BW GasAlertMicro 5/PID/IR User Manual
- June 7, 2024
- Honeywell
Table of Contents
GasAlertMicro 5/PID/IR
1, 2, 3, 4, 5 Gas Detector
User Manual
Limited Warranty and Limitation Liability
BW Technologies LP (BW) warrants the product to be free from defects in
material and workmanship under normal use and service for a period of two
years, beginning on the date of shipment to the buyer. This warranty extends
only to the sale of new and unused products to the original buyer. BW’s
warranty obligation is limited, at BW’s option, to refund of the purchase
price, repair or replacement of a defective product that is returned to a BW
authorized service center within the warranty period. In no event shall BW’s
liability hereunder exceed the purchase price actually paid by the buyer for
the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the
normal wear and tear of the product arising from use; b) any product which in
BW’s opinion, has been misused, altered, neglected or damaged, by accident or
abnormal conditions of operation, handling or use; c) any damage or defects
attributable to repair of the product by any person other than an authorized
dealer, or the installation of unapproved parts on the
product; or The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance
with the product manual instructions and any other applicable recommendations
of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly
making the product available for correction. No goods shall be returned to BW
until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as
the original invoice, bill of sale or packing slip to establish that the
product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY
AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT
LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR
BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries
or states do not allow limitation of the term of an implied warranty, or
exclusion or limitation of incidental or consequential damages, the
limitations and exclusions of this warranty may not apply to every buyer. If
any provision of this warranty is held invalid or unenforceable by a court of
competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.
Warranty Registration
http://www.honeywellanalytics.com/support/product-registration
Contacting BW Technologies by Honeywell
USA & Canada: 1-888-749-8878 Bwa.customerservice@honeywell.com
Europe: 00800-333-222-44 www.honeywellanalytics.com
Other countries: 1-403-248-9226
Table of Contents
Title
Page
Introduction ……………………………………………………………………………………………………………………………………………………. 1 Gases
Monitored …………………………………………………………………………………………………………………………………………. 2
Safety Information – Read First…………………………………………………………………………………………………………………………
2 aCautions …………………………………………………………………………………………………………………………………………………….. 3
Sensor Poisons and Contaminants …………………………………………………………………………………………………………………. 6
Getting Started ……………………………………………………………………………………………………………………………………………….. 8
Parts of the GasAlertMicro 5/PID/IR
…………………………………………………………………………………………………………………. 9 Screen Elements
…………………………………………………………………………………………………………………………………………… 10 Buttons
………………………………………………………………………………………………………………………………………………………… 11 Activating the
Detector ……………………………………………………………………………………………………………………………………………………….. 12
Self-Tests………………………………………………………………………………………………………………………………………………….. 12 Battery
Test ………………………………………………………………………………………………………………………………………….. 12 Datalogging
(Optional)……………………………………………………………………………………………………………………………. 13 Pump Test
……………………………………………………………………………………………………………………………………………. 15 Due-Lock Enabled
…………………………………………………………………………………………………………………………………. 17 Force Calibration
Enabled ………………………………………………………………………………………………………………………. 17 Bump Daily Enabled
………………………………………………………………………………………………………………………………. 18
Self-Test Pass …………………………………………………………………………………………………………………………………………… 19 Self
Test Fail……………………………………………………………………………………………………………………………………………… 19 Battery
Test ………………………………………………………………………………………………………………………………………………. 20 Datalogger
Operation (Optional)…………………………………………………………………………………………………………………… 20
Deactivating the Detector………………………………………………………………………………………………………………………………. 21
i
GasAlertMicro 5/PID/IR User Manual
Title
Page
User Options Menu ……………………………………………………………………………………………………………………………………….. 21 Exit User Options Menu ………………………………………………………………………………………………………………………………. 22 Options Menu…………………………………………………………………………………………………………………………………………….. 22 Backlight ………………………………………………………………………………………………………………………………………………. 23 Confidence Beep …………………………………………………………………………………………………………………………………… 23 Due-Lock ……………………………………………………………………………………………………………………………………………… 23 Latched Alarms ……………………………………………………………………………………………………………………………………… 24 Passcode Protect…………………………………………………………………………………………………………………………………… 24 Safe Display………………………………………………………………………………………………………………………………………….. 25 Sensor Configuration ………………………………………………………………………………………………………………………………….. 25 Sensor Enable/Disable …………………………………………………………………………………………………………………………… 26 Span Gas Value…………………………………………………………………………………………………………………………………….. 27 STEL Period …………………………………………………………………………………………………………………………………………. 28 TWA Method …………………………………………………………………………………………………………………………………………. 28 Resolution …………………………………………………………………………………………………………………………………………….. 29 %Vol CO2 (CO2 Sensors Only)………………………………………………………………………………………………………………… 29 %Vol CH4 (LEL Sensors Only)…………………………………………………………………………………………………………………. 30 Correction Factor (CF) ……………………………………………………………………………………………………………………………. 30 Logger Option ……………………………………………………………………………………………………………………………………………. 33 Clock Option ……………………………………………………………………………………………………………………………………………… 33 Language Selection ……………………………………………………………………………………………………………………………………. 34 Tech Mode ………………………………………………………………………………………………………………………………………………… 35 Sensors………………………………………………………………………………………………………………………………………………… 36 Initialize ………………………………………………………………………………………………………………………………………………… 37 Force Calibration …………………………………………………………………………………………………………………………………… 37 Bump Daily …………………………………………………………………………………………………………………………………………… 38 Stealth Mode…………………………………………………………………………………………………………………………………………. 38
ii
GasAlertMicro 5/PID/IR User Manual
Title
Page
Sleep Mode…………………………………………………………………………………………………………………………………………… 39 Alarms
………………………………………………………………………………………………………………………………………………………….. 40
Gas Exposures Computed …………………………………………………………………………………………………………………………… 43
Viewing Gas Exposures ………………………………………………………………………………………………………………………………. 43
Clearing Gas Exposures ……………………………………………………………………………………………………………………………… 44 Gas
Alarm Setpoints …………………………………………………………………………………………………………………………………… 44 Viewing
the Alarm Setpoints ………………………………………………………………………………………………………………………… 44
Resetting Gas Alarm Setpoints ……………………………………………………………………………………………………………………..
45 Stopping a Gas Alarm …………………………………………………………………………………………………………………………………. 46
Sensor Alarm …………………………………………………………………………………………………………………………………………….. 46 Pump
Alarm ………………………………………………………………………………………………………………………………………………. 46 Low Battery
Alarm………………………………………………………………………………………………………………………………………. 47 Automatic
Deactivation Alarm ………………………………………………………………………………………………………………………. 47 Bump
Test…………………………………………………………………………………………………………………………………………………….. 47 Performing a
Bump Test ……………………………………………………………………………………………………………………………… 47 Calibration and
Setting Alarm Setpoints ………………………………………………………………………………………………………… 48 Guidelines
…………………………………………………………………………………………………………………………………………………. 48 Diagnostics
Testing…………………………………………………………………………………………………………………………………….. 49 Applying Gas to
the Sensors ……………………………………………………………………………………………………………………….. 50
Single Gas Calibration Cap …………………………………………………………………………………………………………………….. 50
Calibration Procedure …………………………………………………………………………………………………………………………………. 51
Start Calibration …………………………………………………………………………………………………………………………………….. 52 Auto
Zero and Oxygen (O2) Sensor Calibration …………………………………………………………………………………………. 52
Zero CO2 (GasAlertMicro 5 IR only) ………………………………………………………………………………………………………….
52 Passcode Protect Activated ……………………………………………………………………………………………………………………..
53 Auto Span …………………………………………………………………………………………………………………………………………….. 54
Successful Span ……………………………………………………………………………………………………………………………………. 56
iii
GasAlertMicro 5/PID/IR User Manual
Title
Page
Setting the Calibration Due Date ……………………………………………………………………………………………………………… 57 Alarm Setpoints …………………………………………………………………………………………………………………………………….. 58 Setting the Remaining Alarm Setpoints …………………………………………………………………………………………………….. 60 Finish Calibration …………………………………………………………………………………………………………………………………… 60 Verification ……………………………………………………………………………………………………………………………………………. 61 Unsuccessful Span ………………………………………………………………………………………………………………………………… 61 Pump ……………………………………………………………………………………………………………………………………………………………. 64 ………………………………………………………………………………………………………………………………………………………………… 64 Identifying the Pump …………………………………………………………………………………………………………………………………… 64 Generation 1: 116885-L3 (yellow) and 118933-L3 (black)……………………………………………………………………………. 64 Generation 2: 130916-L3 (yellow) and 130917-L3 (black)……………………………………………………………………………. 65 Installing the Pump Module …………………………………………………………………………………………………………………………… 66 Replacing the Pump Filter (Generation 2 Pump) …………………………………………………………………………………………….. 67 Replacing the Pump Nozzle (Generation 2 Pump) ………………………………………………………………………………………….. 68 Attaching the Auxiliary Filter ………………………………………………………………………………………………………………………… 68 Confined Space Sampling …………………………………………………………………………………………………………………………… 70 Maximum Hose Length when Confined Space Sampling…………………………………………………………………………….. 71 Datalogger ……………………………………………………………………………………………………………………………………………………. 72 MMC/SD Card Compatibility…………………………………………………………………………………………………………………………… 72 Inserting the MMC/SD Card …………………………………………………………………………………………………………………………. 72 MMC/SD Card Troubleshooting ……………………………………………………………………………………………………………………… 73 Restoring Datalog Files……………………………………………………………………………………………………………………………….. 74 Reformatting the MMC/SD Card …………………………………………………………………………………………………………………… 75 Import Datalogs to Fleet Manager II ……………………………………………………………………………………………………………….. 76 Minimum PC Requirements …………………………………………………………………………………………………………………………. 76 Importing from MicroDock II to Fleet Manager II……………………………………………………………………………………………… 76 Import to Fleet Manager II Using a Card Reader …………………………………………………………………………………………….. 76
iv
GasAlertMicro 5/PID/IR User Manual
Title
Page
View Datalog Files in Spreadsheets
……………………………………………………………………………………………………………….. 77 Example of a Datalog
Spreadsheet ………………………………………………………………………………………………………….. 78
Maintenance …………………………………………………………………………………………………………………………………………………. 81
Battery Cautions ……………………………………………………………………………………………………………………………………….. 81
Charging the Battery …………………………………………………………………………………………………………………………………… 81
Replacing the Alkaline Batteries ……………………………………………………………………………………………………………………
82 Replacing the Lithium Battery Pack
………………………………………………………………………………………………………………. 83 Replacing a Sensor or Sensor
Filter ……………………………………………………………………………………………………………… 83 Photoionization Detector
(PID) …………………………………………………………………………………………………………………….. 85 Clean or Replace the
Lamp …………………………………………………………………………………………………………………….. 85 Replace the
Lamp………………………………………………………………………………………………………………………………….. 86 Replace the
Electrode Stack …………………………………………………………………………………………………………………… 87 WEEE Battery
Directive ………………………………………………………………………………………………………………………………. 87 Removal and
Disposal of the Battery Pack………………………………………………………………………………………………… 87 Removal
and Disposal of the Coin Cell …………………………………………………………………………………………………….. 87
Troubleshooting ……………………………………………………………………………………………………………………………………………. 89
Replacement Parts and Accessories ………………………………………………………………………………………………………………
95 Specifications ……………………………………………………………………………………………………………………………………………….. 97
General Specifications for Datalogger Units
…………………………………………………………………………………………………. 100 GasAlertMicro 5/PID/IR Downloadable
Datalogger ……………………………………………………………………………………….. 101 PID Correction Factor (CF)
Library……………………………………………………………………………………………………………….. 102
v
GasAlertMicro 5/PID/IR User Manual
Title
Page
vi
List of Figures
Figure Title
Page
Parts of the GasAlertMicro 5/PID/IR………………………………………………………………………………………………………. 9
Screen Elements ………………………………………………………………………………………………………………………………. 10
Applying Gas to the Sensors ………………………………………………………………………………………………………………. 50
Single Gas Calibration Cap ………………………………………………………………………………………………………………… 51
Removing the Single Gas Calibration Cap ……………………………………………………………………………………………. 51
Installing the Pump Module ………………………………………………………………………………………………………………… 66
Replacing the Pump Filter (Generation 2 Pump)……………………………………………………………………………………. 67
Replacing the Pump Nozzle ……………………………………………………………………………………………………………….. 68
Attaching the Auxiliary Filter ………………………………………………………………………………………………………………. 68
Attaching the Filter Cord……………………………………………………………………………………………………………………. 69
Attaching the Sample Probe……………………………………………………………………………………………………………….. 70
Inserting/Removing the MMC/SD Card ………………………………………………………………………………………………… 73
Replacing the Alkaline Batteries………………………………………………………………………………………………………….. 82
Replacing the Lithium Battery Pack …………………………………………………………………………………………………….. 83
Replacing a Sensor or Sensor Filter ……………………………………………………………………………………………………. 84
Parts of the PID ………………………………………………………………………………………………………………………………… 86
vii
GasAlertMicro 5/PID/IR User Manual
viii
List of Tables
Table Title
Page
Gases Monitored ………………………………………………………………………………………………………………………………… 2
Sensor Poisons and Contaminants ……………………………………………………………………………………………………….. 6
International Symbols ………………………………………………………………………………………………………………………….. 7
Parts of the GasAlertMicro 5/PID/IR………………………………………………………………………………………………………. 9
Screen Elements ………………………………………………………………………………………………………………………………. 10
Button ……………………………………………………………………………………………………………………………………………… 11
Alarms …………………………………………………………………………………………………………………………………………….. 40
Computed Gas Exposures …………………………………………………………………………………………………………………. 43
Gas Alarm Setpoints …………………………………………………………………………………………………………………………. 44
OSHA Sample Factory Alarm Setpoints……………………………………………………………………………………………….. 45
Applying Gas to the Sensors ………………………………………………………………………………………………………………. 50
Single Gas Calibration Cap ………………………………………………………………………………………………………………… 51
Time Required to Span ……………………………………………………………………………………………………………………… 55
Installing the Pump Module ………………………………………………………………………………………………………………… 66
Generation 1 Pump Maximum Hose Length …………………………………………………………………………………………. 71
Generation 2 Pump Maximum Hose Length …………………………………………………………………………………………. 71
Datalog Spreadsheet Example……………………………………………………………………………………………………………. 78
Datalog Status Codes ……………………………………………………………………………………………………………………….. 79
Datalog Gas and Correction Factor Sensor Codes ………………………………………………………………………………… 80
Replacing the Alkaline Batteries………………………………………………………………………………………………………….. 82
Replacing a Sensor or Sensor Filter …………………………………………………………………………………………………… 85
Parts of the PID sensor ……………………………………………………………………………………………………………………… 85
Troubleshooting………………………………………………………………………………………………………………………………… 89
ix
GasAlertMicro 5/PID/IR User Manual
Replacement Parts and Accessories …………………………………………………………………………………………………… 95
PID Corrections Factor (CF) Library…………………………………………………………………………………………………… 102
x
ISO 9001
GasAlertMicro 5/PID/IR Introduction
GasAlertMicro 5/PID/IR
Introduction
a Warning To ensure personal safety, read the Safety Information Read First
and Cautions before using the detector. The GasAlertMicro 5, GasAlertMicro 5
PID, and GasAlertMicro 5 IR gas detectors (“the detector”) warn of hazardous
gas at levels above user-defined alarm setpoints. The detector is a personal
safety device. It is your responsibility to respond properly to the alarm.
Note Unless reference is made to a specific detector model, the GasAlertMicro
5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR detectors are referred to as
GasAlertMicro 5/PID/IR. Refer to Gases Monitored prior to operating the
detector.
1
GasAlertMicro 5/PID/IR User Manual
Gases Monitored
The following table lists the gases that are monitored by the detector. Table
- Gases Monitored
Gas Detected Oxygen (O2)
Combustible gases (LEL)
Carbon monoxide (CO) Hydrogen sulfide (H2S) Phosphine (PH3) Sulfur oxide (SO2)
Chlorine (Cl2) Ammonia (NH3) Nitrogen dioxide (NO2) Hydrogen cyanide (HCN)
Chlorine dioxide (ClO2) Ozone (O3) Volatile organic compounds (VOC) Carbon
dioxide (CO2) 2
Unit of Measure
% volume a) percent of lower explosive
limit (%LEL) b) percent by volume
methane 0-5.0% v/v parts per million (ppm) parts per million (ppm) parts per
million (ppm) parts per million (ppm) parts per million (ppm) parts per
million (ppm) parts per million (ppm) parts per million (ppm) parts per
million (ppm) parts per million (ppm)
parts per million (ppm)
parts per million or %vol CO2
CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY
QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND THIS USER MANUAL COMPLETELY
BEFORE OPERATING AND SERVICING.
Safety Information – Read First
Use the detector only as specified in this user manual, otherwise the
protection provided by the detector may be impaired.
International symbols used on the detector and in this user manual are defined
in Table 3.
Read the Cautions on the following pages before using the detector.
ec Warning
This instrument contains batteries. Do not mix with the solid waste stream.
Spent batteries must be disposed of by a qualified recycler or hazardous
materials handler. Dispose of lithium cells immediately. Do not disassemble
and do not dispose of in fire. Do not mix with the solid waste stream. Spent
batteries must be disposed of by a qualified recycler or hazardous materials
handler.
aCautions
· Warning: Substitution of components may impair Intrinsic Safety.
· Caution: For safety reasons, this equipment must be operated and serviced by
qualified personnel only. Read and understand this user manual completely
before operating or servicing.
· Do not use the detector if it is damaged. Inspect the detector before using.
Look for cracks and/or missing parts.
· If the detector is damaged or parts are missing, contact BW Technologies by
Honeywell immediately.
· Use only sensor(s) that are specifically designed for the GasAlertMicro
5/PID/IR detectors. Refer to Replacement Parts and Accessories.
· Calibrate the detector before first-time use and then on a regular schedule,
depending on use and sensor exposure to poisons and contaminants. Sensors must
be calibrated regularly and at least once every 180 days (6 months).
· BW recommends to bump test the sensors, before each day’s use, to confirm
their ability to respond to gas by exposing the detector to a gas
concentration that exceeds the alarm setpoints. Manually verify that the
audible and visual alarms are activated. Calibrate if the readings are not
within the specified limits.
· BW recommends the combustible sensor be checked with a known concentration
of calibration gas after any known exposure to catalyst contaminants/poisons
(sulfur compounds, silicon vapors, halogenated compounds, etc).
GasAlertMicro 5/PID/IR aCautions
· The combustible sensor is factory calibrated to 50% LEL methane. If
monitoring a different combustible gas in the % LEL range, calibrate the
sensor using the appropriate gas.
· Warning: High off-scale LEL readings may indicate an explosive
concentration.
· Only the combustible gas detection portion of this instrument has been
assessed for performance by CSA International.
· Protect the combustible sensor from exposure to lead compounds, silicones,
and chlorinated hydrocarbons. Although certain organic vapors (such as leaded
gasoline and halogenated hydrocarbons) may temporarily inhibit sensor
performance, in most cases, the sensor will recover after calibration.
aCautions
· Before using common products around sensors, refer to Sensor Poisons and
Contaminants.
· High concentrations of certain toxic gases, for example H2S, may have an
adverse effect on the LEL sensor. This effect, known as inhibition, is usually
temporary but in extreme circumstances can impair the sensitivity of the LEL
sensor. After any gas exposure that causes an alarm in the toxic gas sensors,
the LEL sensor should be verified with a bump test, and recalibrated if
necessary.
· Any rapid up-scaling reading followed by a declining or erratic reading may
indicate a gas concentration beyond upper scale limit, which may be hazardous.
3
GasAlertMicro 5/PID/IR User Manual
· Calibrate only in a safe area that is free of hazardous gas, in an
atmosphere of 20.9% oxygen.
· Use only BW approved batteries for the GasAlertMicro 5/ PID/IR detectors.
Refer to Replacement Parts and Accessories.
· Charge the detector before first-time use. BW recommends the detector be
charged after every workday.
· Charge the battery pack immediately when a low battery alarm occurs.
· Read and adhere to the battery cautions provided in Battery Cautions.
· Charge the GasAlertMicro 5/PID/IR batteries using the recommended charging
adapter only. Do not use any other charging adapter. Failure to adhere to this
caution can lead to fire and/or explosion.
· Read and adhere to all instructions in the charger user manual. Failure to
do so can result in fire, electrical shock, personal injury, and/or property
damage.
· Extended exposure of the GasAlertMicro 5/PID/IR detectors to certain
concentrations of combustible gases and air may stress a detector element,
which can seriously affect its performance. If an alarm occurs due to high
concentration of combustible gases, recalibration should be performed, or if
needed, the sensor replaced.
aCautions · Protect the PID sensor from exposure to silicone vapors.
· When calibrating O3 and ClO2 sensors that are located in the Toxic 2 sensor
position, a single gas calibration cap must be used to ensure accurate
calibration. For more information, refer to Single Gas Calibration Cap.
· Replace the CO2 sensor only in a safe and non-hazardous area that is free of
hazardous gas.
· The optional pump (M5-PUMP) is certified for use with the GasAlertMicro
5/PID/IR detectors only.
· Do not immerse the detector in liquids. · Do not test the combustible
sensor’s response with a
butane cigarette lighter; doing so will damage the sensor. · Do not expose the
detector to electrical shock or severe
continuous mechanical shock. · Do not attempt to disassemble, adjust, or
service the
detector unless instructions for that procedure are provided in the user
manual and/or that part is listed as a replacement part. Use only BW
Technologies by Honeywell Replacement Parts and Accessories. · The detector
warranty is void if customer, personnel, or third parties damage the detector
during repair attempts. Repair attempts made by non-BW Technologies by
Honeywell repair/service personnel voids this warranty. Lithium Battery Packs
· Warning: The lithium battery (M5-BAT08) may present a risk of fire or
chemical burn hazard if misused. Do not disassemble, heat above 100°C (212°F),
or incinerate.
4
· Do not use any other lithium batteries with the GasAlertMicro 5/PID/IR
detectors. Use of any other cell can cause fire and/or explosion. To order and
replace the M5BAT08 battery, refer to Replacement Parts and Accessories.
· Warning: Lithium polymer cells exposed to heat at 130°C (266°F) for 10
minutes can cause fire and/or explosion.
· Dispose of used lithium cells immediately. Do not disassemble and do not
dispose of in fire. Do not mix with the solid waste stream. Spent batteries
must be disposed of by a qualified recycler or hazardous materials handler.
· Keep lithium cells away from children. · Products may contain materials that
are regulated for
transportation under domestic and international dangerous goods regulations.
Return product in compliance with appropriate dangerous goods regulations.
Contact freight carrier for further instructions.
GasAlertMicro 5/PID/IR aCautions
5
GasAlertMicro 5/PID/IR User Manual
Sensor Poisons and Contaminants
Several cleaners, solvents, and lubricants can contaminate and cause permanent
damage to sensors. Before using cleaners, solvents, and lubricants in close
proximity to the detector sensors, read and adhere to the following caution
and table.
a Caution Use only the following BW Technologies by Honeywell recommended
products and procedures: · Use water based cleaners. · Use non-alcohol based
cleaners. · Clean the exterior with a soft, damp cloth. · Do not use soaps,
polishes, or solvents.
6
The following table lists common products to avoid using around sensors.
Table 2. Sensor Poisons and Contaminants
Cleaners and Lubricants
Brake cleaners
Lubricants
Rust inhibitors
Window and glass cleaners Dishsoaps
Citrus based cleaners Alcohol based cleaners Hand sanitizers Anionic
detergents Methanol (fuels and antifreezes)
Silicones
Silicone cleaners and protectants
Silicone based adhesives, sealants, and gels
Hand/body and medicinal creams that contain silicone
Tissues containing silicone
Mold releasing agents
Polishes
Aerosols Bug repellents and sprays Lubricants
Rust inhibitors
Window and glass cleaners
Table 3. International Symbols
Symbol
n
Description Approved to both U.S. and Canadian Standards by CSA International
GasAlertMicro 5/PID/IR Sensor Poisons and Contaminants
g
European Explosives Protection
X
Conforms to European Union Directives
ATEX IECEx EAC Ex
Conforms to European ATEX Directives
International Electrotechnical Commission Scheme for Certification to
Standards for Electrical Equipment for Explosive Atmospheres
Conforms to Russian Custom Union Certification and Declaration
Conforms to Korea Testing Laboratory (KTL) Certification
Conforms to Brazilian InMetro Certification
Australian Regulatory Compliance Mark 7
GasAlertMicro 5/PID/IR User Manual
Getting Started
The list below provides the standard items included with the detector. If the
detector is damaged or parts are missing, contact the place of purchase
immediately.
· Batteries: Three replaceable alkaline cells with battery pack, or one
rechargeable lithium battery pack
· Sensors: O2, combustible (LEL), toxic, H2S/CO (TwinTox sensor), PID, or CO2
· Calibration cap and hose · Single gas calibration cap · Screwdriver · Quick
reference guide · Quick reference card · Technical Reference Guide and
training on CD-ROM · Fleet Manager II (if applicable)
To order parts, refer to Replacement Parts and Accessories.
The detector is shipped with the sensors, and battery packs installed. To
replace sensors, the pump, or the battery pack, refer to Replacement Parts and
Accessories.
To become oriented with the features and functions of the detector, refer to
the following figures and tables:
· Figure 1. and Table 4. describe the detector’s components.
· Figure 2. and Table 5. describe the detector’s screen elements.
· Table 6. describes the detector’s button.
8
Parts of the GasAlertMicro 5/PID/IR
Figure 1. Parts of the GasAlertMicro 5/PID/IR
GasAlertMicro 5/PID/IR Parts of the GasAlertMicro 5/PID/IR
Table 4. Parts of the GasAlertMicro 5/PID/IR
Item 1 2 3 4
5
6 7 8 9 10 11
Description
Liquid crystal display (LCD) Buttons Audible alarms Toxic 2 sensor Toxic 1/PID
sensor (GasAlertMicro 5 PID) or Toxic 1/IR (CO2) sensor (GasAlertMicro 5 IR)
Visual alarm indicators (LEDs) LEL sensor Oxygen sensor Buttons Battery pack
Alligator clip
9
GasAlertMicro 5/PID/IR User Manual
Screen Elements
Figure 2. Screen Elements Note
If enabled, the backlight automatically activates for 8 seconds when there is
an alarm condition and whenever there is insufficient light to view the LCD.
Any button reactivates the backlight in low light conditions. 10
Table 5. Screen Elements
Item 1 2 3 4 5 6 7 8 9 10 11
12
13
Description
Alarm condition Automatically span sensor Gas cylinder Gas type Battery life
indicator Passcode lock Data transmission Clock Stealth mode Pump indicator
(optional) MMC indicator (optional) Alarm condition (low, high, TWA, STEL, or
multi alarm) or view TWA, STEL and peak (MAX) gas exposures Automatically zero
sensor
Buttons
Button
A
G
H C
GasAlertMicro 5/PID/IR Buttons
Table 6. Button
Description · To activate the detector press A. · To deactivate the detector,
press and hold A until the countdown is complete. · To increment the displayed
value or scroll up, press G. · To enter the user options menu, press and hold
G and H simultaneously until the countdown is complete. · To clear the TWA,
STEL, and peak (MAX) gas exposure readings, press and hold C and G
simultaneously
until the countdown is complete. · To view the date and time, alarm setpoints
(TWA, STEL, low, high) of all sensors, and the LEL/PID correction factor
(if applicable) press G. · To decrement the displayed value or scroll down,
press H. · To initiate calibration and to define alarm setpoints, press and
hold C and H simultaneously until the countdown
is complete. · To view the TWA, STEL, and peak (MAX) gas exposure readings,
press C. · To acknowledge latched alarms, press C. Refer to Latched Alarms on
page 24.
11
GasAlertMicro 5/PID/IR User Manual
Activating the Detector
If using the pump module, attach it and the pump accessories prior to
activating the detector. For illustrations and procedures, refer to the
following:
·. · Confined Space Sampling · Replacing the Pump Filter (Generation 2 Pump) ·
Replacing the Pump Nozzle (Generation 2 Pump) · Attaching the Auxiliary Filter
Note Only activate the detector in a safe atmosphere that is free of hazardous
gas in an atmosphere of 20.9% oxygen.
To activate the detector, press A.
Self-Tests
When the detector is activated, it performs several self-tests. Confirm the
following tests occur.
Note If an error message displays during the self-test, refer to
Troubleshooting.
Battery Test
The detector performs a battery test during startup. If the battery has
insufficient power to operate, the following screen displays before
deactivating.
Replace the batteries and reactivate the detector. 1. All of the LCD elements
display simultaneously as the detector beeps, flashes, vibrates, and briefly
activates the backlight.
12
2. The version and serial number of the detector displays. 3. The date and time displays.
GasAlertMicro 5/PID/IR Activating the Detector
Note
If there is a problem with the MMC/SD card, Datalogger disabled displays. The
detector then automatically continues with the self-test.
If the card requires formatting, the following screen displays as the card is
automatically formatted.
Datalogging (Optional)
4. If the detector is a datalogging unit, it determines if · a MultiMediaCard
(MMC) or Secure Digital (SD) card is inserted, · the detector can communicate
with the card, · the detector supports the size of the card, and · the card
requires formatting.
5. The detector then runs a self-test to verify the sensors and power supply
are operating correctly.
13
GasAlertMicro 5/PID/IR User Manual
Self-test Successful: If successful, the following screen displays.
7. The TWA, STEL, low, and high alarm setpoints then display in the following order (left to right).
TWA
STEL
Self-test Unsuccessful: If a sensor fails the self-test, a warning displays indicating which sensor(s) has failed.
Low
High
6. If correction factors are set in the user options, the LEL or PID (custom) correction factors display.
Note
The alarm setpoints may vary by region. Refer to Resetting Gas Alarm
Setpoints.
14
Pump Test a Warning
If the pump module is installed on the detector, the following three things
must occur during start-up. If any one of the conditions below does not occur,
discontinue use of the detector and contact BW Technologies by Honeywell
immediately. · The detector prompts for a pump test during start-up · The pump
module passes the pump test at start-up when
the pump inlet or sample chain inlet is blocked · The J icon displays on the
LCD 8. If the pump module is attached to the detector, the following
screens display.
When the following screen displays, block the pump inlet.
GasAlertMicro 5/PID/IR Activating the Detector
15
GasAlertMicro 5/PID/IR User Manual
If the pump inlet is not blocked within 10 seconds or the pump test fails, the
following screens display.
9. Unless disabled in user options, the oxygen (O2) sensor is calibrated automatically.
If C is not pressed or the pump is not removed within 25 seconds, the detector
performs the pump test again.
Note
The pump may require a new filter. Refer to Replacing the Pump Filter
(Generation 2 Pump).
If the pump alarm persists, remove the battery pack to deactivate the detector
and then refer to the “Pump Operation” section of Troubleshooting.
If the pump test is successful, the following screen displays and the self-
test continues.
If the calibration is successful, the detector beeps twice.
Note If the automatic O2 calibration feature has been disabled, Automatic O2
span disabled displays.
10. The number of days remaining before calibration is due displays for all
sensors.
16
If any sensor is over due for calibration, the LCD displays the name of the sensor and the number of days past due.
GasAlertMicro 5/PID/IR Activating the Detector
If no passcode is entered, or it is entered incorrectly, the following screen
displays.
Due-Lock Enabled
The Due-lock option ensures that a passcode must be entered when calibration
is past due, otherwise the detector automatically deactivates.
11. If Due-Lock is enabled in the user options, the following screen
displays.
To enable/disable this option, refer to Due-Lock in the user options menu.
Also refer to Passcode Protect.
Force Calibration Enabled
12. If Force cal (force calibration) is enabled in tech mode, calibration is
mandatory before the detector enters normal operation.
Refer to Force Calibration in Tech Mode to enable/disable, and refer to
Calibration Procedure to calibrate.
Enter the correct passcode and press C to confirm.
Note If any sensor is overdue, Y displays continually until calibration is
performed.
If C is not pressed to start calibration, the following screen displays and
the detector deactivates.
17
GasAlertMicro 5/PID/IR User Manual
Successful Bump Test: If the bump test passes, the following screens display.
Bump Daily Enabled a Caution
BW recommends that a bump test to all sensors be performed every 24 hours
prior to the beginning of the work shift.
13. If Bmp Daily (bump daily) is enabled in tech mode, the following screens
display.
The detector waits for the sensor(s) to clear (30 seconds) and then enters
normal operation.
Unsuccessful Bump Test: If the bump test is unsuccessful or the bump test is
not performed, the following screen displays and the detector deactivates.
If a bump test of the LEL and O2 sensor is not performed, the detector will
deactivate.
Apply the LEL gas and then apply the O2 (a higher or lower percentage than the
default 20.9%, such as 18% O2).
18
If additional sensors require a bump test but are not mandatory, the following screens display.
GasAlertMicro 5/PID/IR Activating the Detector
Self-Test Pass
If the detector passes the self-test, it enters normal operation and displays
the ambient gas readings.
Press C Yes to accept and proceed to normal operation.
Or
If A No is pressed, or no buttons are pressed, the sensor(s) that is past due
displays as FAIL when the detector enters normal operation. In the following
screen example, only the SO2, CO, and O2 sensors are operational.
The detector begins recording the peak (MAX) gas exposure and calculating the
short-term exposure level (STEL) and time-weighted average (TWA) exposures.
Self Test Fail
If a sensor fails, FAIL flashes above that sensor on the normal operation
screen. For possible problems and solutions, refer to Troubleshooting.
The self-test is now complete. 19
GasAlertMicro 5/PID/IR User Manual Battery Test
The batteries are tested when the detector is activated and continuously
thereafter. The battery power icon displays continually during normal
operation. If battery power is low, flashes.
Datalogger Operation (Optional) a Caution
Do not remove the battery pack while the detector is activated. Doing so will
prevent the datalogger from logging correctly. Datalogger operation is
automatic and requires no settings. During normal operation the card is tested
every 20 seconds.
Note The MMC/SD card icon (S) displays continuously on the LCD when the card
is inserted. The card is not required for operation; however, if the card is
not inserted the detector will not record data.
20
GasAlertMicro 5/PID/IR Deactivating the Detector
Deactivating the Detector
To deactivate the detector, press and hold A while it beeps and flashes to the
corresponding countdown.
At the end of the countdown, the detector emits an extended beep and flash
then displays 0 before deactivating.
Note If A is not held down for the complete countdown, the detector remains
activated.
User Options Menu
If the detector is passcode protected, a passcode must be entered to access
the user options menu. For more information, refer to Passcode Protect.
The available user options are as follows:
1. Exit
2. Options: Backlight, confidence beep, due-lock, latch, passcode, and safe
mode.
3. Sensors: Sensor enable/disable, span gas, STEL period, TWA method,
resolution, % vol CH4, correction factor, automatic O2 calibration, and % vol
CO2 (CO2 sensor only)
4. Logger
5. Clock
6. Language: English, French, German, Spanish, and Portuguese.
7. Tech mode: Sensors, initialize, forced calibration, daily bump test,
stealth, and zero level (CO2 sensor only).
Note
Tech mode is not visible in the user options menu. To access this option,
refer to Tech Mode.
21
GasAlertMicro 5/PID/IR User Manual
1. To enter the user options menu, press and hold G and H simultaneously as
the detector beeps and flashes to the corresponding countdown.
Note If no buttons are pressed within 20 seconds, the detector returns to
normal operation.
Exit User Options Menu
To exit the user options menu and return to normal operation, scroll to Exit
and press C. The following screen displays.
G and H must be held down for the entire countdown to access the user options
menu. When the countdown is complete, the revision/serial number screen
displays followed by the options menu.
2. To scroll through the options, press H or G. When the cursor displays
beside the desired option, press C.
3. To return to the previous menu, scroll to Back and press C or press A.
The user options menu can also be exited by repeatedly pressing A until the detector returns to normal operation.
Options Menu
Each feature within the Options menu is enabled/disabled by pressing C to toggle between the checkboxes.
Enabled
Disabled
22
Backlight
The Backlght (backlight) option enables the LCD backlight to activate
automatically in low-light conditions. If disabled, the backlight activates
only when the detector is in alarm mode.
Note
In stealth mode, the backlight does not activate.
The detector is shipped with the Backlght option enabled.
Confidence Beep
The Confibeep (confidence beep) option provides continuous confirmation that
the detector is operating properly. When confidence beep is enabled, the
audible alarm beeps once every 10 seconds.
Note
Confidence beep is automatically disabled during a low battery alarm.
The detector is shipped with the Confibeep option disabled.
GasAlertMicro 5/PID/IR User Options Menu
Due-Lock
If the Due-lock (calibration user lockout) option is enabled and a sensor is
overdue for calibration upon startup, the passcode must be entered to access
normal operation.
Note If the correct passcode is not entered, the detector deactivates. The
detector is shipped with the Due-lock option disabled.
23
GasAlertMicro 5/PID/IR User Manual
Latched Alarms
If enabled, the Latch (latched alarms) option causes the low and high gas
alarms (audible, visual, and vibrator) to persist until they are
acknowledgedby pressing C and the gas concentration is below the low alarm
setpoint. The peak concentrations display continually until the alarm
condition no longer exists. Local regulations may require Latch option be
enabled.
The detector is shipped with the Latch option disabled.
Passcode Protect
The Passcode option prevents unauthorized access to the user options menu, the
calibration function, and to adjusting the alarm setpoints.
Note The passcode is provided separately in the supplementary booklet.
24
If passcode protect is enabled and the Enter passcode: 1000 screen displays,
press G or H to scroll to the correct passcode and then press C to confirm.
The detector is shipped with the Passcode option disabled.
If an incorrect passcode is entered or C is not pressed within 5 seconds to
confirm the correct passcode, Passcode incorrect displays. The alarm beeps
three times and the detector either
· resumes normal operation (for user options), or
· deactivates (for calibration if Due Lock is enabled).
Safe Display
When enabled, the Safe option confirms that conditions are normal and there
are no gas hazards present. When all gas levels are normal or below the alarm
setpoints, Safe displays continually on the LCD.
Safe does not display if any active sensor has failed, if any active sensor
has an overdue calibration, and/or if any active sensor has failed a bump
test.
The detector is shipped with the Safe option disabled.
GasAlertMicro 5/PID/IR User Options Menu
Sensor Configuration
The Sensor option provides access to additional options and functions for each
sensor. Depending upon the sensor that is selected, some or all of the
following options are available for configuration:
· Enabling/disabling a sensor · Setting the span gas value · Adjusting the
STEL period (not applicable to LEL and O2 sensors) · Selecting the TWA method
(not applicable to LEL and O2 sensors) · Resolution setting (not applicable to
CO, LEL, O2, and CO2
sensors) · % vol CO2 (CO2 sensor only) · % vol CH4 (LEL sensor only) ·
Selecting the correction factor (LEL and PID sensors only) · Automatic
calibration (O2 sensor only)
25
GasAlertMicro 5/PID/IR User Manual
1. From the option menu screen, scroll to Sensors and press C to access the
following screen.
2. Press G or H to scroll to the desired sensor. Press C to confirm and to
access the menu options specific to the selected sensor. For all sensor
options, if a value is changed but not confirmed within 5 seconds, the
detector emits an audible alarm and displays the following error message.
Sensor Enable/Disable a Warning
Disabling a sensor should only be performed with extreme caution. The disabled
sensor cannot measure and alarm against the applicable gas. If a sensor fails,
disabling the sensor deactivates the fail alarm. The sensor should be replaced
and enabled as soon as possible. The detector will function normally with the
remaining enabled sensors.
Note
Detectors that are configured for a 1, 2, 3, or 4 gases may contain a dummy
sensor in one of the four sensor locations.
After selecting the desired sensor, the following screen displays.
The detector retains the previous setting and returns to the user options menu.
Press C to toggle between enable/disable (sensor can be enabled at any time).
Enabled Disabled
26
If disabled, the readings and the gas type for the sensor do not display when
in normal operation.
If a sensor is enabled but it is not installed in the detector, FAIL flashes
above the gas type of the missing sensor.
Disabled
Enabled/not installed
If all the sensors are disabled, the following screens display.
Enable one or more sensors to exit and access normal operation.
GasAlertMicro 5/PID/IR User Options Menu
Span Gas Value
The Span gas option increases or decreases the gas concentration for calibration (must match the gas concentration listed on the gas cylinder).
After selecting the sensor, press H to scroll to Span gas and press C within 20 seconds to confirm.
Depending upon the sensor selected, a
screen similar to A or B displays. Press
A
G or H to scroll to the desired value and
press C within 5 seconds to confirm.
If C is not pressed within 5 seconds to confirm the new value, the detector retains the previous value and returns to the user options menu.
Note
BW recommends that span concentration
B
values be set between specific ranges.
Refer to the Resetting Gas Alarm
Setpoints.
27
GasAlertMicro 5/PID/IR User Manual
STEL Period
The short-term exposure limit (STEL period) option provides protection for
workers from over exposure to high concentrations of gas, and is based on 5-15
minute intervals.
The STEL period option is available for every toxic sensor. After selecting
the desired sensor, press H to scroll to STEL period and press C within 20
seconds to confirm. The STEL period can be set from 5 to 15 minutes. Press G
or H to scroll to the required value, and then press C within 5 seconds to
confirm. If C is not pressed within 5 seconds to confirm the new value, the
detector retains the previous value and returns to the user options menu. The
detector is shipped with the STEL period set to 15 minutes.
28
TWA Method
The TWA method (time-weighted average) option is a safety measure used to
calculate accumulated averages of gases to notify the user when the maximum
average is accumulated. The options are the OSHA or ACGIH calculating method.
OSHA Method: 8 hour moving average–oldest value (first hour) is replaced by
the newest value (ninth hour). ACGIH Method: Infinite accumulated average to 8
hours–total accumulation, whether it is 2 hours or 8 hours.
After selecting the desired sensor, press H to scroll to TWA method. Press C
within 20 seconds to confirm.
A check displays in the checkbox of the currently selected method. To select
the other method, press H to move the check to the other method. Press C to
confirm the selection.
The detector is shipped with the OSHA method enabled.
Note
If the TWA method has been changed, the TWA, STEL, and MAX values must be
reset to ensure the TWA is calculated correctly. Refer to Clearing Gas
Exposures.
Resolution
The Resolution option displays the gas measurement using Regular or Extra
resolution.
Regular: Displays gas measurement in 1 ppm increments.
Extra: Displays the gas measurement in 0.1 ppm increments.
Note
Regular resolution for O3 and ClO2 sensors is 0.1 ppm, while extra resolution
is 0.01 ppm. The Resolution option is not available for CO, O2, LEL, PID, and
CO2 sensors.
After selecting the desired sensor, press H to scroll to Resolution. Press C
within 20 seconds to confirm.
A check displays in the checkbox of the currently selected resolution. To
select the other resolution, press H to move the check to other resolution.
Press C to confirm the selection.
The detector is shipped with Regular resolution enabled for applicable
sensors.
GasAlertMicro 5/PID/IR User Options Menu
%Vol CO2 (CO2 Sensors Only)
If the %vol CO2 is enabled, the detector displays the carbon dioxide (CO2)
readings as %vol (0.0). From the Sensors option menu, select CO2.
Press H to scroll to %vol CO2. Press C to toggle between enable and disable.
Confirmation is not required. If no buttons are pushed, after 20 seconds the
detector returns to the sensor selection screen. The change is saved
automatically. The detector is shipped with %vol CO2 disabled.
Note Correction factors are not applicable to the CO2 sensor.
29
GasAlertMicro 5/PID/IR User Manual
%Vol CH4 (LEL Sensors Only)
If %vol CH4 is enabled, any currently enabled correction factor is ignored and
the detector operates assuming a methane (CH4) calibration.
From the Sensors menu, select LEL and press C to confirm.
Next, press H to scroll to %vol CH4 and press C within 20 seconds to confirm.
Press C to toggle between enable and disable.
Confirmation is not required. If no buttons are pushed within 20 seconds, the
detector returns to the sensor selection screen. The change is saved
automatically.
Note
If changing the measurement unit from % LEL to % Vol. or from % Vol. to % LEL,
a calibration must be completed and the alarm setpoints changed. For
calibration information and for alarm setpoint information refer to
Calibration and Setting Alarm Setpoints on page 48.
The detector is shipped with %vol CH4 disabled.
30
Correction Factor (CF)
Depending upon the selected sensor, refer to the following sections LEL Sensor
or PID Sensor for more information.
Note Corrections factors are not applicable to CO2 sensors.
GasAlertMicro 5/PID/IR User Options Menu
LEL Sensor
This option is used to enter compensation factors for hydrocarbons other than
methane. The factor can only be applied if the LEL sensor has been calibrated
with methane.
After selecting the LEL sensor, press H to scroll to Correction. Press C
within 20 seconds to confirm and access the LEL correction library.
Scroll to the required gas type and press C. A check displays in the
corresponding checkbox. The detector automatically applies the correction
factor.
To disable the Correction option, press H to scroll to None or to Methane. A
check displays in the corresponding checkbox. If required, select a different
gas type correction factor.
Custom: To enter a correction factor that is not listed in the library, press
H to scroll to Custom and press C within 5 seconds to confirm.
The Custom LEL correction screen displays. Press G or H to select the required
value, and press C within 5 seconds to confirm.
PID Sensor
This option is used to enter compensation factors for selected gas types. The
factor can only be applied if the PID sensor has been calibrated with
isobutylene.
After selecting the PID sensor, press H to scroll to Correction. Press C
within 20 seconds to confirm and access the PID correction library.
Scroll to the required gas type and press C. A check displays in the
corresponding checkbox. The detector automatically applies the correction
factor.
To disable the Correction option, press H to scroll to None or to Isobutyl. A
check displays. If required, select a different gas type correction factor.
Custom: To enter a correction factor for a custom PID sensor, scroll to Custom
and press C. Press G or H to scroll to the required value, and press C within
5 seconds to confirm. Refer to the PID Correction Factor PID Correction Factor
(CF) Library Library for gas types and corresponding correction factor values.
31
GasAlertMicro 5/PID/IR User Manual
Note
The PID sensor is not sensitive enough to detect VOCs Benzene, Butadiene, and
Vynil Chloride before they exceed their toxic threshold limit value. The
MicroDock II is still able to set the Micro 5PID/IR to one of these VOCs. In
this case the Micro5/PID/IR will warn that the selection is not supported and
ask if you wish to continue. If the warning is not acknowledged the Micro
5/PID/IR will fail the PID Sensor until you reboot and accept, or choose a
different VOC.
Automatic Oxygen (O2) Calibration
When the Autocal option is enabled, it forces the detector to automatically
calibrate the oxygen sensor during startup.
If the Autocal option is enabled, ensure the detector is activated in a safe
area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
From the Sensor menu, press H to scroll to O2 and press C within 20 seconds to
confirm.
Press H to scroll to Autocal. Press C to toggle between enable/ disable.
The detector is shipped with the Autocal option enabled.
32
Logger Option
The Logger option is used to define how often the detector records a datalog
sample (once every 1 to 127 seconds).
From the user options menu, press H to scroll to Logger. Press C within 20
seconds to confirm.
GasAlertMicro 5/PID/IR User Options Menu
If C is not pressed within 5 seconds, the following screen displays.
Press G or H to change the current logger rate. When the desired value displays, press C within 5 seconds to confirm the new value.
The detector is shipped with the datalog sample time set to 5 seconds.
Clock Option
The Clock option is used to change the date and time. From the user options
menu, press H to scroll to Clock. Press C within 20 seconds to confirm.
33
GasAlertMicro 5/PID/IR User Manual
The screen displays showing the month highlighted indicating it is selected to
set.
Press G or H to scroll to the desired month and press C within 20 seconds to
confirm. To bypass and retain the current setting, press C. Continue
setting/bypassing the remaining options.
Language Selection
The detector is shipped with English selected as the default language. The
available languages are as follows:
· Français (French) · Deutsch (German) · Español (Spanish) · Prtuguês
(Portuguese)
Press H to scroll to Language and press C within 20 seconds to confirm.
The date/time options are set in the following order: · month · day · year ·
hour · minutes
When the settings have been set or bypassed, the detector beeps twice and
returns to the user options menu.
Note
The detector is shipped with the date and time set to Mountain Standard Time
(MST).
Press G or H to scroll to the desired language and press C. A check displays in the checkbox of the selected language.
34
Wait for 20 seconds until the detector returns to the user options menu, or
press G to scroll to Back (English), Retour (French), Zurück (German), Regreso
(Spanish), or Retornar (Portuguese). All screens now display in the selected
language.
Tech Mode
a Warning Tech mode should only be accessed by authorized personnel. Tech mode
can only be accessed from the Language option. Press H to scroll to Language.
Do not press C until instructed.
GasAlertMicro 5/PID/IR User Options Menu
In the following order, press and continue to hold each button until Tech mode
displays below the Language option.
1. Press and hold H with the right index finger. 2. Press and hold G with the
right middle finger. 3. Press and hold C with the left thumb.
4. Press C to enter Tech mode. The options are as follows:
· Sensors · Initialize · Force Calibration (Force calibration) · Bump Daily
(Bump test daily) · Stealth Mode · Sleep Mode
35
GasAlertMicro 5/PID/IR User Manual
Sensors a Caution
To reconfigure the sensor type, physically change the sensor prior to entering
Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note If a sensor is replaced, the detector will classify the sensor as overdue
for calibration. Calibrate the new sensor immediately.
1. Press H to scroll to Sensors. Press C within 20 seconds to confirm and
access the toxic sensor menu.
2. Press G or H to scroll to Toxic 1 or Toxic 2 and press C within 20 seconds
to confirm.
A corresponding list of toxic sensors displays. A checkbox displays beside the
current toxic sensor.
Note
Toxic 1: List includes the PID and CO2 sensors.
Toxic 2: List includes the H2S/CO COSH sensor.
3. Press G or H to scroll to the new sensor and press C to confirm. A
checkbox displays beside the new sensor. To reconfigure, exit the user options
menu.
The following screen displays. The detector deactivates and immediately
reactivates. It performs the reconfiguration during the startup.
The new sensor must be calibrated. Until the new sensor is calibrated, the due
date displays as OL (over limit) when in normal operation.
4. Calibrate the new sensor immediately.
36
Initialize
The Initialize option restores the original factory default settings of the
detector. 1. Press H to scroll to Initialize and
press C within 20 seconds to confirm.
2. From the Initialize? screen perform one of the following options within 5
seconds. · Press A No to exit, or · Press C Yes to initialize.
If No is selected, the following screen displays and the detector exits the
initialize option.
GasAlertMicro 5/PID/IR User Options Menu
If Yes is selected, the following screen displays while performing the
initializing process.
When initializing is complete, the following screen displays. The detector
deactivates and then immediately reactivates. The detector then reconfigures
to the default settings while it performs the self-test. Verify all settings
and alarm setpoints, and then calibrate the sensors.
Force Calibration
If enabled, the Force cal option automatically forces the detector to enter
calibration if a sensor is overdue upon startup. Press H to scroll to Force
cal. Press C to toggle between enable and disable.
The detector is shipped with the Force cal option disabled.
37
GasAlertMicro 5/PID/IR User Manual
Bump Daily
If enabled, the Bmp daily option forces the detector to perform a daily bump
test to ensure that it is responding to the test gas. Press H to scroll to Bmp
daily. Press C to toggle between enable/disable.
When enabled, the following screen displays during the startup self-test. The
bump test must be performed for the LEL and O2 sensors, otherwise the detector
will deactivate. Refer to Bump Daily Enabled for procedures.
The detector is shipped with the Bmp daily option disabled.
Note If Bmp daily is enabled, the O2 and LEL sensors require daily bump tests
whenever the detector is activated (or reactivated) following 00:00 hours
(midnight).
Stealth Mode
The Stealth option disables the backlight, visual alarms, and audible alarms
when concealment is required.
Only the vibrator and the LCD activate during an alarm condition.
Press H to scroll to Stealth. Press C to toggle between enable and disable.
The detector is shipped with the Stealth option disabled.
38
Sleep Mode
Note
BW recommends the Sleep option be enabled when using rechargeable battery
packs.
If Sleep is enabled, sensor circuits remain active after the detector is
deactivated and inserted into the charging cradle.
Press H to scroll to Sleep. Press C to toggle between enable/disable.
Note
Enabling the Sleep option may require the battery be charged more frequently.
The detector is shipped with the Sleep option disabled.
GasAlertMicro 5/PID/IR User Options Menu
39
GasAlertMicro 5/PID/IR User Manual
Alarms
The following table describes the detector alarms and corresponding screens.
· During an alarm condition, the detector activates the backlight and displays
the current ambient gas reading.
Alarms Low Alarm · Fast beep · Slow flash · L and gas type flash · Vibrator alarm activates High Alarm · Constant beep · Fast flash · L and gas type flash · Vibrator alarm activates
Screen`
· If more than one type or level of alarm exists simultaneously, a multi alarm
results.
· To change the factory-set alarm setpoints, refer to Calibration and Setting
Alarm Setpoints.
Table 7. Alarms
Alarms TWA Alarm · Fast beep · Slow flash · L and gas type flash · Vibrator alarm activates STEL Alarm · Constant beep · Fast flash · L and gas type flash · Vibrator alarm activates
Screen
40
Table 7. Alarms
Alarms Multi Alarm · Alternating low and high alarm
beep and flash
· L and gas types flash
· Vibrator alarm activates Sensor Alarm · One beep every 15 seconds · FAIL
flashes above the failed
sensor
Low Battery Alarm · One beep and two flashes
every 25 seconds
· flashes
Screen
GasAlertMicro 5/PID/IR Alarms
Alarms Over Limit (OL) Exposure Alarm · Fast beep and flash · L and gas type
flash
· Vibrator alarm activates Automatic Deactivation Alarm · Eight beeps and
flashes
· displays
· Vibrator alarm temporarily activates
Normal Deactivation · Three beeps and flashes
Screen
41
GasAlertMicro 5/PID/IR User Manual
Table 7. Alarms
Alarms Confidence Beep · One beep, one flash, and one
vibrate every 10 seconds
Screen
Alarms
Pump Alarm
· Two fast beeps and alternating flashes
· Vibrator alarm activates
· L and J flash
Note If enabled, during an alarm condition the Latch (latching alarms) option
causes the low and high gas alarms (audible, visual, and vibrator) to persist
until the alarm is acknowledged by pressing C and the gas concentration is
below the low alarm setpoint. The peak concentrations display continually
until the alarm condition no longer exists. Local regulations may require the
Latch Alarms option be enabled.
42
Alarms MMC/SD Fail Alarm · One beep every 5 seconds · S flashes
Screen
Screen
If the Stealth option is enabled, the detector only vibrates during an alarm;
the audible and visual alarms are disabled.
a Caution: If the detector emits a low battery alarm, charge the lithium
battery pack immediately.
Gas Exposures Computed
a Warning
To avoid possible personal injury, do not deactivate the detector during a
work shift. TWA and STEL readings reset if the detector is deactivated for
more than 5 minutes.
Table 8. Computed Gas Exposures
Gas Exposure
Description
TWA (toxic only)
Time-weighted average (TWA) based on accumulated exposure to toxic gases averaged over a work day according to OSHA or ACGIH method.
STEL (toxic only)
Short-term exposure limit (STEL) to gas based on a 5-15 minute user selectable period.
*Peak (MAX)
Peak (MAX) concentration exposure during work shift.
*For oxygen, it is the highest or the lowest concentration
encountered.
Viewing Gas Exposures
Press and hold C until the peak (MAX) gas exposures displays.
The TWA gas exposures display next.
The STEL gas exposures display last.
GasAlertMicro 5/PID/IR Alarms
43
GasAlertMicro 5/PID/IR User Manual Clearing Gas Exposures
The exposures automatically clear after 5 minutes when the detector is
deactivated. To clear the MAX, TWA, and STEL exposure readings immediately,
press and hold C and G simultaneously. The detector displays the following
screen during the countdown.
Note Hold C and G for the entire countdown, otherwise the MAX, TWA, and STEL
exposure readings will not clear.
Gas Alarm Setpoints
The gas alarm setpoints trigger the gas alarms and are described in Table 9..
Table 9. Gas Alarm Setpoints
Alarm Low alarm
High alarm
TWA alarm STEL alarm Downscale alarm
Multi alarm
Condition
Toxics and combustibles: Ambient gas level above low alarm setpoint. Oxygen:
Ambient gas level may be set above or below 20.9%.
Toxics and combustibles: Ambient gas level above high alarm setpoint. Oxygen:
Ambient gas level may be set above or below 20.9%.
Toxic only: Accumulated value above the TWA alarm setpoint.
Toxic only: Accumulated value above the STEL alarm setpoint.
Toxic: If sensor reading is negative (half of the TWA setpoint). LEL: If
sensor reading is negative (half of the low alarm setpoint).
Two or more gas alarm conditions.
Viewing the Alarm Setpoints
To view the current alarm setpoints for all of the sensors, press G during
normal operation.
44
GasAlertMicro 5/PID/IR Alarms
The time/date, TWA, STEL, low, and high alarm setpoint screens display in the
following order left to right:.
If a correction factor has been applied to a sensor, one of the following
screens display indicating the sensor and gas type.
Resetting Gas Alarm Setpoints
Note Standard factory alarm setpoints may vary by region.
The following table lists the factory alarm setpoints according to the
Occupational Safety and Health Association (OSHA) settings.
Table 10. OSHA Sample Factory Alarm Setpoints
Gas
O2 LEL CO H2S PH3 SO2 Cl2 NH3 NO2 HCN ClO2 O3 VOC
CO2
TWA
N/A N/A 35 ppm 10 ppm 0.3 ppm 2 ppm 0.5 ppm 25 ppm 2.0 ppm 4.7 ppm 0.1 ppm 0.1
ppm 50 ppm 5,000 ppm
STEL
N/A N/A 50 ppm 15 ppm 1.0 ppm 5 ppm 1.0 ppm 35 ppm 5.0 ppm 10.0 ppm 0.3 ppm
0.1 ppm 100 ppm 30,000 ppm
Low
19.5% vol. 10% LEL 35 ppm 10 ppm 0.3 ppm 2 ppm 0.5 ppm 25 ppm 2.0 ppm 4.7 ppm
0.1 ppm 0.1 ppm 50 ppm 5,000 ppm
High
23.5% vol. 20% LEL 200 ppm 15 ppm 1.0 ppm 5 ppm 1.0 ppm 50 ppm 5.0 ppm 10.0
ppm 0.3 ppm 0.1 ppm 100 ppm 30,000 ppm
45
GasAlertMicro 5/PID/IR User Manual
To change the factory-set alarm setpoints, refer to Calibration and Setting
Alarm Setpoints.
Note To disable an alarm, set the alarm setpoint to 0 (zero).
Stopping a Gas Alarm
The low and high alarms stop when the ambient gas level returns to below the
low alarm setpoint.
Note If alarms are set to latch, press C to acknowledge the alarm. The alarms
will persist until the alarm is acknowledged and the gas concentration is
below the low alarm setpoint.
a Caution Follow all safety procedures as defined by your employer. Confirm
with your supervisor before clearing TWA and STEL alarms.
The TWA and STEL alarms can be stopped either by · clearing the MAX, TWA, and
STEL exposures (refer to Clearing Gas Exposures), or · deactivating the
detector for 5 minutes (minimum) and then reactivating it again.
46
Sensor Alarm
The detector tests for missing or defective sensors during the activation
self-test. If a sensor fails the self-test, FAIL flashes on the LCD above the
failed sensor. Refer to Troubleshooting.
Pump Alarm
The external pump draws air over the sensors continually. If the pump stops
operating or becomes blocked, the detector activates the pump alarm. The
following screens display.
The pump alarm continues until the blockage is cleared or it is acknowledged
by pressing C. If C is pressed, the detector automatically launches a pump
test to reset the pump module.
GasAlertMicro 5/PID/IR Bump Test
Refer to Pump Test for more information. If the pump test is successful, the
detector returns to normal operation, otherwise the pump alarm continues. If
the pump alarm persists, refer to the Pump Operation section in
Troubleshooting.
Low Battery Alarm
The detector tests the batteries upon activation and continuously thereafter.
Battery power is continually displayed during normal operation. If the battery
voltage is low, the detector activates the low battery alarm.
The low battery alarm continues until the batteries are replaced/charged, or
until the battery power is almost depleted. If the battery voltage becomes too
low, the detector deactivates.
Note
Typically, the low battery alarm continues for 30 minutes before the detector
automatically deactivates.
Automatic Deactivation Alarm
If the battery voltage is in immediate danger of falling below the minimum
operating voltage, the audible alarm beeps eight times and the visual alarm
flashes eight times. After 3 seconds, the LCD dims and the detector
deactivates.
To replace or charge the batteries, refer to the following sections:
· Replacing the Alkaline Batteries · Replacing the Lithium Battery Pack
Bump Test
A bump test is the process of applying a small amount of test gas to force the
detector into alarm. A bump test should be performed regularly to confirm the
sensors are responding correctly to gas, and that the audible, visual, and
vibrator alarms activate during an alarm condition. Calibrate if the readings
are not within specified limits.
Performing a Bump Test
a Caution BW recommends to bump test the sensors before each day’s use to
confirm their ability to respond to gas by exposing the sensors to a gas
concentration that exceeds the alarm setpoints.
To perform a manual bump test refer to Applying Gas to the Sensors on page 50
and the following steps 1-6. Follow this procedure when Bump Daily is enabled.
Note
The Calibration icon will flash if a sensor has failed a bump test.
Note
If performing a bump test on a pump unit, connect the calibration hose
directly to the pump module.
47
GasAlertMicro 5/PID/IR User Manual
1. Connect the calibration hose to the 0.5l/min regulator on the gas
cylinder.
Note
Only use the calibration cap for bump tests and calibrations.
2. Connect the calibration hose to the intake inlet on the calibration cap.
Arrows on the calibration cap indicate the direction of gas flow.
3. Attach the calibration cap and tighten the knob. Refer to Applying Gas to
the Sensors on page 50.
Note
Ensure the knob is securely fastened before applying gas. Do not overtighten
the calibration cap. Ensure strong air currents are kept away from the exhaust
side of the calibration cap, as this can affect the bump test accuracy.
4. Apply gas. Verify the visual, audible, and vibrator alarms activate.
5. Close the regulator and remove the calibration cap from the detector. The
detector will temporarily remain in alarm until the gas clears from the
sensors.
6. Disconnect the calibration hose from the cap and regulator.
Calibration and Setting Alarm Setpoints
Guidelines
When calibrating the detector, adhere to the following guidelines:
Recommended gas mixture: CO: 50 to 500 ppm balance N2 H2S: 10 to 100 ppm
balance N2 PH3: 1 to 5 ppm balance N2 SO2: 10 to 50 ppm balance N2 Cl2: 3 to
25 ppm balance N2 NH3: 20 to 100 ppm balance N2 NO2: 5 to 50 ppm balance N2
HCN: 5 to 20 ppm balance N2 ClO2: 0.1 to 1.0 ppm balance N2 O3: 0.1 to 1.0 ppm
balance N2 VOC:100 ppm isobutylene LEL: 50% LEL or 2.5% by vol. methane
balance air CO2: 5000 ppm balance air O2: 18.0%
· To ensure accurate calibration, use a premium-grade calibration gas. Gases
approved by the National Institute of Standards and Technology (NIST) improves
the validity of the calibration.
· Do not use a gas cylinder beyond its expiration date.
48
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
· Calibrate a new sensor before use. Install the sensor, activate the
detector, and allow the sensor to stabilize before starting calibration. Used
sensor: 60 seconds / New sensor: 5 minutes
· Calibrate the detector at least once every 180 days (every 90 days for HCN
sensors) depending upon use and sensor exposure to poisons and contaminants.
For a list of common products that can damage sensors, refer to Sensor Poisons
and Contaminants.
· Calibrate the detector if the ambient gas varies during startup.
· Calibrate the sensor before changing the alarm setpoints.
· Calibrate only in a safe area that is free of hazardous gas in an atmosphere
of 20.9% oxygen.
· To disable an alarm, set the alarm setpoint to 0 (zero).
· If the Auto cal option is enabled, the oxygen (O2) sensor calibrates
automatically every time the detector is activated. Activate the detector in a
safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
· After activating the detector, allow it to stabilize for 1 minute before
performing a calibration or bump test.
· If a certified calibration is required, contact BW Technologies by
Honeywell.
Note
A generator must be used for O3 and ClO2 sensors.
Diagnostics Testing
The detector tests the ambient air (auto zero) and the test gas that is
applied (auto span) to ensure it meets expected values. Auto zero sets the
zero-gas level of the sensor. If ambient gas is present, the zero level will
be incorrect. If excessive target gas is present, the detector displays an
error message and lists the affected sensor.
In auto span, if the target gas is not detected or does not meet expected
values, a message displays that the detector is exiting calibration mode. The
detector retains the previous set values.
49
GasAlertMicro 5/PID/IR User Manual Applying Gas to the Sensors
The calibration cap, single gas calibration cap, and hose are shipped with the
detector. Refer to Figure 3. and Table 11. for installation.
Note The calibration cap and single gas calibration cap must only be used
during the calibration span process.
Figure 3. Applying Gas to the Sensors
Table 11. Applying Gas to the Sensors
Item
1 2 3 4 5
Description
Detector with calibration cap Calibration hose Regulator and gas cylinder
Toxic 2 sensor position Single gas calibration cap
Single Gas Calibration Cap a Caution
If an O3 or ClO2 sensor is located in the Toxic 2 position (refer to Figure 3.
and Table 11.), a single gas calibration cap must be used to ensure accurate
calibration.
To calibrate O3 and ClO2 sensors using the single gas calibration cap, refer
to Figure 4., Table 12., and complete the following procedures.
1. Insert the cap into the Toxic 2 sensor position on the detector (refer to
Figure 3.). Press firmly until the release tabs click.
2. Connect the calibration hose to the gas cylinder and to the intake inlet
on the cap.
3. Proceed to Calibration Procedure.
50
Table 12. Single Gas Calibration Cap
Item
1 2 3 4
Description Intake inlet Calibration hose Gas flow direction arrow Output outlet
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
Removing the Single Gas Calibration Cap Using the thumb, push forward against
both the inlet and the outlet simultaneously to remove the cap from the
detector.
Figure 4. Single Gas Calibration Cap
Note The arrow on the cap indicates the direction of gas flow from intake to
outtake.
Figure 5. Removing the Single Gas Calibration Cap
Calibration Procedure
To calibrate the detector and set the alarm setpoints, perform the following
procedure.
Note To bypass a step during the calibration process (after auto zero), press
A.
51
GasAlertMicro 5/PID/IR User Manual
Start Calibration
Note
Verify that the calibration gas being used matches the span concentration
value(s) that are set for the detector. Refer to Span Gas Value. Correction
factors are not applied during calibration. Correction factors that were set
prior to calibration are restored when the detector returns to normal
operation.
Note
The calibration icon will flash if a sensor has an overdue calibration
a Caution Calibrate only in a safe area that is free of hazardous gas in an
atmosphere of 20.9% oxygen. 1. To enter calibration, press and hold C and H
simultane-
ously as the detector beeps, flashes, and vibrates to the corresponding
countdown.
Auto Zero and Oxygen (O2) Sensor Calibration
flashes while the detector automatically zeroes the
toxic and combustible sensors, and calibrates the O2 sensor.
52
Depending upon the detector being calibrated, the auto zero screens display
differently.
GasAlertMicro 5/PID GasAlertMicro 5 IR
Note Do not apply calibration gas during this process, otherwise the auto zero
step will fail.
Zero CO2 (GasAlertMicro 5 IR only)
If calibrating using the GasAlertMicro 5 or the GasAlertMicro PID, proceed to
Auto Span step #5. If the detector is passcode protected, proceed to Passcode
Protect step #4.
3. For GasAlertMicro 5 IR models, there is a second auto zero that is
performed for the CO2 sensor. When auto zero for the other sensors is
complete, the following screen displays.
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
Press C Yes to zero the CO2 sensor.
Or
Press A No to bypass the CO2 zero and proceed to Auto Span step #5. If the
detector is passcode protected, refer to Passcode Protect Activated step #4.
If C Yes is pressed, the following screen displays.
flashes whiles the detector zeros the CO2 sensor (approximately 30 seconds).
Passcode Protect Activated
4. When auto zero is complete and if the passcode protect option is enabled,
the following screen displays.
a Warning Use nitrogen (N2) to zero the CO2 sensor.
Apply N2 to zero the CO2 sensor. The following screen displays.
The passcode must be entered to proceed. Press G or H to scroll to the correct
passcode and then press C within 5 seconds to confirm. If entered correctly,
the detector beeps twice and proceeds to Auto Span step #5. Incorrect
Passcode: If the passcode is incorrect or is not confirmed within 5 seconds by
pressing C, the following screens display.
53
GasAlertMicro 5/PID/IR User Manual
The detector saves the calibration and returns to normal operation.
Auto Span
5. When auto zero is complete, the following screens display. GasAlertMicro 5
and PID
GasAlertMicro 5 IR
.Note Span sensors in the following order:
· Exotics (NH3, ClO2, O3, Cl2, and CO2)
· Single gas
· Quad gas (H2S, CO, LEL, and O2)
· PID
Note A generator must be used for O3 and ClO2 sensors. To ensure accurate
calibration, a single gas calibration cap must be used to calibrate O3 and
ClO2 sensors.
· ClO2: Use a Tedlar bag as a buffer between the generator and the detector
(using the single gas calibration cap) to regulate the flow rate to ensure
accurate readings.
Allow the Tedlar bag to fill for several minutes before initiating
calibration.
· O3: Calibrate using the generator and the single gas calibration cap. Do not
use a Tedlar bag.
Set the generator to 0.5 ppm at a flow rate of 0.5 l/min.
Depending upon the gas cylinder being used, one or all four sensors can be
calibrated at one time. Apply Span Gas Now
54
Attach the calibration cap (or single gas calibration cap for O3 and ClO2) and
apply gas to the sensor(s). To attach caps, refer to Figure 3..
Refer to the following flow rates:
· 1000 ml/min. for NH3 and Cl2
· 500 ml/min. for CO2
· 250-500 ml/min. (all other sensors)
K flashes as the detector initially detects the calibration gas.
GasAlertMicro 5 and PID
GasAlertMicro 5 IR
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
Exotic toxic gases
5 minutes
LEL (combustibles)
30 seconds
PID gases
2 minutes
Insufficient Level: If a sensor does not attain a sufficient level of expected gas, it is cleared from the LCD and is not spanned.
After 30 seconds the detector beeps and K stops flashing. flashes while
spanning the sensors until the detector
has attained a sufficient level of the expected gas.
Refer to Table 13. for times required to span.
Table 13. Time Required to Span
Gas Type Most toxic gases and CO2
Time Required to Span 2 minutes
55
GasAlertMicro 5/PID/IR User Manual
While the detector is spanning the sensor(s), a countdown of time remaining displays in the lower left of the screen.
GasAlertMicro 5 and PID
GasAlertMicro 5 IR
Successful Span
6. If the sensor(s) has spanned successfully, the audible alarm beeps three
times and the following screens display.
When the span is complete, the following screen displays.
Successful Span: Proceed to Successful Span step #6. Unsuccessful Span: Refer
to Unsuccessful Span for possible solutions.
If there are more sensors to span, remove the current cylinder and connect the
next gas cylinder. Press G and apply gas to span the other sensor(s).
Or Press H to end the span and proceed to Setting the Calibration Due Date
step #7.
When all sensors have spanned, the following screen displays.
If a sensor fails to span, refer to Unsuccessful Span at the end of the
calibration section for possible solutions.
56
Setting the Calibration Due Date
7. When the span is complete, the calibration due date can be set for each
sensor that has spanned successfully. The following screen displays.
Press C to set the calibration due dates. Or Press A to bypass and proceed to
Alarm Setpoints step #8.
Note If a new due date is not entered, the detector automatically resets to
the previously entered number of days.
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
The calibration due dates are set in the following order: · Toxic 1 · Toxic 2
· LEL · O2 If the due date of an unsuccessfully spanned sensor is changed, the
following screens display.
The detector then automatically proceeds to the next sensor. To change the
calibration due date (1-365 days), press H or G until the new value displays.
Press C within 5 seconds to confirm. Or Press A to bypass a sensor and proceed
to the next sensor.
57
GasAlertMicro 5/PID/IR User Manual
Note If a value is changed but C is not pressed within 5 seconds to confirm,
the following screen displays.
Alarm Setpoints
8. When all of the sensor due dates have been set or bypassed, the alarm
setpoints need to be set or bypassed. The following screen displays.
The previous value is automatically retained. The detector proceeds to the
next sensor due date.
Repeat step #7 to set the calibration due dates for the remaining sensors.
When all of the due dates are entered, press C to proceed to Alarm Setpoints
step #8.
Or
Press A to bypass setting the alarm setpoints and proceed to Finish
Calibration step #10.
Factory alarm setpoints may vary by region. Refer to Resetting Gas Alarm
Setpoints for an example.
Alarms can be set to any value within the detection range of the selected
sensor. Refer to Detection Range in Specifications.
Note
To disable an alarm setpoint, set it to 0 (zero).
58
If a new setpoint is entered but not confirmed within 5 seconds by pressing C, the following screen displays.
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
Setting the TWA Alarm Setpoint The current TWA alarm setpoint displays for the
selected sensor (if applicable).
The previous setpoint is retained and the detector proceeds to the next
setpoint. The setpoints are set in the following order:
· TWA (if applicable)
· STEL (if applicable)
· low
· high To bypass a setpoint, press C to save the current value and proceed to
the next setpoint.
Press H or G to change the value for the TWA alarm setpoint. When the required
value displays, press C to confirm.
Setting the STEL Alarm Setpoint
The current STEL alarm setpoint displays for the selected sensor (if
applicable).
Press H or G to change the value for the STEL alarm setpoint. When the
required value displays, press C to confirm.
59
GasAlertMicro 5/PID/IR User Manual
Setting the Low Alarm Setpoint The current low alarm setpoint displays for the
selected sensor.
Setting the Remaining Alarm Setpoints
9. Repeat step #8 to set the alarm setpoints for the remaining sensors. When
complete, the detector emits two quick beeps and proceeds to Finish
Calibration.
Finish Calibration
10. The detector displays the following screen to indicate that calibration
is complete.
Press H or G to change the value for the low alarm setpoint. When the required
value displays, press C to confirm.
Setting the High Alarm Setpoint
The current high alarm setpoint displays for the selected sensor.
Press H or G to change the value for the high alarm setpoint. When the
required value displays, press C to confirm. Proceed to step #9 Setting the
Remaining Alarm Setpoints.
60
Verification
1. After calibration is complete and the detector returns to normal
operation, verify the calibration by using a gas cylinder other than the one
used for calibration.
2. The gas concentration should not exceed the sensor’s detection range.
Confirm that the LCD displays the expected concentration values.
3. To ensure the reading is accurate, apply the verification gas for the same
amount of time as was applied to the sensor when it was calibrated.
Example: SO2 span time 2 minutes therefore, apply verification gas for 2
minutes.
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
Unsuccessful Span
If the sensor(s) did not span successfully, refer to the following sections
for possible solutions:
· Failed Span · No Gas Detected · Did Not Reach Target Span · Large Span
Failed Span If a sensor fails the span, the following error message displays.
If the sensor is not calibrated, it displays as FAIL in normal operation the next time the detector is activated.
61
GasAlertMicro 5/PID/IR User Manual
If all sensors fail the span, the following screen displays.
No Gas Detected
If the detector does not detect any gas within 30 seconds, the following
screens display.
1. Ensure the regulator valve is open and that the gas cylinder is not past
the expiration date. Replace the gas cylinder and/ or regulator if required.
2. Replace the sensor(s) and sensor filter if required.
3. Press A to exit and then calibrate again in a safe area that is free of
hazardous gas in an atmosphere of 20.9% oxygen.
4. If the span fails a second time, reactivate the detector to test the
sensors.
If all sensors fail the span, the calibration due dates cannot be reset. The
following screens display.
1. Ensure the regulator valve is open and the calibration hose is free of
debris.
2. Replace the gas cylinder and/or regulator if required.
3. Replace the sensor(s) and sensor filter if required. 4. Press G to reapply
gas or press H to end the span.
Did Not Reach Target Span If the target span was not achieved, as defined in
the Span Gas Value option, the following screens display.
62
Not reaching the target span can result from · a problem with the span gas, ·
the gas cylinder being past the expiry date, or · a problem with the sensor.
Accept Current Span: If the gas cylinder, regulator, and sensor are operating
correctly, press C to accept the current span. Reject Current Span: Press A to
reject. Verify the gas cylinder, detector settings, and sensor are operating
correctly. Calibrate the sensor again.
Large Span If the span adjustment is unusually large (more than 15%), the
following screens display.
Ensure the correct gas cylinder being used and that the span concentration
value(s) of the detector matches the span concentration value(s) of the gas
cylinder. Refer to Span Gas Value.
Adjustment Expected: If the calibration adjustment is expected, press C to
accept the span. Adjustment Not Expected: If the calibration adjustment is not
expected or the span gas value does not match the gas cylinder, press A to
reject the span and calibrate that sensor again.
GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints
63
GasAlertMicro 5/PID/IR User Manual
Pump
When using the pump module, attach it and the pump accessories prior to
activating the detector.
The pump module can be used independently to measure gas or it can be used
with the sample probe to test for gases in confined spaces. To measure gas
using the sample probe, refer to Confined Space Sampling.
a Warning If the pump module is installed on the detector, the following three
things must occur during start-up. If any one of the conditions below does not
occur, discontinue use of the detector and contact BW Technologies by
Honeywell immediately.
· The detector prompts for a pump test during start-up
· The pump module passes the pump test at start-up when the pump inlet or
sample chain inlet is blocked
· The J icon displays on the LCD
To ensure accurate gas detection, the sensors must be calibrated immediately
when the pump module is replaced by the diffusion cap and vice versa. When
using the sample probe at -10°C to +0°C (14°F to 32°F), keep the sample probe
in your hand.
a Caution A demand flow regulator must be used to manually calibrate the
GasAlertMicro 5/PID/IR detector when the pump module is installed. The
calibration cap is designed for use with the diffusion cap only. It cannot be
used with the pump module.
Identifying the Pump
There are two generations of pump. The best way to differentiate the
Generation 2 pump from the Generation 1 pump is the inline filter that is
visible on the Generation 2 pump on the left-hand side.
Each model of pump has different operating specifications. Refer to Maximum
Hose Length when Confined Space Sampling on page 71.
Generation 1: 116885-L3 (yellow) and 118933-L3 (black)
64
Generation 2: 130916-L3 (yellow) and 130917-L3 (black)
GasAlertMicro 5/PID/IR Pump
65
GasAlertMicro 5/PID/IR User Manual
Installing the Pump Module
Figure 6. Installing the Pump Module
Table 14. Installing the Pump Module
Item 1 2 3 4
Description Pump module Sensor filter Detector Machine screws (2)
To install the pump module, refer to Figure 6., Table 14., and the following procedures.
1. Deactivate the detector.
2. Remove the two machine screws and the sensor cover. Remove the sensor
filter from the sensor cover and insert it into the pump module.
3. Attach the pump module to the detector and replace the two machine screws.
4. If sampling in a confined space, attach the sample probe. Refer to
Confined Space Sampling.
If using an auxiliary filter as an additional filtration system, refer to
Attaching the Auxiliary Filter.
5. Activate the detector. The detector performs a pump test during the
startup self-tests. Refer to Pump Test.
66
Replacing the Pump Filter (Generation 2 Pump)
To replace the Generation 2 pump filter, refer to Figure 7. and the following
steps 1-8.
Note
If using the auxiliary filter on the Generation 1 pump or as an additional
filtration system on the new pump module, refer to Attaching the Auxiliary
Filter.
GasAlertMicro 5/PID/IR Installing the Pump Module
1. Deactivate the detector.
2. Remove the filter window screw.
3. Remove the filter window.
4. Remove the old filter from the filter window.
Important! Ensure the filter cavity and filter window are clean and free of
debris.
5. Insert a new filter.
6. Replace the filter window and screw.
7. Activate the detector. The detector performs a pump test during the
startup self-tests to verify the pump module is operating correctly.
8. To order additional filters, refer to Replacement Parts and Accessories. a
Caution
Filters may need to be replaced more frequently in high particulate areas. BW
Technologies by Honeywell recommends that the auxiliary filter be used as a
additional filtration in high particulate areas.
Figure 7. Replacing the Pump Filter (Generation 2 Pump) 67
GasAlertMicro 5/PID/IR User Manual
Replacing the Pump Nozzle (Generation 2 Pump)
To replace the pump nozzle for the Generation 2 pump, refer to Figure 8. and
step 1-4.
3. Insert the new nozzle. Ensure the nozzle post inserts correctly into the
nozzle gasket.
4. Activate the detector. The detector performs a pump test during the
startup self-tests to verify the pump module is operating correctly.
Attaching the Auxiliary Filter
The Pump Module Auxiliary Filter (“the filter”) is used to filter out
particulates that can decrease the life span of the pump.
Figure 8. Replacing the Pump Nozzle
Note
If the nozzle is damaged, replace it immediately to ensure accurate pump flow.
To order additional pump nozzles, refer to Replacement Parts and Accessories.
1. Deactivate the detector.
2. Gently insert a medium-sized flathead screwdriver into the nozzle slot.
Twist the screwdriver slightly to lift and remove the damaged nozzle.
68
Figure 9. Attaching the Auxiliary Filter
a Caution
The filter is designed to protect the pump. For Generation 1 pumps, the filter
must be connected when the pump is activated. Only remove the filter to
perform calibrations, bump tests, and when using the sample probe.
For new model pumps, the auxiliary filter can be used as an additional
filtration system in high particulate areas.
To insert the auxiliary filter, complete the following:
1. Loop the filter cord through the ring on the alligator clip and pull the
filter through the loop of the filter cord.
GasAlertMicro 5/PID/IR Installing the Pump Module
Note
Attaching the filter cord to the alligator clip ensures the filter remains
with the detector when not in use.
2. Pull the filter cord tight. Ensure the loop knot is secure. 3. Align the
pump quick connector with the filter quick connector
and turn clockwise. Ensure the connection is secure. 4. Change the filter as
required. Environments with heavy air-
borne particulates require the filter to be changed frequently. To order
additional filters, refer to Replacement Parts and Accessories.
Figure 10. Attaching the Filter Cord 69
GasAlertMicro 5/PID/IR User Manual Confined Space Sampling
The sample probe is used to safely test for gas in confined spaces before
entering. Attach the sample probe to the pump module prior to activating the
detector.
a Warning To measure hazardous gas in a confined space, the sample probe must
be used with the pump module.
70
Figure 11. Attaching the Sample Probe
a Caution To prevent the Teflon lining inside the Tygon tubing from causing a
blockage when connecting it to the sample probe, the end of the tubing must be
flared. Refer to steps 3 and 4.
Note
The detector and pump module can also be used with a sintered filter.
1. Deactivate the detector.
2. Attach the pump connector end of the teflon-lined Tygon tubing to the
nozzle on the pump module.
3. Gently insert needle nose pliers into the other end of the teflonlined
Tygon tubing. Using a circular motion, flare the end of the tubing.
4. Connect the flared end of the tubing to the sample probe. Ensure the
teflon lining does not separate from the Tygon tubing, as it will block the
tube and generate a pump alarm.
5. Activate the detector. Ensure all connections are secure before sampling.
6. Insert the sample probe into the confined space.
a Warning Depending upon the length of the tubing and the type of gas in the
confined space, allow a minimum of 3
seconds per foot of tubing to ensure the readings stabilize before entering
the area. Example: 10 ft. = 30 seconds
Maximum Hose Length when Confined Space Sampling
Maximum sample hose length is dependent on pump model, operating temperature,
and sampling accessory. Refer to the tables below.
a Caution Maximum sample hose length is dependent on an inner 1/16″ hose
diameter. Table 15. Generation 1 Pump Maximum Hose Length
Sintered filter Sample Probe
-10°C to +0°C (14°F to 32°F) 3 m (10 ft.)
3 m (10 ft.)
0°C to 50°C (32°F to 122°F) 9.1 m (30 ft.)
3 m (10 ft.)
a Warning
When using the sample probe at -10°C to +0°C (14°F to 32°F), keep the sample
probe in your hand.
GasAlertMicro 5/PID/IR Installing the Pump Module
Table 16. Generation 2 Pump Maximum Hose Length
Sintered filter Sample Probe
-20°C to 50°C (-4°F to 122°F) 20 m (66 ft.)
3 m (10 ft.)
71
GasAlertMicro 5/PID/IR User Manual
Datalogger
Detectors equipped with the datalogger option record information that can be
compiled to create a report. To set how often the detector records a sample
(1-127 seconds), refer to Logger Option in the user options menu. The
following information is recorded in a datalog:
· Date and time · Serial number of the detector · Type of gas the detector
monitors · Gas reading(s) that display · STEL and TWA readings · Sensor status
· Detector status · Passcode protect enabled/disabled · STEL period setting ·
Confidence beep enabled/disabled · Automatic backlight enabled/disabled ·
Stealth mode is enabled/disabled · Latching alarm enabled/disabled ·
Calibration past due option enabled/disabled · Language the detector is set to
display
MMC/SD Card Compatibility
For a list of compatible memory cards, please contact BW Technologies by
Honeywell.
Inserting the MMC/SD Card
To insert the MMC/SD card into the detector, refer to the following steps 1-5
and Figure 12..
1. Deactivate the detector. 2. Release the latch and remove the battery pack.
3. Insert the MMC/SD card (pins face down). 4. Replace the battery pack and
secure the latch. 5. Activate the detector. The MMC/SD card is automatically
for-
matted during the startup self-test.
72
Figure 12. Inserting/Removing the MMC/SD Card
GasAlertMicro 5/PID/IR MMC/SD Card Troubleshooting
MMC/SD Card Troubleshooting
The MMC/SD card is not required for operation in detectors equipped with
datalogging. However, the following two screens display if the card is not
inserted during startup.
A new MMC/SD card is automatically formatted when it is inserted in the detector. When the detector is activated, it begins the self-test and then displays the following screen.
73
GasAlertMicro 5/PID/IR User Manual
Restoring Datalog Files
If the MMC/SD card has been accidentally reformatted or erased by the computer
application, the following screens display when the card is inserted into the
detector.
If the detector successfully restores the logfile, the following screen displays and the startup tests continue.
a Caution
Only erased data files can be restored using the detector. Computer
applications sometimes write data over erased files, and that erased data
cannot be restored by the detector. Always create back up files on the
computer.
To restore the logfile, complete the following:
1. From the detector, press H to restore the logfile. The following screen
displays.
2. From the computer, verify that the logfile has been restored. When the
normal operation screen displays, deactivate the detector.
3. Remove the MMC/SD card and insert it into the card reader.
4. From the computer desktop, double-click My Computer to view the list of
drives.
5. Double-click the Removable Disk drive to access LOGFILE0.csv. Open the
logfile and verify that the data has been restored.
If LOGFILE0.csv does not display, ensure that the MMC/SD card is inserted in
the card reader correctly and that all connections are secure.
6. After verifying that the logfile has been restored, re-insert the MMC/SD
card into the detector.
74
Reformatting the MMC/SD Card
To reformat the MMC/SD card, complete the following:
1. Insert the MMC/SD card into the card reader.
2. From the computer desktop, double-click My Computer to view the list of
drives.
3. Double-click the Removable Disk drive to access LOGFILE0.csv.
4. Select LOGFILE0.csv and delete. 5. Insert the MMC/SD card into the
detector.
6. Activate the detector. The startup self-test begins and the following
screens display.
GasAlertMicro 5/PID/IR MMC/SD Card Troubleshooting
7. Press G to format the MMC/SD card. The following screen displays.
For any additional MMC/SD card errors, refer to Troubleshooting.
75
GasAlertMicro 5/PID/IR User Manual
Import Datalogs to Fleet Manager II
Note Refer to the following minimum requirements before importing datalogs to
Fleet Manager II.
Minimum PC Requirements
· 500 MHz Pentium (or equivalent) · 100 MB free hard disk space · Windows XP
or Vista · USB port
Importing from MicroDock II to Fleet Manager II
If the detector is used with the MicroDock II Station to import datalogs to
Fleet Manager II, refer to the MicroDock II User Manual for complete
instructions.
Import to Fleet Manager II Using a Card Reader
To import a datalog file from the detector to Fleet Manager II, complete the
following:
1. Deactivate the detector.
2. Release the latch and remove the battery pack.
3. Remove the MMC/SD card from the detector. Refer to Figure 12. and 1.
4. Connect the card reader to the USB port on the computer.
5. Insert the MMC/SD card into the card reader (ensure the pins face down).
6. Open Fleet Manager II. Click Start > All Programs > BW Technologies >
Fleet Manager II > BW Technologies Fleet Manager II.
7. From the top task bar, click Help > User Manual. 8. Refer to section
“Importing Data into Fleet Manager II”.
76
View Datalog Files in Spreadsheets
The datalog files can be downloaded from the MMC/SD card into most spreadsheet
applications using a card reader.
Compatible software applications are
· Microsoft® Excel 98 or higher, · Quattro Pro, · Lotus 1-2-3, · Microsoft®
Access, and · Microsoft® Word. To view a datalog file in a software
spreadsheet, complete the following:
1. Deactivate the detector and remove the MMC/SD card (refer to Figure 12.).
2. Insert the MMC/SD card into the card reader.
3. From the computer desktop, double-click My Computer to view the list of
drives.
4. Double-click Removable Disk drive. 5. Double-click LOGFILE0.csv. 6. Refer
to Table 17. for an example of the datalog spreadsheet.
The Unit Config column (far right) in Table 17. contains letter codes. Refer
to Table 18. and Table 19. for definitions of the codes.
GasAlertMicro 5/PID/IR View Datalog Files in Spreadsheets
77
GasAlertMicro 5/PID/IR User Manual
Example of a Datalog Spreadsheet
When datalog information is imported into most spreadsheet software, it
appears similar to the example below.
a Warning: Some compatible software packages have an internal file size limit
and may not load the entire file. Check the software limit.
Note Not all columns are included in this example. Additional Toxic TWA and
Toxic STEL display on a normal spreadsheet.
Table 17. Datalog Spreadsheet Example
Date
Day
dd-mm-yy Mon=1
23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4
Time hh:mm:ss
9:54:25 9:54:30 9:54:35 9:54:40 9:54:45 9:54:50 9:54:55 9:55:00 9:55:05
9:55:10 9:55:15 9:55:20 9:55:25 9:55:30
Toxic1 ppm
5 10 5 10 0 0 0 0 5 10 5 10 0 0
Toxic2 ppm
10 15 10 15 0 0 0 0 10 15 10 15 0 0
Toxic3 ppm
35 50 35 200 0 0 0 0 35 50 35 200 0 0
LEL %CH4 %LEL
10 20 24 24 24 24
10 20 24 24
O2
Toxic 1
Toxic 1
Status
%
TWA ppm STEL ppm Codes
Serial Number
Unit Config
—
—
33——
S104-000001
—
—
44——
S104-000001
19.5 —
—
1111—-
S104-000001
23.5 —
—
2222—-
S104-000001
20.9 —
—
-D-E—D S104-000001 FCEKNL
20.9 0
0
——–
S104-000001 FCEKNL
20.9 0
0
–L—-L
S104-000001 FCEKNL
20.9 0
0
LLH—-M S104-000001 FCEKNL
0
0
LLH—-M S104-000001
0
0
LLL—-M S104-000001
19.5 0
0
-L—–L
S104-000001
23.5 0
0
——–
S104-000001
20.9 0
0
——-B-
S104-000001 FCEKNL
20.9 0
0
——B-
S104-000001 FCEKNL
78
Table 18. Datalog Status Codes
Normal operation
L Low alarm H High alarm T TWA alarm U TWA and low alarm (dual alarms) V TWA
and high alarm (dual alarms) s STEL alarm u STEL and low alarm (dual alarms) f
Fresh air delay @ Zero CO2
P Plugged (blocked) – pump alarm
Batteries OK
L Low alarm H High alarm T TWA alarm
Status Codes G Backlight is on v STEL and high alarm (dual alarms) w TWA and
STEL alarm (dual alarms) x TWA, STEL, and low (triple alarms) y TWA, STEL, and
high (triple alarms) O Overload / sensor is over-ranged C Calibrating F
Failure – sensor failure I Time set
Pump Codes F Failure / pump failure
Battery Status Codes B Low battery alarm
Alarm Status Codes M Multi alarm C Calibration Q Off / quit / manual
deactivation
Note: TWA readings greater than 99 are recorded as OL.
GasAlertMicro 5/PID/IR View Datalog Files in Spreadsheets
1 Alarm setpoint 1 (low alarm) 2 Alarm setpoint 2 (high alarm) 3 Alarm
setpoint 3 (TWA alarm) 4 Alarm setpoint 4 (STEL alarm) D Calibration due date
(days) E Elapsed / last calibration (days) Z Auto zeroing t Testing
C Confidence beep is active
S Automatic shutdown F Failure / self-test fail R RTCC / real-time clock
failure
79
GasAlertMicro 5/PID/IR User Manual
Table 19. Datalog Gas and Correction Factor Sensor Codes
A No sensor
B
F
SO2
G
K NH3
L
Q IR
A Acetaldhyde
B
F Diesel
G
K Isobtyln
L
P Styrene
Q
U Custom
A Acetone
B
F Ethyl_Ace
G
K Isobutylene
L
P Octane
Q
U Custom
V LEL in % by Vol CH4
H2S PH3 ClO2
Gas Sensor Codes
C
H2S COSH
D
H
NO2
I
M
O3
O
CO HCN LEL
Correction Factor Codes for PID (if applicable)
Acetone Ethanol JP8 Toluene
C
Ammonia
H
Ethylene
M
Kerosene
R
Turpentine
D
Benzene
I
Gasoline
N
MEK
S
Vinyl_Cl
Benzene Gasoline Isopropanol Pentane
Correction Factor Codes for LEL
C
Butane
D
H
Heptane
I
M
MEK
N
R
Propane
S
Cyclohexane Hexane Methane Toluene
LEL Unit Codes
L
LEL in % LEL
E
CO COSH
J
Cl2
P
PID
E
Butadiene
J
Hexane
O
Naptha
T
Xylene
E
Ethanol
J
Hydrogen
O
Methanol
T
Turpentine
80
Maintenance
To maintain the detector in good operating condition, perform the following
basic maintenance as required.
· Calibrate, bump test, and inspect the detector at regular intervals. ·
Maintain an operations log of all maintenance, calibrations, bump
tests, and alarm events. · Clean the exterior with a soft damp cloth. Do not
use solvents,
soaps, or polishes. Refer to Sensor Poisons and Contaminants. · Do not immerse
the detector in liquids.
Battery Cautions a Warning
To avoid personal injury and/or property damage, adhere to the following:
· Replace the batteries immediately when the detector emits a low battery
alarm.
· Use only batteries that are recommended by BW Technologies by Honeywell.
Refer to Specifications.
· Ensure the alkaline batteries are properly installed in the detector battery
pack.
· Charge batteries using only a recommended BW charger. Do not use any other
charger. Failure to adhere to this caution can cause fire and/or explosion.
· The detector must be deactivated to charge the battery.
GasAlertMicro 5/PID/IR Maintenance
· Do not calibrate the detector during or immediately after charging.
· Both the rechargeable lithium battery pack and the alkaline battery pack are
user-changeable in hazardous locations, but the alkaline battery cells inside
the pack can only be replaced in a safe area that is free of hazardous gas.
· Warning: The battery pack (M5-BAT08) that is equipped with a lithium polymer
cell may present a risk of fire or chemical burn hazard if misused. Do not
recharge, disassemble, heat above 212ºF (100ºC), or incinerate.
· Warning: Do not use any other lithium batteries with the Micro5/PID/IR
detectors. Use of any other cell can cause fire and/or explosion. To order a
replacement lithium battery pack (M5-BAT08), refer to Replacement Parts and
Accessories.
· Warning: Lithium Polymer cells exposed to heat at 266ºF (130ºC) for 10
minutes can cause fire and/or explosion.
· Dispose of used lithium cells immediately. Do not disassemble and do not
dispose of in fire. Do not mix with the solid waste stream. Spent batteries
must be disposed of by a qualified recycler or hazardous materials handler.
· Keep lithium cells away from children.
Charging the Battery
To charge the battery, refer to the GasAlertMicro 5/PID/IR Charger Instruction
Sheet.
81
GasAlertMicro 5/PID/IR User Manual Replacing the Alkaline Batteries
To replace the alkaline batteries, refer to Figure 13., Table 20., and the
following steps 1-6.
Figure 13. Replacing the Alkaline Batteries
Table 20. Replacing the Alkaline Batteries
Item
1 2 3 4 5 6 7
Description
Detector Latch Battery pack Battery tray Captive screws (2) Alkaline batteries
(3) Battery shell
1. Deactivate the detector. 2. Open the latch on the bottom of the detector.
3. Remove the battery pack by lifting the bottom of the pack away
from the detector. 4. On the battery pack, remove the two captive screws.
Remove
the battery shell from the battery tray. 5. Replace the three alkaline
batteries. 6. Replace the battery pack and the captive screws. 7. Reinsert the
battery pack and secure the latch.
82
Replacing the Lithium Battery Pack
To replace the lithium battery pack, refer to Figure 14. and the following
steps 1-4.
Figure 14. Replacing the Lithium Battery Pack
GasAlertMicro 5/PID/IR Maintenance
1. Deactivate the detector. 2. Open the latch on the bottom of the detector.
3. Remove the battery pack by lifting the bottom of the pack
upward from the detector. a Warning
Do not disassemble lithium battery packs. Read and adhere to the cautions in
Battery Cautions.
4. Replace the battery pack with a fully charged lithium battery pack.
5. Close the latch.
Replacing a Sensor or Sensor Filter
a Warning To avoid personal injury, only use sensors that are specifically
designed for the detector. Refer to Specifications.
a Caution Replace the sensor in a safe and non-hazardous area that is free of
hazardous gas Each sensor has a high degree of resistance to common vapors and
gases. To clear a sensor, place the detector in a clean environment and wait
10 to 30 minutes. Do not expose a sensor to vapors from organic solvents such
as paint thinners and acetone. For a list of common products that can damage
sensors, refer to Sensor Poisons and Contaminants.
83
GasAlertMicro 5/PID/IR User Manual
For sensor problems, refer to Troubleshooting.
To replace a sensor or sensor filter, refer to Figure 15., Table 21., and the following procedures 1-7.
Figure 15. Replacing a Sensor or Sensor Filter Note
Detectors that are configured for 1, 2, 3, or 4 gases may contain a dummy
sensor in one of the four sensor locations.
84
Table 21. Replacing a Sensor or Sensor Filter
Item 1 2 3 4
5
Description Sensor cover Sensor filter Sensors Detector Machine screws (2)
1. Deactivate the detector.
2. Remove the two machine screws on the rear shell and then remove the sensor
cover or pump module.
3. Remove the sensor filter and/or the sensor(s).
4. Insert the new filter and/or sensor. Ensure the sensor posts are aligned
correctly.
5. Re-assemble the detector.
6. If the sensor is changed, such as SO2 to an H2S, the detector must be
reconfigured. Refer to Sensors in the Tech Mode option.
7. A sensor that has been replaced must be calibrated immediately. Refer to
Calibration and Setting Alarm Setpoints.
GasAlertMicro 5/PID/IR Maintenance
Photoionization Detector (PID)
Clean or Replace the Lamp
The PID lamp must be cleaned regularly. Use only the cleaning kit that is
supplied by BW Technologies by Honeywell.
To clean the PID lamp, refer to the illustrations and procedures that are
provided with the PID Lamp Cleaning Kit. To order the kit, refer to
Replacement Parts and Accessories.
a Caution
To ensure proper maintenance and continued accurate readings from the sensor,
use only the PID Lamp Cleaning Kit that is supplied by BW Technologies by
Honeywell.
Table 22. Parts of the PID sensor
Item 1 2 3 4
5
Description PID sensor Diffusion barrier Sensor cover Electrode stack Lamp
85
GasAlertMicro 5/PID/IR User Manual
Replace the Lamp
Replace the lamp when it falls below the acceptable level. Possible indicators
that the lamp requires replacement are as follows:
· The detector will not calibrate. · The startup self-test fails. · The ppm
levels are incorrect. To replace the lamp, refer to the illustrations and
procedures in the PID Lamp Cleaning Kit. If required, contact BW Technologies
by Honeywell for more information.
Figure 16. Parts of the PID 86
Replace the Electrode Stack
Replace the electrode stack when it is contaminated. To replace the electrode
stack, refer to Table 22., Figure 16., and the following procedures.
a Caution Ensure your fingers do not make contact with the diffusion barrier
or the electrodes on the underside of the stack.
1. Remove the sensor cover. 2. Remove the old electrode stack. 3. Insert the
new electrode stack. 4. Replace the sensor cover.
WEEE Battery Directive
Failure to comply with the following battery removal and disposal instructions
may result in battery shorting, battery leakage, and/or other damage. Ensure a
qualified technician completes the following procedures.
Removal and Disposal of the Battery Pack
To remove the alkaline batteries, refer to steps #1 to #3 in Replacing the
Alkaline Batteries on page 82. To remove the lithium battery pack, refer to
steps #1 and #2 in Replacing the Lithium Battery Pack on page 83. Dispose of
the battery pack according to local laws.
Removal and Disposal of the Coin Cell
GasAlertMicro 5/PID/IR Maintenance
87
GasAlertMicro 5/PID/IR User Manual
Item 1 2 3 4 5 6 7 8 9 10
Description Diffusion cap Sensor filter Sensors Front shell Sensor board Back shell Machine screws (for diffusion cap) Battery pack Datalog card Machine screws (for rear shell)
The detector contains a coin cell to power the real-time clock. Only a
qualified technician should complete the following procedure.
1. Deactivate the detector.
2. Open the latch on the bottom of the detector.
3. Remove the battery pack by lifting the bottom of the pack upward from the
detector.
4. Remove the MMC card.
5. Remove the two machine screws on the rear shell and then remove the
diffusion cap or pump module.
88
6. Remove the sensors. 7. Remove the four machine screws in the battery pack
cavity. 8. Remove the rear shell. 9. The coin cell sits on the sensor board.
Gently remove the sensor board. 10. The coin cell is connected to the board by
four leads. Clip the
four leads individually to remove the coin cell. a Caution
Do not touch two or more leads while disconnecting the battery cell. 11.
Dispose of the coin cell according to local laws.
Note Image above shows three of the four battery leads that must be clipped.
GasAlertMicro 5/PID/IR Troubleshooting
Troubleshooting
If a problem occurs, refer to the solutions provided in Table 23. If the
problem persists, contact BW Technologies by Honeywell.
Table 23. Troubleshooting
Problem
Possible Cause
Solution
Startup Troubleshooting The detector does not activate.
The detector immediately enters alarm mode when activated.
No batteries Depleted batteries Damaged or defective detector Sensor needs to stabilize Low battery alarm Sensor alarm Pump alarm
The startup self-test fails.
General fault
Table 21. Troubleshooting
Refer to Replacing the Alkaline Batteries or Replacing the Lithium Battery
Pack.
Refer to Replacing the Alkaline Batteries, or Replacing the Lithium Battery
Pack.
Contact BW Technologies by Honeywell.
Used sensor: wait 60 seconds. New sensor: wait 5 minutes.
Refer to Battery Cautions, Replacing the Alkaline Batteries, Replacing the
Lithium Battery Pack.
Refer to Replacing a Sensor or Sensor Filter.
If the sampling hose is attached, determine if it is obstructed. If not, clean
or replace the pump filter. If the pump alarm persists, refer to the “Pump
Operation” section in Troubleshooting.
Ensure that the sensors and battery pack are installed correctly and then
reactivate the detector. If the fault persists, record the error message and
contact BW Technologies by Honeywell.
89
GasAlertMicro 5/PID/IR User Manual
Problem
Possible Cause
Detector Operation Troubleshooting
Detector does not display normal ambient gas readings after startup self-test.
Sensor not stabilized
Detector requires calibration Target gas is present
Detector does not respond to buttons.
Batteries are depleted
Detector is performing operations that do not require user input
Detector does not accurately measure gas.
Detector requires calibration
Detector is colder/hotter than ambient gas
Sensor filter is blocked
Solution
Used sensor: wait 60 seconds New sensor: wait 5 minutes Calibrate the sensors.
Refer to Calibration and Setting Alarm Setpoints. Detector is operating
properly. Use caution in suspect areas. Refer to Replacing the Alkaline
Batteries or Replacing the Lithium Battery Pack. Button function restores
automatically when the operation ends.
Calibrate the sensors. Refer to Calibration and Setting Alarm Setpoints. Allow
the detector to adjust to ambient temperature before using.
Clean the sensor filter. Refer to Replacing a Sensor or Sensor Filter.
90
Table 21. Troubleshooting
Problem
Possible Cause
MMC/SD Card Troubleshooting
The MMC/SD card is not inserted.
GasAlertMicro 5/PID/IR Troubleshooting
Solution
Insert the MMC/SD card. Refer to Inserting the MMC/SD Card.
The MMC/SD card that is inserted in the detector has a storage size that is not supported by the detector.
Insert an approved MMC/SD card: · 128 MB Delkin MMC · 128 MB Delkin SD card · 128 MB Transcend SD card · 64 MB Unigen SD card
The detector has lost communication with the MMC/ SD card.
Retry communication
Insert a new approved MMC or SD card. Refer to Inserting the MMC/SD Card.
Reformat the MMC or SD card in windows and then reinsert into the detector.
Contact BW Technologies by Honeywell.
91
GasAlertMicro 5/PID/IR User Manual
Table 21. Troubleshooting
Problem
Possible Cause
Alarms Troubleshooting
Detector does not enter alarm mode.
Alarm setpoint(s) are set incorrectly.
Alarm setpoint(s) are set to zero.
Detector requires calibration.
Detector intermittently enters alarm without any apparent reason.
Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas.
Alarms setpoints are set incorrectly.
Missing or faulty sensor.
Detector enters alarm with LEL reading displaying as OL (over limit).
Pump flow rate is set too high.
Automatic Deactivation Troubleshooting
Detector automatically deactivates.
Automatic shutdown activated because of weak batteries.
Calibration is overdue and the Due-lock (calibration user lock-out) option is enabled.
92
Solution
Reset alar
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>