ThermoFisher SCIENTIFIC 88881101 CO2 Plus Digital Orbital Shaker Instruction Manual

June 1, 2024
ThermoFisher SCIENTIFIC

ThermoFisher SCIENTIFIC 88881101 CO2 Plus Digital Orbital Shaker

Product Information

Specifications

  • Product Name: MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker
  • Catalog Numbers: #88881101, #88881102, #88881103, #88881104

Product Usage Instructions

Section 1: Unpacking and Installation
Follow these steps to unpack and install the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker:

  1. Refer to the packing list to ensure all components are included.
  2. Install the universal (aluminium) platform or the 18-inch (aluminium) platform according to the provided instructions.
  3. Alternatively, install the rubber mat platform if preferred.
  4. Connect the instrument following the provided guidelines.

Section 2: Overview
Learn about the key aspects of the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker:

  • Review product specifications, environmental conditions, and safety tips.
  • Ensure proper storage of the signal cable.
  • Understand the maximum shaking speed of the shaker.

Section 3: Operation
Operate the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker effectively:

  1. Familiarize yourself with the control panel operations.
  2. Set desired speed and time settings for your application.
  3. Understand and utilize the alarm system as needed.

Section 4: Application
Explore different application possibilities with the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker.

Section 5: Safety Tips
Adhere to safety guidelines while using the shaker to ensure a secure working environment.

Section 6: Optional Accessories
Discover optional accessories that can enhance your experience with the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker.

FAQ

  • How can I contact customer service for assistance?
    In the United States, call 1-800-766-7000. In Canada, call 1-800-234-7437. For online orders, visit thermofisher.com for more information.

  • What are the available catalog numbers for the MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker?
    The catalog numbers are #88881101, #88881102, #88881103, #88881104.

Do You Need Information or Assistance on Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:

1-740-373-4763
1-800-438-4851
1-877-213-8051
http://www.thermoscientific.com
[email protected]
www.unitylabservices.com

Direct
Toll-Free, U.S. and Canada
FAX
Internet Worldwide Web Home Page Tech Support Email Address Certified Service Web Page

Our Sales Support staff can provide information on pricing and give you quotations. We can take your order and provide delivery information on major equipment items or make arrangements to have your local sales representative contact you. Our products are listed on the Internet and we can be contacted through our Internet home page.
Our Service Support staff can supply technical information about the proper setup, operation or troubleshooting of your equipment. We can fill your needs for spare or replacement parts or provide you with on-site service. We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products.

Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself…over the telephone without a service call. When more extensive service is necessary, we will assist you with direct factory-trained technicians or a qualified service organization for on-the-spot repair. If your service need is covered by the warranty, we will arrange for the unit to be repaired at our expense and to your satisfaction.

Regardless of your needs, our professional telephone technicians are available to assist you Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by telephone or fax.
International customers, please contact your local Thermo Scientific distributor.

Unpacking and Installation

  1. Inspect contents upon receipt of the instrument, if the instrument is upside down, contact the manufacturer immediately.
  2. Check packing list below. If there are any damaged or missing items, contact the manufacturer.
  3. Place the instrument on a level and firm surface, to avoid vibration and noise.

Packing List

ThermoFisher-SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(1\) ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(2\)

Universal (Aluminum) Platform Installation

  1. Place the instrument on a level and firm surface, it must be capable of supporting the weight and load of the instrument, and the impact force generated by the motion. Prepare related parts (see Figure 1-1).
  2. Install the horizontal universal platform on the fixed plate of the Main body, and ensure all four mounting holes in the universal platform and the four threaded mounting holes are in alignment.
  3. Fasten the universal platform and the fixed plate with four M4x14 cross-recessed countersink head screws.

18inch (Aluminum) Platform Installation

  1. Place the instrument on a level and firm surface, it must be capable of supporting the weight and load of the instrument, and the impact force generated by the motion. Prepare related parts (see Figure 1-2).
  2. Install the horizontal universal platform on the fixed plate of the Main body, and ensure all four mounting holes in the universal platform and the four threaded mounting holes are in alignment.
  3. Fasten the universal platform and the fixed plate with four M4x14 cross-recessed countersink head screws.

Rubber Mat Platform Installation

  1. Place the instrument on a level and firm surface, it must be capable of supporting the weight and load of the instrument and the impact force generated by the motion. Prepare related parts (see Figure 1-3).
  2. Install the horizontal platform on the fixed plate of the Main body, and ensure all four mounting holes on the platform and the four threaded mounting holes are in alignment.
  3. Fasten the platform and the fixed plate with four M4x10 cross-recessed countersink head screws.
  4. Place the rubber mat on the platform.

Connections of the Instrument
Refer to Figure 1-4 when connecting the instrument.ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(6\)

Overview

Specifications

  • Rotation Speed
    • Speed Range …………………………………………………..30 to 300rpm
    • Speed Accuracy ………………………………………………………….±1rpm
    • Orbit Diameter …………………………………………………………¢19mm
    • Motion Type ……………………………………………………………….Orbital
    • Speed Display …………………………………………………………………LED
  • Load
    • Maximum Load ………………………………………………6kg @≤300rpm
  • Timer
    • Timing Range …………..……………………………………….0~99h 59min
    • Hour Display.…………….…………………………..…………….……………LED
    • Minute Display ………..…………………..……………………..……………LED
  • Dimensions
    • Main body …………………..360×300×96mm(14.2×11.8×3.8inch)
    • Control Box ………………………164×95×26mm(6.5×3.7×1.0inch)
  • Weight
    • Net Weight (Main body) …………………………….17.5kg(35.27lbs)
    • Net Weight (Control Box) ……………………………….0.3kg(0.66lbs)
    • Gross weight …………………………………………………..19kg(41.9lbs)
  • Power Supply
    • Requirement ……………………………………..AC100-240V, 50/60Hz
  • Others
    • Certification …………………………RoHS, WEEE, cCSAus, CE Mark
    • Noise level …………………………………………………………………<60dB

Environmental Conditions

  • Application Environmental Conditions: indoor use
    • Temperature …………………………………………………………..4 to 60℃
    • Altitude …………………………………………………………………≤2,000 m
    • Humidity ……………………………………………………………. 20% to 99%
    • Co2 Concentration ………………………………………………..0% to 20%
  • Storage Environmental Conditions
    • Temperature …………………………………………………………..5 to 60℃
    • Altitude …………………………………………………………………≤2,000 m
    • Humidity …………………………………20% to 99%, non-condensing

Safety Tips

  1. Unplug and unload before moving the unit.
  2. This instrument uses a single-phase AC100-240V 50/60Hz power supply.
  3. Remove the power supply when the instrument is in maintenance.
  4. Do not clean the unit with flammable or corrosive liquid.
  5. Use specified power supply.
  6. Maximum load 6kg(13.23 lbs) @≤ 300 rpm.
  7. To ensure proper operation, the unit must be free of surrounding obstructions.

Storing the signal Cable
To store the signal cable, simply open the side plate on the main body. Refer to Figure 2-1. The signal cable can be placed inside.

Maximum Shaking Speed
Table 2-1. Maximum Shaking Speed on Rubber Platform in Relation to Flask Size.

Flask size Max Speed(rpm)
Size Liquid volume
50ml ≤50ml
100ml ≤100ml
150ml ≤150ml
200ml ≤200ml
250ml ≤250ml
500ml ≤500ml
1000ml ≤1000ml
2000ml ≤2000ml
3000ml ≤3000ml
5000ml ≤5000ml

Table 2-2. Maximum Shaking Speed and Flask Capacity on Universal Platform.

Flask size Max Flask Capacity Max Speed(rpm)
Size Liquid volume
50ml ≤50ml 30
100ml ≤100ml 15
150ml ≤150ml 15
200ml ≤200ml 15
250ml ≤250ml 15
500ml ≤500ml 9
1000ml ≤500ml 4
≤1000ml 4 240
2000ml ≤1000ml 4
≤2000ml 4 220
3000ml ≤1500ml 3
≤3000ml 3 220
5000ml ≤2500ml 1
≤5000ml 1 200

Note:

  1. The flask clamps (≤2L) can be placed anywhere on the universal (aluminum) platform, even if the load is not balanced. ▲
  2. Please refer to the “Liquid Volume” table in the specifications section for the maximum load. ▲

Table 2-3. Maximum Shaking Speed and Flask Capacity on 18 inch (aluminum) Platform.

Flask size Max Flask Capacity Max Speed(rpm)
Size Liquid volume
50ml ≤50ml 49
100ml ≤100ml 25
150ml ≤150ml 25
200ml ≤200ml 18
250ml ≤250ml 18
500ml ≤500ml 10
1000ml ≤500ml 7
≤1000ml 7 190
2000ml ≤1000ml 4
≤2000ml 4 180
3000ml ≤1500ml 4
≤3000ml 2 150
5000ml ≤2500ml 2
≤5000ml 1 150

Note:

  1. The flask clamps (<2L) can be placed anywhere on the 18inch (aluminum) platform, even if the load is not balanced. ▲
  2. Please refer to the “Liquid Volume” table in the specifications section for the maximum load. ▲

Operation

The unit has been designed with a “soft start/slow down” providing a smooth acceleration/ deceleration during operation. This feature will decrease splashing of solutions from vessels during operation.
Warning: The microprocessor control system may take 1 minute to synchronize the actual RPM of the universal (aluminum) platform to the desired RPM setting. Make sure that all the flasks and test tubes are firmly fixed on the universal platform and the flask clamps are fasten securely with the screws.

Operations on the Control PanelThermoFisher-SCIENTIFIC-88881101-CO2-Plus-
Digital-Orbital-Shaker- \(8\)

  1. Display LED: Shows time or speed
  2. RPM indicator light: LED shows current speed
  3. TIME indicator light: LED shows current time
  4. WARN indicator light: Over-speed alarm
  5. RUN indicator light: The instrument is running
  6. RPM/TIME Key: While instrument is in operation, press to change time and speed display; when setting parameter, press to change minutes, hour or speed.
  7. Up Arrow Key: Increases value. Hold for 3 seconds to increase rapidly
  8. Down Arrow Key: Decreases value. Hold for 3 seconds to decrease rapidly.
  9.  Key: The power is On/Off.
  10.  Key: The instrument is running or stopped.

Speed and Time Settings

  • Speed setting
1. Press RPM/TIME key. The speed indicator light is lit.
2. Press up and down arrow keys to access the speed setting state, speed value starts flashing.
3. Press up and down arrow keys to change the speed value. Speed increases or decreases. The minimum unit is 1rpm.
4. During settings, the unit will reset to current values if no value is input after 3 seconds.
5. Press key to run or stop.
  • Time setting
1. Press RPM/TIME key. The time indicator light is lit.
2. Press up and down arrow keys to access the time setting state. Speed value starts flashing;
3. Press up and down arrow keys to change the time value. Time increases or decreases. The minimum unit is 1 minute or 1 hour.
4. During settings, the unit will reset to current values if no value is input after 3 seconds.
5. Press key to run or stop.  

Notes

  1. When the shaker is running, the LED display will switch between the speed and timer function every 30 seconds automatically. Press RPM/TIME key to manually toggle between the speed and timer value.
  2. In “timer mode” the LED display will show the remaining running time.
  3. In “continuous mode” the LED display will show elapsed run time.
  4. After the run is complete, the LED display will read “end”. Press any key to place the shaker in standby mode.
  • Alarm System
    • Alarm System
    1. During operation, the alarm will sound after 10 seconds if the unit has not begun shaking. In addition, the alarm will sound and stop for the speed over 350 rpm after 2 seconds.
    2. Once the shaker begins to alarm, the RUN indicator light will not be visible and the WARN indicator light will flash, and the unit will stop running.
    • Power failure recovery
      When the power supply is cut off suddenly to the shaker, resulting in a power failure, the unit will automatically run at the previous set parameter upon power restoration.

Application

Application on Lab Benches
To operate the shaker outside of a CO2 incubator, refer to Specifications in Section 2.

Application in CO2 Incubator
Place the main body of the shaker in the CO2 incubator, and use the signal cable to connect the control box through the access port of the CO2 incubator. Refer to Figure 4-1.ThermoFisher-SCIENTIFIC-88881101-CO2-Plus-Digital-
Orbital-Shaker- \(11\)

Warning

  1. Never put the control box in the CO2 incubator.
  2. The back of control box is magnetized, and can be attached on the surface of the CO2 incubator.
  3. When the main body of CO2 Resistant Shaker is in operation in the CO2 incubator, keep a safe distance from the inner wall of the CO2 incubator to avoid collisions.
  4. The main body of the shaker should be connected to the control box wire through the access port of the CO2 incubator.
  5. If the cells are sensitive to vibrations, do not put them in the same incubator or on the shaker.

Safety Tips

  1. Confirm local power supply voltage before use the unit.
  2. The unit must not be plugged into a power supply extension with other pieces of equipment.
  3. The data cable between the main body and control box cannot be extended.
  4. Never modify the supplied power cables.
  5. Service should only be performed by a qualified professional.
  6. In the following situations, the power supply must be unplugged:
    • When the unit is moved.
    • When the control box is opened.
    • When a fuse is replaced.
    • When the equipment is not functioning properly.
    • When the equipment is not in use.

Maintenance
Surfaces can be cleaned with mild detergent and water. Liquid should not be allowed to enter inside the unit.

Clean spill
If accidental spillage of liquids or other materials occur on the platform, shut down the machine and clean up the liquids or other materials immediately.
Note: If liquid has entered the unit, remove the platform and clean it. ▲

  1. Turn off the machine and unplug the power cord.
  2. Remove the platform (with 4 screws).
  3. Remove top plate (with 7 screws).
  4. Use 70% ETOH to clean the inner area.
  5. All allow the unit to dry for one hour, reassemble unit, and operate normally.

Optional Accessories

Table 6-1. Universal (Aluminum) Platform.ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(12\) ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(13\)

Table 6-2. 18inch (Aluminum) Platform.ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(14\) ThermoFisher-
SCIENTIFIC-88881101-CO2-Plus-Digital-Orbital-Shaker-
\(15\)

Troubleshooting

To troubleshoot the unit, refer to the table below. If your problem persists, contact your local sales/service representative.

Error Cause Solution
Indicator off No reaction Power disconnected Connect the power
Power adapter failure Replace power adapter
Switched off Switch on
No rotating of platform Sensor failure Contact local sales/service
Main-board failure Contact local sales/service
Loud noise The unit has been placed on uneven surface Place the unit on

level and sturdy surface
Platform loose| Fasten platform
Accessories loose| Fasten accessories
Other| Keep record for maintenance

WARRANTY

THERMO FISHER SCIENTIFIC STANDARD PRODUCT WARRANTY
The Warranty Period starts two weeks from the date your equipment is shipped from our facility. This allows for shipping time so the warranty will go into effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner during the first-year warranty period.
During the first two (2) years, parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo’s expense, labor included. Installation and calibration are not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to the performance of any repairs. Expendable items, glass, filters and gaskets are excluded from this warranty. Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or equipment. At Thermo’s option, all non-conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB destination.

THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products. Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance. If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. We’re ready to answer your questions on equipment warranty, operation, maintenance, service and special applications. Outside the USA, contact your local distributor for warranty information.

In the United States:
For customer service, call 1-800-766-7000
To fax an order, use 1-800-926-1166
To order online: thermofisher.com

In Canada:
For customer service, call 1-800-234-7437
To fax an order, use 1-800-463-2996
To order online: thermofisher.ca

Find out more at thermofisher.com

For Research Use Only. Not for use in diagnostic procedures. © 2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.

Visit us online to register your warranty www.thermoscientific.com/labwarranty

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ThermoFisher SCIENTIFIC User Manuals

Related Manuals