Inogen One G3 Replacement Columns Instructions
- June 6, 2024
- inogen
Table of Contents
Inogen One G3 Replacement Columns
To change the columns on your Inogen One G4, first:
- Turn off the Inogen One® G4 concentrator by pressing the power button to shut down the device.
- Remove the Inogen One® G4 concentrator from the carrying case (if attached).
- Remove the battery from the Inogen One® G4 concentrator.
- Place the Inogen One® G4 concentrator on its side so that the underside is visible. The metal column assembly can be seen on one side of the device.
- Unlock the column assembly by pushing the latch button away from the columns.
- While holding the button open, slide the column assembly out of the device by pulling on the column pull handle.
- Remove the columns completely from the Inogen One® G4. Both columns are removed as one piece.
- Remove dust caps of new coulmn assembly. Make sure there is no dust or debris where the dust caps were located.
- Insert column assembly into the Inogen One® G4 concentrator. Do not leave the column ends exposed; column assembly should be inserted into the Inogen One G4 as soon as the dust caps have been removed.
- Push the column assembly into the device such that the columns are fully seated into the Inogen One® G4 concentrator. The spring loaded latch button should fully return to the closed position.
- Connect the AC power supply cord to the Inogen One® G4 and plug the power supply AC cord into an electrical outlet. Do Not Power on the Inogen One® G4 concentrator.
- Press and hold the plus (+) and (-) minus buttons for 5 seconds. The screen will display “sieve reset” message. Release buttons once message is displayed on screen.
- Press the alert/bell button once and screen will display “sieve reset success.” Press the power button to turn on the Inogen One® G4, and use it normally.
- If you are using Inogen Connect App, navigate to the Adv anced screen, then to Additional Information screen and click on the Column Reset button.
Note: Column change instructions are only to be used when maintenance is required and are not intended for practice purpose.
Your Equipment Provider
Your Physician:
- 301 Coromar Drive,
- Goleta, CA 93117
- 1-877-4-IMOGEN (Toll Free)
- 1-877-446-6436 (Toll-Free)
- 1-805-562-0515 (International)
- info@inogen.net
- www.inogen.com
References
Read User Manual Online (PDF format)
Loading......