STEP 1 Removing Cartridge
-
Turn off the X-PLORTM by pressing the power button for 2 seconds.
-
Remove the X-PLOR from the carrying bag, if applicable.
-
Remove the battery from the X-PLO₂R.
-
Place the X-PLO₂R on its side to access the underside of the POC.
-
Pull the cartridge out of the POC by pulling on the cartridge pull handle.
-
Remove the cartridge completely from the X-PLO₂R and throw away
STEP 2 Installing Cartridge
-
Remove cartridge from mylar bag. The new cartridge has a dust cap on top. Remove the dust cap. Make sure there is no dust or debris where the dust cap was located.
-
Insert new cartridge top first into the X-PLO₂R. Do not leave the cartridge ends exposed; the cartridge should be inserted into the X-PLO₂R as soon as the used cartridge is removed.
-
Push the cartridge into the X-PLO₂R so it is fully inside.
-
The cartridge pull handle should lay flat once the cartridge is secured.
-
Connect charged battery to the X-PLO₂R.
-
urn on the X-PLO₂R by pressing the power button for 2 seconds. Set the flow setting to your prescribed setting and use normally.