PHOTOS 31 QUICKSTART
Preparation
Confirm that PHOTOS 31 components are intact. For installation, gather a
mounting post and either zip ties or 4-40 bolts. Read the full PHOTOS 31 User
Manual at metergroup.com/phytos31support. All products have a 30-day
satisfaction guarantee.
Connecting
Plug into Data Acquisition System
Connect the stereo plug connector into any METER data logger and configure it
to read the PHOTOS 31 (refer to the PHOTOS 31 User Manual). Select a nonzero
measurement interval to ensure data are being logged. To connect to a non-
METER data logger, see the PHOTOS 31 User Manual.
Verify Readings
Use the SCAN function in the software to show a list of readings. A dry
sensor should have raw counts between 430−445 counts.
ATTENTION
For best results, use the latest versions of METER software and firmware for
the computer or mobile device, products, and sensors. Please use the software
Help menu to find updates. Consult the sensor user manual for more
troubleshooting tips.
Installation
1. Install Sensor to Mounting Post
Install the mounting post. With the PHOTOS 31 electrode traces facing up,
thread zip ties or 4-40 bolts through the mounting holes on the sensor body
and attach them to the mounting apparatus.
NOTE: Wetness duration is affected by the mounting angle. Mount the
sensor at the appropriate angle for the desired measurement.
2. Plug Sensor In and Configure Logger
Plug the sensor into the data logger. Use data logger software to apply
appropriate settings to the sensors plugged into each data logger port.
SUPPORT
Have a question or problem? Our support team can help.
We manufacture, test, calibrate, and repair every instrument in the house. Our
scientists and technicians use the instruments every day in our product
testing lab. No matter what your question