FNIRSi GC-01 Nuclear Radiation Detector User Manual
- June 3, 2024
- FNIRSI
Table of Contents
Nuclear Radiation
Detector User Manual
CATALOG
Notice to user
-
Please read this instruction manual and operation instructions carefully, Follow the instructions in the manual, In order to make the detector function fully.
-
Please keep this manual
-
Don’t use this equipment in a flammable and explosive environment.
-
Replaced used batteries and discarded instruments cannot be
disposed of with household waste. Please handle according to relevant national or local laws. -
When there are any quality problems with the instrument or questions about using the instrument. You can contact “FNIRSI” online customer service or the manufacturer. We will solve it for you as soon as possible
Product Description
This product uses a Geiger-Miller counter. Counter for detecting the intensity
of ionizing radiation (beta particles, gamma rays and x-rays).Use a gas tube
or a small chamber as a probe. When the voltage applied to the probe reaches a
certain range. Each time the ray is ionized in the tube to produce a pair of
ions, it can be amplified to produce an electric pulse of the same size. And
recorded by the connected electronic device. The number of rays per unit time
thus measured. The alarm
threshold measurement rate can be arbitrarily selected.
Key features
- Detect x ray,γ rays and β rays.
- High sensitivity, can be used in various environments.
- Data is saved during shutdown.
- High-definition LCD display. The status display is clear at a glance.
- Light/Vibration/Sound 3 combined alarm modes to choose.
- Real-time clock display.
- The product can preset dose rate and cumulative dose alarm
threshold.
Product parameters
Product name | Nuclear radiation detector |
---|---|
Size | 120x78x27mm |
Types of detection rays | γ rays, x rays, β rays |
Detector | Energy Compensation GM Tube (Geiger Counter Meter) |
dose equivalent rate | 0.00-1000μSv/h(10mSv/h) |
Cumulative dose equivalent | 0.00μSv-500.0mSv |
Energy range | 48keV-1.5Mev≤±30% (for 137Cs-) |
Language | Chinese/English switch |
Sensitivity | 80CPM/μSv For Co-60) |
Dosage unit | μSv/h,μGy/h,mR/ h,cps, cpm Switch |
Power supply | 1100mAh lithium battery |
Alarm method | light, vibration, sound |
The button description
- Left/Back key: Return to the upper menu from the lower menu
- Right key/setting key: Enter the setting menu/enter the lower menu
- Power on/off key/OK key: Switch on/Off/Confirm
- Up key: Switch options up and down
- Down key: Switch options up and down
How to operate
Power on/of
Short press the power button to turn it on. Long press the power button to shut down.
Monitoring interface
Automatically enter this page after booting, and monitor parameters:
- Real-time detection amount, displayed in the upper left panel
- Average value/Maximum value
- Current dose alarm value
- Cumulative dose alarm value
- Cumulative stored dose rate
Press the left and right keys to switch to the waveform detection page as
follows:
●Waveform monitoring
●Current dose rate real-time value
●Maximum value
●Minimum
Settings
Long press the right key/setting key to enter the setting menu. Long press the left button/back button on the setting page to return to the monitoring page. Press the up and down keys to switch the setting options.
Setting Options:
- Unit settings
- Alarm settings
- System clock
- Alarm mode
- Display settings
Unit setting
Press the right button to enter the lower level to set five measurement units:
- μSv/h
- μGy/h
- mR/h
- CPS
- CPM
Alarm settings
Long press the right key/setting key to enter the setting menu. Press the up and down keys to switch the setting options. Press the right button to enter the lower settings to set or change the values of the following options:
- Current dose alarm value
- Cumulative dose alarm value
- The accumulated dose is reset to zero
System Clock
Long press the right key/setting key to enter the setting menu. Press the up and down keys to switch the setting options. Press the right button to enter the lower level settings to set the date and time.
Alarm mode
Long press the right key/setting key to enter the setting menu. Press the up and down keys to switch the setting options. Press the right button to enter the lower level settings. On or off:
- Indicator
- Vibration
- Sound
Display Settings
Long press the right key/setting key to enter the setting menu. Press the
up and down keys to switch the setting options. Press the right button to
enter the lower settings:
●Screen brightness adjustment
●Switch between Chinese/English
Conversion of radioactive units
International Standards (1990)
Radioactive staff: 20mSv/year (10μSv/hour)
General public: 1mSv/year (0.52μSv/hour)
Unit conversion
1μSv/h=100μR/h 1nc/kg.h=4μR/h
1μR=1γ(The unit used for prospecting in the pronuclear industry)
Radioactivity:
1Ci=1000mCi
1mCi=1000μci
1Ci=3.7×1010Bq =37GBq
1mCi=3.7×107Bq =37MBq
1μCi=3.7×104Bq=37KBq
1Bq=2.703×10-11Ci=27.03pci
Metering equivalent:
1Sv=103mSv=106μSv
1Sv=100rem 100μrem=1μSv
Exposure:
1R=103mR=106μR
1R=2.58×10-4c/kg
Absorption metering:
1Gy=103mGy=106μGy
1Gy=100rad 100μrad=1μGy
Radon unit:
1Bq/L=0.27em=0.27×10-10Ci/L
Other:
1Sv is equivalent to 1Gy 1g radium=0.97Ci≈1Ci
Calculation of radioisotope decay values
A=A0eλ-t t=T1/2 ;
A0 The known source strength A is how much time has elapsed, According
to the radioactive decay calculation table look-up table calculation.
The relationship between radioactive source and distance
The intensity of the radioactive source is inversely proportional to
the square of the distance.
X=A.г/R2A: The activity of the point source; R: Distance from source;
г: Exposure rate constant
Note: Ra-226 (t 1608) г=0.825 ren. m2/hour. Curie
Cs—137 (t 29.9 years) г= 0.33 ren. m2/hour. Curie
Co—60 (t 5.23 years) г=1.32 ren. m2/hour. Curie
According to the radioactive decay calculation table, look up the table to calculate the radioactive shielding:
Halved and reduced to 1/10 value (cm) for different substances
Radioactive source|
Pencil
| Iron|
Concrete
| Halving| 1/10| Halving| 1/10| Halving|
1/10
Cesium-137|
0.65
| 2.2| 1.6| 5.4| 4.9|
16.3
Iridium-192|
0.55
| 1.9| 1.3| 4.3| 4.3|
14.0
Cobalt-60|
1.10
| 4.0| 2.0| 6.7| 6.3|
20.3
NOTE
Nuclear radiation detectors are sophisticated instruments. Please be careful. The following recommendations will facilitate instrument maintenance and prolong life.
- Keep as dry as possible during storage and use. Excessive humidity can cause malfunction and damage to the instrument.
- Please don’t use the instrument violently or rudely, prevent dropping, knocking and violent vibration of the instrument. Otherwise, the instrument will be damaged.
- When the power display is too low, it is in an under voltage state and should be charged in time. In case of serious under voltage, the instrument can not be turned on and off, and abnormal phenomena such as blurred screen occur.
※If the instrument cannot work normally, please contact our company after sales. We will solve the problem
Instrument maintenance
- Please keep it dry and wipe off the dirt on the surface of the instrument with a soft cloth before use. Don’t use detergents or solvents
- Please recycle and use damaged instruments, accessories and packaging materials in an environmentally friendly manner.
- Please shut down in time when not in use for a long time
- Don’t disassemble or replace components without permission to avoid failure.
- Please store in a dry place when not in use.
Production information
Product Name: Nuclear Radiation Detector
Brand/Model: FNIRSI/GC-01
Service phone: 0755-83242477
Manufacturer: Shenzhen FRI NI RUI SI Technology Co., Ltd.
URL:www.fnirsi.cn
Factory address: 8th Floor, West of Building C, Weida Industrial Park,
Dalang Street, Longhua District, Shenzhen City, Guangdong Province
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>