HT37 TRMS Digital Multimeter User Manual
- June 5, 2024
- HT
Table of Contents
HT37 TRMS Digital Multimeter
SAFETY PRECAUTIONS AND PROCEDURES
This instrument conforms with safety Standard EN 61010-1 related to electronic
measuring instruments.
For your own safety and to avoid damaging the instrument follow the procedures
described in this instruction manual and read carefully all notes preceded by
this symbol
When taking measurements:
- Avoid doing that in humid or wet places.
- Avoid doing that in rooms where explosive gas, combustible gas, steam or excessive dust is present.
- Keep you insulated from the object under test.
- Do not touch exposed metal parts such as test lead ends, sockets, fixing objects, circuits etc.
- Avoid doing that if you notice anomalous conditions such as breakages, deformations, fractures, leakages of battery liquid, blind display etc.
- Be particularly careful when measuring voltages exceeding 20V to avoid risks of electrical shocks.
The following symbols are used:
PRELIMINARY
- This instrument has been designed for use in environments of pollution degree 2.
- It can be used for VOLTAGE and CURRENT measurements on installations of overvoltage category III 1000V and IV 600V.
- When using the instrument always respect the usual safety regulations aimed at:
- Protecting you against the dangerous electric currents.
- Protecting the instrument against incorrect operations.
- Only the leads supplied with the instrument guarantee compliance with the safety standards in force. They must be in good conditions and, if necessary, replaced with identical ones.
- Do not test or connect to any circuit exceeding the specified overload protection.
- Do not effect measurements under environmental conditions exceeding the limits indicated in paragraph 6.2.1.
- Make sure that batteries are properly installed.
- Before connecting the test probes to the installation, make sure that the function selector is positioned on the required measurement.
- Make sure that LCD and function selector indicate the same function.
DURING USE
Read the recommendations which follow and the instructions in this manual:
CAUTION: An improper use may damage the instrument and/or its components
or injure the operator.
- When changing the range, first disconnect the test leads from the circuit under test in order to avoid any accident.
- When the instrument is connected to measuring circuits never touch any unused terminal.
- When measuring resistors do not add any voltage. Although there is a protection circuit, excessive voltage could cause malfunctioning.
- If during measurement the displayed values remain constant check whether the HOLD function is active.
AFTER USE
- After using the instrument turn it off.
- If you expect not to use the instrument for a long period remove the battery to avoid leakages of battery liquids which may damage its inner components.
MEASURING (OVERVOLTAGE) CATEGORIES DEFINITIONS
The norm EN 61010-1: Safety requirements for electrical equipment for
measurement, control and laboratory use, Part 1: General requirements, defines
what measuring category, usually called overvoltage category, is. On paragraph
6.7.4: Measuring circuits, it says:
(OMISSIS)
circuits are divided into the following measurement categories:
-
Measurement category IV is for measurements performed at the source of the low-voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units. -
Measurement category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation. -
Measurement category II is for measurements performed on circuits directly connected to the low voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment. -
Measurement category I is for measurements performed on circuits not directly connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the norm requires that the transient withstand capability of the equipment is made known to the user.
GENERAL DESCRIPTION
This instrument performs the following measurements:
- DC and AC TRMS Voltage.
- DC and AC TRMS Current.
- Resistance and Continuity test.
- Diode test.
- Frequency.
- Capacitance.
All selectable by means of a 10-position function selector (including OFF
position). FUNCTION keys are also available (see chapter 4.2). An analogical
bargraph is also available. The selected quantity is displayed with indication
of measuring unit and active functions.
The instrument disposes of an Auto Power Off function consisting in an
automatic switching off 30 minutes after last pressure on keys or rotation of
selector. To resume normal operation turn the selector on OFF and switch it on
again.
MEAN VALUE AND TRMS: DEFINITION
Safety testers for AC quantities are divided in two big families:
- MEAN VALUE instruments, measuring only the value of the wave at the fundamental frequency (50 or 60 Hz).
- TRUE ROOT MEAN SQUARE (or “TRMS”) instruments, measuring the true root mean square value of the quantity under test.
In presence of a perfectly sinusoidal wave, both families provide identical results. While in presence of distorted waves, readings are different. Mean value instruments provide only the value of the fundamental wave while TRMS instruments provide the value of the entire wave, including harmonics (within the passband of the instrument). Accordingly, if the same quantity is measured with both kinds of instruments, the measured values are identical only if the wave is purely sinusoidal. Should it be distorted, TRMS instruments provide higher values than MEAN VALUE instruments.
TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR: DEFINITION
The effective current value is defined as follows: “In an interval of time
equivalent to a period, an alternate current with effective value having an
intensity of 1A, by passing on a resistor, disperses the same energy which
would be dispersed in the same period of time by a direct current having an
intensity of 1A”. From this definition comes the numerical
expression : The effective value is indicated as RMS (root mean square).
The Crest Factor is defined as the ratio between the Peak Value of a signal
and its
effective value : . This value varies according to the waveform of the
signal, for a purely sinusoidal wave it’s worth 2 =1.41. In presence of
distortions the Crest Factor assumes higher values as long as the wave
distortion is higher.
PREPARATION FOR USE
INITIAL
This instrument was checked both mechanically and electrically prior to
shipment.All possible cares and precautions were taken to let you receive the
instrument in perfect conditions.Notwithstanding we suggest you to check it
rapidly (eventual damages may have occurred during transport – if so please
contact the local distributor from whom you bought the item).Make sure that
all standard accessories mentioned in paragraph 6.3 are included.Should you
have to return back the instrument for any reason please follow the
instructions mentioned in paragraph 7.
SUPPLY VOLTAGE
The instrument is powered by batteries mod. 9V NEDA1604, JIS006P, IEC6F22
included in the packaging. Battery life is approx. 300 hours. When batteries
are low the symbol ” is displayed. To replace/insert batteries follow the
instructions indicated in paragraph 5.2.
CALIBRATION
The instrument complies with the technical specifications contained in this
manual and such compliance is guaranteed for 12 months.
STORAGE
After a period of storage in extreme environmental conditions exceeding the
limits mentioned in paragraph 6.2.1 let the instrument resume normal operating
conditions before using it.
OPERATING INSTRUCTIONS
INSTRUMENT DESCRIPTION
Front panel
LEGEND:
- LCD display.
- Functions keys (RS-232 and keys for HT39 only).
- Rotary selector.
- A input for AC/DC current measurements.
- HzVΩµA input for other measurements.
- COM input.
FUNCTION KEYS
When pressing a key, the corresponding symbol is displayed with a beep. To resume default state turn the selector on another function.
MIN MAX
By pressing MIN MAX key, maximum and minimum values are measured. Both values
are stored and can be recalled by pressing the same key. The symbol
corresponding to the desired function is displayed: “MAX” for maximum value,
“MIN” for minimum value. MIN MAX key is disabled when HOLD function is active.
To exit this function keep MIN MAX key pressed for at least 1 second or rotate
the selector to another position.
PEAK/∆
This key have the double function of measuring max/min peak values (active for
∼V and A positions of rotary selector) and performing relative measurements
(∆REL) for any position of rotary selector other than ∼V and ∼A. Press
cyclically PEAK/∆ to measure and save peak values. “PMAX” and “PMIN” symbols
on the display correspond to Maximum Peak and Minimum Peak values respectively
which are continuously updated by the meter. By keeping pressed PEAK/∆ key for
at least 3 seconds, “CAL” symbol appears on the display and the meter performs
an auto calibration permitting a higher accuracy on peak measurements. To exit
this function keep pressed PEAK/∆ for at least one second or rotate the
selector on another position.
By pressing PEAK/∆ key, the relative measurement is activated: the meter saves the (offset) value on the display and the “∆REL” symbol is shown. The following measurement will be referred to this offset value. By pressing again PEAK/∆ key the offset value is shown and the “∆REL” symbol is blinking. To exit this function keep pressed PEAK/∆ for at least one second or rotate the selector on another position.
RANGE
By pressing RANGE key the manual mode is activated and the “RANGE” symbol is
displayed. Press RANGE cyclically to change the measuring range and fix the
decimal point on the display. RANGE is disabled on positions / , ∼A and A
performing measurements in Autorange. Keeping press for at least 1 second
RANGE key or moving the rotary switch to exit from this mode.
HOLD
By pressing HOLD the measured value is frozen on the display where the symbol
“HOLD” appears. To exit this function press HOLD again or rotate the selector
to another position.
RS-232 for connection to PC (HT39 only)
HT39 can be connected to a PC by means of the optional software package mod.
SW39 (software + serial cable). To connect HT39 to a PC proceed as follows:
- Insert the connector of the serial cable in the serial input on the upper side of the meter (see Fig. 2).
- Insert the 9 pin connector plug in the PC serial port (COM).
- Press RS-232 key to activate the serial connection. “RS232” symbol is displayed and the “APO” symbol disappears (the AutoPowerOFF function is disabled).
- Run SW39 software and press “Connect” key to activate the real-time visualization of parameters values between the meter and the PC. For any detail please refer to the Help-on-line of SW39.
- Press RS-232 key to exit the RS232 mode.
BACKLIGHT (HT39 only)
By pressing key it’s possible to activate/disable the backlight function on
the display. This function is available on each position of the rotary
selector.
Disable AutoPowerOFF
When the meter is to be used for long periods of time, the operator might want
to disable the AutoPowerOFF function. Once the AutoPowerOFF function is
disabled the meter stays on continuously. To disable the AutoPowerOFF
function:
- Switch OFF the meter.
- Turn ON the meter keeping pressed MIN MAX, PEAK/∆ and RANGE keys.
The AutoPowerOFF function is automatically activated when turning ON again the meter.
MEASUREMENTS
DC Voltage measurement
Caution: Maximum input for DC voltage is 1000V. Do not attempt to take
any voltage measurement that exceeds the limits. Exceeding the limits could
cause electrical shock and damage the multimeter.
- Turn the selector on V.
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 3).
- Connect the red and black test leads to the positive and negative poles of the circuit under test respectively. The voltage value will be displayed with automatic detection of the range.
- Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from display. Press cyclically RANGE to change the position of the decimal point on the display. To activate the Autorange function keep pressed RANGE for at least 1 second.
- The message “OL” means that the voltage exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- The symbol “-” on the display means that voltage has opposite direction than shown in Fig. 3.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
AC Voltage measurement
CAUTION: Maximum input for AC voltage is 750V rms. Do not attempt to
take any voltage measurement exceeding such limit to avoid the risk of
electrical shock and damages to the instrument.
- Turn the selector on ∼V.
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 4).
- Connect the red and black test leads to the positive and negative poles of the circuit under test respectively. The voltage value will be displayed with automatic detection of the range.
- Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from display. Press cyclically RANGE to change the position of the decimal point on the display. To activate the Autorange function keep pressed RANGE for at least 1 second.
- The message “OL” means that the voltage exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- For minimum and maximum value measurement, HOLD function and Peak measurement please refer to paragraph 4.2.
DC Current measurement
CAUTION: Maximum input for DC current is 10A. Do not attempt to take
any current measurement exceeding such limit to avoid the risk of electrical
shock and damages to the instrument.
- Switch OFF the circuit under test.
- Turn the selector on A.
- Insert the red test lead into A jack and the black plug into COM jack (Fig. 5).
- Connect red and black plugs in series with the circuit under test respecting the polarity and current flow indicated in Fig. 5.
- Power the circuit under test. The current value is displayed.
- The message “OL” means that the current exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- If the measured value is lower than 4mA, to get a better resolution:
- Switch off the circuit under test.
- Turn the selector on µA.
- Remove the red test lead from A jack, and insert it into
- Power the circuit under test.
- The symbol “-” on the display means that current has opposite direction than shown in Fig. 5. Press RANGE key for manual range only active in µA position of the rotary selector.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
AC Current Measurement
CAUTION: Maximum input for AC current is 10A. Do not attempt to take
any current measurement exceeding such limit to avoid the risk of electrical
shock and damages to the instrument.
- Switch OFF the circuit under test.
- Turn the selector on ∼A.
- Insert the test leads into the jacks, the red plug into A jack and the black plug into COM jack (Fig. 6).
- Connect red and black plugs in series with the circuit under test as indicated in Fig. 6.
- Power the circuit under test. The current value is displayed.
- The message “OL” means that the voltage exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- For minimum and maximum value measurement, HOLD function and Peak measurement please refer to paragraph 4.2.
Frequency measurement
CAUTION: Maximum input for AC voltage is 750V rms. Do not attempt to take any voltage measurement exceeding such limit to avoid the risk of electrical shock and damages to the instrument.
- Turn the selector on Hz.
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 7).
- Connect the red and black test leads to the positive and negative poles of the circuit under test respectively. The voltage value will be displayed with automatic detection of the range.
- Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from display. Press cyclically RANGE to change the position of the decimal point on the display. To activate the Autorange function keep pressed RANGE for at least 1 second.
- The message “OL” means that the frequency exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
Resistance measurement
CAUTION: Before taking resistance measurements in circuit remove power
from the circuit being tested and discharge all capacitors.
- Turn the selector on Ω.
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 8).
- Connect the red and black test leads to the circuit under test, the resistance value will be displayed with automatic detection of the range.
- Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from display. Press cyclically RANGE to change the position of the decimal point on the display. To activate the Autorange function keep pressed RANGE for at least 1 second.
- The message “OL” means that the resistance exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
Diode test and continuity test
CAUTION : Before taking resistance measurements in circuit remove power
from the circuit being tested and discharge all capacitors.
- Turn the selector on
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 9).
- Connect the red lead to the positive side (anode) of the diode and the black lead to the negative side (cathode). The meter displays the diode voltage to approximately 0.4 ~ 0.9V for good junction.
- Reverse the connections and measure the voltage across the diode again. The message “OL” on the display corresponds to a correct junction.
- The continuity test is always active and the test is performed using the test leads in the same way of resistance measurement (refer to paragraph 4.3.6). The buzzer is on for resistance values <450Ω.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
Capacitance measurement
CAUTION: Before taking any in circuit or capacitance measurement,
remove power from the circuit being tested and discharge all capacitors. Use
the short test lead pair for measurement in order to reduce the stray
capacitance. Before connecting the test capacitor, note the display, which may
show a reading other than zero whenever the range is changed. Subtract this
offset reading from the test result of a capacitor to obtain the true value.
Connect the test capacitor to the input clamps noting the polarity connections
when required.
- Turn the selector on .
- Insert the test leads into the jacks, the red plug into HzVΩµA jack and the black plug into COM jack (Fig. 10).
- Connect the test leads to the capacitor terminals taking care of the shown polarity. The capacitance value will be displayed with automatic detection of the range.
- Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from display. Press cyclically RANGE to change the position of the decimal point on the display. To activate the Autorange function keep pressed RANGE for at least 1 second.
- The message “OL” means that the capacitance exceeds the measuring limits. In this case disconnect the test leads from the circuit under test to avoid damaging the instrument or endanger your own safety.
- For minimum and maximum value measurement, HOLD and Relative ∆REL functions please refer to paragraph 4.2.
MAINTENANCE
GENERAL INFORMATION
- This is a precision instrument. To guarantee its performances be sure to use it according to these instructions and keep it stored on suitable environmental conditions.
- Do not expose it to high temperatures or humidity or direct sunlight.
- Be sure to turn it off after use. If you expect not to use the instrument for a long period remove batteries to avoid leakages of battery liquid which could damage the its inner components.
BATTERY REPLACEMENT
When the symbol ” is displayed, batteries need replacement.
CAUTION: Before replacing batteries disconnect the test leads from any energized circuits to avoid electrical shocks.
LEGEND:
- Turn OFF the meter and disconnect the test leads from the input terminals.
- Remove the protective holster from the meter.
- Unscrew the battery cover and remove the battery.
- Insert a new battery of the same type (9V NEDA1604, JIS006P, IEC6F22) observing the proper polarity, re-screw the battery cover and reposition the protective holster.
FUSE REPLACEMENT
CAUTION: Before replacing fuses, disconnect test leads from any
energized circuit to avoid electrical shock.
LEGEND:
- Turn OFF the meter and disconnect the test leads from the input terminals.
- Remove the protective holster from the meter.
- Unscrew the battery cover and remove the battery.
- Remove the defective fuse and install a new fuse of the same size and rating (fast type 10A/1000V). Make sure the new fuse is centered in the fuse holder. Re-screw the battery cover and reposition the protective holster.
CLEANING
To clean the instrument use a soft dry cloth. Never use a wet cloth, solvents
or water.
END OF LIFE
CAUTION: this symbol indicates that equipment and its accessories shall
be subject to a separate collection and correct disposal.
TECHNICAL SPECIFICATIONS
TECHNICAL FEATURES
The accuracy is indicated as [% of reading + number of digits]. It is referred
to the following environmental conditions: temperature 23°C ± 5°C, relative
humidity < 80%.
DC Voltage
Range| Resolution| Accuracy| Input impedance| Overload
protection
---|---|---|---|---
400.0mV| 0.1mV| ****
±(0.5%rdg+2dgt)
| ****
10MW // less of 100pF
| ****
1000VDC
750VACrms
4.000V| 0.001V
40.00V| 0.01V
400.0V| 0.1V
1000V| 1V
AC TRMS Voltage
Range| Resolution| Accuracy
(50 ¸ 500Hz)
| Input impedance| Overload protection
---|---|---|---|---
400.0mV| 0.1mV| Not specified| ****
10MW // less of 100pF
| ****
1000VDC
750VACrms
4.000V| 0.001V| ±(1.3%rdg+5dgt) (50 ¸ 300Hz)
40.00V| 0.01V| ±(1.2%rdg+5dgt) (50 ¸ 500Hz)
400.0V| 0.1V
750V| 1V
DC Current
Range| Resolution| Accuracy| Output voltage| Overload
protection
---|---|---|---|---
400.0mA| 0.1mA| ****
±(1.0%rdg+2dgt)
| <5mV/mA| ****
600Vrms
4000mA| 1mA
10.00A
| ****
10mA
| ****
2V max
| Fast fuse 10A / 1000V
AC TRMS Current (Autorange)
Range| Resolution| Accuracy
(50 ¸ 500Hz)
| Output voltage| Overload protection
---|---|---|---|---
10.00A| 10mA| ±(1.5%rdg+5dgt)| 2V max| 600Vrms
Frequency
Range| Resolution| Accuracy| Sensitivity| Overload
protection
---|---|---|---|---
4000Hz| 1Hz| ****
±(0.01%rdg+1dgt)
| >1.5VACrms
<5VACrms
| ****
600Vrms
40.00kHz| 0.01kHz
400.0kHz| 0.1kHz
4.000MHz| 0.001MHz| >2VACrms
<5VACrms
40.00MHz| 0.01MHz
Resistance
Range| Resolution| Accuracy| Open voltage| Overload
protection
---|---|---|---|---
400.0W| 0.1W| ±(1.0%rdg+5dgt)| ****
1.3V
| ****
600V rms
4.000kW| 0.001kW| ****
±(0.7%rdg+2dgt)
40.00kW| 0.01kW
400.0kW| 0.1kW
4.000MW| 0.001MW| ±(1.0%rdg+2dgt)
40.00MW| 0.01MW| ±(1.5%rdg+5dgt)
Diode test
Range| Resolution| Accuracy
(0.4 ¸ 0.8V)
| Test current| Open voltage| Overload protection
---|---|---|---|---|---
****| 10mV| ±(1.5%rdg+5dgt)| 1.5mA| <3V| 600V rms
Continuity test
Range | Buzzer | Overload protection |
---|---|---|
**** | R<450W | 600V rms |
Capacitance
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
4.000nF | 0.001nF | ±(3.0%rdg+10dgt) | **** |
600V rms
40.00nF| 0.01nF| ****
±(2.0%rdg+8gt)
400.0nF| 0.1nF
4.000mF| 0.001mF
40.00mF| 0.01mF
400.0mF| 0.1mF
4.000mF| 0.001mF| ±(5.0%rdg+20dgt)
40.00mF| 0.01mF
Electrical specifications
- Conversion: TRMS
- Measuring rate: 1.5 times per second
- Temperature coefficient: 0.15×(accuracy)/°C (<18°C and >28°C)
- NMRR Normal Mode Rejection Ratio: > 50dB for DC parameters and 50/60Hz
- CMRR Common Mode Rejection Ratio: >100dB from DC up to 60Hz on DC > 60dB from DC up to 60Hz on ACV
Safety
- The instrument complies with: EN 61010-1
- Insulation: Class 2, double insulation
- Pollution degree: 2
- Overvoltage category: CAT IV 600V, CAT III 1000V ( V/ W / m A) CAT IV 500V ( A )
- Max height: 2000m
General data
Mechanical characteristics
- Dimensions (with holster): 164x82x44mm
- Weight (including battery): About 400g
Power supply
- Battery type: 9V NEDA1604, JIS006P, IEC6F22
- Low battery indication: ” ” is displayed when battery voltage is too low
- Battery life: About 300 hours
- AutoPowerOFF: 30 minutes after last pressure on keys or rotation of selector
Display
- Specifications: 4 LCD with max. reading 4000 counts + symbol, decimal point and bargraph
- Over range indication: “ OL ” or “ -OL ”
ENVIRONMENT
Environmental conditions
- Reference temperature: 23 ± 5°C
- Working temperature: 0 ÷ 30°C
- Relative humidity: <80%RH
- Storage temperature: -20 ÷ 60°C
- Storage humidity: <80%RH
EMC and LVD
This product conforms to the prescriptions of the European directive on low voltage 2006/95/EEC (LVD) and to EMC directive 2004/108/EEC.
ACCESSORIES
Standard accessories
- HT37 or HT39 meter.
- Test leads.
- Battery (fitted).
- Instruction manual.
Optional accessories
- Carrying bag – Code B80.
- Software + serial cable (only HT39) – Code SW39.
SERVICE
WARRANTY CONDITIONS
This instrument is guaranteed for one year against material or production
defects, in accordance with our general sales conditions. During the warranty
period the manufacturer reserves the right to decide either to repair or
replace the product.
Should you need for any reason to return back the instrument for repair or
replacement take prior agreements with the local distributor from whom you
bought it. Do not forget to enclose a report describing the reasons for
returning (detected fault). Use only original packaging. Any damage occurred
in transit due to non original packaging will be charged anyhow to the
customer.
The warranty doesn’t apply to:
- Accessories and batteries (not covered by warranty).
- Repairs made necessary by improper use (including adaptation to particular applications not foreseen in the instructions manual) or improper combination with incompatible accessories or equipment.
- Repairs made necessary by improper shipping material causing damages in transit.
- Repairs made necessary by previous attempts for repair carried out by non skilled or unauthorized personnel.
- Instruments for whatever reason modified by the customer himself without explicit authorization of our Technical Dept.
- Use not provided by the instrument’s specifications or in the instruction manual.
The contents of this manual may not be reproduced in any form whatsoever without the manufacturer’s authorization. Our products are patented and our logotypes registered. We reserve the right to modify specifications and prices in view of technological improvements or developments which might be necessary.
SERVICE
Shouldn’t the instrument work properly, before contacting your distributor
make sure that batteries are correctly installed and working, check the test
leads and replace them if necessary.
Should you need for any reason to return back the instrument for repair or
replacement take prior agreements with the local distributor from whom you
bought it. Do not forget to enclose a report describing the reasons for
returning (detected fault). Use only original packaging. Any damage occurred
in transit due to non original packaging will be charged anyhow to the
customer.
The manufacturer will not be responsible for any damage to persons or things.