APOGEE ST-100 Thermistor Temperature Sensor Owner’s Manual
- June 5, 2024
- APOGEE
Table of Contents
INSTRUMENTS
OWNER’S MANUAL
TEMPERATURE SENSORS
Models ST-100, ST-110, ST- 150, ST-200, and ST-300
Rev: 14-Apr-2021 APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN,
UTAH 84321, USA
TEL:435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer:
Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: ST-100, ST-110, ST-150, ST-200, ST-300
Type: Temperature Sensor
The object of the declaration described above is in conformity with the
relevant Union harmonization legislation:
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
2015/863/EU Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control, and laboratory
use – EMC requirements
EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous substances
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated biphenyls (PBDE), bis(2-Ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl
phthalate (DIBP). However, please note that articles containing greater than
0.1% lead concentration are RoHS 3 compliant using exemption 6c. Further note
that Apogee Instruments does not specifically run any analysis on our raw
materials or end products for the presence of these substances, but rely on
the information provided to us by our material suppliers.
Signed for and on behalf of:
Apogee Instruments, April 2021
Bruce Bugbee
President
Apogee Instruments, Inc.
INTRODUCTION
Temperature is generally thought of as the relative degree of ‘hotness’ or
‘coldness’ of a specific object or material. In actuality, the temperature is
a measure of the average thermal energy (internal kinetic energy) of an
object. Thermal energy is associated with the motion (kinetic energy) of the
atoms and molecules making up the object/material. Higher temperatures
correspond to higher thermal energies (faster motion of atoms and molecules),
whereas colder temperatures correspond to lower thermal energies (slower
motion of atoms and molecules).
Properties of materials, and nearly all biological, chemical, and physical
processes, are temperature-dependent. Temperature is also a fundamental
weather variable. As a result, temperature is perhaps the most widely measured
environmental variable.
Thermometers are sensors that measure temperature. Thermometers are often
electronic, with multiple options available. The main advantages of
thermistors over other electronic thermometers (thermocouples, platinum
resistance thermometers) are high signal-to-noise ratio, the requirement of
only a single-ended channel for measurement, and low cost, while still
maintaining comparable accuracy.
Apogee Instruments ST-100 temperature sensors consist of a precision
thermistor enclosed in a waterproof rubber covering, a precision bridge
resistor, and lead wires to connect the sensor to a measurement device. ST-100
sensors are weatherproof and are designed for continuous temperature
measurement in air, soil, or water. ST-110 Temperature sensors consist of a
precision thermistor, a precision bridge resistor, and lead wires to connect
the sensor to a measurement device.
ST-110 sensors are weatherproof, have excellent long-term stability, but do
not offer the same level of ingress protection as the ST-100. The ST-110 is
designed for continuous air temperature measurement when housed in the TS-100
aspirated radiation shield. Thermal conductivity to the precision thermistor
is minimized by using constantan wire, which has twenty-fold lower thermal
conductivity than copper wire.
ST-150 is a platinum resistive thermometer (PRT) enclosed in a stainless steel
sheath, and lead wires to connect the sensor to a measurement device. The
sensor is weatherproof and is designed for continuous air temperature
measurement when housed in the TS-100 aspirated radiation shield. The ST-150
is designed to be used with dataloggers that have resistors built-in. ST-200
temperature sensors have a fine-wire precision thermistor, precision bridge
resistor, and lead wires, but the thermistor is not packaged in a weatherproof
housing. ST-200 sensors are weather resistant and are designed for temperature
measurement of delicate surfaces (e.g., leaves, fruits) and small samples
(where a narrow field of view infrared radiometer may not be suitable due to
the integration over the conical field of view), or applications where rapid
response is required. ST-300 is a platinum resistance thermometer (PRT),
consisting of a platinum resistive element enclosed in a stainless steel
sheath, and lead wires to connect the sensor to a measurement device. ST-300
sensor are weatherproof and is designed for continuous air temperature
measurement when housed in the TS-100 aspirated radiation shield.
All Apogee thermistor temperature sensors output an analog voltage (when
supplied with an input voltage) that is related to thermistor resistance.
Resistance is directly related to temperature.
SENSOR MODELS
| The ST-100 temperature sensor includes a rugged, weatherproof housing and
can be exposed to a wide range of environmental conditions, including
submersion in water and burial in soil/porous media.
---|---
The ST-200 temperature sensor does not include the weatherproof housing and is
designed for fast response measurements or measurements of small samples and
fragile surfaces.
The ST-110 thermistor temperature sensor is designed for use inside the Apogee
fan aspirated
radiation shield, model TS-100.
The ST-150 PRT temperature sensor is an accurate and durable sensor for use in
all
applications, including air temperature measurements in the TS-100.
The ST-300 PRT temperature sensor is a highly accurate (1/10 DIN) and a
durable sensor for
use in all applications, including for air temperature measurements in the
TS-100.
| Sensor model number and serial number are located on a label near the
pigtail lead wires. If you need the manufacturing date of your sensor, please
contact Apogee Instruments with the serial number of your sensor.
---|---
SPECIFICATIONS
| ST-100| ST-110| ST-200| ST-150| ST-300
---|---|---|---|---|---
Measurement Range| -60 to 80 C
Measurement Uncertainty| 0.1 C (from 0 to
70 C), 0.2 C (from -25 to 0 C), 0.4 C (from -50 to -25 C)| 0.1 C (from 0 to
70 C), 0.15 C
(from -50 to 0 C)| 0.2 C (from 0 to
70 C), 0.4 C (from
-50 to 0 C)| 0.3 C (from -50
to 70 C), Class A
(see graph on
next page)| 0.1 C (from -60 to
60 C), 1/10 DIN
(see graph on next
page)
Measurement
Repeatability| less than 0.05 C| less than 0.01 C| less than 0.05 C| Less than
0.01 C| less than 0.01 C
---|---|---|---|---|---
Long-term Drift(Non-stability)| less than 0.02 C per year| less than 0.05 C
per year
(when used in non-condensing environments where the annual average temperature
is less than 30 C; continuously high temperatures or continuously humid
environments increase drift rate)
Time Constant ( time for detector signal to reach 63 % following a step
change)| 32 s| 7 s| 2 s| 32 s
---|---|---|---|---
Self-Heating| Less than 0.01 C (typical, assuming pulsed excitation of 2.5 V
DC)| Less than 0.003 C (typical, assuming
pulsed excitation of 2.1 V DC)
Operating
Environment| -60 to 80 C; 0 to 100 % relative humidity
---|---
Input VoltageRequirement| 2.5 V DC excitation (recommended, see OPERATION AND
MEASUREMENT section)| Datalogger
dependent| 2.1 V DC
Output Voltage Range| 0 to 2.5 V DC (assuming input excitation of 2.5 V DC)|
Datalogger
dependent| 16 to 27 mV DC (assuming input
excitation of 2.1 V DC)
---|---|---|---
Current Draw| 0.1 mA DC at 70 C (maximum, assuming continuous input
excitation of 2.5 V DC)| Datalogger
dependent| 0.21 mA DC (maximum, assuming
continuous input excitation of 2.1 V
DC)
Dimension| 100 mm length,
6 mm diameter| 80 mm length,
4 mm diameter| in 25 mm length,
1 mm diameter| 57.in 15 mm length,
3.18 mm diameter
---|---|---|---|---
Mass| 60 g| 95 g
Cable| 5 m of two-conductor, shielded, twisted-pair wire;
additional cable available in multiples of 5 m; TPR jacket (high water
resistance, high UV stability, flexibility in cold conditions); pigtail lead
wires| 5 m of four-conductor, shielded, twisted-pair wire; additional cable
available in multiples of 5 m; TPR
jacket (high water resistance, high UVtability, flexibility in cold
conditions); pigtail lead wires
---|---|---
Calibration Traceability
Apogee ST-100, ST-110, and ST-200 temperature sensors are not factory
calibrated but come with a generic calibration (see Steinhart-Hart
coefficients in OPERATION AND MEASUREMENT section). Custom calibration can be
derived by comparing the temperature from the thermistor to a reference
temperature measurement. Often, a simple offset can be used to make the
measured temperature match the reference temperature. Apogee ST-110, ST-150,
and ST-300 temperature sensors are factory verified to ensure accuracy.
Sensors are compared for absolute temperature against the mean of two
reference PRTs in a constant temperature bath, over a range of approximately
-35 to 60 C. The reference PRT calibrations are directly traceable to NIST.
Accuracy for multiple classifications of PRTs across a temperature range of
-60 to 60 C. Temperature tolerance specifications (y-axis) are defined by the
International electrotechnical Commission (2017). International
Electrotechnical Commission. 2017. IEC 60751:2008. Industrial platinum
resistance thermometers and platinum temperature sensors. International
Electrotechnical Commission.
https://webstore.iec.ch/publication/3400
(verified 10 Oct. 2017). [2017 is year accessed].
DEPLOYMENT AND INSTALLATION
Apogee ST Series temperature sensors are designed to be mounted inside solar radiation shields, such as the model TS-100 fan aspirated radiation shield (see picture below). ST-100, ST-150, and ST-300 sensors can also be buried in soil/porous media, or submerged in water. ST-200 sensors have thin, flexible wires near the fine-wire thermistor and are less sensitive to breakage than similarly-sized fine-wire thermocouples. The temperature measurement returned by a temperature sensor is the temperature for the sensor itself and not that of the environment the sensor is in, unless the sensor is in thermal equilibrium with the environment. In order to get representative temperature measurements, ST series sensors must be in thermal contact with the medium of interest. Accurate air temperature measurement requires a radiation shield (see picture below) to minimize the effects of shortwave radiation absorption (causes warming; occurs during the day) and longwave radiation emission (causes cooling; occurs on clear nights) by the sensor. Proper ventilation is also required to ensure coupling and thermal equilibrium with air. Condensation on air temperature sensors can pose a problem because it is a source of latent heat that can warm the sensor. When the condensed water evaporates, it cools the sensor via the removal of latent heat (evaporation cooling). During the installation of ST-100, ST-150, and ST-300 sensors in soil, care should be taken to minimize soil disturbance, which may potentially alter soil thermal properties.
Top left: SP-230 Heated Pyranometer
Center: TS-100 Fan Aspirated Radiation Shield
Right: RM Young 41303 Static Solar Radiation Shield.
OPERATION AND MEASUREMENT
VERY IMPORTANT: Apogee changed all wiring colors of our bare-lead sensors in
March 2018. To ensure proper connection to your data device, please note your
serial number then use the appropriate wiring configuration below.
Wiring for ST-100 Serial Numbers 2725 and above, ST-110 Serial Numbers 2725
and above, and ST-200 Serial Numbers 1352 and above Wiring for ST-100 Serial
Numbers range 0-2724, ST-110 Serial Numbers range 0-2724, and ST-200 Serial
Numbers range 0-1351
Plug each wire into the appropriate data logger terminal following the
illustrations below: Measurement devices (e.g., datalogger, controller) do not
measure resistance directly, but determine resistance from a half-bridge
measurement, where an excitation voltage is input across the thermistor and an
output voltage is measured across the bridge resistor. NOTE: The wiring
diagram above is based off the new wiring colors for serial numbers 2725 or
1352 and above.
An excitation voltage of 2.5 V DC is recommended to minimize self-heating and
current drain, while still maintaining adequate measurement sensitivity (mV
output from thermistor per C). However, other excitation voltages can be used.
Decreasing the excitation voltage will decrease self-heating and current
drain, but will also decrease thermistor measurement sensitivity. Increasing
the excitation voltage will increase thermistor measurement sensitivity, but
will also increase self-heating and current drain.
Conversion of Thermistor Resistance to Temperature
The thermistor is a resistive element, where resistance changes with
temperature. Thermistor resistance (RT, in Ω) is measured with a half-bridge
measurement, requiring a known excitation voltage input (VEX) and a
measurement of output voltage (VOUT): where 24900 is the resistance of the
bridge resistor in Ω. From resistance, temperature (TK, in Kelvin) is
calculated with the Steinhart-Hart equation and thermistor-specific
coefficients: where A = 1.129241 x 10 -3 , B = 2.341077 x 10 -4 , and C =
8.775468 x 10 -8 (Steinhart-Hart coefficients). If desired, the measured
temperature in Kelvin can be converted to Celsius (TC): ST-150
The ST-150 does not include bridge resistors and therefore requires
measurement with dataloggers or controllers that are capable of current
excitation. Although the ST-150 can be used in a 2-wire or 3-wire measurement
configuration, a 4-wire configuration is recommended for the highest accuracy
(a 4-wire measurement configuration eliminates the influence of lead wires).
In the 4-wire configuration, an excitation current is applied across the PRT
with two of the lead wires, voltage is measured across the other two lead
wires using a differential measurement, and resistance of the PRT is
calculated by dividing the measured voltage by the excitation current (Ohm’s
Law). Current does not flow in the wires where voltage is measured, so the
resistance of the lead wires does not influence the measurement. The
temperature of the PRT is calculated from resistance using equations (2) and
(3) on page 13. If a half-bridge measurement is desired, the same wiring and
bridge resistors detailed below in the ST-300 section can be used to make the
measurement.
Note: as shown in the schematic, the 4-wire PRT can be directly measured.
The black and blue wires are electrically connected to one side of the PRT
while the yellow and white wires are electrically connected to the other side
of the PRT.
ST-300 PRT
Connect the sensor to a measurement device (meter, datalogger, controller)
capable of inputting 2.1 V DC, and measuring and displaying or recording a
millivolt (mV) signal (an input measurement range of 16 to 27 mV is required
to cover the entire temperature range of the sensor). In order to maximize
measurement resolution and signal-to-noise ratio, the input range of the
measurement device should closely match the output range of the thermistor.
VERY IMPORTANT: Apogee changed all wiring colors of our bare-lead sensors in
March 2018. To ensure proper connection to your data device, please note your
serial number then use the appropriate wiring configuration below.
Wiring for ST-300 Serial Numbers 1076 and above
Wiring for ST-300 Serial Numbers range 0-1075
Measurement of PRT resistance is very similar to the measurement of thermistor
resistance, where a half-bridge measurement is used. An excitation voltage is
input across the bridge resistor and an output voltage is measured across the
PRT. NOTE: The wiring diagram below is based off the new wiring colors for
serial numbers 1076 and above.
Note: as shown in the schematic, the 4-wire PRT can be directly measured
without the bridge resistors. The blue and yellow wires are electrically
connected to one side of the
PRT while the green and black wires are electrically connected to the other
side of the PRT.
An excitation voltage of 2.1 V DC is recommended to minimize current drain,
while still maintaining an adequate voltage signal. However, other excitation
voltages can be used. Decreasing the excitation voltage will decrease current
drain, but will also decrease output voltage. Increasing the excitation
voltage will increase output voltage, but will also increase current drain.
Conversion of PRT Resistance to Temperature
The PRT is a resistive element, where resistance changes with temperature. PRT
resistance (RPRT, in Ω) is measured with a half-bridge measurement, requiring
a known excitation voltage input (VEX) and a measurement of output voltage
(VOUT):
where 100 Ω is the resistance of the bridge resistor, V100Ω is the voltage measured across the 100 Ω bridge resistor, and VPRT is the voltage measured across the PRT. From resistance, temperature (TC, in Celsius) is calculated with:
where K = (RPRT / 100) – 1, a = 3.9083 x 10 -3 , d = -2.3100000 x 10 -6 , e = 1.7584810 x 10 -5 , f = -1.1550000 x 10 -6 , g = 1.7909000, h = -2.9236300, i = 9.1455000, and = 2.55819 x 10 2 .
MAINTENANCE AND RECALIBRATION
When sensors are not in use, it is recommended that they be removed from the
measurement environment, cleaned, and stored. ST series temperature sensors
used to measure air temperature should be periodically cleaned to remove all
dust and debris.
Apogee ST-100 temperature sensors are weatherproof and can be submerged in
water or buried in soil/porous media. ST-110 sensors are weatherproof and
designed for air temperature measurements inside radiation shields. ST-200
series sensors are weather-resistant, but not weatherproof. When sensors are
not in use, it is recommended that they be removed from the measurement
environment, cleaned, and stored. ST series temperature sensors used to
measure air temperature should be periodically cleaned to remove all dust and
debris.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee ST Series thermistor temperature sensors yield a resistance
proportional to temperature. A quick and easy check of thermistor/sensor
functionality can be accomplished with an ohmmeter. Connect the lead wires of
the ohmmeter to the red and white wires from the sensor. The resistance should
read 10 kΩ (10,000 ohms) at 25 C. If sensor temperature is less than 25 C,
resistance will be higher. If sensor temperature is greater than 25 C,
resistance will be lower. Connect the lead wires of the ohmmeter to the white
and black wires from the sensor. The resistance should read 24.9 kΩ, and
should not vary. Connect the lead wires of the ohmmeter to the red and black
wires from the sensor. The resistance should be the sum of resistances
measured across the red and white wires, and white and black wires (e.g., 10
kΩ plus 24.9 kΩ at 25 C).
Apogee ST-300 temperature sensors yield a resistance proportional to
temperature. A quick and easy check of sensor functionality can be
accomplished with an ohmmeter. Connect the lead wires of the ohmmeter to the
blue/yellow and green/black wires from the sensor. The resistance should read
100Ω (0.1 kΩ) at 0 C. If the sensor temperature is less than 0 C, the
resistance will be lower. If the sensor temperature is greater than 0 C, the
resistance will be higher. The blue and yellow wires are connected internally,
without a resistor between them; continuity between blue and yellow wires
indicates both wires are functional. Similarly, the green and black wires are
connected internally, without a resistor between them; continuity between
green and black wires indicates both wires are functional. Connect the lead
wires of the ohmmeter to the green/black and white wires from the sensor. The
resistance should read 100 Ω, and should not vary. Connect the lead wires of
the ohmmeter to the white and red wires from the sensor. The resistance should
read 10 kΩ, and should not vary. Connect the lead wires of the ohmmeter to the
blue/yellow and red wires from the sensor. The resistance should be the sum of
the resistances measured across the blue/yellow and green/black wires,
green/black and white wires, and white and red wires (e.g., 100 Ω plus 100 Ω
plus 10 kΩ at 0 C).
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
Measurement of thermistor resistance requires an input excitation voltage,
where 2.5 V DC is recommended.
Measurement of ST-300 PRT resistance also requires an input excitation
voltage, where 2.1 V DC is recommended.
A compatible measurement device should have the capability to supply the
necessary voltage.
The sensitivity (mV output from thermistor per C) of the temperature
measurement varies with the excitation voltage and varies as a function of
temperature. With an excitation voltage of 2.5 V DC, the sensitivity is
lowest near the ends of the measurement range, -60 and 80 C. A compatible
measurement device (e.g., datalogger or controller) should have a resolution
of at least 0.6 mV, in order to produce a temperature resolution of less than
0.1 C across the entire temperature measurement range (less than 0.05 C from
-35 to 45 C). The sensitivity (mV output from PRT per C) of the temperature
measurement from the ST-300 PRT is approximately constant across the entire
measurement range. With an excitation voltage of 2.1 V DC, a compatible
measurement device should have a resolution of at least 0.008 mV, in order to
produce a temperature resolution of less than 0.1 C across the entire
temperature measurement range. An example datalogger program for Campbell
Scientific dataloggers can be found on the Apogee webpage at
http://www.apogeeinstruments.com/content/Thermistor-Temperature-
Sensor.CR1.
Modifying Cable Length
When the sensor is connected to a measurement device with high input
impedance, sensor output signals are not changed by splicing on additional
cable in the field. Tests have shown that if the input impedance of the
measurements device is 1 mega-ohm or higher then there is negligible effect on
ST series temperature sensors, even after adding up to 100 m of cable. See the
Apogee webpage for details on how to extend sensor cable length
(http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/).
For cable extensions, shielded, twisted-pair cable is recommended, in order to
minimize electromagnetic interference. This is particularly important for long
lead lengths in electromagnetically noisy environments. The precision bridge
resistor is located at the pigtail end of the ST-300 cable. Thus, the ST-300
temperature sensor cables should not be shortened, otherwise, the bridge
resistor will be removed.
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as
the product is in new condition (to be determined by Apogee). Returns are
subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from
defects in materials and craftsmanship for a period of four (4) years from the
date of shipment from our factory. To be considered for warranty coverage an
item must be evaluated by Apogee.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content
meters, EE08-SS probes) are covered for a period of one (1) year.
What is Not Covered
The customer is responsible for all costs associated with the removal,
reinstallation, and shipping of suspected warranty items to our factory.
The warranty does not cover equipment that has been damaged due to the
following conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine
recalibration of sensors/meters is considered part of proper maintenance and
is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
- Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking of the service item. Call 435-245-8012 or email techsupport@apogeeinstruments.com with questions.
- For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise, the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee’s standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
- Please write the RMA number on the outside of the shipping container.
- Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders. Apogee Instruments, Inc. 721 West 1800 North Logan, UT 84321, USA
- Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com
to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss. This limited warranty and any disputes arising out of or in
connection with this limited warranty (“Disputes”) shall be governed by the
laws of the State of Utah, USA, excluding conflicts of law principles and
excluding the Convention for the International Sale of Goods. The courts
located in the State of Utah, USA, shall have exclusive jurisdiction over any
Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot by transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision shall
be deemed severable and shall not affect any remaining provisions. In case of
any inconsistency between the English and other versions of this limited
warranty, the English version shall prevail. This warranty cannot be changed,
assumed, or amended by any other person or agreement
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL:435-792-4700 |
FAX:435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM Copyright © 2021 Apogee
Instruments, Inc.