INVACARE Get-U-Up Lift Owner’s Manual
- June 7, 2024
- INVACARE
Table of Contents
- Owner’s Operator and Maintenance Manual
- Get-U-Up™ Lift
- WARNING
- ACCESSORIES WARNING
- SYMBOL LEGEND
- SPECIAL NOTES
- LABEL LOCATIONS
- TYPICAL PRODUCT PARAMETERS
- SECTION 1–GENERAL GUIDELINES
- SECTION 2–ASSEMBLY
- SECTION 3–OPERATION
- Operating the Lift
- Raising/Lowering the Hydraulic Lift
- SECTION 4–LIFTING
- SECTION 5–TRANSFERRING
- SECTION 6–USING THE PATIENT LIFT AS A STANDING AID
- SECTION 7–TROUBLESHOOTING
- SECTION 8–MAINTENANCE
- LIMITED WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Owner’s Operator and Maintenance Manual
Get-U-Up™ Lift
DEALER: This manual MUST be given to the user of the patient lift.
USER: BEFORE using this patient lift, read this manual and save for future reference.
For more information regarding Invacare products, parts, and services, please visit www.invacare.com
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS OR INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT – OTHERWISE, INJURY OR DAMAGE MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
SYMBOL LEGEND
“ATTENTION, see instructions for use”.
CAUTION – Pinch Points, fingers could be pinched.
WARNING – When positioning Lift, be aware of the position of the Swivel Bar and the patient. Injury could occur.
“Date of Manufacture”
Device contains Lead Acid batteries. DO NOT dispose of batteries in normal household waste. They MUST be taken to a proper disposal site. Contact your local waste management company for information.
EMERGENCY Mechanical Lowering – Pull UP on EMERGENCY Button. Push DOWN on Boom.
NOTE: Updated versions of this manual are available onwww.invacare.com.
2
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the table below for definitions of the signal words.
SIGNAL WORD|
MEANING
---|---
DANGER| Danger indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING| Warning indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION| Caution indicates a potentially hazardous situation which, if not
avoided, may result in property damage or minor injury or both.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT
NOTICE.
Check ALL parts for shipping damage. If shipping damage is noted, DO NOT
use. Contact dealer/carrier for further instruction.
MAINTENANCE
Maintenance MUST be performed ONLY by qualified personnel.
5
LABEL LOCATIONS
LABEL LOCATIONS
OPERATION OF THE PATIENT LIFT
The mast (C) includes the steering handles used to move and guide the lift. The mast is supported in a socket in the rear center of the base with casters (A). For transporting the lift in a vehicle, the mast, boom and pump assembly may be lifted from the base socket for convenient storage. Each time the mast is returned to the socket of the base, care should be used to push the mast as far as possible down into the socket. Rotate the mast slightly until it locks in the center position. The bottom of the mast tube is grooved on either side, and those grooves seat on tabs in the base socket to prevent any rotation of the mast during the lift procedure. Secure the mast to the base by securely tightening the screw located at the rear of the base.
Refer to the Patient Sling owner’s manual for complete lifting preparation information.
The horn (D) is made with knobs on both sides and in the center to accept the sling that supports the patient during lift.
There are two controls on the pump assembly (F), which are the pump handle (G) and the control valve (H). With the control valve closed, the pump handle is moved up and down with slow strokes of even pressure to raise the horn. A patient in a sling attached to the horn would be elevated by this operation. The horn is lowered by opening the control valve and moving the control valve away from the pump handle. A safety gate is part of the hydraulic system that prevents sudden lowering of the horn regardless of how far the control valve is opened.
The knee pad (E) can be adjusted for patient support and comfort. Pull both adjustment pins outward at the same time to move the knee pad to the desired position. Make sure adjustment pins are engaged.
To raise the footplate (I), position the knee pad (E) in the highest setting. When the footplate is up, the knee pad can be adjusted to the desired position.
WARNING: For maximum stability and patient safety, operate this lift with the base legs in maximum open position and locked when lifting a patient.
The shifter handle (B) is used to open the legs of the base for stability before lifting a patient. The base legs should be locked in the maximum open position at all times when a patient is on the lift.
To open the base legs, grasp the shifter handle in one hand, and place the opposite hand on the steering handle of the mast for balance. Move the handle to your right to release the lock pin then bring the handle towards you in a complete half-circle. The lock pin should insert into a hole in the plate to lock the legs in full open position and the shifter handle should once again be vertical. To close the legs to a narrow position for storage, reverse the above procedure. If the shifter handle is not vertical (straight up and down), the lock pin is not seated in place.
Rev. A 1148135
WARNING
BEFORE using the Patient Lift, READ and UNDERSTAND the Owner’s Manual for proper operation and safety procedures.
WARNING
WEIGHT LIMITATION 350 lbs. DO NOT roll casterbase over deep carpet, raised carpet bindings, door frames or any uneven surfaces that may cause the Patient Lift to tip over.
DO NOT lock the casters of the Patient Lift when lifting an individual. Casters MUST be left unlocked to allow Patient Lift to stabilize during lifting procedures.
USE ONLY INVACARE SLINGS and LIFT ACCESSORIES.
USE ONLY Get-U-Up™ Model Components for maintenance and replacement.
Refer to Owner’s Manual for periodic maintenance procedures.
Customer Service:
1-800-333-6900
1148134 Rev. A
6
TYPICAL PRODUCT PARAMETERS
TYPICAL PRODUCT PARAMETERS
HEIGHT AT SLING
HOOK-UP – MAX.|
65.7 inches
---|---
HEIGHT AT SLING
HOOK-UP – MIN.|
36 inches
BASE WIDTH OPEN| 37 inches
BASE WIDTH CLOSED| 23 inches
BASE HEIGHT (CLEARANCE)| 4.5 inches
BASE LENGTH| 40.5 inches
OVERALL HEIGHT| 45 inches
OVERALL LENGTH| 44 inches
OVERALL WIDTH| 23 inches
CASTER SIZE (REAR/FRONT)| 5 inches / 3 inches
SLING OPTIONS| Standing or Transport
SLING MATERIAL| Polyester
STANDING SLING
WIDTH
LENGTH|
36 inches
13 inches
TRANSPORT SLING
WIDTH
LENGTH|
38.5 inches
36 inches
WEIGHT CAPACITY| 350 lbs
WEIGHT IN CARTON| 100 lbs
WEIGHT OUT OF CARTON| 88 lbs
7
SECTION 1–GENERAL GUIDELINES
SECTION 1–GENERAL GUIDELINES
WARNING
SECTION 1 – GENERAL GUIDELINES contains important information for the safe operation and use of this product.
Check all parts for shipping damage before using. In case of damage, DO NOT
use the equipment. Contact the dealer for further instructions.
The Invacare patient lift is NOT a transport device. It is intended to
transfer an individual from one resting surface to another (such as a bed to a
wheelchair). Moving a person suspended in a sling over ANY distance is NOT
recommended.
DO NOT attempt to transfer without the approval of the patient’s physician,
nurse or medical assistant. Thoroughly read the instructions in this Owner’s
Manual, observe a trained team of experts performing the lifting procedures
and then perform the entire lift procedure several times with proper
supervision and a capable individual acting as a patient.
Invacare Stand Assist and Transfer slings are specifically designed to be used
in conjunction with Invacare patient lifts. Slings and accessories designed by
other manufacturers are not to be utilized as a component of Invacare’s
patient lift system. Use the sling that is recommended by the individual’s
doctor, nurse or medical assistant for the comfort and safety of the
individual that is being lifted.
Weight Limitation
DO NOT exceed the maximum weight limitation of the patient lift. The Get-U-Up lift has a weight limitation of 350 lbs.
Assembling the Lift
DO NOT overtighten the mounting hardware. This will damage the mounting brackets.
Using the Sling
Individuals that use the Stand Assist Sling MUST be able to support the
majority of their own weight, otherwise injury may occur.
Stand Assist Slings: DO NOT use the stand assist sling in combination with the
patient lift as a transport device. It is intended to transfer an individual
from one resting surface to another (such as a bed to a wheelchair). Moving a
person using the stand assist sling in combination with the patient lift over
ANY distance is NOT recommended.
Stand Assist Slings: Individuals that use the stand assist sling MUST be able
to support the majority of their own weight, otherwise injury can occur.
Before lifting the patient, make sure the bottom edge of the stand support
sling is positioned on the lower back of the patient and the patients arms are
outside the stand assist sling. The belt MUST be snug, but comfortable on the
patient, otherwise the patient can slide out of the sling during transfer,
possibly causing injury.
Transfer Slings: Before lifting the patient, make sure the bottom edge of the
transfer sling is at the base of the spine and the patients arms are outside
the transport sling.
Transfer Slings: DO NOT raise the patient to a full standing position while
using the transfer sling, otherwise injury may occur.
DO NOT use any kind of material (such as a plastic back incontinence pad or
seating cushion) between the patient and sling material that may cause the
patient to slide out of the sling during transferring.
After each laundering (in accordance with instructions on the sling), inspect
sling(s) for wear, tears, and loose stitching.
Bleached, torn, cut, frayed, or broken slings are unsafe and could result in
injury. Discard IMMEDIATELY.
8
DO NOT alter slings.
Be sure to check the sling attachments each time the sling is removed and
replaced, to ensure that it is properly attached before the patient is removed
from a stationary object (bed, chair or commode).
If the patient is in a wheelchair, secure the wheel locks in place to prevent
the chair from moving forwards or backwards.
When connecting slings equipped with color coded straps to the patient lift,
the shortest of the straps MUST be at the back of the patient for support.
Using long section will leave little or no support for the patient’s back. The
loops of the sling are color coded and can be used to place the patient in
various positions. The colors make it easy to connect both sides of the sling
equally. Make sure that there is sufficient head support when lifting a
patient.
Use an Invacare sling that is recommended by the individual’s doctor, nurse or
medical assistant for the comfort and safety of the individual being lifted.
Operating the Lift
Make sure there is an audible click when mounting battery on the battery
charger to confirm proper mounting. Otherwise, injury or damage may occur.
Use the handles to push or pull the patient.
Lifting the Patient
Before positioning the legs of the stand up lift around the patient, make sure
that the patient’s feet are out of the way of the foot plate, otherwise injury
may occur.
Adjustments for safety and comfort should be made before moving the patient.
Patient’s arms should be outside of the sling straps.
Before lifting a patient from a stationary object (wheelchair, commode or
bed), slightly raise the patient off of the stationary object and check that
all sling attachments are secure. If any attachment is not correct, lower the
patient and correct the problem, then raise the patient and check again.
During transfer, with the patient suspended in a sling attached to the lift,
DO NOT roll caster base over objects such as carpet, raised carpet bindings,
door frames, or any uneven surfaces or obstacles that would create an
imbalance of the patient lift to tip over. Use the steering handle on the mast
at ALL times to push or pull the patient lift.
Invacare recommends locking the rear swivel casters ONLY when positioning or
removing the sling (stand assist or transfer) from around the patient.
Transferring the Patient
Before transferring, check that the product’s weight capacity can withstand
the patient’s weight.
Wheelchair wheel locks MUST be in a locked position before lowering the
patient into the wheelchair for transport.
Performing Maintenance
Refer to Maintenance on page 25 for a maintenance schedule and procedures.
Regular maintenance of lifts and accessories is necessary to ensure proper
operation.
After the first 12 months of operation, inspect all pivot points and fasteners
for wear. If the metal is worn, the parts MUST be replaced. Make this
inspection every six months thereafter.
DO NOT overtighten mounting hardware. This will damage mounting brackets.
Casters and axle bolts require inspections every six months to check for
tightness and wear.
If using the lift in the area of a shower or bath, ensure the lift is wiped
clean of any moisture after use. Do not store the lift in a damp area or in a
damp condition. Periodically inspect all components for signs of corrosion and
replace those that are corroded or damaged.
9
SECTION 2–ASSEMBLY
SECTION 2–ASSEMBLY
WARNING
Each time the mast is removed and assembled to the base assembly, the mast MUST be locked into the socket of the base assembly.
Attach the Mast to the Base Assembly
NOTE: For this procedure, refer to FIGURE 2.1.
-
Remove the locking screw from the base assembly.
-
Match the notch at the bottom of the mast with the tabs inside the socket of the base assembly.
-
Insert the mast into the socket and onto the tabs.
-
Try to turn the mast.
NOTE: If the mast turns in the socket, repeat STEPS 2 and 3 until proper alignment is obtained. -
If the mast does not turn in the socket, insert the locking screw into the base and tighten securely.
Side view of the Base Assembly and Mast
FIGURE 2.1 Attach the Mast to the Base Assembly
10
Attach the Horn to the Mast and Actuator
NOTE: For this procedure, refer to FIGURE 2.2.
-
Align the horn bushing with the holes in the horn mounting lugs on the mast.
-
Attach the horn to the mast with a shoulder bolt, two steel washers, two nylon washers and a locknut.
NOTE: The nylon washers MUST be installed between the horn bushing and horn mounting lugs. -
Align the holes in the actuator cylinder rod with the holes in the cylinder mounting lugs on the horn.
-
Attach the actuator cylinder to the horn with a shoulder bolt and locknut.
Check for proper hardware alignment
CAUTION
The horn MUST pivot easily.
- At the attachment point of the horn to the mast assembly, use an Allen wrench to hold the locknut and turn the shoulder bolt by hand (it should rotate without a lot of force being applied).
- Raise the horn.
- Push the control handle away from the pump assembly.
- If the horn pivots easily, tighten the shoulder bolt and locknut.
NOTE: If the horn does not pivot easily or not at all, remove the locknut and shoulder bolt. Check all hardware for proper alignment. Reinstall hardware and repeat STEPS 1-4. If alignment problems continue, contact the Technical Support Department at Invacare.
FIGURE 2.2 Attach the Horn to the Mast and Actuator
11
Attaching/Adjusting the Knee Pad
NOTE: For this procedure, refer to FIGURE 2.3.
WARNING
Never adjust knee pad while patient is in the standing position or while
the lift is moving.
Always make sure that the adjustment pins are fully engaged in the height
adjustment holes before use.
-
Pick a height setting that will be comfortable to the patient and provide the necessary support.
NOTE: The knee pad should be positioned so that the knee portion of the leg contacts the pad.
NOTE: The upper adjustment pins are typically used for shorter individuals. The lower adjustment pins are typically used for taller individuals. -
Using both hands, pull both adjustment pins outward at the same time and hold.
-
Position the knee pad to the desired height and release adjustment pins into the corresponding alignment holes.
-
Check to make sure that both pins are engaged.
FIGURE 2.3 Attaching/Adjusting the Knee Pad
Attaching the Base Shifter Handle
NOTE: For this procedure, refer to FIGURE 2.4.
- Insert the base shifter handle into the cam lock assembly at the back of the base.
- Align the holes of the shifter handle and cam lock assembly.
- Tighten the thumbscrew to secure the shifter handle in place.
NOTE: This should prevent the base shifter handle from being removed.
FIGURE 2.4 Attaching the Base Shifter Handle
12
SECTION 3–OPERATION
SECTION 3–OPERATION
Introduction
WARNING
DO NOT attempt any transfer of a patient without approval of the patient’s physician, nurse, or medical assistant. Thoroughly read the instructions in this owner’s manual, observe a trained team of experts performing the lifting procedures and then perform the entire lift procedure several times with proper supervision and a capable individual acting as a patient.
ONLY operate the lift with the legs of the lift in the maximum open position for optimum stability and safety. If the patient is in a sling and it becomes necessary to move through a narrow passage, close the legs of the stand up lift only as long as it takes to move through the passage. When the stand up lift is through the passage, return the legs to the maximum open position.
If it is necessary to close the legs to maneuver the stand up lift under a bed, close the legs only as long as it takes to position the stand up lift over the patient and lift the patient off the surface of the bed. When the legs of the stand up lift are no longer under the bed, return the legs to the maximum open position.
Invacare slings and lift accessories are specifically designed to be used in conjunction with Invacare lifts. Slings and accessories designed by other manufacturers are not to be utilized as a component of Invacare’s lift system.
DO NOT exceed maximum weight limitation of the lift. The weight limitation for the Get-U-Up Lift is 350 lbs.
ALWAYS keep hands, fingers, feet and toes clear of moving parts to avoid injury.
NOTE: Invacare recommends that two assistants be used for all lifting preparation and transferring to/from procedures; however, the stand up lift can be operated with one assistant. The use of one assistant is based on the evaluation of the health care professional for each individual case.
13
Operating the Lift
NOTE: For this procedure, refer to FIGURE 3.1.
Opening the Legs
- Stand at the rear of the lift and grasp the shifter handle in one-hand and place the opposite hand on the steering handle of the mast for balance. Refer to Details “A” and “B”.
- Push the shifter handle to your right to release the lock pin from its mounting hole.
- Turn the handle clockwise until you are able to secure the lock pin into the opposite mounting hole to fully open the legs of the base. Refer to Detail “C”.
NOTE: The lock pin MUST insert into its mounting hole in the socket assembly to lock the legs in the full open position. The shifter handle will be in the vertical position and parallel with the mast assembly.
WARNING
If the shifter handle is NOT vertical, the lock pin is NOT seated in the socket assembly. DO NOT use the lift until the lock pin is properly seated and locked in place or injury and/or damage may occur.
Closing the Legs
- Stand at the rear of the lift and grasp the shifter handle in one-hand.
- Place the opposite hand on the steering handle of the mast for balance. Refer to Details “A” and “B”.
- Push the shifter handle to your right to release the lock pin from its mounting hole.
- Turn the handle counter-clockwise until you are able to secure the lock pin into the opposite mounting hole to close the legs of the base. Refer to Detail “C”.
DETAIL “A”
DETAIL “B” DETAIL “C”
FIGURE 3.1 Opening/Closing the Legs
14
Raising/Lowering the Hydraulic Lift
NOTE: For this procedure, refer to FIGURE 3.2.
Raising the Lift
NOTE: Ensure the control valve is in the closed position (control valve
positioned towards pump handle).
- Move the pump handle up and down to raise the lift.
Lowering the Lift
NOTE: Ensure the control valve is in the open position (control valve
positioned away from pump handle).
NOTE: The weight of the person on the lifting arm causes the lift arm to lower
when the control valve is open.
NOTE: The rate of descent is controlled by varying the amount that the control
valve is opened. A safety gate is part of the hydraulic system that maintains
a SLOW constant descent of the boom regardless of how far the control valve is
opened.
FIGURE 3.2 Raising/Lowering the Hydraulic Lift
15
SECTION 4–LIFTING
SECTION 4–LIFTING
Introduction
WARNING
DO NOT attempt any transfer without approval of the patient’s physician,
nurse or medical assistant.
Invacare patient slings are made specifically for use with Invacare patient
lifts. For the safety of the patient, DO NOT intermix patient slings and
patient lifts of different manufacturers.
NOTE: Invacare recommends that two assistants be used for all lifting
preparation and transferring to/from procedures; however, our equipment will
permit proper operation by one assistant. The use of one assistant is based on
the evaluation of the health care professional for each individual case.
NOTE: Refer to the patient sling owner’s manual, P/N 1023891, for more
information.
Positioning the Stand Up Lift
NOTE: Before positioning the legs of the patient lift under a bed, make sure that the area is clear of any obstructions.
WARNING
The legs of the stand up lift MUST be in the maximum open position for
optimum stability and safety. If it is necessary to close the legs to maneuver
the stand up lift under a bed, close the legs only as long as it takes to
position the stand up lift over the patient and lift the patient off the
surface of the bed. When the legs of the lift are no longer under the bed,
return the legs to the maximum open position.
Before positioning the legs of the stand up lift near the patient, make
sure the patient’s feet are out of the way of the footplate, otherwise injury
can occur.
- Stand at the rear of the lift and grasp the shifter handle in one-hand and place the opposite hand on the steering handle of the mast for balance.
- Push the shifter handle to your right to release the lock pin from its mounting hole.
- Turn the handle clockwise until you are able to secure the lock pin into the opposite mounting hole to fully open the legs of the base.
NOTE: The lock pin MUST insert into its mounting hole in the socket assembly to lock the legs in the full open position. The shifter handle will be in the vertical position and parallel with the mast assembly.
WARNING
If the shifter handle is NOT vertical, the lock pin is NOT seated in the socket assembly. DO NOT use the lift until the lock pin is properly seated and locked in place or injury and/or damage may occur.
4. Position the stand up lift using the mast handle.
5. Lower the lift arms for easy attachment of the patient sling.
16
Lifting the Patient
WARNING
DO NOT exceed the maximum weight limitation of 350 lbs.
DO NOT use slings and stand up lifts of different manufacturers. Invacare
slings are made specifically for use with Invacare stand up lifts. Injury or
damage may occur.
DO NOT move the patient if the sling is not properly connected to the
attachment points of the stand up lift. Check that the sling is properly
connected to the attachment points prior to lifting a patient. If any
attachments are not properly in place, correct the problem. When the sling is
elevated a few inches off the stationary surface and before moving the
patient, check again to make sure that all sling attachments are secure. If
any attachments are not properly in place, lower the patient back onto the
stationary surface and correct this problem – otherwise, injury or damage may
occur.
Adjustments for safety and comfort should be made before moving the
patient. Patient’s arms should be outside of the straps.
Individuals that use the stand assist sling MUST be able to support the
majority of their own weight, otherwise injury can occur.
Stand Assist Slings – Before lifting the patient, make sure the bottom edge
of the Stand Assist Sling is positioned on the lower back of the patient and
the patient’s arms are outside the standing sling.
Transfer Slings – Before lifting the patient, make sure the bottom edge of
the Transfer Sling is at the base of the spine and the patient’s arms are
outside the sling. DO NOT raise the patient to a full standing position.
Before lifting the patient, the belt MUST be snug but comfortable on the
patient, otherwise the patient can slide out of the sling during transfer,
possibly causing injury.
NOTE: For this procedure, refer to FIGURE 4.1 on page 18.
- Instruct patient to hold onto the handgrips on both sides of the stand up lift. Refer to Detail “A”.
- Instruct the patient to lean back into the stand assist or transport sling.
- Ensure the following:
• Patient’s knees are secure against the knee pad.
• Patients feet are positioned on the footplate.
• The bottom edge of either the standing sling is positioned on the lower back or the transport sling is at the base of the patients spine.
• The patients arms are outside of the standing or transport sling.
• The legs are in the maximum open position.
WARNING
Before transferring, check that the wheelchair weight capacity can
withstand the patient’s weight. Refer to Typical Product Parameters in the
wheelchair owner’s manual.
If transferring from a wheelchair, the wheelchair wheel locks MUST be in
the locked position before lowering the patient into the wheelchair.
Otherwise, injury may occur.
4. If transferring from a wheelchair, lock the wheel locks on the wheelchair.
Refer to Detail “B”.
5. Raise the patient above the surface (bed, wheelchair, or commode) high
enough to clear the surface. The patient’s weight will be fully supported by
the lift. Refer to Detail “C” on page 18 and Operating the Lift on page 14.
17
NOTE: The lower center of gravity provides stability making the patient feel
more secure and the lift easier to move.
NOTE: The lift arms will stay in position until lowered.
DETAIL “A” – LIFT
DETAIL “B” – WHEEL LOCK
DETAIL “C” – LIFTING THE PATIENT
FIGURE 4.1 Lifting the Patient
Moving the Patient
WARNING
DO NOT move a person suspended in a sling any distance. The Invacare lift
is NOT a transport device. It is intended to transfer a seated individual from
one resting surface to another (such as a bed to a wheelchair). Otherwise,
injury or damage may occur.
The legs of the stand up lift MUST be in the maximum open position for
optimum stability and safety. If the patient is in a sling and it becomes
necessary to move through a narrow passage, close the legs of the stand up
lift only as long as it takes to move through the passage. When the stand up
lift is through the passage, return the legs to the maximum open position.
DO NOT roll caster base over objects such as carpet, raised carpet
bindings, door frames, or any uneven surfaces or obstacles that would create
an imbalance of the lift during transfer of a patient suspended in the lift
sling. This could cause the lift to tip over. Use the mast handle at all times
to push or pull the lift.
NOTE: For this procedure, refer to Detail “C” of FIGURE 4.1.
- Ensure the legs of the stand up lift are in the maximum open position. If not, open legs to the maximum open position.
- Move the stand up lift away from the surface using the steering handle.
- Slowly move the patient to the desired surface.
18
SECTION 5–TRANSFERRING
SECTION 5–TRANSFERRING
Introduction
WARNING
DO NOT attempt any transfer of a patient without approval of the patient’s
physician, nurse, or medical assistant.
DO NOT move the patient if the sling is not properly connected to the
attachment points of the stand up lift. Check that the sling is properly
connected to the attachment points prior to lifting a patient. If any
attachments are not properly in place, correct the problem. When the sling is
elevated a few inches off the stationary surface and before moving the
patient, check again to make sure that all sling attachments are secure. If
any attachments are not properly in place, lower the patient back onto the
stationary surface and correct this problem – otherwise, injury or damage may
occur.
Adjustments for safety and comfort should be made before moving the
patient. The patient’s arms should be outside the straps.
Before positioning the legs of the stand up lift around the patient, make
sure the patient’s feet are out of the way of the footplate, otherwise injury
can occur.
DO NOT use slings and stand up lifts of different manufacturers. Invacare
slings are made specifically for use with Invacare stand up lifts. Otherwise,
injury or damage may occur.
The legs of the stand up lift MUST be in the maximum open position before
lifting the patient for optimum stability and safety. If the patient is in a
sling and it becomes necessary to move through a narrow passage, close the
legs of the stand up lift only as long as it takes to move through the
passage. When the stand up lift is through the passage, return the legs to the
maximum open position. If it is necessary to close the legs to maneuver the
stand up lift under a bed, close the legs only as long as it takes to position
the stand up lift over the patient and lift the patient off the surface of the
bed. When the legs of the stand up lift are no longer under the bed, return
the legs to the maximum open position.
DO NOT roll caster base over objects such as carpet, raised carpet
bindings, door frames, or any uneven surfaces or obstacles that would create
an imbalance of the lift during transfer of a patient suspended in the lift
sling. This could cause the lift to tip over. Use the mast handle at all times
to push or pull the lift.
Be sure to check the sling attachments each time the sling is removed and
replaced to ensure that it is properly attached before the patient is removed
from a surface.
NOTE: Invacare recommends that two assistants be used for all lifting preparation and transferring to/from procedures; however, our equipment will permit proper operation by one assistant. The use of one assistant is based on the evaluation of the health care professional for each individual case.
19
Transferring to a Commode
NOTE: For this procedure, refer to FIGURE 5.1.
-
Raise the patient high enough to clear the arms of the commode chair. Their weight will be supported by the lift. Refer to Detail “A”.
-
Guide the patient onto the commode chair.
-
Lower the patient onto the commode chair.
-
Perform one of the following (refer to Detail “B” or “C”):
• Standing Sling – unhook the standing sling from the attachment points on the lift.
• Transport Sling –
i. Unhook the transport sling from the bottom attachment points on the lift.
ii. Lift up on the patient’s legs and remove the thigh supports from underneath the patient.
iii. If desired, unhook the transport sling from the top attachment points on the lift.
NOTE: The patient can remain in the upper portion of the transport sling while using the commode. -
Instruct or assist the patient in lifting their feet off the footplate.
-
Remove the sling from around the patient.
-
Pull the lift away from the commode.
-
When complete, recheck the sling for correct attachments.
-
To lift the patient from the commode, refer to Lifting starting on page 16.
DETAIL “A” – POSITIONING PATIENT
DETAIL “B” – UNHOOKING SLING
DETAIL “C” – UNHOOKING SLING AND STRAPS
FIGURE 5.1 Transferring to a Commode
20
Transferring to a Wheelchair
NOTE: For this procedure, refer to FIGURE 5.2.
NOTE: Before transferring, check that the wheelchair weight capacity can
withstand the patients weight. Refer to Typical Product Parameters in the
wheelchair owner’s manual
- Ensure the legs of the lift with the patient in the sling are in the open position.
- Move the wheelchair into position. Refer to Detail “A”.
- Engage the rear wheel locks of the wheelchair to prevent movement of the wheelchair. Refer to Detail “B”.
WARNING
DO NOT place the patient in the wheelchair if the locks are not engaged. The wheelchair wheel locks MUST be in a locked position before lowering the patient into the wheelchair for transport. Otherwise, injury may result.
4. Position the patient over the wheelchair.
5. Lower the patient into the wheelchair. Refer to Detail “C”.
6. Unhook the sling from all attachment points on the lift. Refer to Detail
“D”.
7. Instruct patient to lift their feet off the footplate. Assist the patient
if necessary.
8. Remove the sling from around the patient.
9. Pull the lift away from the wheelchair.
DETAIL “A” – UNHOOKING SLING AND STRAPS
DETAIL “B” – LOCKING THE WHEELCHAIR
DETAIL “C” – LOWERING
DETAIL “D” – UNHOOKING SLING
FIGURE 5.2 Transferring to a Wheelchair
21
Transferring to a Bed
NOTE: For this procedure, refer to FIGURE 5.3.
NOTE: The lower center of gravity provides stability making the patient feel
more secure and the lift easier to move.
NOTE: The lift arms will stay in position until lowered.
-
Position the patient as far over the bed as possible.
NOTE: If patient is being transferred from a surface that is lower than the bed, raise the patient above the surface of the bed. The patient should be elevated just high enough to clear the bed with their weight fully supported by the lift. -
Lower the patient onto the bed.
-
Unhook the sling from all attachment points on the lift.
-
Instruct the patient to lift their feet off of the footplate. Assist the patient if necessary.
-
Remove the standing or transport sling from around the patient.
-
Pull the lift away from the bed.
DETAIL “A” – LOWERING
DETAIL “B” – UNHOOKING THE SLING
FIGURE 5.3 Transferring to a Bed
22
SECTION 6–USING THE PATIENT LIFT AS A STANDING AID
SECTION 6–USING THE PATIENT LIFT AS A STANDING AID
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: The Get-U-Up lift can be used as an aid to lift a person from a seated
to a standing position. A minimum of two caregivers is required for this
procedure.
NOTE: Invacare recommends that two assistants be used for all lifting
preparation and transferring to/from procedures; however, our equipment will
permit proper operation by one assistant. The use of one assistant is based on
the evaluation of the health care professional for each individual case.
Standing Procedure
-
Ensure that the patient is capable of walking.
-
Fold the footplate toward the mast.
-
Use the adjustment pins to move the kneepad to a comfortable position.
-
Ensure that the patient is in a sitting position to begin the procedure.
A. If the patient is in bed and requires assistance to come to a sitting position, raise the back of the bed.
B. Place the lift over the bed with the lift arms angled toward the patient.
C. Instruct the patient to grasp the lift arms.
D. As the lift arms are raised, the patient should move legs to the side of the bed until a sitting position on the side of the bed is assumed.
NOTE: The caregivers should assist as needed.
E. Pull the lift away as needed to accommodate the patient’s feet and legs over the side of the bed. -
With one caregiver on each side of the patient, place a gait belt on the patient.
-
Ensure that the patient is wearing shoes with skid-proof soles.
-
If the patient is on the bed, raise the bed (if possible) while ensuring that the patient’s feet remain on the floor.
-
If a walker is to be used, place the walker in front of the patient.
-
Advance the lift toward the patient so that the patient can easily reach the lift arms.
-
Instruct the patient to grasp the lift arms.
-
Ensure that the patient’s feet are directly under them.
-
Begin raising the lift arms.
-
On a count of three, instruct the patient to pull and stand-up while the caregivers support with the gait belt.
WARNING
DO NOT lift the patient.
14. Remove the lift to allow the patient to walk.
Note: The upper adjustment pins are typically used for shorter individuals. The lower adjustment pins are typically used for taller individuals.
FIGURE 6.1 Standing Procedure
23
SECTION 7–TROUBLESHOOTING
SECTION 7–TROUBLESHOOTING
SYMPTOMS | FAULTS | SOLUTION |
---|---|---|
Stand up lift feels loose. | Mast/base joint loose. | |
Tie-rods are loose. | Tighten the bolt, washer and locknut that secure the mast |
to the base.
Refer to Assembling the Patient Lift on page 10.
Casters/brakes noisy or stiff.| Fluff or debris in bearings.| Refer to
Hydraulic Pump on page 25.
Noisy or dry sound from pivots.| Needs lubrication.| Refer to Hydraulic Pump
on page 25.
Lift arms will not lower in uppermost position.| Lift arms require a minimum
weight load to lower from the uppermost position.| Pull down slightly on the
lift arms.
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
24
SECTION 8–MAINTENANCE
SECTION 8–MAINTENANCE
- When used for in-home use, a full maintenance safety inspection is required prior to each new user.
- If using the lift in the area of a shower or bath, ensure the lift is wiped clean of any moisture after use. Do not store the lift in a damp area or in a damp condition. Periodically inspect all components for signs of corrosion and replace those that are corroded or damaged.
- The lift is designed to provide a maximum of safe, efficient and satisfactory service with minimum care and maintenance. If you question the safety of any part of the lift, contact your dealer immediately and advise them of the problem.
- There is no adjustment or maintenance of the casters, other than cleaning, lubrication and checking axle and swivel bolts for tightness. Remove all debris, etc. from the caster and swivel bearings. If any parts are worn, replace these parts IMMEDIATELY.
- Regular cleaning will reveal loose or worn parts, enhance smooth operation and extend the life expectancy of the lift.
Maintenance Safety Inspection Checklist
ITEM| INITIALLY| INSTITUTIONAL
INSPECT/ADJUST MONTHLY
| IN-HOME
INSPECT/ADJUST PERIODICALLY
---|---|---|---
THE CASTER BASE
Inspect for missing hardware.
Base opens/closes with ease.
Inspect casters and axle bolts for tightness.
Inspect casters for smooth swivel and roll.
Inspect and clear casters of debris.
Inspect roll pin to ensure secure to base.
| ****
X
X
X
X
X
X
| ****
X
X
X
X
X
X
|
X
X
X
X
X
X
THE MAST
Mast MUST be securely assembled to lift arms.
Inspect for bends or deflections.
Inspect pivot joints for wear.
| ****
X
X
X
| ****
X
X
X
|
X
X
X
THE LIFT ARMS AND LINKAGE
Check all hardware and attachment points.
Inspect for bends or deflections.
Inspect bolted joints of lift arms for wear.
Inspect to ensure that the lift arms are centered between the base legs.
Ensure that the bolt is tightly secured.
Inspect pivot joints for wear.
| ****
X
X
X
X
X
X
| ****
X
X
X
X
X
X
|
X
X
X
X
X
X
SHIFTER HANDLE
Operates smoothly.
Locks adjustable base whenever engaged.
Check roll pin for wear
| ****
X
X
X
| ****
X
X
X
|
X
X
X
CLEANING
Whenever necessary.
| ****
X
| ****
X
|
X
SLINGS AND HARDWARE
Check all sling attachments for wear.
Inspect sling material for wear.
Inspect straps for wear.
Check that all labels are present and legible.
| ****
X
X
X
X
| ****
X
X
X
X
|
X
X
X
X
Hydraulic Pump
All parts of the hydraulic pump are precision machined then carefully assembled and tested to ensure reliable service. The pump assembly is completely enclosed and sealed with Neoprene rings to prevent leakage of hydraulic oil. A small amount of oil (about a drop) will accumulate around the piston from time to time and should be removed with a facial tissue.
WARNING
The hydraulic pump is sealed at the factory and if service is required, the pump unit MUST be returned to the factory for repair. DO NOT attempt to open the hydraulic pump or obtain local service. This will void the warranty and might result in damage and costly repair. Consult your dealer or contact Invacare for information.
25
Lubricating the Lift
A six month check and lubrication should ensure continued safety and
reliability of the patient lift.
The casters MUST swivel and roll smoothly. A light grease (waterproof auto
lubricant) may be applied to the ball bearing swivel of the casters once a
year. Apply more frequently if the casters are exposed to extreme moist
conditions.
Lubricate all pivot points. Wipe all excess lubricant from lift surface.
The connection between the mast and the base should be cleaned and coated with
petroleum jelly prior to assembly.
Detecting Wear and Damage
It is important to inspect all stressed parts, such as slings, sling
attachment knobs, and any pivot for slings for signs of cracking, fraying,
deformation or deterioration. Replace any defective parts immediately and
ensure that the lift is not used until repairs are made.
All sling metal parts should be inspected every three months and if wear is
apparent, replacement must be made.
Check that all labels are present and legible. Replace if necessary.
Cleaning the Sling and the Lift
The sling should be washed regularly in water temperature of 180°F (82°C) and a biological solution. A soft cloth, dampened with water and a small amount of mild detergent, is all that is needed to clean the lift. The lift can be cleaned with non-abrasive cleaners.
Replacing the Knee Pad
NOTE: For this procedure, refer to FIGURE 8.1.
- Remove the four button screws and washers that secure existing knee pad to the lift.
- Remove the existing knee pad from the lift.
- Position the mounting holes in the new knee pad with the mounting holes in the lift.
- Using the existing hardware, secure the new knee pad to the lift.
FIGURE 8.1 Replacing the Knee Pad
26
LIMITED WARRANTY
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW
APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of our products.
This warranty gives you specific legal rights and you may also have other
legal rights which vary from state to state.
Invacare warrants the products manufactured to be free from defects in
materials and workmanship for a period of three years on the lift and one year
on the hydraulic pump from the date of purchase. If within such warranty
period any such product shall be proven to be defective, such product shall be
repaired or replaced, at Invacare’s option. This warranty does not include any
labor or shipping charges incurred in replacement part installation or repair
of any such product. Invacare’s sole obligation and your exclusive remedy
under this warranty shall be limited to such repair and/or replacement.
For warranty service, please contact the dealer from whom you purchased your
Invacare product. In the event you do not receive satisfactory warranty
service, please write directly to Invacare at the address on the back cover,
provide dealer’s name, address, date of purchase, indicate nature of the
defect.
Invacare Corporation will issue a serialized return authorization. The
defective unit or parts MUST be returned for warranty inspection using the
serial number, when applicable as identification within 30 days of return
authorization date. Do not return products to our factory without our prior
consent. C.O.D. shipments will be refused; please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO
SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED,
PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR
STORAGE, PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTEN CONSENT
(INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED
PARTS OR ATTACHMENTS; PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY
COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE, OR TO A PRODUCT DAMAGED BY
CIRCUMSTANCES BEYOND INVACARE’S CONTROL, AND SUCH EVALUATION WILL BE SOLELY
DETERMINED BY INVACARE. THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM
NORMAL WEAR OR FAILURE TO ADHERE TO THE INSTRUCTIONS IN THIS MANUAL.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER
WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED
WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE
SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO
REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PERSUANT TO THE TERMS CONTAINED
HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND
BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL
NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW EXCLUSION OR LIMITATION OF INCIDENTAL OR
CONSEQUENTIAL DAMAGE, OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS, SO
THE ABOVE EXCLUSIONS AND LIMITATIONS MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE OR PROVINCIAL LAWS AND
REQUIREMENTS.
27
Invacare Corporation www.invacare.com
USA Canada
One Invacare Way 570 Matheson Blvd E Unit 8
Elyria, Ohio USA Mississauga Ontario
44036-2125 L4Z 4G4 Canada
800-333-6900
800-668-5324
All rights reserved. Trademarks are identified by the symbols ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
©2009 Invacare Corporation
Part No 1148115 Rev B – 06/09
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>