ResMed Stellar-100 Air Tubing AC Power User Guide
- June 6, 2024
- ResMed
Table of Contents
ResMed Stellar-100 Air Tubing AC Power User Guide
Setup
Noninvasive use
Invasive use
- If not using an external humidifier, connect the HMEF to the ResMed Leak Valve.
- The ResMed Leak Valve or the HMEF can be connected to standardized tracheostomy interfaces including connector pieces like catheter mounts.
The catheter mount and the external humidifier are not part of the ResMed component.
About the control panel
Rear panel
Turning the ventilator on/off
To turn the ventilator on, press the power switch at the rear of the device.
Wait until the Treatment screen is displayed.
To turn the ventilator off, press the power switch at the rear of the device.
Follow the onscreen instructions.
Accessing Clinical mode
Press and simultaneously for three seconds until the duration options appear. Follow the onscreen instructions.
To return to the Patient menu, press and simultaneously for at least three seconds.
Treatment settings
Yellow arrows display at the top or bottom of the pressure bar graph will show
if a machine trigger or cycle occurred. If the inspiratory phase extends
beyond the spontaneous cycle by Ti Min, a red arrow is displayed.
Selecting the mask type and performing a Learn circuit enables to calculate
the impedance and leak for controlled therapy pressure.
Learn Circuit configuration
Do not connect patient interfaces prior to performing the Learn Circuit. Patient interfaces include any components placed after the ResMed Leak Valve (eg, HMEF, catheter mount, tracheostomy tube).
With an external humidifier
Without an external humidifier
Using the menus
Menu overview
Note: iVAPS mode is only available in Stellar 150
Overview Clinical Settings
Note: iVAPS mode is only available in Stellar 150
Overview Alarm Settings and Options
Pathology defaults
Setting| Mode| Obstructive lung disease| Restrictive lung
disease| Obesity Hypoventilation Syndrome| Normal lung mechanics
---|---|---|---|---|---
S| ST| T| PAC| iVAPS
IPAP [cm H2O]| √| √| √| √| | 13| 11| 16| 11
EPAP or PEEP [cm H2O]| √| √| √| √| √| 5| 5| 8| 5
PS [cm H2O]| √| √| √| √| | 8| 6| 8| 6
Rise time [ms]| √| √| √| √| √| 150| 300| 300| 200
Fall time [ms]| √| √| √| √| √| 200| 200| 200| 200
Trigger sensitivity| √| √| | √| √| Medium| Medium| Medium| Medium
Cycle sensitivity| √| √| | | √| High| Low| Medium| Medium
Min PS [cm H2O]| | | | | √| 4| 4| 4| 2
Max PS [cm H2O]| | | | | √| 20| 20| 18| 20
Min EPAP [cm H2O]| | | | | √| 5| 5| 8| 5
Max EPAP [cm H2O]| | | | | √| 15| 15| 15| 15
Ti Min [sec]| √| √| | | √| 0.3| 0.5| 0.5| 0.5
Ti Max [sec]| √| √| | | √| 1.0| 1.5| 1.5| 1.5
- The Rise/Fall Time milliseconds scale is an approximate only.
Note: iVAPS mode is only available in Stellar 150.
See the Stellar Clinical Guide for full details.
Not all devices are available in all regions.
MANUFACTURER
ResMed Germany Inc. Fraunhoferstr. 16 82152 Martinsried Germany
DISTRIBUTED BY
ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
ResMed Corp 9001 Spectrum Center Blvd San Diego CA 92123 USA
ResMed (UK) Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW
UK
See www.resmed.com for other ResMed locations worldwide. For patent information, see www.resmed.com/ip. ResMed and Stellar are trademarks of ResMed Pty Ltd. ResMed and Stellar are registered in U.S. Patent and Trademark Office. © 2015 ResMed. 248804/1 2015-05
References
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Sleep apnea and COPD - learn about symptoms and treatment | ResMed
- Legal disclaimers and intellectual property - ResMed
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>