NEXSENS X2 X2 Data Logger User Guide
- June 4, 2024
- NEXSENS
Table of Contents
X2 X2 Data Logger
User Guide
X2 X2 Data Logger
IMPORTANT – BEFORE FIELD DEPLOYMENT: Completely configure new X2 systems with sensors and a direct connection to the CONNECT software in a nearby work area. Operate the system for several hours and ensure correct sensor readings. Use this test run to become familiar with the features and functions.
-
Visit the following link on the NexSens Knowledge Base to download the CONNECT software and establish a connection with the X2.
a. nexsens.com/connst -
Use the following link to ensure the proper scripts are enabled for each sensor.
a. nexsens.com/conncss -
Power down the X2 and remove the USB cable connection.
a. Remove one blank sensor plug from an 8-pin port (i.e., P0, P1, or P2) for each sensor.
b. Connect all sensors to the desired ports.
Note: Ensure that all SDI-12 and RS-485 sensors have unique addresses. -
Provide 12V power to the X2 and wait up to 5-10 minutes for sensor detection
a. Reconnect the USB cable to the X2 and open CONNECT. -
Once in CONNECT, visit the following article to confirm the X2 sensor configuration and directly download the first few data points.
a. nexsens.com/conndu
b. If the desired sensor configuration is not shown:
• Confirm the proper sensor scripts are enabled and all SDI-12 or RS-485 sensors have unique addresses.
• Confirm all user-configured wiring of sensors.
• Run a new sensor detection in CONNECT.
Figure 1: X2 Environmental Data Logger.
Overview
The X2 includes three sensor ports that provide industry-standard protocols
including SDI-12, RS-232, and RS-485. The center port offers direct
communication (serial to PC) to the CONNECT software and power input.
CONNECT is a software utility that enables users to interface directly with
any NexSens X2-Series data logger using a UW6-USB-485P cable. It supports a
growing number of diagnostic and configuration tools to facilitate system
setup and troubleshooting.
What’s Included?
- (1) X2 data logger
- (1) X2 grounding kit
- (3) Sensor port plugs, sparrings
- (1) Power port plug, sparring
- (1) Oring grease
- (1) Quick start guide
For additional information, please reference the X2 & CONNECT software
resource libraries on the NexSens Knowledge Base.
nexsens.com/x2kb
nexsens.com/connug
937-426-2703
www.nexsens.com
2091 Exchange Court
Fairborn, Ohio 45324
References
- Configure Sensor Scripts - NexSens
- Direct Data Upload - NexSens
- CONNECT Setup - NexSens
- Knowledge Base - NexSens
- Knowledge Base - NexSens
- NexSens Technology Inc. - Better Data. It's what we do