chattanooga M2 Hydrocollator Hot Pack Heating Units User Manual

June 5, 2024
chattanooga

Moving
Rehabilitation Forward™
Hydrocollator®
HEATING UNITS

chattanooga M2 Hydrocollator Hot Pack Heating Units

User Manual
120 – 220/240 Volt Units

Hydrocollator® is a registered trademark of DJO, LLC HotPac™ is a trademark of DJO, LLC Comet® of Prestige Brands, Inc.
Soft Scrub® of the Dial Corporation Scotch-Brite™ of 3M
©2021 DJO, LLC Carlsbad, California USA. Any use of editorial, pictorial, or layout composition of this publication without express written consent from DJO, LLC is strictly prohibited. This publication was written, illustrated, and prepared for distribution by DJO, LLC.

FOREWORD

This manual contains general instructions for operation, precautionary instructions, and maintenance recommendations for the owners and operators of the Hydrocollator SS-2,
M-2, SS, E-2, E-1, and M-4 Heating Units. In order to obtain maximum life and efficiency of the Hydrocollator Heating Units, please read this manual thoroughly and follow all
instructions prior to operating the unit.
Specifications put forth in this manual were in effect at the time of publication. However, owing to DJO, LLC’s policy of continual improvement, changes to these specifications may be made at any time without obligation on the part of DJO, LLC.

PRECAUTIONARY DEFINITIONS

The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definitions of these symbols are as follows:

= CAUTIONS – Text with a “CAUTION” indicator explains possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
** = WARNINGS – Text with a “WARNING” indicator explains possible safety infractions that will potentially cause serious injury and equipment damage.
=
DANGERS – Text with a “DANGER” indicator explains possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
NOTE:** Throughout this manual, “NOTE” may be found. These notes are helpful information to aid in the particular area or function being described.

SAFETY PRECAUTIONS

CAUTION

  • Read, understand and practice the precautionary and operating instructions. Know the limitations associated with the Hydrocollator SS-2, M-2, SS, E-2, E-1, M-4 Heating Units. Observe the cautionary and operational decals installed on the unit.
  • DO NOT operate this unit in an environment where other devices are being used that intentionally radiate electromagnetic energy in an unshielded manner. Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
  • The Hydrocollator Heating Unit generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning the Hydrocollator Heating Unit on and off. Try to correct the interference using one or more of the following: reorient or relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet on a different circuit from that which the other device(s) are connected and consult the DJO, LLC Service Department for help.
  • Do not use accessories other than those supplied with the Hydrocollator Heating Unit or recommended by DJO, LLC. The safety of other products has not been established, and their use could result in injury to the patient.
  • Never adjust the thermostat too high. The thermostat is extremely sensitive and the slightest adjustment will alter the temperature several degrees. The recommended operating temperature is 71° to 74° C (160° F to 165° F). The temperature of the water should be checked with a thermometer after every adjustment, before using the HotPac. Always allow sufficient time for the water temperature to stabilize.
  • The Hydrocollator Heating Unit should not be installed above a surface that is susceptible to casual water damage.
  • Always keep the water level over the top of the HotPacs to avoid burning or scorching of the packs.
  • Never allow the HotPacs to dry after use.
  • Check the water level daily as it has a natural loss due to evaporation.
  • Clean the tank periodically as described in the maintenance portion of this manual.
  • Check the unit periodically for signs of leaking and contact service if leaks are noted.
  • Always replace Hydrocollator HotPacs as soon as they show signs of wear.
  • If the unit is to be left unattended for a period of time, unplug the unit, remove the packs, empty the water, and clean the tank.
  • Do not move the Hydrocollator while filled with heated water. Tipping over could result in burns to the user or others.
  • Always unplug the Hydrocollator from the power source before attempting to empty water from the unit.
  • Always use factory-authorized replacement parts.
  • Always destroy and replace HotPacs that are exposed to contamination: hazardous or cleaning materials, bodily fluids, mold, etc.
  • This unit should be operated, transported, and stored in temperatures between -18° C and 60° C (0° F and 140° F) and atmospheric pressure range between 50 kPa – 160 kPa, with relative humidity ranging from 10% – 100%.
  • Treatment time should not exceed 30 minutes.
  • DO NOT retrieve HotPacs from the tank or replace HotPacs in the tank with an unprotected hand.
  • Avoid direct contact with the mounting rack in the tank.
  • Always wrap HotPac with a towel or terry cover before handling or applying to the patient.
  • Constantly monitor the HotPac application to ensure that the skin is not becoming too hot.
  • Damage to the skin can occur from exposure to extreme heat or cold. Note instructions for proper use.
  • DO NOT apply over insensitive skin or in the presence of poor circulation.
  • Never lay or sit on top of the HotPac.
  • DO NOT use HotPac directly over cuts, abrasions, or wounds.
  • Consult health care professionals before using HotPacs on non-communicative individuals.
  • Exercise extreme caution when using HotPacs on non-communicative individuals.

WARNING

  • Grounding the Hydrocollator SS, SS-2, E-1, E-2, M-2, and M-4 Heating Units: Make certain the unit is electrically grounded by plugging into a hospital-grade electrical outlet with a ground terminal receptacle (U-ground outlet). Follow the National Electrical Code.
  • The water temperature in the Hydrocollator is approximately 71° C (160° F), and the water scalding temperature is approximately 49° C (120° F).
  • Care must be taken when operating the Hydrocollator adjacent to or stacked with other equipment. Potential electromagnetic or other interference could occur to this or other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
  • The Hydrocollator Heating Unit should be placed on a smooth, level surface. The surface should support up to the Filled Weight (weight of heating unit filled with water) found in Technical Specifications pages 8-10 of this manual.
  • Use only accessories that are specially designed for the Hydrocollator. Do not use accessories manufactured by other companies on the Hydrocollator. DJO, LLC is not responsible for any consequence resulting from using products manufactured by other companies. The use of other accessories or cables may result in increased emissions or decreased immunity of the Hydrocollator.
  • Periodically inspect unit for signs of damage and pitting during regular cleaning intervals, as described in the maintenance portion of this manual. Contact service if any signs of damage are discovered.
  • DO NOT apply HotPac directly to skin.
  • Individuals with circulatory problems should consult with a physician before using this product.
  • DO NOT sit or lie on the HotPac.
  • DO NOT add chlorine additive to water.
  • DO NOT allow HotPac to come in contact with hydrocarbons (e.g., Comet®, Soft Scrub®, etc.).

Risk of explosion if used in the presence of flammable anesthetics.

Explanation of Symbols
Refer to Instruction Manual/Booklet
Type B equipment
Thermostat Intensity Control

NOMENCLATURE

WHITE LAMP (where applicable), when illuminated, indicates that the main power is connected, the switch is on and the unit is operating.
RED LAMP (where applicable), when illuminated, warns of an overheat/low water condition. Action is necessary to correct the condition.
GREEN LAMP (220/240-volt units), when illuminated, indicates that the main power is connected, the switch is on and the unit is operating.

chattanooga M2 Hydrocollator Hot Pack Heating Units -
fig1

IMPORTANT: Some models are fitted with a thermal cut-out thermostat which acts as a “non-self-resetting” thermal cut-out. In the unlikely event of a fault condition, this device is designed to open the power circuit between 88° C ± 8° C (190° F ± 14° F). It is not possible to alter its operating range. The operation of the unit may be restored by
depressing the red button after the water/element temperatures have been allowed to cool. If the fault condition persists, please contact DJO, LLC or an authorized agent.

Diagram

1. Heating Element
2. Drain
3. Overtemp Thermostat
4. Thermostat, Hydraulic
5. Bracket (Bracket & Bulb Thermostat A/B Assembly)
6. Lid
7. 3-inch Caster
8. Overtemp Light| 9. Junction Box (J-Box)Cover
10. IEC connector
11. Power Cord
12. Drain Valve (Not Shown)
13. Reset Overtemp Button*
14. 6.3 Amp Fuses
15. Pilot Lamp, Red
16. Pilot Lamp, Clear
---|---

chattanooga M2 Hydrocollator Hot Pack Heating Units -
fig2

NOTE: M-4 model shown; models will vary.

  • All 220-240 Volt and M-4 120 Volt have this button

Diagram (CONTINUED)

17. HotPac
18. Neck Pac Insert
19. Drain, Copper Tube
20. Lid Hinge Nuts
21. Lid Knob/Handle
22. Lid Hinge| 23. Side Handle
24. Thermostat, Hydraulic
25. Thermostat Intensity Control
26. On/Off Switch
27. Internal Wiring Harness
---|---

chattanooga M2 Hydrocollator Hot Pack Heating Units -
fig3

Not illustrated:
28. Stainless Steel Wire Rack
29. DrainValve, Ball – 1/2 in Replacement
30 Drain, Hose, Bib
31. Replacement Tank
32. Outer Jacket
33. Thermo Low water kit
34. Lid Kit Handle
35. Water Thief
36. Bracket, Element Support (2 each) [M-4 only] 37. Plastic Feet
38. 10 amp breaker
39. Drain hose

TECHNICAL SPECIFICATIONS

SS-2

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-out Temp
Heating Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~ 220-240 V~
50/60 Hz
1000 W
49.32 Liters
85 kg (140 lbs)
71° C – 74° C (160° F – 165° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 4 Hours
Time from 70° C (160° F) – 2 Hours
Type B ****
IEC/EN
60601-1
85 cm H x 52 cm L x 39 cm W (33 in H x 21 in L x 16 in W)
30.8 kg (68 lbs)
33.6 kg (74 lbs)
---|---

Standard Accessories : Includes eight Standard size HotPacs.
M-2

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-Out Temp
Heat Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~ 220-240 V~
50/60 Hz
1000 W
68.81 Liters
110 kg (190 lbs)
71° C – 74° C (160° F – 165 ° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 6 Hours
Time from 70° C (160° F) – 3 Hours
Type B ****
IEC/EN
60601-1
85 cm H x 67 cm L x 40 cm W (33 in H x 27 in L x 16 in W)
39 kg (86 lbs)
42.2 kg (93 lbs)
---|---

Standard Accessories: Includes twelve Standard size HotPacs.
M-4

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-out Temp
Heating Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~ 220-240 V~
50/60 Hz
1500 W
135.76 Liters
200 kg (340 lbs)
71° C – 74° C (160° F – 165° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 8 Hours
Time from 70° C (160° F) – 4 Hours
Type B ****
IEC/EN
60601-1
85 cm H x 89 cm L x 50 cm W (33 in H x 35 in L x 20 in W)
61.2 kg (135 lbs)
64.9 kg (143 lbs)
---|---

Standard Accessories: Includes twenty-four Standard size HotPacs.
SS

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-out Temp
Heating Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~, 220-240 V~
50/60 Hz
1000 W
31.64 Liters
56 kg (120 lbs)
71° C – 74° C (160° F – 165° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 3 Hours
Time from 70° C (160° F) – 2 Hours
Type B ****
IEC/EN
60601-1
51 cm H x 46 cm L x 41 cm W (20 in H x 18 in L x 16 in W)
23.1 kg (51 lbs)
24.5 kg (54 lbs)
---|---

Standard Accessories: Includes eight Standard size HotPacs.
E-2

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-Out Temp
Heat Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~, 220-240 V~
50/60 Hz
1000 W
43.23 Liters
60 kg (125 lbs)
71° C – 74° C (160° F – 165° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 4 Hours
Time from 70° C (160° F) – 2 Hours
Type B ****
IEC/EN
60601-1
51 cm H x 38 cm L x 33 cm W (20 in H x 15 in L x 13 in W)
14.1 kg (31 lbs)
15.9 kg (35 lbs)
---|---

Standard Accessories: Includes two Oversize HotPacs, three Standard size HotPacs and 1 Neck Contour HotPac.
E-1

Power Supply
Rated Frequency
Power Consumption
Tank Capacity
Filled Weight
Temperature Range
Thermal Cut-out Temp
Heating Up Time
Cool Down Time
Safety Class
Safety
Tests
Dimensions
Product Weight
Shipping Weight| 110-120 V~, 220-240 V~
50/60 Hz
1000 W
15.44 Liters
25 kg (55 lbs)
71° C – 74° C (160° F – 165° F)
88° C ± 8° C (190° F ± 14° F)
Time to 70° C (160° F) – 2 Hours
Time from 70° C (160° F) – 1 Hour
Type B ****
IEC/EN
60601-1
41 cm H x 33 cm L x 20 cm W (16 in H x 13 in L x 8 in W)
6.4 kg (14 lbs)
7.3 kg (16 lbs)
---|---

Standard Accessories: Includes four Standard size HotPacs.

SETUP

Remove all of the contents inside the Hydrocollator Heating Unit and then replace the wire rack. Make certain the drain valve is closed before filling the tank with water. The
tank should be approximately 3/4 full with packs removed. The tank may be filled from any standard type faucet.
Check the voltage rating on the serial decal plate which is located on the back of the unit. To turn the unit on, simply plug the unit into a 120 Volt, 60 Hz or 220/240 Volt, 50-60 Hz outlet as indicated on the decal plate. To turn the unit off, simply unplug it from the outlet. Do not attempt to use direct current. Follow the procedures indicated in the Precautionary Instructions. Do not attempt to use the unit if it is not properly grounded.

OPERATION

The Hydrocollator Heating Unit is simple and easy to operate. The controls are located in the rear, near the bottom of the unit. The thermostat controls the temperature of the
water. It is located near the bottom of the unit on the control panel. It has been tested and preset at the factory to normal operating temperature. If further adjustment is necessary, it can be easily accomplished by turning the adjusting screw located on the thermostat. Turning the outer screw counterclockwise lowers the temperature while turning it clockwise raises it.
NOTE: Do not turn the inner stem under any circumstance! The stem is preset by the manufacturer. The Hydrocollator Heating Units operate through the use of a hydraulic capillary type thermostat and canbe left plugged in and turned on to provide a ready supply of heated HotPacs. The HotPac rack accommodates different sizes and types of packs.
SS-2 Unit – The unit comes complete with eight standard HotPacs.
M-2 and M-4 Units – The M-2 unit comes complete with 12 standard HotPacs. The M-4
comes complete with 24 standard HotPacs.
SS Unit – Holds eight Standard Size packs. The special rack holds two Neck Contour packs.
E-2 Unit – The E-2 unit comes complete with two Oversize, three Standard Size packs and one Neck Contour pack.
E-1 Unit – With the insert in, the rack accommodates four Standard Size packs.

TROUBLESHOOTING

The following are solutions to minor performance problems. Only qualified personnel should service equipment problems.

PROBLEM POSSIBLE CAUSE POSSIBLE SOLUTION
Unit plugged in, water does not get hot 1. Breaker for electric circuit off

2. Thermostat not functioning
3. Heating element burned out| 1. Turn the breaker on
2. Replace the thermostat
3. Replace element
Cloudiness of water| 1. Filler from packs is seeping
2. Too long a period between cleanings| 1. Replace packs
2. Drain and clean
Packs too hot| 1. Thermostat set too high
2. Thermostat failure| 1. Adjust the thermostat
2.Replace the thermostat
Packs too cool| 1. Thermostat set too low
2. Thermostat failure| 1. Adjust the thermostat
2. Replace the thermostat
Unit boils| 1. Thermostat failure| 1. Replace the thermostat
Rust/Discoloration| 1. Contact with ferrous material
2. Deposits on the inside of the tank
3. Too much chlorine in the water| 1. Remove the ferrous metal item from the unit (staple, paper clip, safety pin, tongs, etc.) and clean thoroughly. Add a dechlorinating agent if local water has high chlorine content.
2. Thoroughly clean the tank and remove all deposits.
3. Clean the tank and rinse thoroughly. Add a dechlorinating agent if local water has high chlorine content.

ACCESSORIES

Listed below are the Power Cords for each Country, their part numbers, and their dimensions.

Ref. Description
21284 Euro [18 AWG, 203.2 cm (80 in), Shielded]
78121 US [18 AWG, 203.2 cm (80 in), Shielded]
20971 Australian [18 AWG, 203.2 cm (80 in), Shielded]
20972 Swiss [18 AWG, 203.2 cm (80 in), Shielded]
20973 UK [18 AWG, 203.2 cm (80 in), Shielded]
20974 Danish [18 AWG, 203.2 cm (80 in), Shielded]
20975 Japanese [18 AWG, 203.2 cm (80 in), Shielded]
20976 Indian [18 AWG, 203.2 cm (80 in), Shielded]
20977 Israeli [18 AWG, 203.2 cm (80 in), Shielded]

MAINTENANCE

Care and Cleaning
The Hydrocollator Heating Unit is equipped with an immersion type heating element and a hydraulic capillary type thermostat which evenly maintains the HotPac temperature in the water and provides a ready supply of heated packs. It is critical to maintaining the water level over the top of the HotPac to avoid damage to the heating element, the stainless steel, or the HotPac. Water is constantly lost during operation due to evaporation. Therefore, it is essential that water be added daily. The tank should also be drained, cleaned, and inspected systematically, at minimum intervals of every two weeks. Always unplug the unit from the power source when emptying or cleaning the unit. Drain the unit by the drain valve located at the bottom rear of the unit. Either move the unit to a floor drain or attach the extension hose to it when draining. (E-1 units have no drain and can be emptied by turning the unit upside down).

WARNING

  • DJO, LLC does not recommend the Hydrocollator Heating Unit be moved while filled with heated water. Tipping over could result in a burn to the user or others. If the Hydrocollator Heating Unit must be moved while filled, exercise extreme caution! It is recommended to allow the water to cool below 49° C (120° F) before moving the unit.
  • Always unplug the Hydrocollator Heating Unit from the power source before attempting to empty water from the unit.

The Hydrocollator Heating Unit is made of high-quality stainless steel that, with proper care and maintenance, could easily last a lifetime. Improper care could allow the stainless steel to wear prematurely.
Follow these simple steps to ensure that the Hydrocollator Heating Unit remains in excellent condition:

  1. Excessive chlorine can be detrimental to the stainless steel in the Hydrocollator Heating Unit. No bleach or any cleaner with high chlorine content should be used under any circumstance.

  2. Clean regularly for optimal service and operation.
    NOTE: Failure to maintain the unit (including regular cleaning) will void your warranty.

  3. Fill daily with water. Water is constantly lost during operation due to evaporation.

  4. Chlorine in regular tap water may be present in high enough concentrations to damage the unit. If you suspect high levels of chlorine in the Hydrocollator Heating Unit, the addition of a de-chlorinator is recommended.

  5. Certain additives (such as herbal teas, essential oils, etc.) will damage the HotPacs and components of the heating unit, including the stainless steel. The Hydrocollator Heating Unit was designed to heat water and HotPacs only. The addition of any such foreign substance (except dechlorinator) to the water in the unit will void the warranty.

  6. Inspect the heating element for pitting and other visible damage during regular cleaning intervals.

  7. Inspect the power cord for wear, fraying, and other visible damage during regular cleaning

To avoid potential for rusting:

  1. Do not use bleach or cleaners with high chlorine content.
  2. Do not use paper clips, safety pins, staples, coat hangers, etc., in, on, or near the unit.
  3. Do not clean the unit with carbon steel wool or a carbon steel brush. Use only soft cloth, ScotchBrite™ type scouring pads of green, blue, or white abrasiveness, or stainless steel wool if absolutely necessary.
  4. Do not use abrasive cleaners which tend to disturb the grain in the finish, especially on the exterior of the unit. For exterior cleaning, use only non-abrasive stainless steel polish.
  5. The water level is critical: Add water daily, if necessary, to keep it full.
  6. Regularly clean and drain the tank (every two weeks). Failure to properly maintain the unit will cause premature wear and will void the warranty.
  7. Remove all deposits from interior surfaces and parts. The deposits are concentrations of minerals that will allow rusting to begin. If allowed to accumulate on the heating element, these deposits will lower the efficiency of the element and increase the power consumption of the unit. However, exercise extreme care if cleaning element or thermostat(s). These components can be easily damaged without proper care or caution.
  8. If you suspect that you have “hard water”, you may need to change the water and clean the unit more frequently. Salt deposits are an indication of hard water that will eventually cause rust.

Cleaning Tips

  1. The interior of the unit should be cleaned, at least every two weeks, using a low abrasive bathroom cleaner with a soft cloth or green, blue, or white Scotch-Brite™ type scouring pad. Stainless steel wool may be used if necessary, but not regular, carbon steel wool. Check for low or no chlorine content in your cleaner and make sure that the residue is thoroughly rinsed away with water.
  2. A strong solution of vinegar and water may be used to dissolve away deposits, which then must be thoroughly rinsed away. Results may vary.
  3. For exterior cleaning, use stainless steel polish and a soft cloth only. Numerous industrial stainless steel polishes are available from various supply companies. Use of any cleaner except stainless steel polish on the exterior of the unit is not recommended. Abrasive cleaners and scouring pads can damage the finish of the unit.

Care of HotPacs
Always return the Hydrocollator HotPac to the hot water of the heating unit after each treatment. There it is heated, kept clean, and ready for immediate use. The HotPac may be boiled; however, this hastens the deterioration of the pack. The pack may also be cleaned by scrubbing the pack with soap and water. The simplified method of keeping the pack clean is to keep it immersed in water.
The Hydrocollator HotPac, under constant daily use, should have a useful life of at least one year. When the pack begins to wear out, the filler oozes and leaks through the fabric of the cloth, and sections of the pack appear to be loose and empty and will not retain heat properly. The pack should be replaced when it reaches this state. Original replacement Hydrocollator HotPacs may be ordered from an authorized DJO, LLC dealer.
NOTE: If HotPacs are to be stored for extended periods, they may be placed – while wet – in plastic bags and stored in a freezer.

WARRANTY

DJO, LLC (“Company”) warrants that the Hydrocollator Models SS-2, M-2, SS, E-1, E-2, and M-4 Heating Units (“Products”) are free of defects in material and workmanship. This warranty shall remain in effect for one year (12 months) from the date of the original consumer purchase. If these products fail to function during the one-year warranty period due to a defect in material or workmanship, at the Company’s option, the Company or the selling dealer will repair or replace this Product without charge within a period of thirty days from the date on which the Product is returned to the Company or the dealer.
All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized centers or groups will void this warranty.
To participate in warranty coverage, this Product’s warranty registration card (included with the Product) must be filled out and returned to the Company by the original owner within ten (10) business days of purchase.

This Warranty Does Not Cover:

  • Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a certified Company service agent.
  • Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service agent.
  • Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and necessary maintenance or any use that is inconsistent with the Product User Manual.

THE COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.

Some states do not allow the exclusion or limitation of incidental or consequential damages, so
the above limitation or exclusion may not apply to you.
To obtain service from Company or the selling dealer under this warranty:

  1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to:
    DJO, LLC
    5919 Sea Otter Place,
    Suite 200
    Carlsbad, CA 92010 • USA

  2. The Product must be returned to the Company or the selling dealer by the owner. This warranty gives you specific legal rights and you may also have other rights which vary from state to state. The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product. Any representation or agreement not contained in the warranty shall be void and of no effect.

THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OR  MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

ISO 13485 Certified
DJO, LLC
5919 Sea Otter Place,
Suite 200
Carlsbad, CA 92010 • USA
© 2021 DJO, LLC
22323 REV K 2021/12/03