ThermoFisher SCIENTIFIC 88870008 Touch Screen Dry Bath and Block Heater User Manual
- June 8, 2024
- ThermoFisher SCIENTIFIC
Table of Contents
- ThermoFisher SCIENTIFIC 88870008 Touch Screen Dry Bath and Block Heater
- Preface
- Models
- Safety Information
- Introduction
- Features
- Regulatory Compliance
- Product Safety
- FCC Statement (USA)
- Preparation Work
- Structure Interface
- Operation Guidance
- Files saving interface
- Maintenance
- Failure Analysis and Troubleshooting
- Modular Block Accessories
- Warranty
- FAQ Section
- WEEE Compliance
- Documents / Resources
ThermoFisher SCIENTIFIC 88870008 Touch Screen Dry Bath and Block Heater
IMPORTANT! Before using this product, read this entire operation manual
carefully. Users should follow all of the operational guidelines contained in
this manual and take all necessary safety precautions while using this
product. Failure to follow these guidelines could result in potentially
irreparable bodily harm and/or property damage. Caution all internal
adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for information purposes only. Thermo Fisher
Scientific is committed to a continuing program of product development and
improvement, and reserves the right to change information, such as
specifications, appearance, and dimensions, described in this document without
notice. Thermo Fisher Scientific makes no representations or warranties with
respect to this manual. In no event shall Thermo Fisher Scientific be held
liable for any damages, direct or incidental, arising out of or related to the
use of this manual. No part of this manual may be reproduced or transmitted in
any form or by any means, including photocopying, recording, or using
information storage and retrieval systems, for any purpose other than the
purchaser’s own use, without the express written permission of the
manufacturer. Any other product names and services identified in this manual
are trademarks or registered trademarks of their respective owners. No such
use, or the use of any trade name, is intended to convey endorsement or other
affiliation with Thermo Fisher Scientific.
©2021 Thermo Fisher Scientific. All rights reserved.
All rights reserved. Without prior written consent from Thermo Fisher
Scientific, any part of the manual should not be duplicated, reproduced or
translated into any other languages.
Caution: All rights reserved. Thermo Fisher Scientific reserves the right
to modify this manual at any time without notice.
Before the initial use of this instrument, please carefully read this
manual.
Preface
Our Sales Support staff can provide information on pricing and give you
quotations.We can take your order and provide delivery information on major
equipment items or make arrangements to have your local sales representative
contact you. Our products are listed on the Internet and we can be contacted
through our Internet home page.
Our Service Support staff can supply technical information about proper setup,
operation or troubleshooting of your equipment.We can fill your needs for
spare or replacement parts or provide you with on-site service. We can also
provide you with a quotation on our Extended Warranty for your Thermo
Scientific products. Whatever Thermo Scientific products you need or use, we
will be happy to discuss your applications. If you are experiencing technical
problems, working together, we will help you locate the problem and, chances
are, correct it yourself over the telephone without a service call. When more
extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If
your service need is covered by the warranty, we will arrange for the unit to
be repaired at our expense and to your satisfaction. Regardless of your needs,
our professional telephone technicians are available to assist you Monday
through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. International customers, please contact your local Thermo
Scientific distributor.
Models
This manual covers the models shown below.
Table 1. Models
NA Model | EU Model | Voltage | Description |
---|---|---|---|
88870007 | N/A | 100-120 Vac | Drybath Stdrd 1 blck 100-120 V |
88870008 | N/A | 100-120 Vac | Drybath Stdrd 2 blck 100-120 V |
88870009 | N/A | 100-120 Vac | Drybath Stdrd 4 blck 100-120 V |
N/A | 88870010 | 200-240 Vac | Drybath Stdrd 1 blck 200-240 V |
N/A | 88870011 | 200-240 Vac | Drybath Stdrd 2 blck 200-240 V |
N/A | 88870012 | 200-240 Vac | Drybath Stdrd 4 blck 200-240 V |
Safety Information
Alert Signals
WARNING: The warning message requires extremely careful operation of a
certain step. Failure to observe the instruction may result in serious
personal injury.
CAUTION: Important information is contained in any item and should be
carefully read. Failure to observe the instruction would result in damage or
abnormal function of the instrument
Note: Notes alert you to pertinent facts and conditions.
WARNING: If the Thermo Scientific Touch Screen Drybath is used in a
manner not specified by the manufacturer, the protection provided by the unit
may be impaired.
During operation, maintenance and repair of this instrument, the following
basic safety notes should be observed. In case of failure to follow these
instructions, the warnings or notes indicated here in, the basic protection
provided by the instrument, its safety criteria of design and manufacture, and
its predicted use range would be impaired. If the equipment is used in a
manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
Thermo Fisher Scientific is not responsible for any injury as a result of the
user’s failure to observe the following requirements.
For your own safety, user should observe the following safety warning signs.
The warning signs indicate potential sources of danger. They also inform on
how to correct action can avert danger.
CAUTION: This instrument is for indoor use.
-
The ground connection
To avoid the electric shock, the input power line must be reliably grounded. The instrument is equipped with the three-pin plug that has the third pin (the pin connects the ground), therefore, the plug should be used with the grounded power socket only. This is a kind of safety device. If the plug cannot be plugged into the power socket, please ask the electrician to install a correct power socket, so as to make the grounded plug to work for safety. -
Observe voltage/power requirements Before the instrument is connected with the AC power source, the voltage of the power source should be the same with the required voltage of the instrument (a deviation of 10%is allowed). e rated load of the power socket should not be lower than the requirements of the instrument.
-
The power cord
The instrument should use the power line cord attached to it. If the power line is destroyed, it must be replaced but not be repaired. The replacement should be carried out with the power line of the same type and same specification (reach out to technical service team for further information). No items should be put on the power cord when the instrument is in operation. -
Connecting and disconnecting from power source
The user should hold the plug to remove from power source. When connecting the plug, user should make sure it has been fully plugged in; when removing the plug, do not pull the power line forcefully. Do not position the equipment so that it is difficult to operate the disconnecting device (plug). -
Placement of the instrument
This instrument should be fixed in a low RH and low dust place away from water source (e.g. sink or water pipeline) and the room should be well ventilated, and free of corrosive gas or interference of strong magnetic field. The instrument should not be placed in a wet or dusty location.
The openings on this instrument are for ventilation circulation and in order to avoid over-heat of the instrument body they should not be blocked or covered. When a single set of instrument is used, the interval between ventilation opening before and after the instrument and its nearest object should not be less than 25 cm. Also, don’t use the instrument on loose or soft surface, or the air inlet of instrument bottom might be blocked. Excessive temperature will impair performance or result in failure of instrument. This instrument should not be used in location subjected to direct sun light. The instrument should be kept away from hot gas, oven and all other heat sources. If the instrument is to be stored for a long time, the power plug should be withdrawn and the instrument covered with soft cloth or plastic film to avoid entrance of dust. The product is powered by connecting the mains plug to a standard socket-outlet. Always place the product in such a way that it is easy for the operator to disconnect the product from the mains supply. -
Explanation of symbols
Attention, read user manual before use. -
WARNING: There is a sign of
“CAUTION: HOT SURFACE! ATTENTION: SURFACE HOT” on the instrument. The metal part (module) near the sign should not be touched with any part of the body when the instrument is operating in a high temperature state or just finished operation to avoid burns! -
Alternating Current
-
Protective Conductor Terminal
-
CAUTION: In any of the following cases, immediately turn off the power supply, withdraw the power plug from the power socket, contact the supplier for service:
-
Liquid drops into the inside of the instrument.
-
The instrument is rained on or water is spilled on it.
-
The instrument works abnormally, especially if generating an abnormal sound or odor.
-
The instrument is dropped or its casing is damaged.
-
The function of instrument obviously changes.
Introduction
CAUTION: After unpacking, immediately check the goods against the packing
list. In case of damaged or missed goods, please immediately contact the
distributor in which the product was purchased from. After unpacking, the
packing box and packing materials should be well kept for future repair. For
the damage of instrument occurring during transporting to the repair
department due to improper package, Thermo Fisher Scientific will not be held
responsible for damages.
The product is dry bath with advanced microprocessor control, which can be
widely applied to sample reservation, enzyme reservation and reaction, DNA
amplification, electrophoresis degeneration and serum coagulation, etc. The
characteristics of the product are as follows:
- The digital display and control of the temperature.
- Heat blocks help avoid possible contamination while containing the sample inside a tube.
- The heat blocks are easily replaced, cleaned or disinfected and are suitable for various tubes.
- Built-in over-temperature protection device warrants sample and user safety.
- Temperature deviation adjustment.
Features
This chapter introduces the usage, transportation, storage conditions of the
instrument, as well as its basic parameters, performance and functions.
Normal Working Conditions
- Ambient temperature 10°C ~ 30°C
- Relative humidity ≤ 80%
- Power supply 100-120 V ~ 50/60 Hz or
- 200-240 V ~ 50/60 Hz
- Altitude Up to 2000 m
- Pollution Degree 2
- Overvoltage category II Indoor use
- Mains supply voltage fluctuations ±10%
- CAUTION: Before using the instrument, make sure the working conditions meet the above requirements.
Transportation and Storage Conditions
- Ambient temperature: -20°C~+55°C
- Relative humidity: ≤ 80%.
Basic Parameters
Table 2. 100-120 V
Parameters / Type| Touch Screen 1 block 100-120 V| Touch Screen 2
block 100-120 V| Touch Screen 4 block 100-120 V
---|---|---|---
Temperature Range| Ambient Temperature +5°C~130°C (ambient temperature 25°C)
Temperature Uniformity| ±1.0°C
Temperature Accuracy| ±0.5°C@37°C
Temperature Fluctuation| ±0.5°C
Temperature Rise Time| ≤ 20 min (rise from 30°C to 130°C)| ≤ 25 min (rise from
30°C to 130°C)
Size (L × W × H)
| 373 × 200 × 100 mm
4.7 × 7.87 × 3.94 in
| 403 × 200 × 100 mm
15.9 × 7.87 × 3.94 in
| 535 × 200 × 100 mm
21.1 × 7.87 × 3.94 in
Weight| 3.4 kg (7.5 lbs)| 3.7 kg (8.16 lbs)| 5.1 kg (11.24 lbs)
Fuse Protector (ø5×20)| F 250 V 2.5 A| F 250 V 5 A| F 250 V 8 A
Table 3. 220 V
Parameters / Type| Touch Screen 1 block 200-240 V| Touch Screen 2
block 200-240 V| Touch Screen 4 block 200-240 V
---|---|---|---
Temperature Range| Ambient Temperature +5°C~130°C(ambient temperature 25°C)
Temperature Uniformity| ±1.0°C
Temperature Accuracy| ±0.5°C@37°C
Temperature Fluctuation| ±0.5°C
Temperature Rise Time| ≤ 20 min (rise from 30°C to 130°C)| ≤ 25 min (rise from
30°C to 130°C)
Size (L × W × H)
| 373 × 200 × 100 mm
14.7 × 7.87 × 3.94 in
| 403 × 200 × 100 mm
15.9 × 7.87 × 3.94 in
| 535 × 200 × 100 mm
21.1 × 7.87 × 3.94 in
Weight| 3.4 kg (7.5 lbs)| 3.7 kg (8.16 lbs)| 5.1 kg (11.24 lbs)
Fuse Protector (ø5×20)| F 250 V 2.5 A| F 250 V 2.5 A| F 250 V 3.15 A
Regulatory Compliance
European Union
The European voltage models of this product meet all the applicable requirements of the European Directives and therefore display the CE Marking. These Directives are captured in the EU Declaration of Conformity which may be obtained from the manufacturer.
Product Safety
This product family has been tested to applicable product safety standards by a Nationally Recognized Test Laboratory (NRTL) and may bear the NRTL’s mark of safety compliance to those applicable standards.
- CEEMC EN 61326-1
- CE Safety EN 61010-1 CE
- Safety EN 61010-2-010
- UL 61010-1/CSAC22.2 NO.61010-1
- ROHS2.0 2011/65/EU
Electromagnetic Compatibility
FCC Statement (USA)
- This device complies with Part 15 Subpart B of the FCC Rules. Operation is subject to the following two conditions:
- this device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
- Any changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canadian ISED IC Notice
- This ISM digital apparatus complies with Canadian ICES-001, Class A
- Cet appareil ISM est conforme á la norme NMB-001 du Canada, Classe A
End of Life Care
- Some considerations and suggestions are listed below for proper disposal of this product. While addressing these actions for safe recycling and disposal, please follow all guidelines, Safety Data Sheets (SDS), or regulations applicable to your country and region.
- This product has materials and components that may be recycled or reused according to local guidelines and regulations.
- Remove any batteries present before disposal. Batteries, battery packs, and accumulators should not be disposed of as unsorted household waste. Please use the public collection system to return, recycle, or treat them in compliance with the local regulations.
- Remove all samples and items before defrosting a unit to room ambient temperatures.
- Clean up any chemical or biological safety hazards using appropriate methods.
- Have a certified technician remove the Refractory Ceramic insulation from the unit then dispose properly.
Preparation Work
This chapter introduces the structure of the instrument, user interface, and functions of all buttons and preparation before startup. Read the content in the chapter carefully before startup when using the instrument for the first time.
Structure Interface
Note:
- When the external temperature sensor is connected, the internal temperature sensor will disconnect automatically, and the temperature control and display will relay on the external temperature sensor.
- DO NOT plug in or remove the external temperature sensor without the power being off.
Operation Guidance
The chapter introduces the display window and the operation of buttons of the
dry bath with constant temperature.
Examination Before Start-Up
Confirm the following before inserting the plug into the power socket.
- Ensure the power source is in accordance with the voltage requirement of the instrument (refer to “Features” section for voltage requirements).
- Make sure the plug has been completely inserted into the power socket.
- Ensure the power line has been reliably grounded.
CAUTION: If the display of the instrument is abnormal after startup, turn off the power source immediately and contact Thermo Fisher Scientific.
StartUp
Press the power switch and the instrument will be powered on. The LCD screen of the operation panel will show the main interface.
Operating Instructions
The Main Interface
Press “Single point set” to enter into the single setting interface.
Single Setting Interface
- Press “BACK” to go back to the main interface.
- Press “RUN” to enter into the single-running interface.
- Press “Temp” to set the temperature.
- Press “Time” to set the time.
- Instruction: There is a limitation for temperature and time setting. The setting temperature range is 15~135°C; the time range is 0~99: 59 and “∞ ”.
Single Running Interface
- Press “STOP”, and the interface of asking you whether to stop the program will appear.
- Press “YES” to stop the program and go back to the main interface.
- Press “NO” to go back to the single-running interface.
Calibration
- Press “CAL” to calibrate the temperature adjustment interface. (If the external sensor is being used, then there would be no “CAL.”, and the temperature is unable to be adjusted.)
- Input the correct temperature into the box that pops out.
- To ensure the accuracy of temperature, the instrument should be calibrated after temperature stabilizes for 30 minutes.
- Calibrate the instrument with the qualified Grade II standard filled thermometer.
- Calibration point: The middle hole of the heat block. Pour paraffin oil into the well, and soak the thermometer bulb.
- Instruction: During the operation of the instrument, the temperature and the time should not be changed. If it they should, stop the program and enter into the setting interface to change them.
Listed Files Interface
- Press “BACK” to go back to the main interface.
- Enter into the setting interface of the specific file in accordance with its name. If the file has a password, then enter into the password inputting interface.
Password Inputting Interface
- Input the password into the box.
- CAUTION: If the password has been forgotten, user can use 000000 to ROOT and then the file will be reset to the initial state (the name of the file would not be changed).
Program Setting Interface
- Press “BACK” to go back to the interface of the listed file.
- Press “SAVE” to enter into the files saving interface
- Press “RUN” to enter into the program’s running interface. If the alteration of the program has not been saved yet, a warning box will pop up.
- Press “YES”, and the program will run in the changed form, but the changes won’t be saved. Press “NO” to go back to the program setting interface.
- The user can set up 1-5 sections of the program.
- Press to delete the corresponding section (when there is only one section, this step would be of no use).
- Press to add sections. Choose the corresponding box to set the corresponding temperature and time
- CAUTION: This instrument doesn’t have a cooling function. If the user wants to set multiple sections, set the temperature from low to high. Otherwise, the temperature will be very slowly decreasing while the blocks cool down.
Files saving interface
- Press “BACK” to go back to the program setting interface.
- Press “SAVE” to save the file.
- Press “RUN” to enter into the program running interface. If the changed program has not been saved, a warning box will pop up.
- Choose “File name” and “Password” to set the file name and the password. By pressing Figures from 2-9 quickly, the input method will be changed to English to set the file name.
Program Running Interface
- Press “CAL.” to call for the temperature adjustment interface. ((If the external sensor is being used, then there would be no “CAL”, and the temperature is unable to be adjusted). Refer to ‘Single Running Interface’.
- Press “SKIP” to skip to the next section. If it is the last section, the interface of asking you whether to stop the program will appear. Refer to ‘Single Running Interface’.
- Press “PAUSE”, the program will be paused and the temperature will be maintained to the set temperature and the pause interface will pop up.
- Press “STOP”, the cancel interface will pop up. Refer to ‘Single Running Interface’.
- Instruction: During the operation of the instrument, the temperature and the time should not be changed. If needed, stop the program and enter into the setting interface to change them.
Maintenance
The holes of the blocks should be regularly cleaned with the damp cloth to
ensure the test tube be well contacted with the wall of the holes so as to
have good heat conduction. If the surface of the instrument is polluted, it
can be cleaned with a slightly damp soft cloth. Generally unit falls under
protective system of IP20 pursuant to IEC 60529.
WARNING: When cleaning the instrument, the power supply should be shut
off and unplugged. The instrument surface should be cleaned with a non-
corrosive cleaning agent.
Failure Analysis and Troubleshooting
This chapter details possible failures of this instrument and possible causes
with troubleshooting.
Table 4. Failure Analysis and Troubleshooting
No. | Error | Cause | Solution |
---|
1
| ****
Display window is not turned on after switching on power supply
| Power supply fails to be connected| Check power supply and connect it
Fuse burned| Replace fuse
Damage of the switch| Replace the switch
Others| Contact Thermo Fisher Scientific
2| The LCD shows “Err code: 0x1” and the buzzer alarms in a di…di…sound| Short
circuit of the sensor| Check whether the connecting wire of the sensor is
damaged and replace it
3| The LCD shows “Err code: 0x2” and the buzzer alarms in a di…di…sound| Open
circuit of sensor| Check whether the connecting wire of the sensor is damaged,
and replace it
4
| ****
The LCD shows “Err code: 0x3” and the buzzer alarms in a di…di…sound
| Sensor is damaged or the heat blocker is damaged| Contact supplier or
factory
Use the external sensor without putting it into the module| Put the external
sensor correctly into the corresponding place of the module
5| Temperature display does not conform to actual temperature| Sensor is
damaged or in poor contact| Contact supplier or factory
CAUTION: The user is not allowed to open the housing of the instrument for inspection during the warranty period. If any failures result from opening the housing for inspection, contact Thermo Fisher Scientific.
Modular Block Accessories
Table 5. Modular Block Accessories
Table 5. Modular Block Accessories
Warranty
The Warranty Period starts two weeks from the date your equipment is shipped
from our facility. This allows for shipping time so the warranty will go into
effect at approximately the same time your equipment is delivered. The
warranty protection extends to any subsequent owner during the first year
warranty period.
During the first two (2) years, component parts proven to be non-conforming in
materials or workmanship will be repaired or replaced at Thermo’s expense,
labor included. Installation and calibration are not covered by this warranty
agreement. The Technical Services Department must be contacted
for warranty determination and direction prior to performance of any repairs.
Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty
shall not extend the warranty to either the equipment or to the component part
beyond the original warranty period. The Technical Services Department must
give prior approval for return of any components or equipment.
At Thermo’s option, all non-conforming parts must be returned to Thermo Fisher
Scientific postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER
WRITTEN, ORAL OR IMPLIED. NO WARRANTIES.
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo
shall not be liable for any indirect or consequential damages including,
without limitation, damages relating to lost profits or loss of products. Your
local Thermo Sales Office is ready to help with comprehensive site preparation
information before your equipment arrives. Printed instruction manuals
carefully detail equipment installation, operation and preventive maintenance.
Warranty Scope
The above mentioned warranty does not apply to damages incurred by improper
use and maintenance, using under unqualified conditions, unauthorized
maintenance or modification.
Thermo Fisher Scientific does not provide any other warranty expressions
otherwise; and does not bear liabilities for business promotion under special
circumstances.
FAQ Section
- Scenario 1: Does unit turn off once the program time is complete:
- Example- If unit is set for one program to run at 45degC for 2 hrs.
- Q-Would the unit turn off upon completion of this 2 hrs?
- A- The instrument doesn’t turn off by itself, it just stops running.
- Q-If not- would the unit hold the set temperature beyond this 2 hrs?
- A- After this 2 hrs, the unit doesn’t hold the set temperature. It will cool to ambient temp.
- Q-Would unit return to ambient temp?
- A- The unit will return to ambient temp.
- Scenario 2: Suppose the unit is running a set program and there is power outage. Once the power resumes-
- Q-Would unit restart automatically once power is restored
- A- The instrument will automatically turn on.
- Q-Would unit resume the program from where it was left?
- A- It can’t.
- Q-Would customer need to start the program manually?
- A- Yes.
- Q-Would the unit keep the saved program in memory?
- A- The unit has no memory function.
- Q-Would customer need to create new program?
- A- Need to reset the program.
WEEE Compliance
WEEE Compliance. This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol. Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on our compliance with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at t hermofisher.com/WEEERoHS.
-
North America’s Technical Support Team
-
Phone: 1-866-984-3766, Option#2
-
Service Email: Servicesupport.led.asheville@thermofisher.com Presales Email: LED.Presales@thermofisher.com
-
Webpage:www.unitylabservices.com
-
Europe Middle East Africa Technical Support Team
-
German national toll free: 0800 1 112 110,
-
Fax: 0800 1 112 114 German international: +49 6103 408 0,
-
Fax: +49 6103 408 12 13 Email:unity.de@thermofisher.com and let.support@thermofisher.com
-
Singapore Technical Support Team
-
Service Hotmail is Service.LPG.ROA1@thermofisher.com
-
Thermo Fisher Scientific Inc.
-
22 Alpha Rd,
-
Chelmsford, MA
-
01824, USA
-
Find out more atthermofisher.com/drybaths
-
Australia: (613) 9757-4300 In Australia: (1300) 735-295 China: (86) 21-6865-4588 Germany: (49) 6184-90-6321 India: (91) 22-4157-8800 Japan: (81) 045-453-9175 North America: 1-978-232-6000 Toll-Free: 1-800-225-1480 Singapore: (65) 6778-6876
-
© 2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. 1021
Documents / Resources
|
ThermoFisher SCIENTIFIC 88870008 Touch Screen Dry Bath and Block
Heater
[pdf] User Manual
0000849, 88870008 Touch Screen Dry Bath and Block Heater, 88870008, Touch
Screen Dry Bath and Block Heater, Dry Bath and Block Heater, Bath and Block
Heater, Block Heater, Heater, Touch Screen Drybath
---|---
References
- Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Dry Baths | Thermo Fisher Scientific - US
- EPM Weee Compliance | Thermo Fisher Scientific - US
- Thermo Fisher Scientific - US
- Home
- EPM Weee Compliance | Thermo Fisher Scientific - US
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>