GIMA 28209 VEGA UNO SUCTION ASPIRATOR User Manual
- June 6, 2024
- GIMA
Table of Contents
PROFESSIONAL MEDICAL PRODUCTS
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy
gima@gimaitaly.com –
export@gimaitaly.com
www.gimaitaly.com
Made in Italy
VEGA UNO SURGICAL ASPIRATOR is a medical device powered by 230V ~ 50Hz electricity, to be used for nasal, oral, tracheal aspiration of body fluids, such as mucus, phlegm, and blood, in adults or children. Luminaire is designed to offer ease of transport and continuous use. Thanks to these features and its performance, this product is particularly suitable for use in hospital wards, for minor surgery applications, and for post-operative treatments at home. Built with a plastic body with high thermal and electrical insulation in compliance with European safety regulations, the appliance is supplied with a complete suction tank in sterilizable polycarbonate, with an overflow valve, and is equipped with a suction regulator and vacuum gauge placed on the front panel.
GENERAL WARNING
Read the instruction manual carefully before use. The device is for use by
qualified personnel (surgeon / professional nurse/assistant) The use of the
device at home is restricted to an adult in full possession of mental
faculties and/or home carers The instrument must not be disassembled. For
technical service always contact Gima S.p.A.
IMPORTANT SAFETY RULES
-
Check the condition of the unit before each use. The surface of the unit should carefully be inspected for visual damage. Check the mains cable and do not connect to power if the damage is apparent;
-
Before connecting the appliance always check that the electric data indicated on the data label and the type of plug used, correspond to those of the mains electricity to which it’s to be connected;
-
Respect the safety regulations indicated for electrical appliances and particularly:
– Use original components and accessories provided by the manufacturer to guarantee the highest efficiency and safety of the device;
-The device can be used only with the bacteriological filter;
– Never immerse the appliance into the water;
– Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin. In the event it is accidentally dropped, do not attempt to remove the device from the water whilst the plug is still connected: disconnect the mains switch, remove the plug from the power supply and contact the GIMA technical service department. Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the GIMA technical service department.
-Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;
– To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which may expose the inner parts of the energized device, do not connect the plug to the electrical socket? Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the GIMA technical service department.
– Don’t use in the presence of inflammable substances such as anesthetic, oxygen, or nitrous oxide;
– Don’t touch the device with wet hands and always prevent the appliance from coming into contact with liquids;
– Don’t leave the appliance connected to the power supply socket when not in use;
– Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;
– Store and use the device in places protected against the weather and far from any sources of heat. After each use, it is recommended to store the device in its own box away from dust and sunlight.
– In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be necessary, you must use ones that are in compliance with safety regulations, however, take care not to exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.
– Prevent children from using the device without proper supervision;
– Never leave the appliance near water, do not immerse it in any liquid. If the device has fallen into the water, unplug it before you hold it. Do not use the appliance if the plug or AC / DC power supply is damaged or wet (send it immediately to an authorized service center or technical service). -
For repairs, exclusively contact technical service and request the use of original spare parts. Failure to comply with the above can jeopardize the safety of the device;
-
Use only for the purpose intended. Don’t use for anything other than the user defined by the manufacturer. The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not complying with current regulations.
-
The medical device requires special precautions regarding electromagnetic compatibility and must be installed and used in accordance with the information provided with the accompanying documents: the VEGA UNO SUCTION UNIT device must be installed and used away from mobile and portable RF communication devices (mobile phones, transceivers, etc.) that may interfere with the said device.
-
Instrument and accessory discharging must be done according to current regulations In the coun try of use.
-
WARNING: Do not change this equipment without the permission of the manufacturer GIMA S.p.A. None of the electric or mechanical parts have been designed to be repaired by customers or end-users. Don t open the device, do not mishandle the electric/mechanical parts. Always contact technical assistance
-
Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety and the technical characteristics of the same.
-
The medical device is in contact with the patient by means of a disposable probe (not supplied with the device). Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standard on material biocompatibility.
-
The product and its parts are biocompatible in accordance with the requirements of regulation EN 60601-1.
-
operation of the device is very simple and therefore no further explanations are required other than those indicated in the following user manual.
-
Use in Home-Care: Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed.
-
Do not leave the device unattended in places accessible to children and/or persons not in full possession of mental faculties they may strangle themselves with the patient’s tube and/or the power cable.
The manufacturer cannot be held liable for accidental or indirect damages should the device be modified, repaired without authorization or should any of its components be damaged due to ac cident or misuse. Any minimal modification/repair on the device voids the warranty and does not guarantee compliance with the technical requirements provided by the MDD 93/42/EEC (and sub sequent changes) and it’s normative.
CONTRAINDICATIONS
– Before using the VEGA UNO, consult the instructions for use: failure to read
all the instructions in this manual can be harmful to the patient.
– The device cannot be used to drain chest fluids;
– The device must not be used for suction of explosive, corrosive, or easily
flammable liquids.
– VEGA UNO is not suitable for MRI. Do not introduce the device in MRI
environments.
TECHNICAL CHARACTERISTICS
Model | VEGA UNO SUCTION ASPIRATOR |
---|---|
Typology (MDD 93/42/EEC) | Medical device Class IM |
Classification UNI EN ISO 10079-1 | High Vacuum / Low Flow |
Main Voltage | 230 V∼/ 50 Hz |
Power consumption | 184 VA |
Fuse | F1 x 1.6AL250 V |
Maximum suction aspiration (without jar) | -75kPa (- 0.75bar) |
Regolable from -75kPa (-0.75 bar) to -10kPa (-0.10 bar)
Maximum flow (without jar)| 15 Vmin
Weight| 2.2 Kg
Dimension| 250 x 190 (h) x 160 mm
Functioning| NON-STOP OPERATED
Accuracy of Vacuum Indicator| ± 5%
Working Condition| Room temperature: 5 =35°C
Room humidity percentage: 10 + 93% RH
---|---
Atmospheric pressure: 700 + 1060 hPa
Conservation condition and Transport| Room temperature: -25 + 70°C
Room humidity percentage: 0 + 93% RH Atmospheric pressure: 500 + 1060 hPa
The technical specifications may change without notice.
CLEANING OF THE DEVICE
Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device’s external parts always use a cotton cloth dampened with detergent. Don’t use abrasive or solvent detergents. Before carrying out any cleaning and/or maintenance operation, disconnect the appliance from the power supply, unplug it or turn off the switch on the device
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protective gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
- Complete aspiration jar 1000ml
- Conical fitting
- Tubes set 6mm x 10mm (trasparent silicon)
- Aspiration probe CH20
- Antibacterial and hydrophobic filter
Available under request with different versions with complete jar 2000m1.
Anti – bacterial and hydrophobic filter: designed for the individual patient to protect patient and machine from cross-infections. Prevents the liquids, that come into contact with it, from passing through it. Replace it whenever you suspect that it may be contaminated and/or it becomes wet or discolored. Replace the filter every time it is used if the suction pump is used on patients in unknown pathological situations and where an assessment of indirect contamination is not possible. The filter is not manufactured to be decontaminated, dismantled, and/or sterilized. If, however, the patient’s pathology is known and/or there is no risk of indirect contamination, the filter should be replaced after every work shift or once a month even if the device is not used.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-infections. Don’t use after the lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standard on material biocompatibility.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be replaced if there are visible signs of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of
cleaning cycles are strictly linked to the employment of the said component.
Therefore, after each cleaning cycle, it is up to the final user to verify
whether the fitting is suitable for reuse. The component must be replaced if
there are visible signs of decay in the material constituting the said
component.
Service life of the device : More than 850 hours of operation (or 3
years) in accordance with the standard conditions of testing and operation.
Shelf life: maximum 5 years from the date of manufacture.
CLEANING OF ACCESSORIES
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories. Washing and/or cleaning the autoclavable jar is to be carried out as follows:
-
Wear protective gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances.
-
Disconnect the tank from the device and remove the said container from the support of the device.
-
Separate all the parts of the cover (overflow device, washer).
-
Disconnect all tubes from the jar and the protection filter.
-
Empty and dispose of the contents of the suction vessel (also comply with regional regulations);
-
Wash each part of the container from secretions under cold running water and then clean every single part in hot water (temperature not exceeding 60°C). Once again, carefully wash every single part using, if necessary, a non-abrasive brush to remove any deposits.
-
Rinse with hot running water and dry all parts with a soft cloth (non-abrasive).
-
Dispose of the aspiration catheter according to the provisions of local laws and
The jug and lid can be further disinfected using a common disinfectant, strictly following the instructions and dilution values provided by the manufacturer. At the end of cleaning operations, leave to air dry in a clean environment.
The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed 60°C). After cleaning, leave the parts to dry in an open, clean environment.
When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:
- Place the overflow valve into its seat in the cover (under VACUUM connector)
- Insert floating valve keeping the o-ring towards the opening of the cage
- Place the o-ring into its seat around the cover
- After completing assembling operations always make sure that the cover seals perfectly to avoid vacuum leakages or liquid exit
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure — 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit, the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at an environmental temperature of the parts make sure that these are not damaged. The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure —15 min).
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure —15 min).
PERIODICAL MAINTENANCE CHECKS
The VEGA UNO SUCTION ASPIRATOR does not need maintenance or lubrication. It
is, however, necessary to inspect the unit before each use. With regard to
training, given the information contained in the user manual and since it is
easy to understand the said device, it doesn’t appear to be necessary.
Unpack the instrument and always check the integrity of plastic parts and
feeding cable, they might have been damaged during previous use. Connect the
cable to the electrical network and turn the switch on. Close the aspirator
outlet with your finger and with the suction regulator at maximum check that
the vacuum indicators reach at least -75kPa (-0.75 bar). Rotate the knob from
right to left. The vacuum indicator should go down -25kPa (-0.25 bar).
Check that no loud noises are present. A protection fuse ( F 1 x 1.6 A L
250V ) reachable from the exterior and situated in the plug protects the
instrument. For fuses replacing, always the type and the range. Before
changing the fuse, disconnect the plug from the power supply socket.
Fault type | Cause | Solution |
---|---|---|
1. The suction unit doesn’t work | Cable is damaged External power source | |
failure | Replace the cable Check the external power source | |
2. No aspiration | Jar Cap not screwed on correctly | Unscrew the cap, and re- |
screw it
3. No aspiration| Lid seal is not in its seat| Unscrew the cap and insert the
seal properly in its seat
4. The Vacuum power on the patient side is either very low or absent|
- Vacuum regulator set to minimum
- Protection filter blocked or damaged
- Connection tubes blocked, kinked or disconnected
- Shut-off valve blocked or damaged
- Pump motor damaged
|
- Turn the vacuum regulator clockwise and check the value of the vacuum on the gauge
- Replace the filter
- Replace or reconnect the tubes, check the jar connections
- Empty the jar, or disconnect the tube from the jar and unblock the shut-off valve. The unit will only work in the upright position
- Refer to authorized service personnel
5. The float doesn’t close| If the cap has been washed, ensure that the float
is not partially detached| Fit the float into its place
6. The float doesn’t close| The float it’s covered by dirty material|
Unscrewed the cap, leave the and put it on autoclave
7. Low suction| The foam inside the jar| Fill the jar to 1/3 full of ordinary
water
8. No aspiration due to flow leakage of mucus| Filter blocked| Replace filter
Faults 1 – 2 – 3 – 4 – 5 – 6 – 7 – 8| None of the procedures have
achieved the desired results| Contact GIMA customer service
If the overfill security system it’s activated, don’t proceed with the liquid
aspiration. If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be
stopped by the bacteriological filter who avoiding the liquid penetration
inside the device.
2° case – If both the security system doesn’t work, there is the possibility
that liquid comes inside the device, in this case, return the device to GIMA
technical service.
Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting instructions, and any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE A GUARANTEE IF THE INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO TAMPER.
INSTRUCTION FOR USE
- The device must be checked before each use in order to detect malfunctions and/or damage caused by transport and/or storage.
- The working position must be such as to allow one to reach the control panel and to have a good view of the empty indicator, the jar, and the antibacterial filter.
- It is recommended not to keep the device in your hands and/or to avoid prolonged contact with the body of the apparatus.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and better efficiency of the overflow device.
- Connect one end of the short silicon tube, with an antibacterial filter, to the suction connector on the lid of the jar.
- The other tube, connected to the filter on one end, must be attached to the connector on the flask cover to which the float is secured inside. (overflow device). The overflow device starts working (the float closes the connector on the cover) when the maximum volume of liquid is reached, so no liquid can enter the machine (90% of the flask’s total volume), thus ensuring that the liquid cannot penetrate inside the machine. The device must be used on a flat worktop.
Filter assembling
Make sure the filter is assembled with the arrows on the side of the
patient.
WARNING: The inside of the medical device must be regularly checked for the
presence of liquids or other visible contamination (secretions). In the
presence of liquids or other visible contamination, immediately replace the
medical device due to the risk of an insufficient vacuum flow rate. These
products have been designed, tested, and manufactured exclusively for single
patient use and for a period no longer than 24 hours.
- Connect the long silicone tube to the “PATIENT” jar outlet
- Connect the other end of the long silicon tube to the probe plastic connector, then connect the suction probe to it.
- Connect the power cord to the device then connect the plug to the electrical mains supply.
- Push switch on the position I to start the suction.
- Unscrew the lid of the jar and fill the jar 1/3 full of ordinary water (this assists the unit to reach peak vacuum performance and makes clean-up easier) then re-screw the lid on the jar correctly.
- During operation, the jar has to be in a vertical position to avoid an overflow valve to cut off aspiration. Should this happen, switch off the device and disconnect the tube from the jar cover (from the “VACUUM” outlet).
- Once finished push the switch on O position and unplug.
- Remove the accessories and clean.
- At the end of each use, place the device in its box away from dust.
WARNING: The power supply cable plug is the element of separation from the electrical mains system: even if the units are equipped with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to allow a further method of disconnection from the mains supply system.
NEVER USE THE DEVICE WITHOUT A JAR AND / OR PROTECTION FILTER MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT SURVEILLANCE
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES
This section contains information regarding the conformity of the compliance
with the EN 60601-1-2 Standard (2015). The VEGA UNO surgical aspirator is an
electro-medical device that requires particular precautions regarding
electromagnetic compatibility and must be installed and commissioned according
to the electromagnetic compatibility information supplied. Portable and mobile
radio communication devices (mobile phones, transceivers, etc.) may interfere
with the medical device and should not be used in close proximity with,
adjacent to, or on top of the medical device. If such use is necessary and
unavoidable, special precautions should be taken so that the electro-medical
device functions properly in its intended operating configuration (for
example, constantly and visually checking for the absence of anomalies or
malfunctions). The use of accessories, transducers, and cables differing from
those specified, with the exception of transducers and cables sold by the
appliance and system manufacturer as spare parts, can lead to an increase in
emissions or in a decrease of the immunity of the device or system. The
following tables supply information regarding the EMC (Electromagnetic
Compatibility) characteristics of the electro-medical device.
Guidance and manufacturer’s declaration — electromagnetic Emissions
The VEGA UNO SUCTION UNIT is intended for use in the electromagnetic
environment specified below. The customers or the user of the VEGA UNO SUCTION
UNIT should make sure that it’s used in such an environment.
Emissions test| Compliance| Electromagnetic environment –
guidance
Irradiated / Conducted emissions CISPR11| Group 1| The VEGA UNO SUCTION UNIT
only used RF energy only for its internal functioning. Therefore, its RF
emissions are very low and are not cause interference in the proximity of any
Electronic appliances.
Irradiated / Conducted emissions CISPR11| Class [B]| The VEGA UNO SUCTION UNIT
can be used in all environments, including domestic and those connected
directly to the public mains distribution that supplies power to environments
used for domestic scopes.
Harmonic emissions EN 61000-3-2| Class [A]
Voltage fluctuations / flicker emissions EN 61000-3-3| Complies
Guidance and manufacturer’s declaration — Immunity Emissions
The VEGA UNO SUCTION UNIT is intended for use in the electromagnetic
environment specified below. The customers or the user of the VEGA UNO SUCTION
UNIT should make sure that it’s used in such an environment.
Immunity Test| Level indicated by the EN 60601-1-2|
Compliance Level| Electromagnetic environment – guidance
Electrostatic
discharge (ESD) EN 61000-4-2
| +/-8kV on contact +/-15kV in air| The device doesn’t change its state|
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic materials, the relative humidity should be at least 30%.
Electrical fast transient/burst EN 61000-4-4| ± 2kV power supply lines
± 1 kV for input /
output lines
| The device doesn’t change its state| Mains power quality should be that of a
typical commercial environment or hospital.
Surge
EN 61000-4-5
| ± 1 kV differential mode +/-2 kV ordinary mode| The device doesn’t change
its state| Mains power quality should be that of a typical commercial
environment or hospital.
Loss of voltage, brief voltage interruptions and variations
EN 61000-4-11
| <5% UT (>95% dip UT) for 0,5 cycle
40% UT (60% dip UT) for 5 cycle
70% UT (30% dip UT) for 25 cycle
<5% UT (>95% dip UT) for 5 sec
| – –| Mains power quality should be that of a typical commercial environment
or hospital If the user of the VEGA UNO SUCTION UNIT request that the
appliance operates continuously, the use of a continuity unit is
recommended.
A magnetic field with network frequency (50/60 Hz)
EN 61000-4-8
| 30A/m| The device doesn’t change its state| The power frequency magnetic
field should be measured in the intended installation location to make sure
that it’s sufficiently low.
Note UT is the value of the power supply voltage
Guidance and manufacturer’s declaration – Immunity Emissions
The VEGA UNO SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the VEGA UNO SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test| Level indicated by the EN 60601-1-2| Compliance Level|
Electromagnetic environment – guidance
Conducted
Immunity
EN 61000-4-6
| 3Vrms 150kHz to 80Mhz(for non-life-supporting devices)| V1 = 3 V rms| The
portable and mobile RF communication devices, including cables, must not be
used closer to the VEGA UNO SUCTION UNIT device, then the separation distance
is calculated by the equation applicable to the transmitter frequency.
Recommended separation distance
Radiated
Immunity
EN 61000-4-3
| 10V/m 80MHz to 2.7GHz (for non-life-supporting devices)| E1 = 10 V / m
| | Where P is the maximum nominal output voltage of the transmitter in Watt (W) depending on the manufacturer of the transmitter and the recommended separation distance in meters (m). The intensity of the field from the fixed RF transmitters, as determined by an electromagnetic study of the site could be lower than the level of conformity of each frequency interval b). It is possible to check for interference in proximity to devices identified by the following symbol:
---|---|---
Note 1 : At 80 MHz and 800 MHz the interval with the highest frequency is
applied
Note 2 : These guidelines may not be applicable in all situations. The electromagnetic propagation is influenced by absorption and by reflection from buildings, objects, and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial mobile radio, amateur radio devices, radio AM and FM transmitters, and TV transmitters can not be theoretically and accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic study of the site should be considered. If the field intensity measured in the place where the device will be used surpasses the above-mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnormal performance arises, additional measures such as changing the device’s direction or positioning may be necessary. b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 10 V/m.
Recommended separation distance between portable and mobile radio- communication devices and the monitor
The VEGA UNO SUCTION UNIT surgical aspirator is intended to operate in an
electro-magnetic environment where RF irradiated interferences are under
control. The client or operator of the VEGA UNO SUCTION UNIT device can help
prevent electro-magnetic interference by keeping a minimum distance between
the portable and mobile RF communication devices (transmitters) and the VEGA
UNO SUCTION UNIT device, as recommended below, in relation to the radio-
communication maximum output power.
Maximum nominal output power of the Transmitter W| Separation
distance from the frequency transmitter (m)
150KHz to 80MHz
| 80MHz to 800MHz
| 800MHz to 2,7GHz
0.01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters with a maximum nominal output power not shown above, the
recommended separation distance in meters (m) can be calculated using the
equation applicable to the transmitter frequency, where P is the maximum
nominal output power of the transmitter in Watt (W) depending on the
transmitter’s manufacturer.
Note 1:At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guidelines may not be applicable in all situations. The electromagnetic propagation is influenced by the absorption and by the reflection from buildings, objects, and people
SYMBOLS
| Caution: read instructions (warnings) carefully| | Follow instructions for
use
---|---|---|---
| Keep in a cool, dry place| | Keep away from sunlight
| Manufacturer| | Date of manufacture
| Product code| | Lot number
| Medical Device complies
with Directive 93/42/EEC
| | Type BF applied part
| WEEE disposal| | Class II applied
| Serial number| | Temperature limit
| Fuse| | Atmospheric Pressure limit
IP21
| Covering Protection rate| | Alternating current
| ON|
Hz
| Mains frequency
| OFF| | Humidity limit
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>