Honeywell BW Max XT ll Multi Gas Detector with Pump User Guide
- June 5, 2024
- Honeywell
Table of Contents
BW Max XT ll Multi Gas Detector with Pump
Limited Warranty and Limitation Liability
BW Technologies LP (BW) warrants the product to be free from defects in
material and workmanship under normal use and service for a period of two
years, beginning on the date of shipment to the buyer. This warranty extends
only to the sale of new and unused products to the original buyer. BW’s
warranty obligation is limited, at BW’s option, to refund of the purchase
price, repair or replacement of a defective product that is returned to a BW
authorized service center within the warranty period. In no event shall BW’s
liability hereunder exceed the purchase price actually paid by the buyer for
the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the
normal wear and tear of the product arising from use; b) any product which in
BW’s opinion, has been misused, altered, neglected or damaged, by accident or
abnormal conditions of operation,
handling or use; c) any damage or defects attributable to repair of the
product by any person other than an authorized dealer, or the installation of
unapproved
parts on the product; or The obligations set forth in this warranty are
conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance
with the product manual instructions and any other applicable recommendations
of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly
making the product available for correction. No goods shall be returned to BW
until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as
the original invoice, bill of sale or packing slip to establish that the
product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY
AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT
LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR
BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries
or states do not allow limitation of the term of an implied warranty, or
exclusion or limitation of incidental or consequential damages, the
limitations and exclusions of this warranty may not apply to every buyer. If
any provision of this warranty is held invalid or unenforceable by a court of
competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.
Contacting BW Technologies by Honeywell
USA: 1-888-749-8878
Canada: 1-800-663-4164
Europe: +44(0) 1295 700300
Other countries: +1-403-248-9226
Email us at: info@gasmonitors.com Visit BW Technologies by Honeywell’s website at: www.gasmonitors.com
Introduction
The Quick Reference Guide provides basic information to operate the
GasAlertMaxXT II gas detector. For complete operating instructions, refer to
the GasAlertMaxXT II Technical Reference Guide provided on the CD-ROM. The
GasAlertMaxXT II gas detector (“the detector”) is designed to warn of
hazardous gas levels above user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond
properly to the alarm.
Note
The detector is shipped with English as the default displayed language.
Additional languages provided are French, German, Spanish, and Portuguese. The
screens for the additional languages are displayed on the detector and in the
corresponding operator’s manual.
Zeroing the Sensors
To zero the sensors, refer to steps #1-3 in Calibration on page 9.
GasAlertMaxXT II
Safety Information – Read First
Use the detector only as specified in this operator’s manual and the technical
reference guide, otherwise the protection provided by the detector may be
impaired.
Read the following Cautions before using the detector.
a Cautions · Warning: Substitution of components may impair Intrinsic
Safety. · Caution: For safety reasons, this equipment must be
operated and serviced by qualified personnel only. Read and understand the
technical reference guide completely before operating or servicing. · Charge
the detector before first-time use. BW recommends the detector be charged
after every workday. · Calibrate the detector before first-time use and then
on a regular schedule, depending on use and sensor exposure to poisons and
contaminants. The sensors must be calibrated regularly and at least once every
180 days (6 months).
1
GasAlertMaxXT II
Quick Reference Guide
· Calibrate only in a safe area that is free of hazardous gas in an atmosphere
of 20.9 % oxygen.
· The combustible sensor is factory calibrated to 50% LEL methane. If
monitoring a different combustible gas in the % LEL range, calibrate the
sensor using the appropriate gas.
· Only the combustible gas detection portion of this instrument has been
assessed for performance by CSA International.
· Protect the combustible sensor from exposure to lead compounds, silicones,
and chlorinated hydrocarbons. Although certain organic vapors (such as leaded
gasoline and halogenated hydrocarbons) may temporarily inhibit sensor
performance, in most cases the sensor will recover after calibration.
· BW recommends that the combustible sensor be checked with a known
concentration of calibration gas after any exposure to contaminants/poisons
such as sulfur compounds, silicon vapors, halogenated compounds, etc.
· BW recommends to bump test the sensors before each day’s use to confirm
their ability to respond to gas by exposing the detector to a gas
concentration that exceeds the alarm setpoints. Manually verify that the
audible and visual alarms are activated. Calibrate if the readings are not
within the specified limits.
· Caution: High off-scale LEL readings may indicate an explosive
concentration.
· Any rapid up scaling reading followed by a declining or erratic reading may
indicate a gas concentration beyond upper scale limit, which can be hazardous.
2
· For use only in potentially explosive atmospheres where oxygen
concentrations do not exceed 20.9% (v/v).
· Extended exposure of the GasAlertMax XT II to certain concentrations of
combustible gases and air may stress the detector element that can seriously
affect its performance. If an alarm occurs due to a high concentration of
combustible gases, recalibration should be performed, or if needed, the sensor
replaced.
· The BW pump (XT-RPUMP-K1) is certified for use with the GasAlertMax XT II
only.
· Warning: The lithium battery (MX-BAT01) may present a risk of fire or
chemical burn hazard if misused. Do not disassemble, heat above 212°F (100°C),
or incinerate.
· Warning: Do not use any other lithium batteries with the GasAlertMax XT II
detector. Use of any other cell can cause fire and/or explosion. To order and
replace the MX-BAT01 lithium battery, contact BW Technologies by Honeywell.
· Warning: Lithium polymer cells exposed to heat at 266°F (130°C) for 10
minutes can cause fire and/or explosion.
· Dispose of used lithium cells immediately. Do not disassemble and do not
dispose of in fire. Do not mix with the solid waste stream. Spent batteries
must be disposed of by a qualified recycler or hazardous materials handler.
· Keep lithium cells away from children.
· Calibrations and bump tests can be performed using a 0.5 l/min. regulator or
a demand flow regulator. If the demand flow regulator is used, it must meet
the following maximum inlet pressure specifications:
· Disposable cylinders 0-3000 psig/70 bar · Refillable cylinders 0-3000
psig/70 bar
· Products may contain materials that are regulated for transportation under
domestic and international dangerous goods regulations. Return product in
compliance with appropriate dangerous goods regulations. Contact freight
carrier for further instructions.
Informations sur la sécurité – À lire avant toute chose
Respectez strictement les instructions d’emploi figurant dans ce
manuel d’utilisation ainsi que dans le guide technique, sinon, la
protection offerte par le détecteur pourrait être diminuée.
Lisez les Précautions suivantes avant d’utiliser le détecteur.
a Précautions
· Avertissement: Le remplacement d’un composant de l’appareil peut altérer sa
sécurité intrinsèque.
· Précaution: Pour des raisons de sécurité, cet appareil doit être
exclusivement utilisé et entretenu par du personnel qualifié. Lisez
attentivement le guide technique avant
· d’utiliser l’appareil ou d’en assurer l’entretien, et assurezvous d’en avoir
bien compris les instructions.
· Chargez le détecteur avant sa première utilisation. BW recommande de
recharger le détecteur après chaque journée d’utilisation.
· Étalonnez le détecteur avant sa première utilisation, puis de manière
régulière, en fonction de l’utilisation et de l’exposition du capteur aux
poisons et autres contaminants. Les capteurs doivent être étalonnés
régulièrement et au moins une fois tous les 180 jours (6 mois).
· Veillez à effectuer l’étalonnage dans une zone sûre, exempte
GasAlertMaxXT II
Informations sur la sécurité – À lire avant toute chose
de gaz dangereux, dans une atmosphère contenant 20,9 % d’oxygène.
· Le capteur de gaz combustible est étalonné en usine au méthane à une
concentration de 50% LIE. Si la surveillance porte sur un autre gaz
combustible dans la plage de %LIE, étalonnez le capteur en utilisant le gaz
approprié.
· Sur cet appareil, seule la détection de gaz combustible a fait l’objet d’une
évaluation des performances par CSA International.
· Protégez le capteur de gaz combustible des expositions aux composés de
plomb, aux silicones et aux hydrocarbures chlorés. Bien que certaines vapeurs
organiques (comme l’essence au plomb ou les hydrocarbures halogénés) puissent
neutraliser provisoirement les performances du capteur, dans la plupart des
cas, le capteur retrouvera son fonctionnement normal après étalonnage.
· Si le capteur de gaz combustible a été exposé à des contaminants/poisons
(composés de soufre, vapeurs de silicium, produits halogénés, etc.), il est
conseillé de vérifier son bon fonctionnement en le mettant en présence d’une
concentration connue d’un gaz.
· BW recommande d’effectuer un ” test fonctionnel ” des capteurs, avant chaque
utilisation quotidienne, afin de vérifier qu’ils réagissent bien au gaz,
notamment lorsque le détecteur est exposé à une concentration de gaz qui
dépasse les seuils d’alarme. Vérifiez manuellement que les alarmes sonore et
visuelle sont activées. Étalonnez l’appareil si les lectures sont en dehors
des limites spécifiées.
· Précaution: Des lectures élevées hors échelle peuvent indiquer la présence
d’une concentration explosive.
3
GasAlertMaxXT II
Quick Reference Guide
· Une lecture qui augmente rapidement, puis qui baisse, ou une lecture
fantaisiste peuvent être représentatives d’une concentration de gaz excédant
la limite d’échelle supérieure et risquant donc d’être dangereuse.
· Cet appareil est uniquement destiné aux atmosphères potentiellement
explosives où la concentration d’oxygène ne dépasse pas 20,9 % (v/v).
· L’exposition prolongée du GasAlertMax XT II à certaines concentrations de
gaz combustible et d’air peut fortement éprouver l’élément du détecteur et
nuire gravement à ses performances. En cas d’alarme, suite à une forte
concentration de gaz combustible, il convient d’effectuer un réétalonnage,
voire de remplacer le capteur si nécessaire.
· La pompe BW (XT-RPUMP-K1) est certifiée pour une utilisation avec le
GasAlertMax XT II uniquement.
· Avertissement: La batterie au lithium (MX-BAT01) peut présenter un risque
d’incendie ou de brûlure chimique en cas de mauvaise utilisation. Elle ne doit
jamais être démontée, incinérée ni chauffée au-delà de 100 °C.
· Avertissement: N’utilisez jamais d’autres piles au lithium avec le détecteur
GasAlertMax XT II, au risque de provoquer un incendie et/ou une explosion.
Pour commander une batterie au lithium MX-BAT01 de rechange, contactez BW
Technologies by Honeywell.
· Avertissement: Les piles au lithium polymère exposées à une température
supérieure à 130 °C pendant plus de 10 minutes peuvent provoquer un incendie
et/ou une explosion.
· Mettez immédiatement au rebut les piles au lithium usagées. Ne les démontez
jamais et ne les jetez pas au feu. Ne les
4
mélangez pas aux autres déchets solides. Les piles usagées doivent être
éliminées par un centre de recyclage agréé ou un centre de traitement des
matières dangereuses.
· Gardez les piles au lithium hors de portée des enfants.
· Avertissement: BW Technologies recommande l’utilisation de gaz d’étalonnage
de haute qualité et de bouteilles répondant aux normes nationales. Les gaz
d’étalonnage doivent correspondre à la précision du détecteur.
· Les tests fonctionnels et les étalonnages peuvent être effectués en
utilisant un régulateur 0,5 l/min. ou un régulateur de débit à la demande. Si
vous utilisez le régulateur de débit à la demande, ce dernier doit répondre
aux spécifications de pression d’entrée maximale suivantes :
· Bouteilles jetables 0-3 000 psig/70 bar
· Bouteilles rechargeables 0-3 000 psig/70 bar
· Les produits peuvent contenir des matériaux qui sont réglementés pour le
transport en vertu des règlements nationaux et internationaux de marchandises
dangereuses. Retourner le produit conformément à la réglementation sur les
marchandises dangereuses appropriées. Contactez transporteur pour plus
d’instructions.
Parts of the GasAlertMaxXT II
GasAlertMaxXT II Parts of the GasAlertMaxXT II
Item 1 2 3 4 5 6 7 8 9 10
Description
Visual alarm indicators (LEDs) Pump quick connector Pump filter and moisture
filter Pushbutton Alligator clip Charging connector and IR interface Diffusion
cover locking screw (1) Diffusion cover Audible alarm Liquid crystal display
(LCD)
5
GasAlertMaxXT II Quick Reference Guide Display Elements
6
Item 1 2 3 4 5 6 7 8 9
Description
Alarm condition Automatically zero sensor Numeric values Battery life
indicator Pump indicator Heartbeat indicator Gas type identifiers Gas cylinder
Automatically span sensor
Pushbutton Pushbutton
C
GasAlertMaxXT II Pushbutton
Description · To activate the detector, press C. · To deactivate the detector,
press and hold C until the OFF countdown is complete. · To view the date/time,
TWA, STEL, and maximum (MAX) readings, press C twice rapidly. To clear
the TWA, STEL, and MAX readings, press C when the LCD displays CLEAR ALL. · To
initiate calibration, press and hold C while the detector performs the OFF
countdown. Continue
holding C while the LCD briefly deactivates and then reactivates to begin the
CAL countdown. Release C when the CAL countdown is complete. · To activate the
backlight, press C and release. · To acknowledge latched and pump alarms,
press C. · To acknowledge a low alarm and disable the audible alarm, press C.
The Low Alarm Acknowledge option must be enabled in Fleet Manager II. · To
acknowledge any of the Due Today alarms (calibration, bump test, block test)
and disable the audible alarm, press C.
7
GasAlertMaxXT II Quick Reference Guide Calibration and Bump Test Installation
8
Calibration
a Caution
Calibrate only in a safe area that is free of hazardous gas in an atmosphere
of 20.9 % oxygen. Do not calibrate the detector during or immediately after
charging is complete.
Note
The maximum hose length for calibration is 3 ft. (1 m).
1. Press and hold C as the detector performs the OFF countdown. Continue to hold C when the detector briefly deactivates.
2. The detector activates again and performs the CAL countdown. Continue to hold C until the countdown is complete to enter calibration.
flashes while the detector
zeroes all of the sensors.
If a sensor fails to auto zero, it cannot be spanned. When auto zero is complete, the LCD displays APPLY GAS.
GasAlertMaxXT II Calibration
4. Refer to Calibration and Bump Test Installation (page 6). Attach the 0.5
l/min. regulator or the demand flow regulator and apply gas.
Note
The diffusion cover must be attached to the detector to calibrate.
K flashes on the LCD. After a sufficient amount of gas has been detected (30
seconds), the detector beeps.
flashes while the detector completes the span.
5. The LCD displays CAL DUE. Next, a screen displays showing the number of
days remaining before calibration is due for each sensor. Last, the LCD
displays (e.g.) CAL DUE 180 d. As some sensors require more frequent
calibrations, the LCD displays the earliest calibration that must be
performed.
Note
Calibration can be aborted at any time. To abort calibration, press C. The CAL
ABORTED screen displays.
9
GasAlertMaxXT II Quick Reference Guide
Alarms
Refer to the following table for information about alarms and corresponding
screens.
Alarm
Screen
Alarm
Low Alarm
· Slow siren · Slow alternating flash · L and target gas bar flash · Vibrator
alarm activates
TWA Alarm
· Fast siren · Fast alternating flash · L and target gas bar flash · Vibrator
alarm activates
High Alarm
· Fast siren · Fast alternating flash · L and target gas bar flash · Vibrator
alarm activates
STEL Alarm
· Fast siren · Fast alternating flash · L and target gas bar flash · Vibrator
alarm activates
Multi-Gas Alarm
· Alternating low and high alarm siren and flash
· L and target gas bars flash · Vibrator alarm activates
Over Limit (OL) Alarm · Fast siren · Fast alternating flash · L and target gas bar flash · Vibrator alarm activates
Screen
Note If the Low Alarm Acknowledge option is enabled in Fleet Manager II, the
low alarm can be acknowledged and the audible alarm deactivated by pressing C.
If the alarm escalates to a high, STEL, TWA, or multi-gas alarm, the audible
alarm reactivates.
10
GasAlertMaxXT II Alarms
Alarm Sensor Alarm · ERR displays
Screen
Low Battery Alarm
· Sequence of 10 rapid sirens and alternating flashes with 7 seconds of silence in between (continues for 10 minutes)
·
and L flashes
· LOW BATTERY displays and the vibrator alarm activates
· After 10 minutes, a sequence of 10 rapid sirens and alternating flashes with 1 second of silence in between (reactivates seven times)
· TURNING OFF displays before the detector deactivates
Alarm
Automatic Deactivation Alarm · Sequence of 10 rapid sirens and alter-
nating flashes with 1 second of silence in between (reactivates seven times) ·
L flashes and the vibrator alarm activates
· TURNING OFF displays before the detector deactivates
Pump Alarm · Two beeps and two flashes · J and L flashes
· HIGH displays · The vibrator alarm activates
Screen
Note:
If enabled, during an alarm condition the Latched Alarms option causes the low
and high gas alarms (audible, visual, and vibrator) to persist until the alarm
is acknowledged (by pressing C) and the gas concentration is below the alarm
setpoint. The LCD continues to display the high peak concentration until the
alarm condition no longer exists. Enable/disable Latching Alarms in Fleet
Manager II. Local regulations may require the Latching Alarms option be
enabled.
11
GasAlertMaxXT II Quick Reference Guide
Options Menu
The detector, IR Link adapter, and Fleet Manager II software are required to
define options. Refer to the GasAlertMax XT II Technical Reference Guide and
Fleet Manager II Operator’s Manual. For user and sensor options, refer to the
following:
Device Configuration
· Startup Message Top Line: Enter a line of text to display on the LCD during
startup (maximum 25 characters).
· Startup Message Bottom Line: Enter a line of text to display on the LCD
during startup (maximum 25 characters).
· Lockout on Self-Test Error (sensor alarm lock): If a sensor fails during
startup and the Lockout on Self-Test option is enabled, Safety Lock On
displays on the LCD and the detector deactivates.
· Safe Mode: If enabled, SAFE displays continuously on the LCD unless an alarm
condition occurs.
· Confidence Beep: If enabled, the confidence beep provides continuous
confirmation that the detector is operating correctly. To define how often the
detector beeps (every 1-120 seconds), enter the value in the Confidence
Interval field. Confidence beep is automatically disabled during a low battery
alarm.
· Latching Alarms: Enable to ensure an alarm persists until the alarm is
acknowledged and the gas concentrations are below the alarm setpoint. The
audible alarm can be temporarily deactivated for 30 seconds by pressing C, but
the LCD continues to display the high peak concentration until the alarm
condition no longer exists.
· Force Calibration: If enabled, the detector must be calibrated if a sensor
is overdue upon startup. User defined (0365 days) in the Calibration Interval
field.
· Cal IR Lock (must use IR device to calibrate): If enabled, the detector
automatically auto zeros, but the sensors must be
12
spanned using the IR Link or MicroDock II station with Fleet Manager II. ·
Force Bump: If enabled, a bump test must be performed to ensure the sensor(s)
are responding correctly to the test gas. User defined (0-365 days) in the
Bump Interval field. · Location Logging: If enabled, a series of numbers
(1-999) can be entered on the detector to identify gas wells, plants, and
other areas that identify the location where the detector is being used. ·
Force Block Test: If enabled, a pump block test must be performed during the
startup tests. · Datalog Interval (seconds): Define how often the detector
records a sample (every 1-120 seconds). · Confidence Interval (seconds):
Define how often the detector beeps (1-120 seconds) when the Confidence Beep
option is enabled. · Language: The LCD displays the screens in English,
Français (French), Deutsch (German), Español (Spanish), or Português
(Portuguese). Select the language from the drop-down menu in Fleet Manager II.
Sensor Configuration (H2S, CO, LEL, and O2)
· Sensor Disabled: Enables/disables the selected sensor. · Calibration Gas
(ppm): Define the span gas concentration
for each sensor. The span gas concentration must match the span value on the
gas cylinder. · Calibration Interval (days): Define the number of days (0365)
when the next calibration is due. · Bump Interval (days): Define the number of
days (0-365) when the next bump test is due. · Low Alarm (ppm): Define the low
alarm setpoint for each sensor. · High Alarm (ppm): Define the high alarm
setpoint for each sensor.
· TWA Alarm (ppm): Define the time-weighted average (TWA) alarm setpoint
(toxic sensors only).
· STEL Alarm (ppm): Define the short-term exposure limit (STEL) alarm setpoint
(toxic sensors only).
· STEL Interval (minutes): Define the short-term exposure limit (STEL) from
5-15 minutes (toxic sensors only).
· TWA Period (hours) (TWA moving average (hours): The TWA Period option is
used to define a time-weighted moving average of accumulated gases over a
period of 4-16 hours, to ensure the worker leaves the area when the defined
maximum average is accumulated.
· Correction Factor (%): Enter the compensation factors for hydrocarbons other
than methane. The factor can only be applied if the LEL sensor has been
calibrated with methane (LEL only).
· 50% LEL = (%CH4): Enter a percentage value to display the LEL reading as
%vol., assuming a methane environment (LEL only).
· Auto-Zero on Startup: Enable/disable the detector to automatically zero the
sensor(s) during startup (H2S, CO, LEL, and O2).
· LEL by Volume CH4: If enabled, the detector operates assuming a methane
(CH4) calibration. Enable to read and display %CH4 values. Disable to read and
display %LEL values.
· 5% LEL Over-span: If enabled, the detector automatically over-spans the LEL
sensor by 5% LEL above the span gas concentrations to ensure the detector is
in compliance with CAN/CSA C22.2 No. 152.
· 20.8% Base Reading: When enabled, the detector is configured to detect 20.8%
O2 as ambient air. When disabled, the detector is configured to detect 20.9%
O2 as ambient air.
· Low Alarm Acknowledge: If enabled, the audible alarm can be temporarily
disabled during a low alarm
GasAlertMaxXT II Maintenance
by pressing C. The vibrator, alarm LEDs, and LCD remain operational (toxic and
LEL only).
Maintenance
To maintain the detector in good operating condition, perform the following
basic maintenance as required.
· Calibrate, bump test, and inspect the detector on a regular schedule.
· Maintain an operations log of all maintenance, bump tests, calibrations, and
alarm events.
· Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or
polishes.
· Do not immerse the detector in liquids. Charging the Detector
a Warning Charge only in a safe area that is free of hazardous gas within
temperatures of 32°F-113°F (0°C-45°C).
Charge the battery after each workday. To charge the battery, refer to the
following procedures and illustration:
1. Deactivate the detector. 2. Insert the charging adapter plug into an AC
outlet. 3. Connect the charging adapter to the detector IR
receptacle. Refer to the following illustration. 4. Allow the battery to
charge for 6 hours.
13
GasAlertMaxXT II Quick Reference Guide
5. To reach full battery capacity, allow a new battery to fully charge and
discharge three times.
Replacing a Sensor or Sensor Filter
a Warning
To avoid personal injury and/or property damage, only use sensors that are
specifically designed for the detector.
Note
Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor
in one of the four sensor locations.
To replace a sensor or sensor filter, refer to the following figure, table and
procedures #1-13.
14
GasAlertMaxXT II Maintenance
15
GasAlertMaxXT II Quick Reference Guide
Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14
Description Front shell LEL sensor PCB PCB screws (2) Pump Rear shell Machine screws (6) Pump inlet Pump filter (particulate) Moisture filter CO sensor H2S sensor O2 sensor Sensor filter
1. Deactivate the detector. 2. Remove the pump inlet screw and the pump inlet. 3. Remove the six machine screws from the rear shell.
16
4. Because the pump hose is connected to the rear shell and front shell pump,
carefully remove the rear shell by lifting upward and tilting to the left.
Both the rear and front shell are laying flat side by side.
5. Remove the two PCB screws. 6. Lift the PCB upward and tilt to the left.
Lay the PCB
(sensors facing up) onto the rear shell.
7. Slide the LEL, CO, and H2S sensor(s) outward to remove. To remove the
oxygen sensor, gently insert a screwdriver to the back of the oxygen sensor to
push out the sensor.
8. Insert the new sensor(s) and replace the sensor filter. 9. Reassemble the
detector and replace the two PCB
screws. 10. Replace the six machine screws. 11. If required, replace the pump
and moisture filter. 12. Replace the pump inlet and the pump inlet screw.
GasAlertMaxXT II Specifications
13. Activate the detector and then calibrate the new sensor(s). Refer to
Calibration.
Specifications
Instrument dimensions: 13.1 x 7.0 x 5.2 cm (5.1 x 2.8 x 2.0 in.) Weight: 328 g
(11.6 oz.) Operating temperature: -20°C1 to +50°C (-4°F to +122°F) Storage
temperature: -40°C1 to +60°C (-40°F to +140°F) Operating humidity: 10% to 100%
relative humidity (non-condensing) Dust and moisture ingress: IP66/IP67 Alarm
setpoints: May vary by region and are user-defined Detection range: H2S: 0 –
200 ppm (1 ppm increments) CO: 0 – 1000 ppm (1 ppm increments) O2: 0 – 30.0%
vol. (0.1% vol. increments) Combustible (LEL): 0 – 100% (1% LEL increments) or
0-4.4% v./v CH4
1. Standard release data. This product has extended characteristics. For
specifications refer to EAC Ex certificate and Pattern Approval document.
17
GasAlertMaxXT II Quick Reference Guide
Sensor type: H2S, CO, O2: Single plug-in electrochemical cell Combustibles:
Plug-in catalytic bead
O2 measuring principle: Capillary controlled concentration sensor
Alarm conditions: TWA alarm, STEL alarm, low alarm, high alarm, multi-gas
alarm, over limit (OL) alarm, low battery alarm, confidence beep, automatic
deactivation alarm, and pump alarm
Audible alarm: 95 dB+ at 30 cm variable pulsed beeper with full battery charge
Visual alarm: Red light-emitting diodes (LEDs)
Display: Alphanumeric liquid crystal display (LCD)
Backlight: Activates upon startup and when the pushbutton is pressed;
deactivates after 10 seconds. Also activates during an alarm condition and
remains lit until alarm ceases
Self-test: Initiated at activation and tests continuously
Calibration: Automatic zero and automatic span
User field options: Startup message, lockout on self-test error, safe mode,
confidence beep, latching alarms, force calibration, cal IR lock, force bump,
location logging, force block test, set datalog interval, set confidence
interval, language selection
Sensor options: Sensor enable/disable, set span calibration values, set
calibration interval, set bump interval, set alarm setpoints, set STEL
interval, set TWA period, auto zero at
18
startup enable/disable, 5% over span, low alarm acknowledge, oxygen measurement, and combustible gas measurement
Battery operating time: One rechargeable lithium polymer: 13 hours (typical)
Year of manufacture: The detector’s year of manufacture is determined from the
serial number. The second and third num-
ber after the second letter determines the year of manufacture.
E.g., MA 110-000001 = 2010 year of manufacture
Approved batteries:
North America Approved batteries for GasAlertMax XT II product: Lithium-ion
polymer battery as per standards EN50020, UL913,
C22.2 No. 157
Rechargeable battery (MX-BAT01)
Temperature code
Lithium polymer
-20°C Ta +50°C
T4
Battery charger: charging adapter
First-time charge: 6 hours
Normal charge: 6 hours
Warranty: 2 years including sensors
Approvals: Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152
ANS/UL – 913 and ANSI/ISA – S12.13 Part 1
CSA ATEX
IECEx
INMETRO KTL IP
Class I, Division 1, Group A, B, C, and D CE 0539 g II 1 G Ex da ia IIC T4 Ga
KEMA 08 ATEX 0001 EN 60079-0, EN 60079-1, EN 60079-11 Ex da ia IIC T4 Ga CSA
07.0012 IEC 60079-0, IEC 60079-1, IEC 60079-11 Ex ia IIC T4 Ga DNV 12.0135
12-KB4BO-0054a IP66/IP67
a. http://www.honeywellanalytics.com/~/media/honeywell-analytics/products /gasalertmax-xt-ii/documents/certifications/ koreanexcertificate_bw_gamaxxtii_12kb4bo0054.pdf?la=en
GasAlertMaxXT II Specifications
try to correct the interference by one or more of the following measures:
· Reorient or relocate the receiving antenna. · Increase the separation
between the equipment and receiver. · Connect the equipment into an outlet on
a circuit different from
that to which the receiver is connected. · Consult the dealer or an
experienced radio/TV technician for
help.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian
EMI requirements. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to
19
GasAlertMaxXT II Quick Reference Guide
20
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>