Celltrion DiaTrust™ COVID-19 Ag Home Test
Healthcare Provider Instructions for Use
For use under the Emergency Use Authorization (EUA) only
For in vitro diagnostic use
[INTENDED USE]
Celltrion DiaTrust™ COVID-19 Ag Home Test is a lateral flow immunoassay
intended for the qualitative detection of nucleocapsid protein and receptor-
binding domain (RBD) of the SARS-CoV-2 spike proteins in mid-turbinate swabs
from the SARS-CoV-2. This test is authorized for non-prescription home use
with self-collected and adult-collected direct mid-turbinate swab samples from
individuals aged 14 years or older with symptoms of COVID-19 within the first
seven days of symptom onset.
This test is also authorized for non-prescription home use with self-collected
and adult-collected midturbinate swab samples from individuals aged 14 years
or older with or without symptoms or other epidemiological reasons to suspect
COVID-19 when tested twice over three days with at least 24 hours (and no more
than 48 hours) between tests.
The Celltrion DiaTrust™ COVID-19 Ag Home Test does not differentiate between
SARS-CoV and SARSCoV-2.
Results are for the identification of SARS-CoV-2 nucleocapsid protein antigen
and/or receptor-binding domain (RBD). These antigens are generally detectable
in mid-turbinate swabs during the acute phase of infection. Positive results
indicate the presence of viral antigens, but clinical correlation with a past
medical history and other diagnostic information is necessary to determine
infection status. Positive results do not rule out bacterial infection or co-
infection with other viruses. The agent detected may not be the definite cause
of the disease. Individuals who test positive with the Celltrion DiaTrust™
COVID19 Ag Home Test should self-isolate and seek to follow up care with their
physician or healthcare provider as additional testing and public health
reporting may be necessary.
Negative results are presumptive, do not rule out SARS-CoV-2 infection, and
should not be used as the sole basis for treatment or patient management
decisions, including infection control decisions. Negative results should be
considered in the context of an individual’s recent exposures, history, and
the presence of clinical signs and symptoms consistent with COVID-19 and
confirmed with a molecular assay, if necessary, for patient management.
For serial testing programs, additional confirmatory testing with a molecular
test for negative results may be necessary, if there is a high likelihood of
SARS-CoV-2 infection, such as an individual with close contact with COVID-19
or with suspected exposure to COVID-19 or in communities with a high
prevalence of infection. Additional confirmatory testing with a molecular test
for positive results may also be necessary if there is a low likelihood of
SARS-CoV-2 infection, such as in individuals without known exposures to
COVID-19 or residing in communities with a low prevalence of infection.
Individuals who test negative and continue to experience COVID-19 like
symptoms of fever, cough, and/or shortness of breath may still have SARS-CoV-2
infection and should seek to follow-up care with their physician or healthcare
provider.
Individuals should provide all results obtained with this product to their
healthcare provider for public health reporting. All healthcare providers will
report all test results they receive from individuals who use the authorized
product to relevant public health authorities in accordance with local, state,
and federal requirements, using appropriate LOINC and SNOMED codes, as defined
by the Laboratory In Vitro Diagnostics (LIVD) Test Code Mapping for SARS-CoV-2
Tests provided by CDC.
The Celltrion DiaTrust™ COVID-19 Ag Home Test is authorized for non-
prescription self-use or a lay user testing another person 14 years or older
in a non-laboratory setting. The Celltrion DiaTrust™ COVID-19 Ag Home Test is
only for use under the Food and Drug Administration’s Emergency Use
Authorization.
[SUMMARY AND EXPLANATION]
A coronavirus is a group of viruses that belongs to the Family Coronaviridae;
a type of RNA virus of 27 32 kb commonly found in birds and mammals including
humans. Coronavirus is divided into four genera: alpha, beta, gamma, and
delta. The virus causes illnesses ranging from the common cold to more severe
diseases such as Middle East Respiratory Syndrome (MERS-CoV) and Severe Acute
Respiratory Syndrome (SARS-CoV).
Coronavirus disease 2019 (COVID-19) is a new strain caused by severe acute
respiratory syndrome coronavirus 2 (SARS-CoV-2). The disease originated in
Wuhan city of China in December 2019. The World Health Organization (WHO)
publicly named this virus ‘COVID-19’ and declared it a pandemic and a Public
Health Emergency of International Concern. The infection is typically spread
from one person to another via direct contact or respiratory droplets from
cough or sneeze. The latent period from exposure to onset of symptoms is
between one to fourteen days (four to seven days on average). Common symptoms
and signs of infection include fever, cough, shortness of breath, and
breathing difficulties. In severe cases, infections can cause pneumonia,
severe acute respiratory syndrome, kidney failure, and even death.
Due to the wide variety of symptoms, it is difficult to differentiate COVID-19
from other existing respiratory viruses or bacteria. Diagnosing COVID-19
through isolating the virus or detecting specific genes from the collected
respiratory droplet specimens is a challenge in terms of time and
accessibility as it requires long hours, a well-equipped laboratory, and
advanced technology which are often not available to many public. The test is
designed to detect antigen to SARS-CoV-2, and it will help assess if an
individual has COVID-19 antigen within 15 minutes r.
[TEST PRINCIPLE]
The Celltrion DiaTrust™ COVID-19 Ag Home Test is a lateral flow immunoassay
test. The Celltrion DiaTrust TM COVID-19 Ag Home Test is designed to detect
antigens from the SARS-CoV-2 from direct mid-turbinate swab samples from
individuals with or without symptoms or other epidemiological reasons to
suspect COVID-19 when tested twice over two or three days with at least 24
hours but not more than 48 hours between tests. This test is also authorized
to detect antigens from the SARS-CoV2 from direct mid-turbinate swab samples
from individuals aged 14 years or older with symptoms of COVID-19 within the
first seven days of symptom onset. This test is authorized for non-
prescription home use with mid-turbinate nasal swab specimens from individuals
aged 14 years and older. The Celltrion DiaTrust™ COVID-19 Ag Home Test is
validated for use from direct specimens testing without transport media.
A nitrocellulose membrane strip in the device has a test line and a control
line, wherein the test line is pre-coated with anti-mouse monoclonal antibody
to SARS-CoV-2 to detect SARS-CoV-2 nucleocapsid and RBDs from the SARS-CoV-2
spike proteins, and the control line is coated with goat anti-mouse IgG. When
the extracted swab specimen is dispensed into the sample well, the specimen
migrates towards the conjugate pad, which contains conjugated antibodies with
colloidal gold directed against the SARS-CoV-2 antigen. When the sample
contains SARS-CoV-2 antigens, an antigen-antibody conjugate complex is formed.
The sample-conjugate complex then passes over the membrane until it reaches
the capture zone (test line). Here, the complex is bound to immobilized
antibodies and form a visible colored band in the test line. The sample then
migrates across the membrane along the strip until it reaches the control line
where excess conjugate binds and produces a second visible line on the
membrane. This control line indicates that the sample has migrated across the
membrane as intended and indicates that the test was correctly performed. This
test does not use biotin-Streptavidin/avidin chemistry in any of the steps for
coupling reagents.
[MATERIALS SUPPLIED]
Kit components
| Quantity
---|---
1 Test Kit| 2 Tests Kit| 5 Tests Kit| 25 Tests Kit
Test cassette with a test strip| 1 ea/box| 2 ea/box| 5 ea/box| 25 ea/box
Extraction buffer (0.3 mL / test tube)1| 1 ea/box| 2 ea/box| 5 ea/box| 25
ea/box
Filter cap| 1 ea/box| 2 ea/box| 5 ea/box| 25 ea/box
Swab| 1 ea/box| 2 ea/box| 5 ea/box| 25 ea/box
Instructions for Use| 1 ea/box| 1 ea/box| 1 ea/box| 1 ea/box
¹Extraction buffer is provided in the sealed test tube.
[MATERIALS REQUIRED BUT NOT PROVIDED]
- Celltrion DiaTrust™ COVID-19 Ag Home Test Application (Celltrion SafeKey)
- Smartphone for using App, Celltrion SafeKey (Android 10 or newer, iOS 14.2 or newer)
- A compatible computer for web-based App (https://celltrion.safekey.tools)
[PRECAUTIONS AND WARNINGS]
- For in vitro diagnostic use only
- This product has not been FDA cleared or approved but has been authorized by FDA under an Emergency Use Authorization (EUA).
- The emergency use of this product is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostics for the detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food, Drug, and Cosmetic Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner.
- All the results within the United States and its territories are required to be reported to the appropriate public health authorities.
- This product has been authorized only for the detection of proteins from SARS-CoV-2, not for any other viruses or pathogens.
- Do not use the test device beyond the expiration date.
- Keep sealed until usage, and once opened use immediately.
- Test samples immediately after collection.
- Do not use the test device if the pouch is damaged or open.
- Do not re-use the device.
- This test is intended for the diagnosis of coronavirus infection by detecting COVID-19 antigen, but should not be used as a sole criterion for the determination of SARS-CoV-2 infection. Other laboratory tests and clinical information (signs and symptoms) should be used and considered for diagnosis.
- Inadequate or inappropriate sample collection may yield false test results.
- To obtain accurate results, the test must be performed as indicated in this Instructions for Use
- Results should be read within 15 minutes. If the test is read before 15 minutes or after 20 minutes, false-negative or false-positive results may occur.
- Inadequate or improper nasal swab sample collection may result in false-negative test results.
- Do not touch the swab head when handling the swab.
- Do not ingest
- Keep out of reach of children
- Avoid contact with skin and eyes
- If contact with the body occurs, rinse with water. If irritation persists, seek medical advice.
- Discard Celltrion DiaTrust™ COVID-19 Ag Home Test in accordance with local, state, and federal regulations or accreditation requirements.
[LIMITATIONS]
- Do not use this test for individuals under 14 years of age. The swab included in the kit is designed for the collection of samples from adults and additional safety measures are needed for safe collection in children under 14 years of age.
- Testing for asymptomatic individuals should be performed at least twice over three days, with at least twenty-four hours and no more than 48 hours between tests. You may need to purchase additional tests to perform this serial (repeat) testing.
- There is a higher chance of false-negative results with antigen tests than with laboratory-based molecular tests. This means that there is a higher chance this test will give you a negative result when you have a COVID-19.
- Serial testing (i.e., testing every day or every other day) is more likely to detect COVID-19, especially when you do not have any symptoms
- This test detects both viable (live) and non-viable SARS-CoV-2. Test performance depends on the amount of virus (antigens) in the sample and may or may not correlate with viral culture results performed on the same sample.
- A negative test result may occur if the level of antigens in a sample is below the detection limit of the test or if the sample was collected or transported improperly.
- Failure to follow the Test Procedure may adversely affect test performance and/or invalidate the test result.
- Test results must be evaluated in conjunction with other clinical data available to the physician.
- Positive test results do not rule out co-infections with other pathogens.
- Negative test results are not intended to rule in other non-SARS-CoV-2 viral or bacterial infections.
- Negative results should be treated as presumptive and confirmed with an FDA-authorized molecular assay, if necessary, for clinical management.
- Performance of nasal swabs collected from patients without symptoms or other epidemiological reasons to suspect COVID-19 infection or for serial screening, when tested twice over two to three days with at least 24 but not more than 48 hours between tests has not yet been determined; a study to support use will be completed.
- If the differentiation of specific coronaviruses and strains is needed, additional testing, in consultation with state or local public health departments, is required.
- The amount of antigen in a sample may decrease as the duration of illness increases. Specimens collected after seven days are more likely to be negative compared to RT-PCR.
- The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between March and July of 2021. The clinical performance has not been established in all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation. Performance at the time of testing may vary depending on the variants circulating, including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time.
- The Celltrion DiaTrust™ COVID-19 Ag Home Test does not differentiate between SARS-CoV and SARS-CoV-2. CHEMICAL HAZARD AND
SAFETY INFORMATION
Hazardous ingredients for the extraction buffer
Chemical Name (CAS)| Material Safety Data Sheet| GHS Code for
each ingredient| Conc.
---|---|---|---
Sodium Azide (26628-22-8)| Material Safety
Data
Sheet|
Acute Tox.2 (oral), H300
Acute Tox.1 (dermal), H310
|
0.09%
The extraction buffer solution in the extraction buffer tube contains a
hazardous ingredient as shown in the above table. If the extraction buffer
solution contacts the skin or eye, immediately wash with plenty of running
water. In case the irritation persists, please seek medical advice at:
https://www.poison.org/contact-us or
1-800-222-1222.
[STORAGE AND STABILITY]
An unopened test device should be stored at 2 – 30°C (36 – 86°F). It is stable until the expiration date is marked on the label. Do not open the aluminum pouch until you are ready to use the test device. Use the test device immediately after opening the aluminum pouch. If the tests were refrigerated, keep them at room temperature for 30 minutes prior to use.
[QUALITY CONTROL]
A procedural internal control is built in the ‘control line (c)’ of the device and is used to ensure that the applied specimen has migrated well into the device. It is coated with goat anti-mouse IgG and a red-colored line will always appear when the test is performed properly. External run controls are not required to use the Celltrion DiaTrust™ COVID-19 Ag Home Test.
[TEST PROCEDURE]
When opening the test device, download the mobile application (Celltrion SafeKey) using a QR code from Instructions for Use provided with the test kit and follow the instructions as described in the mobile application.
- Test Preparation
Following the instruction in the mobile application, when you are ready to proceed with the test, tear open the two aluminum pouches.
- Prepare the aluminum pouch containing the test device and place it on the testing surface along with the reagents from the second aluminum pouch – test tube filled with the extraction buffer and filter cap. In case the tests were refrigerated, keep them ambient for 30 minutes to let them reach room temperature.
- Testing should be completed within 30-60 minutes of opening the pouch.
- Remove the test device, test tube, and filter cap from the aluminum pouches and place ita a flat surface just prior to starting the test.
- Scan the QR code on the test device through your mobile phone camera. If you are have difficulty scanning the QR on the test device, you may type the serial number into the input box below. The serial number is printed on the test device.
- Fill out the requested personal information and symptoms about the person who will test.
- Specimen collection (CDC guideline):
Use only the swabs provided with the test kit (FA/FANAB01 and Miraclean Technology, Item No. 96000) for specimen collection following the instruction on the mobile application.
- Make sure extraction buffer tube and filter cap are also readily available before starting sample collection, as the collected swab sample must be immediately inserted into the extraction buffer tube for sample extraction. After swabbing, immediately insert the swab into the extraction buffer tube. Do not leave the sampled swab dry in the open air as it may result in incorrect test results.
- Look for the “PEEL HERE” sign to peel open the swab package halfway. Make
sure the soft tip is still covered with the packaging. Identify the breakpoint
on the swab and break off the handle. Remove the swab from the package. Do not
touch the soft tip or lay it down on any surfaces. You
will see two notches on the handle. Make sure to hold the swab at the second notch.
- Insert the entire soft end of the swab straight back into your nostril less
than one inch (about 2cm) or until resistance is felt. Slowly rotate the swab,
gently rubbing it along the insides of your nasal passage several times.
Gently remove the swab. Using the same swab, repeat this process in your other
nostril with the same end of the swab.
Note: The swab included in the kit is designed for the collection of samples from adults. Do not collect swabs from children under 14 years of age.
- Test method
- Put the tip of the swab into the test tube. Move the swab up and down at least 10 times to ensure a sufficient sample is extracted.
- Remove the swab while pressing against the sides of the tube to ensure the maximum amount of liquid has been squeezed from the swab.
- Avoid swabbing and inserting an excessive amount of mid-turbinate nasal specimen into the test tube, as it may block the filter cap when dispensing sample extracts.
Note: False-negative results can occur if the specimen is not properly mixed or too vigorously mixed.
- Put the filter cap on the opening of the test tube and immediately dispense
three drops of sample extracts (100 μL) into the sample well of the device.
– Collection of buffer fluid –
– Dispensation of three drops into sample well –
- If you have dropped the test device after the sample application, please discard the test device and restart the test using a new test device.
Note: Adding only one drop of solution or the entire vial may result in false-negative results.
- Click the “Completed” button and the 15-minute timer will start on the mobile application.
- Read results 15 minutes after applying the sample. Do not read results
after 20 minutes. Click the “YES” or “NO” button on the application for the
presence of red-colored lines in the device window next to each of the two
letters, C (Control) and T (Test). Follow the instructions based on
your test result.
Note: False-negative or false-positive results can occur if results are read before 15 minutes or after 20 minutes.
- You will be instructed to take a picture of the test device. Please note that the photos are not processed and have no effect on reported results. They are collected for future reference only.
- Dispose the remainder of the test in general waste.
[INTERPRETATION OF RESULTS]
– Negative result: If no red-colored line appears in the test line (T) and a
red-colored line is present
on the control region (C), then the result is negative. A negative result
indicates viral antigens were
not detected in the specimen and the individual is presumed negative for
COVID-19.
- Negative results do not rule out COVID-19.
- In case of negative test result: Continue to follow all social distancing recommendations and take protective measures. If suspicions of infection persist and/or your first test is negative, repeat the test after 1 – 2 days and consult your healthcare provider or local COVID-19 center.
- Note: A negative result is presumptive and confirmation with a molecular assay, if necessary, for patient management may be performed. For serial testing programs, additional confirmatory testing with a molecular test for negative results may be necessary after second negative result for asymptomatic patients, if there is a high likelihood of SARS-CoV-2 infection, such as in an individual with as close contact with COVID-19 or with suspected exposure to COVID-19 or in communities with a high prevalence of infection. Additional confirmatory testing with a molecular test for positive results may also be necessary, if there is a low likelihood of SARS-CoV-2 infection, such as in individuals without known exposures to SARS-CoV-2 or residing in communities with a low prevalence of
infection.
– Positive result: If the red-colored line is visible in the test line (T) and control line (C), the result is positive. A positive result indicates that viral antigens from COVID-19 were present in the specimen and the individual is positive for COVID-19.
- Persons who test positive with the Celltrion DiaTrust™ COVID-19 Ag Home Test should self-isolate and seek to follow-up care with their physician or healthcare provider as additional testing and public health reporting may be necessary.
– Invalid result: If there is no red-colored line in the control region (C), the result is invalid.
- In case of an invalid test result: Repeat the test using a new test kit. If the test result is still invalid, contact your doctor or local COVID-19 center. An invalid result does not indicate if the individual did or did not have COVID-19 and should be repeated.
| NEGATIVE
If no colored line appears in the test line (T Line Level 0) and a colored
line is present on the control region (C Line Level 1 – 11), then the result
is negative. |
---|---|---
| POSITIVE
If the colored line is visible in the test line (T Line Level 1 – 11) and
control line (C Line Level 1 – 11), the result is positive.
| INVALID
If there is no colored line in the control region (C Line Level 0), the result
is invalid.
√ If invalid results are obtained, please discard the device and re-do
the testing using a new device.
√ Dispose of the test in general waste.
These are photos of actual positive results. Please note that the test line can show up faintly. This faint line still indicates a positive result.
[PERFORMANCE CHARACTERISTICS]
Analytical testing is conducted with the nasopharyngeal swab specimen, and the matrix equivalence study is conducted to support the mid-turbinate nasal swab as the specimen type.
1) Limit of detection (LoD)
LoD studies determine the lowest detectable concentration of SARS-CoV-2. The
LoD was determined by limiting dilution studies using SARS-CoV-2 virus
inactivated by beta-Propiolactone (BPL).
A negative sample was prepared by collecting nasopharyngeal swab samples from
healthy donors (negative clinical matrix) eluted in PBS.
The positive standard materials are prepared with the six different
concentrations of SARS-CoV-2 inactivated virus (Conc. 6.3 × 10⁵TCID50/mL, NMC-
nCoV02 #24) that are serially diluted in PBS and negative clinical matrix.
The diluted positive standard materials are applied to the swab tip with 100
μL of approximate absorption volume. The extraction buffer tubes are prepared
and each swab sample is inserted into each extraction buffer tubes. The swab
was moved up and down inside the tube 10 times and taken out by pressing to
remove the extracted liquid. The filter cap was equipped onto the test tube,
then three drops of extracts (100 μL) were dispensed into the sample inlet.
The result was read 15 minutes after applying the sample.
Serial dilutions of the inactivated SARS-CoV-2 were tested in 5 replicates.
The lowest concentration at which all 5 replicates were positive was treated
as the tentative LoD for each test. Based on this testing, the tentative LoD
was 3.2 × 10¹ TCID50/mL.
The LoD of each test was then confirmed by testing 20 replicates with
concentrations near the tentative limit of detection. The final LoD of
Celltrion DiaTrust™ COVID-19 Ag Home Test was determined to be the lowest
concentration resulting in positive detection more than 95% of the time, which
is at least 19 out of 20 replicates.
In conclusion, the limit of detection (LoD) of Celltrion DiaTrust™ COVID-19 Ag
Home Test for NP swab is 3.2 × 10¹ TCID50/mL.
3) Cross-reactivity (Analytical specificity) and Microbial Interference
Studies Wet-testing: The study was performed to evaluate the cross-reactivity
of the Celltrion DiaTrust™ COVID-19 Ag Home Test.
Nasopharyngeal swab samples from healthy donors (negative clinical matrix)
were collected and eluted in extraction buffer to be used as a negative
standard material. For each test, the diluted sample was added to a sterile
nasal swab before conducting the test according to the instruction for use.
Positive standard materials (NMC-nCoV02 #24, 6.3 × 10¹ TCID50/mL) were spiked
into the negative sample and were diluted to make low concentration level (6.3
× 10⁵
TCID50/mL, approx. 2xLoD) for testing.
Potential cross-reactive organisms listed in the below table were prepared at the concentration of 10⁶ PFU/mL or higher for viruses and 10⁵ CFU/mL or higher for bacteria. They were spiked into the negative and low positive samples and were tested in 3 replicates. A total of 31 pathogens are listed in the below table showed no cross-reactivity with the Celltrion DiaTrust™ COVID-19 Ag Home Test.
List of organisms
|
Testing conc.
| Test result
---|---|---
Negative
(No. of negative/ No. of replicates)
| Low Positive (No. of positive/ No. of replicates)
Other high priority pathogens from the same virus family
| Coronavirus OC43| 4.4 × 107 PFU/mL| 3/3| 3/3
Coronavirus 229E| 3 × 106 PFU/mL| 3/3| 3/3
Coronavirus NL63| 1 × 105 TCID50/mL| 3/3| 3/3
MERS-coronavirus| 1.183 × 105
TCID50/mL
| 3/3| 3/3
Other high priority organisms
| Human adenovirus 1| 7 × 10 7 PFU/mL| 3/3| 3/3
Human adenovirus 3| 2.4 × 10 6 PFU/mL| 3/3| 3/3
Human adenovirus 5| 4.0 × 10 7 PFU/mL| 3/3| 3/3
Human adenovirus 7| 2.0 × 10 8 PFU/mL| 3/3| 3/3
Respiratory syncytial virus A| 8.0 × 10 5 PFU/mL| 3/3| 3/3
Respiratory syncytial virus B| 2.4 × 10 6 PFU/mL| 3/3| 3/3
Parainfluenza 1| 2.8 × 10 5 PFU/mL| 3/3| 3/3
Parainfluenza 2| 2 × 10 7 PFU/mL| 3/3| 3/3
Parainfluenza 3| 8 × 105 PFU/mL| 3/3| 3/3
Parainfluenza 4a| 1.3 × 108 PFU/mL| 3/3| 3/3
Rhinovirus 1| 1.4 × 105 PFU/mL| 3/3| 3/3
Metapneumovirus| 6 × 105 PFU/mL| 3/3| 3/3
Human enterovirus| 1 × 105 PFU/mL| 3/3| 3/3
Influenza A H1N1| 2 × 105 PFU/mL| 3/3| 3/3
Influenza A H3N2| 4.9 × 106 PFU/mL| 3/3| 3/3
Influenza B| 1 × 106 PFU/mL| 3/3| 3/3
Mycoplasma pneumonia
(whole organism)
| 1 × 107 CFU/mL| 3/3| 3/3
Streptococcus pyogenes| 1 × 106 CFU/mL| 3/3| 3/3
Bordetella pertussis| 1 × 106 CFU/mL| 3/3| 3/3
Streptococcus pneumoniae| 1 × 106 CFU/mL| 3/3| 3/3
Legionellapneumophila| 1 × 106 CFU/mL| 3/3| 3/3
Haemophilus influenza| 1 × 106 CFU/mL| 3/3| 3/3
Candida albicans| 1 × 106 CFU/mL| 3/3| 3/3
Chlamydiapnuemoniae| 2.0 × 107 TCID50/mL| 3/3| 3/3
Pooled human nasal wash| 100%| 3/3| 3/3
Staphylococcus epidermidis| 1 × 106 CFU/mL| 3/3| 3/3
Staphylococcus aureus| 1 × 106 CFU/mL| 3/3| 3/3
Human coronavirus HKU1 spike protein at the concentration of 10 μg/mL was
spiked into negative and positive samples. It was tested in 3 replicates using
the Celltrion DiaTrust™ COVID-19 Ag Home Test, and no cross-reactivity was
observed.
In-silico: To estimate the likelihood of cross-reactivity with SARS-CoV-2
virus in the presence of organisms that were not available for wet testing, in
silico analysis was used to assess the degree of protein sequence homology.
- Human coronavirus HKU1: 25% homology was found between SARS-CoV-2 Receptor Binding Domain spike proteins and HKU1 spike protein, and 44% homology was found between SARS-CoV2 Nucleocapsid protein and HKU1 Nucleocapsid protein. Therefore, cross-reactivity cannot be ruled out.
- Pneumocystis jirovecii: No significant similarity was found between SARS-CoV-2 RBD spike protein/nucleocapsid protein and P. jirovecii. But minor similarity was found between some partial proteins of P. jirovecii RU 7and SARS-CoV-2 RBD spike protein/nucleocapsid protein. Therefore, cross-reactivity cannot be ruled out.
- Mycobacterium tuberculosis: No significant similarity was found between M. tuberculosis and SARS-CoV-2 RBD spike protein/nucleocapsid protein despite increasing expect threshold.
- SARS-CoV: 72% homology was found between SARS-CoV-2 Receptor Binding Domain spike proteins and SARS-CoV spike protein, and 96% homology was found between SARS-CoV-2 Nucleocapsid protein and SARS-CoV Nucleocapsid protein. Therefore, cross-reactivity is highly likely.
- The Celltrion DiaTrust COVID-19 Ag Home Test does not differentiate between SARS-CoV and SARS-CoV-2. TM
Endogenous interference substances study: Test to evaluate the interference of the Celltrion DiaTrust™ COVID-19 Ag Home Test was performed. Extraction buffer was used as a negative sample. Positive standard materials were spiked into the negative sample and were diluted to make low concentration levels (6.3 × 10¹ TCID50/mL, approx. 2xLoD) for testing. Potential interfering substances were added to the negative and positive samples and were tested using the Celltrion DiaTrust TM COVID-19 Ag Home Test in 3 replicates. The test results demonstrated that 48 interfering substances did not affect the performance of the Celltrion DiaTrust TM COVID-19 Ag Home Test.
No.
|
Interfering substances
|
Testing conc.
|
Negative
| Negative + Interfering substances| Low positive| Low
pos. + Interfering substances
---|---|---|---|---|---|---
1| Whole blood| 4%| 3/3| 3/3| 3/3| 3/3
2| Mucin| 0.5%| 3/3| 3/3| 3/3| 3/3
3| Chloraseptic| 1.5 mg/mL| 3/3| 3/3| 3/3| 3/3
4| NeilMed NasoGel| 5% v/v| 3/3| 3/3| 3/3| 3/3
5| CVS Nasal drops| 15% v/v| 3/3| 3/3| 3/3| 3/3
6| Afrin (Oxymetazoline)| 15% v/v| 3/3| 3/3| 3/3| 3/3
7| Sodium cromoglycate| 15% v/v| 3/3| 3/3| 3/3| 3/3
No.
|
Interfering substances
|
Testing conc.
|
Negative
| Negative + Interfering substances| Low positive| Low
pos. + Interfering substances
---|---|---|---|---|---|---
| (CVS nasal spray, Cromolyn)| | | | |
8| Zicam| 15% v/v| 3/3| 3/3| 3/3| 3/3
9| Homeopathic (Alkalol)| 1:10 dilution| 3/3| 3/3| 3/3| 3/3
10| Sore throat Phenol Spray| 15% v/v| 3/3| 3/3| 3/3| 3/3
11| Tobramycin| 5 μg/mL| 3/3| 3/3| 3/3| 3/3
12| Mupirocin| 10 mg/mL| 3/3| 3/3| 3/3| 3/3
13| Fluticasone Propionate| 5% v/v| 3/3| 3/3| 3/3| 3/3
14| Tamiflu (Oseltamivir Phosphate)| 5 mg/mL| 3/3| 3/3| 3/3| 3/3
15| Albumin, human| 3000 mg/dL| 3/3| 3/3| 3/3| 3/3
16| Bilirubin| 500 μmol/L| 3/3| 3/3| 3/3| 3/3
17| Hemoglobin| 500 mg/dL| 3/3| 3/3| 3/3| 3/3
18| Cholesterol| 20 μmol/L| 3/3| 3/3| 3/3| 3/3
19| Triglyceride| 1000 mg/dL| 3/3| 3/3| 3/3| 3/3
20| Biotin| 0.75 mg/mL| 3/3| 3/3| 3/3| 3/3
21| Sodium citrate| 25 mg/mL| 3/3| 3/3| 3/3| 3/3
22| Heparin| 100 U/mL| 3/3| 3/3| 3/3| 3/3
23| EDTA| 5 μmol/L| 3/3| 3/3| 3/3| 3/3
24| K3-EDTA| 20 mg/mL| 3/3| 3/3| 3/3| 3/3
25| Diphenhydramine hydrochloride| 5 mg/mL| 3/3| 3/3| 3/3| 3/3
26| Acetaminophen| 199 μmol/L| 3/3| 3/3| 3/3| 3/3
27| Acetylsalicylic acid| 3.62 mmol/L| 3/3| 3/3| 3/3| 3/3
28| Ibuprofen| 2.425 mmol/L| 3/3| 3/3| 3/3| 3/3
29| Olopatadine hydrochloride| 5 mg/mL| 3/3| 3/3| 3/3| 3/3
30
| Hanmi Ko-and-Cool Nasal Spray (Chlorpheniramine Maleate 250 mg/ 100 mL, Xylometazoline
Hydrochloride 0.1 g/100 mL)
|
10%(v/v)
|
3/3*
|
3/3*
|
3/3**
|
3/3**
31
| Samchundang Narista-S Nasal Spray (Chlorpheniramine Maleate 2.5 mg/mL, Dipotassium Glycyrrhizinate 3
mg/mL, Naphazoline Hydrochloride 0.5 mg/mL)
|
10%(v/v)
|
3/3*
|
3/3*
|
3/3**
|
3/3**
32| Sodium chloride| 20 mg/mL| 3/3| 3/3| 3/3| 3/3
33| Zanamivir| 5 mg/mL| 3/3| 3/3| 3/3| 3/3
No.
|
Interfering substances
|
Testing conc.
|
Negative
| Negative + Interfering substances| Low positive| Low
pos. + Interfering substances
---|---|---|---|---|---|---
34| Oseltamivir| 10 mg/mL| 3/3| 3/3| 3/3| 3/3
35| Artemether- lumefantrine| 50 μmol/L| 3/3| 3/3| 3/3| 3/3
36| Doxycycline hyclate| 70 μmol/L| 3/3| 3/3| 3/3| 3/3
37| Quinine| 150 μmol/L| 3/3| 3/3| 3/3| 3/3
38| Lamivudine| 1 mg/mL| 3/3| 3/3| 3/3| 3/3
39| Erythromycin| 81.6 μmol/L| 3/3| 3/3| 3/3| 3/3
40| Ciprofloxacin| 30.2 μmol/L| 3/3| 3/3| 3/3| 3/3
41| Rheumatoid factor positive plasma| 10%(v/v)| 3/3| 3/3| 3/3| 3/3
42| Neutrogena lotion (glycerin)| 1% (v/v)| 3/3| 3/3| 3/3| 3/3
43| Hand sanitizer (ethyl alcohol)| 1% (v/v)| 3/3| 3/3| 3/3| 3/3
44| Hand soap (benzalkonium chloride)| 1% (v/v)| 3/3| 3/3| 3/3| 3/3
45
| Laundry detergent (C12- 15 pareth-7 and sodium laureth-12 sulfate)|
1% (v/v)
|
3/3*
|
3/3*
|
3/3**
|
3/3**
46| Bleach (sodium hypochlorite)| 1% (v/v)| 3/3| 3/3| 3/3| 3/3
47| Surface sanitizer (citric acid)| 1% (v/v)| 3/3| 3/3| 3/3| 3/3
48| Dish-washing liquid (sodium lauryl sulfate)| 1% (v/v)| 3/3| 3/3| 3/3|
3/3
*: Negative / **: Positive
4) High-dose Hook effect
Pooled nasopharyngeal specimens were used as clinical matrix, and SARS-CoV-2
virus inactivated by beta-Propiolactone (BPL) was spiked to make various high
concentration levels of SARS-CoV-2 antigens. Prepared samples of each
concentration level were tested using Celltrion DiaTrust™ COVID-19 Ag
Home Test in 3 replicates following instructions. No high-dose hook effect was
observed up to 6.3 × 10⁵ TCID50/mL, approx. 20,000xLoD.
SARS-CoV-2 inactivated virus
( 6.3 105 TCID50/mL )
TCID 50 /mL
(concentration)
| Test results
(No. of positives/ No. of replicates)
Lot 1| Lot 2
3.2 101 [1xLoD]| 3/3| 3/3
1.3 102 [4xLoD]| 3/3| 3/3
1.5 104 [500xLoD]| 3/3| 3/3
6.3 105 [20,000xLoD]| 3/3| 3/3
5) Flex study
The robust use of Celltrion DiaTrust™ COVID-19 Ag Home Test was demonstrated
by ten (10) Flex studies: temperature and humidity, delay in sample testing,
delay in result reading, extraction buffer volume variability, swab mixing
expression variability, disturbance during testing, testing on non-level
surface, the impact of light sources, test device held at 90° angle, and
disturbance during analysis – receiving a phone call while the mobile app is
running.
6) Clinical performance
The clinical evaluation of the Celltrion DiaTrust™ COVID-19 Ag Home Test was
evaluated by testing a total of 492 prospectively collected direct mid-
turbinate nasal swab samples, consisting of 45 positive and 447 negative
samples from suspected COVID-19 patients in the United States, aged 14 years
and older at four clinical sites. Mid-turbinate nasal swabs were collected and
tested by each study participant, eluted in the extraction buffer, and tested
with the device immediately, using only the QRI and App. The results of each
sample were confirmed by FDA EUA RT-PCR.
According to the test results, clinical performance results of the Celltrion
DiaTrust COVID-19 Ag Home Test was as follows:
Table 1. Demographic and Clinical Characteristics
Characteristic
| ****
Total number
| Total Positive by RT-PCR| ****
% Positive
---|---|---|---
Age Range
| 14-24| 88| 12| 12/88 (13.6%)
25-64| 381| 31| 31/381 (8.1%)
≥65| 23| 2| 2/23 (8.7%)
Sex| | |
Female| 266| 18| 18/266 (6.8%)
Male| 226| 27| 27/226 (11.9%)
Total| 492| 45| 45/492 (9.1%)
Table 2. Observations of All subjects
All Data
| Reference PCR Results
---|---
Positive| Negative| Total
DiaTrustTM COVID-19 Ag Home Test
| Positive| 39| 1| 40
Negative| 6| 446| 452
Total| 45| 447| 492
PPA:86.7 % (95% CI: 73.8%-93.7%)
NPA: 99.8 % (95% CI: 98.7%-100.0%)
Table 3. Observations of Symptomatic subjects
Symptomatic Data
| Reference PCR Results
---|---
Positive| Negative| Total
DiaTrustTM COVID-19 Ag Home Test
| Positive| 31| 1| 32
Negative| 5| 174| 179
Total| 36| 175| 211
PPA: 86.1% (95% CI: 71.3% – 93.9%)
NPA: 99.4% (95% CI: 96.8% – 99.9%)
Table 4. Observations of Asymptomatic subjects
Asymptomatic Data
| Reference PCR Results
---|---
Positive| Negative| Total
DiaTrustTM COVID-19 Ag Home Test
| Positive| 8| 0| 8
Negative| 1| 272| 273
Total| 9| 272| 281
PPA: 88.9 % (95% CI: 56.8%-98.0%)
NPA: 100.0% (95% CI: 98.6%-100.0%)
Table 5. PPA and NPA by days since onset of symptoms
Days since symptom onset | PPA (95% CI) | NPA (95% CI) |
---|---|---|
Asymptomatic | 88.9% (8/9) | |
(95% CI: 56.5% – 98.0%) | 100% (272/272) |
(95% CI: 98.6% – 100.0%)
1| 75.0% (3/4)
(95% CI: 30.1% – 95.4%)| 95.8% (23/24)
(95% CI: 79.8% – 99.3%)
2| 100.0% (8/8)
(95% CI: 67.6% – 100.0%)| 100.0% (40/40)
(95% CI: 91.2% – 100.0%)
3| 100.0% (9/9)
(95% CI: 70.1% – 100.0%)| 100.0% (38/38)
(95% CI: 90.8% – 100.0%)
4| 85.7% (6/7)
(95% CI: 48.7% – 97.4%)| 100.0% (30/30)
(95% CI: 88.6% – 100.0%)
5| 66.7% (2/3)
(95% CI: 20.8% – 93.9%)| 100.0% (24/24)
(95% CI: 86.2% – 100.0%)
6| 100.0% (2/2)
(95% CI: 34.2% – 100.0%)| 100.0% (12/12)
(95% CI: 75.8%-100.0%)
7| 33.3% (1/3)
(95% CI: 6.1%-79.2%)| 100.0% (7/7)
(95% CI: 64.6%-100.0%)
[ASSISTANCE]
If you have any questions regarding the use of this product or if you want to report a test system problem, please contact Humasis Co., Ltd. (via email: info@humasis.com, via phone: +82-31-80856284) or Celltrion USA, Inc. (via email: celltrionusa.CS@celltrion.com, or via phone: 201-499-1844). Test system problems may also be reported to the FDA through the MedWatch medical products reporting program (phone: 800.FDA.1088; fax: 800.FDA.0178; http://www.fda.gov/medwatch).
[REFERENCES]
[1] Zhu N, Zhang D, Wang W, et al. A Novel Coronavirus from Patients with
Pneumonia in China, 2019. N Engl J Med 2020.
[2] Huang C, Wang Y, Li X, et al. Clinical features of patients infected with
2019 novel coronavirus in Wuhan, China. Lancet 2020.
[3] Kang CK, Song KH, Choe PG, et al. Clinical and Epidemiologic
Characteristics of Spreaders of
Middle East Respiratory Syndrome Coronavirus during the 2015 Outbreak in
Korea. J Korean Med Sci
2017; 32:744-9.
[4] WHO, Novel Coronavirus (2019-nCoV) situation reports. Available at:
https://www.who.int/emergencies/diseases/novel-
coronavirus-2019/situationreports/ (Accessed at 2 Feb, 2020).