weigh Scales SW-II Instructions
- June 6, 2024
- weigh
Table of Contents
weigh Scales SW-II Instructions
Note: Below the scale there is a breathable knob. Its default setting is for weigh. If you want to wash tighten the knob, after wash release the knob for weigh
When want to weigh, release screw for getting normal pressure
When want to wash, tight screw for getting seal
Parameters
Enter setting mode
To set parameters, first turn on the weighing scales, hold the tare key during
self-checking, until display show “F1 CAL”.
set resolution
After enter setting parameter, the display shows “F1 CAL”, press
tare key, display shows “F2 RES “, press ZERO key to enter, press tare
key to select resolution,
Change capacity
After enter setting parameter, the display shows “F1 CAL”, press
tare key, until display shows “F3 CAP “, press ZERO key to enter, press tare
key to select capacity , press zero key to sure.
Check internal counts
After enter setting mode, display shows “F1 CAL”, press tare key
select , until display shows “F4 INP”, press zero key to enter, display shows
internal counts, press TARE key to return setting mode.
Automatic sleep
In the weighting mode, on the waterproof scale weight don’t move
and change on the pan within a minute. The scales will Automatic sleep. If
users want to waken scale increase 4d weight on the pan.
INTRODUCTION
The SW-II series of waterproof scale are ABS plastic housing and stainless
steel platform.
The SW-II series of waterproof scale have excellent outside design of china
patented IP67 waterproof technology. A/D electron circuitry use humidity
resistant design in damp environment use natural.
The SW-II series of waterproof scale are super low power consumption.
All the keypads are light touch switches and the displays are large easy to
read liquid crystal type displays (LCD). The LCD’s size is 25mm, the LCD’s
without a backlight, this method is used for economize electric power. If user
want to use, User select optional 20mm rear display. All units include
automatic zero tracking, tare, unit shift etc function.
Key function
Zero
Set the zero point for all subsequent weighing. The display shows zero.
Tare
Tares the scale. Stores the current weight in memory as a tare value,
subtracts the tare value from the weight and shows the results. This is the
net weight.
Unit
Press tare key and zero key to together in the weighing mode.
ON/OFF
Turn on or turn off the power.
OPERATION
Zero the display
You can press the ZERO key at any time to set the zero point from which all
other weighing and counting is measured, within 2% of power up zero. This will
usually only be necessary when the pan is empty.
When the zero point is obtained the display will show indicator for zero. The
scale has an automatic rezeroing function to account for minor drifting or
accumulation of material on the platform. However you may need to press the
ZERO key to rezero the scale if small amounts of weight are shown when the
platform is empty.
Taring
Zero the scale by pressing the ZERO key if necessary. The zero indicator will
be on.
Place a container on the platform, a value for its weight will be displayed.
Press the TARE key to tare the scale. The weight that was displayed is stored
as the tare value and that value is subtracted from the display, leaving zero
on the display. The “TARE” indicator will be on.
As product is added only the weight of the product will be shown. When the
container is removed a negative value will be shown. If the scale was tared
just before removing the container this value is the gross weight of the
container plus all product that was removed.
The zero indicator will also be on because the platform is back to the same
condition it.
Change unit
Press TARE key and Zero key to together will change weighing unit in the
weight mode, relevant weighing unit mark turn on (kg or lb).