  • Product Name : ID NOWTM COVID-19 2.0
  • Features : Flexible Testing, Flexible Workflow
  • Additional Test : Influenza A & B can be added with the same patient sample
  • Software Update : ID NOWTM software update to version 7.1 is required for sequential workflow capability

Product Usage Instructions

Test for COVID-19

  1. Obtain a single patient swab sample (nasal or nasopharyngeal).
  2. Follow the step-by-step instructions on the ID NOW screen to perform the COVID-19 test.

Add On Influenza A & B

  1. After the COVID-19 test, choose to add on an Influenza A & B test or conclude testing.
  2. Select the desired option on the ID NOW screen: Add on Influenza A & B or No Influenza A & B test.
  3. If adding on Influenza A & B:
    • Select Actions.
    • Select COMBO: COVID-19 + Flu A & B.
    • Conclude the COVID-19 test by removing and discarding used components.
    • Attach the USED COVID-19 test base/transfer cartridge to the UNUSED Influenza A & B sample receiver for disposal.
    • Use a new Influenza A & B test base and transfer cartridge with the SAME COVID-19 2.0 sample receiver to complete the add-on test.
  4. Read the final result and follow the Product Insert instructions for safe handling and disposal of components.

FAQ

  • Q: Can I use the same sample receiver for both COVID-19 and Influenza A & B tests?
    • A : Yes, you can use the same sample receiver for both tests by following the specified workflow and using new test components as instructed.
  • Q: Is a software update required to run the Influenza A & B test in sequential workflow?
    • A : Yes, an ID NOWTM software update to version 7.1 is required for sequential workflow capability when adding on Influenza A & B after the COVID-19 test.

Test for COVID-19, and easily add on Influenza A & B with the same patient sample.

Materials required for sequential workflow

Abbott-COVID-19 2.0-Flexible-Testing -FIG1

  • ID NOW™ Influenza A & B 2 test time to result when run in sequential workflow after the ID NOW™ COVID-19 2.0 test. ID NOW™ software update to version 7.1 required for sequential workflow capability.

TEST FOR COVID-19, AND ADD ON INFLUENZA A & B USING A SINGLE PATIENT SWAB

Before performing these tests, refer to the ID NOW™ COVID-19 2.0 Product Insert for the complete test procedure and additional information.

Follow the step-by-step instructions on the ID NOW screen to perform a COVID-19 test. After the COVID-19 test, choose to add on an Influenza A & B test or conclude testing.

  • Option 1: Add on Influenza A & B.
  • Option 2: No Influenza A & B test.

Abbott-COVID-19 2.0-Flexible-Testing -FIG2

Abbott-COVID-19 2.0-Flexible-Testing -FIG3

  • Do not remove the used sample receiver.
  • Take an UNUSED sample receiver from the Influenza A & B test kit, and carefully remove the seal.
  • Lift the USED COVID-19 test base/transfer cartridge from the instrument, and attach it to the UNUSED Influenza A & B sample receiver for disposal.
  • Abbott-COVID-19 2.0-Flexible-Testing -FIG3
  • Use a new Influenza A & B test base and transfer cartridge  with the SAME COVID-19 2.0 sample receiver to complete an add-on test
  • © 2023 Abbott. All rights reserved. All trademarks referenced are trademarks of either the Abbott Group of companies or their respective owners. Any photos displayed are for illustrative purposes only. Any person depicted in such photos is a model. COL-21623-01 09/23
    • Read result.