PHILIPS 1109290 Mini Nebulizer Compressor
Compressor nebulizer system
CAUTION:
Federal (USA) law restricts this device to sale by or on the order of a
physician or licensed healthcare professional.
InnoSpire Mini InnoSpire Mini compressor (back
view)
Read these instructions before using this compressor. If you do not understand any portion of these instructions, contact your healthcare provider or call Philips Respironics Customer Service at 1-800-345-6443 or 1-724-387-4000.
Intended use statement
The Inspire Mini is intended to provide a source of compressed air for
medical purposes. It is to be used with a pneumatic (jet) nebulizer to produce
aerosol particles of medication for respiratory therapy for both children and
adults.
Unpacking your InnoSpire Mini system
When you unpack your InnoSpire Mini from its shipping case:
- Carefully examine the contents against items listed and for damage. Make sure you have all the content items listed and that they are not damaged.
- Save these instructions for use for reference.
- Report anything missing or damaged to Philips Respironics.
Note:
Retain carton and packing materials for storing the unit or for product returns.
Read these warnings and cautions before using your InnoSpire Mini
Warnings and cautions
It is recommended to have a backup device (e.g. MDI) for respiratory
delivery in case a situation arises when your nebulizer cannot be used (e.g.
during a power outage or if your compressor and/or battery becomes inoperable
for any reason).
DANGER – To reduce the risk of electrocution:
- Always unplug this compressor immediately after using.
- Do not place or store product where it can fall or be pulled into a tub or sink.
- Do not place in or drop into water or other liquid.
- Do not reach for product that has fallen into water. Unplug immediately.
Electromagnetic information
Portable and RF communication devices such as cellular phones, pagers, etc.
can interrupt operation of electrical medical equipment. For this reason, your
compressor must be placed far enough away from these devices to prevent
interference. This device meets the IEC60601-1-2 Standard for Electromagnetic
Compatibility (EMC). EMC data sheets are available upon request from Philips
Respironics Customer Service at
1-800-345-6443 or
1-724-387-4000. The CE mark on the
product denotes compliance with all applicable EU Directives. Note that the
Notified Body number does not apply to the RoHS (restriction of the use of
certain hazardous substances in electrical and electronic equipment)
Directive.
WARNING – To reduce the risk of burns, electrocution, fire or injury to
persons
-
Do not operate compressor while it is in carrying case.
-
A compressor should never be left unattended when plugged in.
-
Adult supervision is necessary when this product is used by, on, or near children or physically challenged individuals.
-
Use this product only for its intended use as described in this manual. Failure to do so
may compromise performance. Do not use nebulizers or accessories not recommended by the manufacturer. -
Never operate this product if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water.
-
Do not use with an extension cord.
-
Do not use this device in-line in a ventilator circuit.
-
Keep the power cord away from heated surfaces.
-
This unit has been designed for intermittent use: 30 minutes ON / 30 minutes OFF. Failure to operate appropriately may lead to device damage.
-
To avoid the risk of possible electrical shock, the compressor must be operated using the specified power sources.
-
Do not remove the compressor housing or disassemble unit. There are no serviceable parts.
-
Never use while sleeping or drowsy.
-
Do not use while operating a vehicle.
-
InnoSpire Mini and all its parts must be disposed of according to the local regulations in force.
-
Do not stack the compressor with other equipment.
-
Small parts can be inhaled, swallowed or dangerous. Cable, due to its length, may result in provoking strangulation or asphyxiation. Do not leave the device alone with a small child or physically challenged individual or individual with learning difficulties.
-
People who require assistance might underestimate the risks involved, such as strangulation, therefore they should be supervised.
-
Device should not be used around flammable substances, in an oxygen rich environment, or in the presence of a flammable anaesthetic mixture with air, oxygen or nitrous oxide.
-
Use the product in a clean environment, one that is free from dirt, dust, pet hair etc.
-
Do not use the compressor on a bed/blanket to avoid overheating.
-
In case of accidental spillage of water or any liquid, dry the compressor immediately, and then allow any residual moisture to dry completely before operating.
How to use your InnoSpire Mini
- Check the nebulizer and compressor for damage. Read the nebulizer instructions prior to use.
- Place your InnoSpire Mini compressor on a firm, flat surface near an electrical outlet (if using AC power) and where you can be seated comfortably during your treatment. Do not hold the compressor in hand during treatment.
- To use the compressor with the AC power adapter, connect the AC power adapter into the power plug-in on the back of the compressor. Then connect the other end into the electrical outlet.
- alternatively, the unit can be used with the rechargeable battery pack or 12V DC (car) adapter. Read the section relating to the rechargeable battery pack carefully. Use only with 12V vehicle electrical system.
- Do not operate the unit without the filter or with a clogged (dirty) or wet filter. Operating the unit with a dirty filter can impair proper operation and cause damage that is not covered by the warranty.
- Wash your hands before beginning your treatment.
- Assemble your nebulizer and add your prescribed medicine following your prescription and the separate nebulizer instructions.
- Connect one end of the tubing to the bottom of the nebulizer and the other end of the tubing to the air outlet on the compressor.
- Turn the power on/off switch to begin your treatment. A green light appears indicating that the unit is on.
- When your treatment is complete, turn the power on/off switch and unplug the unit if usingAC power or 12V DC (car) adapter.
- Clean the nebulizer according to its instructions.
InnoSpire Mini Rechargeable lithium-ion battery
Read these instructions before using this battery The battery will not charge when it is connected to the compressor Read these warnings and cautions before using this battery Warnings and cautions
- Use only with Philips Respironics Innospire Mini compressor and power adapter.
- Do not immerse in water.
- Do not use battery if accidentally dropped in water.
- Do not touch with wet or moist hands.
- Do not dispose of in fire.
- Do not expose to direct sunlight, chemicals, high temperatures or humidity.
- Do not allow metal objects to touch metal parts of the battery as it may result in a short circuit.
- Do not disassemble.
- Do not try to operate the compressor while charging the battery as the compressor will not operate.
How to attach and detach the rechargeable lithium-ion battery
- Shift the contact cover on the bottom of the Inspire Mini compressor so the battery contacts are uncovered.
- To attach the battery, align the back end of the Innospire Mini compressor into the back end of the rechargeable lithium-ion battery.
- Fit the front end of the compressor down until it engages with a “click” into the front end of the rechargeable lithium ion battery.
- To remove the rechargeable lithium-ion battery, lift the front of the compressor off the battery while pressing the release button on the rechargeable lithium ion battery.
Charging the battery
- Charge the battery pack fully prior to first use.
- Connect the AC or 12V DC car adapter to the power adapter plug-in on the front side of the battery.
- Plug the other end of the AC or DC car adapter into the wall outlet or 12V DC car outlet.
- The battery indicator light will turn orange.
- Charge the battery until the battery indicator turns green. It may take up to 6 hours to fully charge the battery.
- Unplug the power adapter from both the battery power plug-in and the wall outlet or 12V DC car outlet.
Storing the battery
- Store the battery fully charged in a cool dry place.
- Every 1-2 months, fully charge the battery to maintain maximum battery capacity.
- It may take up to 5 charge/discharge cycles to return the battery to maximum capacity after prolonged storage.
Battery indicator lights
- The battery indicator light will turn orange when the battery is being charged.
- The battery indicator light will turn green when the battery is fully charged while still connected to charging source.
- The battery indicator light will blink orange when the battery is almost exhausted/drained.
- The light is off under the following conditions:
- when the battery has charge and is not connected to a charging source
- when battery is fully discharged
- when battery enters a fault condition
CAUTION:
Clean the device, according to instructions, between patients when used on
multiple patients. Care and cleaning of your InnoSpire Mini
CAUTION:
To reduce risk of electric shock, DO NOT DISASSEMBLE and never submerge or
place under running water.
- Unplug the compressor before cleaning.
- Use a damp cloth to wipe down the outside of the compressor as required. Do not use abrasive cleaners.
- Check the filter periodically and replace the filter when dirty.
- To change the filter:
- Use a coin to unscrew the filter holder (Note: Turn clockwise to unscrew).
- Pull out filter and replace with new filter.
- Screw filter holder back on by turning counterclockwise.
- After each cleaning, check that the unit is working properly by verifying that air still comes out of the nozzle when the unit is turned to the “on” position and that the unit’s noise level has not increased. (Increased noise indicates improper filter positioning.)
- The compressor is oil-free. Do not lubricate.
Troubleshooting
Problems | Solutions |
---|
Compressor overheats. The compressor has an internal thermal protector to prevent the device from overheating. If the internal temperature reaches the preset level, the unit will automatically shut down, and the ON/OFF button light will turn red.| The thermal protector has been activated due to:
a. The device operating outside the conditions outlined in this manual.
b. The device has been operating near heat sources or in hot environments.
Allow the compressor to cool down for a minimum of
40 minutes. Wait until the red light turns off before turning the compressor on again.
The compressor does not operate.| Check all power sources to the compressor.
Push the ON/OFF button again.
Medication does not nebulize properly.| Check for proper connection of the
tubing adapter. Make sure the nebulizer is not empty.
Clean the nebulizer as described in the manual. Replace the nebulizer.
Check the intake filter and replace it if it is dirty.
The device is noisy.| Check that the filter cover is correctly positioned.
The compressor does not operate when connected to battery.| Make sure battery
contacts cover is removed
Check the assembly of the battery pack to compressor
Compressor is disabled when battery is charging. Remove charger from battery to operate the compressor.
Ensure battery pack has adequate charge. Recharge battery.
Short circuit protection may have been activated. This fault condition can be re-set by charging the battery pack.
Specifications
Power supply | See adapter rating label |
---|---|
Weight | 0.93 lbs. |
Size | 4.5” x 3.5” x 2.1” (114.3 mm x 88.9 mm x 53.3 mm) |
Noise level | 50 +/_ 4dBA |
Average flow rate | 3.4 LPM @12.7 psi |
Maximum pressure | 30 psi |
Maximum flow rate | ~ 8 LPM |
Mode of operation | Intermittent use 30 min ON / 30 min OFF |
Class II device (double safety insulation).
Type BF device (device with specific protection against electrical hazards).
Storage conditions
Temperature | MIN -4° F MAX 140° F (MIN -20° C MAX +60° C) |
---|---|
Humidity | MIN 10% RH –MAX 95% RH |
Atmospheric pressure | Performance may vary depending on altitude above sea |
level,
barometric pressure and temperature
Operating conditions
Temperature | MIN 41° F MAX 104° F (MIN +5° C MAX +40° C) |
---|---|
Humidity | MIN 10% RH –MAX 95% RH |
Atmospheric pressure | Performance may vary depending on altitude above sea |
level,
barometric pressure and temperature
Expected service life
Compressor 3 years from date of purchase Battery 6 months from date of
purchase
Replacement parts and accessories
Filter 1109488
AC power/battery charger 1109441
12V DC (car) adapter 1109455
Rechargeable lithium ion battery 1110726
Symbol glossary
- Attention: Read accompanying documents ON (power)
- OFF (power)
- Follow instructions for use
- Temperature limitation
- Serial number
- Alternating current
- Direct current
- Atmospheric pressure
- Separate collection
- Double insulated
- Type BF applied parts
- Indoor use only
- Humidity
- Housing Ingress Protection Rate: IP22 (Protected against solid foreign objects of 12.5 mm Ø and greater; Protected against vertically falling water drops when ENCLOSURE tilted up to 15°)
Optional rechargeable battery
Operating time with battery | Up to 90 minutes |
---|---|
Number of treatments per charge | Up to 12 treatments when nebulizing 2.5 ml of |
medication
Normal time to charge| Approximately 6 hours