ETS-LINDGREN 7001-0XX EMSwitch RF Switch Plug In Card
EMSwitchTM RF Switch Plug-In Card
Specifications :
- Models: 7001-0XX
- Trademark: ETS-Lindgren Inc.
- Revision Record: Part #399343, Rev. K
- Copyright: ETS-Lindgren Inc.
- Trademark: ETS-Lindgren logo (registered), EMCenter, EMSwitch, TILE!, EMQuest
Product Information
The EMSwitch RF Switch Plug-In Card is designed to provide seamless switching capabilities in RF applications. It is a reliable and efficient solution for managing signal routing and control.
Safety Information:
-
High Voltage: Indicates hazardous voltage. Follow proper safety measures.
-
Protective Earth Ground (Safety Ground):
Ensure a safety earth ground connection for safe operation. -
Laser Warning: Laser (class 1M) is part of the device’s operating system.
-
Waste Electrical and Electronic Equipment (WEEE) Directive: Dispose of the product at an appropriate recycling facility at the end of its useful life.
Notes, Cautions, and Warnings:
- Note: Provides helpful information for better product use.
- CAUTION: Indicates potential hazards resulting in minor injury or property damage if instructions are not followed.
- WARNING: Indicates severe hazards resulting in personal injury or property damage if instructions are not followed.
Product Usage Instructions
Standard Configuration:
The EMSwitch RF Switch Plug-In Card is designed for easy installation in
compatible systems. Follow these steps for standard configuration:
- Ensure the power source is disconnected before installation.
- Insert the card into the designated slot carefully to avoid damage.
- Secure the card in place according to the system’s instructions.
- Connect any required cables or interfaces to the card.
- Power on the system and configure the switch settings as needed.
EMSwitch Card Models:
The EMSwitch RF Switch Plug-In Card is available in various models to suit
different RF applications. Refer to the product manual for specific details on
each model’s features and compatibility.
Other Optional Items:
Additional optional items such as remote relay modules or accessories may
enhance the functionality of the EMSwitch RF Switch Plug-In Card. Check with
the manufacturer for compatible add-ons.
FAQ:
-
Q: Can I use the EMSwitch RF Switch Plug-In Card with other brands of equipment?
A: The compatibility of the EMSwitch card with other equipment may vary. It is recommended to consult with ETS-Lindgren or refer to the product manual for detailed compatibility information. -
Q: How do I troubleshoot if the switch card is not functioning properly?
A: If you encounter issues with the switch card, first check the connections and power supply. Ensure proper installation and configuration according to the manual. If problems persist, contact customer support for assistance.
EMSwitch™ RF Switch Plug-In Card
Models 7001-0XX
Product Manual
Copyright and Trademark
- ETS-Lindgren Inc. reserves the right to make changes to any products herein to improve functioning or design. Although the information in this document has been carefully reviewed and is believed to be reliable, ETS-Lindgren does not assume any liability arising out of the application or use of any product or circuit described herein; nor does it convey any license under its patent rights nor the rights of others. All trademarks are the property of their respective owners.
- © Copyright 2014–2024 by ETS-Lindgren Inc. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren Inc.
Trademarks used in this document: The ETS-Lindgren logo is a registered trademark, and EMCenter, EMSwitch, TILE!, and EMQuest are trademarks of ETS- Lindgren Inc.
Revision Record
MANUAL, EMSWITCH | Part #399343, Rev. K
Revision | Description | Date |
---|---|---|
A | Initial Release | JUN, 2014 |
B | Updated EMCenter models | JUN, 2014 |
C | Added Model 7001 005; updated EMSwitch Remote Relay | JUL, 2015 |
D | Updated Typical Data with main specifications | MAR, 2016 |
E | Updated physical specifications | DEC, 2016 |
F | Corrected specs on page 39-40; added replacement part information | MAY, |
2020
G| Removed incorrect warranty info| SEP, 2020
H| Add 7001-004 Commands| JUN, 2022
J| Updated format| SEP, 2022
K| Update| FEB, 2024
Safety Information
- High Voltage: Indicates presence of hazardous voltage. Unsafe practice could result in severe personal injury or death.
- Protective Earth Ground (Safety Ground): Indicates protective earth terminal. You should provide uninterruptible safety earth ground from the main power source to the product input wiring terminals, power cord, or supplied power cord set.
- Laser Warning: Denotes a laser (class 1M) is part of the operating system of the device.
- Waste Electrical and Electronic Equipment (WEEE) Directive: (European Union) At end of useful life, this product should be deposited at an appropriate waste disposal facility for recycling and disposal. Do not dispose of with household waste.
Notes, Cautions, and Warnings
- Note: Denotes helpful information intended to provide tips for better use of the product.
- CAUTION: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures.
- WARNING: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures.
Introduction
- The ETS Lindgren EMSwitch™ RF Switch Plug in Card is a general purpose multi-channel switch matrix used to switch the RF path of equipment for RF measurement applications, including immunity, emissions, and wireless measurements. EMSwitch is designed for use with the EMCenter™ Modular RF Platform.
- The first relay of each EMSwitch card can be used as a standard relay or as a safety interlock relay. When being used as a safety interlock relay, the RF input signal to the RF amplifier can be switched off to prevent personnel from being subjected to high RF fields. For example, the RF interlock input can be connected to a switch mounted on the entrance door of the test chamber.
Standard Configuration
Standard configuration may vary by model; contact ETS-Lindgren for additional
information.
-
EMCenter Modular RF Platform
Some Models may require a specific processor; See Processor Requirements for more information. -
EMSwitch Plug-In card
EMSwitch cards switch RF signals up to 40 GHz, depending on the model, with powers up to 240 W (3 GHz) directly or any RF power switches indirectly. When high power RF amplifiers are used in a test system, the EMSwitch card can be connected to an optional EMSwitch Remote Relay Module. -
Interlock
When one of the relays is used as a safety interlock relay, it can be used to switch of the RF input signal to the RF amplifier. A switch mounted on the entrance door of the test chamber can for example, activate the RF interlock input. This can prevent personnel from being subjected to high RF fields. This interlock input is fully integrated in the hardware; software cannot overwrite this safety system.
EMSwitch Card Models
| Number of Relays| Frequency Band| Processor
---|---|---|---
7001- 001| 2x SPDT| ****
18 GHz
|
7001- 002| 4x SPDT|
7001- 003| 2x SP6T|
7001- 005| 1xSP6T|
7001- 011| 2x SPDT| 40 GHz|
7001- 012| 4x SPDT|
7001- 013| 2x SP6T|
7001- 015| 1xSP6T|
7001- 021| 1x SPDT| 12.4 GHz|
7001 022| 1x SPDT| 3 GHZ|
7001- 023| 2x SPDT| 50 GHZ| ARM Only *
See Processor Requirements for more information
7001- 024| 4x SPDT
7001- 025| 1x SPDT
7001- 026| 2x SPDT| 67 GHZ
7001- 027| 4x SPDT
Other Optional Items
- The Model 7001-004 Remote Relay is an optional 19-inch 2U rack mountable device to control up to four external (coax) relays using the touchscreen on the EMCenter or with one of these software products:
- ETS-Lindgren TILE!™ (Totally Integrated Laboratory Environment)
- ETS-Lindgren EMQuest™ Data Acquisition and Analysis Software
- Other test automation software
- It has an internal power supply to power 12 VDC/28 VDC relays.
ETS-Lindgren Product Information Bulletin
See the ETS-Lindgren Product Information Bulletin included with your shipment
for the following:
- Safety, regulatory, and other product marking information
- Steps to receive your shipment
- Steps to return a component for service
- ETS-Lindgren calibration service
- ETS-Lindgren contact information
Processor Requirement
- Note: The 2-slot and 7-slot EMCenter require an ARM processor board with firmware version 3.3.3 or higher to some EMSwitch interface cards. 2-slot and 7-slot EMCenter systems utilizing an X86 processor board and/or firmware version 3.3.0 or older are not supported and will not communicate with the interface card.
- To view the 2-slot or 7-slot EMCenter version information, navigate to the main screen and press the Info button. Slot 8 indicates the processor version of this EMCenter as shown in the graphic below.
- Note: If the EMCenter is not operating the 7000-008 or 7000-009 processor card, contact ETS-Lindgren to purchase a new 2-slot or 7-slot EMCenter or a new processor card compatible with EMSwitch.
Model Number | Processor Type |
---|---|
7000-008 | EMCENTER PROCESSOR BOARD STD (ARM) |
7000-009 | EMCENTER PROCESSOR BOARD + GPIB (ARM) |
Operation
Installation
- CAUTION: Before connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
- WARNING: Always unplug the unit before starting maintenance to prevent electrical shock. Maintenance includes removal of the plug-in cards or the top panel.
Changing Plug-In Cards
Note: Due to the width of the 7001-003, 7001-005, 7001-013, and 7001-015
EMSwitch cards, two consecutive empty slots are required for installation.
- Before installing and inserting a new plug-in card make sure that the EMCenter is turned OFF and disconnect the AC Mains power cord.
- Determine in which empty slot of the EMCenter you want to install the Plug-in Card. Looking at the back of the EMCenter, the slots are numbered 1 through 7 from left to right.
- Remove the blank panel from the slot by removing the two screws at the top of the blank panel and the two screws at the bottom. Be careful not to lose the screws.
- Carefully insert the card into the slot of the EMCenter. Position the plug- in card into the slot and slowly push it, using the lower part of the plug-in card. When it reached the end of the rail, gently push and lock the plug-in card into the backplane socket.
- Secure the card by tightening the four previous screws using a Pozi type screwdriver head PZ1.
- Plug the interlock into the connector on the back of the EMCenter.
- Connect the desired device(s) to the correct plug-in card port.
- Optional: Connect additional local interlock to the correct plug-in card.
- Optional: Connect the EMCenter to a computer using Ethernet or GPIB.
- Re-connect the AC mains power cord and turn ON the EMCenter. It can now be started by tapping the touch screen. The EMCenter will automatically detect the newly installed card.
- The card installation is complete and the EMCenter is now ready for use. You can control all cards through the touchscreen or sending remote commands.
Remote Relay
- CAUTION: Before placing into operation, follow the information provided in Safety Information document.
- For applications that require to switch higher power signals (>700 Watts), it is not possible to use internal relays of the switch cards; in those cases, special dedicated relays are needed in the test setup. For these external relays, EMSwitch™ Model 7001 004 Remote Relay, which is capable of driving two external relays, is available.
- Model 7001 004 can drive any relay with a supply voltage of 12V, 24V, or 28V, and from SPDT to SP6T. For each relay a driver current of 0.5A is available, or 1A if only one relay is connected.
- The type of relay and the usage of indicator contacts can be configured in software. The connection for relay 1 and relay 2 are identical.
Front and Back Panel Connectors
- The front panel includes a key switch and an LED. If the EMSwitch™ Remote Relay is connected to a 220 V AC power supply and the on/off switch on the back panel is set to the I position, you can power on the remote relay by turning the key clockwise. The LED will illuminate.
- The Remote Relay must be connected to a 220 V AC power supply via the mains lead, connected at the AC inlet on the back panel.
Relays
- You can connect four relays to the remote relay. It is possible to connect relays with 1 input and 2, 3, 4, 5, or 6 outputs.
- To connect external relays you must use the mating connector set included with the remote relay.
- The relay x drive connectors on the back panel of the remote relay will switch the relays into position.
- For a 24 V relay, connect pin 1 (24 V DC output) to the Common of the relay power terminals.
- For a 12 V relay, connect pin 2 (12 V DC output) to the Common of the relay power terminals.
- Connect pin 3–8 to the power terminals 1–6 of the relay.
- The number of power terminals to connect depends on the type of relay you use. For example, a 1–6 relay requires all of the pins, and a 1–3 relay requires three pins.
Readback Function
The relay x readback connectors on the back panel remote relay can verify if
the external relay is set into position.
- Connect pin 1 (3v3 out) to the indicator common of the relay.
- Connect pin 2 to the first indicator of the relay.
- Connect pin 3 to the second indicator; continue until complete.
The number of indicator terminals to connect depends on the type of relay you
use. Pin 8 is not in use.
Set the readback function on or off in the EMCenter™ Modular RF Platform
depending on the number of relay(s) with readback function that you use.
Set Up RS 232 Address
You can connect up to four external switch boxes to one EMCenter. Interconnect
the switch boxes with straight RS 232 cables. The DIP switches on the back
panel of the EMCenter allow you to set the individual addresses.
DIP Switch | EMSwitch Remote Relay Address |
---|---|
1 | 2 |
On | On |
Off | On |
On | Off |
Off | Off |
Manual Control
Back Panel
The EMSwitch has the following Connections:
- SPDT 1x and 4x
- SP6T
- External Relay
Relay Number and Contact Definitions
7001-001, 7001-002, 7001-011, 7001-012
Relay number| Indicated below each relay, A to D.
Common| The center SMA connector of each relay, labeled COM.
Normally Open| The left SMA connector of each relay, labeled NO.
Normally Closed| The right SMA connector of each relay, labeled NC.
7001-003, 7001-005, 7001-013, 7001-015
Common| The center connector; the six contacts are labeled J1 to J6.
7001- 021
Common| The center N connector, labeled COM.
Normally Open| The bottom N connector, labeled NO.
Normally Closed| The top N connector, labeled NC.
SPDT Connections
- The connections of the SPDT (Single Pole Double Throw) relays are marked with “NO”, “COM” and “NC” for respectively Normally Open, Common, and Normally Closed.
- These terms indicate the position of the relay. When not activated, the relay connects the COM to the NC connector. When energized, the relay connects the COM to the NO connector.
SP6T Connections
The center connector is the COM (Common). The other connections are numbered
from one to six; this is visible in the illustration below and on the relay
itself.
External Relay Plug-In Card
When the models with internal relays do not fit the needs (E.G. high-power switching), there is the 7001-004; this plug-in card is designed to use with external relays and can drive up to two (2) SP6T relays. It provides a user selectable supply of 12V, 24V, or 28V, with a maximum current of 0.5A per relay (voltage independent). The 7001-004 is also equipped with indicator contacts; these can provide feedback to the user on the state of the relay.
- Connector
- The 7001-004 is shipped with four (4) connectors* to mate with the plug-in card connectors. A 2×8-way terminal block is used for each relay driver. The left row is used for energizing the coils (please note that only one coil can be active at a time).
- The right row is used for the readback contacts of the relays. The connections for the indicator contacts are optional and can be enabled or disabled in the configuration screen of the switch card.
- Coil connections
- The relay coil(s) must be connected on the left row between the V+ (common) and one of the six coil connections. The V+ is a user selectable 12V, 24V, or 28V.
- The voltage can be set in the configuration screen of the plug-in card.
- WARNING: Make sure that the correct voltage is set before connecting the relay. Wrong settings may result in permanent damage to the relay!
- The coil connections are open collector outputs, which connect the coil to ground when, activated, thus energizing the relay coil.
- Indicator connections
- Some relays are equipped with indicator contacts; these contacts indicate the position of the relay. The indicator contacts can be connected on the right row. The indicator contacts can be connected between the +3.3V and the corresponding input channel (the input channel corresponds to the same coil number).
- Interlock
- The interlock connector provides two floating contacts, which require shorting for the first relay.
- If the interlock is open, the first relay will illuminate red on the EMCenter™ Modular RF Platform screen and the relay cannot be used. Use the supplied connector to wire to the emergency switch of your site. Connect the two pins of the connector.
- Note: For replacement interlock connectors, contact ETS-Lindgren to order part number 1608925.
Touchscreen
- Home Screen To change relay position settings, on the Home screen press the status box to the right of the slot number for the installed EMSwitch plug in card. This will display the following switch screen:
- EMControl Main Settings Screen
- (COM — Common; NO — Normally Open; NC — Normally Closed)
- Press a switch button (switch A, switch B, and so on) to toggle between NO and NC. In case of a failure, an error will be displayed in the screen.
- Safety Interlock Relay
- Relay A on each EMSwitch plug in card can be used as a safety interlock relay. As long as the connector is shorted (by an external interlock switch), the relay can be activated.
- However, as soon as the interlock circuit opens, relay A will return to the idle position.
- Note: If you do not want to use this relay as a safety relay, the connector must be shorted in order to use it as a standard relay.
- When the safety interlock relay is activated, the power supply to the first relay of the card will be interrupted (fully hardware controlled), causing the relay to switch to the normally open position. The EMCenter will detect this interlock status and display a message. For safety reasons, the software is not able to overrule this interlock condition.
- After the interlock has been closed again, the relay will switch automatically to its original condition.
- Sample Interlock Error Screen
The interlock hardware is implemented in a redundant way for maximum safety. The interlock relay can be used to interrupt the signal path between the signal generator and the power amplifier. An interlock switch mounted on the door of the test chamber can be used to activate the safety relay, preventing personnel from being subject to high field strengths when entering the test chamber.
Relay Errors
The EMSwitch plug in card checks for the following error conditions of the
internal relays:
- Over temperature — A temperature sensor close to each relay checks for excessive heating of the relay. A high temperature error will display if the relay temperature exceeds 85°C (185°F).
- Relay errors — The EMSwitch checks the following error conditions of the internal relays.
- Switching error — Each internal relay has a set of control contacts, which are used to check if a relay has changed position. A switching error message will appear if this check fails.
- Interlock open for safety interlock relay — When the interlock of relay A is opened, this will result in a safety interlock open error.
To use the EMSwitch remote commands, read the manual document 399342 EMCenter for examples and a full command list.
EXAMPLE: SET EXTERNAL RELAY
To set the external relay 3 from the EMSwitch Remote Relay with address 4,
connected to the EMSwitch card at slot 2, position 5: S2:N14RELAY_3_5
Character | Definition |
---|---|
S | Device character of EMSwitch card |
2 | Board number of EMSwitch card |
N1 | Indicates EMSwitch card to pass message to external interface |
4 | Address of EMSwitch Remote Relay (configurable with DIP switches) |
RELAY_ | Message to switch relay |
3 | Number of switch on EMSwitch Remote Relay |
_ | Separator |
5 | Instruction to put switch in position 5 |
Maintenance
- CAUTION: Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
- WARNING: Maintenance of the EMCenter is limited to external components such as cables or connectors. If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service.
- ****If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service.
Fiber Optic Maintenance
Fiber optic connectors and cables can be damaged from airborne particles,
humidity and moisture, oils from the human body, and debris from the
connectors they plug into. Always handle connectors and cables with care,
using the following guidelines.
- CAUTION: Before performing any maintenance, disconnect fiber optic cables from the unit and turn off power.
- When disconnecting fiber optic cables, apply the included dust caps to the ends to maintain their integrity.
- Before connecting fiber optic cables, clean the connector tips and in-line connectors.
- Before attaching in-line connectors, clean them with moisture-free compressed air.
- Failure to perform these tasks may result in damage to the fiber optic connectors or cables.
Service Procedures
-
Contacting ETS-Lindgren
Note: Please see www.ets-lindgren.com for a list of ETS-Lindgren offices, including phone and email contact information. -
Sending a Component for Service
For the steps to return a system or system component to ETS-Lindgren for service, see the Product Information Bulletin included with your shipment. -
Calibration Services and Annual Calibration
See the Product Information Bulletin included with your shipment for information on ETS-Lindgren calibration services.
Specifications
Note: The first switch on each EMSwitch card can be used as a true interlock switch.
Performance Specifications
| Number of Relays| Frequency Band| RF Switching Capacity
---|---|---|---
7001- 001| 2x SPDT| 18 GHz|
- DC to 3 GHz: 240 W
- 3 to 8 GHz: 150 W
- 8 to 12.4 GHz: 120 W
- 12.4 to 18 GHz: 100 W
7001- 002| 4x SPDT
7001- 003| 2x SP6T
7001- 005| 1xSP6T
7001- 011:| 2x SPDT| 40 GHz|
- DC to 6 GHz: 80 W
- 6 to 12.4 GHz: 60 W
- 12.4 to 18 GHz: 50 W
- 18 to 26.5 GHz: 20 W
- 26.5 to 40 GHz: 10 W
7001- 012:| 4x SPDT
7001- 013:| 2x SP6T|
- DC to 6 GHz: 40 W
- 6 to 12.4 GHz: 30 W
- 12.4 to 18 GHz: 25 W
- 18 to 26.5 GHz: 15 W
- 26.5 to 40 GHz: 5 W
7001- 015:| 1xSP6T
7001- 021:| 1x SPDT| 12.4 GHz|
- DC to 1 GHz: 700 W
- 1 to 2 GHz: 500 W
- 2 to 3 GHz: 400 W
- 3 to 8 GHz: 250 W
- 8 to 12.4 GHz: 200 W
Lifetime Relays
- SPDT relays, SMA or 2.92mm (k) — 10,000,000 cycles
- SP6T relay SMA — 5,000,000 cycles
- SP6T relay 2.92mm (k) — 2,000,000 cycles
- N type relay — 1,000,000 cycles
Electrical Specifications
All EMSwitch™ RF Switch Plug in Cards perform to the following electrical
specifications:
- Note: The supply voltage for the Model 7001 004 Remote Relay is 230 VAC.
- Supply Voltage (Volts) — Through EMCenter
- Power Consumption (Max Watts) — 30 W
Physical Specifications
| Exterior Dimension| RF Connectors| Remote Control External
Relays
---|---|---|---
7001- 001| One slot| SMA type| Fiber optic link
7001- 002
7001- 003| Two slots
7001-004 (H x W x D)| 2U x 250 mm x 82.6 mm 2U x 9.8 in x 19 in| SMA
type
7001- 005| Two slots| SMA type
7001- 011| One slot| k type 2.92 mm
7001- 012
7001- 013| Two slots
7001- 015
7001- 021| One slot| N type
| |
Environmental Specifications
All EMSwitch cards perform to the following environmental specifications:
- Temperature Range — 0°C to 40°C (32°F to 104°F)
- Relative Humidity — 10% to 90% (non-condensing)
Typical Data **
**
Specifications of plug-in cards with SMA connectors
Specification | SMA, 18 GHZ, SPDT relays |
---|---|
Life Time | 10.000.000 cycles |
Frequency | GHz |
18
VSWR| | 1.10| 1.20| 1.20| 1.40
Insertion Loss| dB| 0.15| 0.20| 0.25| 0.35
Isolation| dB| 80| 75| 65| 60
Average Power| W| 240| 150| 120| 100
Specification| SMA, 18 GHZ, SP6T relays
---|---
Life Time| 5.000.000 cycles
Frequency| GHz| 0 to 3| 3 to 8| 8 to 12.4| 12.4 to
18
VSWR| | 1.20| 1.30| 1.40| 1.50
Insertion Loss| dB| 0.20| 0.30| 0.40| 0.50
Isolation| dB| 80| 75| 65| 60
Average Power**| W| 240| 150| 120| 100
Specifications of plug-in cards with 2.92 mm connectors
Specification | K 2.92 mm, 40 GHz, SPDT relays |
---|---|
Life Time | 10.000.000 cycles |
Frequency | GHz |
to 6.5** | 26.5 to 0 |
VSWR | |
Insertion Loss | dB |
Isolation | dB |
Average Power | W |
Specification | K 2.92 mm, 40 GHz, SP6T relays |
--- | --- |
Life Time | 2.000.000 cycles |
Frequency | GHz |
to 6.5** | 26.5 to 0 |
VSWR | |
Insertion Loss | dB |
Isolation | dB |
Average Power | W |
Power Handling
- The RF power rating is the capability of handling RF power (CW power) through closed contacts. The RF power should be removed (turned off) before switching the relay.
- Power ratings assume the following specifications; see the power handling data chart above.
- Unity VSWR (matched load)
- Room temperature (25°C / 77°F)
- Sea level pressure (14.7 psi)
- Cold switching
- Changes to these specifications require power derating; see the VSWR data chart below.
VSWR
The average power input must be reduced for load VSWR above 1:1.
Appendix A: EC Declaration on Conformity
ETS-Lindgren Inc. declares these products to be in conformity with the following standards and provisions:
-
Product EMSwitch RF Switch Plug In Card
-
Models:
-
Directives: EMC Directive 2014/30/EU
RoHS Directive: 2015/863/EU -
Emission: EN 61326-1:2013, Class A1
Electrical equipment for measurement, control and laboratory use. -
Immunity: EN 61326-1:2013, Industrial level, performance criteria A
Electrical equipment for measurement, control and laboratory use.
Technical Construction Files are available upon request.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>