These release notes contain important information about version 1.6.1 of the AFG31000 software.
Introduction
This document provides supplemental information regarding the behavior of the AFG31000 software. This information is grouped into six categories:
Revision history| Lists the version of software, the document version,
and the date of the software release.
---|---
New features / enhancements| Summary of each significant new feature
included.
Problem fixes| Summary of each significant software/firmware bug fix
Known problems| Description of each significant known problem and ways to work
around it.
Installation instructions| Detailed instructions describing how to
install the software.
Appendix A – Previous versions| Contains information about previous
versions of the software.
Revision history
This document is periodically updated and distributed with releases and service packs to provide the most up-to-date information. This revision history is included below.
Date | Software version | Document number | Version |
---|---|---|---|
3/23/2021 | V1.6.1 | 0771639 | 02 |
12/3/2020 | V1.6.0 | 0771639 | 01 |
9/30/2019 | V1.5.2 | 0771639 | 00 |
11/15/2018 | V1.4.6 |
VERSION 1.6.1
Problem fixes
Issue number | AFG-676 |
---|---|
Models affected | AFG31XXX |
Symptom | Modulation issues on single-channel units. |
Resolution | This issue has been corrected. |
VERSION 1.6.0
New features/enhancements
Issue number | AFG-648 |
---|---|
Models affected | AFG31XXX |
Enhancement | Added a new SCPI command to get the MAC address of an |
AFG31XXX instrument: SYSTem: MAC ADDress?.
Problem fixes
Issue number | AFG-471 |
---|---|
Models affected | AFG31XXX |
Symptom | System could crash when running Instaview and then immediately |
changing the system
language setting.
Resolution| This issue has been corrected.
Issue number| AFG-474
Models affected| AFG31XXX
Symptom| Step 9 of the firmware installation section of the User’s
manual is incorrect.
Resolution| This issue has been corrected.
Issue number| AFG-484 / AR63489
Models affected| AFG31XXX
Symptom| A previously-installed feature license will disappear if the
system time zone setting is
changed to more than two of hours difference than originally set.
Resolution| This issue has been corrected.
Issue number| AFG-497 / AR63922
Models affected| AFG31XXX
Symptom| When two channels are both in Pulse mode, the pulse width
setting of one channel might
affect the other channel when an irrelevant pulse parameter is changed.
Resolution| This issue has been corrected.
Issue number| AFG-505
Models affected| AFG31XXX
Symptom| When using Burst mode with an external delay, the Trigger delay
value does not affect the
displacement of the waveform. This issue was introduced in the v1.5.2 release.
Resolution| This issue has been corrected.
Issue number| AFG-506 / AR63853
Models affected| AFG31XXX
Symptom| Incorrect PM Output formula in the “Modulate a waveform” topic
in the User’s manual.
Resolution| This issue has been corrected.
Issue number| AFG-508 / AR64101
Models affected| AFG31XXX
Symptom| Two-channel waveforms’ phases are not aligned in modulation and
sweep modes. The Align
Phase button does not function correctly in these modes.
Resolution| This issue has been corrected. The Align Phase button will re-align two-channel waveforms’
phases when pressed in continuous, modulation, and sweep modes.
Issue number| AFG-588 / AR64270
Models affected| AFG31XXX
Symptom| The displayed string length update routine was limited to
filenames less than 18 characters in length.
Resolution| The filename string length has been increased to 255
characters.
Issue number| AFG-598
Models affected| AFG31XXX
Symptom| The word “Frequency” is not translated into Chinese correctly.
Resolution| This issue has been corrected.
Issue number| AFG-624
Models affected| AFG31XXX
Symptom| SCPI command: SEQuence:ELEM[n]:WAVeform[m] does not default the
m parameter to 1 when unspecified.
Resolution| This issue has been corrected.
Issue number| AFG-630
Models affected| AFG31XXX
Symptom| The TRACE:DATA command example shown in the manual is
incorrect.
Resolution| This issue has been corrected.
Issue number| AFG-653 / AR64599
Models affected| AFG31XXX
Symptom| Not all settings are recalled when a setup is overwritten.
Resolution| This issue has been corrected.
Known problems
Issue number | AFG-663 |
---|---|
Models affected | AFG31XXX |
Symptom | The predefined equations in ArbBuilder will not compile with |
the default settings
Workaround| Change the range or the number of points for the equation to
compile properly.
Issue number| AFG-663
Models affected| AFG31XXX
Symptom| When running the Refresh Relay action using the Utility front
panel hard key, the display operations are not locked out, allowing other
functions to run.
Workaround| It is recommended that the Refresh Relay action be run using
the touch screen menus. If it is run using the Utility front panel hard key,
then do not select any other options from the touch screen until the action
completes.
Installation instructions
You can use the front-panel USB Type A connector to update your instrument firmware using a USB flash drive. This task is done using the front-panel touchscreen.
| CAUTION. Updating your instrument firmware is a sensitive operation; it
is critical that you follow the instructions below. If you do not, you may
cause damage to your instrument. For example, to prevent damage to the
instrument, do not remove the USB flash drive at any time while updating the
firmware, and do not power off the instrument during the update process
---|---
To update your instrument firmware:
-
Visit tek.com and search for the Series 31000 firmware.
-
Download the compressed .zip file to your computer.
-
Unzip the downloaded file and copy the .ftb file to the USB flash drive root directory.
-
Insert the USB in the AFG31000 Series instrument front panel.
-
Press the Utility button.
-
Select Firmware > Update.
-
Select the USB icon.
-
Select the file that you are using to update your instrument.
-
Select OK. You will see a message asking to confirm this update.
-
Verify that the instrument powered off and powered on to install the update.
-
Remove the USB drive.
NOTE. When using InstaView, each time a cable is changed, the firmware is upgraded, or the instrument is power-cycled, the cable propagation delay must be auto-measured or manually updated to make sure InstaView works properly.
Appendix A – Previous versions
V1.5.2
New features/enhancements
Issue number | AFG-131 / AR62531 |
---|---|
Models affected | AFG31XXX |
Enhancement | The AFG31000-RMK Rack Mount Kit is available for AFG31XXX |
models. Visit tek.com for details.
Issue number
Models affected
Enhancement| AFG-336
AFG31XXX
Updated language translations for the user interface.
Issue number
Models affected
Enhancement| AFG-373
AFG31XXX
Added SYSTem: RESTart SCPI command to reboot the instrument.
Issue number| AFG-430
Models affected| AFG31XXX
Enhancement| Waveform preview images will update immediately after
entering new values in Standard waveform view.
Issue number
Models affected
Enhancement| AFG-442
AFG31XXX
Display default brightness is now 100%.
Problem fixes
Issue number | AFG-21 / AR-62242 |
---|---|
Models affected | AFG3125X |
Symptom | Cannot create DC offset waveform in ArbBuilder for the AFG3125x |
in Sequence mode
Resolution| This issue has been corrected.
Issue number| AFG-186
Models affected| AFG3125X
Symptom| An application crash can occur when cancelling the Recall
Default Setup dialog, after closing the virtual keyboard, and when editing the
Point Draw Table of ArbBuilder.
Resolution| This issue has been corrected.
Issue number| AFG-193
Models affected| AFG31XXX
Symptom| Trig out should stay disabled when transitioning to DC
waveform.
Resolution| This issue has been corrected.
Issue number| AFG-194
Models affected| AFG31XXX
Symptom| In Burst mode, the graphical green arrow would not display when
starting to modify the Interval parameter.
Resolution| This issue has been corrected.
Issue number| AFG-198
Models affected| AFG31XXX
Symptom| On-screen keyboard crashes in some situations.
Resolution| This issue has been corrected.
Issue number| AFG-199
Models affected| AFG31XXX
Symptom| Graph refresh issue when selecting an ARB waveform for Mod
Shape using the Modulation function in Basic Mode.
**|
Resolution| This issue has been corrected.
Issue number| AFG-264
Models affected| AFG31XXX
Symptom| You should be prompted with a warning when you try to delete a
file that is not empty.
Resolution| This issue has been corrected.
Issue number| AFG-290
Models affected| AFG31XXX
Symptom| Screen capture functionality not working.
Resolution| This issue has been corrected. Press and hold both the left
and right keys in either order, then release either key.
Issue number| AFG-291 / AR62720
Models affected| AFG31XXX
Symptom| SCPI license commands not fully implemented.
Resolution| This issue has been corrected. See the AFG31000 Series
Arbitrary Function Generator Programmer’s Manual, available from tek.com.
Issue number| AFG-300
Models affected| AFG31XXX
Symptom**| Dual channel waveform alignment problems under the following
conditions:
- When using the manual trigger in burst mode with a positive slope setting, there is a 40 ns delay between channel 1 and channel 2.
- When using the manual trigger in burst mode with negative slope setting, there is a one- second delay between channel 1 and channel 2.
- If booting the unit in burst mode, the channel-to-channel phase alignment is incorrectly set to 2 ns.
Resolution| These issues have been corrected.
Issue number| AFG-303 / AR62139
Models affected| AFG31XXX
Symptom| When using the Japanese language setting in Basic mode,
changing from a Sine waveform to another type may cause the unit to hang.
Resolution| This issue has been corrected.
Issue number| AFG-308 / AR62443
Models affected| AFG31XXX
Symptom| This update fixes an issue setting the generated waveform using
the Recall feature in Basic mode. The pulse width was not always set properly,
causing unexpected results.
Resolution| This issue has been corrected.
Issue number| AFG-310 / AR62352
Models affected| AFG31XXX
Symptom| The user will not get the expected waveform when they attempt
an AM modulation with an Arb file greater than 4,096 points. The max points
for AM modulation using an Arb waveform is 4,096 points.
Resolution| This issue has been corrected. The product datasheet has
been updated.
Issue number| AFG-316 / AR62581
Models affected| AFG31XXX
Symptom| Undesired glitches could occur in burst mode idle state or when
turning the output on and off.
Resolution| This issue has been corrected.
Issue number| AFG-324
Models affected| AFG31XXX
Symptom| Instrument Ethernet connection using DHCP mode is unstable,
repeatedly disconnecting and reconnecting, for a long time with certain
network configurations.
Resolution| This issue has been corrected.
Issue number
Models affected
Symptom| AFG-330
AFG31XXX
Grammar and typographical errors in the automatic calibration dialog.
Resolution| This issue has been corrected
Issue number
Models affected
Symptom| AFG-337
AFG31XXX
Grammar and typographical errors in the self-diagnostics dialog.
Resolution| This issue has been corrected.
Issue number| AFG-352 / AR62937
Models affected| AFG31XXX
Symptom| In Sequence mode, the idle value of the signal is always the
offset of the waveform (or example, 2.5 V of a 0 to 5 Vpp waveform), This will
ultimately distort the customer’s desired waveform shape.
Resolution| Changed the default sequence mode from idle value to be 0 V
if the waveform can reach 0 V. Otherwise the idle value will be the offset.
Issue number| AFG-356
Models affected| AFG31XXX
Symptom| The ArbBuilder equation editor allows equation lines up to 256
characters in length to be entered, but it is limited to 80 characters per
line in the compiler.
Resolution| This issue has been corrected. The compiler now supports up
to the full 256 characters per line.
Issue number| AFG-374
Models affected| AFG31XXX
Symptom| Keyboard can appear partially off-screen.
Resolution| This issue has been corrected. This fix limits keyboard
positioning so that the keyboard is always displayed within the screen
boundaries.
Issue number| AFG-376
Models affected| AFG31XXX
Symptom| Advanced Sequence view incorrectly allowed selection of .tfw
files
Resolution| This issue has been corrected. .tfw files are not supported
in the Advanced sequence view.
Issue number| AFG-391
Models affected| AFG31XXX
Symptom| Advanced Sequence menu sometimes left the New and Save buttons
selected.
Resolution| This issue has been corrected.
Issue number| AFG-411
Models affected| AFG31XXX
Symptom
Resolution
| Scrolling the sequence table is too sensitive.
This issue has been corrected.
Issue number| AFG-422
Models affected| AFG31XXX
Symptom| Running the Refresh Relay operation is too lengthy.
Resolution| The issue has been corrected. The Refresh Relay operation
has been reduced to 250 cycles.
Issue number| AFG-427
Models affected| AFG31XXX
Symptom
Resolution
| The 123 button of the soft alpha-numeric keyboard does not work with some
plugins. This issue has been corrected.
Issue number| AFG-437
Models affected| AFG31XXX
Symptom
Resolution
| Selecting x on the small numeric virtual keyboard should issue a cancel
request and close the dialog.
This issue has been corrected.
Known problems
Issue number
Models affected
Symptom| AFG-380
AFG31XXX
The predefined equations in ArbBuilder will not compile with the default
settings.
---|---
**Workaround| ** Change the range or the number of points for the
equation to compile properly.
V1.4.6
Issue number
Models affected
Enhancement| 1
AFG31151, AFG31152, AFG31251, and AFG31252
Support the AFG31151, AFG31152, AFG31251, and AFG31252 models.
---|---
Issue number
Models affected Enhancement| 2
AFG31151, AFG31152, AFG31251, and AFG31252
Optimized user interface.