Rebotec TRE-1860010 Fixi (Small) and Fox (Junior) Folding Pediatric Anterior Walker Instructions
- June 5, 2024
- REBOTEC
Table of Contents
Maintenance Instructions
Rollators and Walkers Images show a selection of rollators and walkers
Important advisory!
Store the maintenance inspection reports carefully. This will allow you to
keep an
overview of the product’s maintenance status. ****
Maintenance of the rollator and walker
1.1 Maintenance / inspection requirements
In this maintenance instructions pamphlet, you’ll find information and
advisories on the safe and professional maintenance and/or repair work on
rollator and walker.
The products are Class I medical products as to the 93/42/EEC and the
German MPG/Medical Products Act.
As to the requirements of the German MPBetreibV Medical Products Ordinance,
the products must be maintained and the hygiene requirements must be adhered
to. Check your local requirements.
The product’s lifetime is in large part affected by how it is used. The
frequency of use and manner of use reduces the lifetime as does improper
handling and inadequate maintenance.
In order to ensure safe use, a visual inspection and function check must be
done at least once per year. Shorter time intervals between maintenance may be
necessary when the frequency of use or the condition of the product exists due
to safety reasons.
The operator has to carry out the maintenance on his own or must instruct a
competent person (specialized supplier, medical supply store, or a medical-
maintenance service) if he has not had the needed expertise. Inspection
results must be documented and confirmed by a signature. The product is
recommended with an inspection sticker that states the next inspection. The
product operator is responsible for assuring regularly-conducted inspections.
1.2 Maintenance scope
The product’s inspection extends from the general condition to the individual
components, function, and safety.
Necessary repair works must be done without delay. If immediate repair work is not possible, then it is in the patient’s interest and health to put the product out of service.
1.3 Maintenance advisories
When making repairs, use only original spare parts from REBOTEC.
Do not undertake any modifications or conversions to the product on your own.
These will adversely impact the product’s safety and function. REBOTEC accepts
no liability in such cases.
When performing maintenance and repairs, please use the product’s User
Instructions booklet as an aid and follow the instructions concerning the
proper use, re-use, lifetime, guarantee, care, and maintenance.
After completing maintenance or repair work on the product, follow the
requirements for hygiene for re-use.
1.4 Maintenance reports and overview
In the appendix, you will find a form for documenting maintenance. You can
make copies of the form. This allows you to keep an overview of the product’s
maintenance status.
Maintenance protocol (single report) It is a form for proof of a single
maintenance session.
Maintenance overview (total report) It is a form for proof of executed
maintenance sessions.
1.5 Disposal
For disposing of the product, follow all the advisories in the product’s
User Instructions booklet. Disposal must be done at a recycling company.
Contact your local recycling company for more information.
1.6 Transport and storage
For transporting or storing the product follow the instructions of the
particular user manual.
Advisory on other documents
At our www.rebotec.de website, you can download the
User Instructions booklets, forms, and other information.
Maintenance protocol No.:
Product information / Rollator, Walker (medical product class I)
REBB02.11.149-17-06
Product| |
---|---|---
Serial no.| Year of| | Use| |
Manufacturer| |
User| |
Location of| |
Maintenance- General Condition
P | Checks | Scope of work | Remarks | **Finding*** |
---|---|---|---|---|
1. | Visual inspection | General damage, wear, and corrosion | ||
2. | Connection elements | Completeness, seat, function (screws, nuts, etc.) | ||
3. | Functions | Folding mechanism | ||
Height adjustment | ||||
Brake (permanent lock/normal brake) |
Maintenance- Component checking
P | Article | Scope of work | Remarks | Finding |
---|---|---|---|---|
1. | rollator frame | Damages | ||
2. | grip bow | Function, fastening | ||
3. | swivel castors | Function, damage, wear | ||
4. | brake wheels | Function, damage, wear | ||
5. | brake levers | Function, damage, wear | ||
6. | hand switch | Function, damage, wear | ||
7. | wood seat | Stability, damage, wear | ||
8. | cushion | Damage | ||
9. | accessory | Function, wear | ||
10. | YANO-support | Function, fastening | ||
11. | YANO-handle | Function, damage, wear | ||
12. | Castor brakes | Function, damage, wear | ||
13. | YANO-accessory | Function, damage, wear |
* Findings: I=inspected; E=exchanged’ R=repaired, i.o. = in order; n.i.o. = not in order
The following parts were replaced: (Use only original parts from the manufacturer)
Qty. | Article description |
---|---|
Maintenance results / summary
Maintenance overview
Product information / Rollator, Walker (medical product class I)
REB02.11.149-17-06
Product|
---|---
Serial no.| | Year manuf.| Use start|
Manufacturer|
User|
Location of use| | Inventory no.|
1st Maintenance Date| Maintenance company| Report| Signature
---|---|---|---
| | |
Remarks:
2nd Maintenance Date| Maintenance company| Report| Signature
| | |
Remarks:
3rd Maintenance Date| Maintenance company| Report| Signature
| | |
Remarks:
4th Maintenance Date| Maintenance company| Report| Signature
| | |
Remarks:
5th Maintenance Date| Maintenance company| Report| Signature
| | |
Remarks:
REBOTEC / WA-No: 02.11.152/2017-06