  • A space is used as thousands separator (example: 10 000 specimens).
  • A period is used to separate the integer part from the fractional part of a number written in decimal form (example: 3.12%).

ARCHITECT and related brand marks are trademarks of Abbott. Other trademarks are the property of their respective owners.

Abbott Laboratories
Diagnostics Division
Abbott Park, IL 60064 USA

PRODUCED FOR ABBOTT BY

Fujirebio Diagnostics, Inc.
201 Great Valley Parkway
Malvern, Pennsylvania 19355, USA

Customer Service: Contact your local representative or find country- specific contact information onwww.corelaboratory.abbott

Created April 2020. ©2020 Abbott Laboratories