Tektronix P6616 16 Channel General-Purpose Logic Probe Instructions
- June 5, 2024
- Tektronix
Table of Contents
P6616
General-Purpose Logic Probe
Instructions
Product Description
The P6616 general-purpose logic probe connects the Tektronix MSO/DPO5000
Series and MSO4000B Series of mixed-signal oscilloscopes to digital buses and
signals on your target system. The probe contains 16 data channels split
between two lead sets (GROUP 1 and GROUP 2).
All leads include a ground connection at the tip. You can connect the probe
leads separately to the target system, or group the leads together using the
probe tip holders.
Connecting the Probe to the Oscilloscope
Connect the probe as shown in the illustration below.
- Insert the probe label-side up into the connector on the oscilloscope.
- To remove the probe, squeeze the buttons on the side and pull out the probe.
Connecting the Probe to Your Circuit
Attach the probe to the circuit using the connectors and adapters shown on the
back of these instructions. Select the best method for your needs, and then
proceed to Setting up the Probe.
Setting Up the Probe
To set and view the digital channel parameters, do the following:
On MSO/DPO5000 Series instruments, select Vertical>Digital Setup.
On MSO/DPO4000B Series instruments, push the D15D0 button.
The parameters listed below can be set on each digital channel:
- Threshold voltage and vertical position (the default threshold setting is 1.4 V)
- Signal height and position (set once for all 16 channels)
- Channel label
To set and view bus characteristics, do the following:
On MSO/DPO5000 Series instruments, use the controls in the Bus Setup screen.
On MSO/DPO4000B Series instruments, push the B1 to B4 buttons.
The setup screens allow you to set and view bus characteristics such as:
- Clock type
- Bus type (Serial or Parallel)
- Bus width
- Display format (Hex, Binary, or ASCII symbols)
Parallel bus setup information is resident on some oscilloscope models. However, for other buses such as SPI and I2C, you must have the appropriate option. See your oscilloscope manual or product data sheet for nomenclature and ordering details.
Functional Check
Logic activity immediately displays on all connected, active channels. If you do not see an active signal:
- Push the Trigger button.
- Select Edge for trigger type.
- Select the channel that you are setting up as the source.
- On MSO/DPO4000B Series instruments, push the Autoset button.
If you do not see an active signal, try another probe channel (or analog probe) to verify circuit activity at the test point.
Typical Application
- Use the P6616 probe to view digital signals on a system bus.
- Use an analog probe, such as the TPP0500 or TPP1000 active probe to view analog waveform information.
Accessories
The following standard accessories ship with the probe and are shown in the illustration on the following page.
Item | Description | Quantity | Part number |
---|---|---|---|
— | Logic probe accessory kit | Item 1–6 | 020-2662-XX |
1 | Extension ground tip | 1 set of 20 | 020-2711-XX |
2 | Probe tip | 1 set of 10 | 131-5638-11 |
3 | IC grabber | 1 set of 20 | 020-2733-XX |
4 | Probe tip holder | 2 ea | 352-1115-XX |
5 | 8” Ground lead | 1 set of 2 | 020-2713-XX |
6 | 3” Ground lead | 1 set of 8 | 020-2712-XX |
Instructions1 | 1 ea | 071-2831-XX |
1 Instruction are included with the probe, but not in the accessory kit. The instructions can be downloaded at www.tektronix.com/manuals.
These optional accessories can be ordered for your probe:
Description | Part number |
---|---|
P6960 Probe D-MAX Footprint to Square Pin Header Adapter | NEX-P6960PIN |
Specifications
Table 1: Electrical and mechanical specifications
Characteristic | Description |
---|---|
Input channels | 16 digital |
Input resistance | 100 kΩ ±1.0% |
Input capacitance | 3.0 pF |
Input signal swing
Minimum
Maximum
| 400 mV p-p
30 V p-p, ≤200 MHz (centered around the DC threshold voltage) at the probe tip
10 V p-p, ≥200 MHz (centered around the DC threshold voltage) at the probe tip
Maximum non-destructive input signal| 30 V p-p, 42 V peak, ±50 VDC
Threshold voltage| ± 40 V
Minimum detectable pulse width| 1ns
Maximum input toggle rate| 500 MHz
Digital channel-todigital channel skew| 200 ps
Probe length| 1.0 m (3.28 ft)
Table 2: Environmental specifications
Characteristic | Description |
---|
Temperature
Operating
Nonoperating
| 0 °C to +50 °C (+32 °F to +122 °F) 55 °C to +75 °C (–67 °F to +167 °F)
Humidity
Operating
Nonoperating
| 5% to 95% relative humidity 10% to 95% relative humidity
Altitude
Operating
Nonoperating
| 4.6 km (15,092 ft) maximum 15 km (50,000 ft) maximum
Equipment Recycling. This product complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Safety Summary
Connect and Disconnect Properly.
Connect the probe output to the measurement instrument before connecting the
probe to the circuit under test. Disconnect the probe input and the probe
ground from the circuit under test before disconnecting the probe from the
measurement instrument.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markings on the
product. Consult the product manual for further ratings information before
making connections to the product.
Do not Operate Without Covers.
Do not touch exposed connections and components when power is present.
Avoid Exposed Circuitry.
Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures.
If you suspect there is damage to this product, have it inspected by qualified
service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could
result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could
result in damage to this product or other property.
Symbols on the Product.
This symbol may appear on the product:
Contacting Tektronix
Web site: | www.tektronix.com |
---|---|
Phone: | 1-800-833-9200 |
Address: | Tektronix, Inc. |
Department or name (if known) 14200 SW Karl Braun
Drive P.O. Box 500, Beaverton, OR 97077, USA
Email:| techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty.
Connecting the Probe to the Circuit
Copyright © Tektronix, Inc. All rights reserved.
www.tek.com
References
- Test and Measurement Equipment | Tektronix
- Test and Measurement Equipment | Tektronix
- Product Support and Downloads | Tektronix
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>