AMPROBE ULD-400 Ultrasonic Leak Detector User Manual
- June 5, 2024
- AMPROBE
Table of Contents
- AMPROBE ULD-400 Ultrasonic Leak Detector
- Limited Warranty and Limitation of Liability
- PRECAUTIONS AND SAFETY MEASURES
- INTRODUCTION
- KIT COMPONENTS
- KIT COMPONENTS
- KIT COMPONENTS
- MAIN APPLICATIONS
- MAIN APPLICATIONS
- MAINTENANCE
- SPECIFICATIONS
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
AMPROBE ULD-400 Ultrasonic Leak Detector
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for
one year from the date of purchase unless local laws require otherwise. This
warranty does
not cover fuses, disposable batteries or damage from accident, neglect,
misuse, alteration, contamination, or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other warranty on the
behalf of Amprobe. To obtain service during the warranty period, return the
product with proof of purchase to an authorized Amprobe Service Center or to
an Amprobe dealer or distributor. See Repair Section for details. THIS
WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES – WHETHER EXPRESS, IMPLIED
OR STATUTORY – INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR
PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR
LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of incidental
or consequential damages, this limitation of liability may not apply to you.
Repair
All Amprobe returned for warranty or non-warranty repair or for calibration
should be accompanied by the following: your name, company’s name, address,
telephone number, and proof of purchase. Additionally, please include a brief
description of the problem or the service requested and include the test leads
with the meter. Non-warranty repair or replacement charges should be remitted
in the form of a check, a money order, credit card with expiration date, or a
purchase order made payable to Amprobe.
In-warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting
repair. During the warranty period, any defective test tool can be returned to
your Amprobe distributor for an exchange for the same or like product. Please
check the “Where to Buy” section
on amprobe.com for a list of distributors near you. Additionally, in the
United States and Canada, in-warranty repair and replacement units can also be
sent to an Amprobe Service Center (see address below).
Non-warranty Repairs and Replacement – United States and Canada
Non-warranty repairs in the United States and Canada should be sent to an
Amprobe Service Center. Call Amprobe or inquire at your point of purchase for
current repair and replacement rates.
PRECAUTIONS AND SAFETY MEASURES
SYMBOLS
Safety information
The product complies with:
- IEC 61326-1
CENELEC Directives
The instrument conforms to CENELEC Electromagnetic compatibility directive
2014/30/EU.
WARNINGS AND PRECAUTIONS
- Not for use on explosive gases.
- Use extreme care when using near pressurized air/gas.
- Use extreme care when using near rotation equipment.
- Use extreme care when using near electrical equipment.
- Use only 4 x AA batteries for the ULD-400-R Receiver and only 2 x AAA size batteries for the ULD-400-T Transmitter, properly installed in the battery compartment, to power the Product (see Section 5: Maintenance).
- Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 122 °F (50 °C). If the batteries are not removed, battery leakage can damage the Product.
- Follow all battery care from the battery manufacturer.
INTRODUCTION
Ultrasonic sound, or ultrasound, is a sound wave with frequencies above 20
kHz, higher than the upper audible limit of human hearing. Ultrasound can be
generated when turbulence created by air or gas is forced through a small
orifice. Leaking air or gas is generally considered to be viscous flow, and as
the flow velocity increases, the frequency of the ultrasound emitted becomes
higher. Vibrating, moving objects or electric discharge will also create an
ultrasonic wave, which is very directional in nature and can be used to
pinpoint the exact location of a leak, vibration or discharge.
The ULD-400-R Receiver detects ultrasounds within 20 kHz to 90 kHz frequency,
then amplifies and converts these ultrasonic sounds to frequencies and levels
that the human ear can hear through headphones and show it on the LCD screen.
The 20 kHz to 90 kHz frequency range is the optimal range for detecting a
variety of leakage events in assets such as HVAC systems and pneumatic lines.
A change in the ultrasound produced by an asset may be indicative that an
asset is beginning to fail.
KIT COMPONENTS
Kit Components
Your shipping box should include:
| ULD- 410| ULD- 420
---|---|---
ULD-400-R Receiver| 1| 1
ULD-400-T Transmitter| –| 1
Headphones| 1| 1
Earbuds (for use with hard hat)| 1| 1
PB-1 Power Parabola| 1| 1
TEA-1 Flexible Tubing Adapter| 1| 1
TE-1 Tubular Extension| 1| 1
CC-ULD-400 Hard Carrying Case| 1| 1
AA Batteries (Receiver)| 4| 4
AAA Batteries (Transmitter)| –| 2
Manual| 1| 1
Note: Batteries are not pre-installed in the Receiver or Transmitter.
KIT COMPONENTS
KIT COMPONENTS
ULD-400-T Transmitter
The ULD-400-T Transmitter is included with the ULD-420 kit and is optional for
the ULD-410 kit.
When a leak is not pressurized sufficiently the Receiver will not be able to
detect it. In these circumstances, the ULD-400-T Transmitter can be used to
emit an ultrasonic sound that the Receiver can read. The Transmitter is
programmed with three signal levels for precise pinpointing of leaks
.
Accessories
The ULD-400 comes supplied with additional Receiver accessories that are
helpful in leak identification. Plug the headphones into the Receiver to
audibly hear the leak and verify its source (for example hissing sound of an
air leak versus ticking sound of an electric discharge). Use the Parabola
attachment in situations where there is a high level of background noise to
help direct the ultrasound towards the sensor. Use the Tubular Extension with
the Adapter in hard to reach areas for additional reach.
Note: There is no speaker on the Receiver. Without headphones, no noise
will be audible.
MAIN APPLICATIONS
Using the ULD-400-R Ultrasonic Leak Detector Receiver
- Turn on the Receiver and plug the headphones into the jack located on the front of the Receiver. Any standard set of 3.5 mm jack headphones are compatible.
- Before moving to the target area, press “+” or “-” sensitivity buttons to adjust signal strength sensitivity to the highest possible level where bargraph still shows either 0 or a value close to 0. If signal strength cannot be adjusted down and the LCD still shows a maximum value regardless on sensitivity adjustments, press the Filter button.*
- Scan the target area with the microphone sensor.
- As you move nearer to the source of the leak, vibration or electric discharge, the signal strength will increase. This will be indicated on a screen with increasing signal strength number and level of the bargraph.
- The bargraph is a relative measurement only, so when the signal strength reaches maximum, lower the sensitivity by pressing the “-” sensitivity button until the displayed signal strength is less than 75. Repeat this process until you have isolated the source of the ultrasound.
- The audible sound emitted via headphones will help to verify the source of the leak, for example hissing sound of the air leak versus ticking sound of the electric discharge. The Receiver screen alone will not provide an indication of the leak source.
Note:
- For surroundings with a high level of background noise use the Parabola (PB-1) to direct the ultrasound towards the sensor.
- For locations that you cannot point the Receiver directly at the leak, the Tubular Extension can be used (TE-1 with the TEA-1 Adapter).
MAIN APPLICATIONS
*ULD-400-R Filter Operation
In some situations, there might be strong ultrasonic noise generated by
running machinery, motion sensors or other equipment. This noise will cause
the Receiver to read the maximum signal strength on the display regardless of
the sensitivity settings and make it unusable for detecting leaks.
In such cases, press the “Filter” button. The firmware will automatically
detect the main noise frequency and filter it out by applying digital non-pass
bandwidth filter within +/- 5 kHz. If the signal level does not decrease down
close to 0, then the firmware will automatically repeat the process for other
noise frequencies. Up to three frequencies can be filtered out automatically.
The filter icon will be displayed on the screen and will indicate the number of
filters applied, from zero (no noise detected) to a maximum of three.
Note: The filter will be applied to the visual signal indication on a
screen (signal strength number and bargraph). The filter will NOT be applied to
the audio frequency in order to preserve original sound of the leak and allow
the user to better determine the source of the leak.
Note: When working with the Transmitter, make sure to apply the filter on
the Receiver before the Transmitter is turned on, or in an area where the
Transmitter signal can not be detected. Otherwise, the Transmitter signal
frequency will be filtered out and Receiver will not be able to pick it up.
Using the ULD-400-T Ultrasonic Leak Detector Transmitter
The Transmitter allows for ultrasonic detection of openings in locations where
there is no gas or air pressure, or the pressure is not sufficient to detect
the leak with the Receiver alone.
Typical applications include verifying tightness or pinpointing places of air,
water or gas leaks in:
- Fluid or gas tanks
- Building windows, doors or roofs
- Car windows and windshields
- HVAC vents
- Refrigerant pipelines
The process involves placing the Transmitter inside an object (such as tank, house or a car), sealing the entryways and scanning the object from the outside with the Receiver to verify tightness and pinpoint potential leaks.
- Turn on the Transmitter.
- Select the output signal level using the “+” or “-” signal level adjustment buttons. Note: High setting is the default setting. For most applications and specifically for large objects this setting is particularly useful. The High signal may cause the Receiver to read maximum value away from the place of the leak even when lowest sensitivity was selected. Press “-” once to get to Medium signal and press “-” again to get to Low signal strength. Adjust signal level to allow for more precise pinpointing of the leaks.
- Place the Transmitter inside the object to be verified and make sure entryways are sealed.
- Preform leak detecting with the Receiver as described in section 4.1.
MAINTENANCE
Changing the Receiver Batteries
The ULD-400-R uses four 1.5 V AA (LR6) batteries (supplied). To replace the batteries, follow these steps:
- Make sure that the Receiver is turned off.
- Use a screw driver to unscrew the captive screw.
- Remove the battery cover.
- Replace the batteries as shown in Figure 5.1. Observe the battery polarity shown in the battery compartment.
- Replace the battery cover and secure it with the provided screw.
Changing the Transmitter Batteries
The ULD-400-T uses two 1.5 V AAA (LR03) batteries (supplied with ULD-420). To
replace the batteries, follow these steps:
- Make sure that the Transmitter is turned off.
- Loosen the battery-door lock with a flat screwdriver.
- Remove the battery cover.
- Replace the batteries as shown in Figure 5.2. Observe the battery polarity shown in the battery compartment.
- Replace the battery cover to the lock position.
Cleaning
The only maintenance the ULD-400 requires is inspection and cleaning.
Periodically wipe the exterior with a mild solution of detergent and water.
Apply sparingly with a soft cloth and allow to dry completely before using. Do
not use aromatic hydrocarbons, gasoline or chlorinated solvents for cleaning.
SPECIFICATIONS
Features | ULD-400-R | ULD-400-T |
---|---|---|
Sensitivity Adjustment | Yes | N/A |
Volume Adjustment | Yes | N/A |
Signal Level Adjustment | N/A | Yes |
__
Earphone Jack
| Yes
(compatible with 3.5 mm audio jack)
| N/A
Display Size| LCD 2.5 in (6.35 cm)| N/A
Display Dimensions| 1.45 x 1.93 in (36.72 x 48.96 mm)| N/A
Display Resolution| 240(RGB) x 320 pixels| N/A
Display Type| TFT-LCD (262 K)| N/A
Display Color| True, 16bit/color| N/A
Frequency Range| 20 kHz to 90 kHz| Typical 40 kHz squarewave
__
Filter
| ±5 KHz of main noise frequency, up to three filters| N/A
Power Supply| 4 x 1.5 V AA (LR6) alkaline batteries| 2 x 1.5 V AAA
(LR03) alkaline batteries
Power Consumption (typical)| 75 mA| 33 mA
Battery Life (typical)| 105 hours (Alkaline)| 60 hours (Alkaline)
Low battery indication| | Yes (Red LED)
APO function| 60 minutes when in idle| 60 minutes when in idle
Weight| Approx. 0.518 lb (0.235 kg)| Approx. 0.335 lb (0.152 kg)
Dimensions
| 7.547 x 2.984 x 1.791 in
(183 x 75 x 43 mm)
| 5.295 x 2.559 x 1.326 in
(137 x 65 x 33 mm)
Operating Temperature| -4 °F to 122 °F (-20 °C to 50 °C)
Storage Temperature| -4 °F to 158 °F (-20 °C to 70 °C)
Operating Humidity| <80% RH
Pollution Degree| 2
Protection| IP40
Certifications| CE
__
__
__
Electromagnetic Compatibility (EMC)
| EN 61326-1
Korea (KCC): Class A Equipment (Industrial Broadcasting & Communication Equipment) [1]
[1] This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, MA 01801
TestEquipmentDepot.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>