ets-lindgren 8000-035 RF Power Amplifier User Manual
- June 5, 2024
- ETS LINDGREN
Table of Contents
ets-lindgren 8000-035 RF Power Amplifier
Introduction
The ETS-Lindgren Model 8000-035 RF Power Amplifier is capable of supplying a minimum of 35 watts into a 50-ohm load over the frequency range 1MHz to 1000MHz. The amplifier is designed with sufficient gain so that it may be used with normal output levels of signal sources.
- A safety interlock on the rear panel is also provided, which will mute the amplifier when grounded.
- The unit is powered by a switched-mode power supply for high efficiency, high power factor, and a wide voltage range operation. The unit is air-cooled with integral fans and is protected against faulty cooling by excess temperature sensing.
- A front panel fault indicator is provided to indicate over-temperature.
The amplifier is designed for rugged operation into a variety of loads and is primarily intended for use as an RF power source for EMC susceptibility testing, but is also applicable to other systems requiring a wide-band linear amplifier. This amplifier is designated as Professional Equipment and should not be operated by untrained staff. As this product has the capability to generate high levels of RF energy it is not intended for use in a residential environment.
Standard Configuration
- The 8000-035 amplifier is supplied for fitting into a 19-inch rack.
- The RF connectors are N-type which are located on the rear panel of the amplifier.
Maintenance
If you have any questions concerning maintenance or if the amplifier system is misoperating, contact ETS-Lindgren Customer Service. See Service Procedures for contact information.
CAUTION:
Before performing any maintenance, follow the information provided in
Before You Begin on page 12.
WARNING: There are no user-serviceable parts within the 8000-035
amplifier system.
To prevent electrical shock DO NOT remove the covers. Warranty will be void if
the covers are opened. ALWAYS UNPLUG THE UNIT BEFORE CLEANING. Clean the
exterior of the cabinet using a damp cloth.
Note: The cables which form part of the amplifier are matched lengths and
are critical to its performance. If replacement is necessary use only genuine
spares.
CONTACTING ETS-LINDGREN
Note: Please see ets-lindgren.com for a list of ETS-Lindgren offices, including phone and email contact information.
SENDING A COMPONENT FOR SERVICE
- Contact ETS-Lindgren Customer Service to obtain a Service Request Order (SRO).
- Briefly describe the problem in writing. Give details regarding the observed symptom(s) or error codes, and whether the problem is constant or intermittent in nature. Please include the date(s), the service representative you spoke with, and the nature of the conversation. Include the serial number of the item being returned.
- Package the system or component carefully. If possible, use the original packing materials or carrying case to return a system or system component to ETS-Lindgren.
CALIBRATION SERVICES
- Annual Calibration—For reliable and repeatable long-term performance, annual recalibration of your measuring device by an ETS-Lindgren experienced technician is recommended. The ETS-Lindgren calibration team will calibrate most any type or brand of measuring device. Additional information is available at ets-lindgren.com.
- Calibration Services Plus!™—Calibration Services Plus! is a customized, optional program to help you expertly maintain your test and measurement assets. It is designed to address the challenges of managing the calibration and repair of test and measurement components, including scheduling and planning to ensure peak performance. Additional information is available at ets-lindgren.com.
Specifications
Electrical Specifications
Note: The third-order intercept point is a nominal value as its
calculation depends upon the power level at which distortion measurements are
made.
Note: Output VSWR tolerance is specified for excitation within the
permitted levels and frequency range.
Frequency Range (Instantaneous) | 1MHz – 1000MHz |
---|---|
Rated Output Power | 35W Minimum |
Output Power at 1dB Gain Compression | 20W Minimum |
Gain | 50dB Minimum |
Third Order Intercept Point | 57dBm |
Gain Variation with Frequency | +/- 3dB |
Harmonics at Minimum Linear Output Power |
Better than -20dBc
Output Impedance| 50 Ohms
Stability| Unconditional
Output VSWR Tolerance| Infinity:1
Input VSWR| 2:1
RF Connector Style| Type N Female
Safety Interlock| BNC Female, S/C, and O/C to Mute
USB/GPIB/Ethernet Interface| Standard
Physical Specifications
Case Dimensions | 19-inch, 3U Rack, 440mm deep |
---|---|
Weight | 13 kg (28.6 lbs) |
Operating Temperature Range | 0-40°C |
--- | --- |
Altitude | Up to 2000m above sea level |
Humidity | 95% (Non-condensing) |
Before You Begin
Unpacking
The 8000-035 amplifier is supplied with the following contents:
- Power Cable (10A)
- Spare Fuse (T 6.3A H 250V 20mm)
- User Manual (This document)
- Remote Control Interface Manual
- Remote Control Interface CD
- USB Cable
- Calibration Test Report
If any signs of damage are found no attempt should be made to install or operate the amplifier system. Contact ETS-Lindgren Customer Service immediately to report the problem. If the shipping carton has been damaged, retain the shipping carton and packing material for the carrier’s inspection. Check that the equipment is complete and is in an undamaged condition, and then contact ETS-Lindgren Customer Service for further advice.
Safety Information
- This RF amplifier has been designed and tested in accordance with BS EN61010-1 and has been supplied in a safe condition. This manual contains warning information which must be followed to ensure safe operation and to retain the apparatus in a safe condition.
- This apparatus does not incorporate components liable to explode or implode during normal operating conditions.
- In normal operating conditions this apparatus does not liberate injurious or poisonous gases.
- Sound levels of this apparatus in a rack are below 85dBA as required by EN61010-1.
- If local regulations have a lower limit, appropriate action should be taken in line with the local regulations so as to comply.
- This apparatus is of Installation Category 2.
Warnings
This apparatus is capable of delivering harmful levels of radio frequency power. Ensure at all times during operation that the RF output is properly terminated with an adequately rated termination or transducer and that the cables and connectors attached to the apparatus are in good condition. The use of suitably rated mains cords must only be used as specified within this document, see Rear Panel, Mains Input on Page 15. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor. The opening of covers or removal of parts is likely to expose live parts. This apparatus must only be serviced by qualified service personnel and must be disconnected from all voltage sources before it is opened for adjustment, replacement, maintenance, or repair. Make sure that only fuses of the required rated current and of the specified type are used for replacement. The use of makeshift fuses and the short-circuiting of fuse holders are prohibited.
Installation
CAUTION: Before connecting any components, follow the information provided in Before You Begin on page 12.
Mains Input
- Mains input is via a 10A power cable (supplied).
- Access for mains disconnection must be maintained at all times.
Amplifier Protection
- The amplifier is protected by a fuse of the following rating, T 6.3A H 250V 20mm.
- This instrument must be earthed.
WARNING: If the rack mounting option has been selected, the unit must be installed in the rack using a suitable tray or runners. The amplifier must not be supported only by the front panels.
Operation
CAUTION: Before placing into operation, follow the information provided in Before You Begin on page 12.
Note: Make sure you are satisfied with the contents and condition of your amplifier prior to placing it into operation.
Warning: Avoid exceeding the stated maximum input level as this can cause a risk of damaging the equipment.
Warning: There is a significant risk of an electrical or burn hazard; to avoid this, always ensure the output is terminated into a suitable load prior to switching on the equipment.
Equipment Use
- The equipment must only be used for its intended purpose.
- All parts of the equipment must remain connected to ensure that protective circuits operate. Disconnecting parts of the equipment may impair operation and safety.
Mains Switch
The amplifier is switched on or off by pressing the mains switch on the rear panel of the amplifier.
Standby/Operate Switch
Pressing the standby/operate switch toggles the state of the amplifier between Standby and Operate modes. The blue halo indicator flashes in Standby mode, whereas steady blue indicates Operate mode.
Fault Indicator (Red LED)
Should a fault be detected, the red LED on the amplifier front panel will be illuminated. The amplifier will be forced into standby mode and the unit will cease to function. This action is latching, therefore only when the fault has been cleared can the amplifier be reset; this is done by pressing the standby/operate switch. The reasons for the system fault light to illuminate are as follows:
- Excess module temperature
- Air cooling failure (fans failing to operate)
Interlock Indicator (Orange LED)
When the rear panel interlock is activated the interlock indicator will illuminate and the amplifier gain stages will all be turned off. This action is latching; therefore, the amplifier will return to Standby mode when the interlock is released.
Local Lockout Indicator
This indicator is used with the USB/GPIB/Ethernet interface. Refer to the separate Remote Control Interface manual supplied.
RF Input (Rear Panel Mounted)
The RF input will accept a signal from an RF generator. This input must be within the operating frequency range of the 8000-035 amplifier. An amplitude of up to 0dBm will be sufficient to saturate the amplifier. Operation outside the specified frequency range should not be attempted as it may subject the amplifier to undue internal stress.
RF Output (Rear Panel Mounted)
WARNING: The center conductor of the RF output represents a severe burn hazard to personnel. Never operate the amplifier without a proper termination or with defective cables or connectors.
This connector must be suitably terminated at all times during operation.
Ensure that the cable is capable of handling the power available, which may be
as much as
85 watts.
Air Inlet
This amplifier requires a free air supply for cooling; the front air inlet must not be restricted.
Safety Interlock
- A dual interlock system can be operated by applying either short circuit (S/C) or open circuit (O/C) to mute the amplifier. In order for the amplifier to operate, the Interlock BNC connector must be S/C and the Interlock must be O/C. If the Interlock is O/C or the Interlock is S/C the amplifier will be muted.
- The amplifier is delivered without the interlock applied and will be ready to operate.
- The interlock function is latching so when the interlock is removed the amplifier will be in standby mode. Press the standby/operate switch to return to normal operation.
- Current drawn from the interlock BNC connectors under short circuit conditions will typically be 2.8mA.
Fan Outlet
This amplifier requires a free air supply for cooling; the rear air outlets must not be restricted.
Appendix A: EC Declaration of Conformity
ETS-Lindgren Inc. declares these products to be in conformity with the following standards, following the provisions of EMC Directive 2014/30/EU and Low Voltage Directive 2014/35/EU:
Technical Construction Files are available upon request.