NEUATION iTHERM D150-1 Single Block Dry Bath User Manual

June 5, 2024
NEUATION

iTHERM D150-1
PRODUCT USER MANUAL

INTRODUCTION

This manual provides important safety information for this Dry Bath. It should be kept near the equipment for quick & easy reference. The Dry bath is equipped with a better display for safe operations. This dry bath has range from ambient + S°C to 120°C making it useful for a wide range of applications. The probe is provided for monitoring the accurate temperature of samples.

INTENDED USE
The dry bath is intended to use for specific applications such as Enzymatic/kinetic reactions Denaturation of DNA/RNA and protein samples, Labeling of nucleic acids and proteins, quantification assays, Lysis reactions, Immunol precipitation, etc.
NOTE: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and Notts repair. For repair please contact the supplier.

FEATURES

  • Dry heating block with temperature control range from ambient +5C to 120C
  • Aluminum-made blocks are resistant to chemical spillage & temperature shocks.
  • 1 heating block with a long-timer range from lmin- 19hr.59mins & infinite mode
  • A clear display shows critical parameters.
  • Can be used as a bead bath &water bath
  • Various types of tube capacity with convenient lifting gear for cleaning & disinfecting.
  • Rapid heating & constant temperature across the blocks throughout the operation.
  • Build-in safe temperature protection

STANDARD ACCESSORIES

  • Power adapter -Input voltage 100-240 VAC S0/60Hz, (Attached to the unit)
    Output voltage 24V ****4A

  • Product user manual & warranty card

  • Power cord, Dry block, Dry block lifter

TECHNICAL SPECIFICATIONS

Temperature Range Ambient +5 to 120C
Temperature Uniformity ±0.5C
Temperature Stability ±0.5°€
Set-up Block material Aluminum alloy
Number of Blocks 1 Nos
Display Digital
Timer 1min – 19 hr:59mins & infinite
Dimension with Stack Lid 190 x 87 x 145 mm
Input Voltage 100-240 VAC, 50 / 60Hz
Power Consumption 85W

SAFETY PRECAUTIONS

  • Never use the instrument in any manner not specified in this manual or by the manufacturer might result in the laps of warranty.
  • Repairs must only be performed by an authorized service technician.
  • Only use recommended original spare parts for best result & product safety.
  • If liquids are spilled in blocks, the instrument must be cleaned carefully and properly before being used again.
  • Prior to operation, the tubes should be visually inspected for material damage.
  • Damaged tubes must not be used as this can result in sample loss and the contamination of the product.
  • This product must be used for specified applications only. It must not be operated in a hazardous or flammable environment and must not be used to mix explosive or highly reactive substances.
  • Do not place the potentially hazardous material within the clearance area/ envelope.
  • Wear your personal protective equipment in accordance with the hazards category of the medium to be processed. There is risk of liquid splashing.
  • Do not move the instrument while it is operating or connected to the main power supply.
  • Do not touch the Block surface when the temperature of the dry bath is over 50°C, this could result in serious burns or injury. Pay attention to the residual heat after switching off.
  • Properly lift the device with both hands while moving or installing. Also, the device should only be moved from its position once it attains the room temperature.
  • Ensure that only closed tubes are used in thermal reactions.
  • Do not lean on the equipment. It may damage the equipment or harm the operator.
  • Do not fill tubes near the device. Liquid spillage may harm the device.
  • In the event of contamination caused by aggressive agents, the device must be cleaned immediately using a natural cleaning liquid. If any damage is seen, contact the service technician.
  • Only process media which will not react dangerously to the extra energy produced by mixing.
  • Do not operate the appliance in an explosive atmosphere with hazardous substances or underwater.

INSTALLATION

Gently remove the upper packaging and take out the dry bath by holding it from the bottom. When this equipment is used for the first time, ensure that all the packaging accessories are removed from the product.
LOCATION & MOUNTING
Place the device on a flat and leveled surface and ensure that all four feet stand on the surface firmly. Avoid installing on a slippery surface or surface prone to vibration.

  1. The ideal ambient temperature is 20°C ± 5°C; avoid placing the device in direct sunlight.
  2. Keep a clearance of at least 30 cm from all sides to guarantee cooling efficiency.
  3. Keep this away from heat or water to avoid sample temperature issues.
  4. Do not place the device in such a place that it becomes difficult to operate.
  5. Lid extended can be used with 15 and 50m1 blocks to increase the height of the dry bath ensuring the closure of the lid to maintain the Temp. uniformity

REMOVING & INSTALLING BLOCKS

REMOVING BLOCKS

Device blocks can be removed/ changed by a simple screwing device into the blocks for lifting/removing/changing the blocks. The above image shows the process to remove blocks.
INSTALLING BLOCK
Follow the exact reverse process to insert the blocks back into the dry bath. The above image shows the step-by-step procedure to replace the blocks back in the unit.

USER INTERFACE & DISPLAY

CONTROL PANEL OPERATION

Buttons Function
Start /stop button for operation start and stop.
1. Probe/ plate mode selection (long press standby condition)

2. Temp/ time change (single press)
| (+)and the (-) button will increase and decrease the value of the TEMPERATURE/TIME and Prag command.

All LEDs will blink as per the status of command.

OPERATION SWITCH ON THE DEVICE

After connecting the power cord, switch ON the main power supply. Make sure that tubes are placed properly in the wells.
10.1 START/ STOP OPERATION

  • Single Press this button to “START” Operation (as per set Time, Temp & Mode).
  • With the start of the operation, the temperature will increase on display. Press this button again to “STOP”

10.2 MODE

  • When we select the “MODE” button we can enable Time and Temperature Mode alternate. To change its value, Press the (+)and (-) buttons.
  • You can set time from 1 min to 19 hr: 59 min or infinite” ][“Mode. The TIME value will increase by 1 min by a single pressing (+) button.
  • After adjusting temp & time, the value will be saved after 5 times blink and then it will show you the home screen alternatively.

2.1 HEATING FUNCTION

  • You can Set the Temperature from Ambient + 5°Cto 120°C.
  • The temperature value will increase by 0.1 °C by the single press ( +) button & decrease with the (-) button. The temperature remains in CELSIUS Unit.
  • With the start of the unit, the display shows the value for the temperature function default. By pressing the (+)or(-) button, the temp Set value increases or decreases. Set Value blinks for 5 times & save the value. After that, you will see “Room temperature” on-screen.
  • For fast increment and decrement of temperature setting, long press (+)or (­ -) button continuously. (after setting the desired value in the window, the value will blink 5 times and then the blink will stop, It means the value is saved.

Note: The heating  function works only when the set temperature is above room temperature

2. 2 PROBE SELECTION

  • By Long pressing the MODE Key (about 1.5 sec), the Probe mode will be activated.
  • Probe LEDwill flash with the probe mode activation.
  • Without probe mode, the device will automatically work in plate mode. Users can change Plate/Probe mode by long-pressing the MODE button only at standby

2. 3 TIME FUNCTION

  • By pressing the MODE button the time led will flash. It will show the Time
    command activation and we can change time parameters in this situation.

  • “19.59”(19hr:59min) value will display on the window at first-time initiated machine

  • To set the Time value, press(+) and (-) for increment and decrement of Value from 00.01 to 19hr:59 min. After set desired value in the window, the Value will blink 5 times and then stop, it means the value is saved.

  • The time value will increment in 1 minute with a single press. The Time unit will be in (hour. minute) as shown upon display.

  • In running conditions, if time changes, the timer will reset & count from the
    After 19hr:59 mins, there is an infinite mode which is shown as “It “.

CALIBRATION (PROBE MODE)

  • Press (MODE) and ( + ) buttons together After the Set temperature is reached in a display of Unit
  • Set reference temperature as per master instrument temperature, the display will blink 5 times and the value is saved.

Example: set value is 100 reference value is 101.2. then set 101.2 in calibration mode and calibration is done once the value is saved.

  • In case of reset of calibrated value, press (Mode) and(+) together and change the value of “set temperature”

Note: In Probe calibration for the perfect result it is suggested to calibrate twice.

CLEANING & MAINTENANCE

  • As the dry bath has no mechanical moving part, it requires very little maintenance and cleaning.
  • Cleaning should be done with mild detergent and water
  • Unplug the unit from its power source before any cleanup process. In case of any spillage allow the device to cool before any cleanup or removal of blocks.
  • The holes of the blocks should be regularly cleaned with a damp cloth to ensure the tube be well contacted with the wall to have good heat
  • Use a slight Damp cloth for the cleaning process.
  • Wipe the exterior of the machine with a Damp Cloth.
  • Cover the samples with the device lid over the block, this cover provides thermal insulation and reduces the water condensation.

Note: User is solely responsible for decontamination of the unit in case of spillage of hazardous material.

TROUBLESHOOTING

Error code Cause Effect Solution
Er 01 Break in the safety circuit Heating off Switch off the device & allow

it to cool down. Otherwise to be carried by an authorized service person
Error code| Cause| Effect| Solution
---|---|---|---
Er 02| Heating not progress| Heating off| Switch off the device. Otherwise to be carried by an authorized service person
Er 03| NTC probe sensor failed| Heating off| Switch off the device. Only to be carried by an authorized service person
Er 04| Probe not in liquid medium| Heating off| Put the probe in liquid medium
Er 05| Heater NTC 100k failed| Heating off| Switch off the device. Only to be carried by an authorized service person

WARRANTY STATEMENT

This product is warranted to be free from defects in material and workmanship for a period of two (2) years from the date of purchase. Your product will be duly repaired upon prompt notification in compliance with the following conditions: This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in this instruction manual. This warranty does not cover damage caused by accident, neglect, misuse, improper service, natural forces or other causes not arising from defects in original material or workmanship. This warranty does not cover any incidental or consequential damages, commercial loss or any other damages from the use of this product. All items returned for service should be set postage prepaid in the original packaging or another suitable carton, added to avoid damage. This warranty is valid only if the warranty is registered with the supplier within 30 days from the date of purchase.

For your reference, make a note of the serial number, date of purchase and supplier here.
Serial No. Purchase Date
Supplier

PRODUCT DISPOSAL

In case the product is to be disposed of, the relevant legal regulations are to be observed. Information on the disposal of electrical and electronic devices in the European Community. The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2012/19/EU on waste electrical and electronics equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business to a business sphere, to which this product is assigned, may no longer be disposed of in the municipal or domestic waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country please contact your supplier if necessary.

Negation Technologies Pvt. Ltd.
Plot No. 15, GIDC Electronic Park SEZ Kolavada Road,
Gandhinagar — 382026, Gujarat, India
Website: www.neuation.com