COXO CX235 Contra Angle User Manual
- June 1, 2024
- coxo
Table of Contents
COXO CX235 Contra Angle
Specifications
- Model: CX235
- Ratio: 1:5
- Product Composition: Inner channel contra angle speed increase 1:5, Optic fiber, Push button
Product Usage Instructions
Cartridge Replacement
- Use a screwdriver to remove the screw from the tail of the contra angle and pull out the small head from the outside.
- Use the unloading tool to open the rear cover, remove the shaft, and take out the cartridge set. Replace with a new cartridge.
- After replacing the cartridge, use the loading tool to tighten the back cover and shaft. Ensure proper insertion into the fuselage.
Contra Angle and Motor Connections
- Straightly insert the motor into the connection of contra angle until you hear a cracking sound.
- To separate, hold the bending machine and pull out the motor.
- Align the groove to insert the contra angle into the air motor.
Lubrication
After daily use, spray lubricant for 1-2 seconds by inserting the lubricant
nozzle fully into the drive air tube. For heavy use, lubricate at noon and
afternoon after daily use. Lubricate the head, gear, and inside. When
lubricating the head, open the latch chuck. Remove contra handpiece from
motor, install E spray nozzle on the lubricant, insert E spray nozzle on the
back of contra handpiece to spray 2-3 seconds until clean lubrication comes
out. Keep contra handpiece standing for a while before reconnecting with the
motor.
FAQ
-
Q: How often should I lubricate the contra angle?
A: After daily use, lubricate for 1-2 seconds. For heavy use, lubricate at noon and afternoon after daily use. -
Q: How do I replace the cartridge?
A: Use a screwdriver to remove the screw from the tail of the contra angle, open the rear cover, replace the cartridge set with a new one, and tighten the back cover and shaft. -
Q: How do I connect the contra angle to the motor?
A: Straightly insert the motor into the connection of contra angle until you hear a cracking sound. To separate, hold the bending machine and pull out the motor.
CONTRA ANGLE
MODEL:CX235
Please read this operation Manual carefully and file for future reference.
https://www.alandental.com
Preface
Thanks for purchasing our dental Contra angle!
In order to fully exert the function of this equipment, safely and effectively
use this equipment, please read this instruction carefully before using , and
please keep this manual well for reference anytime.
Under regulations’request, we have been committed to the optimization of our
product, this may cause some parts of change, when you find the product
appears some difference from description in the manual, please contact the
dealers or our after-sales center for details .
Technical specifications
- Protection of liquids-normal liquids – protected equipment
- Temperatures range:-40°C–+135°C
- Relative humidity:10% to 80%,non-condensing
- Noise:≤70dB
WARNING
- Please read this operation manual carefully before use and file it for future reference .
- This product is spcialized for abnormal stomatology treatment,don’t use it in other ways.
- This product is only for doctors.
- Every time you use it,please check first.If find anything abnormal,stop using immediately,and contact your supplier or our after-sales center for details.
- Before operating low speed handpiece,pull contra angle handpiece with hand to make sure handpiece connect firmly and safely.
- Head can not running without insert bur,otherwise head or motor would be damaged.
- Ensure working pressure before running,Air pressure for four bole air motor is 0.3Mpa,for two hole air motor is 0.25Mpa.
- Don’t immerse handpiece in any chemical solvent or solution or by dry-heat disinfection.Sterilization by autoclave at(135±2)°C&0.22MPa is sugguested.
Product Composition
Inner channel contra angle speed increase 1:5
Technical parameters
- Inner channel contra angle low speed handpiece ;
- Burs applicable:for CA burs Φ1.6;
- Max.speed of contra angle:200,000rpm
- Speed: speed up to 5times
- Chuck speed:push button
- Spray:4 way spray
Cartridge Replacement
- Use a screwdriver to make the screw out from the tail of the contra angle and pull out small head from outside, as shown in figure 1-1.
Technical parameters
- Inner channel contra angle low speed handpiece ;
- Burs applicable:for CA burs Φ1.6;
- Max.speed of contra angle:200,000rpm
- Speed: speed up to 5times
- Chuck speed:push button
- Spray:4 way spray
Cartridge Replacement
- Use a screwdriver to make the screw out from the tail of the contra angle and pull out small head from outside, as shown in figure 1-1.
- Use the unloading tool to open the rear cover, and make shaft out from small head .as shown in figure 1-3.
- Take the cartridge and cartridge set out ,replace the new cartridge. as shown in figure 1-4.
- After Replacing the new cartridge ,use the loading tool to tighten the back cover and shaft ,You must aim for insert of the fuselage, as shown in figure 1-4, tighten screw of the tail part
Contra angle and motor connections
- Straightly insert the motor into the connection of contra angle, until you hear a “cracking” sound can be.
- Hold the bending machine and pull out the motor, it will be separated[fig.3.2].
- Alignment the groove,contra angle can insert into the air motor[fig.3.3].
Lubrication
-
After daily use,spray lubricant for 1 to 2 seconds by inserting lubricant nozzle all the way into drive air tube.
-
If work heavily,lubrication should be implemented at noon and afternoon’s after using everyday.
-
Lubricate the head,gear and inside.[Fig.16] When lubricate the head,please open the latch chuck.
-
Dismount contra handpiece from motor.
-
Install E spray nozzle on the lubricant.
-
Insert E spray nozzle on the back of contra handpiece to spray2-3 seconds till the lubricant through out of head,Stop until clean lubrication comes out.
-
Keep contra handpiece standing for a while,and then connect with the motor after the lubricant inside come out.
Caution
Be sure to hold the handpiece firmly to prevent the handpiece from slipping
out by the spray pressure when lubricating. Supply lubricant until it comes
out of the handpiece nose
(for approx 2 seconds)
Keep the SPRAY can upright
Clean-head system
- Scrub dirt and debris from the handpiece.
- Stop the handpiece and wipe it dry.If the dirt could not remove from the hole, remove the cartridge and clean by brush.
Notice!
Use only clean water to maintain the clean head system.
Cleaning -bulb
Wipe clean the Cellular optical bulb entry point with an alcohol-immersed
cotton swab.Remove all debris and oil.
Sterilization and cleaning
- Steam autoclave is recommended.
- Sterilization requirement after using as noted below.
- Autoclave Procedure:
- Scrub dirt and debris from the handpiece,and wipe clean with alcoholic cotton.Do not use a wire brush(Fig.18.4)
- Lubricate with SPRAY or Lubricant oil.
- Lubrication.
- Insert into an autoclave pouch.Seal the pouch.
- Autoclavable up to a max.135°C(EX.)Autoclave for 20 min. at 121°C or 15 min.at 132°C(Fig.18.2/18.3).
NOTE:After treatment or before autoclave,please do the lubrication.
NOTE:After treatment or before autoclave,please do the lubrication .
Trouble Shooting
Trouble | Possible cause | Solution |
---|---|---|
Body of straight head or contra angle rotates during motor running | O-ring on | |
the nose of motor wears | Change o-ring | |
The handpiece fail to rotate | Contra angle chucking with bur but lacth plate | |
fail to be fixed in the proper position. dirt get into the handpiece | Clean |
and lubricate, move it by hand
Our service center can offer technical assistance to you.
Warranty
COXO grants the user a 12 months guarantee for its complete product range,except ball bearing(3 months guarantee) from the date of invoice issued, Maintenance over the term of guarantee will be at the customer’s charge. COXO will not be responsible for damage or injury resulting from:
- excessive use
- improper manipulation of the product,or modification to product carried out by persons not authorizes by COXO.
- fail to follow the instruction to install,operate and maintenance the handpiece.
- damage of chemical,electrical or electrolysis due to improper autoclaving and storing.
- improper working pressure.
The technical specifications and dimension in these instruction are given only as a guide. They may not be subject of any claim.
Standard Symbols
On the instrument label appear standard symbols as follows:
Transportation and deposited condition
Environmental temperature range:-40°C–+50°C;Relative humidiy≤80%;The scope of
air pressure:500hPa—1060hPa.
Foshan COXO medical instrument Co.,Ltd
Address:21 Wufeng Si Road,Foshan,Guangdong,China
Wellkang Ltd.
Address:SuiteB,29 Harely Street,London W1G9QR,United Kingdom
https://www.alandental.com
References
- Home of the (M)obil (I)nformation (N)etwork
- Dental Instruments Australia |Dental Equipment Supplies For Sale - Alandental.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>