HT ITALIA DT-8806H Medical Thermometers User Manual
- June 4, 2024
- HT ITALIA
Table of Contents
HT ITALIA DT-8806H Medical Thermometers
General description
The Non-Contact Forehead IR Thermometer is specially designed to take the body temperature of a person regardless of room temperature. Depending on various skin types and thickness, there may be temperature differences.
Safety information
- This device must only be used for the purposes described in this instruction manual
- This device must only be used in an ambient temperature range between 10°C and 40°C
- Do not expose this thermometer to electric shocks.
- Do not expose this thermometer to extreme temperature conditions of > 50°C or < 0°C
- Do not use the device in relative humidity higher than 85%RH
- Do not use the device near large electromagnetic fields such as found with cordless or celi phones.
- Keep the device away from water and heat, including direct sunlight.
- Do not drop or knock the device, and do not use if damaged .
- lt may affect the accuracy of measurements when the forehead is covered by hair, perspiration, cap or scarf.(See Part 10-4)
- Keep the Measuring distance as 5cm-15cm (2in-5.9in).(See Part 10-4)
- When the body infrared thermometer should be left in that room during 15 to 20 minutes before using.
- lt may affect the accuracy of measurements when the forehead is covered by perspiration or other factors, please take the temperature behind the ear lobe .(See Part 10-5)
- Clean the glass with a cotton bud lightly moistened with 70% alcohol
IMPORTANT REMARKS:
- Before taking of the temperature make sure to remove hair and perspiration from the forehead .
- Selecting “Body” mode to measure the body temperature; Selecting “Surface” mode to measure the surface temperature.
- Use of this thermometer is not intended as a substitute tor consultation with your physician .
- Should a problem occur with your device,please contact your retailer.
- Do not attempt to repair the device yourself.
- According to EMC standard, the medicai electronic products should be maintained specially
Features
- Precise non-contact measurements
- User selectable °C or °F
- Selectable Body and Surface temp
- Set Alarm value
- Memorization of the last 32 measurements
- Automatic Data Hold & Auto power off
- Automatic selection range and Display Resolution 0.1°(0.1°F)
- Backlight LCD displa
General
Non-Contact Forehead IR Thermometer is designed for body surface and forehead
temperature measurement for infants and adults without contact to human body.
Non-Contact Forehead IR Thermometer can also be used to measure the
temperature of a baby-bottle or bath, or room temperature (by using the
Surface Temp function).
NORMAL TEMPERATURES BASED TO MEASUREMENT METHOD
Measurement Method | Temperature (°C) | Temperature (°F) |
---|---|---|
Rectal | 36.6 to 38 | 97.8 to 100.4 |
Oral | 35.5 to 37.5 | 95.9 to 99.5 |
Axillary | 34.7 to 37.3 | 94.4 to 99.1 |
Ear | 35.8 to 38 | 96.4 to 100.4 |
NORMAL TEMPERATURES ACCORDING TO AGE
Age | Temperature (°C) | Temperature (°F) |
---|---|---|
0-2 years | 36.4 to 38.0 | 97.5 to 100.4 |
3-10 years | 36.1 to 37.8 | 97.0 to 100.0 |
11-65 years | 35 .9 to 37 .6 | 96.6 to 99.7 |
> 65 years | 35.8 to 37.5 | 96.4 to 99.5 |
Instrument description
CAPTION
- IR sensor
- LCD display
- Mode selection
- DOWN key
- UP key
- MODE key
- Measurement Trigger
- Battery cover
CAPTION
- Surface mode symbol
- Body mode symbol
- Result of measurement
- Low battery indication
- Memory location number
- Saved data value
- Measurement unit (°C/°F)
- Activated alarm symbol
Description of symbols on the instrument and manual
Technical Specifications
General specifications
Display Resolution| 0.1°C (0.1°F)
Operating Temperature| 10°C ÷ 40°C (50°F ÷ 104°F)
Storage Temperature| 0°C ÷ 50°C (32°F ÷ 122°F)
Humidity Rate| <85%RH
Power supply| 3V DC (2 x 1.5V type AA IEC LR06)
Battery life:| 4 hours (continue test)
Dimensions (L x W x H)| 149 x 77x 43mm / 5.9 x3 x 1.7in
Weight (included batteries)| 195g (7 ounces)
Mechanical protection:| IP40
Measurement range|
---|---
Body mode| 32.0°C ÷42 .5°C (90°F ÷ 108°F)
Surface mode| 0°C ÷ 60°C (32°F ÷ 140°F)
Accuracy Surface mode| ±0.3°C (0.54°F)
Measurement distance| 5cm ÷ 15cm (2in ÷ 5.9in)
Response time| 0.5s
Auto Power OFF| 7s
Accuracy Body mode
32°C÷35.9°C/ 93.2°F÷96.6°F| ±0.3°C/0.5°F| According to ASTM Standard
E1965-1998 (2003)
---|---|---
36°C ÷ 39°C / 96.8°F÷102.2°F| ± 0.2°C/0.4°F
39°C÷ 42.5°C/102.2°F÷108.5°F| ± 0.3°C/0.5°F
Initial calibration
For stable and reliable results, The thermometer provides a user-correction function, the operation of the method is as follows: 1st step take the temperature of a person using a conventional thermometer; you will get 37.5°C (99.5°F) for instance 2nd step take the temperature of the same person using the Noncontact Body lnfrared Thermometer keeping the 5 to 15 cm (2 into 5.9 in) distance between the thermometer and the forehead (Take care to remove any obstacle which could alter the measurement (hair, perspiration …). lf you get 37.5°C (99.5 °F), the Non-contact Body lnfrared Thermometer is properly set and ready for use lf you get a lower temperature, 36.4°C (97 .4°F) for example, your difference is 1.1°C (2 .2°F). You should adjust the temperature on the Non-contact Body lnfrared Thermometer and add the difference, i.e. 1.1°C (2 .2°F): Press MODE key for 2 seconds, the screen displays “F1” Press the MODE key again until you get “F3” Press UP key in order to add the difference (in our example, 1.1°c – 2.2°F) To check, take the temperature again using the Non-contact Body mode
Use of instrument
- Insert the batteries
- For the first use or when inserting new batteries wait between 10 minutes for the warm-up of the apparatus and when inserting the new batteries)
- lf the device is not used for a long time, once you turn it on again, the device will test the room temperature first and will delay turning on for one or two seconds.
- Aim towards the forehead (see the diagram below for the positioning), from a distance of 5cm(2in), press the measuring key, the temperature is displayed immediately
- Making sure there is no hair, perspiration, cosmetic or cap covered on the forehead.
- When the room temperature is significant difference, or there is perspiration on the forehead, You can take the temperature behind the ear lobe
- Making sure there is no hair, perspiration, cosmetic or cap covered
Settings
-
Unit selection Function F1
-
Press MODE key for 2 seconds, the screen displays “F1”
-
Press the DOWN key for degrees Celsius or UP key for degrees Fahrenheit
-
Alarm setup Function F2
-
Press MODE key for 2 seconds, the screen displays “F1”
-
Press twice MODE touch to get “F2” . Press UP key to increase the threshold by 0.1°C (0.1°F) or DOWN key to reduce it by 0.1°C (0.1°F).
NOTE: The alarm threshold default value is 38°C (100.4°F) -
Temperature adjustment Function F3
To adjust the total variation of your the Non-contact Body lnfrared Thermometer: -
Press MODE key for 2 seconds, the screen displays “F1”
-
Press “MODE” button two times until shows the “F3”
-
Press UP” to increase the difference by 0.1°C (0.1 °F), “DOWN” to reduce it by 0.1°C (0.1 °F)
-
ln case of seasonal or environmental changes a verification and adjustment should be carried out. NOTA: This function is only effective Body
-
Buzzer ON/OFF – F4 Menu
-
Press MODE key for 2 seconds, the screen displays “F1”
-
Premere il tasto MODE fino a visualizzare il messaggio “F4”
-
Press UP key to activate the buzzer (a sound icon ” ” is displayed on the LCD screen) or press DOWN key to disable it (the icon ” ” disappear) Exit Setting mode Press MODE key until the screen turns off
-
Measurement modes
The instrument is especially designed to take the body temperature of a human being. For this, use the Body mode. Measurement range for Body mode: 32°C to 42.5°C (86°F to 108°F). You can also use the Non-contact Body lnfrared Thermometer to measure the temperature of an area or an object, a food, a liquid or a room temperature. For this, use the Surface mode. Measurement range for Surface mode: 0°C to 60°(32°F to 140°F).
IMPORTANT NOTE : The area temperature differs from the internal body temperature. To obtain the internal temperature always use the Body mode. Please make sure to select the Body mode for an internal temperature reading and the Surface mode for an external area reading (bottle, bath, room…) -
Internal memory Data memory automatically after temperature measurements, which will display at the right corner of the LCD. Press UP key or DOWN key to display the last temperature measurement. Deleting internal memory
-
With instrument switched off press UP and DOWN keys together tor 2s to display the last temperature measurement
-
Change the order number to “0” and press MODE key in order to delete all memory data
-
Battery replacement When the ” ” symbol is displayed is necessary replace batteries.
-
Open the battery cover
-
Remove the batteries and replace with new ones of the same type (see § 8). Taking great care to respect the correct polarity. Never use rechargeable batteries. Use only batteries for single usage.
-
Restore battery cover
-
Remove the battery from the instrument if it is not required tor extended periods of time in order to avoid damage to the thermometer resulting from a leaking battery
-
Do not dispose of used batteries in the environment. Use the appropriate containers for disposal
-
Longevity use The Non-contact Body lnfrared Thermometer was conceived for an intense and professional use, its longevity is guaranteed for 40000 takings.
Warnings
- he protective glass over the lens is the most important and fragile part of the thermometer, please take great care of it.
- Do not recharge nonrechargeable batteries, do not throw in fire. · Do not expose the thermometer to sunlight or water
Maintenance and Cleaning
-
The infrared Sensor is the most precise part, must be protected carefully.
Clean the device with a cotton bud lightly moistened with 70% alcohol. -
Do not clean the device with corrosive detergent.
-
Keep the device away from water or other liquid .
-
Store the device in a dry environment, and keep it away from dust and direct sunlight
Troubleshooting
lf you happen to have one of the following problems while using the non- contact forehead IR thermometer please refer to this breakdown service guide to help resolve the problem. lf the problem persists please contact our customer service.
-
The screen displays the body temperature inferior to 32°C (89 .6°F)
A lf you’re on Surface mode the 32°C(89.6 °F) temperature displayed is showing the external temperature that your body releases) -
The screen displays the “Hl” message
When using the Non-contact Body lnfrared Thermometer the message Hl can show on the screen. The analysis is above the measurement range selected, either superior to 42.5°C (108 °F) in Body mode or superior to 60°C (140°F) in Surface mode -
The screen displays the “Lo” message
When using the Non-contact Body lnfrared Thermometer the message “Lo” can show on the screen. The temperature analyzed is under the measuring range selected, either less than 32°C (90°F) in Body mode or less than 0°C (32°F) in Surface mod
List of possible wrong conditions during a measurement
Temperature reading hampered by hair, perspiration. . .| Make sure that there is no
obstruction prior to taking a temperature.
---|---
Temperature hampered by an air flux.| Make sure there is no air flux as
this could interfere with the infrared system.
The measuring distance is too far| Please respect the measuring distance (between 5 to 15cm
– 2 to 5.9 in).
From high/low temperature condition to room temperature| Waiting for 10 minutes before taking the body temperature
Reference guidelines
The instrument is compliance with the below guidelines::
- EN12470-5 and ASTM E1965-1998
- EN980: Graphical symbols tor use in the labeling of medical devices
- EN1041: Information supplied by the manufacturer with medical devices
- EN60601-1 : Medical electrical equipment Part 1: General requirements tor safety (IEC:60601-1 :1998)
- EN60601-1-2 : Medical electrical equipment Part 1-2 : General requirements tor safety Collateral standard Electromagnetic compatibility Requirements and test (IEC 60601-1-2:2001)
EMC Directive
This device has been tested and homologated in accordance with EN60601-1-2:2007 tor EMC. This does not guarantee in any way that the device will not be affected by electromagnetic interference. Avoid using the device in a high electromagnetic environment. The MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS.
The manufacturer reserves the right to alter the specifications of the product
without prior notification .
The manufacturer allows himself the right to modify without any preliminary
opinion the technical specifications of the product.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>