Tzone TempU06 Temp Data Logger User Manual
- June 6, 2024
- Tzone
Table of Contents
Temp Data Logger User Manual
—TempU06 Series
Model:
TempU06
TempU06 L60
TempU06 L100
TempU06 L200
- *External Temperature Probe
- Back Splint
- USB Interface
- LCD Screen
- Stop Button
- Start/View/Mark Button
- Please note the Model TempU06 is with built-in temperature sensor, it do not have external temperature probe
LCD Display Instruction
1| OK
Alarm
| 8| Bluetooth*
---|---|---|---
2| ►Start recording
■ Stop recording
| 9| Flight mode
3&14| Alarm zones
↑,H1, H2 (High) ↓, L1, L2 (Low)
| 10| Bluetooth communicating
4| Delay recording| 11| Unit
5| Password (AccessKey) protected| 12| Reading
6| Stop button disabled| 13| Data cover
7| Remaining battery level| 15| Statistics
- Please note the Model TempU06 do not have bluetooth function
Product Introduction
The TempU06 series are mainly used to monitor and record temperature data of vaccines, pharmaceuticals, foods, and other products during storage and transportation. The Bluetooth connectivity of the TempU06 series and Temp Logger App brings customers the advantages of tracking data for data viewing. And you can enable quick connection with PC to get data by Temperature Management Software, no cable or reader required for downloading data.
Feature
-
Bluetooth and USB connection. Dual interface ensures convenience and stability*
-
Large LCD screen with powerful indicators
-
External temperature probe for low temperature condition, up to -200°C*
-
Flight mode for air transportation*
-
FDA 21 CFR Part 11, CE, EN12830, RoHS, NIST traceable calibration
-
No need any software to get the PDF and CSV file
-
Please note the Model TempU06 do not have bluetooth function or flight mode
-
For the temperature range, please refer to the datasheet
LCD Screens
Home Screens
1 Initialization 2 Above upper & lower limit
3 Log interface 4 Mark interface
5 Max Temp interface 6 Min Temp interface
Error Screens
If there E001 or E002 on the screen, please check
- If the sensor is not connected or broken
- If over the temperature detect range
Download Report Screen
Connect data logger to USB port, it will automatically generate reports.
Connecting to USB
How to use
a.Start recording
Press and hold the left button for more than 3s until the led light turns
green, and the “►” or “WAIT” displays on the screen, which indicates the
logger is started.
(For the model with external temperature probe, please make sure the sensor is
completely inserted into the device.)
b.Mark
When the device is recording, press and hold the left button for more than 3s, and the screen will switch to the “MARK” interface. The number of “MARK” will increase by one, indicating data was marked successfully.
c.Stop recording
Press and hold the right button for more than 3s until the led light turns red, and the “■” displays on the screen, indicating stop recording successfully.
d.Turn on/off Bluetooth
Press and hold the two buttons for more than 3s at the same time, until the
red light flashes quickly, and the “ ” displays on the screen or disappears,
indicating the bluetooth was turn on or off.
(when the device in flight mode, press and hold the two buttons for more than
3s will quit the flight mode)
e.Get report
After recording, there are two ways to get report: connect the device with USB port of PC or using Temp Logger App on smart phone, it will automatically generate PDF and CSV report.
Configure Device
**Configure device via App***
Please scan this QR code to download the app.
Configure device via temperature management software
Please download the temperature management software from : http://www.tzonedigital.com/d/TM.zip
- Please note the Model TempU06 do not have bluetooth function
Battery Status Indication
Battery | Capacity |
---|---|
Full | |
Good | |
Medium | |
Low (Please replace battery) |
Battery replacement
a.Remove rear cover
I . Pull out the external sensor
II. Remove the screw
b.Replace the rear cover
III . Take out the rear cover
IV. Replace the battery
V. Replace the rear cover
- Put old batteries in special sorting bins
Cautions
- Please read the manual carefully before using the logger.
- When the logger is recording, do not move external temperature probe, otherwise may get error data.
- Do not bend or press the end of the external temperature probe, as this may damage it.
- Please recycle or dispose the data logger in accordance with local laws and regulations.
TZ-TempU06 datasheet
Tzone TempU06 Temperature Data Logger Suite
Industry leading temperature data logger suite provides different kinds of temperature range devices to provide a complete temperature recording solution.
Model| TempU06
| TempU06 L60
| TempU06 L100| TempU06 L200
| Technical Information
Dimension| 115mm50mm20mm
Sensor Type| Build in temp sensor| External temp sensor
Battery Life| Typically 1.5 years| Typically 1 year
Bluetooth| Not support| Support
Weight(s)| 100g| 120g
Connectivity| USB 2.0| USB 2.0 and Bluetooth 4.2
Detecting Temperature Range| -80°C~+70°C| -60°C~+120°C| -100°C~+80°C|
-200°C~+80°C
Temperature Accuracy| ±0.5°C| ±0.3°C (-20°C~+40°C)
±0.5°C (-40°C~-20°C/+40°C~+60°C)
±1.0°C (-80°C~-40°C)
| ±0.5°C
Temperature Resolution| 0.1°C
Data Storage Capacity| 32000
Start Mode| Push-To-Start or Timing Start
Logging Interval| Programmable (10s ~ 18h) [Default:10mins]
Alarm Range| Programmable [Default: <2°C or >8°C]
Alarm Delay| Programmable (0 ~ 960mins) [Default: 10mins]
Report Generation| Automatic PDF/CSV Report Generation
Software| Temp (RH) Management Software
(For Windows,21 CFR 11 Compliant)
|
Temp Logger APP
Temp (RH) Management Software (For Windows,21 CFR 11 Compliant)
Protection Grade| IP65