Royce Water Technologies TU 8525 Turbidity Electronics Instructions
- June 1, 2024
- Royce Water Technologies
Table of Contents
Royce Water Technologies TU 8525 Turbidity Electronics
Product Information
Technical Specifications
- Scalability factor: 10/100% 4/20 mA
- Sensitivity: 70/130%
- Zero Resolution: 0.001 FTU
- Power supply: 9-36 VDC
- Accuracy: 0.2% of the full-scale selected
- Repeatability: 0.1%
- Non-linearity: 0.1%
- Check signal: 0/200.0%
- Load: 600 Ohm max. at 24 Vdc
- Digital output: RS 485 isolated
- Current loop: 4/20 mA isolated
- Protocols: B&C ASCII and Modbus RTU (03, 06, 16 functions)
- Baud rate: 2400 / 4800 / 9600 / 19200 baud
- Probes ID: 01/99 (B&C protocol), 01/243 (Modbus protocol)
- Probes network: 32 max.
- Operating temperature:
- TU 8525: L=143 mm, D=40 mm
- TU 8325: L=165 mm, D=60 mm
- Weight:
- TU 8525 Body: 160 g, cable 640 g
- TU 8325 Body: 420 g, cable 640 g
- Body Material: PVC-C (TU 8525.5 model in PVDF is available)
- Cable: 10 m (100 m max.), PVC sheath
- Protection: IP 68
Product Usage Instructions
Installation
- Choose a suitable location for the sensor installation.
- Ensure the power supply matches the required specifications.
- Connect the sensor following the provided wiring diagram.
- Secure the sensor in place using appropriate mounting hardware.
Calibration
- Power on the sensor and allow it to stabilize.
- Access the calibration menu as per the user manual.
- Follow the steps to calibrate sensitivity, zero resolution, and other parameters.
- Verify calibration results with known reference samples.
Maintenance
Regularly inspect the sensor for any signs of damage or contamination.
Clean the sensor as recommended in the manual to ensure accurate readings.
Frequently Asked Questions (FAQ)
-
What is the maximum cable length supported by the sensor?
The sensor supports a cable length of up to 100 meters. -
How do I change the communication protocol of the sensor?
Refer to the user manual for detailed instructions on changing the communication protocol of the sensor.
INTRODUCTION
These unique sensors have been designed to measure Turbidity based on nephelometric method (ISO 7027 – EN 27027). The sensors are available for submersible and in-pipe installations.
The measuring system consists of:
- Infrared light source
- 90 degree scattered light detector
- Detector of the clean lens status
- 2-wire 4/20 mA analog output
- Modbus Output
- Nozzle for the autoclean by external pressured air (TU 8325)
Principle of operation
The Turbidity follows the nephelometric method (ISO 7027 – EN 27027). A light
beam is sent to the sample through an optical lens. The 90 degree scattered
light by suspended particle is collected by the sensor through a second lens
and it is converted in an electric signal proportional to the Turbidity of the
sample. The probe uses an infrared light and the measuring is not affected by
the color of the sample.
Accessories for TU 8555 / TU 8525
TU 910 Overflow cell
Technical Specifications
- Turbidity ranges 0/4.000 – 0/40.00 – 0/400.0 NTU
- Scalability factor 4/20 mA 10/100 %
- Sensitivity 70/130 %
- Zero ± 0.400 NTU
- Resolution 0.001 FTU
- Power supply 9/36 VDC
- Accuracy 0.2 % of the full-scale selected
- Repeatability 0.1%
- Non-linearity 0.1 %
- Check signal 0/200.0 %
- Load 600 Ohm max. a 24 Vdc
- Digital output RS 485 isolated
- Temperature limit 0/50 °C
- Dual filter software 2/220 seconds
- Current loop 4/20 mA isolated
- Protocols B&C ASCII e Modbus RTU (03, 06, 16 functions)
- Baud rate 2400 / 4800 / 9600 / 19200 baud
- Probes ID
- 01/99 (B&C protocol)
- 01/243 (Modbus protocol)
- Probes network 32 max.
- Operating temperature 60 °C max.
- Operating pressure 6 bar at 25 °C (TU 8525) 1 bar at 25 °C (TU 8325)
- Dimensions TU 8525 L=143 mm, D=40 mm
- Dimensions TU 8325 L=165 mm, D=60 mm
- Weight TU 8525 Body 160 g, cable 640 g
- Weight TU 8325 Body 420 g, cable 640 g
- Body PVC-C (TU 8525.5 model in PVDF is available)
- Cable 10 m (100 m max.), PVC sheath
- Protection IP 68