MICHELL Instruments MDM300 I.S. Advanced Dew Point Hygrometer Instruction Manual
- June 5, 2024
- MICHELL Instruments
Table of Contents
MDM300 I.S.
Advanced Dew-Point Hygrometer
Sensor & Battery Replacement Manual
97469 Issue 1
October 2014
Please fill out the form(s) below for each instrument that has been purchased.
Use this information when contacting Michell Instruments for service purposes.
Instrument|
---|---
Code|
Serial Number|
Invoice Date|
Location of Instrument|
Tag No|
MDM300 I.S.
For Michell Instruments’ contact information please go to
www.michell.com
© 2014 Michell Instruments
This document is the property of Michell Instruments Ltd. and may not be
copied or otherwise reproduced, communicated in any way to third parties, nor
stored in any Data
Processing System without the express written authorization of Michell
Instruments Ltd.
Safety
The manufacturer has designed this equipment to be safe when operated using
the procedures detailed in this manual. The user must not use this equipment
for any other purpose than that stated. Do not apply values greater than the
maximum value stated.
This manual contains operating and safety instructions, which must be followed
to ensure the safe operation and to maintain the equipment in a safe
condition. The safety instructions are either warnings or cautions issued to
protect the user and the equipment from injury or damage. Use qualified
personnel and good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and
accessories supplied by the manufacturer for use with the instrument. The
instrument is powered by an internally mounted rechargeable battery. The input
power supply voltage limits for the battery charger supplied with the
instrument are 100 to 240 V AC, 47/63Hz.
Caution: No other battery charger unit, other than that supplied with the
instrument should be used.
Pressure Safety
High-pressure gases can be extremely dangerous and only trained personnel
should attempt to connect and use the MDM300 I.S. with such gases. DO NOT
permit pressures greater than the safe working pressure to be applied to the
instrument. The specified safe working pressure (SWP), for this instrument, is
350 bar (5000 psi).
Toxic Materials
The use of hazardous materials in the construction of this instrument has been
minimized. During normal operation, it is not possible for the user to come
into contact with any hazardous substance which might be employed in the
construction of the instrument. Care should, however, be exercised during
maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained either by the manufacturer or an accredited
service agent. Refer to www.michell.com for details
of Michell Instruments’ worldwide office’s contact information.
Calibration
The recommended calibration interval for this instrument is 12 months. The
instrument should be returned to the manufacturer, Michell Instruments Ltd.,
or one of their accredited service agents for re-calibration. Alternatively, a
freshly calibrated exchange sensor can be purchased and easily fitted, see
Section 1.2.3.
Safety Conformity
This product carries the CE mark and meets the requirements of all the
relevant European and U.S. safety directives.
Abbreviations
The following abbreviations are used in this manual:
AC | alternating current |
---|---|
bar | pressure unit (=100 kP or 0.987 atm) (gauge) |
Hz | Hertz |
psig | pound(s) per square inch (gauge) |
V | Volts |
Warnings
The following general warnings listed below are applicable to this instrument.
They are repeated in the text in the appropriate locations.
Where this hazard warning symbol appears in the following sections, it is used
to indicate areas where potentially hazardous operations need to be carried
out.
MAINTENANCE
The MDM300 I.S. requires very little maintenance. The only user-replaceable parts are the battery, the internal sensor, and the top hat filter located behind the inlet orifice fitting. The following sections detail the routine and corrective maintenance procedures.
1.1 Routine Maintenance
The only routine maintenance required is to periodically clean the casing,
display, and keyboard of the instrument with a damp cloth and a mild
detergent. The required frequency of cleaning will depend upon instrument
usage and whether or not it is used in a carrying case.
Do not use acetone or any other type of solvent as this could damage the casing, display and keyboard.
1.2 Replacement of Sensor and Battery Pack
To gain access to the internal sensor and the battery housing, proceed as
follows (refer to Figure 1).
1.2.1 Open Casing
When the casing is open, the circuit board, which contains electrostatically
sensitive devices, is exposed. Take appropriate precautions i.e. wear an
earthing wristband, in order to prevent possible damage.
To open the casing, proceed as follows:
-
Switch the instrument OFF.
-
Disconnect any connections to the gas ports.
-
Place a sheet of card onto a firm surface in order to protect the instrument’s display and casing from possible scratching and place the instrument face down.
-
Using an Allen key, sequentially remove the five cap screws (1) and washers (2) (see Figure 1).
-
Open the two halves of the casing.
NOTE: There is a seal between the two halves so the top and bottom halves may need to be teased apart. There should also be an O-ring (see Figure 2, (5)) on each of the five pillars – keep these safe for later use. -
Place the two halves of the casing (3) and (4) side by side (refer to Figure 2).
1.2.2 Close Casing
To close the casing, proceed as follows:
- Place the two halves of the casing together, checking that no cables are trapped between the two halves and that there is an O-ring on each of the five pillars.
- Initially replace each of the cap screws and washers without fully tightening them.
- Tighten up each cap screw. NOTE: Do not overtighten as this could cause the case to crack.
1.2.3 Internal Sensor Removal and Replacement
Sensor removal
To remove the sensor proceed as follows:
- Open the instrument case as described in Section 1.2.1.
- Remove the connectors, (3), (4), and (5) from the back of the sensor (see Figure 3). NOTE: Each connector is locked in position by a clamp situated at the back of the connector. Before attempting to pull the connector out, push this clamp away from the connector in order to release it.
- Use a spanner to unscrew the locknuts (1) securing the sensor to the gas inlet and outlet ports.
- Lift the sensor (2) clear.
Sensor replacement
To replace the sensor, proceed as follows:
- Place the replacement sensor back into the gas connections, ensuring that it is re-installed in the same orientation as previously, and locate the olives in the gas ports.
- Tighten each of the locknuts (1) finger tight.
- Tighten each union an additional ¾ of a turn.
- Re-fit the connectors (3), (4), and (5). The connectors are polarized and cannot be inserted incorrectly. Ensure that the connectors lock into position.
- Close the casing as described in Section 1.2.2.
- Switch on the instrument and check that it is operating satisfactorily.
1.2.4 Battery Replacement
Battery Replacement
To replace the battery, proceed as follows:
- Open the instrument case as described in Section 1.2.1.
- Loosen the 3 battery clamp retaining screws (2).
- Release the battery (1) at connector (3) and remove.
- Re-connect new battery to connector (3).
- Replace and tighten battery clamp retaining screws (2).
- Close the casing as described in Section 1.2.2.
- Switch on the instrument and check that it is operating satisfactorily.
- Charge the battery, when necessary, as described in the User’s Manual.
NOTE: No other parts of the MDM300 I.S. are user-serviceable – please
contact your Michell representative if further help is needed.
Refer to www.michell.com for details of Michell
Instruments’ worldwide office’s contact information.
Appendix A
Quality, Recycling & Warranty Information
A.1 Pressure Equipment Directive (PED) 97/23/EC
The above Directive has been implemented in United Kingdom Law by the Pressure
Equipment Regulations 1999.
The Regulations require that all pressure equipment and assemblies within the
scope of the Pressure
Equipment Directive must be safe when placed on the market or put into
service.
Michell Instruments’ products have been assessed and, as referenced against
the Classification Charts detailed in Annex II of the Directive, do not fall
into the requirements for CE marking and comply with the Pressure Equipment
Directive.
Article 3, paragraph 3 states that any product containing a pressurized fluid
that does not qualify for compliance should, nevertheless, be constructed with
Sound Engineering Practice (SEP).
Michell Instruments attests here that its products have been designed,
manufactured & tested to assure safe operation, and in accordance with Sound
Engineering Practices.
A.2 Recycling Policy
Michell Instruments is concerned with the protection of the environment. It is
our commitment to reduce and eliminate from our operations, wherever possible,
the use of substances that may be harmful to the environment. Similarly, we
are increasingly using recyclable and/or recycled material in our business and
products wherever it is practical to do so.
To protect natural resources and to promote material reuse, please separate
batteries from other types of waste and recycle responsibly. If batteries are
not properly disposed of, these substances can cause harm to human health and
the environment
The product that you have purchased may contain recyclable and/or recycled
parts and we will be happy to provide you with information on these components
if required. For further information please see the following sections.
A.3 WEEE Compliance
Directive 2012/19/EU 4 July 2012 on Waste Electronic and Electrical
Equipment (WEEE)
The Waste Electronic and Electrical Equipment (WEEE) Directive places rules
upon European manufacturers of electrical and electronic equipment. The
directives’ aim is to reduce the impact that electronic devices have on the
environment.
Michell Instruments is in full compliance with the WEEE Directive and is
registered with an approved recycler (Registration No. WEE/JB0235YW) and
treats the requirement of the directive and the protection of the environment
with the utmost importance. All Michell Instruments’ products are
appropriately marked indicating their requirement for recycling. It may be
required to return certain instruments for treatment at the end of their
working life.
A.4 RoHS2 Compliance
Directive 2011/65/EU of the European Parliament and of the Council of 8 June
2011
The Restriction of Hazardous Substances (RoHS) Directive places rules upon
European manufacturers of electrical and electronic equipment. The directives’
aim is to reduce the impact that electronic devices have on the environment.
According to the EC Directive 2002/95/EC, Michell Instruments’ products
qualify as Category 9, Control and Monitoring Equipment. Under the 2002/95/EC
Directive, Category 9 products are exempt from compliance with the Directive.
However, the careful design of all Michell Instruments’ products takes into
consideration the requirements of the Directive and, wherever possible,
compliance is achieved. All future products will be developed entirely using
compliant materials. Furthermore, Michell Instruments is taking active steps
to remove non-compliant materials and components from existing products
wherever these may occur. Presently, none of the non-compliant materials are
known to occur in Michell Instruments’ products.
The new Directive 2011/65/EU (RoHS2) entered into force on 21 July 2011 and
required all Member States to transpose the provisions into their respective
national laws by 2 January 2013. The provisions of the RoHS2 EU Directive
2011/65/EU (Article 3, [24]) define ‘Control and Monitoring Equipment’
specifically as ‘monitoring and control instruments designed exclusively for
industrial or professional use’.
RoHS2 EU Directive 2011/65/EU states the closing date for compliance of any
Control and Monitoring Equipment product sold into the EU marketplace as 22nd
July 2017.
However, the careful design policy of all Michell Instruments’ products
continues to attain compliance in the shortest practical timescales and
strives to ensure that less than 0.1% of total mass per product, of all non-
compliant materials, appear within them. Michell Instruments continues to
monitor suppliers and material sources to ensure that compliance of goods
provided is maintained.
A.5 Warranty
Unless otherwise agreed, the Supplier warrants that, as from the date of
delivery for a period of 12 months, the goods and all their component parts,
where applicable, are free from any defects in design, workmanship,
construction, or materials.
The Supplier warrants that the services are undertaken shall be performed
using reasonable skill and care, and be of a quality conforming to generally
accepted industry standards and practices. Except as expressly stated, all
warranties whether express or implied, by operation of law or otherwise, are
hereby excluded in relation to the goods and services to be provided by the
Supplier. All warranty services are provided on a return to base basis. Any
transportation costs for the return of a warranty claim shall reside with the
Customer.
A.6 REACH Compliance
Regulation (EC) No. 1907/2006
Registration, Evaluation, Authorisation, and Restriction of Chemicals (REACH)
Michell Instruments is a manufacturer of moisture measurement and gas analysis
instrumentation and is a ‘downstream’ user of chemicals, as described by the
EU Council Directive 76/769/EEC. The products we supply are not raw chemical
products (goods). Under normal and reasonably foreseeable circumstances of
application, the goods supplied to you shall not contain or release any
prohibited chemicals. No listed SVHC (Substances of Very High Concern) appear
within products manufactured by Michell Instruments. Therefore the 0.1% mass
per product, or total usage of 1 tonne/year, will never be exceeded. For these
reasons we are neither required by obligation for registration nor for the
creation of material safety data sheets (MSDS) for our products. Our continued
review of the SVHC Candidate List and latest additions is to ensure we remain
compliant.
Michell Instruments maintains a hazardous material register in which MSDS data
sheets are collated, and we will check that our suppliers will comply to REACH
requirements for all materials and substances we use in the processes of our
manufacturing. In the unlikely event that any chemicals of concern appear in
our products in quantities greater than 0.1% of total mass per product, we
will immediately inform you by correspondence according to the REACH Article
33 requirements. Our current appraisal is, however, that we do not expect or
foresee such an incidence.
A.7 Calibration Facilities
Michell Instruments’ calibration facilities are among the most sophisticated
in the world and have been recognized for their excellence.
Traceability to the National Physical Laboratory (NPL) the UK is achieved
through our UKAS Accreditation (Number 0179). This covers dew points over the
range -90 to +90°C (-130 to +194°F) and also Relative Humidity.
Dew-point calibrations are also traceable to the National Institute for
Standards & Technology (NIST) the USA over the range of -75 to +20°C (-103 to
+68°F).
NOTE: Standard traceable calibration certificates for instruments and
sensors are not issued under our UKAS accreditation. UKAS certificates are
usually to special order and
are clearly identified.
A.8 Return Policy
If a Michell Instruments’ product malfunctions within the warranty period, the
following procedure must be completed:
- Notify a Michell Instruments’ distributor, giving full details of the problem, the model variant and the serial number of the product.
- If the nature of the problem indicates the need for factory service then the instrument should be returned to Michell Instruments, carriage prepaid, preferably in the original packaging, with a full description of the fault and the customer contact information.
- Upon receipt, Michell Instruments will evaluate the product to determine the cause of the malfunction. Then, one of the following courses of action will be taken:
• If the fault is covered under the terms of the warranty, the instrument will be repaired at no cost to the owner and returned.
• If Michell Instruments determines that the fault is not covered under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs, at standard rates, will be provided. Upon receipt of the owner’s approval to proceed, the product will be repaired and returned.
A.9 Manufacturing Quality
Michell Instruments is registered with the British Standards Institute for
Quality Assurance to:
BS EN ISO 9001: 2008
Rigorous procedures are performed at every stage of production to ensure that
the materials of construction, manufacturing, calibration, and final test
procedures meet the requirements laid down by our BSI-approved Quality System.
Please contact Michell Instruments (www.michell.com)
if the product does not arrive in perfect working order.
Appendix B
Return Document & Decontamination Declaration
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>