RS PRO RS1710 Wired Digital Thermometer Instruction Manual

June 3, 2024
RS PRO

Stock number: 205-0959
Stock number: 205-0960
RS1710 Wired Digital Thermometer

Instruction Manual

RS1710 Wired Digital Thermometer

RS PRO RS1710 Wired Digital Thermometer

About this documentation

1.1 Purpose of the document

  • For this reason, read the operating manual before operating the product for the first time.

1.2 Legal notices
This document is entrusted to the recipient for personal use only. Any impermissible transfer, duplication, translation into other languages or excerpts from this operating
manual is prohibited.
The manufacturer assumes no liability for print errors.
1.3 Further information
The software version of the product:

  • V1.2 or later

For the exact product name, refer to the type plate on the rear side of the product.
NOTE
For information about the software version, press and hold the ON button to switch on the product for longer than 5 seconds. The series is shown in the main display and the software version of the product is shown in the secondary display.

Safety

2.1 Explanation of safety symbols
DANGER
This symbol warns of imminent danger, which can result in death, severe bodily injury, r severe property damage in case of non-observance.
CAUTION
This symbol warns of potential dangers or harmful situations, which can cause damage to the device or to the environment in case of non-observance.
NOTE
This symbol indicates processes, which can have a direct influence on the operation or can trigger an unforeseen reaction in case of non-observance.
2.2 Foreseeable misuse
The fault-free function and operational safety of the product can only be guaranteed if applicable safety precautions and the device-specific safety instructions for this document are observed.
If these notices are disregarded, personal injury or death, as well as property damage can occur.
DANGER
Incorrect area of application!
In order to prevent erratic behavior of the product, personal injury, and property damage, the product must be used exclusively as described in the chapter Description in the operating manual.

  • The product is not suitable for use in explosion-prone areas!
  • The product must not be used for diagnostic or other medical purposes on patients!
  • Not suitable for use with requirements on functional safety, e.g. SIL!

2.3 Safety instructions
CAUTION
Stab injury!
Products with insertion probes entail the risk of stab injuries due to the pointed probe design.

  • Handle insertion probes with care!
  • Fit a protective cap on the measuring probe!

NOTE
This product does not belong in children’s hands!
NOTE
The sensor handle, connecting cable, and product housing are not designed for continuous contact with foods.
Designed for continuous contact with foods in accordance with EC Regulation 1935 / 2004:

  • The temperature sensor from the measuring tip to approx. 1 cm before the end of the stainless steel tube.

2.4 Intended use
The product is a water-protected thermometer. It is designed for precise and instantaneous temperature measurements in the following media:

  • Food
  • Liquids
  • Gases
  • Soft plastic materials
  • Bulk material

The product at a glance

3.1 RS 1700 series

3.2 Display elements
Display

| Battery indicator| Evaluation of the battery status
---|---|---
| Unit display| Display of units or type of mode, min/max/hold
| Main display| Measurement of the current temperature or value
for min/max/hold
| Auxiliary display| Measurement of the current temperature in min/max/hold mode with unit

3.3 Operating elements
On / Off button

Press briefly Switch on the product Activate / deactivate lighting
Long press Switch off the product Reject changes in a menu

Up / Down button

Press briefly| Display of the min/max value Change the value of the selected parameter
---|---
Long press| Reset the min/max value of the current measurement
Both simultaneously| Rotate display, overhead display

Function key

Press briefly| Freeze measurement (Hold)
Return to measurement display
Call up the next parameter
---|---
Long press, 2s| Start menu configuration, (ONF appears in the
display

Operation

4.1 Opening the configuration menu

  1. Press the Function key for 2 seconds to open the Configuration menu.
  2. ( appears in the display. Release the Function key.
Parameter Values Meaning
Alarms
No active alarm
Alarm alerting via text display, acoustic signal and

flashing of the backlighting
| | Alarm alerting via text display and acoustic signal
| | Alarm alerting via text display and flashing of the
backlighting
| |
| | Min. alarm limit; a min. the alarm is triggered when the
value is undercut, e.g. at -94.0 °F
| |
| |
| | Max. alarm limit; a min. the alarm is triggered when the
value is exceeded, e.g. at 482.0 °F
Shut-off time| |
---|---|---
| |
| | No automatic shut-off
| | Automatic shut-off after a selected time in minutes, during which no buttons have been pressed
Backlight| |
---|---|---
| |
| | Backlight deactivated
| | Automatic shut-off of the backlight after a selected
time in seconds, during which no buttons have been pressed
| | No automatic shut-off of the backlight
Display unit| |
---|---|---
| |
| | Temperature display in °C
| | Temperature display in °F
Factory settings| |
---|---|---
| |
| | Use current configuration
| | Reset the product to factory settings. appears in the display

4.2 Adjustment of the measuring input

The temperature input can be adjusted with the zero point correction and the gradient correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signaled with the or display text when the product is switched on.

  1. The product is switched off
  2. Press and hold the Down button.
  3. Press the On/Off button to switch on the product and open the Configuration menu. Release the Down button. The display shows the first parameter.
Parameter Values Meaning
Zero-point correction of the temperature
No zero-point correction
Zero point correction in °C. and/or at °F -9.00 .. 9.00
Gradient correction of the temperature
--- --- ---
No gradient correction
Gradient correction in %

Formula:

Zero point correction: Displayed value = measured value –
Gradient correction °C: Display = (measured value – ) * (1 + / 100)
Gradient correction °F: Display = (meas. value – 32 °F – ) * (1 + / 100) + 32

°F

Operation and maintenance

5.1 Operating and maintenance notices
NOTE

The product and temperature probe must be handled with care and used in accordance with the technical data. Do not throw or strike.
NOTE
Plugs and sockets must be protected from soiling.
NOTE
If the product is stored at a temperature above 50 °C or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as
a result.
5.2 Battery
5.2.1 Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time.
If the BAT display text appears in the main display, the battery voltage is no longer adequate for the operation of the product. The battery is fully depleted.
5.2.2 Changing battery
DANGER

The danger of explosion!
Using damaged or unsuitable batteries can generate heat, which can cause the batteries to crack and possibly explode!

  • Only use high-quality and suitable alkaline batteries!

CAUTION
Damage!
If the batteries have different charge levels, leaks and thus damage to the product can occur.

  • Use new, high-quality batteries!
  • Do not use different types of batteries!
  • Remove depleted batteries and dispose of them at a suitable collection point.

NOTE
This symbol indicates processes, which can have a direct influence on the operation or can trigger an unforeseen reaction in case of non-observance.
NOTE
Read the following handling instructions before replacing batteries and follow them step by step.
If disregarded, the product could be damaged or the protection from moisture could be diminished.

  1. Unscrews the Phillips screws (A)and remove the cover.
  2. Carefully replace the two Mignon AA batteries (B). Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
  3. The O-ring (C) must be undamaged, clean, and positioned at the intended depth. In order to facilitate assembly and avoid damage, suitable grease can be applied.
  4. Fit the cover on evenly. The O-ring must remain at the intended depth!
  5. Tighten the Phillips screws (A).

Error and system messages

Display Meaning Possible causes Remedy
No suitable measuring probe connected Incorrect measuring
probe Connect a suitable measuring

probe
Measurement far outside of the measuring range| Measuring probe or
product defect| Measurement leaves the
permissible range  Send in for repair
| Sensor cable defect| Cable breakage| Send in for repair
Sensor or probe defect| Defective sensor or
probe| Send in for repair
Measuring range exceeded or undercut| Measurement
outside of the measuring
range|
No display, unclear characters
or no response when buttons  are pressed| Battery depleted| Battery depleted| Replace battery
System error| Error in the product| Send in for repair
Product is defective| Product is defective|
| Battery depleted| Battery depleted| Replace battery
| Measuring range exceeded| Measurement too high| Stay within the allowable measurement range
Incorrect measuring probe connected| Check to measure probe Send in for repair
Measuring probe or product defect|
| The measuring range is undercut| Measurement too low| Stay within the allowable measurement range
Measuring probe or product defect| Check measuring probe Send in for repair
| System error| Error in the product| Switch product on/off Replace batteries
Send in for repair

Technical data

Measuring range temperature -70.0 .. +250.0 °C (-94.0 .. +482.0 °F)
Accuracy temperature -20 .. +100 °C: ±0.1 K ±1 digit

otherwise: ±0.2 To of measured value ± 2 digits

Response time T90 water (0.4 m/s)| approx. 3 s
Measuring cycle| approx. 2 measurements per second
Probe| Immersion sensor 0 3 mm, Pt1000 permanent 2-wire connection, V4A, 1 m cable (RS 1710)
Durable insertion sensor 0 3 mm, Pt1000 permanent 2-wire connection, V4A, 1 m cable (RS 1720)
Response time T 90| 0 3 mm: Water 0.4 m/s <2 s (RS 1710)
0 3 mm: Water 0.4 m/s <2 s (RS 1720)
Display| 3-line segment LCD, additional symbols, illuminated (adjustable white, permanent illumination)
Additional functions| Min/max/hold, alarm (optical and acoustic)
| Compensation| Offset and gradient correction
Housing| | Break-proof ABS housing
---|---|---
Protection rating| IP65 / IP67
Dimensions L’W’H [mm]| 108′ 54′ 28 mm without kink protection
Weight| 150 g, incl. battery, and sensor
Operating conditions Storage temperature| -20 to 50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.) -20 to 70 °C
Current supply| | 2*AA battery (included in the scope of delivery)
Current requirement/

battery life

| approx. 0.4 mA, approx. 2 mA with lighting

Service life > 5000 hours with alkaline batteries (without backlighting)

Battery

indicator

| 4-stage battery status indicator,

Replacement indicator for depleted batteries: “BAT”

Auto-power-OFF function
Directives and standards| The device switches off automatically if this is activated

The devices conform to the following Directives of the Council for the harmonization of legal regulations of the Member States:

2014/30/EU EMC Directive 2011/65/EU RoHS

Applied harmonized standards:
EN 61326-1:2013        Emission limits: Class B
Immunity according to Table 2 Additional errors: < 0.5 % FS
EN 50581:2012
The device is intended for mobile use and/or stationary operation in the scope of the specified operating condi­tions without further limitations.

Service

8.1 Manufacturer
If you have any questions, please do not hesitate to contact us:
Contact

RS Components Limited
Birchington Road
Corby
Northamptonshire
NN17 9RS
WEE/GF0002ZR