ETS-LINDGREN EM8 Modular Test System Instruction Manual
- June 1, 2024
- ETS LINDGREN
Table of Contents
ETS-LINDGREN EM8 Modular Test System
Copyright and Trademark
ETS-Lindgren Inc. reserves the right to make changes to any products herein to improve functioning or design. Although the information in this document has been carefully reviewed and is believed to be reliable, ETS-Lindgren Inc. does not assume any liability arising out of the application or use of any product or circuit described herein; nor does it convey any license under its patent rights nor the rights of others. All trademarks are the property of their respective owners.
© Copyright 2023-2024 by ETS-Lindgren Inc. All Rights Reserved. No part of
this document may
be copied by any means without written permission from ETS-Lindgren Inc.
Trademarks used in this document: The ETS-Lindgren logo is a registered
trademark, and EM Center,
EM Switch, EM Power, EM Control, EM Sense, EM Sense 10/40, EM Link, EM Field,
TILE!, and EM Quest are trademarks of ETS-Lindgren Inc.
Revision Record
MANUAL, EM8 | Part #1788365, Rev. A
Revision | Description | Date |
---|---|---|
A | Initial Release | FEB, 2024 |
Safety Information
High Voltage: Indicates presence of hazardous voltage. Unsafe practice could result in severe personal injury or death.
Protective Earth Ground (Safety Ground): Indicates protective earth terminal. You should provide uninterruptible safety earth ground from the main power source to the product input wiring terminals, power cord, or supplied power cord set.
Laser Warning : Denotes a laser (class 1M) is part of the operating system of the device.
Waste Electrical and Electronic Equipment (WEEE) Directive: (European Union) At end of useful life, this product should be deposited at an appropriate waste disposal facility for recycling and disposal. Do not dispose of with household waste.
Notes, Cautions, and Warnings
Note: Denotes helpful information intended to provide tips for better use of the product.
CAUTION: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures.
WARNING: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures.
Introduction
ETS-Lindgren’s EM Center™ EM8 is the latest member of the EM family of
products capable of significantly higher data collection speeds.
The EM Center™ EM8 Modular Test System forms the heart of many different EMC
and RF test systems. The EM Center has the ability to store up to eight (8)
instrument plug-in cards, which can all be controlled and used simultaneously.
The 1 Gigabit LAN interface, providing some of the fastest acquisition in the
industry, makes the EM Center EM8 the ideal hardware platform for applications
requiring multiple channels and fast measurements, like automotive mode-stir
(reverberation chamber) testing or switching applications.
Product Features
-
Fast Measurements
- The EM Center EM8 is equipped with a 1 Gigabit LAN interface. Each instrument that is installed in the eight slots can be directly controlled at a designated IP address, allowing for fast measurement throughout.
-
Laser safety
- The EM Center EM8 is equipped with a laser ON/OFF switch to prevent accidental operation and/or starting the laser power supply of the EM Sense™ probes.
-
Flexible
- Each of the eight slots can be configured by the user, allowing any combination of
plug-in cards. All cards can be commanded and operated using TILE! software.
- Each of the eight slots can be configured by the user, allowing any combination of
-
Scalable
- The EM Center can be scaled using a wide array of available plug-in cards including:
- EM Sense™ LASER powered E-Field Sensors
- EM Field™ Integrated EMC Immunity Test Solution
- EM Power™ RF power meters for EMC application
- EM Gen™ RF signal generators
- EM Switch™ for switching one, two, four or six RF signal
8-Slot Example Application
The international standard ISO11451-5 vehicle test methods for electrical
disturbances from narrowband radiated electromagnetic energy – part 5:
reverberation chamber describes the measurement of E-fields at eight reference
points as part of the immunity test in a reverb chamber.
For this application, the EM Center EM8 can be configured with eight EM Sense
laser powered E-field probes, allowing the fast and accurate measurement of
the isotropic field strength within the working volume of the chamber.
Utilizing multiple switches, the EM Center EM8 allows for simultaneous
communication, increasing the rate of switch state changes and reducing test
lag time.
Operation
Installation
CAUTION: Before connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
WARNING: Always unplug the unit before starting maintenance to prevent electrical shock. Maintenance includes removal of the plug-in cards or the top panel.
CAUTION: Leave an empty slot or space 1U in height beneath the EM Center to allow sufficient cooling through the bottom air inlets of the cabinet.
Changing Plug-In Cards
- Before installing and inserting a new plug-in card make sure that the EM Center is turned OFF and disconnect the AC Mains power cord.
- Determine in which empty slot of the EM Center you want to install the Plug-in Card. Looking at the back of the EM Center, the slots are numbered 1 through 8 from left to right.
- Remove the blank panel from the slot by removing the two screws at the top of the blank panel and the two screws at the bottom. Be careful not to lose the screws.
- Carefully insert the card into the slot of the EM Center. Position the plug- in card into the slot and slowly push it, using the lower part of the plug-in card. When it reached the end of the rail, gently push and lock the plug-in card into the backplane socket.
- Secure the card by tightening the four previous screws using a Pozi type screwdriver head PZ1.
- Plug the interlock into the connector on the back of the EM Center.
- Connect the desired device(s) to the correct plug-in card.
- Connect the EM Center to a computer using Ethernet.
- Re-connect the AC mains power cord and turn ON the EM Center. It can now be started by tapping the touch screen. The EMC enter will automatically detect the newly installed card.
- The card installation is complete and the EM Center is now ready for use. You can control the cards by sending remote commands.
Interface and Cable Setup
CAUTION : Never disconnect the fiber optic cables when one of the lasers is active.
CAUTION : Before placing into operation, follow the information provided in Safety Information on page v.
CAUTION : Prior to operation, verify that the mains voltage is within the operating range of the equipment.
Front Panel
Power Button
How long you press the power button determines which function is used:
- ON/OFF (1 sec)
- To power ON the EM8, press and hold the power button for one second until the LED ring around the button lights up in white. After a brief moment, the LED ring will begin flashing, indicating the booting process. Once the LED stops flashing and continues to burn steadily, the EM8 is now ready for use.
- To power OFF the EM8, press and hold the power button one second until an audible beep is heard. After a brief pause, the LED ring will initiate flashing, signaling the shutdown process. Once the LED ceases flashing and turns off, the EM8 is fully powered down.
- Reset IP addresses to default (10 secs)
- While the EM8 is in normal operation, hold the power button for about 10 seconds. Release the button within two seconds after the two short audible beeps to initiate the reset of the IP addresses to their default values.
- NOTE: It is important not to continuously press the button, as doing so will prevent the IP address reset.
- Full Factory reset
- In the event of lost credentials or the need for reconfiguration of the EM8, a reset to default settings is possible. Factory Reset includes default IP addresses, default software version, default password.
- To perform a factory reset, follow these steps:
- Begin by shutting down the EM8.
- Turn off the mains power, wait for 20 seconds, and then turn the power back on.
- Hold the power button for approximately 30 seconds. Release the button immediately after hearing two short audible beeps.
- It’s crucial not to keep pressing the button, as this will prevent the factory reset. Releasing the button within two seconds after the two beeps will execute the factory reset.
LASER Button
When the EM8 is equipped with plugin cards, including one or multiple plugin
cards containing lasers, the LASER ON button takes on a critical role. Its
primary function is to activate the lasers safely simultaneously on all
installed plugin cards.
To maintain safety standards, initiating the laser power in the plugin cards
follows a twostep process: entering a designated ‘laser code’ and manually
activating the lasers.
- To activate the laser(s) in the EM8, follow these steps:
- Press the ‘Laser ON’ button on the front of the EM8 and hold it.
- Five loud ‘beep’ sounds should be heard.
- On the fifth ‘beep,’ the laser source is activated, and the red LED ring around the button lights up.
- Release the ‘Laser ON’ button.
- This means that:
- If the activation process is to be interrupted, release the ‘Start’ button (before the fifth ‘beep’). The laser will not be activated.
- If the activation process is interrupted (and released ‘Start’ button by accident), the laser will not be activated.
- The user is warned by an audible warning when the start button is pressed by accident (i.e., without the intention to activate the laser).
- To deactivate the lasers, press the laser-on button once, and all lasers will be powered off.
- NOTE : If an error occurs with one of the laser plugin cards, an audible beep will notify you. To resolve this error, simply press the laser on button briefly. This action will clear the error, and the plugin card will stop beeping.
Back Panel
Mains
Plug the power supply cable into the mains inlet of the EM8, power the system by toggling the mains switch located beneath the mains inlet.
Ethernet
There is no display, so everything is controlled exclusively via the Ethernet connection. Therefore, an Ethernet cable must be connected to the rear of the EM8 and your network for proper functionality.
- The box itself a LAN address to change the system settings.
- Each slot can have its own LAN address and therefore be regarded as its own instrument.
Interlock Plug
The EM8 uses a hardware interlock plug (6,3mm ¼” stereo jack) which can be
used as a safety interlock. The hardware interlock plug provides two floating
contacts which require connection for certain plugin cards to operate.
By default, the supplied interlock plug is shorted and must be inserted into
the interlock input at the back side of the EM8. When a safety interlock is
required in your system, please cut the RED wire, and connect the open ends to
your interlock system.
Network configuration
The EM8 is factory configured with standard static Ethernet addresses:
Slot number
|
IP Address
|
Port number
---|---|---
EM8 system
|
192.168.1.200
|
80
Slot 1
|
192.168.1.201
|
27531
Slot 2
|
192.168.1.202
|
27531
Slot 3C
|
192.168.1.203
|
27531
Slot 4
|
192.168.1.204
|
27531
Slot 5
|
192.168.1.205
|
27531
Slot 6
|
192.168.1.206
|
27531
Slot 7
|
192.168.1.207
|
27531
Slot 8
|
192.168.1.208
|
27531
Subnet address
|
255.255.255.0
|
–
Remote Control
To use the EM Center remote commands, read the manual document 399342 EM Center for examples and a full command list.
EXAMPLE: Connect Via Pu TTY
If you want to establish a raw connection without specifying a particular
protocol in Pu TTY, you can follow these steps:
-
Open Pu TTY.
-
On the left menu select “Terminal”:
a. Enable “Implicit CR in every LF” and “Implicit LF in every CR” to automatically send a Carriage Return after every Line Feed and vice versa. -
On the left menu select “Session”:
a. Enter the IP address “192.168.1.201” in the “Host Name (or IP address)” field.
b. In the “Port” field, enter “27531” as the port number for the default configuration.
c. Under “Connection type,” choose the “Raw” option.
d. Under “Saved Sessions,” enter a name for your session in the “Saved Sessions” field (e.g., “EM8 Slot 1”).
e. Click the “Save” button to save your configuration.
f. Click the “Open” button to start the raw connection.
The provided configuration enables a raw connection without specifying a
particular protocol, facilitating direct interaction with the plugin card in
an unprocessed manner. This setup allows for the transmission of commands and
reception of responses from the plugin card.
Specifically, each plug-in card supports the command “*IDN?” for retrieving
its identification information. When the configuration and connection are
correct, issuing this command to the plug-in card will prompt it to return the
identifier string (ID) associated with the plug-in card.
Specifications
Physical Specifications
Height: | 132 mm (5.2 in)(3U) |
---|---|
Width: | 19 in rack mountable |
Depth: | 312 mm (12.3 in) excluding rear panel |
Weight: | ~15.4 lbs. (7kg) empty |
Environmental Specifications
Temperature Range: | 0°C to 40°C (32°F to 104°F) |
---|---|
Relative Humidity: | 5% to 95% (non-condensing) |
Power Specifications
Supply Voltage: | 115/230 VAC |
---|---|
Power Consumption: |
- Standby : < 3 W
- Empty : 33 W
- Max load : 225 W
Interface and Cable Specifications
Backplane|
- **** Ethernet
- 6.35mm (1/4″) jack
- Fused and switched IEC 320 C13 socket
- 8 blanking plates (replace with plug in cards)
---|---
Cables:| IEC power cord
Interfaces:| LAN
Interlock:| 6.35 mm jack plug connection to safety disables interlock &
interlocked laser outputs when triggered
Mounting:| Desktop or 19-in rack mountable
Slots:| 8 modular card slots
Appendix A: EC Declaration on Conformity
ETS-Lindgren Inc. declares these products to be in conformity with the
following standards and
provisions:
Product Models: | EM Center Modular RF Platform |
---|---|
Directives: | EMC Directive 2014/30/EU |
Low Voltage Directive 2014/35/EU RoHS Directive: 2015/863/EU
Emission:| EN 61326-1:2013, Class A1
Electrical equipment for measurement, control and laboratory use.
Immunity:| EN 61326-1:2013, Industrial level, performance criteria A
Electrical equipment for measurement, control and laboratory use.
Safety:| EN 61010-1:2010, Safety requirements for electrical equipment
for measurement, control, and laboratory use.
RoHs:| EN 63000:2018, Technical documentation for the assessment of
electrical and electronic products with respect to the restriction of
hazardous substances.
Technical Construction Files are available upon request.