NEXSENS X2-I Environmental Data Logger User Guide

June 6, 2024
NEXSENS

NEXSENS X2-I Environmental Data Logger

NEXSENS-X2-I-Environmental-Data-Logger

For best results, setup your new system in the office/lab. Collect a few data points and become familiar with the system prior to field deployment.

  1. Go to WQDataLIVE.com/Getting-Started

  2. Create a new account or sign into an existing account. Choose or create a project by selecting the Projects link in the bottom right footer of the page.

  3. Go to the ADMIN tab located at the top of the project dashboard and click Settings.NEXSENS-X2-I-Environmental-Data-Logger-1

  4. From there, choose the Project/Site pull down menu and select the site for the new data logger. If a site has not been created, select New Site.

  5. Enter the claim code listed below into the space provided under Assigned Devices.

  6. Click Add Device.

    • The logger will now be visible in the Assigned Devices list.
  7. Scroll down to the Advanced Device Remote Configuration tab, open the drop-down menu, and select the logger within the site name.

    • Enter key ‘nexsens’ to access the Advanced Device Remote Configuration settings.
  8. Scroll down to Transmission Settings and choose Iridium Satellite from the drop-down menu under New Value.NEXSENS-X2-I-Environmental-Data-Logger-3

  9. Enter the following settings into the available fields and then click SAVE:

    • Modem IMEI
    • Email Server: imap.gmail.com
    • (Insert this server name if using Gmail; however, any host domain will work.)
    • IMAP Port: 993 (Usually 143, or 993 when using SSL)
    • Pull down and select ‘Yes’ and input the email address and password for the Custom Transmit Email, along with the email address for the Custom Command Email.
    • Custom Transmit Email
    • Receives emails sent from the logger to WQData LIVE.
    • Custom Command Email
    • Sends queue commands from WQData LIVE to the logger.
    • Select ‘Yes’ for Use SSL for IMAPNEXSENS-X2-I-Environmental-Data-Logger-4
  10. Move the system outside to an area devoid of overhead obstructions.

  11. Connect the iridium antenna, which is included within the X2 shipment, to the antenna port of the logger.NEXSENS-X2-I-Environmental-Data-Logger-5

  12. Remove the required blank sensor plugs (P0/P1/P2) and connect all sensors.

  13. If ports arrive labeled by sensor, establish the connections as indicated.

  14. Remove the COM/SOLAR blank plug and connect the solar panel cableNEXSENS-X2-I-Environmental-Data-Logger-6

  15. Wait up to 20 minutes for sensor detection to complete.

  16. Refresh WQData Live and confirm all sensor parameters are shown.

  17. Return to the Advanced Device Remote Configuration tab.

  18. Scroll down to the Transmit Flag section and expand the list for all sensors.NEXSENS-X2-I-Environmental-Data-Logger-7

  19. Change the New Value next to each parameter to Yes for any parameters that should be uploaded via satellite and No to disable unwanted parameters.
    Note that the number of transmitted parameters, sensor logging intervals, and the system transmission intervals all affect satellite data usage and telemetry costs.
    Ensure that the configured iridium data plan can support the desired system settings.

  20. Click SAVE at the bottom of the configuration menu.

    • At the next scheduled transmission interval, the parameter settings will be applied. Data will begin to upload on the transmission interval after these changes are applied.

2091 Exchange Court Fairborn, Ohio
45324 937-426-2703
www.NexSens.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

NEXSENS User Manuals

Related Manuals