chattanooga FLU110D Standard Single Extremity Unit User Manual
- June 1, 2024
- chattanooga
Table of Contents
chattanooga FLU110D Standard Single Extremity Unit
FOREWORD
- Thank you for purchasing a Fluidotherapy Dry Heat Therapy Unit.
- This manual contains general safety, operating, maintenance, and care instructions for the owners and operators of the Fluidotherapy Dry Heat Therapy Units.
- The specifications put forth in this manual were in effect at the time of the publication. However, owing to DJO, LLC’s policy of continuous improvement, changes to these specifications may be made at any time without obligation on the part of DJO, LLC.
- Read, understand, and follow the information contained in this manual.
- Stay current with the latest clinical developments in the field of Dry Heat Therapy and observe all applicable precautionary measures for treatment.
- Keep informed on appropriate indications and contraindications for the use of Dry Heat Therapy.
- This equipment is to be used only under the prescription and supervision of a licensed practitioner.
ABOUT DRY HEAT THERAPY
PRECAUTIONARY INSTRUCTIONS
The precautionary instructions found in this section and throughout this
manual are indicated by specific symbols. Understand these symbols and their
definitions before operating this equipment. The definitions of these symbols
are as follows:
- CAUTION – Text with a “CAUTION” indicator will explain possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
- WARNING – Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
- DANGER – Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
- EXPLOSION HAZARD – Text with an “Explosion Hazard” indicator will explain possible safety infractions if this equipment is used in the presence of flammable anesthetics.
- NON-IONIZING ELECTROMAGNETIC RADIATION – Text with a Non-Ionizing Electromagnetic Radiation” indicator informs the user of possible hazards resulting from elevated, potentially dangerous, levels of non-ionizing radiation.
- Protective Earth (ground)
NOTE: Helpful information marked as “NOTE” may be found throughout this manual. Each “NOTE” is beneficial in the aid of a particular area or function being described for this product.
DESCRIPTION OF DEVICE MARKINGS
- Degree of Protection Against Electrical Shock ………………………………………………..
- Complies with UL 2601-1 CSA C22.2 No. 601.1 Medical Electrical Equipment ………………………………………………………
- Refer to Instruction Manual/Booklet ………………………………………..
CAUTION
-
Read, understand, and practice the precautionary and operating instructions found in this manual. Know the limitations and hazards associated with using any electrical device. Observe the precautionary and operational decals placed on the unit.
-
DO NOT operate this unit in an environment where other devices
are being used that intentionally radiate electromagnetic energy in an unshielded manner. Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. -
Portable and Mobile RF Communications equipment can affect Medical Electrical Equipment.
-
This unit generates, uses, and can radiate radio frequency energy and, if not installed and used by the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this unit on and off. Try to correct the interference using one or more of the following: reorient or relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet on a different circuit from that which the other device(s) are connected and consult the DJO, LLC Service Department for help.
-
Clean Inlet Filter(s) daily before unit startup.
-
Turn unit OFF before positioning a patient or removing a patient from the unit.
-
Place the patient in a comfortable position allowing for the correct placement of the limb being treated.
-
Secure all entry ports before turning the unit ON.
-
Use only Cellex® Dry Heat Medium in the Fluidotherapy units.
-
Refill unit daily to proper fill level with Chattanooga Cellex Dry Heat Medium.
-
Use only fingers to operate button controls on the control panel(s). The use of sharp objects such as pencils or pens will result in damage to the unit.
-
Check unit temperature before treating the patient to ensure the correct temperature.
-
This unit should be transported and stored in temperatures between -40 °F and 158 °F (-40 °C and -70 °C) with Relative Humidity between 10% – 100% to prevent damage to the unit or its components.
-
This unit should be operated in temperatures between 110 °F and 125 °F (43 °C and 52 °C) to prevent damage to the unit or its components.
-
DO NOT operate the unit when connected to any unit other than Chattanooga devices.
-
Change Cellex Dry Heat Medium every six (6) months.
-
Do not use accessories other than those supplied with the unit, or recommended by DJO, LLC. The safety of other products has not been established, and their use could result in injury to the patient.
WARNING
- Federal law restricts this device to sale by, or on the order of, a physician or licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner.
- Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the applicable national and local electrical codes.
- Before administering any treatment to a patient, you should become acquainted with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings, and precautions. Consult other resources for additional information regarding the application of Dry Heat Therapy.
- The Fluidotherapy Dry Heat Therapy Unit should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the Fluidotherapy Dry Heat Therapy Unit should be observed to verify normal operation in the configuration in which it will be used.
- For continued protection against fire hazards, replace fuses only with ones of the same type and rating.
- This device should be kept away from children.
- To prevent electrical shock, disconnect the unit from the power source before attempting any maintenance procedures.
- Use only processed dry heat medium in the unit such as Cellex to prevent excessive dusting.
- Use only accessories that are specially designed for this unit. Do not use accessories manufactured by other companies on this unit. DJO, LLC is not responsible for any consequence resulting from using products manufactured by other companies. The use of other accessories or cables may result in increased emissions or decreased immunity of this unit.
- Dispose of all products under local and national regulations and codes.
- Fluid therapy equipment not in use should be protected against unqualified use.
- The Fluidotherapy Dry Heat Therapy Unit is not suitable in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
- Perform all Preventive Maintenance as described in this manual. Failure to perform the Preventive Maintenance could result in the Cellex medium entering the heat chamber of the unit and cause severe injury to patients as well as smoke damage to the facility and the Fluidotherapy unit.
- The solvents of adhesives and flammable solutions used for cleaning and disinfecting the unit should be allowed to evaporate before the unit is used.
Indications & Contraindications
Indications
- Relief of local pain
- Increase of local blood circulation
- Treatment in range of motion when combined with exercise
- Treatment for minor pain and stiffness of non-rheumatoid arthritis
Contraindications
- This device should not be used for symptomatic pain relief unless etiology is established or unless a pain syndrome has been diagnosed.
- This device should not be used when cancerous lesions or open wounds are present in the treatment area.
- Other counterindications are patients suspected of carrying serious infectious diseases or diseases where it is advisable, for general medical purposes, to suppress heat or fevers.
WARNING
- Adequate precautions should be taken when treating individuals with suspected or diagnosed medical conditions or diseases such as heart problems, epilepsy, diabetes, etc.
- Prior to treatment, consult a medical professional familiar with the precautionary measures to be taken for patients who may experience allergic reactions to dust and pollen.
NOMENCLATURE
FLU110D & 110DE Unit Familiarization
-
RESERVOIR LID & TREATMENT VIEWING WINDOW Access to add/change medium, view patient treatment, and clinician access to the patient treatment area
-
MEDIUM/TREATMENT RESERVOIR
Treatment access and medium reservoir -
HAND/ARM ACCESS PORT
Treatment port for Hand/Arm Treatment (end of unit) -
REPLACEABLE INLET FILTERS
Filters room air entering the Fluidotherapy unit (1 on each side) -
FIVE POINT BASE
Casters included to convert to a Mobile Base -
BLOWER HOUSING
Houses Blower & Heater -
CONTROL PANEL
See page 8 for a detailed description of the controls -
ELBOW/FOOT TREATMENT ACCESS PORT
Treatment port for Elbow and Foot
FLU115D & FLU115DE Unit Familiarization
-
ELBOW/FOOT ACCESS PORT
Treatment port for Elbow and Foot Treatment, access to add/change medium, clinician access to patient treatment area -
HAND/ARM ACCESS PORTS
Treatment ports for Hand/Arm Treatment (end of unit) -
INLET FILTERS
Filters room air entering the Fluidotherapy unit (1 on each side) -
MEDIUM/TREATMENT TUB
Treatment access and medium reservoir -
MOBILE BASE
-
LOCKING CASTERS
-
BLOWER HOUSING
Houses Blower & Heater -
CONTROL PANEL
See page 8 for a detailed description of controls
Operating Controls (All Units)
-
DISPLAY
Displays Treatment Time, Temperature, Airspeed, Pulse Time,
Preheat the Timer, and Clock when the respective Mode is indicated -
PREHEAT TIMER ON
Indicator light for Preheat Timer -
INCREASE/HOUR
Use to increase Mode parameters and to set the Hour when setting Clock -
DECREASE/MINUTE
Use to decrease Mode parameters and use to set the Minutes when setting Clock -
STOP
Use to stop treatment and Preheat Timer -
MOTOR FUSE
-
START
Use to start treatment -
INDICATOR BAR
Indicates Mode as they are chosen -
HEATER FUSE
-
PREHEAT TIMER BUTTON
Press once to turn the Preheat Timer On. Press twice to start the Preheat function. -
MODE SELECT
Use to select the desired Mode -
PM
PM indicator for Clock
SPECIFICATIONS
FLU110D & FLU110DE
MODES OF OPERATION
Continuous (Default)
Variable Adjustments for Time, Temp, and Air Speed
- Pulse Mode……………………………………………………………OFF to 6 Sec ON/OFF
- TREATMENT TIME………………………………………………………………1 to 99 minutes
- OPERATING TEMPERATURE………………………….110 °F (43.3 °C) to 125 °F (51.6 °C)
- AIR SPEED…………………………………………………….0% to 100% (5% increments)
- PREHEAT TIMER………………………………….115 °F (46.1 °C) with 50% Air Flow
- MEDIUM CAPACITY…………………………………..Approximately 30 lb (13.6 kg)
- INPUT POWER (FLU110D)……………………………………….120V, 50/60 Hz, 12A
- INPUT POWER (FLU110DE)……………………………….230-240V, 50/60 Hz, 8A
- PHYSICAL DIMENSIONS
- Cabinet Length………………………………………………………………..34.0” (86.4 cm)
- Cabinet Width………………………………………………………………….11.5” (29.2 cm)
- Height……………………………………………………………………………………31.5” (80 cm)
- Weight………………………………………………………….77 lbs (35 kg) Less Medium
- Shipping Weight………………………………………………………………121 lbs (55 kg)
- DEGREE OF PROTECTION AGAINST INGRESS OF WATER……………IPX0
- ELECTRICAL CLASS………………………………………………………………………….CLASS I
- DEGREE OF PROTECTION AGAINST ELECTRICAL SHOCK……TYPE B
- POWER CORD……………………………….. 2-meter cable length, 14 AWG, shielded
WARNING
Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the applicable national and local electrical codes.
FLU115D & FLU115DE
MODES OF OPERATION
Continuous (Default)
Variable Adjustments for Time, Temp, and Air Speed
- Pulse Mode……………………………………………………………OFF to 6 Sec ON/OFF
- TREATMENT TIME………………………………………………………………1 to 99 minutes
- OPERATING TEMPERATURE…………………………110 °F (43.3 °C) to 125 °F (51.6 °C)
- AIR SPEED…………………………………………………….0% to 100% (5% increments)
- PREHEAT TIMER………………………………….115 °F (46.1 °C) with 50% Air Flow
- MEDIUM CAPACITY…………………………………Approximately 40 lbs (18.1 kg)
- INPUT POWER (FLU115D)………………………………………..120V, 50/60 Hz, 12A
- INPUT POWER (FLU115DE)……………………………..230-240V, 50/60 Hz, 10A
- PHYSICAL DIMENSIONS
- Cabinet Length………………………………………………………………..34.0” (86.4 cm)
- Cabinet Width…………………………………………………………………..18.5” (47.0 cm)
- Height…………………………………………………………………………………33.0” (83.8 cm)
- Weight……………………………………………………….60 lbs (27.2kg) Less Medium
- Shipping Weight…………………………………………………………..100 lbs (45.4 kg)
- DEGREE OF PROTECTION AGAINST INGRESS OF WATER. …………..IPX0
- ELECTRICAL CLASS………………………………………………………………………….CLASS I
- DEGREE OF PROTECTION AGAINST ELECTRICAL SHOCK………….TYPE B
- POWER CORD……………………………….. 2-meter cable length, 14 AWG, shielded
WARNING
Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the applicable national and local electrical codes.
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES
TABLE 1: GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Fluidotherapy-Dry-Dry-Dry-Dry – Dry Heat Therapy Unit is intended for use in the electromagnetic environment specified in the table below. The user of the Fluidotherapy – Dry Heat Therapy Unit should ensure that it is used in such an environment.
Emission Tests| Compliance| Electromagnetic Environment
– Guidance
---|---|---
RF Emissions CISPR 11| Group 1| The Fluidotherapy – Dry Heat Therapy Unit uses
RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
RF Emissions CISPR 11| Class A|
The Fluidotherapy-Dry-Dry-Dry-Dry – Dry Heat Therapy Unit is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations IEC 61000-3-3| Complies
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
TABLE 2 & 3: GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
IMMUNITY
The Fluidotherapy-Dry-Dry-Dry-Dry-Dry-Dry – Dry Heat Therapy Unit is intended
for use in the electromagnetic environment specified in the table below. The
user of the Fluid therapy – The Dry Heat Therapy Unit should ensure that it is
used in such an environment.
Immunity Test| IEC 60601
Test Level
| Compliance Level| Electromagnetic Environment –
Guidance
---|---|---|---
| | | Portable and mobile RF communications equipment should be used no
closer to any part of the Fluidotherapy-Dry-Dry-Dry-Dry – Dry Heat Therapy
Unit, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
| | | Recommended separation distance:
Conducted RF IEC 61000-4-6| 3 Vrms
150 kHz to 80 MHz
| 3 V| d = [3.5] √ P
V 1
Radiated RF IEC 61000-4-3
|
3 V/m
80 MHz to 2.5 GHz
|
3 V/m
| d = [3.5]__ √ P 80 MHz to 800 MHz
E 1
d = [7]__ √ P 800 MHz to 2.5 GHz
E 1
| | | where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). a
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey , should be less than the compliance level in each
frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcasts and TV broadcasts cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Fluidotherapy – Dry Heat
Therapy Unit is used exceeds the applicable RF compliance level above, the
Fluidotherapy – Dry Heat Therapy Unit should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Fluidotherapy-Dry-Dry-Dry –
Dry Heat Therapy Unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
TABLE 2 & 3: GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
IMMUNITY
The Fluidotherapy-Dry-Dry – Dry Heat Therapy Unit is intended for use in the
electromagnetic environment specified in the table below. The user of the
Fluidotherapy – The Dry Heat Therapy Unit should ensure that it is used in
such an environment.
Immunity Test IEC 60601 Compliance Level Electromagnetic Environment – Guidance Test Level
Electrostatic discharge (ESD) IEC 61000-4-2| ±6 kV contact
±8 kV air
| ±6 kV contact
±8 kV air
| Floors should be wood, concrete, or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines
±1 kV for input/output lines
| ±2 kV for power supply lines
±1 kV for input/output lines
| Mains power quality should be that of a typical commercial or hospital
environment.
Surge (1)
IEC 61000-4-5
| ±1 kV differential mode
±2 kV common mode
| ±1 kV differential mode
±2 kV common mode
| Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| <5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
| <5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
| Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Fluidotherapy – Dry Heat Therapy Unit requires
continued operation during power mains interruptions, it is recommended that
the Fluidotherapy – Dry Heat Therapy Unit be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60Hz) magnetic field
IEC 61000-4-8
| 3 A/m| 3 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage before application of the test level.
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
TABLE 4: RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE FLUIDOTHERAPY-DRY – HEAT THERAPY UNIT
(CONTINUED)
The Fluidotherapy-Dry-Dry-Dry – Dry Heat Therapy Unit is intended for use in
an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Fluidotherapy – Dry Heat Therapy
Unit can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the Fluidotherapy – Dry Heat Therapy Unit as recommended
below.
Rated Maximum Output Power of Transmitter
W
| Separation Distance According to Frequency of Transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = [3.5] √ P d = [3.5] √ P _d = [ 7 ]_√ P
V1 _E1 E1_
---|---
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
SETUP
TREATMENT MODE PARAMETERS
ATTENTION!
ATTRITION OF PARTICLES DURING SHIPMENT MAY RESULT IN INITIAL DUSTING AND SMALL
PARTICLE RELEASE. DUST AND ODOR WILL DISSIPATE WITH USE.
Use the following instructions to set the various Mode Parameters to the desired settings.
TREATMENT TIME
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“TREATMENT TIME”.
NOTE:
The default time of “20:00” will be displayed.
Press the “INCREASE” or “DECREASE” buttons until the desired Treatment Time is
displayed.
NOTE:
The “TREATMENT TIME” can be adjusted in one-minute increments from 1 to 99.
TREATMENT TEMPERATURE
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“TEMPERATURE”.
NOTE:
The default temperature of “115 °F (46.1 °C) will be displayed when the unit
is first plugged into the wall outlet.
The “F” or “C” and “TEMPERATURE” indicators will flash while the programmed
Treatment Temperature is being displayed.
When the Bed Temperature is displayed, the “F” or “C” and the “TEMPERATURE”
indicator will illuminate steadily.
TREATMENT TEMPERATURE (CONTINUED)
Press the “INCREASE” or “DECREASE” buttons until the desired Treatment
Temperature is displayed.
NOTE:
°F can be changed to °C and vice versa as desired. See page 20 for
instructions. Treatment Temperature can be adjusted in 1° increments from 88
°F to 130 °F (31.1 °C to 54 °C).
AIR SPEED
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“AIR SPEED”.
NOTE:
The default speed of “50” will be displayed.
Press the “INCREASE” or “DECREASE” buttons until the desired Air Speed is
displayed.
NOTE:
The “AIR SPEED” is adjusted in increments of 5 from 0 to 100.
PULSE TIME
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“PULSE TIME”.
NOTE:
The “PULSE TIME” allows the unit to operate by pulsing the medium during
treatment. If selected and set, the medium will pulse on and off in equal
increments. The Factory Default is “OFF”.
EXAMPLE: “PULSE TIME” is set at “5”. The unit will pulse the medium
during treatment, five seconds on and five seconds off.
PULSE TIME (CONTINUED)
Press the “INCREASE” or “DECREASE” buttons until the desired Pulse Time is
displayed.
NOTE:
“PULSE TIME” is adjusted in one-second increments from “OFF” to 6 seconds.
STOP TREATMENT
Press the “STOP” button.
TIME CONTROLLED PARAMETERS
PREHEAT TIMER
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“PREHEAT TIMER”.
NOTE:
The “PREHEAT TIMER” allows the unit to preheat the medium. The unit will run
at 50% Air Speed until the unit reaches the default preheat temperature or 90
minutes, whichever comes first. The “PREHEAT TIMER” will operate Monday
through Friday only and can be set to automatically start at a predetermined
time and heat the medium to a predetermined temperature. See page 21 for
setting the default parameters of the “PREHEAT TIMER.
Press the “PRESS ONCE TO TURN ON PREHEAT TIMER” button to illuminate the
“PREHEAT TIMER ON” indicator light. Press again to start the “PREHEAT TIMER”.
NOTE:
The “PREHEAT TIMER” indicator light must be set at the end of each day in
order for it to automatically come on the next day. The “STOP” button is used
to turn off the Preheat if desired.
CLOCK
Press the “PRESS TO CHANGE MODE” button until the indicator light is beside
“CLOCK”.
NOTE:
The “PREHEAT TIMER” will not function until the Clock is set.
Press the “INCREASE” button to set hours.
Press the “DECREASE” button to set minutes.
NOTE:
The “PM” indicator light will illuminate when PM hours are reached.
PREFERENCE MODE DEFAULT PARAMETERS
ENTERING “PrEF” MODE (PREFERENCE MODE)
Simultaneously press the “PRESS TO CHANGE MODE”, “INCREASE”, and “DECREASE”
buttons. “PrEF” will be displayed.
NOTE:
The “PrEF” mode allows the user to change the default settings.
TREATMENT TIME DEFAULT
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing
beside “TREATMENT TIME”.
Press the “INCREASE” and “DECREASE” buttons to set unit to the desired default
treatment time.
Press the “PRESS TO CHANGE MODE” button to save the new setting.
NOTE:
The default setting is 20:00 minutes.
TEMPERATURE READING DEFAULT
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing
beside “TEMPERATURE”.
Press the “INCREASE” button to set unit to display °F or °C as desired for the
default setting.
Press the “PRESS TO CHANGE MODE” button to save the new setting.
NOTE:
The default setting is °F.
TREATMENT TEMPERATURE DEFAULT
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing
beside “TEMPERATURE” and the existing default temperature is displayed.
- Press the “INCREASE” and “DECREASE” buttons to set unit to the desired default treatment temperature.
- Operating Temperature Range is: 110 °F – 125 °F (43.3 °C – 51.6 °C)
- Press the “PRESS TO CHANGE MODE” button to save the new setting.
PULSE MODE (Enable/Disable)
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing
beside “PULSE TIME”.
“EnAb” (enable) or “dISA” (disable) will display.
- Press the “INCREASE” button to set unit to display “EnAb” or “dISA” as desired.
- Press the “PRESS TO CHANGE MODE” button to save the new setting.
PREHEAT TIMER MODE (Enable/ Disable)
-
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing beside “PREHEAT TIMER”.
-
“EnAb” (enable) or “dISA” (disable) will display.
-
Press the “INCREASE” button to set unit to display “EnAb” or “dISA” as desired. Press the “PRESS TO CHANGE MODE” button to save the new setting.
NOTE: -
If Clock is disabled, the “PREHEAT TIMER” will automatically be disabled.
CLOCK (Enable/Disable)
Press the “PRESS TO CHANGE MODE” button until the indicator light is flashing
beside “CLOCK”.
“EnAb” (enable) or “dISA” (disable) will display.
Press the “INCREASE” button to set unit to display “EnAb” or “dISA” as
desired. Press the “PRESS TO CHANGE MODE” button to save the new setting.
NOTE:
If the Clock is disabled, the “PREHEAT TIMER” will automatically be disabled.
SETTING DAY OF WEEK
Press the “PRESS TO CHANGE MODE” button until “dAy” (= day code) is
displayed.
DAY CODE CHART
WEEKDAY| DAY CODE
Sunday| dAyu
Monday| dAyn
Tuesday| dAyE
Wednesday| dAyd
Thursday| dAyr
Friday| dAyF
Saturday| dAyA
- Press the “INCREASE” or “DECREASE” button to set unit to display the desired day code.
- Press the “PRESS TO CHANGE MODE” button to save the new setting.
EXITING “PrEF” MODE
Press the “PRESS TO CHANGE MODE” buttons until “PrEF” is displayed.
Simultaneously press the “INCREASE”, “DECREASE”, and “STOP” buttons. The
default treatment time will display and the indicator will be beside
“TREATMENT TIME”.
OPERATION
PATIENT PREPARATION
DANGER
- Clean the patient treatment area thoroughly with an antimicrobial soap and clean water as per industry, facility, regulatory standards, and universal skin washing procedures.
- Following the skin washing procedure, apply a hospital grade antiseptic skin cleanser, according to the cleanser manufacturer’s recommended instructions for use.
- Reference the maintenance schedule and make certain the preventive maintenance has been performed prior to starting the unit.
STARTING TREATMENT
STARTING
Make certain all control panel settings have been made, unit is preheated, and
patient is in proper position with sleeve(s) secure prior to starting the
treatment. Refer to pages 15 – 17 for setup of the unit.
Press “START” button to begin treatment.
STOPPING TREATMENT
STOPPING
Treatment will automatically stop when the treatment time reaches zero. Should
it be desired to stop treatment before the timer reaches zero, push the “STOP”
button.
NOTE:
Should the treatment be stopped before the timer reaches zero, it will be
necessary to re-set the “TREATMENT TIME” in order to complete the prescribed
treatment.
PREVENTIVE MAINTENANCE
DAILY MAINTENANCE
WARNING
- Before any maintenance is performed or attempted, unplug the unit from the power source to prevent the possibility of electrical shock
CLEAN INLET FILTERS
At the end of each work day, unplug the unit and clean the Inlet Filters on
the unit. Carefully remove the filter retainer and wash the filter and screen
with a mild antibacterial soap and water. Thoroughly dry the filter and screen
before placing back on the unit.
NOTE
Should your unit have the earlier style inlet filters, gently clean the filter
using a soft bristle brush. Be careful not to puncture or damage the filter.
Should the filter become damaged, torn, or punctured, call your dealer for
replacement of the filter before resuming operation.
REFILL WITH CELLEX MEDIUM
Refill the unit with Cellex Dry Heat Medium to approximately one (1) inch
above the bottom of the arm treatment ports of the unit.
NOTE:
Use of other than Cellex Medium may cause premature failure of the
Fluidotherapy unit(s).
INSPECT SLEEVE CONDITION
Inspect the port sleeves for tears, rips, and weak seams. Replace all sleeves
that show signs of tears, rips, weak, or loose seams, or excessive wear.
Keeping the sleeves in excellent condition prevents excessive spillage of the
Cellex medium and prevents the medium from entering the heat chamber of the
unit.
ARM PREHEAT TIMER
Plug the unit back into an approved power source. If the “Preheat Timer” is
used, press the “PREHEAT TIMER” button once and the “PREHEAT TIMER ON”
indicator will illuminate.
NOTE:
The “PREHEAT TIMER ON” indicator must be illuminated in order for the Preheat
Timer to automatically come on at the default time the next morning.
The Fluidotherapy Preheat Timer does not operate on Saturday or Sunday.
However, for Monday morning preheat, set the unit as described above on Friday
evening.
WEEKLY MAINTENANCE
Each week all sleeves of the Fluidotherapy unit should be laundered in a mild
antibacterial detergent. Allow the sleeves to air dry or dry on a low
temperature setting. Drying the sleeves in high temperatures could cause the
sleeves to shrink or become distorted, resulting in the sleeve(s) not properly
fitting when placing them back onto the unit.
REMOVING TOP SLEEVES (All Models)
- Unplug the unit from the power source.
- Carefully remove the rectangular bezel(s) from the top of the unit.
- Carefully pull the hook and loop fasteners apart to remove the sleeve from the unit.
- Replace in reverse order making certain the corners are well seated to form the seal required to prevent excessive escape of medium.
REMOVING END SLEEVE (Model FLU110D & FLU110DE)
- Carefully pull the hook and loop fastener on the sleeve to remove sleeve from unit.
- Replace in reverse order making certain the corners are well seated to form the seal required to prevent excessive escape of medium.
REMOVING END SLEEVES (Model FLU115D & FLU115DE)
NOTE:
The sleeve in these photos has been modified (cut away) for clarity.
Remove the top sleeves to gain access to the inside of the unit. Refer.
Pull the medium back and away from the arm ports.
NOTE
It is extremely important to move the medium away from the ports to prevent
the medium entering the space between the inner and outer walls of the unit
when the sleeves are removed and replaced.
-
Dust away all Cellex medium from inside the sleeve.
-
Pull the retainer ring from the sleeve and lay aside for re-installation.
-
Pull the sleeve away from the sleeve extrusion in the inner housing of the unit.
NOTE: -
There is a ring folded into the sleeve that fits over the sleeve extrusion on the inside of the unit.
DANGER
-
Cellex medium entering the heat chamber of the unit can cause severe injury to patients as well as smoke damage to the facility and the Fluidotherapy unit.
-
Remove the ring from the folded end of the sleeve.
-
Launder sleeves as described on page 26.
-
Vacuum out any of the Cellex medium that may have entered the cavity between the inner and outer walls of the unit.
WARNING
Removal of all Cellex medium that has entered the cavity is essential to maintain proper and safe operation of the unit. If the medium reaches the heating chamber of the unit, immediately call a certified technician for removal of any suspected medium in the heat chamber. Do not place the unit into service if medium is suspected to be in the heat chamber.
INSTALLING 115 Model SLEEVE
Fold the elastic end of the sleeve over the ring.
Place sleeve through port hole. Place the ring/sleeve assembly over the sleeve
extrusion on the inside of the unit.
NOTE:
Position the sleeve seam toward the top of the unit and push the ring/sleeve
assembly completely onto the sleeve extrusion.
Install the inner sleeve retainer ring making certain the ring split will be
toward the top of the unit.
Push the retainer ring over the ring/sleeve assembly until it is completely
seated.
NOTE:
The retainer ring should fit tightly over the sleeve. If necessary, use a pair
of pliers to tighten the ring in the area where it slides over the sleeve.
From the outside of the unit, reach through the sleeve and install the outer
sleeve retaining ring.
NOTE:
Start installation of the outer ring at the top of the unit. As the ring is
being seated, pull the area of the sleeve tight. This will aid in the
prevention of excessive medium spillage and patient comfort.
ADDITIONAL FLUIDOTHERAPY PREVENTIVE MAINTENANCE REQUIREMENTS
The following additional preventive maintenance requirements must be scheduled
and performed as described to ensure that the unit is operating efficiently,
safely, and functioning at optimum level. A blank Maintenance Record is
provided on page 31 to aid in the scheduling and record keeping of this
prescribed preventive maintenance program. The following preventive
maintenance procedures must be performed by a DJO, LLC qualified service
technician trained in the maintenance requirements of the Chattanooga
Fluidotherapy units.
QUARTERLY (Every 3 Months)
The following Preventive Maintenance must be performed on all
Fluidotherapy units quarterly by a certified service technician.
- INTERNAL CAVITY INSPECTION AND CLEANING
- FULL FUNCTIONAL AND PERFORMANCE TESTS
BI-ANNUAL (Every 6 Months)
The following Preventive Maintenance must be performed on all
Fluidotherapy units every six months in addition to the Quarterly
Maintenance requirements by a certified service technician.
- CHANGE CELLEX MEDIUM
ANNUAL (Once per Year)
The following Preventive Maintenance must be performed on all
Fluidotherapy units annually in addition to the Quarterly and Bi-Annual
Maintenance requirements by a certified service technician.
- CALIBRATION
AS NEEDED
The following Preventive Maintenance must be performed only if performance
test results indicate replacement is necessary.
- INTAKE FILTER REPLACEMENT
- DISTRIBUTOR REPLACEMENT
- BLOWER MOTOR(S) REPLACEMENT
DANGER
- If the level of Cellex medium suddenly drops one inch or more below the operating level of the unit, immediately pull the unit out of service and contact a certified service technician.
- A sudden level drop in the Cellex medium may indicate that medium is leaking into the heating chamber of the unit and must be repaired before the unit is placed back into service.
FLUIDOTHERAPY MAINTENANCE RECORD
- UNIT SERIAL NUMBER…………………………UNIT MODEL NUMBER…………………………………DATE PLACED IN SERVICE………………………………………….
- DEALER…………………………………PHONE…………………………………….CONTACT………………………………………………………………
DATE | MAINTENANCE PERFORMED | TECH INITIALS |
---|---|---|
DATE | MAINTENANCE PERFORMED | TECH INITIALS |
--- | --- | --- |
CLEANING THE FLUIDOTHERAPY UNIT
After each use, clean the unit and its accessories using a soft, clean cloth
dampened with water and a mild antibacterial detergent.
Avoid the use of abrasive materials and cleaning solvents.
SERVICE
When the Fluidotherapy Dry Heat Therapy unit requires service or preventive
maintenance, contact the selling dealer or DJO, LLC Service Department.
All units returned to the factory for service must include the following:
WARRANTY REPAIR/OUT OF WARRANTY REPAIR
- Written statement containing the following information:
- RA Number- Obtain from Factory
- Unit Model Number
- Unit Serial Number
- Contact person with Phone and Fax Numbers
- Billing Address (for Out of Warranty Repair)
- Shipping Address (Where to Ship Unit after Repair)
- Detailed Description of Problem or Symptoms
- Copy of original invoice issued at purchase of the unit.
- Ship unit to Factory in the original container with all accessories and information as required in item 1 above to:
- DJO, LLC
- Chattanooga Repair Center
- 47492 SD Hwy 22
- PO Box 709
- Clear Lake, SD 57226 USA
- chattgroup.com
- Service to these units should be performed only by service technicians certified by DJO, LLC.
- Council Directive 2002/96/EC concerning Waste Electrical and Electronic Equipment (WEEE). Indicates a requirement not to dispose of WEEE as municipal waste. Contact your local distributor for information regarding disposal of the unit and accessories.
ACCESSORIES
REPLACEMENT ACCESSORIES
When ordering additional accessories for Chattanooga Fluidotherapy units, use
the following part numbers and descriptions.
HENLEY No. CGI No.
MODEL FLU110D & FLU110DE
- TOP SLEEVE SLE0003 31775
- TOP SLEEVE BEZEL FRA0003 31456
- ARM SLEEVE SLE0002 31774
MODEL FLU115D & FLU115DE
- TOP SLEEVE SLE0003 31775
- TOP SLEEVE BEZEL FRA0003 31456
- INNER SLEEVE RETAINER TRI0034 31883
- SLEEVE RING PPL0013 31679
- OUTER SLEEVE RETAINER TRI0035 31834
- ARM SLEEVE SLE0001 31773
CELLEX Dry Heat Medium 10 lbs (4.5 kg) MED0001
MEDIUM
Use only Chattanooga’s Cellex Dry Heat Medium in the Fluidotherapy unit. The
Cellex medium is designed specifically for use in the Chattanooga
Fluidotherapy units to ensure optimal and efficient operation of all the
Fluidotherapy products.
WARRANTY
- DJO (“Company”) warrants that the Fluidotherapy Dry Heat Therapy units (“Product”) are free of defects in material and workmanship. This warranty shall remain in effect for one year (12 months) from the date of original consumer purchase. If this Product fails to function during the one year warranty period due to a defect in material or workmanship, at the Company’s option, the Company or the selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product
- is returned to the Company or the dealer.
- All repairs to the Product must be performed by a service center certified by the Company. Any modifications or repairs performed by unauthorized centers or groups will void this warranty.
- The warranty period for accessories is 90 days. Accessories consist of sleeves and use replaceable intake filter(s).
- The warranty period for the motor(s) and distributor is 180 days.
- To participate in warranty coverage, this Product’s warranty registration card (included with Product) must be filled out and returned to the Company by the original owner within ten (10) business days of purchase.
This Warranty Does Not Cover:
- Replacement parts or labor furnished by anyone other than the Company, the selling dealer, or a certified Company service technician.
- Defects or damage caused by labor furnished by someone other than Company, the selling dealer, or a certified Company service technician.
- Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and preventive maintenance or any use that is inconsistent with the Product User’s Manual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
Some states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation or exclusion may not apply to
you.
To Obtain Service From Company or the selling dealer under this warranty:
-
A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to:
DJO, LLC
5919 Sea Otter Place, Suite 200 Vista, CA 92081 USA and -
The Product must be returned to the Company or the selling dealer by the owner.
This warranty gives you specific legal rights and you may also have other
rights which vary from state to state.
The Company does not authorize any person or representative to create for it
any other obligation or liability in connection with the sale of the Product.
Any representation or agreement not contained in the warranty shall be void
and of no effect.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE.
CONTACT
- DJO, LLC | 5919 Sea Otter Place, Suite 200 | Carlsbad | CA 92010 | U.S.A.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>