APOGEE SQ-100X Sun Calibration Quantum Sensor Owner’s Manual
- June 5, 2024
- APOGEE
Table of Contents
- SQ-100X Sun Calibration Quantum Sensor
- CERTIFICATE OF COMPLIANCE
- INTRODUCTION
- SENSOR MODELS
- SPECIFICATIONS
- Calibration Traceability
- Spectral Response
- Cosine Response
- DEPLOYMENT AND INSTALLATION
- CABLE CONNECTORS
- Instructions
- OPERATION AND MEASUREMENT
- Spectral Errors
- Underwater Measurements and Immersion Effect
- MAINTENANCE AND RECALIBRATION
- TROUBLESHOOTING AND CUSTOMER SUPPORT
- RETURN AND WARRANTY POLICY
- WARRANTY POLICY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
SQ-100X Sun Calibration Quantum Sensor
OWNER’S MANUAL
QUANTUM SENSOR
Models SQ-100X Series
Rev: 1-May-2021
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA TEL: 435-792-4700 | FAX: 435-787-8268 | WEB: APOGEEINSTRUMENTS.COM Copyright © 2021 Apogee Instruments, Inc.
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer:
Apogee Instruments, Inc. 721 W 1800 N Logan, Utah 84321 USA
for the following product(s):
Models: SQ-100X Type: Quantum Sensor
The object of the declaration described above is in conformity with the relevant Union harmonization legislation:
2014/30/EU 2011/65/EU 2015/863/EU
Electromagnetic Compatibility (EMC) Directive Restriction of Hazardous Substances (RoHS 2) Directive Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory
use EMC requirements EN 50581:2012 Technical documentation for the
assessment of electrical and electronic products with respect to
the restriction of hazardous substances
Please be advised that based on the information available to us from our raw material suppliers, the products manufactured by us do not contain, as intentional additives, any of the restricted materials including lead (see note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls (PBB), polybrominated diphenyls (PBDE), bis(2-ethylhexyl) phthalate (DEHP), butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl phthalate (DIBP). However, please note that articles containing greater than 0.1% lead concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on our raw materials or end products for the presence of these substances, but rely on the information provided to us by our material suppliers.
Signed for and on behalf of: Apogee Instruments, May 2021
Bruce Bugbee President Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is often expressed as photosynthetic photon flux density
(PPFD): photon flux in units of micromoles per square meter per second (µmol
m-2 s-1, equal to microEinsteins per square meter per second) summed from 400
to 700 nm (total number of photons from 400 to 700 nm). While Einsteins and
micromoles are equal (one Einstein = one mole of photons), the Einstein is not
an SI unit, so expressing PPFD as µmol m-2 s-1 is preferred.
The acronym PPF is also widely used and refers to the photosynthetic photon
flux. The acronyms PPF and PPFD refer to the same variable. The two terms have
co-evolved because there is not a universal definition of the term “flux”.
Some physicists define flux as per unit area per unit time. Others define flux
only as per unit time. We have used PPFD in this manual because we feel that
it is better to be more complete and possibly redundant.
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation.
Typical applications of quantum sensors include incoming PPFD measurement over
plant canopies in outdoor environments or in greenhouses and growth chambers,
and reflected or under-canopy (transmitted) PPFD measurement in the same
environments.
Apogee Instruments SQ-100X series quantum sensors consist of a cast acrylic
diffuser (filter), interference filter, photodiode, and signal processing
circuitry mounted in an anodized aluminum housing, and a cable to connect the
sensor to a measurement device. Sensors are potted solid with no internal air
space, and are designed for continuous PPFD measurement in indoor or outdoor
environments. SQ-100X series sensors output an analog voltage that is directly
proportional to PPFD. The voltage signal from the sensor is directly
proportional to radiation incident on a planar surface (does not have to be
horizontal), where the radiation emanates from all angles of a hemisphere.
SENSOR MODELS
This manual covers the unamplified model SQ-100X. Additional models are covered in their respective manuals.
Model | Signal |
---|---|
SQ-100X | Self-powered |
SQ-202X | 0-2.5 V |
SQ-204X | 4-20 mA |
SQ-205X | 0-5 V |
SQ-420X | USB |
SQ-421X | SDI-12 |
SQ-422X | Modbus |
Sensor model number and serial number are located near the pigtail leads on the sensor cable. If you need the manufacturing date of your sensor, please contact Apogee Instruments with the serial number of your sensor.
SPECIFICATIONS
Power Supply | Self-powered |
---|---|
Sensitivity |
0.1 mV per μmol m-2 s-1
Calibration Factor (Reciprocal of Sensitivity)|
10 μmol m-2 s-1 per mV
Calibration Uncertainty|
± 5 % (see Calibration Traceability below)
Calibrated Output Range| 0 to 400 mV
Measurement Repeatability| Less than 0.5 %
Long-term Drift (Non-stability)|
Less than 2 % per year
Non-linearity|
Less than 1 % (up to 4000 μmol m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range|
370 to 650 nm (wavelengths where response is greater than 50 % of maximum; see Spectral Response below)
Directional (Cosine) Response|
± 5 % at 75° zenith angle (see Cosine Response below)
Temperature Response| – 0.04 % per C
Operating Environment|
10 to 60 C; 0 to 100 % relative humidity; can be submerged in water up to depths of 30 m
Dimensions|
24 mm diameter, 33 mm height
Mass|
90 g (with 5m of lead wire)
Cable|
5 m of two conductor, shielded, twisted-pair wire; TPR jacket (high water resistance, high UV stability, flexibility in cold conditions); pigtail lead wires; stainless steel (316), M8 connector located 25 cm from sensor head
Calibration Traceability
Apogee SQ series quantum sensors are calibrated through side-by-side comparison to the mean of transfer standard quantum sensors under a reference lamp. The reference quantum sensors are recalibrated with a 200 W quartz halogen lamp traceable to the National Institute of Standards and Technology (NIST).
Spectral Response
Mean spectral response of four SQ-100x series quantum sensors compared to PPFD weighting function. Spectral response measurements were made at 10 nm increments across a wavelength range of 350 to 800 nm in a monochromator with an attached electric light source. Measured spectral data from each quantum sensor were normalized by the measured spectral response of the monochromator/electric light combination, which was measured with a spectroradiometer.
Cosine Response
Directional, or cosine, response is defined as the measurement error at a specific angle of radiation incidence. Error for Apogee SQ100X series quantum sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and 75°, respectively.
Mean cosine response of five SQ100X series quantum sensors. Cosine response measurements were made by direct side-by-side comparison to the mean of seven reference SQ-500 quantum sensors.
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided to prevent galvanic corrosion. To accurately measure PPFD incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 Leveling Plate is recommended to level the sensor when used on a flat surface or being mounted to surfaces such as wood. To facilitate mounting on a mast or pipe, the Apogee Instruments model AL-120 Solar Mounting Bracket with Leveling Plate is recommended.
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 1 %, but it is easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for the sensor when it is not in use.
CABLE CONNECTORS
Apogee sensors offer cable connectors to simplify the process of removing sensors from weather stations for calibration (the entire cable does not have to be removed from the station and shipped with the sensor).
The ruggedized M8 connectors are rated IP68, made of corrosion-resistant marine-grade stainless-steel, and designed for extended use in harsh environmental conditions.
Inline cable connectors are installed 25 cm from the head
Instructions
Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins are used for every sensor. There may also be unused wire colors inside the cable. To simplify datalogger connection, we remove the unused pigtail lead colors at the datalogger end of the cable.
If a replacement cable is required, please contact Apogee directly to ensure ordering the proper pigtail configuration.
Alignment: When reconnecting a sensor, arrows on the connector jacket and an aligning notch ensure proper orientation.
A reference notch inside the connector ensures proper alignment before tightening.
Disconnection for extended periods: When disconnecting the sensor for an extended period of time from a station, protect the remaining half of the connector still on the station from water and dirt with electrical tape or other method.
When sending sensors in for calibration, only send the short end of the cable and half the connector.
Tightening: Connectors are designed to be firmly finger-tightened only. There is an o-ring inside the connector that can be overly compressed if a wrench is used. Pay attention to thread alignment to avoid cross-threading. When fully tightened, 1-2 threads may still be visible.
Finger-tighten firmly
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying or recording a millivolt signal (an input measurement range of approximately 0-250 mV is required to cover the entire range of PPFD from the sun). In order to maximize measurement resolution and signal-to-noise ratio, the input range of the measurement device should closely match the output range of the quantum sensor. DO NOT connect the sensor to a power source. The sensor is self-powered and applying voltage will damage the sensor.
Wiring
White: Positive (signal from sensor)
Black: Negative (signal from sensor)
Clear: Shield/Ground
Sensor Calibration
All Apogee unamplified quantum sensor model (SQ-100X) have a standard PPFD calibration factor of exactly:
10.0 µmol m-2 s-1 per mV
Multiply this calibration factor by the measured mV signal to convert sensor output to PPFD in units of µmol m-2 s-1:
Calibration Factor (10.0 µmol m-2 s-1 per mV) * Sensor Output Signal (mV) = PPFD (µmol m-2 s-1)
10.0 * 200 = 2000
Full Sunlight (2000 µmol m-2 s-1)
Sensor Output 200 mV
Example of PPFD measurement with an Apogee quantum sensor. Full sunlight yields a PPFD on a horizontal plane at the Earth’s surface of approximately 2000 µmol m-2 s-1. This yields an output signal of 200 mV. The signal is converted to PPFD by multiplying by the calibration factor of 10.00 µmol m-2 s-1 per mV.
Spectral Errors
Apogee SQ-100X series sensors can measure PPFD for sunlight and electric light
with a single calibration factor. However, errors occur in various light
sources due to changes in spectral output. If the light source spectrum is
known then errors can be estimated and used to adjust the measurements. The
weighting function for PPFD is shown in the graph below, along with the
spectral response of Apogee SQ-100X series quantum sensors. The closer the
spectral response matches the defined PPFD spectral weighting functions, the
smaller spectral errors will be. The table below provides spectral error
estimates for PPFD measurements from light sources different than the
calibration source. The method of Federer and Tanner (1966) was used to
determine spectral errors based on the PPFD spectral weighting functions,
measured sensor spectral response, and radiation source spectral outputs
(measured with a spectroradiometer). This method calculates spectral error and
does not consider calibration, cosine, and temperature errors.
Federer, C. A., and C. B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654657.
McCree, K. J., 1972. The action spectrum, absorptance and quantum yield of
photosynthesis in crop plants. Agricultural Meteorology 9:191-216.
Spectral Errors for PPFD Measurements with Apogee SQ-100X Series Quantum Sensors
Quantum sensors can be a very practical means of measuring PPFD from multiple radiation sources, but spectral errors must be considered. The spectral errors in the table above can be used as correction factors for individual radiation sources.
Underwater Measurements and Immersion Effect
When a quantum sensor that was calibrated in air is used to make underwater
measurements, the sensor reads low. This phenomenon is called the immersion
effect and happens because the refractive index of water (1.33) is greater
than air (1.00). The higher refractive index of water causes more light to be
backscattered (or reflected) out of the sensor in water than in air
(Smith,1969; Tyler and Smith,1970). As more light is reflected, less light is
transmitted through the diffuser to the detector, which causes the sensor to
read low. Without correcting for this effect, underwater measurements are only
relative, which makes it difficult to compare light in different environments.
The SQ-100X series sensors have an immersion effect correction factor of 1.15.
This correction factor should be multiplied to measurements made underwater.
Further information on underwater measurements and the immersion effect can be
found at http://www.apogeeinstruments.com/underwater-par-measurements/.
MAINTENANCE AND RECALIBRATION
Moisture or debris on the diffuser is a common cause of low readings. The
sensor has a domed diffuser and housing for improved self-cleaning from
rainfall, but materials can accumulate on the diffuser (e.g., dust during
periods of low rainfall, salt deposits from evaporation of sea spray or
sprinkler irrigation water) and partially block the optical path. Dust or
organic deposits are best removed using water or window cleaner and a soft
cloth or cotton swab. Salt deposits should be dissolved with vinegar and
removed with a soft cloth or cotton swab. Never use an abrasive material or
cleaner on the diffuser.
Although Apogee sensors are very stable, nominal accuracy drift is normal for
all research-grade sensors. To ensure maximum accuracy, we generally recommend
sensors are sent in for recalibration every two years, although you can often
wait longer according to your particular tolerances.
To determine if your sensor needs recalibration, the Clear Sky Calculator
(www.clearskycalculator.com) website and/or smartphone app can be used to
indicate the total shortwave radiation incident on a horizontal surface at any
time of day at any location in the world. It is most accurate when used near
solar noon in spring and summer months, where accuracy over multiple clear and
unpolluted days is estimated to be ± 4 % in all climates and locations around
the world. For best accuracy, the sky must be completely clear, as reflected
radiation from clouds causes incoming radiation to increase above the value
predicted by the clear sky calculator. Measured values of total shortwave
radiation can exceed values predicted by the Clear Sky Calculator due to
reflection from thin, high clouds and edges of clouds, which enhances incoming
shortwave radiation. The influence of high clouds typically shows up as spikes
above clear sky values, not a constant offset greater than clear sky values.
To determine recalibration need, input site conditions into the calculator and
compare total shortwave radiation measurements to calculated values for a
clear sky. If sensor shortwave radiation measurements over multiple days near
solar noon are consistently different than calculated values (by more than 6
%), the sensor should be cleaned and re-leveled. If measurements are still
different after a second test, email calibration@apogeeinstruments.com to
discuss test results and possible return of sensor(s).
Homepage of the Clear Sky Calculator. Two calculators are available: one for
quantum sensors (PPFD) and one for pyranometers (total shortwave radiation).
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in
middle of page and an estimate of PPFD is returned on right-hand side of page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee SQ-100X series quantum sensors are self-powered devices and output a
voltage signal proportional to incident PPFD. A quick and easy check of sensor
functionality can be determined using a voltmeter with millivolt resolution.
Connect the positive lead wire from the voltmeter to the white wire from the
sensor and the negative (or common) lead wire from the voltmeter to the black
wire from the sensor. Direct the sensor head toward a light source and verify
the sensor provides a signal. Increase and decrease the distance from the
sensor head to the light source to verify that the signal changes
proportionally (decreasing signal with increasing distance and increasing
signal with decreasing distance). Blocking all radiation from the sensor
should force the sensor signal to zero.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SQ-100X series quantum sensors are calibrated with a standard calibration
factor of 10.0 µmol m-2 s-1 per mV, yielding a sensitivity of 0.1 mV per µmol
m-2 s-1. Thus, a compatible measurement device (e.g., datalogger or
controller) should have resolution of at least 0.1 mV in order to provide PPFD
resolution of 1 µmol m-2 s-1.
An example datalogger program for Campbell Scientific dataloggers can be found
on the Apogee webpage at http://www.apogeeinstruments.com/content/Quantum-
Sensor-Unamplified.CR1.
Cable Length
When the sensor is connected to a measurement device with high input
impedance, sensor output signals are not changed by shortening the cable or
splicing on additional cable in the field. Tests have shown that if the input
impedance of the measurements device is greater than 1 mega-ohm there is
negligible effect on the calibration, even after adding up to 100 m of cable.
All Apogee sensors use shielded, twisted pair cable to minimize
electromagnetic interference. For best measurements, the shield wire must be
connected to an earth ground. This is particularly important when using the
sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See Apogee webpage for details on how to extend sensor cable length:
(http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/).
Unit Conversion Charts
Apogee SQ-100X series quantum sensors are calibrated to measure PPFD in units
of µmol m-2 s-1. Units other than photon flux density (e.g., energy flux
density, illuminance) may be required for certain applications. It is possible
to convert the PPFD value from a quantum sensor to other units, but it
requires spectral output of the radiation source of interest. Conversion
factors for common radiation sources can be found on the Unit Conversions page
in the Support Center on the Apogee website (http://www.apogeeinstruments.com
/unit-conversions/; scroll down to Quantum Sensors section). A spreadsheet to
convert PPFD to energy flux density or illuminance is also provided on the
Unit Conversions page in the Support Center on the Apogee website
(http://www.apogeeinstruments.com/content/PPFD-to-Illuminance-Calculator.xls).
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as the product is in new condition (to be determined by Apogee). Returns are subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered All products manufactured by Apogee Instruments are warranted
to be free from defects in materials and craftsmanship for a period of four
(4) years from the date of shipment from our factory. To be considered for
warranty coverage an item must be evaluated by Apogee.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content
meters, EE08-SS probes) are covered for a period of one (1) year.
What is Not Covered The customer is responsible for all costs associated with
the removal, reinstallation, and shipping of suspected warranty items to our
factory.
The warranty does not cover equipment that has been damaged due to the
following conditions:
1. Improper installation or abuse.
2. Operation of the instrument outside of its specified operating range.
3. Natural occurrences such as lightning, fire, etc.
4. Unauthorized modification.
5. Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine
recalibration of sensors/meters is considered part of proper maintenance and
is not covered under warranty.
Who is Covered This warranty covers the original purchaser of the product or
other party who may own it during the warranty period.
What Apogee Will Do At no charge Apogee will:
1. Either repair or replace (at our discretion) the item under warranty.
2. Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense. How
To Return An Item 1. Please do not send any products back to Apogee
Instruments until you have received a Return Merchandise
Authorization (RMA) number from our technical support department by submitting
an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-
repairs/. We will use your RMA number for tracking of the service item. Call
435-245-8012 or email
techsupport@apogeeinstruments.com with questions.
2. For warranty evaluations, send all RMA sensors and meters back in the
following condition: Clean the sensor’s exterior and cord. Do not modify the
sensors or wires, including splicing, cutting wire leads, etc. If a connector
has been attached to the cable end, please include the mating connector
otherwise the sensor connector will be removed in order to complete the
repair/recalibration. Note: When sending back sensors for routine calibration
that have Apogee’s standard stainless-steel connectors, you only need to send
the sensor with the 30 cm section of cable and one-half of the connector. We
have mating connectors at our factory that can be used for calibrating the
sensor.
3. Please write the RMA number on the outside of the shipping container.
4. Return the item with freight pre-paid and fully insured to our factory
address shown below. We are not responsible for any costs associated with the
transportation of products across international borders.
Apogee Instruments, Inc. 721 West 1800 North Logan, UT 84321, USA
5. Upon receipt, Apogee Instruments will determine the cause of failure. If
the product is found to be defective in terms of operation to the published
specifications due to a failure of product materials or craftsmanship, Apogee
Instruments will repair or replace the items free of charge. If it is
determined that your product is not covered under warranty, you will be
informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss.
This limited warranty and any disputes arising out of or in connection with
this limited warranty (“Disputes”) shall be governed by the laws of the State
of Utah, USA, excluding conflicts of law principles and excluding the
Convention for the International Sale of Goods. The courts located in the
State of Utah, USA, shall have exclusive jurisdiction over any Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot by transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision shall
be deemed severable and shall not affect any remaining provisions. In case of
any inconsistency between the English and other versions of this limited
warranty, the English version shall prevail.
This warranty cannot be changed, assumed, or amended by any other person or
agreement.
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA TEL:
435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
References
- apogeeinstruments.com/content/Quantum-Sensor-Unamplified.CR1
- How to Make a Weatherproof Cable Splice
- Recalibration and Repair | Apogee Instruments
- Underwater PAR Measurements | Apogee Instruments
- Unit Conversions | Apogee Instruments
- Clear Sky Calculator | Apogee Instruments Inc.
- Clear Sky Calculator | Apogee Instruments Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>