Dwyer RP3 Wireless Thermo-Hygrometer Probe Instruction Manual
- June 6, 2024
- Dwyer
Table of Contents
Dwyer RP3 Wireless Thermo-Hygrometer Probe
The Model RP3 Wireless Thermo-Hygrometer Probe measures relative humidity, temperature, dew point, and wet bulb temperatures when combined with the UHH4 or newer universal handheld device, or your own mobile device. In order to protect the sensor in higher air flow ducts, each probe comes with a membrane filter. Factory calibrated replacement humidity sensors can plug directly into the probe without requiring additional calibration. Wireless probes can take measurements up to 50 feet away from the base unit. A bicolor LED flashes on the handle of the probe to indicate communication status with the base unit. The battery is rechargeable via the mini-USB connector on the bottom of the probe.
NOTICE: This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions:
- This device may not cause harmful interference. and
- this device must accept any interference received, including interference that may cause undesired operation.
This Class B digital apparatus complies with Canadian ICES-003.
CHARGING BATTERY
NOTICE : It is required prior to the initial usage to charge the battery for 12 hours.
When the charge of the battery is almost used up, the LED on the handle of the wireless probe will turn solid red. There is approximately 5 minutes of battery life left at this point. If the probe is not already paired to the UHH, it will not pair while in low battery condition.
Step 1: Open the USB cover on the bottom of the probe handle.
Step 2: Plug the mini-USB connector end of the cable into the probe
handle.
Step 3: Plug the USB connector end of the cable into the port on the
charger or PC.
Step 4 : Plug the charger into either an electrical outlet or car
charging port (LED on the charger and the handle should both light up).
Step 5 : The LED on the handle will turn off when fully charged.
Step 6 : Remove charger from electrical outlet.
Step 7: Remove USB connectors from the handle and the charger.
Step 8 : Replace the USB cover on the bottom of the probe handle.
Dimensions
SPECIFICATIONS
Service : Clean air.
Temperature Limits: Process: -4 to 140°F (-20 to 60°C); Ambient: 5 to
125°F (-15 to 51°C); Battery charging: 32 to 113°F (0 to 45°C).
Range : RH: 0 to 100% (non-condensing); Temperature: -22 to 140°F (-30 to
60°C).
Accuracy : RH: ±2% @ 25°C (10 to 90% RH); ±4% (0 to 10, 90 to 100% RH);
Temperature : ±0.54°F @ 77°F (±0.3°C @ 25°C).
Response Time : 1.5 s.
Probe Length: 8˝ (203 mm) insertion.
Power Requirements: 3.7 V YT562447 lithium ion battery, installed
functional, user replaceable.
Maximum Wireless Distance : 50´ (15 m).
Handle Enclosure: Thermoplastic elastomer over polycarbonate.
Supplied With: Wrist strap.
Weight: 11.2 oz (331.22 g).
Supported Devices: UHH4 or newer universal handheld, and smart devices
that support Bluetooth® LE (Bluetooth® Low Energy).
Supported Operating Systems: Android® firmware version 3.X or later, iOS®
firmware version 5.X or later.
WARNING : Lithium ion polymer batteries are very volatile and can cause a fire if punctured or severely damaged. Only use a Dwyer Instruments, Inc. approved charging device in a well ventilated area away from any flammable materials or gases. Do not incinerate the battery. Only charge between 32 to 113°F (0 to 45°C).
USING MOBILE METER® APP
To use the wireless probe with a personal device not supplied by Dwyer
Instruments, Inc. you must first download the Mobile Meter® app. Once
installed, the wireless probes can be connected by following the directions
below.
USING WIRELESS PROBE
To power the probe on press the button on the handle once. The LED on the probe will change according to the below key.
Flashing Red – Probe is in pairing mode.
Flashing Green – Probe is connected to a mobile device.
To power the probe off press and hold the button on the handle until the LED
turns off.
CONNECTING TO WIRELESS PROBES
- Verify that the universal handheld, UHH4 or newer, or device has Bluetooth® technology turned on and is within range. Open the Mobile Meter® app by clicking on the icon.
- Press “Without Mobile Gateway”.
- Power on wireless probe.
- Once discovered, the probe name will appear on the device screen.
- Select the check box next to the desired probe name(s).
- Press the back arrow to return to probe list, selected probes will connect automatically.
Notice : It is recommended that only four probes be selected due to the bandwidth limitations.
FCC/INDUSTRY CANADA NOTICE
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular use case. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between equipment and receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
NOTICE : This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions:
- This device may not cause harmful interference;
- This device must accept any interference received, including interference that may cause undesired operation of the device.
CAUTION : Pursuant to FCC 15.21 of the FCC rules, changes or modifications not expressly approved by Dwyer Instruments, Inc. may void the user’s authority to operate the equipment.
RF NOTICE
This product complies with FCC OED Bulletin 65 and Industry Canada’s RSS-102
radiation exposure limits set forth for an uncontrolled environment. This
Class B digital apparatus complies with Canadian ICES-003.
CAUTION : The antenna used for this transmitter must maintain a
separation of at least 20 cm from all persons and must not be co-located or
operating in conjunction with any other antenna or transmitter.
INFORMATION TO THE USER
- Power Output: 6 mW
- Operating Frequency: 2.4 GHz
- Operating Mode: IEEE 802.15.4, BLE, Direct Sequence Spread Spectrum
- Data Rate: Up to 250 kbps
- Intended Use: Industrial/commercial HVAC
- Antenna Connection: Internal only, non-tunable
BATTERY REMOVAL
If, for some reason, the wireless probe needs to be returned to Dwyer Instruments, Inc. for maintenance or repair, the rechargeable lithium ion battery needs to be removed prior to shipping the unit. Before attempting to remove the battery ensure the probe has been powered down. To remove the battery, set the wireless probe face down on a non-abrasive surface. There are four Phillips head screws that secure the two halves of the handle together. Two are located under the serial number label at the top of the probe and two are located at the bottom of the unit. When lifting the ends of the serial label to access the screws be careful to not damage the label such that the serial number cannot be read. Remove the four screws with a suitable Phillips screw driver and set aside the back half of the housing. Remove the battery by grasping the black plastic connector on the battery wire harness and pull straight out while securing the black plastic connector on the circuit board with the other hand. Be careful to not puncture the battery as this may cause a fire. Remove the battery and set aside. Securely replace the four screws to hold the handle of the probe together. The unit may now be packaged for shipping back to Dwyer Instruments, Inc.
MAINTENANCE/REPAIR
Upon final installation of the Model RP3, no routine maintenance is required.
The Model RP3 is not field serviceable and should be returned if repair is
needed. Field repair should not be attempted and may void warranty. This
symbol indicates waste electrical products should not be disposed of with
household waste. Please recycle where facilities exist. Check with your Local
Authority or retailer for recycling advice.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem, plus any additional application notes.
Android® is a registered trademark of Google, Inc. Bluetooth® is a registered trademark of Bluetooth SIG, Inc. iOS® is a registered trademark of Cisco Systems, Inc.
Phone:219-879-8000
Fax:219-872-9057
www.dwyer-inst.com
e-mail: info@dwyermail.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>