SQ-515 Quantum Sensor
Owner’s Manual
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer:
Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: SQ-614
Type: ePAR Sensor
The object of the declaration described above is in conformity with the
relevant Union harmonization legislation:
2014/30/EU
Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
2015/863/EU Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory
use – EMC requirements EN 50581:2012 Technical documentation for the
assessment of electrical and electronic products with respect to the
restriction of hazardous substances
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated diphenyls (PBDE), bis(2- ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl
phthalate (DIBP). However, please note that articles containing greater than
0.1% lead concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on
our raw materials or end products for the presence of these substances, but
rely on the information provided to us by our material suppliers.
Signed for and on behalf of:
Apogee Instruments, March 2021
Bruce Bugbee
President
Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is almost universally quantified as photosynthetic photon
flux density (PPFD), the sum of photons from 400 to 700 nm in units of
micromoles per square meter per second (µmo m-2 s-1, equal to microEinsteins
m-2 s-¹). While microEinsteins and micromoles are equal (one Einstein = one
mole of photons), the Einstein is not an SI unit, so expressing PPFD as pmol
n–r2 s’ is preferred. Daily total PPFD is typically reported in units of moles
of photons per square meter per day (mol m-2 d–‘) and is often called daily
light integral (DLI).
The acronym PPF is also used and refers to the photosynthetic photon flux. The
acronyms PPF and PPFD refer to the same variable. Both terms are used because
there is not a universal definition of the term flux. Flux is sometimes
defined as per unit area per unit time and sometimes defined as per unit time
only. PPFD is used in this manual.
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation.
Typical applications of quantum sensors include measurement of incident PPFD
on plant canopies in outdoor environments or in greenhouses and growth
chambers, and reflected or under-canopy (transmitted) PPFD measurement in the
same environments.
Apogee Instruments SQ series quantum sensors consist of a cast acrylic
diffuser (filter), photodiode, and signal processing circuitry mounted in an
anodized aluminum housing, and a cable to connect the sensor to a measurement
device. SQ-500 series quantum sensors are designed for continuous PPFD
measurement in indoor or outdoor environments.
SENSOR MODELS
This manual covers the amplified voltage output quantum sensors, models SQ-512 and SQ-515 (listed in bold below). Additional models are covered in their respective manuals.
Model | Signal |
---|---|
SQ-512 | 0-2.5 V |
SQ-515 | 0-5 V |
SQ-500 | 0-40 mV |
SQ-520 | USB |
SQ-521 | SDI-12 |
SQ-522 | Modbus |
Serial Numbers 3636 (SQ-512), 3717 (SQ-515) and above: Sensor model
number and serial number are located on the bottom of the sensor.
If you need the manufacturing date of your sensor, please contact Apogee
Instruments with the serial number of your sensor.
Serial Number 0-3635 (SQ-512), 0-3716 (SQ-515) Sensor model number and serial number are located near the pigtail leads on the sensor cable. If you need the manufacturing date of your sensor, please contact Apogee Instruments with the serial number of your sensor.
SPECIFICATIONS
| SQ-512|
SQ-515
---|---|---
Power Supp| 5 to 24 V DC| 5.5 to 24 V DC
Current Draw| 12 V is 57 µA
Sensitivity| 0.625 mV per µmol m-2 s-1| 1.25 mV per µmol m -2 s-1
Calibration Factor (Reciprocal of Sensitivity)| 1.6 µmol m-2 s-1 per mV| 0.8
µmol m-2 s-1 per mV
Calibration Uncertainty| ± 5 % (see calibration Traceability below)
Measurement Range| 0 to 4000 µmol m-2 s-1
Measurement Repeatability| Less than 1 % (up to 4000 µmol m-2 s-1)
Calibrated Output Range| 0 to 2.5 V| 0 to 5 V
Long-term Drift (Non-stability)| Less than 2 % per year
Non-linearity| Less than 1 % (up to 4000 µmol m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range| 389 to 692 nm ± 5 nm (wavelengths where response is greater
than 50 %)
Spectral Selectivity| Less than 10 % from 412 to 682 ± 5 nm (see Spectral
Response below)
Directional (Cosine) Response| ± 2 % at 45º zenith angle, ± 5 % at 75º zenith
angle (see Directional Response below)
Azimuth Error| Less than 0.5 %
Tilt Error| Less than 0.5 %
Temperature Response| -0.11 ± 0.04 % per C (see Temperature Response below)
Uncertainty in Daily Total| Less than 5 %
Detector| Blue-enhanced silicon photodiode
Housing| Anodized aluminum body with acrylic diffuser
IP Rating| IP68
Operating Environment| -40 to 70 C; 0 to 100 % relative humidity; can be
submerged in water up to depths of 30 m
Dimensions Serial # 3636 (SQ-512), 3717 (SQ-515) and above| 30.5 mm diameter,
37 mm height
Dimensions Serial # 0-3635 (SQ-512), 0- 3716 (SQ-515)| 24 mm diameter, 37 mm
height
Mass (with 5 m of cable) Serial # 3636 (SQ- 512), 3717 (SQ-515) and above| 140
g
Mass (with 5 m of cable) Serial # 0-3635 (SQ-512), 0-3716 (SQ-515)
Cable| 100 g
5 m of shielded, twisted-pair wire; TPR jacket (high water resistance, high UV
stability, flexibility in cold conditions); pigtail lead wires; stainless
steel (316), M8 connector
Calibration Traceability
Apogee Instruments SQ-500 series quantum sensors are calibrated through side-
by-side comparison to the mean of four transfer standard quantum sensors under
a reference lamp. The reference quantum sensors are recalibrated with a
quartz halogen lamp traceable to the National Institute of Standards and
Technology (NIST).
Spectral Response
Mean spectral response measurements of six replicate Apogee SQ-100 (original) and SQ-500 (full-spectrum) series quantum sensors. Spectral response measurements were made at 10 nm increments across a wavelength range of 300 to 800 nm with a monochromator and an attached electric light source. Measured spectral data from each quantum sensor were normalized by the measured spectral response of the monochromator/electric light combination, which was measured with a spectroradiometer.
Temperature Response
Mean temperature response of ten SQ500 series quantum sensors (errors bars represent two standard deviations above and below mean). Temperature response measurements were made at 10 C intervals across a temperature range of approximately -10 to 40 C in a temperature controlled chamber under a fixed, broad spectrum, electric lamp. At each temperature set point, a spectroradiometer was used to measure light intensity from the lamp and all quantum sensors were compared to the spectroradiometer. The spectroradiometer was mounted external to the temperature control chamber and remained at room temperature during the experiment.
Cosine Response
Directional (cosine) response is defined as the measurement error at a specific angle of radiation incidence. Error for Apogee SQ-500 series quantum sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and 75°, respectively.
Mean directional (cosine) response of seven apogee SQ-500 series quantum sensors. Directional response measurements were made on the rooftop of the Apogee building in Logan, Utah. Directional response was calculated as the relative difference of SQ-500 quantum sensors from the mean of replicate reference quantum sensors (LI-COR models LI-190 and LI-190R, Kipp & Zonen model PQS 1). Data were also collected in the laboratory using a reference lamp and positioning the sensor at varying angles.
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided. To accurately measure PPFD incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 leveling plate is recommended for this purpose. To facilitate mounting on a cross arm, an Apogee Instruments model AL-120 mounting bracket is recommended.
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 0.5 %, but it is easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for the sensor when it is not in use.
CABLE CONNECTORS
Apogee started offering cable connectors on some bare-lead sensors in March
2018 to simplify the process of removing sensors from weather stations for
calibration (the entire cable does not have to be removed from the station
and shipped with the sensor).
The ruggedized M8 connectors are rated IP68, made of corrosion-resistant
marine-grade stainless-steel, and designed for extended use in harsh
environmental conditions.
Instructions
Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins
are used for every sensor. There may also be unused wire colors inside the
cable. To simplify datalogger connection, we remove the unused pigtail lead
colors at the datalogger end of the cable.
If a replacement cable is required, please contact Apogee directly to ensure
ordering the proper pigtail configuration.
Alignment: When reconnecting a sensor, arrows on the connector jacket and an
aligning notch ensure proper orientation.
Disconnection for extended periods: When disconnecting the sensor for an extended period of time from a station, protect the remaining half of the connector still on the station from water and dirt with electrical tape or other method.
Tightening: Connectors are designed to be firmly finger-tightened only. There is an o-ring inside the connector that can be overly compressed if a wrench is used. Pay attention to thread alignment to avoid crossthreading. When fully tightened, 1-2threads may still be visible.
WARNING: Do not tighten the connector by twisting the black cable or sensor head, only twist the metal connector (yellow arrows).
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller)
capable of measuring and displaying or recording a millivolt signal (an input
measurement range of approximately 0-2.5 V (SQ-512) or 0-5 V (SQ-515) is
required to cover the entire range of PPFD from the sun). In order to maximize
the measurement resolution and signal-to- noise ratio, the signal input range
of the measurement device should closely match the output range of the quantum
sensors. The amplification circuit requires a power supply of 5.5 to 24 V DC.
NOTE: To prevent sensor damage, DO NOT connect the sensor to a power source
greater than 24 VDC.
VERY IMPORTANT: Apogee changed all wiring colors of our bare-lead sensors
in March 2018 in conjunction with the release of inline cable connectors on
some sensors. To ensure proper connection to your data device, please note
your serial number or if your sensor has a stainless-steel connector 30 cm
from the sensor head then use the appropriate wiring configuration below.
Wiring for SQ-512 & SQ-515 Serial Numbers 1690 and above or with a cable
connector
Wiring for SQ-512 & SQ-515 Serial Numbers within Range 0-1689
Sensor Calibration
Apogee amplified full-spectrum quantum sensors, models SQ-212 and SQ-215, have
a standard PPFD calibration factor of exactly:
SQ-512: 1.6 µmol m-2 s-1per mV
SQ-512: 0.8 µmol m-2 s-1per mV
Multiply this calibration factor by the measured voltage to convert sensor
output to PPFD in units of pm& nr2
Calibration Factor (0.8 Imot m-2 s-1 per mV) Sensor Output Signal (mV) =
PPFD (pmol m-2 s-1)
0.8 2500 = 2000
Example of PPFD measurement with an Apogee quantum sensor. Full sunlight
yields a PPFD on a horizontal plane at the Earth’s surface of approximately
2000 pmol rry2 s-‘. This yields an output signal of 2500 mV for the 0-5 V
option or an output signal of 1250 mV for the 0-2.5 V option. The signal is
converted to PPFD by multiplying by the calibration factor.
SQ-512 Sensor Output: 1250 mV Conversion Factor: 1.6 Imo’ m-2 s-1 per mV
SQ-515 Sensor Output: 2500 mV Conversion Factor: 0.8 pmol m-2 s-l per mV
Spectral Error
The combination of diffuser transmittance, interference filter transmittance,
and photodetector sensitivity yields spectral response of a quantum sensor. A
perfect photodetector/filter/diffuser combination would exactly match the
defined plant photosynthetic response to photons (equal weighting to all
photons between 400 and 700 nm, no weighting of photons outside this range),
but this is challenging in practice. Mismatch between the defined plant
photosynthetic response and sensor spectral response results in spectral
error when the sensor is used to measure radiation from sources with a
different spectrum than the radiation source used to calibrate the sensor
(Federer and Tanner, 1966; Ross and Sulev, 2000).
Spectral errors for PPFD measurements made under common radiation sources for
growing plants were calculated for Apogee SQ-100 and SQ-500 series quantum
sensors using the method of Federer and Tanner (1966). This method requires
PPFD weighting factors (defined plant photosynthetic response), measured
sensor spectral response (shown in Spectral Response section on page 7), and
radiation source spectral outputs (measured with a spectroradiometer). Note,
this method calculates spectral error only and does not consider calibration,
directional (cosine), temperature, and stability/drift errors. Spectral error
data (listed in table below) indicate errors less than 5 % for sunlight in
different conditions (clear, cloudy, reflected from plant canopies,
transmitted below plant canopies) and common broad spectrum electric lamps
(cool white fluorescent, metal halide, high pressure sodium), but larger
errors for different mixtures of light emitting diodes (LEDs) for the SQ-100
series sensors.
Spectral errors for the SQ-500 series sensors are smaller than those for
SQ-100 series sensors because the spectral response of SQ-500 series sensors
is a closer match to the defined plant photosynthetic response.
Quantum sensors are the most common instrument for measuring PPFD, because
they are about an order of magnitude lower cost the spectroradiometers, but
spectral errors must be considered. The spectral errors in the table below
can be used as correction factors for individual radiation sources.
Spectral Errors for PPFD Measurements with Apogee SQ-100 and SQ-500 Series Quantum Sensors
Radiation Source (Error Calculated Relative to Sun, Clear Sky)| SQ-100
Series
PPFD Error [%]| SQ-500 Series
PPFD Error [%]
---|---|---
Sun (Clear Sky)| 0| 0
Sun (Cloudy Sky)| 0.2| 0.1
Reflected from Grass Canopy| 3.8| -0.3
Transmitted below Wheat Canopy| 4.5| 0.1
Cool White Fluorescent (T5)| 0| 0.1
Metal Halide| -2.8| 0.9
Ceramic Metal Halide| -16.1| 0.3
High Pressure Sodium| 0.2| 0.1
Blue LED (448 nm peak, 20 nm full-width half-maximum)| -10.5| -0.7
Green LED (524 nm peak, 30 nm full-width half-maximum)| 8.8| 3.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| 2.6| 0.8
Red LED (667 nm peak, 20 nm full-width half-maximum)| -62.1| 2.8
Red, Blue LED Mixture (80 % Red, 20 % Blue)| -72.8| -3.9
Red, Blue, White LED Mixture (60 % Red, 25 % White, 15 % Blue)| -35.5| -2
Cool White LED| -3.3| 0.5
Warm White LED| -8.9| 0.2
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657.
Ross, J., and M. Sulev, 2000. Sources of errors in measurements of PAR.
Agricultural and Forest Meteorology 100:103-125.
Yield Photon Flux Density (YPFD) Measurements
Photosynthesis in plants does not respond equally to all photons. Relative
quantum yield (plant photosynthetic efficiency) is dependent on wavelength
(green line in figure below) (McCree, 1972a; Inada, 1976). This is due to the
combination of spectral absorptivity of plant leaves (absorptivity is higher
for blue and red photons than green photons) and absorption by non-
photosynthetic pigments. As a result, photons in the wavelength range of
approximately 600-630 nm are the most efficient.
Defined plant response to photons (black line, weighting factors used to calculate PPFD), measured plant response to photons (green line, weighting factors used to calculate YPFD), and SQ-500 series quantum sensor response to photons (sensor spectral response).
One potential definition of PAR is weighting photon flux density in units of pmol m-2 s-¹ at each wavelength between 300 and 800 nm by measured relative quantum yield and summing the result. This is defined as yield photon flux density (YPFD, units of pmol m2 s-¹) (Sager et al., 1988). There are uncertainties and challenges associated with this definition of PAR. Measurements used to generate the relative quantum yield data were made on single leaves under low radiation levels and at short time scales (McCree, 1972a; Inada, 1976). Whole plants and plant canopies typically have multiple leaf layers and are generally grown in the field or greenhouse over the course of an entire growing season. Thus, actual conditions plants are subject to are likely different than those the single leaves were in when measurements were made by McCree (1972a) and Inada (1976). In addition, relative quantum yield shown in the figure above is the mean from twenty-two species grown in the field (McCree, 1972a). Mean relative quantum yield for the same species grown in growth chambers was similar, but there were differences, particularly at shorter wavelengths (less than 450 nm). There was also some variability between species (McCree, 1972a; Inada, 1976).
McCree (1972b) found that equally weighting all photons between 400 and 700 nm and summing the result, defined as photosynthetic photon flux density (PPFD, in units of pmol m-2 s-¹), was well correlated to photosynthesis, and very similar to correlation between YPFD and photosynthesis. As a matter of practicality, PPFD is a simpler definition of PAR. At the same time as McCree’s work, others had proposed PPFD as an accurate measure of PAR and built sensors that approximated the PPFD weighting factors (Biggs et al., 1971; Federer and Tanner, 1966). Correlation between PPFD and YPFD measurements for several radiation sources is very high (figure below), as an approxismation, YPFD = 0.9PPFD. As a result, almost universally PAR is defined as PPFD rather than YPFD, although YPFD has been used in some studies. The only radiation sources shown (figure below) that don’t fall on the regression line are the high pressure sodium (HPS) lamp, reflection from a plant canopy, and transmission below a plant canopy. A large fraction of radiation from HPS lamps is in the red range of wavelengths where the YPFD weighting factors (measured relative quantum yield) are at or near one. The factor for converting PPFD to YPFD for HPS lamps is 0.95, rather than 0.90. The factor for converting PPFD to YPFD for reflected and transmitted photons is 1.00.
Correlation between photosynthetic photon flux density (PPFD) and yield photon flux density (YPFD) for multiple different radiation sources. YPFD is approximately 90 % of PPFD. Measurements were made with a spectroradiometer (Apogee Instruments model PS-200) and weighting factors shown in the previous figure were used to calculate PPFD and YPFD.
Biggs, W., A.R. Edison, J.D. Eastin, K.W. Brown, J.W. Maranville, and M.D.
Clegg, 1971. Photosynthesis light sensor and meter. Ecology 52:125-131.
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657.
Inada, K., 1976. Action spectra for photosynthesis in higher plants. Plant and
Cell Physiology 17:355-365.
McCree, K.J., 1972a. The action spectrum, absorptance and quantum yield of
photosynthesis in crop plants.
Agricultural Meteorology 9:191-216.
McCree, K.J., 1972b. Test of current definitions of photosynthetically active
radiation against leaf photosynthesis data. Agricultural Meteorology
10:443-453.
Underwater Measurements and Immersion Effect
Sager, J.C., W.O. Smith, J.L. Edwards, and K.L. Cyr, 1988. Photosynthetic
efficiency and phytochrome photoequilibria determination using spectral data.
Transactions of the ASAE 31:1882-1889.
When a quantum sensor that was calibrated in air is used to make underwater
measurements, the sensor reads
Immersion Effect Correction Factor
When a radiation sensor is submerged in water, more of the incident radiation
is backscattered out of the diffuser than when the sensor is in air (Smith,
1969; Tyler and Smith, 1970). This phenomenon is caused by the difference in
the refractive index for air (1.00) and water (1.33), and is called the
immersion effect. Without correction for the immersion effect, radiation
sensors calibrated in air can only provide relative values underwater (Smith,
1969; Tyler and Smith, 1970). Immersion effect correction factors can be
derived by making measurements in air and at multiple water depths at a
constant distance from a lamp in a controlled laboratory setting.
Apogee SQ-500 series quantum sensors have an immersion effect correction
factor of 1.25 (SQ-512 serial # 3656 and above; SQ-515 serial # 3857 and
above) or 1.32 (SQ-512 serial # 0-3655; SQ-515 serial # 0-3856). This
correction factor should be multiplied by PPFD measurements made underwater to
yield accurate PPFD.
Further information on underwater measurements and the immersion effect can be
found on the Apogee webpage (http://www.apogeeinstruments.com/underwater-par-
measurements/).
Smith, R.C., 1969. An underwater spectral irradiance collector. Journal of
Marine Research 27:341-351.
Tyler, J.E., and R.C. Smith, 1970. Measurements of Spectral Irradiance
Underwater. Gordon and Breach, New York, New York. 103 pages
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward- looking sensor can block the optical path in three common ways:
- Moisture or debris on the diffuser.
- Dust during periods of low rainfall.
- Salt deposit accumulation from evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments upward-looking sensors have a domed diffuser and housing
for improved self-cleaning from rainfall, but active cleaning may be
necessary. Dust or organic deposits are best removed using water, or window
cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved
with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be
removed with solvents such as alcohol or acetone. Use only gentle pressure
when cleaning the diffuser with a cotton swab or soft cloth to avoid
scratching the outer surface. The solvent should be allowed to do the
cleaning, not mechanical force. Never use abrasive material or cleaner on the
diffuser.
Although Apogee sensors are very stable, nominal accuracy drift is normal for
all research-grade sensors. To ensure maximum accuracy, we generally recommend
sensors are sent in for recalibration every two years, although you can often
wait longer according to your particular tolerances.
To determine if a specific sensor needs recalibration, the Clear Sky
Calculator (www.clearskycalculator.com)
website and/or smartphone app can be used to indicate PPFD incident on a
horizontal surface at any time of day at any location in the world. It is most
accurate when used near solar noon in spring and summer months, where accuracy
over multiple clear and unpolluted days is estimated to be ± 4 % in all
climates and locations around the world. For best accuracy, the sky must be
completely clear, as reflected radiation from clouds causes incoming
radiation to increase above the value predicted by the clear sky calculator.
Measured PPFD can exceed PPFD predicted by the Clear Sky Calculator due to
reflection from thin, high clouds and edges of clouds, which enhances incident
PPFD. The influence of high clouds typically shows up as spikes above clear
sky values, not a constant offset greater than clear sky values.
To determine recalibration need, input site conditions into the calculator and
compare PPFD measurements to calculated PPFD for a clear sky. If sensor PPFD
measurements over multiple days near solar noon are consistently different
than calculated PPFD (by more than 6 %), the sensor should be cleaned and re-
leveled. If measurements are still different after a second test, email
calibration@apogeeinstruments.com
to discuss test results and possible return of sensor(s).
Homepage of the Clear Sky Calculator. Two calculators are available: one for quantum sensors (PPFD) and one for pyranometers (total shortwave radiation).
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in middle of page and an estimate of PPFD is returned on right-hand side of page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee SQ-614 ePAR Sensors provide a 4-20 mA output that is proportional to
extended photosynthetically active radiation for the 380-760 nm wavelength
range. A quick and easy check of sensor functionality can be determined using
a DC power supply and an ammeter. Power the sensor with a DC voltage by
connecting the positive voltage signal to the red wire from the sensor and
the negative (or common) to the black wire from the sensor. Use the ammeter to
measure across the white wire (signal output) and green wire (signal ground).
Direct the sensor head toward a light source and verify the sensor provides a
signal. Increase and decrease the distance from the sensor head to the light
source to verify that the signal changes proportionally (decreasing signal
with increasing distance and increasing signal with decreasing distance).
Blocking all radiation from the sensor should force the sensor signal to 4
mA.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SQ-614 ePAR Sensors are calibrated with a standard calibration factor
of 250 pimol m-251 per mA, yielding a sensitivity of 0.004 mA per pmol m-2
s-1. Thus, a compatible measurement device (e.g., datalogger or controller)
should have resolution of at least 0.004 mA in order to provide photon flux
density resolution of 1 limo! m-25-1.
The 4-20 mA circuit design allows the output to drive a resistive load (RL) to
within 2 volts of the supply voltage to the sensor (VS), at 20 mA (0.02 A).
The equation to calculate resistive load is RL = [VS — 2 V] / 0.02 A. For
example, a sensor with a supply voltage of 12 V DC can drive a maximum load of
500 0 (RL = [12 V — 2 V] / 0.02 A = 500 0). The output voltage from the sensor
is calculated by adding the wire resistance to the input resistance of the
data collection system, and then multiplying by 0.02 A.
An example datalogger program for Campbell Scientific dataloggers can be found
on the Apogee webpage at
https://www.apogeeinstruments.com/downloads/#datalogger.
Cable Length
Shortening or splicing on additional cable in the field is generally not a
problem for the current output of the SQ614. However, adding cable will
result in a greater resistive load, which should be taken into consideration
when determining the maximum resistive load that the sensor will drive (see
section above on Compatible Measurement Devices). All Apogee sensors use
shielded, twisted pair cable to minimize electromagnetic interference. For
best measurements, the shield wire must be connected to an earth ground. This
is particularly important when using the sensor with long lead lengths in
electromagnetically noisy environments.
Modifying Cable Length
See Apogee webpage for details on how to extend sensor cable length:
http://www.nogeeinstruments.com/how-to-make-a-weatherproof-cable-solice/).
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as
the product is in new condition (to be determined by Apogee). Returns are
subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from
defects in materials and craftsmanship for a period of four (4) years from the
date of shipment from our factory. To be considered for warranty coverage an
item must be evaluated by Apogee.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content
meters, EE08-SS probes) are covered for a period of one (1) year.
What is Not Covered
The customer is responsible for all costs associated with the removal,
reinstallation, and shipping of suspected warranty items to our factory.
The warranty does not cover equipment that has been damaged due to the
following conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of proper maintenance and is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
-
Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking of the
service item. Call 435-245-8012 or email techsupport@apogeeinstruments.com with questions. -
For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee’s standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
-
Please write the RMA number on the outside of the shipping container.
-
Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders.
Apogee Instruments, Inc.
721 West 1800 North Logan, UT
84321, USA -
Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under arranty, you will be informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com
to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss.
This limited warranty and any disputes arising out of or in connection with
this limited warranty (“Disputes”) shall be governed by the laws of the State
of Utah, USA, excluding conflicts of law principles and excluding the
Convention for the International Sale of Goods. The courts located in the
State of Utah, USA, shall have exclusive jurisdiction over any Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot by transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision
shall be deemed severable and shall not affect any remaining provisions. In
case of any inconsistency between the English and other versions of this
limited warranty, the English version shall prevail.
This warranty cannot be changed, assumed, or amended by any other person or
agreement
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
References
- Software Downloads - Datalogger Programs | Apogee Instruments
- How to Make a Weatherproof Cable Splice
- Support | Apogee Instruments
- Recalibration and Repair | Apogee Instruments
- Underwater PAR Measurements | Apogee Instruments
- Clear Sky Calculator | Apogee Instruments Inc.
- Clear Sky Calculator | Apogee Instruments Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>