ALPHAMED Infrared Forehead Thermometer UFR106 Instruction Manual

June 6, 2024
ALPHAMED

ALPHAMED
Instruction manual
lnfrared Forehead Thermometer
Model:UFR106

ALPHAMED Infrared Forehead Thermometer UFR106

ALPHAMED Infrared Forehead Thermometer UFR106 -

Introduction

  • Your new infrared forehead thermometer uses advanced infrared (IR) technology to measure temperature instantly and accurately on the forehead or object.

  • Easy to use and less measurement time
    This thermometer does not need to contact the body or object to ensure safety and hygiene. Its ergonomic design makes this thermometer be simple and very easy to check the temperature. It only takes 1 second to take measurement and reading.

  • Bod_y and object modes
    This thermometer supports the measurement of the temperature of the body and object.
    The measuring range of object mode is 0.0 to 100.0 F ). That means except to take body temperature, it also supports taking the surface temperature of an object as below, The surface temperature of milk in the bottle;
    The surface temperature of the baby bath;
    Environment temperature.

  • Coler visible indication of alarm and alarm sounds
    When the body temperature is over 37.5 ° C color indicator will show red color and an alarm sounds 10 times.

  • Memory function
    It stores up to 20 sets of recent measurement data.

  • Auto power off
    Automatically power off if left idle for Approx 60s.

  • Measurement
    The measurement time interval is 15s and the measurement distance is within 1-3cm.
    Please read” the manual carefully before you use the until, and keep it for future­ reference.

  • Intended use
    The infrared Thermometer is intended for the intermittent measurement and monitoring of human body temperature from forehead or object. The device is indicated for use by people of all ages at-home care and in hospitals.

Safety Information

To assure the correct use of the product, basic safety measures should always be followed including the warning and the caution listed in the instruction manual:
Symbol descriptions
The following symbols may appear in this manual, on the label,on the device, or on it’s accessories. Some of the symbols represent standards and compliances associated with the device and its use.

| WARNING: This alert identifies hazards that may cause serious personal injury or death.
---|---
| CAUTION: This alert identifies hazards that may cause minor personal injury, product damage, or property damage.
| Type BF applied part
| Manufacturer

SN

| Specifies serial number
| DISPOSAL: Do not dispose of this product as unsorted municipal waste. Collection
of such waste separately for special treatment is necessary.
| Direct current
| Follow instructions for use
| CAUTION: Consult accompanying documents

This thermometer is not intended to substitute for a consultation with your physician. The forehead scan temperature serves as a reference only. It cannot be a judgment on fever.
Basic safety precautions should always be observed, especially when the thermometer is used on or near children and disabled persons.
Please place the device unreachable by young.
Avoid direct sunlight.
Do not touch the lens.
No modification of this device is allowed.
The swallowing of small parks like packing bags, batteries, battery covers and so on may cause suffocation.
Please do not use a dilution agent, alcohol, or petrol to clean the unit. Please
treat is gentle and prevents falling from a high place.
Please do not immerse it in liquid.
Never leave the battery in the battery compartment for a long time without use, as they may leak and cause damage to the unit.
Please take off the battery if you do not intend to use it within 3 months. Replace with new batteries if the unit displays a low battery symbol.
Do not mix the old and new batteries together.
Do not use it during transportation.

WARNING:
Do not dispose of electrical appliances as unsorted municipal waste, use – separate collection facilities. Contact your local government for information regarding the collection systems available. If electrical appliances are disposed of in landfills or dumps, hazardous substances can leak into the groundwater and get into the food chain, damaging your health and well-being.

Classification

  1. Internally powered equipment;
  2. Type BF applied part;
  3. Protection against ingress of water or Particulate matter: IP21;
  4. Not category AP/APG equipment.
  5. Mode of operation: Continuous operation.

The user must check that the equipment functions safely and see that it is in proper working condition before being used.

Product Structure

Display

ALPHAMED Infrared Forehead Thermometer UFR106 -
Display

1. Body temperature indicator
2. Object temperature indicator
3. Memory mode| 4. Temperature reading
5. Temperature unit
6. Low battery indicator
---|---

Body

ALPHAMED Infrared Forehead Thermometer UFR106 - Body

1 . Auto-sense Probe
2. Display Screen
3. Display Screen
4.°C/°F Button| 4. Mode Button
6. START/STOP Button
7. Battery Cover
---|---

Battery Installation

Remove the battery cover from the battery compartment, insert the battery.

  1. Remove the battery cover.the battery compartment as the arrow direction accordingly.

  2. Insert 2 AAA powerful batteries into the compartment and ensure each battery is in the proper direction, Positive(+) and Negative(-) are displayed on the back of the battery cover.

  3. Close the battery cover.

Low battery and replacement
When power is on, the low battery symbol , will display once the unit starts to work, and you must replace it with new batteries, otherwise, the unit can’t work.
Battery type and replacement
Please use 2pcs AAA identical 1. 5V alkaline batteries. Do not use the batteries beyond their expiry date. Please remove the batteries if you do not need to use them for a long time.
WARNING
Dispose of the battery in accordance with all federal, state, and local laws. To avoid fire and explosion hazards, do not bum a incinerate the battery.

Setting Mode

How To Set

  1. Mode setting: Press the Mode button when power on after you hear 1 beep, the screen will display or , Mode will change between and .when the button is pressed will confirm the Body Mode automatically when you_switchio and confirm the Object Mode automatically when you switch to e. Then Mode setting is finished.
    ALPHAMED Infrared Forehead Thermometer UFR106 - Mode
setting Note: The Body Mode is used to measure the forehead temperature, then the Object Mode is used to measure the object temperature.

  2. Unit setting:
    Press Mode button °C/°F when power on after you hear 1 beep, the screen will display °[ or °F, Mode will change between °[ and °F .when the button “C/”F is pressed. It will confirm the Celsius Unit automatically when you switch to °F, Ari’s confirm the Fahrenheit Unit automatically when you switch to °F. Then Mode setting is finished.
    ALPHAMED Infrared Forehead Thermometer UFR106 - Mode setting
1

Proper Use Of The Unit

Pre-measurement
About Normal Body Temperature&Fever
The temperature in the forehead and temple area differs from the internal temperature, which is taken orally or rectally. Vasoconstriction, an effect that constricts the blood vessels and cools the skin, can occur during the early stages of a fever. ln this case, the temperature measured by the Infrared thermometer may be unusually low. If the measurement, therefore, does not match the patient’s own perception or is unusually low, re-peat the measurement every 15 minutes. As a reference, you can also measure the internal body temperature using a conventional oral or rectal thermometer.
Body temperature can vary from one individual/person to b next.
It also varies by location on the body and time of day. Below shows the statistical normal ranges from different sites.
Please keep in mind that temperatures measured from different sites, even at the same time, should not be directly compared. Fever indicates that the body temperature is higher than normal. This symptom may be caused by infection, overdressing, or immunization. Some people may not experience fever even when they are ill.
These include, but are not limited to, infants younger than 3 months old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin, ibuprofen, acetaminophen), or persons with certain chronic illnesses Please consult your physician when you feel ill even if you do not have a fever.
Table*1 Body Site Normal Temperature Range

Body Site Normal Temperature Range
Oral 0.6°C (1°F) or more

above or below 37°C (98.6°F)
Rectal/ear| 0.3°C to 0.6°C (0.5°F to 1°F) higher than an oral temperature
Axillary (armpit)| 0.3°C to 0.6°C (0.5°F to 1°F) lower than an oral temperature

Note: Body Temperature at WebMD;
website: http://firstaid.webmd.com/body-temperature;retrieved at 2010 Jan 7.
As a Body Thermometer

ALPHAMED Infrared Forehead Thermometer UFR106 - As a Body
Thermometer

  1. .Press the START/STOP button, all symbols appear on the display, you can hear 1 short beep, then unit ° C/°F on the display start to flash.
  2. Press the Mode button to elect Body Mode, the temperature unit flashes.
  3. Move the probe close to the forehead make sure the probe is flat and take measurements with a distance between 1-3cm. Press the START/STOP button, The meter will take measurements.
  4. The measurement will be finished in 3 seconds, when it has been completed you can learn 3short beeps in the meantime, the reading will appear on the display with green LED blinks.

Note:

  • If the readingis<37.5°C(99.5°F ), the display will show together with a green LED.
  • If the reading is37.5°C(99.5°F)and<43°C(109.4°F), the display will show together with red LED and ten short beeps.
  • Only after the unit °C/°F starts to flash again, it can continue to take a measurement.
  • As the forehead measurement temperature is likely to be affected by sweat, oil, and the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close to the forehead measurement, the reading will be affected by the surrounding temperature.
  • Babies’ skin reacts very quickly in ambient temperature. Therefore, do not take their temperature with the non-contact thermometer during/after breastfeeding, because the skin temperature is maybe lower than the internal body temperature.

As an Object Thermometer

ALPHAMED Infrared Forehead Thermometer UFR106 - As a Object
Thermometer

  1. Press the START/STOP button, all symbols appear on the display, you can hear 1 short beep, then unit ° C/ °F on the display start to flash.
  2.  Press the Mode button to select Object Mode, the temperature unit flashes.
  3. Move the probe close to the object, make sure the probe is flat and take measurements with a distance between 1-3cm. Press the START/STOP button, The meter will take measurements.
  4. The measurement will be finished in 3 seconds, when ii has been completed, you can hear 3 short beeps, in the meantime, the reading will appear on the display with green LED blinks.

Note:

  • As the object scan temperature is likely to be affected by the surroundings, the reading shall to taker ara reference only.
  • If the probe is placed at an angle close to the object scan, the reading will be affected by the surrounding temperature.

Memory-recall of measurements
This Non-contact forehead thermometer automatically stores 20 sets of measurements value, the oldest record will be replaced by the latest measurement value when more than 20 sets.
Read memory record
Press the Mode Button .f/Q to enter the memory mode when power off. Each time you press the Mode button, a number (from 1 to 20) will be displayed along with the symbol [Ml, 1 second later, the measurement will be shown, as well as subsequent measurements, can be displayed one after the other by pressing the Mode Button.f/Q each lime.
Note:
The reading number “1” is the latest measurement, “20” is the oldest measurement. When the oldest measurement is read, pressing the start/SIQ2 Button again will read the reading “1” again
ALPHAMED Infrared Forehead Thermometer UFR106 - recall of
measurements

Memory-clear Of Measurements

If you are sure that you want to remove all stored memories permanently. Long press the Mode Button for 5 seconds until CLR appears when power off, CLR will flash for 3 times to clear all the memories along with 3 short beeps.

Exceptional Situation

Symbol Correction
In Body Mode, the Measured temperature is above the Measuring range of 43 °

C/109.4  ° F it is shown on the  CD with a red LED.
| In Body Mode, the Measured temperature is below the Measuring range of 32 ° C/89.6 °F, ii is shown on the LCD with red LED
| In Object Mode, Measured temperature is above the Measuring range 100.0°C/199.9°F or environmental temperature is above the system Operating range 40 °C/104.0°F, it is shown on the LCD with red LED.
| In Object Mode, Measured temperature is below Measuring range 0.0°C/32.0°F, or environmental emperature is been low system operating range 5 °C/41.0 °F, it is shown on the LCD with red LED.
| Low battery, replace all the worn batteries with new ones.
| Thermometer system fails or affected by electric-magnetic field

Please contact the distributor if you can not solve the problem, do not disassemble the unit by yourself!

Care And Maintenance

Care for the main unit

  • Keep the unit in the storage case when not in use.
  • Clean the unit with a soft dry cloth. Do not use any abrasive or volatile cleaners.
  • Never immerse the unit or any of its’ components in water.

Maintenance

Do not clean the unit with naphtha, thinner or gasoline, etc.| Store the unit in a clean and dry location. not subject the unit to extreme hot or cold temperature, humidity, and direct sunlight.
---|---
|
Remove the batteries if the unit will not be used in 3 months or longer.| Do not use the unit under the influence of electromagnetic interference (nearby cellphones, microwave, etc.)
|

Note: We will not be responsible for any quality problem if you do not care and maintain the product as instructed.

Specification

Description noncontact traced Eareheadihecmonuttec- _
Display LCD digital display
Measuring localization Forehead and object surface
Measurement range Body mode 32.0°C- 43.0°C (89.6°F~109.4°F);

Object mode 0.0~100.0°C (32.0°F-199.9°F);
Temperature unit| °C/°F
display resolution| 0.1°C/°F
Accuracy| ±0.2°C±0.4°F
(within 36.00-39.0°C/ 96BI-102.2°F)
Memory function| 20 sets memory of measurement values
Beep alarm| One short beep when power on and start measurement
One long beep when measurement reading is below 37.5C199.5°F. 10 short beeps when measurement reading is g more real than or equal to 37.5Y/99.5°F
3 short beeps when the system fails
LED color indicator| Green: Temperature < 37.5.099.5°F Red: Temperature-37.5°C
Power source| 2pcSAAA alkaline battery
Automatic power-off| In the 60s
Main unit weight| Approx. Mg (batteries not included)
Main unit size| L151 mm x W 53mm x H41mm
Battery life| Could be used 300 times for the normal condition
-ACCES.SoWeS”| Instruction manual
Operating environment| Body mode 10.0°C-40.0 °C150.0°F -104.0 °F Object mode 5C-40.0.041.0V -104.0V
Storage environment| Temperature: -20°C-4.50.0-4.0V-+122.0°F Humidity: 15% -95% avoid crash. sun bum or rain during transportation

Clinical Measurement Accuracy And Safety Verification:
The product has passed clinical trials. The measured results of the infrared forehead thermometer were compared with the measured results of mercury thermometers, the deviation average.lib = 0.011 °C not exceeding 0.3°C; the clinical repeatability of the infrared forehead thermometer Δtv =0.100 ° C, not exceeding 0.3 ° C. The measured results p to the laboratory standard and the clinical standard. Therefore, the deviation average and the clinical repeatability of the infrared forehead thermometer comply with the regulatory requirement ISO 80601-2-56. The conclusions are drawn from the clinical trials, the accuracy and safety comply with the regulatory requirement.

Warranty Information

  • The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the date of purchase.

  • For repair under this warranty. Our authorized service agent must be advised of the fault within the period of the warranty. This warranty only covers parts and labor service under normal operations. Any defect resulting from natural causes, eg. flood, hurricane, etc, are not covered in this guarantee.
    This guarantee also does not cover damage incurred by use of the unit notion accordance with the instructions, accidental damage, or being tampered with or serviced by unauthorized service agents.

  • The following will be excluded from this warranty if the thermometer has been misused, abused, or neglect in following the manual’s instructions on purpose and unauthorized repair or modifications.

  • The device requires no calibration.

  • The device is not repairable and contains no user-serviceable parts.

EMC

Table 1

Guidance and manufacturer’s declaration-electromagnetic emissions

The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared forehead thermometer should assure that it is used in such an environment
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The Infrared forehead thermometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely b cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class 181| The Infrared forehead thermometer is suitable for use in all
establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2| Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3| Complies

Table2

Guidance and manufacturer’s declaration-electromagnetic emissions

The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared forehead thermometer should assure that it is used in such an environment
Immunity Test| IEC 60601
Test level| Compliance
level| Electromagnetic
environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2| ±8 kV contact ±15 KV air| ±8 kV contact ±15 KV air| Floors should be wood, concrete, or learning tile. If floors are covered with synthetic material, the relative humidity should be at at least 30 %
Electrical fast transient/burst IEC 610004-4| Power supply lines: t.2 kV input/output lines: ±1 kV| Power supply lines: ±2 kV input/output lines: ±1 kV| The main power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| line(s) to line(s): t1 kV. line(s) to earth: ±2 kV. 100 kHz repetition frequency| line(s) to line(s): ±1 kV.
line(s) to earth: ±2 kV. 100 kHz repetition frequency| The main power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply input lines IEC 61000-4-11| 0% 0.5 cycle At 0°, 45 °,
90 0,135 0, 180 °, 225 0,
2700 and 315 ° 0% 1 cycle And 70% 25/30 cycles
Single-phase: at 0 0% 300 cycle| 0% 0.5 cycle At 0°, 45 °, 90 0,135 °, 180 °, 225 °, 270° and 315 ° 0% 1 cycle And 70% 25/30 cycles Single phase: at 0 0% 300 cycle| The main power quality should be that of a typical commercial or hospital environment.
Power frequency (50/60Hz) magnetic field
IEC 61000-4-8| 30 A/m 50Hz/60Hz| 30 A/m 50Hz/60Hz| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
NOTE UT is the a.c. mains voltage prior to application of the test level.

Table 3

Guidance and manufacturer’s declaration-electromagnetic emissions

The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared forehead thermometer should assure that it is used in such an environment
Immunity Test| ICE 60601 Test level| Compliance level| Electromagnetic environment – guidance
Conducted RF
IEC61000- 4-6| 150KHz to 80MHz:
3Vmis 6Vrms (in
ISM and amateur
radio bands)
80% Am at 1kHz| 150KHz to 80MHz:
3Vmis 6Vans (in
ISM and amateur
radio bands)
80% Arm at 1kHz| Portable and mobile RF communications equipment should be used no closer to any part of the Infrared forehead thermometer, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: d=0.35VP;
d=1.2 VP;
Radiated RF
IEC61000- 4-3| 10V/m, 80% Am at 1kHz| 10V/m, 80% Am at 1kHz| 80MHz to 800MHz: d=1,2,/p;
800MHz to 2.7GHz:
d=2.3VP;| Where, P is the maximum output power rating of the transmitter in watts (W)according to the transmitter manufacturer, d is the recommended separation distance in meters (m) Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency
range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the h higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM, and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Infrared forehead thermometer is used exceeds the applicable RF compliance level above, the Infrared forehead thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Infrared forehead thermometer.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Table 4

Recommended separation distances between portable and mobile RF communications equipment and the

The infrared forehead thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the infrared forehead thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the infrared forehead thermometer as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
the output power of the transmitter
W| Separation distance according to the frequency of the transmitter
m
150 kHz to 80 MHz
d=3.5V-P.| 80MHz to 800MHz
d=1.2145;| 800MHz to 2.7GHz
d=2.3 VP;
0,01| /| 0.12| 0.23
0,1| /| 0.38| 0.73
1| /| 1.| 2.
10| /| 4.| 7.
100| /| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Table 5

Guidance and manufacture’s declaration electromagnetic emissions

The Infrared forehead thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared forehead thermometer should assure that it is used in such an environment
Rad61ia0 ted 00-RF4-3 IEC
(Test
specifications for
ENCLOSURE
PORT
IMMUNITY to
RF Wireless
communications
equipment)| Test
Frequency(MHz)| Banda)
(MHz)| *Sends a)| Modulo
n b)| Modulation b)
(W)| Distance
(m)| IMMUNITY TEST
LEVEL (V/m)**
385| 380-390| TETRA 400| Pulse
modulation b)
18 Hz| 1,8| 0,3| 27
450| 380-390| GMRS 460,
Fits 460| FM c)± 5IcHz
deviation 1 kHz
sine| 2| 0,3| 28
710| 704- 787| LTE Band 13,
17| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
745
780
810| 800- 960| GSM 8001900,
TETRA 800′ iDE N 820′ CDMA eso, LTE Band 5| Pulse modulation o) 18 Hz| 2| 0,3| 28
870
930
1720| 1700-1990| GSM 1800; CDMA 1990; GSM 1900;
DECT; LTE Band 1, 3, 4, 25; UMTS| Pulse
modulation  b)
217 Hz| 2| 0,3| 28
1845
1970
2450| 2400 -2570| Bluetooth, MAN,
802.11 b/gln,RFID 2450,
LTE Band 7| Pulse
modulation b)
217 Hz| 2| 0,3| 28
5240| 5100-5800| WlAN 80211
a/n| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
5240
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b)-The carrier shall be moderated using-a a 50-% duty cycle square-wave signal.
c)As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be the worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher  IMMUNITY TEST LEVEL .S shall be calculated using the following equation:
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVEL .S shall be calculated using the following equation:
E=6/a√P
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.

Manufacturer
Shenzhen Urion Technology Co., Ltd.
Floor 4-61h of Building D, Jiale Science&Technology Industrial
Zone, No.3, ChuangWei Road, Heshuikou Community, Marian Street,
GuangMing New District,518106 ShenZhen, PEOPLE’S REPUBLIC OF CHINA
Tel: (86) -755-29231308 E-mail: [email protected]
MADE IN CHINA
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg,Germany
Tel:+49-40-2513175 E-mail:[email protected]
ALPHAMED Infrared Forehead Thermometer UFR106 - ceREV:1.1

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ALPHAMED User Manuals

Related Manuals